You are on page 1of 272

GENIUS HF 0434

65 – 80 kW

Service Manual

Version 13 June 2001 (Rev. 3)


SERVICE MANUAL
Revision history

Revision history

Rev. Date Page/s Modification description

3 13.06.01 All New SW release and new Epron on MPU (Master


Processing Unit) board.
New Display board.
New IGBT type (TRENCH) on inverter assy.
New manual layout.
(Ref. RDM 4824, RDM 5215)

4
5
6

(Rev. 3) GENIUS HF
SERVICE MANUAL
Revision history

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF (Rev. 3)
SERVICE MANUAL
Contents

Contents
1 FOREWORDS 1-1
1.1. Warning ...............................................................................................1-1
1.2. General precautions.............................................................................1-2
1.3. Risks involved with the use of this equipment......................................1-3
1.4. Icons appearing in the manual.............................................................1-4

2 GENERAL 2-1
2.1. Installation cards .................................................................................2-1
2.2. Electrical factors ................................................................................ 2-11
2.2.1. General characteristics ...................................................................2-11
2.2.2. Power supply ..................................................................................2-12
2.2.3. Condition of use .............................................................................2-12
2.2.4. Charge factors – Intermittent state ..................................................2-13
2.2.5. Charge factors – Continuos state.....................................................2-14
2.2.6. External safety system....................................................................2-14
2.2.7. Accuracy and precision of the radiography parameters ....................2-15
2.2.8. Charge factors measurement method ..............................................2-16
2.3. Compatibility ..................................................................................... 2-17
2.3.1. X-ray tubes ....................................................................................2-17
2.3.2. Measuring chambers (AEC).............................................................2-18
2.3.3. Examination tables.........................................................................2-18
2.3.4. Traditional TV systems or Digital Image TV systems ........................2-18
2.4. Environmental factors........................................................................ 2-19
2.4.1. Assembly........................................................................................2-19
2.4.2. Temperature requirements (in functions).........................................2-19
2.4.3. Heat generation ..............................................................................2-19
2.4.4. Environmental condition for Storage and Transportation .................2-20
2.5. Environmental risks........................................................................... 2.21
2.5.1. Electromagnetic compatibility .........................................................2-21
2.5.2. Flammability ..................................................................................2-21
2.5.3. Environmental risks and disposal ...................................................2-22
2.5.4. Information regarding generator oils ...............................................2-24
2.6. Symbols ............................................................................................. 2-29
2.6.1. Electric symbols .............................................................................2-29
2.6.2. General symbols .............................................................................2-30
2.6.3. Graphic symbols.............................................................................2-31
2.6.4. Commands console symbols ...........................................................2-32

(Rev. 3) i GENIUS HF
SERVICE MANUAL
Contents

2.7. Labels ................................................................................................2-35


2.7.1. Commands console ........................................................................2-35
2.7.2. Power rack .....................................................................................2-35
2.7.3. High voltage transformer – 1 tube and 3 tubes ................................2-36
2.7.4. Labels – General description ...........................................................2-37
2.8. IN-OUT signals...................................................................................2-39

3. COMPOSITION 3-1
3.1. Packaging.............................................................................................3-2
3.1.1. Command console ............................................................................3-2
3.1.2. High voltage transformer packaging ..................................................3-2
3.1.3. Power rack packaging.......................................................................3-3
3.2. Shipping and storage ...........................................................................3-4
3.3. Visual inspections ................................................................................3-4
3.4. Positioning of the components..............................................................3-5

4. INSTALLATION 4-1
4.1. Command console – Power rack connection .........................................4-2
4.2. Assembly of the separate components ..................................................4-3
4.2.1. Primary power supply connection cables...........................................4-4
4.2.2. Ground cable connection ..................................................................4-4
4.2.3. kV signal cable connection ...............................................................4-4
4.2.4. mA signal cable connection...............................................................4-5
4.2.5. Large and small focus power supply cables connection......................4-5
4.2.6. High voltage commutator cables connection
(only for 3-tube transformer model)...................................................4-6
4.3. Mains connection .................................................................................4-7
4.4. High voltage cables connection.............................................................4-8
4.5. Rotating anodes connection .................................................................4-9
4.6. Potter-Bucky connections ..................................................................4-10
4.7. Radiography contacts connections .....................................................4-11
4.8. Continued and pulsed fluoroscopy contacts connections ...................4-12
4.9. Other connections functions ..............................................................4-13
4.9.1. X-ray tubes selection contacts ........................................................4-13
4.9.2. Pulsed fluoroscopy exposure contact...............................................4-13
4.9.3. Active preparation contacts ............................................................4-13
4.9.4. Radiography exposure contact ........................................................4-14
4.9.5. Continuos fluoroscopy exposure contact .........................................4-14
4.9.6. Potter Bucky request contacts (grid release) ....................................4-15
4.9.7. Workstation contacts......................................................................4-15
4.9.8. KHUT alarm contact .......................................................................4-16

GENIUS HF ii (Rev. 3)
SERVICE MANUAL
Contents

4.10.Input signals connections .................................................................. 4-17


4.10.1. Door safety contacts .......................................................................4-17
4.10.2. kV-mA adjustment analogic input in automatic fluoroscopy ............4-18
4.10.3. kV-mA adjustment analogic input in manual fluoroscopy ................4-19
4.10.4. Continued/Pulsed fluoroscopy selection..........................................4-19
4.10.5. Decrease in step digital angiography ...............................................4-19
4.10.6. X-rays alarm without request..........................................................4-20
4.10.7. Tubes thermal protection................................................................4-20
4.10.8. Tomography program......................................................................4-20
4.10.9. External X-rays protection ..............................................................4-21
4.11.Services line connections ................................................................... 4-22
4.11.1. Monophase power supplies .............................................................4-22

5. CONFIGURATIONS 5-1
5.1. General ................................................................................................5-1
5.2. Workstations........................................................................................5-2
5.2.1. Number of workstations................................................................... 5-2
5.2.2. Examination table name .................................................................. 5-2
5.2.3. X-ray tube number .......................................................................... 5-2
5.2.4. Fluoroscopy..................................................................................... 5-2
5.2.5. Maximum fluoroscopy current value ................................................ 5-3
5.2.6. Fluoroscopy current value at maximum kV ...................................... 5-3
5.2.7. Total fluoroscopy time...................................................................... 5-3
5.2.8. Fluoroscopy on large focus............................................................... 5-3
5.2.9. Pulsed fluoroscopy........................................................................... 5-4
5.2.10. AEC device ...................................................................................... 5-4
5.2.11. Radiography buttons activation ....................................................... 5-5
5.2.12. Unrequested X-rays - Safety ............................................................ 5-5
5.3. Automatic exposure device (AEC) parameters ......................................5-6
5.3.1. Number of dominants ...................................................................... 5-6
5.3.2. Chamber number ............................................................................ 5-6
5.3.3. Base sensitivity of the chamber and correction curves ...................... 5-6
5.3.4. Minimum level of instant dose ......................................................... 5-6
5.4. Radiogenic tubes characteristics ..........................................................5-7
5.4.1. Maximum voltage of the radiogenic tube .......................................... 5-7
5.4.2. Anode rotation speed ....................................................................... 5-7
5.4.3. Anode cooling curve......................................................................... 5-8
5.4.4. Maximum anode load ...................................................................... 5-9
5.5. Configuration data input.................................................................... 5-10
5.5.1. Workstation data input...................................................................5-10
5.5.2. Radiogenic tubes data input ...........................................................5-14
5.5.3. Radiogenic tubes library .................................................................5-20
5.5.4. AEC chambers sensitivity calibration ..............................................5-22

(Rev. 3) iii GENIUS HF


SERVICE MANUAL
Contents

5.6. H.S. starter configuration...................................................................5-25


5.6.1. Mains voltage matching ..................................................................5-25
5.6.2. Braking activation at 1000 revs/minute..........................................5-26
5.6.3. Jumper location for tube selection ..................................................5-27
5.6.4. Jumper arrangement for high speed tube........................................5-28
5.6.4.1. Comet tubes ...................................................................5-28
5.6.4.2. IAE tubes........................................................................5-30
5.6.4.3. Philips tubes...................................................................5-31
5.6.4.4. Siemens tubes ................................................................5-32
5.6.4.5. Dunlee tubes ..................................................................5-33
5.6.4.6. Varian tubes ...................................................................5-34
5.6.4.7. CGR tubes ......................................................................5-35
5.6.4.8. Shimadzu tubes .............................................................. 5.36
5.6.4.9. Unlisted tubes ................................................................5-37
5.7. Tomography .......................................................................................5-38

6. CALIBRATION 6-1
6.1. General ................................................................................................6-1
6.2. High voltage kV coefficient value ..........................................................6-2
6.3. Filament heating characteristic values .................................................6-4
6.3.1. Stand-by current..............................................................................6-4
6.3.2. Minimum radiography current..........................................................6-4
6.3.3. Maximum radiography current .........................................................6-4
6.4. Fluoroscopy heating calibration ...........................................................6-5
6.5. Radiography small focus calibration.....................................................6-8
6.5.1. Automatic calibration .......................................................................6-9
6.5.2. Small focus manual calibration ......................................................6-12
6.6. Radiography large focus calibration ...................................................6-17
6.7. Calibration copy .................................................................................6-18
6.8. Generator regulation loop...................................................................6-20
6.8.1. kV regulation loop ..........................................................................6-21
6.8.2. mA regulation loop .........................................................................6-22

7. INTERFACE WITH DIGITAL SYSTEM 7-1


7.1. Characteristics.....................................................................................7-1
7.2. Signals .................................................................................................7-2
7.3. Signals timing ......................................................................................7-3
7.3.1. Pulsed fluoroscopy timing (high contrast fluoroscopy) .......................7-3
7.3.2. Continuous fluoroscopy timing .........................................................7-3
7.3.3. Digital radiography timing ................................................................7-4
7.4. Configuration program .........................................................................7-5
7.5. 0 points technique ...............................................................................7-6
7.5.1. Activation at a workstation ...............................................................7-6
7.5.2. Use of the 0 point technique .............................................................7-6

GENIUS HF iv (Rev. 3)
SERVICE MANUAL
Contents

8. GENERATOR STATUS INFORMATION


(DISPLAY ON MPU) 8-1

9. SYNOPTIC TABLES FOR EQUIPMENT SET-UP 9-1


9.1. Language change .................................................................................9-3
9.2. Configuration of the work stations .......................................................9-5
9.3. Input of the X-ray tube data............................................................... 9-11
9.3.1. Set-up of the tube stored in the memory .........................................9-13
9.3.2. Manual setting of the tube data ......................................................9-14
9.4. Radiography calibration ..................................................................... 9-21
9.5. Fluoroscopy calibration...................................................................... 9-27
9.6. Fluoroscopy ABC parameters configurations...................................... 9-31
9.7. AEC configuration.............................................................................. 9-35
9.8. Data and Time configuration.............................................................. 9-39
9.9. Historical errors display ..................................................................... 9-41
9.10.Calibration copy................................................................................. 9-43
9.11.Changing the kV transfer tables......................................................... 9-45

10. ALARM MESSAGES 10-1


10.1.Alarm messages in calibration ........................................................... 10-1
10.2.Alarm messages in anatomical technique........................................... 10-4
10.3.Alarms and equipment status of PCB Supervisor (PCB 99991030)..... 10-9
10.3.1. Status Led’s ...................................................................................10-9
10.3.2. Alarm Led’s ..................................................................................10-10

11. MAINTENANCE 11-1


11.1.Operator maintenance ....................................................................... 11-2
11.1.1. Cleaning.........................................................................................11-2
11.1.2. Daily tests ......................................................................................11-2
11.1.3. Weekly tests ...................................................................................11-2
11.1.4. Monthly tests .................................................................................11-3
11.2.Technical servicing maintenance........................................................ 11-4
11.2.1. Condition of the cables and connections .........................................11-6
11.2.2. Current leakage test .......................................................................11-6
11.2.3. Dust removal..................................................................................11-6
11.2.4. Cleaning the high voltage cable terminals........................................11-7
11.2.5. Cleaning the contacts .....................................................................11-7
11.2.6. Oil check ........................................................................................11-7
11.2.7. Alarms control................................................................................11-8
11.2.7.1. Check the inverter cards CCB and QUAD DRIVER ...........11-8
11.2.7.2. Inverter power supply check ............................................11-9
11.2.7.3. Inverter LED check..........................................................11-9
11.2.7.4. Power electronic leds check .............................................11-9

(Rev. 3) v GENIUS HF
SERVICE MANUAL
Contents

11.2.8. Fuse check................................................................................... 11-10


11.2.9. Radiography parameters check ..................................................... 11-10
11.2.9.1. Parameters accuracy tests............................................. 11-11
11.2.9.2. Parameters repeatability test ......................................... 11-14
11.3.EEprom management.......................................................................11-16
11.3.1. Data recording ............................................................................. 11-16
11.3.2. Loading data from the EEprom ..................................................... 11-17

12. WIRING DIAGRAMS 12-1


12.1.Wiring diagrams – Electronic PCBoard ...............................................12-1
12.2.Wiring diagrams – General terminal board .....................................12-107
12.3.Wiring diagrams – Power and high voltage circuits .........................12-127

13. SPARE PARTS LIST 13-1

This publication can only be reproduced, transmitted, transcribed or translated into


any human or computer language with the written consent of VILLA SISTEMI
MEDICALI S.p.a.
This manual is the English translation of the Italian original manual version.

GENIUS HF vi (Rev. 3)
SERVICE MANUAL
Forewords

1. FOREWORDS

NOTE:
/ The present manual is updated for the product it is sold with in order to
grant an adequate reference in performing diagnostics and repair
operations normally carried out by the service engineer.
The manual may not reflect changes to the product not impacting service
operations.

1.1. Warning

The safety of the patient and that of the operator has been given priority
consideration during the design and construction phase of this system.
This system complies with all the standards in accordance with which it
has been approved and will function correctly if it is used, maintained
and repaired in accordance with the instructions herein. Incorrect use
could jeopardize the patient, the operator and/or the equipment. VILLA
SISTEMI MEDICALI declines all liability for generator malfunctions due
to incorrect operations, installation, maintenance or repair.

(Rev. 3) 1-1 GENIUS HF


SERVICE MANUAL
Forewords

1.2. General precautions

This appliance is part of a system that emits ionizing radiation (X-rays).


This radiation may be hazardous both to the patient and to the operator
if the procedures for radiation protection as prescribed by the law are not
complied with. VILLA SISTEMI MEDICALI recommends the following
rules:
• only qualified persons who are appropriately trained in the use of
ionizing radiation may use this machine
• read and comply with the instructions provided in the Service and
User’s manuals
• read and comply with all the instructions on the labels on the
equipment
• follow the rules for radiation protection
• do not expose parts of the body to radiation unnecessarily
• comply with the local or national laws and regulations on ionizing
radiation
• if there is any doubt, consult a qualified specialist.
The above rules must not be considered as a substitute for the
procedures or rules already applied or for those prescribed by national or
international laws and policies.
Installation precautions:
• before installation, the radiation protection systems must be
implemented by specialized personnel, in accordance with the latest
laws on protection from ionizing radiation
• only personnel authorized by VILLA SISTEMI MEDICALI may perform
installation, maintenance or repair services on the equipment
• do not exceed the rating values of the radiogenic tube as provided in
tube's technical manual. The appliance's life span can only be
guaranteed if the generator is used in accordance with its technical
characteristics.
Mechanical parts precautions:
• each component of the appliance must be handled with care and
inspected periodically
• appropriate tools must be used during the assembly and disassembly
of the parts.
Electrical precautions:
• only appropriately trained and qualified persons may have access to
the internal parts of the generator
• exposed electrical terminals are a hazard. Ensure that the mains
power is disconnected before gaining access to the equipment's
internal parts
• do not remove the high voltage cable terminals from the tube or from
the high voltage transformer and do not access live parts before
having disconnected the mains switch
• follow all servicing requirements prescribed by this manual.

GENIUS HF 1-2 (Rev. 3)


SERVICE MANUAL
Forewords

1.3. Risks involved with the use of this equipment

LIABILITY:
The producer declines all liability for damages to things and/or persons
due to the equipment being used for reasons other than its intended use
or due to incorrect installation.

This equipment constitutes a high voltage generator that, when used


with a radiogenic tube and when used under the operational conditions
described in the User Manual, issues radiation within the frequency
spectrum typical of X-rays of the power and duration as set via the
command console.
The equipment must be used exclusively for diagnostic radiography and
fluoroscopy.
The user is responsible for the correct use of the generator. The
manufacturer shall supply the necessary information for its correct use
and regarding the risks involved in the use of the high voltage generator,
but declines all liability for the operations and safety and diagnostic
techniques used after the sale.
The manufacturer declines all liability for the over-exposure to radiation
of the operators and/or patients due to incorrect procedures or
operational techniques and declines all liability in the case of incorrect
installation and maintenance of the generator due to noncompliance with
the procedures provided in this Manual or for malfunctions due to
changes made by the installer.
This apparatus is sold with the acknowledgment that the producer, his
agent and representative decline all liability for the over-exposure to
X-rays by patients or personnel.
There exists no design technique that could safeguard operators or
servicing personnel who do not comply with the safety regulations.
Only authorized and suitably trained personnel is authorized to
operated these X-ray generators.
The technical service personnel must be aware of the dangers associated
with the maintenance of high voltage equipment and the danger
associated with high doses of X-rays.
Exposure to X-rays may cause injury to health with cumulative effects
that could surface after many months or even years.
Both the maintenance and hospital personnel who use this machine
should avoid exposure to the primary beam and be aware of the
dangers of the secondary beam and diffused radiation.

(Rev. 3) 1-3 GENIUS HF


SERVICE MANUAL
Forewords

1.4. Icons appearing in the manual

/
This icon indicates a NOTE: please read thoroughly the items marked
by this picture.

This icon indicates a WARNING: the items marked by this icon refer to
1 the safety aspects of the patient and/or of the operator.

GENIUS HF 1-4 (Rev. 3)


SERVICE MANUAL
General

2. GENERAL

2.1. Installation cards

The following cards refer to the equipment to which this Service Manual
relates to.
There are four cards.
The first refers to configurations set at the factory during the testing of
the equipment.
The other three refer to tubes that are calibrated in the factory and
delivered with the generator.
If the tube/s are not delivered together with the generator because they
were not available at the time of testing of the generator, or because
VILLA SISTEMI MEDICALI is not aware of the installation to which the
generator is to be connected, these four cards will not be filled in.

(Rev. 3) 2-1 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-2 (Rev. 3)


SERVICE MANUAL
General

GENIUS HF Model: ❑ 65 ❑ 80

Serial Number .................. Manufactured ....../....../......

Installation site ....................................................................

WORKING ACCESSORY FLUORO


TUBE AEC NOTE
STATION NAME N° kV min

1
2
3
4
5

FLUORO kV: - = disactive 1 = 100 kV


2 = 110 kV 3 = 120 kV

FLUORO min: + = stop every 5 minutes


1 = stop every 90 minutes, beeper every 5 minuets

AEC: + = active - = disactive

SERIAL NUMBER

INVERTER:

HT TRASFORMER

STARTER HS

(Rev. 3) 2-3 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-4 (Rev. 3)


SERVICE MANUAL
General

GENIUS HF Model: ❑ 65 ❑ 80

Serial Number .................. Manufactured ....../....../......

TUBE 1
Type ...................... Model .........................

Focus ................. Power .............. max kV .................

Cooling curve kHU


0 min 2 min 4 min 6 min 8 min NOTE

Power limit 10 x kW

TIME SMALL FOCUS LARGE FOCUS


sec. 3000 g/min 9000 g/min 3000 g/min 9000 g/min
0.01
0.1
0.2
0.5
1
2
3
5

Filament current mA
FOCUS STAND-BY MIN RADIO MAX RADIO NOTE

SMALL
LARGE

(Rev. 3) 2-5 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-6 (Rev. 3)


SERVICE MANUAL
General

GENIUS HF Model: ❑ 65 ❑ 80

Serial Number .................. Manufactured ....../....../......

TUBE 2
Type ...................... Model .........................

Focus ................. Power .............. max kV .................

Cooling curve kHU


0 min 2 min 4 min 6 min 8 min NOTE

Power limit 10 x kW

TIME SMALL FOCUS LARGE FOCUS


sec. 3000 g/min 9000 g/min 3000 g/min 9000 g/min
0.01
0.1
0.2
0.5
1
2
3
5

Filament current mA
FOCUS STAND-BY MIN RADIO MAX RADIO NOTE

SMALL
LARGE

(Rev. 3) 2-7 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-8 (Rev. 3)


SERVICE MANUAL
General

GENIUS HF Model: ❑ 65 ❑ 80

Serial Number .................. Manufactured ....../....../......

TUBE 3
Type ...................... Model .........................

Focus ................. Power .............. max kV .................

Cooling curve kHU


0 min 2 min 4 min 6 min 8 min NOTE

Power limit 10 x kW

TIME SMALL FOCUS LARGE FOCUS


sec. 3000 g/min 9000 g/min 3000 g/min 9000 g/min
0.01
0.1
0.2
0.5
1
2
3
5

Filament current mA
FOCUS STAND-BY MIN RADIO MAX RADIO NOTE

SMALL
LARGE

(Rev. 3) 2-9 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-10 (Rev. 3)


SERVICE MANUAL
General

2.2. Electrical factors

2.2.1. General characteristics

Characteristics 65 kW 80 kW

Manufacturer name VILLA SISTEMI MEDICALI same


20090 Buccinasco (MI)
Italia
Appliance name GENIUS 65 HF GENIUS 80 HF
Mains frequency 50 - 60 Hz same
Nominal mains voltage 400V 3~ 400V 3~
Absorbed peak current 141A @ 400V 173A @ 400V
Absorbed power 98 kVA apparent, 120 kVA apparent,
82 kW active 100 kW active
Nominal electric power 65 kW 80 kW
Insulation class CLASS 1 same
Applied parts TYPE B same
Conforming to standard I.E.C. 601-2-7 same
Original language of manual Italian same
Protection index IP20 same

(Rev. 3) 2-11 GENIUS HF


SERVICE MANUAL
General

2.2.2. Power supply

NOMINAL
MAINS
MAINS MAINS FREQUENCY MAINS VARIATION
IMPEDENCE
VOLTAGE
65 80 65 80 65 80
380V 3~ 50/60 Hz 50/60 Hz -5%/+10% -5%/+10% 0.20 Ω 0.20 Ω
400V 3~ 50/60 Hz 50/60 Hz -10%/+10% -10%/+10% 0.20 Ω 0.20 Ω
415V 3~ 50/60 Hz 50/60 Hz -15%/+6% -15%/+6% 0.20 Ω 0.20 Ω
220V 3~ ① 50/60 Hz 50/60 Hz -5%/+5% -5%/+5% 0.05 Ω 0.05 Ω

① this power is not considered standard and so a transformer is


required.

NOTE
/ For voltages below 400V nominal power may be reduced with respect to
the nominal value given in the table. This reduction is in ratio with the
mains voltage and impedance values.

2.2.3. Condition of use

The appliance has been designed to function with continuous charge or


intermittent charge. The charge power is in function of the radiogenic
tube power and the thermal charge accumulated during operations.

GENIUS HF 2-12 (Rev. 3)


SERVICE MANUAL
General

2.2.4. Charge factors – Intermittent state

Characteristics 65 kW 80 kW

Nominal high voltage 150kV (400mA) 150kV (500mA)


Max. current in tube for 800mA (81kV) 1000mA (80kV)
HV
Current and voltage at 800mA @ 81kV, 1000mA @ 80kV,
tube for nominal power 630mA @ 103kV 630mA @ 126kV
Nominal electric power 65kW, 80kW,
81kV @ 800mA @ 0.1sec, 100kV @ 800mA @ 0.1sec,
92kV @ 700mA @ 0.1sec 114kV @ 700mA @ 0.1sec
Exposure time range 0,001 ÷ 16sec same
Minimum current-time 0.4mAs, same
product 200mA @ 0.002sec
Maximum current-time 630mAs same
product
Current-time reference 40mAs, same
product 70kV @ 400mA @ 0.1sec
20mAs,
100kV @ 200mA @ 0.1sec
5mAs,
150kV @ 200mA @ 0.1sec
Charge factors range 40 ÷ 150kV (steps of 1kV) 40 ÷ 150kV (steps of 1kV)
10-13-16-20-25-32-40-50- 10-13-16-20-25-32-40-50-
63-80-100-125-160-200- 63-80-100-125-160-200-
250-320-400-500-630- 250-320-400-500-630-
800mA 800- 1000mA
Minimum nominal 2msec. same
radiation time

(Rev. 3) 2-13 GENIUS HF


SERVICE MANUAL
General

2.2.5. Charge factors – Continuos state

Characteristics 65 kW 80 kW

Nominal high voltage 120kV (5mA) same


Max. nominal current at 5mA (120kV) same
tube
Current and voltage at tube 120kV (5mA) same
for nominal power

NOTE:
/ The nominal performances of the equipment cannot be guaranteed if the
installation specifications are not complied with.
The above performance values are typical for a high voltage generator
and will be limited if the radiogenic tube has an inferior performance.

2.2.6. External safety system

The GENIUS HF must be protected by an external electric safety system


with the following characteristics:
• Type IEC328 Tripolar Braker with a minimum air space
between the 400V AC conductors of 3 mm
• Capacity 63 A high speed.

GENIUS HF 2-14 (Rev. 3)


SERVICE MANUAL
General

2.2.7. Accuracy and precision of the radiography parameters

PARAMETER
PARAMETER
PARAMETER VARIATION ACCURACY LIMITS
INDICATION
INTERVAL

kV 3-figure luminous From 40 to 150 kV 5% from 40 to


display located at left of peak 150 kV
the kV indication (the
unit is the peak kV)
Time 4-figure luminous From 0.001 to 16 s 1%
display located on left The accuracy also
of the s indication (the depends on the mA
unit is the second, s.) value and the length
of the cables
mA 4-figure luminous From 10 to ±10% on the entire
display located on left 1000 mA range
of the mA indication
(the unit is the
milliampere, mA)
mAs Top right part of the From 0.4 to ±5% on the entire
F.I.P. parameter 630 mAs range
represented by 3
figures plus a decimal
preceded by the
indication mAs (the
unit is the
milliampere/second,
mAs)

Definition of the radiography parameters:


- kV: Peak kV after all initial overshoots
- Time: Time during which the high voltage (between anode and cathode) is
greater than 75% of the kV
- mA: Average value of the current that circulates between cathode and anode
during exposure time
- mAs: Milliampere/seconds product (mA * Time).

(Rev. 3) 2-15 GENIUS HF


SERVICE MANUAL
General

2.2.8. Charge factors measurement method

This generator has been approved in accordance with the IEC 601-2-7
standard.
The test instruments used are:

• DYNALIZER III
This is an invasive high voltage measuring instrument in compliance
with the measuring methods required by the IEC 601-2-7 standard.
For further information contact:

RadCal Corporation
426 west duarte road
Monrovia - California 91016 USA
818-357-7921

• ASPN
This is a measuring system produced by AIR KERMA and with a
500 cc chamber.

All tests are performed using the instrument DYNALIZER III.

INSTRUMENTS INSIDE THE


TYPE OF TEST TEST INSTRUMENTS
GENERATOR

High voltage measurement DYNALIZER III HV divider on the secondary circuit.


The measurement is made by an
amplifier in instrument configuration
mA measurement DYNALIZER III The measurement of the mA values is
made by an amplifier in instrument
configuration
Time measurement DYNALIZER III Double clock inside the generator
Radiation measurement ASPN Telemat type AEC produced by
Comet AG (Berne, Switzerland)

GENIUS HF 2-16 (Rev. 3)


SERVICE MANUAL
General

2.3. Compatibility

2.3.1. X-ray tubes

This generator has been designed to operate with a wide number of


radiogenic tubes.
Following is presented a compatibility table.

Manufacturer Type of tube

VARIAN A102 - A119 - A132 - A186 - A192 - A152 -


A252 - A286 - A282 A292 - G292 - A482
COMET DI7 - DI9 - DI10H - DI10HS - DI81HS - DI91H -
DI93H - DI93HS - DI101H - DI101HS -
DI105HS - DI104HS
EUREKA RAD21 - RAD14 - RAD56 - RAD60
PHILIPS SRO 33/100
IAE X50H - RTM92H - RTM101H
CGR MN641 - MS740G - MSN742

The above-listed tubes are tested in the factory.


Thanks to its versatility and precision in the measurement and
adjustment of the fundamental parameters, it is possible to connect the
GENIUS HF generator to other tubes not included in the compatibility
list.
The characteristics required for a tube to be connected to the GENIUS HF
generator are:
• rotating anode
• bipolar
• maximum 2 focuses with maximum heating current not exceeding
9 amperes.

(Rev. 3) 2-17 GENIUS HF


SERVICE MANUAL
General

2.3.2. Measuring chambers (AEC)

It is possible to connect the generator to the following measuring devices:

Manufacturer Type of measuring chambers

PHILIPS Ionization and semiconductor


chamber
COMET Semiconductor chamber
ZHIEM Semiconductor chamber

2.3.3. Examination tables

The Genius HF generator has been designed for interface with the tables
currently produced by VILLA SISTEMI MEDICALI. Its versatility also
enables it to interface with VILLA SISTEMI MEDICALI tables no longer in
production.
If the generator needs to be interfaced with VILLA SISTEMI MEDICALI
tables which are no longer in production and with other accessories
found on the market, the manufacturer must first examine the possibility
of doing so and give his approval.

2.3.4. Traditional TV systems or Digital Image TV systems

The GENIUS HF generator has been studied and perfected to interface


with TV chains and/or digital acquisition systems currently produced by
VILLA SISTEMI MEDICALI. Its versatility also enables it to interface with
VILLA SISTEMI MEDICALI tables no longer in production.
If the generator needs to be interfaced with VILLA SISTEMI MEDICALI
tables which are no longer in production and with other accessories
found on the market, the manufacturer must first examine the possibility
of doing so and give his approval.

GENIUS HF 2-18 (Rev. 3)


SERVICE MANUAL
General

2.4. Environmental factors

2.4.1. Assembly

Both the 65 kW and 80 kW versions of the GENIUS HF generator


comprise:
• Command console 1 unit
• Power rack 1 unit
• High voltage generator 1 unit
• Set of cables connecting the console-rack (18 m) 1 unit

2.4.2. Temperature requirements (in functions)

Operating area Temperature (°C) Humidity Precautions

Consolle 5 ÷ 40 30 ÷ 75 % No condensation must


form in either
Rack 5 ÷ 40 30 ÷ 75 % operating zone.

NOTE:
/ If the above values are not complied with, it will not be possible to
guarantee the correct functioning of the equipment.

2.4.3. Heat generation

Operating area Apparatus in use Quantity of heat generated

Command Generator console Approx. 95 kcal/h


Operating Power rack and generator Approx. 930 kcal/h

NOTE:
/ If the above values are not complied with, it will not be possible to
guarantee the correct functioning of the equipment.

(Rev. 3) 2-19 GENIUS HF


SERVICE MANUAL
General

2.4.4. Environmental condition for Storage and Transportation

Equipment part Storage Transportation

Transformer Store in a very dry place not Do not turn upside down
subject to temperature during transportation. Do
variations to prevent not shake and handle with
formation of condensation. great care.
The temperature must not
fall below -10 degrees and
must not exceed 60 with a
humidity range between 30
and 85% not condensed.
Power rack Store in a very dry place not Do not shake and handle
subject to temperature with great care. The rack
variations to prevent contains highly fragile
formation of condensation. electronic material.
The temperature must not
fall below -10 degrees and
must not exceed 60 with a
humidity range between 30
and 85% not condensed.
Command console Store in a very dry place not Do not shake and handle
subject to temperature with great care. The console
variations to prevent contains highly fragile
formation of condensation. electronic material.
The temperature must not
fall below -10 degrees and
must not exceed 60 with a
humidity range between 30
and 85% not condensed.

GENIUS HF 2-20 (Rev. 3)


SERVICE MANUAL
General

2.5. Environmental risks

2.5.1. Electromagnetic compatibility

The apparatus is in conformance with the electromagnetic compatibility


standard EN 60601-1-2:1993.
There are no risks arising from the coupling of this electronic device with
other electronic devices, verification of which has been made in
accordance with provisions of the standard.
Although the appliance has been designed to give a reasonable degree of
protection from electromagnetic interference in compliance with the
European IEC standards, it must be installed at a certain distance from
electric energy transformer rooms, static continuity units, air
conditioning systems and other systems that could cause considerable
interference to the electrical lines.
Cellular phones and portable two-way radios may only be used at a
distance of over 3 meters away from any part of the equipment.
All instruments and equipment for professional use used near the
GENIUS HF must be in conformance with the requirements of the
Electromagnetic Compatibility standards. Those instruments not in
conformance with said standard and which are known to have little
immunity to electromagnetic fields must be installed at a distance of over
3 meters from the Genius HF and must be powered by a separate electric
line.
The Genius HF must be switched off when acusectors or similar
appliances are used.

2.5.2. Flammability

The flammable components in the generator may be the following:


• Oil. For further information on the flammable level of the oil
contained in the high voltage transformer see the paragraph 2.5.4.
“Information Regarding Generators Oils”, in this document.
• Cables. All the cables in the generator are of the type NO 7V-K IEC
20-22 II.
• Printed circuits. The basic laminate used in the production of the
printed circuits must correspond to the FR 4 V0 level of the
NEMA/L1 standard.
• Transformers. (They are, however, protected against short circuits).

The equipment is not suitable for use in the presence of anesthetic


mixtures that are inflammable in air , oxygen or nitrous oxide.

(Rev. 3) 2-21 GENIUS HF


SERVICE MANUAL
General

2.5.3. Environmental risks and disposal

The components used in this equipment do not present any hazard to the
surrounding environment.
This appliance is not suitable for use in the presence of anesthetic
mixtures that are flammable in air, oxygen or nitrous oxide.
Since the high voltage transformer contains oil (especially formulated by
SHELL for transformers) the standards for use and safety given in
paragraph 2.4.4. “Information Regarding Generator Oils” must be
complied with.
Do not dispose of oil in the environment; local laws regarding the
disposal of oils must be complied with.
The following regulations must be followed with regard to disposal:
• For the transformer:
– remove the oil from the high voltage transformer. Separate the
plastic and metal parts from those parts impregnated with oil.
The parts impregnated with oil comprise the high voltage
capacitor and the primary and secondary circuits. The primary
and secondary circuits may in turn be disassembled in order to
reach the plastic components such as spools, etc., and the
metallic components such as copper or iron present in the
windings and in the nucleus, and the capacitor must be removed
from its plastic container
– once the plastic and metallic parts have been separated from the
oil or from the parts impregnated with oil, the latter must be sent
to oil disposal centers or to companies specialized in the disposal
of oils, together with a list of the characteristics of the oil as per
the card in paragraph 2.4.4. “Information Regarding Generator
Oils”
– the methods used for the disposal of metallic and/or plastic parts
are the same as for ferrous and/or plastic parts. The high voltage
diode rods and printed circuit located on the lid of the high
voltage transformer must be sent to firms specialized in the
disposal of metal and plastic scrap.

• For the generator:


– separate the metallic parts from: printed circuits, plastic parts,
single components of large dimensions such as electrolyte
capacitors and/or 600 Vdc films, IGBT power semiconductors
and three-phase bridge mains and magnetothermal switches,
fuses and transformers
– dispose of ferrous materials such as sheet metal, copper, brass,
etc. in the same manner as ferrous scrap
– dispose of plastic and bakelite supports in the same manner as
plastic scrap
– power semiconductors, mains switches, large capacitors and
transformers must be sent to companies authorized in the
collection and recovery of ferrous-plastic materials, in accordance
with the laws in force
GENIUS HF 2-22 (Rev. 3)
SERVICE MANUAL
General

– printed circuits must be sent to companies authorized in the


collection and recovery of printed circuits, in accordance with the
laws in force.

The above procedures constitute a support for the disposal of the


generator produced by the manufacturer and must not be considered as
substitutes for existing national laws and regulations but only as a
complement to these. Priority must be given to the compliance of the
laws and regulations in force in the country and/or region where the
product is disposed of.

VILLA SISTEMI MEDICALI declines all responsibility for the disposal


of the product eliminated by the customer.

(Rev. 3) 2-23 GENIUS HF


SERVICE MANUAL
General

2.5.4. Information regarding generator oils

• Type: SHELL DIALA OIL DX


Shell Diala Oil DX is a type of dielettic oil produced by the
German Division of Shell.

• Properties:
Shell Diala Oil DX is a dielectric oil with a very high degree of
stability against oxidation, a good dielectric characteristic and
excellent resistance to low temperatures. It contains an anti-oxidant
(inhibitor) which guarantees its resistance efficiency to materials.

• Application:
Shell Diala Oil DX is used in transformers, oil-cooled rectifiers and
commutators with high oil temperatures. Shell Diala Oil DX is also
recommended for recharging equipment in which the existing
dielectric oil must be substituted.

• Standards:
Shell Diala Oil DX conforms to the requirements of the DIN 57370
Part 1 and VDE 0370 Part 1 Class A standards. These
specifications do not include the requirements relative to the
oxidation stability for dielectric oils. Shell Diala Oil DX complies with
the IEC 296, Class II A standard.

GENIUS HF 2-24 (Rev. 3)


SERVICE MANUAL
General

• Characteristics:

Purity DIN 57370 no solid substances, limpid


Density DIN 51757
at 15°C kg/m3 880
at 20°C kg/m3 877
Kinetic viscosity DIN 51562
at 20°C mm²/s 17
at -30°C mm²/s 750
Flash point (PM) DIN 51758 136
KOH/g mg DIN 57370 not determinable
neutralization no.
Corrosive sulphur DIE none
kV insulation voltage DIN 57370 approx 60
(after pre-treatment)
Dielectric dissipation DIN 57370 0,001
factor (after pre-
treatment)
Oxidation stability DIN 51554 28 d 110°C
KOH/g mg 0,20
saponification
Dielectric dissipation 0,050
factor at 90°C
Effusion point DIN lSO 3016

(Rev. 3) 2-25 GENIUS HF


SERVICE MANUAL
General

SHELL DIALA OIL DX SAFETY CARD - NOVEMBER 1989

Composition
Solvent refined naphthenic type mineral oil containing an antioxidant additive
Use
Insulating oil for electrical transformers and equipment
Chemical-Physical Characteristics
Physical state at 20°C liquid
Color straw-yellow
Density at 15°C 0.872 kg/l
Viscosity at 40°C 8.8 mm2/s
Boiling point I.B.P. > 280°C
Solubility in water not miscible
Flammability / Fire Prevention Measures
Flammable point 144°C (can open)
Extinguishing means carbon dioxide, chemical powder, foam
Reactivity
Chemical-physical stability stable at environmental temperature
Possible polymerization does not polymerize
Toxicity
Maximum concentration of oil mist 5 mg/m3
admissible in the atmosphere (TLV)
Acute oral toxicity LD 50 rat > ml / kg (estimate).
Acute percutaneous toxicity LD 50 rat > ml / kg (estimate)
Effects on the skin Frequent and prolonged contacts may cause
irritation to the skin and more serious effects
on more sensitive persons (allergies,
dermatitis, etc.)
Effects on the eyes Possible slight irritation
Effects if ingested If a significant quantity of the product is
swallowed anomalous digestive symptoms
could occur
Effects due to inhalation of the vapours Prolonged exposure to a high concentration
of TLV could cause irritation to the
respiratory tract

GENIUS HF 2-26 (Rev. 3)


SERVICE MANUAL
General

Emergency Measures
Contact with skin Wash with soap and water
Contact with the eyes Wash with plenty of water for ten minutes.
Consult a doctor
Ingestion Do not induce vomiting. To prevent
aspiration, seek medical assistance and show
this card.
Contact a poison-prevention center
Inhalation of the vapours Move person to the open air.
If the problem persists, seek medical
assistance
Aspiration If aspiration is suspected, take person to
hospital immediately
For further information to provide to the doctor assisting the victim of any one of the above
problems contact:
SHELL ITALIA S.p.A. –Direzione lubrificanti
Via Chiese, 74 – 20126 Milano
Tel. 02/38004461-2 (24 ore)
Preventive Measures
Do not mix with flammable substances or foodstuffs
Avoid the formation or diffusion of oil mist in the atmosphere
Avoid repeated and prolonged contact with the skin
Wash the skin carefully with soap and water. Do not use irritant or solvent products
Do not dry the skin with dirty or oily cloths
Keep the work clothing clean
Avoid contact of the oil with parts of the body that are injured or affected by infectious
diseases
Personal Protection Equipment
Water-proof gloves
Glasses
Water-proof overalls
Protection shields in case of splashes
Aspirators, in case of oil mists being formed
Applicable Laws
D.P.R. no. 547/1955: Regulations for the prevention of industrial accidents
D.P.R. no. 303/1956: General regulations for hygiene at the workplace
Risk 47: Reference is made to the attachment of this law "For workers assigned to
operations that require contact with ...omission... mineral oils on a regular basis ...omission a
half-yearly check-up must be made and immediate medical intervention must be assured
when the worker reports or shows signs of neoplasia”.

(Rev. 3) 2-27 GENIUS HF


SERVICE MANUAL
General

Storage and Losses


Store in closed containers, under cover and far from sources of heat and flames
In the case of leaks, absorb the product with earth, sand, sawdust or cloths that must be
then burnt. The contaminated area must be degreased with solvents to prevent the
formation and stagnation of vapours, with adequate protection provided to the relevant
personnel. Do not pollute drains, rivers or streams
To eliminate spent product, send to an authorized company for the collection and recovery
of spent oils, in accordance with the laws in force.

GENIUS HF 2-28 (Rev. 3)


SERVICE MANUAL
General

2.6. Symbols

2.6.1. Electric symbols

Three-phase

Closed (connected to mains)

Open (disconnected from mains)

Phase 1

Phase 2

Phase 3

Caution - High Voltage

Do not touch electrical parts when equipment is operating

Ground

(Rev. 3) 2-29 GENIUS HF


SERVICE MANUAL
General

2.6.2. General symbols

CE European mark

Type B appliance in accordance with the IEC 601 standards

Attention - Consult attached documentation

Radiography preparation

Radiography

Fluoroscopy

Emission of ionizing radiation

GENIUS HF 2-30 (Rev. 3)


SERVICE MANUAL
General

2.6.3. Graphic symbols

Start sequence

Press key

Key may be pressed

Acoustic signal

Reading dictionary

Press radiography button

Press fluoroscopy pedal

(Rev. 3) 2-31 GENIUS HF


SERVICE MANUAL
General

2.6.4. Commands console symbols

Switch on

Switch off

Alarm reset

Workstation no. 1

Workstation no. 2

Workstation no. 3

Workstation no. 4

Workstation no. 5

Child

Slight build

Normal build

Heavy build

GENIUS HF 2-32 (Rev. 3)


SERVICE MANUAL
General

Small focus

Large focus

Dominant chamber no. 1

Dominant chamber no. 2

Dominant chamber no. 3

AEC screen no. 1

AEC screen no. 2

AEC screen no. 3

Fluoroscopy reset time

High dose selection in fluoroscopy

Decrease associated parameter

Increase associated parameter

(Rev. 3) 2-33 GENIUS HF


SERVICE MANUAL
General

Memory blocks

Body part selection

Alarm reset

High speed forcing anode

Temperature anode

Transfer tables from fluoroscopy

GENIUS HF 2-34 (Rev. 3)


SERVICE MANUAL
General

2.7. Labels

2.7.1. Commands console

The label on the back of the command console identifies the model, serial
number and date of production of this piece.

2.7.2. Power rack

The label on the bottom left of the front of the rack identifies the model,
serial number and date of production of this piece.

(Rev. 3) 2-35 GENIUS HF


SERVICE MANUAL
General

2.7.3. High voltage transformer – 1 tube and 3 tubes

The label on the cover of the transformer near connection terminal no. 4
identifies the serial number and date of production of this piece.

GENIUS HF 2-36 (Rev. 3)


SERVICE MANUAL
General

2.7.4. Labels – General description

The following three figures show the meanings of the various parts of the
labels.

• Command console label

Month of production

Serial number,
month and year Year of production
of production
Power supply
data
Console model

Markings

• Rack label

The blackened Month of production


field indicates the
power of the
equipment
Year of production

Serial number, Power supply


month and year data
of production

Power rack Markings


model

(Rev. 3) 2-37 GENIUS HF


SERVICE MANUAL
General

• High voltage transformer label

Month of production

Year of production
Serial number,
month and year
of production Power supply
data

Transformer Markings
model

GENIUS HF 2-38 (Rev. 3)


SERVICE MANUAL
General

2.8. IN-OUT signals

The list here below shows the meanings of the IN and OUT signals given
in the text or on the electrical diagrams.

Signal name Type Description

IBO IN General inverter malfunction


ich IN High current in radiography
10 V = 1500 mA
ICK IN Recognition at MPU
icl IN Current in fluoroscopy 10 V = 10 mA
icm IN Average current in radiography
10 V = 150 mA
IDF IN Fluoroscopy kV decrease in digital mode
IDS IN Angio step program
iec+ IN Analogic 1 positive input
iec- IN Analogic 1 negative input
iep+ IN Analogic 2 positive input
iep- IN Analogic 2 negative input
IFS IN Start continuous fluoroscopy
IHS IN High speed anode rotation safety
IIF IN Fluoroscopy kV increase in digital mode
ilf IN Real current in large focus 10 V = 10 A
INS IN Normal speed anode rotation safety
IPA IN External radiography stop
IPB IN AEC activation
IPC IN Angio step kV decrease
IPD IN Stop radiography for ray alarm
IPE IN Manual or automatic fluoroscopy mode
IPF IN Start pulsed fluoroscopy
IPR IN Preparation request
IPW IN DC bar charged
IRM IN High voltage present at 75% of kV
IRP IN Potter reply
isf IN Real current in small focus 10 V = 10 A
ISR IN Start radiography

(Rev. 3) 2-39 GENIUS HF


SERVICE MANUAL
General

Signal name Type Description

IST-1 IN Tube 1 safety


IST-2 IN Tube 2 safety
IST-3 IN Tube 3 safety
ITA IN 7° tomography angle
ITB IN 20° tomography angle
ITC IN 30° tomography angle
ITD IN 45° tomography angle
ITF IN High speed tomography
ITS IN Slow speed tomography
IUB IN Inverter jumper not balanced
COM OUT Generator common ON-OFF
OAB OUT Equipment switched off at MPU
OAT OUT Active with anode rotating
OBF OUT Filament safety
OBS OUT Potter call
OFF OUT Generator OFF
OFR OUT Anode rotation in fluoroscopy
OFT OUT Active with continuous fluoroscopy on
OHS OUT Start anode rotation at high speed
OHT OUT High voltage start
okv OUT High voltage reference 10 V = 200 kV
olf OUT Large focus reference current 10 V = 10 A
ON OUT Generator ON
OPT OUT Active with pulsed fluoroscopy on
ORA OUT Data to converter card
ORB OUT High voltage deactivation
ORC OUT Clock to converter card
OREI OUT Alarm reset
ORP OUT Preparation call
ORR OUT Start anode rotation at normal speed
ORS OUT Radiography time
ORT OUT Active with radiography on

GENIUS HF 2-40 (Rev. 3)


SERVICE MANUAL
General

Signal name Type Description

OSD OUT Active with braking anode from HS to NS


osf OUT Small focus reference current 10 V = 10 A
OSM OUT Watchdog from MPU
OTC OUT Clock from MPU
OTD OUT Data from MPU
OTU-1 OUT Tube 1 select
OTU-2 OUT Tube 2 select
OTU-3 OUT Tube 3 select
OWS-1 OUT Workstation 1 select
OWS-2 OUT Workstation 2 select
OWS-3 OUT Workstation 3 select
OWS-4 OUT Workstation 4 select
OWS-5 OUT Workstation 5 select

(Rev. 3) 2-41 GENIUS HF


SERVICE MANUAL
General

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 2-42 (Rev. 3)


SERVICE MANUAL
Composition

3. COMPOSITION
The Genius HF 65-80 kW comprises two main parts with relative
connecting cables.

• Command console
The command console is a keyboard with or without a support
column with push buttons for the selection of the functions,
numerical luminous indicators (Displays) and an alphanumerical
luminous indicator (FIP). The command console is connected to a
command rack via a flexible cable of 10 mm diameter and a useful
length of 18 meters.

• Power rack
The power rack is a metallic rack of the following dimensions:

Base Height Weight


Power rack only
55 x 52 cm. 195 cm 161 kg.
(without HV transformer)
1 tube option 55 x 52 cm. 195 cm 195 kg.
3 tubes option 55 x 52 cm. 195 cm 210 kg.

The following parts are located inside the rack:


– control computer
– all the command and control circuits that receive information
from the microcomputer together with the command circuits for
the High Frequency inverter group
– the power inverter
– The high voltage transformer for the connection of a radiogenic
tube. Inside are located the high voltage transformer and two
filament transformers immersed in insulating oil
– The frequency converter for the high and/or normal speed
rotation of the anode in the radiogenic tubes. This converter is
optional. Included in the basic version of the generator is the
normal speed starter.

(Rev. 3) 3-1 GENIUS HF


SERVICE MANUAL
Composition

3.1. Packaging

The Genius 65 - 80 kW is packed in three plywood crates.

3.1.1. Command console

Dimensions:
• Base: 58 x 109 cm
• Height: 56 cm
• Net weight 34 kg
• Gross weight 57 kg

The following components are found herein:


• command console with two connecting cables for connection with the
power rack, both of 18 meters useful length
• a User’s Manual
• a Service Manual for the installer and maintenance technician

3.1.2. High voltage transformer packaging

Dimension:
• Base: 54 x 62 cm
• Height: 67 cm
• Net weight 34 kg (1 tube) 49 kg (3 tubes)
• Gross weight 54 kg (1 tube) 69 kg (3 tubes)

The high voltage transformer is packed in a wooden crate. Under no


circumstances must the crate be laid on its side because the insulating
oil could leak out of the High Voltage Transformer.

NOTE:
/ Even a partial loss of insulating oil could jeopardize the transformer's
function.
In the case of loss of insulating oil, contact the manufacturer, informing
him of the distance of the oil level from the lid. To obtain this
measurement, insert a metallic millimeter graduated rod into the oil refill
hole. The rod must be previously cleaned and degreased so that no
polluting substances are introduced into the insulating oil.

GENIUS HF 3-2 (Rev. 3)


SERVICE MANUAL
Composition

3.1.3. Power rack packaging

Dimensions:
• Base: 62 x 207 cm
• Height: 62 cm
• Net weight 161 kg
• Gross weight 201 kg

The crate contains the power rack packed in a horizontal position.

(Rev. 3) 3-3 GENIUS HF


SERVICE MANUAL
Composition

3.2. Shipping and storage

Great care should be taken during the shipment of this equipment.


Special attention must be given to the position of the high voltage
transformer during shipment, since it contains oil.
It must not be shaken and must be treated with great care as it contains
extremely fragile electronic materials.
A location with a relative non condensed humidity between 30 and 75%
must be used for storage in the case of especially long storage times
(more than two weeks).
The temperature must not fall below -10 degrees or rise higher than
60 degrees.

3.3. Visual inspections

A visual inspection of the packaging cases is necessary upon receipt of


the equipment.
If the packaging is damaged, the parts contained therein may also be
damaged.
Ensure that the case containing the high voltage transformer shows no
traces of oil (indicating that oil has spilled out of the transformer).
Complaints of damage must be made immediately to the shipping
company in writing, and a copy sent to VILLA SISTEMI MEDICALI.
Once the components has been unpacked, VILLA SISTEMI MEDICALI
must be informed of any missing materials or damages suffered during
shipment.

GENIUS HF 3-4 (Rev. 3)


SERVICE MANUAL
Composition

3.4. Positioning of the components

There are no special requirements with regard to the positioning of the


components.
The only criterion to be observed is the ease of use and easy access for
servicing interventions.
The power rack must be positioned in such a way that it can be moved
forward by at least 60 centimeters to enable the servicing personnel to
reach the components located at the back of the rack.
See drawing here below.

Figure 3-1

(Rev. 3) 3-5 GENIUS HF


SERVICE MANUAL
Composition

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 3-6 (Rev. 3)


SERVICE MANUAL
Installation

4. INSTALLATION
PRECAUTIONS DURING INSTALLATION
This apparatus contains an inverter system. Its oscillation frequency
(from few Hz to several kHz) may fall within the audibility range of the
human ear.
A high-pitched note will therefore be issued during the radiography
phase.

Before installing the power rack, the following regulations should be


followed:
• ensure that the equipment is installed in a room at a suitable
distance from power installations such as large conditioning systems,
liquid pumping systems or from large electromechanical installations
that could cause significant disturbance
• ensure that the TV chain cables do not run close to the generator
cables and that no function of the TV chain is supplied with power by
the suppliers in the generator. These may be disturbed by the noise
produced by the inverter
• the rack must be located approximately 60 cm. from the wall from all
sides, and all sides must be quickly and easily accessible to facilitate
the work of servicing personnel.

PRECAUTIONS WHEN OPERATING


• a number of large electrolytic capacitors, are located in the mains
interfacing section at the back of the rack (behind the General
Terminal Boards). The load stored during the functioning of the
equipment remains active for a brief period of time even after the
machine has been switched off.
If it is necessary to work in this vicinity, ensure that at least
10 seconds have elapsed after the machine has been switched off in
order to avoid contact with the loaded capacitors.
This ensures that the capacitors are no longer charged
• do not perform any fluoroscopy or X-rays when the high voltage
cables are not connected. This could damage the H.V. transformer
and the inverter
• Before checking or reconnecting the H.V. cables from the generator or
from the X-ray tube, ensure that the generator is switched off and
that the H.V. cables have been discharged by placing the cable
terminals on a metallic part connected to ground for at least
5 seconds.

(Rev. 3) 4-1 GENIUS HF


SERVICE MANUAL
Installation

4.1. Command console – Power rack connection

• Type:
Shielded cable with 25 conductors.
3 mm² unipolar ground cable.
• Function:
Transmission of signals between the microprocessor located on the
command console and the microcomputer located in the power rack.
All the commands selected by the operator via the keyboard are
transmitted in serial format.
• Connections:
This cable has a Cannon DB25 connector that must be connected to
connector J4 located on the MPU PCB 9999102700 (paragraph 12.1).

The following figure shows the connections via the cables supplied with
the appliance of the various modules that comprise the Genius HF.

Figure 4-1

GENIUS HF 4-2 (Rev. 3)


SERVICE MANUAL
Installation

4.2. Assembly of the separate components

Remove the high voltage transformer from its packaging. Remove any
traces of oil and check the oil level as indicated in paragraph 3.1.2.
The maximum distance between the level of the oil and the lid must not
exceed 25 mm.
Position the transformer inside the power rack in its place at the bottom
of the rack.
The slots for the high voltage terminals must be at the front of the rack.

Figure 4-2

The electrical connections can be made once the transformer is in


position.

(Rev. 3) 4-3 GENIUS HF


SERVICE MANUAL
Installation

4.2.1. Primary power supply connection cables

Four 10 mm² unipolar cables connect the high voltage transformer to the
power inverter.
These are marked 1, 2, 3, 4 and have M8 screw terminals.
They must be connected using the appropriate screws issuing from the
transformer lid (see Figure 4-2).
Their electrical function is shown in diagram PCB 5812 power circuit
(paragraph 12.3).

4.2.2. Ground cable connection

The 10 mm² unipolar cable connects the transformer to the general


ground.
This cable is yellow-green in color with an M6 screw terminal.
This cable must be connected to the ground terminal on the transformer
lid.

4.2.3. kV signal cable connection

This is a shielded, three-conductor cable, with a 4-pole green AMP


connector marked J2. (See PCB 5823 Internal connection map X-ray
generator, paragraph 12.1).
The cable must be connected to the corresponding connector located on
the appropriate printed circuit in the center of the transformer lid.
The other end marked J3 is already connected to the corresponding
connector located on the control card PCB 9999102500 Converter
Control Board (paragraph 12.1).
The shield of this cable (yellow/green cable with red terminal 4) must be
connected to one of the M4 screws that fixes the PCB 5802 HV tank
service GTB (paragraph 12.3).
The cable sends the real high voltage value (kV) to the PCB 9999102500
Converter Control Board (paragraph 12.1).

GENIUS HF 4-4 (Rev. 3)


SERVICE MANUAL
Installation

4.2.4. mA signal cable connection

This is a screened, three-conductor cable, with a green AMP connector


marked J4.
The mA signal cable must be connected to the corresponding connector
located on the printed circuit in the center of the transformer lid.
The other end of the cable terminates with a connector marked J2 which
is already connected to the connector located on the control card PCB
9999102500 Converter Control Board (paragraph 12.1).
The screen of this cable (yellow/green cable with red terminal 4) must be
connected to one of the M4 screws that fixes the PCB 5802 HV tank
service GTB (paragraph 12.3).
The cable sends the real high voltage current value (mA) to the PCB
9999102500 Converter Control Board (paragraph 12.1).

4.2.5. Large and small focus power supply cables connection

Two, three-conductor shield cables, with a green AMP connector marked


J1 (see PCB 5823 Internal connection map X-ray generator, paragraph
12.1).
Connector J1, with twelve poles, must be connected to the corresponding
connector located on the PCB 5802 HV tank service GTB (paragraph
12.3) in the center of the H.V. transformer lid.
At the other end of the two cables is a connector marked J2 which is
connected to the corresponding connector located on the PCB
9999102600 filament inverter driver (paragraph 12.1).
These conductors are used for the power supply to the filament
transformer of the small and large focus.
The shield (yellow/green cables with red terminal 4) must be connected
to one of the M4 screws that fixes the PCB 5802 HV tank service GTB
(paragraph 12.3).
See diagram PCB 5824 Filament Heating Circuit (paragraph 12.3) for
their electrical function.

WARNING:
1 After all the wires have been connected as described above, the red M6
vent screw located on the high voltage transformer lid must be loosened.

(Rev. 3) 4-5 GENIUS HF


SERVICE MANUAL
Installation

4.2.6. High voltage commutator cables connection (only for 3-tube


transformer model)

This component is supplied only upon specific request.


Follow the instructions for the oil level check and the positioning of the
transformer provided for the 1-tube transformer.
Follow the above instructions for the cables connection.
The only difference lies with the cables for the connection of the high
voltage commutator.

• High voltage commutator power supply cable


This is a 4-conductor cable with a 1 mm² cross section and green 10-
pole AMP connector marked J3.
This cable supplies power to the 3 magnets located in the high power
commutator.
The J3 connector must be connected to the corresponding
connection on the PCB 5802 HV tank service GTB (paragraph 12.3)
in the center of the H.V. transformer lid.

• High voltage commutator safety cable


This is a 4-conductor cable with a 1 mm² cross section.
Each conductor has 6mm faston terminals.
Their function is to detect the closure of the 3 magnets located in the
commutator.
The four conductors are marked as follows (see paragraph 12.2):
– C polarization common at 0 Vdc from PCB 9999102900 GTB 409
– 1 no.1 tube safety level at PCB 9999102900 GTB 410
– 2 no.2 tube safety level at PCB 9999102900 GTB 414
– 3 No. 3 tube safety level at PCB 9999102900 GTB 418.
The four faston terminals must be connected to the appropriate
microswitches as indicated in 5813 Main Service Transformer
(paragraph 12.3).

GENIUS HF 4-6 (Rev. 3)


SERVICE MANUAL
Installation

4.3. Mains connection

The generator has been set for connection to a 3~ 400 Volt power supply
at a frequency of 50 Hz.
With this supply the maximum mains resistance is 0.20 ohm.

• Connection to 3~ 400V
The mains supply must be connected to the mains filter via three M8
screw terminals located at the back of the power rack.
A cable cross section area between 10 mm² and 25 mm² for each
conductor must be used.
Connect the yellow/green ground conductor with the same cross
section as the mains connectors to the corresponding terminal.

• Connection to non-standards power mains


The generator may be connected to mains with voltages other than
400 V AC only if the optional three-phase auto-transformer has been
requested.
The auto-transformer may be connected after installation has been
completed and must be installed outside the power rack.
The possible power supply voltage values (all three-phase) are:
– 220V (with reduction of maximum power provided to the
radiogenic tube)
– 435V
– 465V
– 500V.

(Rev. 3) 4-7 GENIUS HF


SERVICE MANUAL
Installation

4.4. High voltage cables connection

WARNING:
1 Always disconnected the generator from the mains supply before taking
any action on the high voltage cable terminals.

The high voltage cable terminals must be of the three-conductor type


with IEC 526 terminals.
The number and polarity of the tube are marked on the high voltage
commutator lid at the terminal housing slots and on the Figure 4-2.
The greatest care is recommended when cleaning and lubricating the
cable heads and before their insertion into the transformer housing and
into the X-ray casing. Check the condition of the pins for the filament
connection in the negative terminal.
The recommended grease to use is SILICON COMPOUND 4.
Tighten the cable head to the housings on the transformer and the RX
casing via the appropriate ring nut.
A perfect connection of the filament power supply guarantees a constant
efficiency.
Connect a 1 mm² yellow/green cable between the shield of each H.V.
cable (high voltage transformer end) and the H.V. transformer ground
clamp.
This guarantees a perfect ground connection between the H.V. cables
and the transformer in case of high voltage discharge.

Figure 4-3

GENIUS HF 4-8 (Rev. 3)


SERVICE MANUAL
Installation

4.5. Rotating anodes connection

The Genius HF generator allows the supply to a maximum of three


rotating anode tubes.
The standard starter is supplied for normal speed and a starter for high
and low speed with automatic 3000/9000 rotations commutation may
also be connected.
With the normal speed starter the following function modes are possible:
• Fluoroscopy: start at medium voltage equal to that of running speed
• Radiography preparation: start for a duration of 1.2 seconds, then
maintenance of rotation for the entire duration of the preparation of
the radiography.
During this phase the rotation parameters are checked and
radiography is allowed only if all the parameters are correct.

The rotating anode connections for both the simple and the double speed
starters are made on the general terminal boards GTB 100 - 200 - 300
(see diagram 5813 Main Service Transformer, paragraph 12.3).

• Tube n.1:
– common (9) GTB 103
– inductive arm (7) GTB 109
– capacitive arm (8) GTB 110
PCB 9999103600 (paragraph 12.2)

• Tube n.2:
– common (9) GTB 203
– inductive arm (7) GTB 209
– capacitive arm (8) GTB 210
PCB 9999103700 (paragraph 12.2)

• Tube n.3:
– common (9) GTB 303
– inductive arm (7) GTB 309
– capacitive arm (8) GTB 310
PCB 9999102800 (paragraph 12.2).

(Rev. 3) 4-9 GENIUS HF


SERVICE MANUAL
Installation

4.6. Potter Bucky connections

The PCB 9999103600 and 9999103700 General Terminal Board 100-200


diagrams (paragraph 12.2) show how, by making changes to the terminal
board, it is possible to change the power supply voltage of an external
potter and obtain the feedback on the same power supply voltage (X-ray
consent) or with an insulated contact.
There are four possible methods:
• 3-wire potter supplied with 230 V AC (supplied by the generator)
• 4-wire potter supplied with 110 V AC (externally supplied)
• 4-wire potter supplied with 24 V AC (externally supplied)
• 0 Vdc (24 Vdc high logic level, supplied by the generator).

The 230 V AC power supply to the potter which is protected by fuse F2


(3.15 Amps) on the PCB 5855, (see diagram 5813 Main Service
Transformer, paragraph 12.3) is located on the GTB 102 and 201 with
common 101 and 201 (PCB 9999103600 and PCB 9999103700
paragraph 12.2).
Two insulated contacts are provided for each GTB to supply the external
potters (to be polarized by the relay request for the workstation selection).
Pure contacts are provided for the potter's power supply (to be selected
with the relays at the workstations) for the GTB 100 and GTB 200:
• GTB 100 first contact 112 – 113
• GTB 100 second contact 114 – 115
• GTB 200 first contact 212 – 213
• GTB 200 second contact 214 – 215.

GENIUS HF 4-10 (Rev. 3)


SERVICE MANUAL
Installation

4.7. Radiography contacts connections

A two-position command may be connected for radiography.

• Connection of an X-ray command for all workstations:


– common: connect to terminal GTB 421 0 V dc
– first release (preparation): connect to terminal GTB 422
– second release (radiography): connect to terminal GTB 425
PCB 9999102900 (paragraph 12.2)

• Connection of X-ray command for each workstation:


– common: supply 0 V dc via the potter and workstation contacts.
Power is present on the command whose common refers to the
selected workstation.
– first release (preparation): connect to terminal GTB 422
– second release (radiography): connect to terminal GTB 425
PCB 9999102900 (paragraph 12.2).

At this point the GTB 432 input (PCB 9999102900 paragraph 12.2) must
be polarized to 0 V dc to start exposure.
This external X-ray request input (potter) must be activated by the
configuration program as described in paragraph 9.2.

(Rev. 3) 4-11 GENIUS HF


SERVICE MANUAL
Installation

4.8. Continued and pulsed fluoroscopy contacts


connections

One or more continuous fluoroscopy commands may be connected in


parallel or selected from the workstation.
The connection points are GTB 428 0 V dc and GTB 429 optoinsulator of
the signal called fluoroscopy request (PCB 9999102900 paragraph 12.2).
In this case, fluoroscopy will be activated independently of the selected
workstation (if this function is activated in the configuration program). To
distinguish this function on the basis of the selected workstation, follow
the instructions given in the previous paragraph regarding the activation
of the workstation functions.
To request pulsed fluoroscopy, the connection points are GTB 433 0 V dc
and GTB 434 optoinsulator of the signal called pulsed fluoroscopy
request (PCB 9999102900 paragraph 12.2).

GENIUS HF 4-12 (Rev. 3)


SERVICE MANUAL
Installation

4.9. Other connections functions

These functions include the signals output from the generator and used
by the various external devices.
Normally, these comprise relay contacts insulated at 2500 V ac to
ground.
These contacts may be loaded with a maximum current of 3 Amp in
alternate current. For interfacing with electronic systems, optoinsulators
with a driving capacity of approximately 10 mA are included for the most
important signals.
It is also possible to select the logic level via jumpers. (See PCB
9999102800 General Terminal Board 300 paragraph 12.2).

4.9.1. X-ray tubes selection contacts

These contacts are normally open and close when the X-ray tube has
been selected.
The contacts are located on the GTB terminal boards at the following
numbers:
• 104-105 and 106-107 for tube 1 PCB 9999103600 paragraph 12.2
• 204-205 and 206-207 for tube 2 PCB 9999103700 paragraph 12.2
• 304-305 and 306-307 for tube 3 PCB 9999102800 paragraph 12.2

4.9.2. Pulsed fluoroscopy exposure contact

This signal leads to an optoinsulator whose polarity may be selected via


jumpers.
It is active when a pulsed fluoroscopy is in course. This signal is present
on the GTB 300 terminal board at point 330 (PCB 9999102800
paragraph 12.2).

4.9.3. Active preparation contacts

These are two contact couples that are usually open and used to inform
the external user that a radiography preparation is in course.
The two couples on the GTB 300 terminal board (PCB 9999102800
paragraph 12.2) are:
• 314-315
• 316-317.

(Rev. 3) 4-13 GENIUS HF


SERVICE MANUAL
Installation

4.9.4. Radiography exposure contact

These are two contact couples that are usually open and used to inform
the external user that a radiography exposure is in course.
The two terminal couples on the GTB 300 terminal board (PCB
9999102800 paragraph 12.2) are:
• 318-319
• 320-321.

This signal is also present on the GTB 328 terminal (PCB 9999102800
paragraph 12.2) and differs from the above contacts since the signal
derives from an optoinsulator whose activation and deactivation speed is
greater. The polarity may be selected via jumpers.

4.9.5. Continuous fluoroscopy exposure contact

These are two groups of two contact couples each one of which is usually
open; the other is normally closed and used to inform the external user
that a continuous fluoroscopy is in course.
The two terminal couples on the GTB 300 terminal board (PCB
9999102800 paragraph 12.2) are:
• First group
– 323-324 contact open
– 323-325 contact closed

• Second group:
– 326-325 contact open
– 326-327 contact closed.

This signal is also present on the GTB 329 terminal (PCB 9999102800
paragraph 12.2) and differs from the above contacts since the signal
derives from an optoinsulator whose activation and deactivation speed is
greater. The polarity may be selected via jumper contacts.

GENIUS HF 4-14 (Rev. 3)


SERVICE MANUAL
Installation

4.9.6. Potter Bucky request contacts (grid release)

The group of contacts that are normally open and that con supply power
to the Potter Bucky are two.
The first is located on GTB 100 board (PCB 9999103600 paragraph 12.2)
and the second on the GTB 200 board (PCB 9999103700 paragraph
12.2).
Closure of these contacts occurs at the end of the radiography
preparation time once all the procedures have been terminated correctly
and if the operator requests radiography by pressing the second position
on the X-ray push-button.

4.9.7. Workstation contacts

There are two groups of contacts for the selection of the five workstations
on the GTB 100 (PCB 9999103600 paragraph 12.2) and GTB 200
(PCB 9999103700 paragraph 12.2) terminal boards.
The two groups are available on the following terminals of the general
terminal board:

• First group

GTB 100 GTB 200


common 119 219
first workstation 120 220
second workstation 122 222
third workstation 124 224
fourth workstation 126 226
fifth workstation 128 228

• Second group

GTB 100 GTB 200


common 130 230
first workstation 121 221
second workstation 123 223
third workstation 125 225
fourth workstation 127 227
fifth workstation 129 229

(Rev. 3) 4-15 GENIUS HF


SERVICE MANUAL
Installation

4.9.8. kHUT alarm contact

During a sequence of radiography or fluoroscopy exposures, either


normal or pulsed, once 80% of the maximum kHUT of the anode is
reached, this contact is closed, indicating that the maximum limit is
being reached.
This contact, which is normally open, is present on the GTB 300 general
terminal board, at points 312-313 (PCB 9999102800 paragraph 12.2).
We recommend a sound or luminous signal be connected to this contact

NOTE:
/ The kHUT percentage is displayed on the leds on the command console.
When the safety limit has been reached the exposure sequence is
interrupted by the microprocessor.

This contact may therefore be used to connect an accessory (buzzer,


lamp, etc.) to provide further information to the user.

GENIUS HF 4-16 (Rev. 3)


SERVICE MANUAL
Installation

4.10. Input signals connections

These are signals input to the generator from external devices or from
push-buttons or pedals.
The majority of these signals are activated when polarized at 0Vdc.
The external contacts that perform these closures must be suitable for a
continuous voltage of 24Vdc and a current of 10 mA.

4.10.1. Door safety contacts

It is possible to connect a series of contacts relative to the doors of entry


to the room where the equipment is installed. These contacts are usually
closed and located on the GTB 400 general terminal board at input
points 435 0V dc and 436 (PCB 9999102900 paragraph 12.2).
When input 436 is not polarized at 0 V dc, the preparation, radiography
or fluoroscopy exposures are suspended and the message DOOR
SAFETY is displayed on the command console.
If there are no door contacts to connect, the above two terminals must be
short-circuited.

(Rev. 3) 4-17 GENIUS HF


SERVICE MANUAL
Installation

4.10.2. kV – mA adjustment analogic input in automatic fluoroscopy

The possibility of an analogic or digital command is dispose for this


function of increase or decrease of the kV-mA of automatic adjustment
fluoroscopy.
• Digital command (active at 0):
– common GTB 439 (0V dc power supply)
– increases kV-mA GTB 440
– decreases kV-mA GTB 441
PCB 9999102900 (paragraph 12.2)
• Analogic command:
The inputs for analogic adjustment in continuous and pulsed
fluoroscopy (analogic signal 0 / 1.4 V dc) are:
– GTB 401 0Vdc
– GTB 402 V ref
PCB 9999102900 (paragraph 12.2).
The system works as follows:
– if V ref = balance voltage (700 mV) stabilized system
– if V ref > balance voltage (700 mV) dose increased
– if V ref < balance voltage (700 mV) dose decreased.
The Vref stabilization value and the adjustment range may be
selected during installation in the configuration procedures of the
workstations as described in paragraph 9.2.
During the configuration procedure of the fluoroscopy dose level, the
parity value [C] must be input (the parity level is the voltage value for
which we consider the optimum dose level reached).
This value is typical to the TV camera or the TV system being used,
and the variation interval is 0/1.4 V dc.
It is therefore necessary to input a variation range of this signal [B],
i.e. the window around the parity value on which the signal changes.
The variation interval is +/-10 Volts. This couple of values is good for
both the pulsed and the continuous fluoroscopy. In addition, if the
adjustment is unstable, adjust the last digit of the increasing values
variation interval until it is stabilized, and do the same for the last
digit of the parity value.
The typical values for standard VILLA SISTEMI MEDICALI equipment
are:
– [B] = +0703
– [C]= 1003
This means that the parity value is set at 0.7 Volts and the band at
2 Volts. Values greater than 2 V are normalized to 2 V.
If the adjustment is unstable even if the Kp value is changed, a
number from 0 to 9 may be introduced into the last digit of [B] and
[C] using the following rule:
– increase the last [B] digit if the adjustment is unstable
– increase the last [C] digit if the signal is disturbed.

GENIUS HF 4-18 (Rev. 3)


SERVICE MANUAL
Installation

4.10.3. kV – mA adjustment analogic input in manual fluoroscopy

To activate this function instead of the automatic function the input on


the general terminal board GTB 400 at point 454 (PCB 9999102900
paragraph 12.2) must not be polarized at 0V dc.
The inputs for the analogic adjustment in manual fluoroscopy (analogic
signal 0/6 V dc) are:
– GTB 403 0Vdc
– GTB 404 V ref
PCB 9999102900 (paragraph 12.2).

If it is necessary to connect an external potentiometer for manual


adjustment, see the following diagram.

Figure 4-4

4.10.4. Continued/Pulsed fluoroscopy selection

To select between continued and pulsed fluoroscopy, the input on the


general terminal board GTB 400 (PCB 9999102900 paragraph 12.2) at
point 434 must be open (continued fluoroscopy) or polarized at 0 V dc at
point 433 (pulsed fluoroscopy).
The pulsed fluoroscopy function is possible with a digital acquisition
system.

4.10.5. Decrease in step digital angiography

The activation of this function is made by polarization point 452 of the


general terminal board GTB 400 (PCB 9999102900 paragraph 12.2) at
0 V dc.
When this function is active, at each radiography preparation release the
kV value changes in function of the values programmed at the beginning
of the program.
The program can perform a total of 5 reductions.
The values are displayed on the first line of the FIP.

(Rev. 3) 4-19 GENIUS HF


SERVICE MANUAL
Installation

4.10.6. X-rays alarm without request

If the input at pin 453 on the GTB 400 (PCB 9999102900 paragraph
12.2) is not polarized at 0 V dc, the fluoroscopy or radiography exposure
is halted.
This function is activated from the workstations via the configuration
program described at paragraph 9.2.

4.10.7. Tubes thermal protection

There are three couples of terminals (one for each tube) to connected to
the tube thermal protection.
The points are the following:
• Tube 1 GTB 411 0Vdc
• Tube 1 GTB 412 thermal input
• Tube 2 GTB 415 0Vdc
• Tube 2 GTB 416 thermal input
• Tube 3 GTB 419 0Vdc
• Tube 3 GTB 420 thermal input
PCB 9999102900 (paragraph 12.2).

4.10.8. Tomography program

To activate the tomography program whose timing is automatically set in


function of the speed and angle selected on the tomograph, the relative
inputs must be polarized to 0 V dc.
For each program, two inputs must be simultaneously active: 1 input is
relative to the speed and 1 input is relative to the angle.
The points on the GTB 400 terminal board (PCB 9999102900 paragraph
12.2) relative to the tomographic program are:
• GTB 443 Tomograph slow speed
• GTB 444 Tomograph high speed
• GTB 445 7 degree angle
• GTB 446 20 degree angle
• GTB 447 30 degree angle
• GTB 448 45 degree angle.

GENIUS HF 4-20 (Rev. 3)


SERVICE MANUAL
Installation

4.10.9. External X-rays protection

The input on the GTB 400 terminal board (PCB 9999102900 paragraph
12.2) at point 450 must always be polarized at 0 V dc (GTB 449) to
activated the X-rays emission. This protection must always be connected
to an external contact that inhibits all emission.

(Rev. 3) 4-21 GENIUS HF


SERVICE MANUAL
Installation

4.11. Services line connections

It is possible to connect a three-phase 400 V ac line via 6 mm² conducts


to the terminals of the LC contactor numbers 2, 4 and 6.
The maximum drawing current is 20 A for each phase with a maximum
three-phase load of 13.8 kVA.
This power supply is present when the equipment is ON.
Check the Power Circuit diagram 5812 (paragraph 12.3).

4.11.1. Monophase power supplies

Monophase voltages are present on the general GTB terminal boards that
may be used to supply external services.
These supplies may be used only for loads of less than 3 A (see the
protection fuse value on drawing 5812 paragraph 12.3).
It must be remembered that these supplies show a decrease in voltage
during the initial preparation phase when the normal speed starter
present inside the generator.
The voltages available are the following.
• 0Vac GTB 101 and GTB 201
• 230Vac GTB 102 and GTB 202
PCB 9999103600 and PCB 9999103700 (paragraph 12.2)
• 0Vdc GTB 449
• 24Vdc GTB 407 and GTB 408
PCB 9999102900 (paragraph 12.2).

GENIUS HF 4-22 (Rev. 3)


SERVICE MANUAL
Configurations

5. CONFIGURATIONS

5.1. General

The configuration phase allows the characteristics of the accessories


connected to the generator to be memorized and the operating selections
to be set. The data input refer to the X-ray tube, the workstation and the
automatic exposure device (AEC) etc.
These data may already be present in the memory if they are known
when the generator is tested at the manufacturer's, otherwise default
data are set.

(Rev. 3) 5-1 GENIUS HF


SERVICE MANUAL
Configurations

5.2. Workstations

5.2.1. Number of workstations

The workstations may be five in number and must be connected to three


X-ray tubes.
The number of workstations with which the configuration will be made
must be set.

5.2.2. Examination table name

The possible names to be set are given in the vocabulary list of the User
Manual.
Select the name of the workstation from the vocabulary list using the
appropriate keys, as described in paragraph 9.2.

5.2.3. X-ray tube number

The number of the X-ray tube (1, 2, 3) to which a workstation is to be


matched and the anode rotation mode (hs) must be set. The "hs"
selection ensures that the tube at that workstation always rotates at high
speed independently of the load selected.

5.2.4. Fluoroscopy

Set the fluoroscopy function to a workstation. This depends upon the


presence of a TV chain.
If fluoroscopy is permitted at a particular workstation, the maximum
value of the fluoroscopy kV must be given, and selected from the
following values: 100, 110, 120 kV.
This setting does not exclude the settings of the X-ray tube
characteristics.

GENIUS HF 5-2 (Rev. 3)


SERVICE MANUAL
Configurations

5.2.5. Maximum fluoroscopy current value

The maximum value of the fluoroscopy current may be selected from the
following values:
0.5 - 1.0 - 1.5 - 2.0 - 2.5 - 3.0 - 4.0 - 5.0 mA

5.2.6. Fluoroscopy current value at maximum kV

These values are the same as for paragraph 5.2.5.

5.2.7. Total fluoroscopy time

If fluoroscopy is set, the time value at the end of which fluoroscopy is


interrupted must be set. The maximum value possible is 5 mins.

5.2.8. Fluoroscopy on large focus

For special applications, fluoroscopy may be made using the large focus
instead of the small focus.
This option is activated with the addition of the JP8 jumper option.
The jumper is located on the PCB 9999104200 printed circuit of the PPU
Peripheral Processing Unit (paragraph 12.1) located inside the console.

Figure 5-1

(Rev. 3) 5-3 GENIUS HF


SERVICE MANUAL
Configurations

5.2.9. Pulsed fluoroscopy

Pulsed fluoroscopy is a fluoroscopy with a high current value (from 30 to


300 mA), very short exposure "pulses" (from 2 to 10 milliseconds) and a
rapid frequency (from 1 to 30 exposures per second).
The "pulses" are synchronized with the filming system and image
digitalization via the potter's feedback signal [IRP].
The voltage and current values are presented on the fluoroscopy display
and when this mode is active, substitute the continuous fluoroscopy
values.

• Switch-over from continuous to pulsed fluoroscopy:


– Activation of the input signal [ICS] (GTB 318 – PCB 9999102800
paragraph 12.2)

• Variation of the kV-mA value:


– the current (mA) value is fixed and corresponds to the value set
during the configuration phase
– the voltage (kV) value may be adjusted manually or automatically
via the analogic inputs as for continuous fluoroscopy
– the value of each pulse corresponds to that set during the
configuration phase.

5.2.10. AEC device

The fact that the workstation is equipped with an exposure chamber


must be memorized.
The automatic exposure device comprises two components: the pre-
amplifier, one for each chamber up to a maximum of three, and the
relative exposure chamber of the same number as the pre-amplifiers.
The exposure chamber is the TELEMAT made by COMET.
When the workstation is ready to use the exposure chamber the
dominant leds (1 or 3) light up.

GENIUS HF 5-4 (Rev. 3)


SERVICE MANUAL
Configurations

5.2.11. Radiography buttons activation

The "preparation" and "radiography" functions may be activated from the


console or from an accessory connected to the general terminal board
GTB 400 (PCB 9999102900 paragraph 12.2).
The following configurations are possible:
• C = "prep" and "radiography" activated via the console’s commands
• P = "prep" and "radiography" activated via the commands on the
GTB 400
• C-P = "prep" and "radiography" activated via both the console and
the GTB 400.
To activate the function from the console, press the "prep" button.
To activate the function from the terminal board, press the "radiography"
button.

5.2.12. Unrequested X-rays - Safety

This safety function is activated via the AEC blackening key (+).
The activation is evidenced by the lighting up of the 7 AEC blackening
leds.
This safety function is deactivated via the AEC blackening key (-).
With this function activated, the radiography and fluoroscopy exposures
are possible only if the IPD input signal is active (GTB 453 – PCB
9999102900 paragraph 12.2 - connected to 0 V dc).

(Rev. 3) 5-5 GENIUS HF


SERVICE MANUAL
Configurations

5.3. Automatic exposure device (AEC) parameters

When this function is activated as described in paragraph 5.2.10, the


following data must be memorized.
This must be repeated for each one of the devices installed.

5.3.1. Number of dominants

The number of dominants in a chamber may be one or three.

5.3.2. Chamber number

The chamber number is in function of the connector to which it is


connected.
The chamber number must be associated to the workstation.

5.3.3. Base sensitivity of the chamber and correction curves

• Sensitivity: this value goes from 1 to 9999.


This value determines the base blackening value according to the
screen/film combination. Three different combinations to which to
associate different blackening values are possible.

• kV Correction Curve: this adjustment compensates the blackening


according to the set kV value. This function comprises 2 curves with
different behaviours in function of the kV value (< = 80 kV and > 80
kV).

5.3.4. Minimum level of instant dose

This is the minimum level, expressed in millivolts, of the instant dose


under which radiography is blocked.
This safety measure makes it possible to interrupt an X-ray that would
reach the maximum time limit due to incorrect radiological parameters
that do not generate a dose that could integrate the measuring chamber.

GENIUS HF 5-6 (Rev. 3)


SERVICE MANUAL
Configurations

5.4. Radiogenic tubes characteristics

Radiogenic tubes with a rotating anode and two high voltage polarities
may be connected to the Genius HF generator.
The starter supplied with the generator allows the anode to rotate at
3000 revs.
A high speed starter (9000 revs.) may be installed upon request.
The use of the high or low rotation speed depends upon the maximum
load or on the selections made by the operator.

5.4.1. Maximum voltage of the radiogenic tube

The maximum kV value allowed by the radiogenic tube must be input


into the computer and this data will be used to protect the tube.

5.4.2. Anode rotation speed

The following three configurations are possible with the high speed
starter:
1. Only normal speed (3000 revs.).
2. Only high speed (9000 revs.).
3. Both speeds (3000 and 9000 revs.) according to the load or selection.

NOTE:
/ If an incorrect value is selected (e.g. high speed with normal starter
installed), the data is accepted but the function will not be activated. If
the "hs" value has been input (paragraph 5.2.3), the anode will turn at
high speed at that workstation, irregardless of the number input in this
field.

(Rev. 3) 5-7 GENIUS HF


SERVICE MANUAL
Configurations

5.4.3. Anode cooling curve

The Anode's cooling data expressed in kilo-thermal-units (kHU) at the


following times must be input as follows:
• 0 minutes maximum thermal capacity storage
• 2 minutes thermal residue after 2 minutes
• 4 minutes thermal residue after 4 minutes
• 6 minutes thermal residue after 6 minutes
• 8 minutes thermal residue after 8 minutes.

In some cases the cooling curves are expressed in kilo-joules (kJ also
called kilo Watt secs. kWs).
The conversion value to transform this unit in kHU is as follows:

kHU = 1,33 x kJ

GENIUS HF 5-8 (Rev. 3)


SERVICE MANUAL
Configurations

5.4.4. Maximum anode load

This table makes it possible to check the maximum load.


The maximum power values allowed for each radiogenic tube and each
focal spot on the tube must be input.
When the double speed (3000 and 9000 revs.) anode rotation system has
been selected, the maximum load values of the tube for both rotation
speeds must be input.
The maximum loads for the following combinations must then be input:
• Small focus normal speed SF NS
• Small focus high speed SF HS
• Large focus normal speed SF NS
• Large focus high speed SF HS.
When the rotation system for only one speed is required, only the power
values referring to that speed and for both focal spots shall be input.
The power values for each combination required must be input for the
following radiography time values:
• 0,010 seconds
• 0,100 seconds (nominal power)
• 0,200 seconds
• 0,500 seconds
• 1 seconds
• 2 seconds
• 3 seconds
• 5 seconds
The maximum load curves are normally given as a series of curves with
different kV values that vary in function of time (abscises) and of the
current value (mA) (ordinates).
The calculation of power in time must therefore be performed. To obtain
this value, the value of the maximum current value allowed with a
voltage of 100 kV must be detected from the maximum load curve
relative to the focus and rotation speed, for each of the above eight
radiography times.
Then the kV value (100) is multiplied by the value of the current
displayed expressed in A read on the small curve, to obtain the kW value
for that particular value of radiography time.

(Rev. 3) 5-9 GENIUS HF


SERVICE MANUAL
Configurations

5.5. Configuration data input

An additional computer is not required for the input of data; the console
is used as access terminal to the principal computer. The keys assume a
specific function for this function.

5.5.1. Workstation data input

1. Switch on the equipment via the ON key. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack and move the switch located
on the front part of the PCB 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press one of the vertical anatomical selection keys.
4. Press the Child key to enter the configuration.
5. The following is displayed on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

• Changing the language


It is possible to change the user-generator interface language.
Using the key located below the writing that indicates the current
language set, scroll through the eight possible languages
available.
6. By pressing the key ORG 1 located in correspondence of the lower-
case letter of the word CONFiGURATION, the leds associated with the
workstations keys will start to flash.
By pressing the desired workstation key, the FIP will be modified as
follows:

CP 058 REMOTE TA T1 kp Cont FL = 03


P Dose : DIG kV Gra 00 kV Sco C Dose : ANA

GENIUS HF 5-10 (Rev. 3)


SERVICE MANUAL
Configurations

• Changing the workstation


Using key ORG 2 it is possible to scroll forward through the
vocabulary.
Using key ORG 1 it is possible to scroll backward through the
vocabulary.
• Selecting the tube
The selection is made using key ORG 3.
The tubes and the selections of the rotation speeds (T1 - T1hs -
T2 - T2hs - T3 - T3hs) are displayed cyclically.
• Selecting the “prep” and “radiography” mode
C indicates that the key located on the console is activated at the
workstation
P indicates that the GTB 400 terminal board input is activated
- indicates the deactivation of the "prep", "radiography" function.
• Proportional coefficient for automatic fluoroscopy
adjustment

kp Cont FL = 20

The operating mode (pulsed or continued) is set via key ORG 4.


The Kp value (adjustment proportional coefficient) between 0 and
99 can be scrolled cyclically via key ORG 5.
• Selecting the control mode for automatic adjustment of the
fluoroscopy dose
The continued or pulse fluoroscopy mode control may be changed
via the M and +mA keys respectively.
DIG indicates digital mode. ANA indicates analogic mode.

P Dose : DIG C Dose : ANA

(Rev. 3) 5-11 GENIUS HF


SERVICE MANUAL
Configurations

• Technical activation at 0 points digital


This option is activated via the +sec. key. The 00 digit indicates
that the function is deactivated. All digits other than 00 indicate
that the function is activated.

kV Gra 00

• Modifying the pulsed fluoroscopy parameters


Press the lateral anatomic selection key to display the pulsed
fluoroscopy parameters on the three numerical fluoroscopy
displays and enable these values to be changed.
The mA display shows the value at which the fluoroscopy will be
performed.
It is possible to vary this value from a minimum of 30 to a
maximum of 300 mA using the DS1 key.
The time display will show the time of each "pulse".
It is possible to change this value from a minimum of 2 msecs to
a maximum of 10 msecs using key DS2.
• Anode rotation braking delay
Upon release of the fluoroscopy or the radiography preparation, it
is possible to maintain rotation for a selected period of time from
0 to 99 seconds. Using the keys Reset min fluoroscopy and >5R
the value on the fluoroscopy kV display expressed in seconds can
be changed.
• Introduction of an exposure meter to one or more fields
The selection of a field is shown by the lighting up of the led on
the relative key. By pressing one of the dominants keys the led is
lit or switched off.
If the chamber is not present at the workstation being changed,
the three leds relative to the dominants must be off.

GENIUS HF 5-12 (Rev. 3)


SERVICE MANUAL
Configurations

• Changing the continuous fluoroscopy parameters


Press the front anatomic selection key to display the continuous
fluoroscopy parameters on the three numerical fluoroscopy
displays and enable these values to be changed.
The kV display presents the maximum voltage value at which
continuous fluoroscopy can occur.
It is possible to change this value between 100, 110 or 120 using
the fluoroscopy Reset min. key.
The mA display shows the maximum current value reached by
continuous fluoroscopy.
This value can be changed via key >5R. The possible values to be
selected are: 0.5, 1, 1.5, 2, 2.5, 3, 4, 5.
7. To complete the set-up program press key M1, the FIP shows the
following:

CONFIRM ESCAPE

When the key ORG 2 is pressed, the above settings are memorized on
EEprom.
When the key ORG 4 is pressed, the above settings are cancelled and
the preceding settings reset.
In both cases the program returns to the phase in which the
workstation leds flash.
Press the key of the workstation to be changed, otherwise press key
Sturdy Build to exit this subroutine.

(Rev. 3) 5-13 GENIUS HF


SERVICE MANUAL
Configurations

5.5.2. Radiogenic tubes data input

1. Switch on the equipment via the ON key. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack and move the switch located
on the front part of the PCB 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press one of the vertical anatomical selection keys.
4. Press the Child key to enter the configuration.
5. The following is displayed on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

When key ORG 1 is pressed, the leds associated to the workstations


flash.
Press the key for the desired workstation and confirm the data via
key M1.
The FIP displays the following:

Tube

6. Press key +sec. located under the word TUBE, the FIP is changed as
follows:

FORwARD BAcK CONfIRM EScAPE MANUaL

GENIUS HF 5-14 (Rev. 3)


SERVICE MANUAL
Configurations

7. Press key ORG 5 above the word manual. The FIP changes as follows:

COOLING curve (kHU) 0 2 4 6 8 min


0040 0240 0200 2175 0150

In addition, the following information is displayed on the fluoroscopy


display:
– kV display = maximum kV value that may be set
in radiography
– mA display = maximum mA values of small and
large focus
– fluoroscopy min display = anode rotation speed.

• Changing kVmax, mAmax and anode rotation speed


Using the keys >5R and Reset min. Fluoroscopy the kVmax
radiography value can be increased or decreased.
With keys DS1 and DS2 the maximum mA value for the focus
with the lighted led is changed (small focus from 100 and 300
mA, for large focus from 500 to 1000 mA). The rotation speed
may be changed using key DS3 (1 only slow speed, 2 slow and
high speed, 3 only high speed).
• Changing the cooling curve values
With this FIP configuration, the active keys are the +mA and +kV
as shown in the figure.
These keys enable the user to highlight a required field by
scrolling from left to right and vice versa.
After having selected the desired value (flashing value), the
datum can be changed using the AEC blackening keys (increase
or decrease).
The amount of increase or decrease for each action on the above
keys depends upon which of the below-listed leds is active:
– left AEC chamber led active = 1 unit
– central AEC chamber led active = 10 units
– right AEC chamber led active = 100 units.

(Rev. 3) 5-15 GENIUS HF


SERVICE MANUAL
Configurations

• Changing the small focus low speed maximum load values


(SF NS)
Press key ORG 1 to access the maximum load setting routine.
The FIP changes as follows:

CAR MAX kV*10 FP NS .01 .1 .2 .5 1 2 3 5


0230 0220 0210 0200 0180 0170 0150 0130

With this configuration, key ORG 1 is used to select the


maximum load parameters and the cooling curve according to the
following combinations:
– SF NS = Small focus - Slow speed
– LF NS = Large focus - Slow speed
– SF HS = Small focus - High speed
– LF HS = Large focus - High speed.
Select SF NS.
To determine the correct value to introduce in place of the one
indicated, see paragraph 5.4.4.
In this FIP configuration the activated keys are +mA and +kV as
indicated in the figure.
A field can be highlighted by scrolling from left to right and vice
versa with these keys.
After having selected the desired value (flashing), the data can be
changed using the AEC blackening keys (increase or decrease).
The amount of increase of decrease for each of the above actions
via the keys depends upon which of the below listed leds are
activated:
– left AEC chamber led active = 1 unit
– central AEC chamber led active = 10 units
– right AEC chamber led active = 100 units.

GENIUS HF 5-16 (Rev. 3)


SERVICE MANUAL
Configurations

• Changing the large focus low speed maximum load values


Press key ORG 1 to access the maximum load setting routine.
The FIP changes as follows (see following figure):

CAR MAX kV*10 SF NS .01 .1 .2 .5 1 2 3 5


0230 0220 0210 0200 0180 0170 0150 0130

To set the parameters for this mode follow the instructions as for
the SF NS mode.
• Changing the small focus high speed maximum load values
When high speed has been set, press ORG 1 to display the SF HS
settings.
The FIP displays the following:

CAR MAX kV*10 SF HS .01 .1 .2 .5 1 2 3 5


0230 0220 0210 0200 0180 0170 0150 0130

To set the parameters for this mode follow the instructions as for
the SF NS mode.

(Rev. 3) 5-17 GENIUS HF


SERVICE MANUAL
Configurations

• Changing the large focus high speed maximum load values


Press key ORG 1 and select the LF AN configuration on the FIP
(see following figure).

CAR MAX kV*10 LF HS .01 .1 .2 .5 1 2 3 5


0230 0220 0210 0200 0180 0170 0150 0130

To set the parameters for this mode follow the instructions as for
the SF NS mode.
8. To quit this settings routine press key M1.
The FIP changes as follows:

CONFIRM ESCAPE

9. Press key ORG 2 to memorize the data set in EEprom.


Press key ORG 4 to cancel the data.
10. Press Sturdy Build to exit the routine.

GENIUS HF 5-18 (Rev. 3)


SERVICE MANUAL
Configurations

EXAMPLE OF CALCULATION TO DETERMINE THE MAXIMUM LOADS:

Tube COMET type DX104


Small Focus 0.6 mm
Rotation speed 9000 revs./min
Radiography time 0.01 sec
We see from the load curve that at 100 kV the radiography mA limit value
is 395.

So proceeding with the following calculation:


kW = 100 kV x 1 radiography (A) = 100 x 0.395 = 39.5
The 39.5 kW value will be input into the computer with four significant
digits, where the most significant are the hundreds of kW and the least
significant are the kW decimals. So our value of 39.5 kW will be changed
to 0395.

The power values for the focal spot are the following:

n° memorized by
Times secs mA max kW max
Genius HF
0.010 395 39.5 0395
0.100 370 37.0 0370
0.200 347 34.7 0347
0.500 315 31.5 0315
1.000 270 27.0 0270
2.000 225 22.5 0225
3.000 190 10.0 0190
5.000 165 16.5 0160

(Rev. 3) 5-19 GENIUS HF


SERVICE MANUAL
Configurations

5.5.3. Radiogenic tubes library

The data relative to the most common radiogenic tubes are memorized in
the PCB Peripheral Processing Unit's EPROM (paragraph 12.1).
It is therefore possible to use this data and associate them to the
installed tube.
The data memorized are:
• The anode's cooling curve
• The low speed load curve
• The high speed load curve (if the tube is set accordingly)
• The small focus and large focus calibration tables.
The calibration tables refer to the standard calibration which is then
adapted to the relevant tube characteristics.

1. Switch on the equipment via the ON key. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack and move the switch located
on the front part of the PCB 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press one of the vertical anatomical selection keys.
4. Press the Child key to enter the configuration.
5. The following is displayed on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

When key ORG 1 is pressed, the leds associated to the workstations


flash.
Press the key for the desired workstation and confirm the data via
key M1.
The FIP displays the following:

Tube

GENIUS HF 5-20 (Rev. 3)


SERVICE MANUAL
Configurations

6. Press key +sec. located under the word TUBE, the FIP is changed as
follows:

FORwARD BAcK CONfIRM EScAPE MANUaL

7. Press key ORG 1 or ORG 2 to scroll the tubes library forward or


backwards.
At each pressure of the keys the name of the tube whose data are
memorized is displayed on the FIP. When the name of the tube
corresponding to the one presently installed is displayed, press key
ORG 3 to confirm.
8. The FIP displays the following:

TUBE PARAMETERS WILL BE OVERWRITTEN


Confirm Escape

providing the data of the selected tube in RAM and EEprom.


9. To perform this operation, press the key located under the message
"Confirmation" and wait until the data memorization operation is
concluded.

(Rev. 3) 5-21 GENIUS HF


SERVICE MANUAL
Configurations

5.5.4. AEC chambers sensitivity calibration

1. Switch on the equipment via the ON key. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack and move the switch located
on the front part of the PCB 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press one of the vertical anatomical selection keys.
4. Press the Child key to enter the configuration.
5. The following is displayed on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

When key ORG 1 is pressed, the leds associated to the workstations


flash.
Press the key for the desired workstation and confirm the data via
key M1.
The FIP displays the following:

AEC camera

6. Press key +mA, the FIP is changed as follows:

AEC camera Amplif kV<80 kV>80


1 0050 0 0

GENIUS HF 5-22 (Rev. 3)


SERVICE MANUAL
Configurations

• Chamber number
The keys + and - mA select the number of AEC chamber (1,2, 3).
to be associated to the selected workstation.
• Amplification value
The keys + and - sec. set the amplification value (1 ... 9999).
This setting characterizes basic blackening for each couple of
film-screens.
The led corresponding to the film-screen key to be changed must
be active with this setting.
• Blackening compensation according to the kV value
The Reset and ORG 4 keys decrease or increase the value (-4 ÷
+4) of the blackening compensation for the <80 kV values for
each couple of film-screens.
The led corresponding to the film-screen key to be changed must
be active with this setting.
The M and ORG 5 keys decrease or increase the value (-4 ÷ +4) of
the blackening compensation for the >80 kV values for each
couple of film-screens.
The led corresponding to the film-screen key to be changed must
be active with this setting.
• Minimum dose value
The minimum dose (expressed in mV) is displayed on the
numerical kV fluoroscopy display, and must be read by the
exposure meter during emission in order to proceed with the
radiography.
If, during the initial milliseconds of exposure the measured dose
is lower than the set dose (caused by a too high density for the
set parameters or by lack of radiation of the selected
dominant[s]), radiography is immediately suspended and the
generator enters into the alarm state "AEC DOSE SAFETY".
The variation range for this parameter is between 0 and 3000
mV, which may be changed via the keys:
>5R increase 1 mV
Reset min. fluoroscopy decrease 1 mV
DS1 decrease 100 mV
DS2 increase 100 mV.

(Rev. 3) 5-23 GENIUS HF


SERVICE MANUAL
Configurations

7. Press key M1 to exit the AEC calibration routine.


The following message appears on the FIP:

CONFIRM ESCAPE

Press key ORG 2 to confirm the settings and memorize them.


Press key ORG 4 to cancel and exit the program.

GENIUS HF 5-24 (Rev. 3)


SERVICE MANUAL
Configurations

5.6. H.S. starter configuration

5.6.1. Mains voltage matching

JP27

JP23

The above figure shows the position of jumpers JP23 and JP27 for the
matching of the anode operation power to the three-phase mains voltage
in accordance with the following table (1 = connected, 0 = disconnected):

JP23 JP27

Mains < 370 V ac 1 1


370 V ac <= Mains < 390 V ac 1 0
390 V ac <= Mains < 410 V ac 0 0
Mains >= 410 V ac 0 1

(Rev. 3) 5-25 GENIUS HF


SERVICE MANUAL
Configurations

5.6.2. Braking activation at 1000 revs/minute

The following illustration shows the position of the jumpers for the
activation of the braking at 1000 revs. per minute for the various tubes.

JP3 JP11 JP16

Frenat. tubo 3
Frenat. tubo 1
Frenat. tubo 2

• JP 3 connected = activates the braking of tube T1


• JP11 connected = activates the braking of tube T2
• JP16 connected = activates the braking of tube T3.

GENIUS HF 5-26 (Rev. 3)


SERVICE MANUAL
Configurations

5.6.3. Jumper location for tube selection

FEDCBA FEDCBA FEDCBA

Tubo1 Tubo2 Tubo3

The above figure shows the position of the jumpers for the various
selections relative to the piloting of the anode for tubes T1, T2, T3.

There are three groups of jumpers:


• Tube 1 JP3 - JP4 - JP5 - JP6 - JP7 - JP8 - JP9
• Tube 2 JP11 - JP10 - JP26 - JP12 - JP13 - JP14 - JP15
• Tube 3 JP16 - JP17 - JP19 - JP18 - JP25 - JP21 - JP20
If present, jumpers JP3, JP11, and JP16 activate the braking at
1000 revs./min. of tubes 1, 2 and 3 respectively.

The arrangement of the jumpers for the function of the starter with the
various types of tubes is illustrated here below.
To simplify the illustration, the reference to the jumpers for each tube is
made with the initials JP followed by a letter.
This letter corresponds to the number of the jumper according to the
tube, as shown in the following figure:

JP A JP B JP C JP D JP E JP F

Tube 1 JP9 JP8 JP7 JP6 JP5 JP4


Tube 2 JP15 JP14 JP13 JP12 JP26 JP10
Tube 3 JP20 JP21 JP25 JP18 JP19 JP17

(Rev. 3) 5-27 GENIUS HF


SERVICE MANUAL
Configurations

5.6.4. Jumper arrangement for high speed tube

5.6.4.1. Comet tubes

Type of
Type of Tube JP A JP B JP C JP D JP E JP F
stator

Dx 71 H HS Stator 20-50 X l
Dx 91 H HS Stator 20-50 l X l
Dx 92 H HS Stator 20-50 l X l
Dx 93 H HS Stator 20-50 l X l
Dx 10 H HS Stator 20-50 l X l
Dx 101 H HS Stator 20-50 l X l
Dx 104 H HS Stator 20-50 l l X l
Dx 105 H HS Stator 20-50 l l X l
Dx 700 H HS Stator 20-50 l l X l

Legend:
l = jumper present
X = Read NOTES for instructions

NOTE:
/ Connect the jumper to position JP C as indicated if there is a
deceleration during fluoroscopy or if the alarm message "ANOTE
ROTATION ERROR" appears on the FIP.
If this situation persists, remove jumper JP C and connect it in JP D.

Anode connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 9


Inductive GTB *09 7 (20 Ω)
Capacitive GTB *10 8 (50 Ω)

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

GENIUS HF 5-28 (Rev. 3)


SERVICE MANUAL
Configurations

Relay configuration activation on starter card

Jumpers Speed Capacity K1 K2 K3 K4 K5

Slow 30 µF off off off on on


JP C open High 7,5 µF on on on off on
JP C closed High 10 µF on on on off on

(Rev. 3) 5-29 GENIUS HF


SERVICE MANUAL
Configurations

5.6.4.2. IAE tubes

Type of
Type of Tube JP A JP B JP C JP D JP E JP F
stator

RTM 102 HS Stator 20-42 l X l

Legend:
l = jumper present
X = Read NOTES for instructions

NOTE:
/ Connect the jumper to position JP C as indicated if there is a
deceleration during fluoroscopy or if the alarm message "ANOTE
ROTATION ERROR" appears on the FIP.

Anode connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 1


Inductive GTB *09 2 (20 Ω)
Capacitive GTB *10 2c (42 Ω)

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Jumpers Speed Capacity K1 K2 K3 K4 K5

Slow 30 µF off off off on on


JP C open High 7.5 µF on on on off on
JP C closed High 10 µF on on on off on

GENIUS HF 5-30 (Rev. 3)


SERVICE MANUAL
Configurations

5.6.4.3. Philips tubes

Type of Tube Type of stator JP A JP B JP C JP D JP E JP F

SRO 22 / 50 Stator ROT 350 l


SRO 32 / 100 Stator ROT 350 l
SRO 33 / 100 Stator ROT 350 l
SRM 35 / 100 Stator ROT 501 l

Legend:
l = jumpers present

Anode connections

ROT 350 ROT 501


Generator Terminal Board
Terminal Board Terminal Board

Common GTB *03 107-109-208-210 105-106


Inductive GTB *09 202-204 110
Capacitive GTB *10 101-103 102

the asterisk sign "*" substitutes numbers 1, 2 or 3 according to the


tube to be connected

Relay configuration activation on starter card

Speed Capacity K1 K2 K3 K4 K5

Slow 40 µF on off off on off


High 10 µF on on on off on

(Rev. 3) 5-31 GENIUS HF


SERVICE MANUAL
Configurations

5.6.4.4. Siemens tubes

Type of
Type of Tube JP A JP B JP C JP D JP E JP F
stator

xxxxxxxxxxxx Stator 14-30 l

Legend:
l = jumper present

Anode connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 0 - Ia


Inductive GTB *09 II (14 Ω)
Capacitive GTB *10 I - Ib (30 Ω)

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Speed Capacity K1 K2 K3 K4 K5

Slow 20 µF off off off off off


High 5 µF off off on on off

GENIUS HF 5-32 (Rev. 3)


SERVICE MANUAL
Configurations

5.6.4.5. Dunlee tubes

Type of Tube Type of Stator JP A JP B JP C JP D JP E JP F

PX 1400 Series housing 15-30 l l

Legend:
l = jumper present

Anode connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 White


Inductive GTB *09 Black (15 Ω)
Capacitive GTB *10 Red (30 Ω)

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Speed Capacity K1 K2 K3 K4 K5

Slow 40 µF off off off on off


High 7,5 µF on on on off on

(Rev. 3) 5-33 GENIUS HF


SERVICE MANUAL
Configurations

5.6.4.6. Varian tubes

Type of Tube Type of Stator JP A JP B JP C JP D JP E JP F

A 192 Stator B130 "STD" l X l


A 292 Stator B130 "STD" l X l

Legend:
l = jumper present
X = Read NOTES for instructions

NOTE:
/ Connect the jumper to position JP C as indicated if there is a
deceleration during fluoroscopy or if the alarm message "ANOTE
ROTATION ERROR" appears on the FIP.

Anode Connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 1 White


Inductive GTB *09 2 Black
Capacitive GTB *10 3 Green

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Jumpers Speed Capacity K1 K2 K3 K4 K5

Slow 40 µF off off off on off


JP C open High 7,5 µF on on on off on
JP C closed High 10 µF on on on off on

GENIUS HF 5-34 (Rev. 3)


SERVICE MANUAL
Configurations

5.6.4.7. CGR tubes

Type of
Type of Tube JP A JP B JP C JP D JP E JP F
Stator

550 Statorix 200 l l l l

Legend:
l = jumper present

Anode Connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 C


Inductive GTB *09 P
Capacitive GTB *10 A

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Speed Capacity K1 K2 K3 K4 K5

Slow 10 µF off off off on off


High 2,5 µF off off on off on

(Rev. 3) 5-35 GENIUS HF


SERVICE MANUAL
Configurations

5.6.4.8. Shimadzu tubes

Type of Tube Type of Stator JP A JP B JP C JP D JP E JP F

DI10HS in P38 Housing P38 l l


Hous.

Legend:
l = jumper present

Anode Connections

Generator Terminal Board Tube Terminal Board

Common GTB *03 X


Inductive GTB *09 Y
Capacitive GTB *10 Z

the asterisk sign "*" substitutes numbers 1, 2 or 3 according


to the tube to be connected

Relay configuration activation on starter card

Speed Capacity K1 K2 K3 K4 K5

Slow 30 µF off off off on on


High 2,5 µF off off on off on

GENIUS HF 5-36 (Rev. 3)


SERVICE MANUAL
Configurations

5.6.4.9. Unlisted tubes

If the characteristics of the tubes are not know, request the electrical
characteristics of the stator and the mechanical characteristics of the
rotating anode before connection.

(Rev. 3) 5-37 GENIUS HF


SERVICE MANUAL
Configurations

5.7. Tomography

The tomographical program is automatically activated when an input


relative to the set speed input and an input relative to the set angle are
present on the GTB 400 terminal board.
The following table provides the automatically set tomography times
displayed in function of the activation of the various inputs.

Speed Angle Tomography


input input time

GTB 443 (slow) GTB 445 (7°) 0.6 sec


GTB 443 (slow) GTB 446 (20°) 1.8 sec
GTB 443 (slow) GTB 447 (30°) 2.6 sec
GTB 443 (slow) GTB 448 (45°) 4.0 sec
GTB 444 (fast) GTB 445 (7°) 0.3 sec
GTB 444 (fast) GTB 446 (20°) 0.9 sec
GTB 444 (fast) GTB 447 (30°) 1.3 sec
GTB 444 (fast) GTB 448 (45°) 2.0 sec

GENIUS HF 5-38 (Rev. 3)


SERVICE MANUAL
Calibration

6. CALIBRATION

6.1. General

This chapter supplies all the necessary information for setting the
parameters for the X-ray tubes function.
The regulation of these parameters: focusing and correction of the high
voltage kV values are discussed here below.

(Rev. 3) 6-1 GENIUS HF


SERVICE MANUAL
Calibration

6.2. High voltage kV coefficient value

If the kV value detected by an instrument outside the generator is


different to the set value, it can be corrected using this coefficient.
Calibration of the kV is made at the manufacturer's upon the final
testing and verified via a standard instrument.
The base correction value (100) can be changed using a lower datum to
reduce the value of the high voltage kV.
A higher kV will increase the value.
Every unit above or below 100 will effect a percentage variation to the
high voltage kV value.
1. Via the ON key, switch on the equipment. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack, turn the switch located on
the front of PCB 9999102700 Master Processing Unit (paragraph
12.1) to the right.
3. Press a vertical anatomical selection key.
4. Press the Child key to access the configuration.
5. The following message appears on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

Press the key ORG 3.


The FIP displays the following message:

CALIBRATE
k*kV

GENIUS HF 6-2 (Rev. 3)


SERVICE MANUAL
Calibration

6. Press the Reset key. The message Calibration k*kV will appear on the
FIP.
The value of the kV correction parameter will appear on the
fluoroscopy kV display.
To increase or decrease this value, press the keys >5R and Reset
min. fluoroscopy.
The default value is 100. This means that the multiplying factor of
the theoretical value of the kV is 1.
With this condition, the real value of the high voltage kV is equal to
the calibration value obtained during the testing of the equipment.
To verify the correspondence between the set kV values and the
detected values, connect an oscilloscope between ground and the
signal [ikv] on test point TP5 of PCB 9999192500 Converter Control
Board (paragraph 12.1).
The ratio for the reading of the reference value is 10 V = 200 kV.
Perform radiography with a current value of 100 mA.
If the value of the kV reading is less than 2% that of the set value,
the number to input is 102.
The same concept is used if the voltage value is higher.
The percentage error in excess must be calculated and this
percentage detracted from 100.

EXAMPLE:
Radiography data 90 kV - 100 mA - 0.8 sec
Value read on oscilloscope 4.65 V (ratio 10 V = 200 kV)
Real kV value 93
Error 3.3 %
Datum to input 103.

7. To confirm the changes made, press the key M1.


The changed data input will be saved on EEPROM. Confirmation of
the save will be testified by a beep.
By pressing the Aux key the changes will be cancelled.
8. Press Sturdy Build to quit the configuration program.
The generator will return to operating mode and new verifications can
be made.

(Rev. 3) 6-3 GENIUS HF


SERVICE MANUAL
Calibration

6.3. Filament heating characteristic values

These are the current values for the filament heating expressed in amps
and obtained from the filament characteristics supplied by the
manufacturer of the radiogenic tube.
These values must be input into the computer and are used to inform the
relevant circuit to regulate the value of the filament heating current.

6.3.1. Stand-by current

This current value must be given for both focuses and for all the tubes
installed.
The typical stand-by current value for tubes with filament of a maximum
current of 5.5 A and 2.5 A.
In all cases, this is the value that must be set so as not to cause an issue
of high voltage current.

6.3.2. Minimum radiography current

This value corresponds to the value of the filament current that enables
us to obtain approximate 6 mA in radiography on the small focus and
25 mA in radiography for the large focus.
This value is obtained from the filament heating curves given by the
manufacturer of the radiogenic tube and may be adjusted during the
calibration procedure of the radiography current.

6.3.3. Maximum radiography current

The maximum filament current in radiography is a value given by the


manufacturer of the radiogenic tube.
The typical value is 5.5 A.
The minimum and maximum filament current value for a particular
focus is used by the computer for the calculation of the heating current
during the radiography preparation.
If one of the two values is changed after calibration is made, all the
values obtained during said calibration are cancelled.
The calibration of that particular focus and that particular tube must
therefore be repeated.
Caution must be taken to ensure that the maximum filament current
value does not exceed 8.5 A.

GENIUS HF 6-4 (Rev. 3)


SERVICE MANUAL
Calibration

6.4. Fluoroscopy heating calibration

Before commencing calibration exposure, the maximum heating current


value must be input.
This current value ranges between 2800 mA and 3400 mA for filaments
with a maximum heating current of 5.5 Amps.
As explained in paragraph 6.3.1, the preheating value of the large focus
must not cause an issue of current.
If this occurs, calibration must be performed otherwise it would be
impossible to bring the fluoroscopy current to the minimum value of
0.5 mA.
In this case, the relative alarm messages will be generated during
fluoroscopy calibration (see paragraph 10.1).
The maximum kV value must also be selected from the range
90 - 120 kV.

1. Via the ON key, switch on the equipment. After 5 seconds the console
will display the data program.
2. After opening the front panel of the power rack, turn the switch
located on the front of 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press a vertical anatomical selection key.
4. Press the Child key to access the configuration.
5. The following message appears on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

Press the key ORG 3.


The FIP displays the following message:

CALIBRATE
FLUORO

(Rev. 3) 6-5 GENIUS HF


SERVICE MANUAL
Calibration

6. Press the key +kV to access the fluoroscopy calibration program.


The FIP provides the selection of the tube on which to calibrate the
fluoroscopy:

FLUORO CALIBRATION
TuBE 1

Use key +mA to select the tube to calibrate.


Keys M1 and Sturdy Build are also active and have the function to
confirm or cancel the selection.

7. When key M1 is pressed, the following message appears on the


fluoroscopy display:
• display kV: the maximum kV value that the microprocessor
reaches during fluoroscopy calibration.
It is possible to change this value (from 90 to 120) using the keys
>5R (decrease) and Reset min. fluoroscopy (increase)
• display mA: the maximum heating value that the microprocessor
reaches during fluoroscopy calibration.
Transfer from one kV step to the next is generated by reaching
the set heating value. This value may be changed using the DSI 2
key to increase, or the DS1 key to decrease.
8. When the key M1 is pressed, the word READY appears on the FIP.
At this point the fluoroscopy procedure is ready to begin. By
shortcircuiting via pushbuttons or the accessory connected to points
GTB 427 and GTB 429 exposure commences.
The automatically set values are:
• kV = 50 kV (value indicated by the fluoroscopy kV display)
• mA = stand-by value of the small focus.
When fluoroscopy starts, the mA fluoroscopy reading must be
between 0.10 and 0.25 mA. This shows that the two focuses are at a
non-emission value.
The small focus heating is automatically increased and at the same
time the heating and radiogenic tube emission values are read and
memorized.
The emission value is given on the fluoroscopy mA display. Once the
maximum heating current value fluoroscopy is reached, the heating
value is automatically brought to stand-by value and the kV value to
80 kV.

GENIUS HF 6-6 (Rev. 3)


SERVICE MANUAL
Calibration

The above sequence is repeated and once the maximum mA values


are reached, the kV are set to the maximum value. Fluoroscopy is
automatically stopped at the end of the last cycle and the FIP
displays the message "FLUOROSCOPY CALIBRATION OK".
Release the rays command and press the Sturdy Build button to quit
the calibration routine.

NOTE:
/ The release of the fluoroscopy command blocks emission and
consequently blocks the heating detection procedure. The procedure is
resumed at the point of interruption via the fluoroscopy command.

(Rev. 3) 6-7 GENIUS HF


SERVICE MANUAL
Calibration

6.5. Radiography small focus calibration

The small focus calibration consists in the memorization of the data


relative to the filament heating values in function of the kV values.
These data are used during the radiography preparation phase to set the
filament heating value in function of the selected kV and mA.
A table containing the emission current values for the different heating
values and voltages is therefore necessary.
To generate this table inside the memory, the interval between the
minimum and maximum radiography current has been divided into 9
equal parts. Radiographs with different kV values must be performed for
each one of the ten current values derived therefrom.
The kV - mA couples values detected at each emission are memorized as
described here below.
A minimum of 3 value couples must be recognized for each filament
current value.
This makes it possible to calculate 3 coefficients for each one of the ten
filament current values which shall be called STEPS.

WARNING:
1 Because this is a manual calibration, data for the performance of
calibration radiographs will be provided throughout this entire phase.
It is the operator's responsibility to ensure that the limit values of the
radiogenic tube's maximum load are not exceeded.

Forecast of the kV and mA values obtained in the next exposure shall be


made as per the following description. Consequently, once he has
obtained the forecasted values, the operator shall decide whether it is
possible to proceed, or, if the data risk reaching or exceeding the
maximum load values, to finish the STEP.
With regard to the kV value, the equipment will propose the value to be
used, while the forecast of the mA emission value linked to the STEP
depends upon the value of the STEP itself.
At STEP 0, the filament is heated to the minimum radiography current
value and the forecast value is in function of this data. Once the first
exposure with STEP 0 at 40 kV has been made, the forecast of the
current value for the second exposure at the same STEP at 60 kV will be
an increase of 20% of the value read during the first radiography.
The increase of 20% will also be valid for the successive increases of high
voltage. The forecast of current increase for the successive STEPS shall
be of 50% of the current value detected at 40 kV in the previous STEP.
The calibration method described is the method for a tube whose
characteristics are not contained in the EEprom tube library.
This method is called the "manual method".
If the tube to be calibrated is contained in the library, the tube's data can
be loaded and, if necessary, the basic calibration data can be changed.

GENIUS HF 6-8 (Rev. 3)


SERVICE MANUAL
Calibration

6.5.1. Automatic calibration

As described above, if the tube to be calibrated has been memorized, it is


possible to effect a semi-automatic calibration of the tube.
This calibration consists in adapting the basic calibration to the installed
tube. Proceed with the loading of the installed tube's data as described in
paragraph 5.5.3.
Once this procedure has been completed, perform the following
procedure:

1. Via the ON key, switch on the equipment. After 5 seconds the console
will display the data program.
2. After opening the front panel of the power rack, turn the switch
located on the front of the PCB 9999102700 Master Processing Unit
(paragraph 12.1) to the right.
3. Press a vertical selection key.
4. Press the Child key to enter configuration.
5. The following message appears on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

Press key ORG 3.


The FIP displays the following:

CALIBRATE
X-RaY

(Rev. 3) 6-9 GENIUS HF


SERVICE MANUAL
Calibration

6. Press the key +mA to access the radiography calibration program.


The FIP changes as follows:

X-RAY CALIBRATION
TuBE 1 SMaLL FOCUS

The tube to be calibrated can be selected using the +mA key.


The focus to be calibrated can be selected using the Reset key.
Press key M1 after having set the desired values.
To quit this program at any time, press key Sturdy Build.
7. When key M1 is pressed, the following is displayed on the FIP:
– in correspondence to the radiography kV keys is displayed the
maximum heating value used during a given focus calibration
procedure (STEP 9). This value can be changed via the
radiography kV keys (see paragraph 6.3.3)
– in correspondence to the radiography mA keys is displayed the
minimum heating value used as starting point for the focus
calibration (STEP 0).
This value can be changed via the radiography mA keys (see
paragraph 6.3.2)
– in correspondence to the radiography sec. keys is displayed the
stand-by value for the selected focus.
This value can be changed via the radiography sec. keys (see
paragraph 6.3.1)
– fluoroscopy kV display. Here is displayed the variation coefficient
of the basic calibration.
This coefficient has a variation interval within the range of 0.900
and 1.100.
This data is used as variation factor for the filament heating
value.
If 1.000 is the basic calibration, a lower number would reduce
the mA emission value, while a greater number would increase
this value
– fluoroscopy mA display. Here is displayed the parameter that
automatically corrects the basic calibration.
This data assumes two values:
0 = inactive correction
1 = active correction.
If this condition is activated, the difference between the set mA
value and the real mA value is calculated at the end of each
exposure.
If the difference is greater than 2.5%, the calibration variation
coefficient is automatically adjusted.

GENIUS HF 6-10 (Rev. 3)


SERVICE MANUAL
Calibration

8. To perform this procedure it is necessary to measure the radiography


mA value.
Use an mA meter connected to the "LINK" connector (after having
removed the jumper) located above the high voltage transformer on
the PCB 8502 (paragraph 12.3).

WARNING:
1 Under no circumstances must exposure be made without the closure of
the "LINK" jumper.
This could damage the power inverter.

Once the tube's data has been loaded in memory, perform a


radiography with the following data.
60 kV - 100 mA - 40 ms.
The following three conditions may occur after this operation:
1. The generator is switched off by the circuit breaker.
Switch on again and reduce the coefficient significantly (e.g. from
1.000 to 0.900).
2. The generator is blocked by "mA Safety".
– The mA readings are too low, i.e. lower than those set. In this
case increase the coefficient.
– The mA readings are high, i.e. higher than those set. In this
case decrease the coefficient.
3. The generator Performs the radiography.
Correct the calibration coefficient if the mA value does not lie
within the limits required (10%) performing a radiography
releasing the prep. at each exposure and waiting for the
generator to return automatically to the limits. Repeat this point
until the desired mA value is reached.
Perform the radiographs with all the mA and kV values to verify
that all are within the desired tolerance limits.
9. To perform calibration of the large focus, perform the same procedure
as described for the calibration of the small focus.
At Point 6, select large focus.

(Rev. 3) 6-11 GENIUS HF


SERVICE MANUAL
Calibration

6.5.2. Small focus manual calibration

The calibration procedure is as follows:


1. Via the ON key, switch on the equipment. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack, turn the switch located on
the front of PCB 9999102700 Master Processing Unit (paragraph
12.1) to the right.
3. Press a vertical anatomical selection key.
4. Press the Child key to access the configuration.
5. The following message will appear on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ENGLISH MODIF PROG 0 PrOGRAMS ERASE

Press key ORG. 3.


The FIP displays the following:

CALiBRATE
X-RaY

6. Press the key +mA to access the radiography calibration program.


The FIP changes as follows:

X-RAY CALIBRATION
TuBE 1 SmaLL FOCUS

The tube to be calibrated can be selected using the +mA key.


The focus to be calibrated can be selected using the Reset key.
Press key M1 after having set the desired values.
To quit this program at any time, press key Sturdy Build.

GENIUS HF 6-12 (Rev. 3)


SERVICE MANUAL
Calibration

7. When key M1 is pressed, the following is displayed on the FIP:


– in correspondence to the radiography kV keys is displayed the
maximum heating value used during a given focus calibration
procedure (STEP 9). This value can be changed via the
radiography kV keys (see paragraph 6.3.3).
– in correspondence to the radiography mA keys is displayed the
minimum heating value used as starting point for the focus
calibration (STEP 0).
This value can be changed via the radiography mA keys (see
paragraph 6.3.2).
– in correspondence to the radiography sec. keys is displayed the
stand-by value for the selected focus.
This value can be changed via the radiography sec. keys (see
paragraph 6.3.1).
As mentioned above, to change one of the above parameters the
relative key must be pressed, bearing in consideration how much
variation is required as follows:
– left AEC led chamber ON = variation: 1 unit
– center AEC led chamber ON = variation: 10 units
– right AEC led chamber ON = variation: 100 units.
The variation of the radiography maximum or radiography minimum
value, causes the cancellation of the previous calibration made.
After having set the desired values, press the M1 key.
8. Press the key ORG 2 to memorize the new settings made or ORG 4 to
cancel.
9. The STEPS will now be displayed on the FIP:

STEP 0 = 3600
+000.00000 +000.00000 +000.00000

The first line of the FIP indicates the STEP number and the value of
the minimum filament current expressed in mA (3600). When the
second line of the FIP has the three values equal to zero, this means
that the STEP is not calibrated and calibration must be made.
The selection of the step is made via the ORG 1 (increase) and +mA
(decrease) keys.

(Rev. 3) 6-13 GENIUS HF


SERVICE MANUAL
Calibration

10. Once selection of the desired STEP has been made, press key M1 to
confirm.
The first line of the FIP displays the following information:

STEP 0 = 3600 kV = 040 kV = 000 mA = 0000


10 mA 40 kV 100 ms

– STEP 0 = 3600: indicates the number of the selected step and


the filament current heating value (3.6 A) used in this STEP
– kV = 040: indicates the kV value that will be used in the next
radiography
– kV=000 mA=0000: are couples of values that will be read by the
computer during radiography calibration and displayed in the
appropriate spaces. These values are given without decimals.
It is possible to change the kV value with which to perform the next
radiography using the + and - kV keys. This change may be
necessary to reduce the kW value and keep away from the maximum
load value.

The second line of the FIP gives the following information:


– mA RADIOGRAPHY: this means that the circuit breaker that
controls the flow of current during radiography is set at a
threshold of 100 mA increased by 40%. This value may be
changed up to 300 mA via radiography mA selection keys.
The manual setting of the current value must be made by
selecting a value close to the forecast current value.

WARNING:
1 The selection of a high current value with respect to the real value may
be due to a kV waveform with a strong oscillation at the start. The
selection of a low current value with respect to the real value may be due
to a kV waveform with a high ripple (greater than 10%) or to the
automatic switch-off of the equipment due to the circuit breaker (Led
DL21, PCB 9999103000 Safety Supervisor, paragraph 12.1).

It is useful to remember that the protection of the radiography


current is set as follows:
– 10 / 100 nominal mA: protected with a current of 100 mA
increased by 40%. Breaker intervention = 140 mA
– 150-200 nominal mA: protected with a current of 200 mA
increased by 40. Breaker intervention = 280 mA
– 320 nominal mA: protected with a current of 320 mA increased
by 40%. Breaker intervention = 440 mA
– 0.100 s radiography time: this time may be changed using the
radiography time selection keys.

GENIUS HF 6-14 (Rev. 3)


SERVICE MANUAL
Calibration

The selection of the radiography time during calibration depends


upon the speed of the instruments used to detect the kV and mA
couple values. Using the automatic reading system of the generator
the time of 0.1 seconds (1 tenth of a second) is optimum.
For the calibration of the large focus with high voltage current values
that exceed 300 mA, it is preferable to reduce the exposure time to
0.05 seconds because for greater times the instant current value
decreases.
As described here below, the calibration radiographs must be
performed using current values lower than 10 mA for the small focus
and lower than 32 mA for the large focus. When deciding the value of
the calibration radiography time, it is appropriate not to change the
times of the exposures within the same STEP.
If the high speed selection key LED is lit, this means that the
9000 revs. speed has been selected.
During this phase the operator decides which mode to accept (high or
low speed) according to the forecasted radiography values.
At the end of radiography the state or alarm state is displayed in
numeric codes on the fluoroscopy minutes numerical display.
At the end of radiography, the kV and mA values detected are
associated to the step of the kV used. A new kV value must then be
selected.
11. When the kV is changed (by pressing the key ORG 3), the FIP
changes as follows:

STEP 0 = 3600 kV = 060 kV = 000 mA = 0000


10 mA 60 kV 100 ms

Perform radiography with a voltage value of 60 kV.


The above procedure must also be applied to this phase.
12. Press key ORG 3 to increase the kV value.
The FIP changes as follows:

STEP 0 = 3600 kV = 080 kV = 000 mA = 0000


10 mA 80 kV 100 ms

Perform radiography with a voltage value of 80 kV.


The above procedure must also be applied to this phase.
3 couples of values are memorized at this point and it is possible to
proceed to the next STEP.

(Rev. 3) 6-15 GENIUS HF


SERVICE MANUAL
Calibration

We recommend however, to proceed with the calibration of the STEP


with the kV values of 100, 120 and 140, as proposed automatically
by the appliance in order to memorize a greater quantity of data and
guarantee a higher precision.
Confirm the data by pressing key M1.
13. The following will be displayed on the FIP:

STEP 0 = 3600 kV = 140 kV = 000 mA = 0000


+AAA.AAAAA +BBB.BBBBB +CCC.CCCCC

The real kV and mA fields (indicated as zero in the figure) will, in fact
indicate the values measured with the last exposure before pressing
the M1 key.
The second line of the FIP displays the three coefficients A, B and C
that were calculated during the last exposure.
By pressing the M1 key, the three calculated coefficients, including
their positive or negative sign, are memorized.
A beep confirms that we are ready to proceed with the calibrations
and with another STEP.
14. The following STEP is then proposed.
The procedure considered until now is valid for all STEPS through to
STEP 9.
Each time we confirm the values of the coefficients calculated by the
computer, the successive STEP will be proposed up the last which is
STEP 9.
If the maximum load of the radiogenic tube does not allow all the
STEPS to be completed, this is because the maximum radiography
current for the filament is too high.
Changing this value means the resetting of all the coefficients
calculated and the consequent total repetition of the calibration
exposures.
It is not necessary to complete the calibration of all ten STEPS but,
as mentioned above, the greater the volume of data memorized the
higher the precision of the emission during normal operations.
As well as being influenced by the thermal state of the anode and the
cathode of the radiogenic tube, the precision of the radiography
current value depends upon the precision of the calculations that the
appliance will perform during the radiography preparation phase

NOTE:
/ To reduce the effect due to the variation of the emission current, we
recommend performing the first radiographs approximately 30 minutes
after the focus has been switched on from stand-by.

To quit the program at the end of the calibration, press the Sturdy Build
key and set the service switch on the MPU to OFF.

GENIUS HF 6-16 (Rev. 3)


SERVICE MANUAL
Calibration

6.6. Radiography large focus calibration

The procedure for the calibration of the large focus is more or less the
same as for the small focus.
Follow the procedure as per the preceding paragraph, with the following
variations:
1. Select the large focus when the FIP displays the selection of the focus
to be calibrated.
2. The minimum emission current value at STEP 0 is not between 6 and
7 mA as for the small focus but between 20 and 25 mA.
This is because the heating values for a minimum value of 32 mA in
radiography are calculated.
3. The default value of the radiography calibration mA is not 100 mA as
for the small focus but 300 mA.
This value may be changed from 32 to 800 (1000 mA in the case of
the Genius HF 80 kW).
The circuit breaker value is the value selected on the display
increased by 40%.
4. Special attention must be given to the forecast value so that the
selection of the radiography times is such as not to cause exposures
with an overload of anodic power.
This could occur for high filament current values (high STEPS).
Short time exposure values are preferable for these STEPS.
Preferable times are lower or equal to 0.1 s with an optimum time
being 0.05 s.
5. If it is impossible to memorize three valid kV-mA couples with high
heating values and consistent emission current variation with voltage
values between 40 - 60 - 80 kV, the voltage value automatically
proposed by the program can be changed using the +/- kV keys.
With this possibility, the first exposure of the STEP can be performed
at 40 kV (a value automatically proposed by the program), the second
at 50 kV (set manually instead of 60 kV) and the third at 60 kV.
In this way the emission current variations due to increases in the kV
values will be limited and will not exceed the 1000 mA or 1200 mA
emission values that automatically block radiography.
The 800 mA current is the blocking limit for the Genius 65 kW
generator.
The 1000 mA current is the blocking limit for the Genius 80 kW
generator.

(Rev. 3) 6-17 GENIUS HF


SERVICE MANUAL
Calibration

6.7. Calibration copy

If there are two tubes of the same characteristics connected in one


generator, the calibration can be copied (not the parameters that are
characteristic to the tube) from one tube to another.
Proceed as follows:
1. Via the ON key, switch on the equipment. After 5 seconds the console
will display the data program.
2. Open the front panel of the power rack, turn the switch located on
the front of PCB 9999102700 Master Processing Unit (paragraph
12.1) to the right.
3. Press a vertical anatomical selection key.
4. Press the Child key to access the configuration.
5. The following message will appear on the FIP:

CONfIG ALArMDB CALiBRATE EePROM TIME


ITALIANO MODIF PROG 0 PrOGRAMS ERASE

Press key ORG. 3.


The FIP displays the following:

CALiBRATE
COpY

6. Press the key Reset to access the data copy program.


The FIP now proposes the selection of the tube from which to
copy the calibration (source) and the selection of the tube on
which to copy the calibration (destination):

COPY SOUrCE DESTiNATION


1 1

GENIUS HF 6-18 (Rev. 3)


SERVICE MANUAL
Calibration

7. Press the key +kV to change the number of the tube from which to
copy the calibration.
Press the key Reset to change the number of the tube on which to
copy the calibration.
Once the selection has been made confirm by pressing the key M1.
8. While the data is being copied, the FIP displays the message "PLEASE
WAIT", after which the FIP displays the following:

CONfIG ALArMDB CALiBRATE EePROM TIME


ITALIANO MODIF PROG 0 PrOGRAMS ERASE

9. Press Sturdy Build to quit the calibration program.

(Rev. 3) 6-19 GENIUS HF


SERVICE MANUAL
Calibration

6.8. Generator regulation loop

After having completed the calibrations described in the above


paragraphs, it is important to know how the generator regulation loop is
set by the computer.
The procedure is described here below.

The logic diagram of the 65 or 80 kW generator is given on Figure 6-1.


The blocks represent the physical subdivision of the generator (see
Figure 6-3).

• Block 1
This block constitutes the interface between the inverter with the
mains. Here there are the three-phase jumper safety switches, the
capacitors and inductors for the power supply to the DC-RAIL and to
enable the transfer of power to the inverter.

• Block 2
This block constitutes the power inverter and the converter control
board PCB 9999102500.

• Block 3
This block constitutes the H.V. transformer.

• Block 4
This block constitutes the electronic group.
This group contains:
– the PCB 9999103000 supervisor card and the power supply card
– the normal speed starter anode card PCB 9999103200
– the inverter filament driver PCB 9999102600
– the optional normal and high speed starter PCB 9999103400 /
PCB 9999103300.

• Block 5
This block constitutes the principal microprocessor (Master
Processing Unit) PCB 9999102700 and the A.E.C. PCB 9999104300.

• Block 6
This block is positioned at the front of the generator and is called the
general terminal board (GTB). It is the interfacing block with the
outside accessories.

GENIUS HF 6-20 (Rev. 3)


SERVICE MANUAL
Calibration

6.8.1. kV regulation loop

Refer to the Figure 6-2 for a better understanding of the description here
below.
The kV regulation loop is completely analogic.
This means that the microprocessor does not enter into this regulation
directly. It recognizes the real value of the KV and the state of the
generator thanks to the digital signals received from the inverter.

When a preparation is required, the microprocessor prepares the signals


[IRA], [IRC].
It then checks that the [IPW] and the [IUX] signals are active.
The kV reference value is then set under the form of an analogic signal
(OKV) with a scale of 1 V dc = 20 kV.
All the signals are for the external devices and the generator supervisor.

When the [OHV] signal is sent to the CCB radiography commences.


From this moment until the end of the exposure, the kV control is made
by the CCB.
The microprocessor waits for the [IRM] signal. If this should fail during
radiography, the state of alarm is activated and all the operating cycles
are interrupted.
The same check is made on the analogic signal [IKV].
When the [OHV] signal reaches the CCB, the kV ref. signal (OKV) is
compared with the real kV value.
The "regulated" signal reaches a voltage - frequency converter generating
the [FRE] signal.
The [FRE] signal is the real command signal of the power devices.
Before reaching the IGBT, it passes through an activation logic and
finally becomes either [GF1-4] (command of one side of the jumper) or
[GF2-3] (command of the other side of the jumper).
If 166kV is exceeded in output, an alarm signal blocks the inverter.

(Rev. 3) 6-21 GENIUS HF


SERVICE MANUAL
Calibration

6.8.2. mA regulation loop

Refer to the Figure 6-2 for a better understanding of the description here
below.
The mA regulation loop is completely controlled by the microprocessor.
This means the microprocessor recognizes the mA emission value that at
all times during exposure.
The microprocessor can change the emission value by changing the
filament heating value.
The measurements involved in this regulation are:
• Current measurement circuit
– (ich) for high radiography current
– (icm) for medium radiography current
– (ifc) for fluoroscopy current.
The current signal is taken from the H.V. transformer and taken to
the CCB by the J4 connector.
Here, it is measured by a circuit and amplified according to the
following scales:
– (ich): 10 V dc = 1500 mA
– (icm): 10 V dc = 150 mA
– (icl): 10 V dc = 10 mA.
• MPU card
A/D of the microprocessor, used to convert the analogic signal of the
mA to a digital signal and, D/A, to generate the reference analogic
signal of the filament heating.
• Inverter filament driver card
This card provides for the heating of the filaments.

In order to recognize the characteristics of the tube emission, the


microprocessor must recognize the tube calibration.
In this way it can calculate the value at which it must heat the filament
for the kV-mA couple selected by the operator. The microprocessor
requires this value only for the start of the exposure and this value must
be as near as possible to the value of the selected mA. After a few
seconds from the start of exposure the microprocessor starts to monitor
the emission mA values and consequently regulates the heating.
If the heating mA value is too far from that selected (25% for low values,
12.5% for high values), the microprocessor halts radiography and enters
into alarm mode. This it because it decides that regulation is no longer
possible for reaching the correct emission value and therefore achieve a
good result.
The alarm state may occur for two reasons:
– uncalibrated tube
– incorrect reading of the emission mA values caused by a malfunction
in the mA measurement loop or heating regulation loop.

GENIUS HF 6-22 (Rev. 3)


SERVICE MANUAL
Calibration

Figure 6-1

(Rev. 3) 6-23 GENIUS HF


SERVICE MANUAL
Calibration

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 6-24 (Rev. 3)


SERVICE MANUAL
Calibration

Figure 6-2
(Rev. 3) 6-25 GENIUS HF
SERVICE MANUAL
Calibration

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 6-26 (Rev. 3)


SERVICE MANUAL
Calibration

Figure 6-3

(Rev. 3) 6-27 GENIUS HF


SERVICE MANUAL
Calibration

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 6-28 (Rev. 3)


SERVICE MANUAL
Interface with digital system

7. INTERFACE WITH DIGITAL SYSTEM

7.1. Characteristics

The characteristics of the Genius HF 65 - 80 for interfacing with digital


image acquisition appliances are the following:
• Images per second during radiography
– max. no.: 50 Imgs./sec
– output power at 50 Imgs./sec: limited by the tube power

• During continued fluoroscopy


– max. current: 5 mA
– max. voltage: 120 kV

• During pulsed fluoroscopy


– no. images/max sec: 50 Imgs./sec
– mA current: selectable from 30 to 300 mA in 7 steps
– max. voltage: 120 kV
– pulse time: selectable from 2 to 10 msecs

• Automatic dose control for continued and pulsed fluoroscopy


modes
– signal: analogic
– range: 0 ÷ 1.4 V with SW regulation of the balance value

• Zero points technique


– type of transfer: automatic setting of radiography parameters
derived from fluoroscopy
– transfer tables: 3

• Step digital angiography


– mode: kV decrease at the end of each station.

(Rev. 3) 7-1 GENIUS HF


SERVICE MANUAL
Interface with digital system

7.2. Signals

The connection between the Genius HF and the VDR 2000 digital system
has been optimized and simplified thanks to the highly advanced
software developed.
Few signals are therefore used to synchronize the generator with the
digital system.
These in/out signals are present on the PCB.

Position Type Notes


GTB 421 / 424 / 0 V dc This voltage is used as ground reference of the
427 / ecc interface with the digital system
GTB 429 IFS Start command for continued fluoroscopy
GTB 434 IPF Start command for pulsed fluoroscopy
GTB 422 IPR Start command for preparation
GTB 426 ISR The ready release or input from the external
accessory
GTB 113 PLS Polarization contact closed when the generator
is ready for radiography start
GTB 432 IRP Rays start command
GTB 328 ORT Rays present during radiography
GTB 329 OPT Rays present during continued fluoroscopy
GTB 330 OSC Rays present during pulsed fluoroscopy
GTB 401 / 402 iec + / iec - Analogic signal for kV regulation command
during continued and pulsed fluoroscopy

GENIUS HF 7-2 (Rev. 3)


SERVICE MANUAL
Interface with digital system

7.3. Signals timing

The term signals timing indicates the time that those signals involved in
a particular sequence (radiography and pulsed or continuous
fluoroscopy) require to achieve a correct functioning of the machine.

7.3.1. Pulsed fluoroscopy timing (high contrast fluoroscopy)

The following tables provides a summary of the sequence.

Position Type Notes


GTB 434 IN The pulsed fluoroscopy request signal is sent
from the accessory
GTB 429 IN The fluoroscopy start signal is sent from the
digital system and together with the 434
generates the prep. necessary for the launch of
the anode and the heating of the filament at its
correct value
GTB 113 OUT The generator informs that the preparation
operations have been successfully completed
and that everything is ready for emission
GTB 432 IN The rays command is sent from the digital
system
GTB 328 OUT The generator informs the digital system of the
presence of the rays
GTB 410/402 IN The analogic ref. automatically regulates the kV

7.3.2. Continuous fluoroscopy timing

The following tables provides a summary of the sequence.

Position Type Notes


GTB 429 IN The continuous fluoroscopy request signal is
sent from the accessory
GTB 329 IN The generator informs the digital system of the
presence of the rays
GTB 410/402 IN The analogic ref. automatically regulates the kV

(Rev. 3) 7-3 GENIUS HF


SERVICE MANUAL
Interface with digital system

7.3.3. Digital radiography timing

Position Type Notes


GTB 422 IN The signal is sent from the digital system for
the launch of the anode and the heating of the
filament at its correct value
GTB 113 OUT The generator informs that the preparation
operations have been successfully completed
and that everything is ready for emission
GTB 432 IN The rays command is sent from the digital
system
GTB 328 OUT The generator informs the digital system of the
presence of the rays

GENIUS HF 7-4 (Rev. 3)


SERVICE MANUAL
Interface with digital system

7.4. Configuration program

As described in paragraph 5.5 the workstation on which the digital


system is operating must be configured.
In summary, the digital system must operate with:
• Continued and pulsed fluoroscopy
• Max. continuous current = 5 mA
• Pulsed current = 30 mA
• Max. voltage for both = 120 kV
• Regulation for both = analogic
• Regulation mode = T1 (single channel for both continuous and
pulsed)
• Regulation coefficient proportional for both Kp = from 20 to 25
depending on the system
• Correction activation window around the parity value B = 1003
• Reference signal value C = +0703.

In the configuration operations the dose level for the fluoroscopy must be
regulated.
The parity value [C] defined as that voltage value for which we consider
reached the optimal dose level is intrinsic to the TV camera or TV system
used. The variation interval is 1.5 Volts.
The variation range of this signal [B] must then be set, i.e. the window
around the parity value on which the signal is piloted.
This couple of values must be input for both the pulsed and the
continuous fluoroscopy.
As mentioned, the typical values of these data are B=1003 and C=+703
as they are closely linked to the VDR 2000 digital device used.
The above numbers are compound numbers. Their significance is as
follows:
• C= ±abcd
a= thousands of equal mV
b= hundreds of equal mV
c= tens of equal mV
d= sampling interval with values from 0 = 16 msecs to
9 = 200 msecs
• B= efgh
e= window for reading in thousands mV around a parity value
f= window for reading in hundreds mV around a parity value
g= window for reading in tens mV around a parity value
h= dead band with values from 0 = 7 bit to 9 = 210 bit.

(Rev. 3) 7-5 GENIUS HF


SERVICE MANUAL
Interface with digital system

7.5. 0 points technique

The term Zero Point Technique indicates a working technique where the
radiography parameters are automatically set by the system on the basis
of the value obtained either manually or automatically during
fluoroscopy.
The fluoroscopy kV value is automatically regulated according to the
density of the patient during fluoroscopy.
The radiography time displayed on the generator in this mode is to be
considered as the limit time.

7.5.1. Activation at a workstation

The 0 point technique may be activated at the desired workstation.


Activation or deactivation of this mode is made by pressing the key +s.
The 00 value indicates that the transfer is not active.
Any value other than 00 activates this function.

kV GRA 00

7.5.2. Use of the 0 point technique

If the kV transfer mode has been activated at the selected workstation,


one of the three keys DS1 or DS2 or DS3 must be pressed in order to use
the associated transfer tables.

GENIUS HF 7-6 (Rev. 3)


SERVICE MANUAL
Generator status information

8. GENERATOR STATUS INFORMATION


(DISPLAY ON MPU)
On the PCB 9999102700 MPU inside the power rack, there is an
alphanumerical display on which are displayed all the input, output,
setting, analogic and digital data handled by the central microcomputer.
All the data are arranged in pages, each of may be accessed by pressing
the button located on the right hand side of the display.
The change page command is given by pressing up éto display the
previous page or down ê to display the next page.

When the equipment is switched on in the display show the following:

GENERATOR
MVB xxx.yy
Where xxx represents the level of the version of the appliance and yy the
software revision level.

Using the selector you can scroll through the pages stored in the
memory.
Upon switch-on, the command up é on the switch will not have effect
since we are in the first page of the memory, the command down ê will
display the switching on values of the focuses (in mA) where:
• SF = small focus
• LF = large focus
• o = reference value
• i = the real switch-on value.

SF o=2500 i=2499
LF o=2000 i=2000

(Rev. 3) 8-1 GENIUS HF


SERVICE MANUAL
Generator status information

By pressing up é, you go to the previous page, by pressing down ê you


display the current selections:
• W1 = workstation 1
• T1 = selected tube
• NS = the selected anode rotation speed calculated in function of the
radiography parameters set (NS = low speed, HS = high speed)
• kVr = kV reference value (10 V = 200 kV).
On the second line the focus heating selected on the console (in
radiography) is displayed.

W1 T1 NS kVr = 0000
LF o=2000 i=2000
The command down ê will display:
• Bk = the breaker values sent to the safety values in relation to the
radiography parameters set.
The second line shows the gain value for the exposure meter circuit
calculated by the microprocessor on the basis of the preset and
configuration values:

Rp = xxxx Bk = xxxx
AEC = xxxx xxxxxxxx
The command down ê will display the mA value, the kV value, the time
and the mAs of the last radiography.

mA = xxxx kVp = xxxx


ms = xxxx mAs = xxxx
At this point there follow three pages for future use and not yet
implemented.

NFH = xxxx kVr = xxxx


mA = xx.xx kVa = xxxx
On the page after the three unused pages are diplayed the digital inputs
sent to the MPU in the form of logic states 0 / 1.

GENIUS HF 8-2 (Rev. 3)


SERVICE MANUAL
Generator status information

The first line shows the numbers associated with the inputs.
In order to recognize the function of the input, consult the PCB
9999102700 MPU diagram (paragraph 12.1) dedicated to digital inputs.
The second line shows the state of the inputs with the sequence from
right to left.
Since it is not possible to display all the inputs in a single page, they are
divided into 9 different screens.

IN Dig. 31 .. 24
0 1 0 1 1 1 0 0
Following the last page dedicated to inputs are 9 pages containing the
state of the digital outputs issued by the MPU.
The first line shows the numbers associated to the outputs (for
codification see diagram 9999102700 paragraph 12.1).
The second line shows the state of the signals in sequence from right to
left.

OUT Dig. 23 .. 16
0 1 0 1 1 1 0 0
Following the 9 pages containing the output signals there are 7 pages
containing the analogic in and out values that depend on the MPU.
To identify the function consult PCB 9999102700 (paragraph 12.1).

U1 = xxxx U2 = xxxx
U3 = xxxx U4 = xxxx

(Rev. 3) 8-3 GENIUS HF


SERVICE MANUAL
Generator status information

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 8-4 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9. SYNOPTIC TABLES FOR EQUIPMENT


SET-UP
This chapter contains the synoptic tables that provide a concise
summary of the instructions provide in the configurations, set-up and
calibration chapters.

These tables contain the icons that represent the actions to be performed
and the images relative to the keys on the command console that in these
operations have the functions of the input and confirmation of data.

The relative symbols are described in paragraph 2.6.

(Rev. 3) 9-1 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 9-2 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.1. Language change

Turn the switch located on the


1 front part of the MPU card to the
right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the mA increase key


(indicated in bold type in point 4)
5 to scroll through the languages
available in memory: Italiano,
Deutsch, Français, Español, etc.

Press the M1 key to perform the


selected set-up.
6

Press the sturdy build key to quit


the configuration routine.
7

Turn the switch located on the


front part of the MPU card to the
8 left (access to service routine) to
return to the normal operating
conditions.

(Rev. 3) 9-3 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 9-4 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.2. Configuration of the work stations

Turn the switch located on the


1 front part of the MPU card to the
right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the ORG1 key (indicated in


bold type in point 4) to access the
5 work stations configuration
program.

Select the work station to be


configured. The leds associated
6 with the relative work station
÷ keys will flash.

The FIP will indicate the name to


associate to the selected work
station.
7

Using the ORG1 and the ORG2


keys, scroll through the names
that may be associated to the
8 workstations. The ORG1 key
scrolls forward, the ORG2 scrolls
backwards.

(Rev. 3) 9-5 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the preparation key. The


FIP displays the letter "C"
indicating that the rays command
of the console key has been
activated. If the key is pressed
again, the character "-" is
displayed, indicating that the
command rays control is
deactivated. Press the rays key.
The FIP displays the letter "P"
indicating that the rays command
of an external accessory has been
9 activated. If the key is pressed
again, the character "-" is
displayed indicating that the
external rays command has been
deactivated.
The possible combinations are:
CP = activation from console and
external
C- = activation only from the
console
-P = activation only externally
-- = deactivated workstation.

The ORG3 key selects the tube to


associate to the workstation to be
configured. The tube number and
the anode's rotation mode must
be set as follows:
10 T"n" = Tube "n" with low and
high rotation (if the
tube allows it)
T"n"HS = Tube "n" with high
rotation always (if the
tube allows it).

If fluoroscopy is activated at the


workstation under configuration,
the dose adjustment mode in
continuous fluoroscopy is set
using key M. The possible modes
are:
ANA = adjustment using an
11 analogic reference signal
DIG = adjustment using the
input UP and DOWN
inputs
When configurating the appliance
as indicated in this manual, the
mode to select is always the ANA
mode.

GENIUS HF 9-6 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

In the case of continuous


fluoroscopy with ANA adjustment
set at the workstation under
configuration, the coefficient of
the proportional error may be
12 adjusted using the ORG4 and
ORG5 keys. The values that may
be set are within the range of 0 to
99. The ORG4 key increases the
value, the ORG5 key decreases
the value.

If the digital VDR is present at


the workstation under
configuration, the 0-point
technique (setting of the
radiography parameters in
function of the kV values of the
13 previous fluoroscopy) may be set
using the sec. increase key
(indicated in bold type in the
figure). The 00 datum indicates
that the 0-point technique is not
activated, any other number
activates the technique.

Press the front/back projection


selection key to set the kV, mA
and maximum time values for
14 continuous fluoroscopy.

Press the fluoroscopy time reset


key to set the maximum kV value
obtainable in continuous
fluoroscopy. The 100, 110, 120
and ---- values are shown
15 cyclically each time the key is
pressed and displayed on the
fluoroscopy kV DISPLAY. The
symbol ---- excludes the
fluoroscopy mode from that
workstation.

Press the 5R key to set the mA


values obtainable at the
maximum kV. The values 0.5, 1,
1.5, 2, 2.5, 3, 4 and 5 are shown
cyclically each time the key is
16 pressed and displayed on the
fluoroscopy mA DISPLAY.

(Rev. 3) 9-7 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the DSI1 key to set the


maximum mA values. The values
0.5, 1, 1.5, 2, 2.5, 3, 4 and 5 are
shown cyclically each time the
key is pressed and displayed on
17 the fluoroscopy min DISPLAY.

Press the DSI2 key to set the time


after which fluoroscopy is
stopped. When the key is pressed
the relative led lights up. This
18 means that the limit time is set at
90 minutes. When the key is
pressed again the led switches off
indicating that the limit time is
set at 5 minutes.

Press the lateral projection


selection key to set the kV, mA
and pulse time values for pulsed
19 fluoroscopy.

Press the fluoroscopy time reset


key to increase the anode rotation
time in fine fluoroscopy or fine
radiography mode. When the key
5R is pressed, the time is
20 decreased. The time expressed in
seconds within the range of 0 to
99 is displayed on the
fluoroscopy kV DISPLAY.

If the workstation under


configuration has been activated
for pulsed fluoroscopy, the pulse
mA value must be set. Press the
key DSI1. The value selected
21 within the range of 30 to 300 mA
is displayed on the fluoroscopy
mA DISPLAY. The symbol ----
indicates that the function has
been deactivated.

GENIUS HF 9-8 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The pulse time value must be set.


Press the key DSI2. The value
selected within the range of 2 to
10 msecs is displayed on the
fluoroscopy min DISPLAY. The
22 symbol ---- indicates that the
function has been deactivated.

If during steps 21 and 22 values


were set to activate pulsed
fluoroscopy, the dose adjustment
mode in pulsed fluoroscopy is set
using the mA increase key.
The possible modes are:
ANA = adjustment using an
23 analogic reference signal
DIG = adjustment using the
input UP and DOWN
inputs
When configurating the appliance
as indicated in this manual, the
mode to select is always the ANA
mode.

In the case of continuous


fluoroscopy with ANA adjustment
set at the workstation under
configuration, the coefficient of
the proportional error may be
24 adjusted using the ORG4 and
ORG5 keys. The values that may
be set are within the range of 0 to
99. The ORG4 key increases the
value, the ORG5 key decreases
the value.

If a measuring chamber is
present at the workstation under
configuration, the areas relative
to the dominants must be
activated. This is made by
25 pressing the key relative to the
dominant to be activated. The
lighting up of led on the key
pressed confirms the activation of
the dominant.

(Rev. 3) 9-9 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the increase key to activate


the "rays alarm" function (see
paragraph 4.10.6). Confirmation
26 is given when the 7 AEC fine
adjustment leds light up. To
deactivate (leds off) press the
decrease key.

To end the configuration routine


press the key M1.
27

The FIP is displayed as indicated


in the figure, requesting the
confirmation or cancellation of
the settings made during the
28 above phases.

If the key ORG2 is pressed, the


settings will be confirmed. A
buzzer will sound at the end of
29 the memorization phase
signalling that the operation has
been successfully completed.

If the key ORG4 is pressed, the


settings made during the above
30 phases will be cancelled and the
memorized data will not be
changed.

Press the sturdy build key to quit


the configuration routine.
31

Turn the switch located on the


front of the MPU card to the left
32 to return to normal operating
conditions.

GENIUS HF 9-10 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.3. Input of the X-ray tube data

Turn the switch located on the


1 front part of the MPU card to the
right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the ORG1 key (indicated in


bold type in point 4) to access the
5 workstations configuration
program.

Select the workstation to which


the tube to be configured is
6 associated. The leds of the keys of
÷ the workstations flash.

Press the key M1 to continue


with the configuration routine.
7

The FIP display is illustrated in


the figure.

Press the sec. increase key


(indicated in bold type in the
9 figure in point 8 above) to display
the tube data change access
screen.

(Rev. 3) 9-11 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

The figure shows the FIP display.


The function of keys ORG1, 2, 3,
4 and 5 are described here below.
10

Press the key ORG1 to select


(scroll forward) one of the tubes
11 whose characteristics are stored
in the EEprom memory.

Press the key ORG2 to select


(scroll backward) one of the tubes
12 whose characteristics are stored
in the EEprom memory.

Press the key ORG3 to confirm


the selection (as described here
13 below).

If the key ORG4 is pressed, the


operations made until this point
14 will be cancelled and the program
will return to point 4.

The key ORG5 changes the tube


data and gives access to the
15 manual calibration program.

As described in the above points, it is possible to select a tube that has


been stored in the memory and perform an automatic calibration of the
tube or set the data and proceed with a manual calibration of the tube.

GENIUS HF 9-12 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.3.1. Set-up of a tube stored in the memory

Using the ORG1 and ORG 2


tubes, scroll the EEprom memory
containing the data of the various
memorized tubes forwards or
backwards and select the desired
1 tube. The model of the tube is
displayed on the bottom line of
the FIB. Once the desired tube
has been displayed, press the key
ORG3 to confirm the selection.

The FIP display is illustrated in


the figure.

Press the kV increase key to


confirm the selection of the tube.
A buzzer indicates the end of the
3 memorization phase and
successful completion thereof.

If the RESET button is pressed,


all the settings made during the
4 above phases will be cancelled
and the data in the memory will
not be changed.

Press the sturdy build key to quit


the configuration routine.
5

Turn the switch located on the


front part of the MPU card to the
6 left to return to normal operating
conditions.

(Rev. 3) 9-13 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

9.3.2. Manual setting of the tube data

Press the key ORG5 to access the


manual setting of the data
1 relative to the installed tube that
has not yet been stored in the
memory.

Press the front/back selection


key. The values described in
points 3, 5, 7 and 8 are displayed
2 on the fluoroscopy DISPLAY. The
FIP changes as indicated in point
9.

The maximum kV value


reachable during radiography is
displayed on the kV DISPLAY.
Press the key 5R to decrease the
3 value and the fluoroscopy time
reset key to increase it.

Select fine focus using the


relative key.
4
The maximum mA value
reachable during radiography for
fine focus is displayed on the mA
DISPLAY. Press the key DSI1 to
5 decrease the value and the key
DS12 to increase it.

Select large focus using the


relative key.
6

The maximum mA value


reachable during radiography for
large focus is displayed on the
mA DISPLAY. Press the key DSI1
7 to decrease the value and the key
DS12 to increase it.

GENIUS HF 9-14 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The rotation values of the tubes


during radiography are displayed
on the fluoroscopy min. DISPLAY.
To effect the settings press the
key DSI3. The information
displayed is detailed herebelow:
1 = Rotation only at 3000
8 turns/min (low speed)
2 = Rotation only at 9000
turns/min (high speed)
3 = Rotation at both 3000 and
9000 turns/min in function
of the load set (high and low
speed).

The FIG shows the values of the


cooling curve for 0, 2, 4, 6, and
8 mins.
9

With the mA increase and kV


increase keys, move the cursor
and position it in correspondence
with the field to be changed. The
10 key mA increase moves the
cursor from right to left, the key
kV increase moves the cursor
from left to right.

The AEC fine adjustment keys


change the value of the field on
which the cursor has been
positioned. If we associate the
lighting of one of the leds relative
to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
11 variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.
NOTE:
The difference between 2 values
must be equal by at least 5%. If
not, the datum is not accepted.

(Rev. 3) 9-15 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the ORG1 key to display


the values relative to the
12 maximum fine focus load with a
rotation of 3000 turns/min.

The FIP display is illustrated in


the figure. The maximum load
value in function of the exposure
13 time with a voltage value
(kV = 100) must be calculated as
described on paragraph 5.5.2.

With the mA increase and kV


increase keys, move the cursor
and position it in correspondence
with the field to be changed. The
14 key mA increase moves the
cursor from right to left, the key
kV increase moves the cursor
from left to right.

The AEC fine adjustment keys


change the value of the field on
which the cursor has been
positioned. If we associate the
lighting of one of the leds relative
to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
15 variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.
NOTE:
The difference between 2 values
must be equal by at least 5%. If
not, the datum is not accepted.

Press the ORG1 key to display


the values relative to the
16 maximum large focus load with a
rotation of 3000 turns/min.

The FIP display is illustrated in


the figure. The maximum load
value in function of the exposure
17 time with a voltage value (kV =
100) must be calculated as
described on paragraph 5.5.2.

GENIUS HF 9-16 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

With the mA increase and kV


increase keys, move the cursor
and position it in correspondence
with the field to be changed. The
18 key mA increase moves the
cursor from right to left, the key
kV increase moves the cursor
from left to right.

The AEC fine adjustment keys


change the value of the field on
which the cursor has been
positioned. If we associate the
lighting of one of the leds relative
to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
19 variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.
NOTE:
The difference between 2 values
must be equal by at least 5%. If
not, the datum is not accepted.

If the tube can rotate at high


speed, press the key ORG1 to
20 display the values of the
maximum fine focus load at a
rotation of 9000 turns/min.

The FIP display is illustrated in


the figure. The maximum load
value in function of the exposure
21 time with a voltage value (kV =
100) must be calculated as
described on paragraph 5.5.2.

With the mA increase and kV


increase keys, move the cursor
and position it in correspondence
with the field to be changed. The
22 key mA increase moves the
cursor from right to left, the key
kV increase moves the cursor
from left to right.

(Rev. 3) 9-17 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

The AEC fine adjustment keys


change the value of the field on
which the cursor has been
positioned. If we associate the
lighting of one of the leds relative
to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
23 variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.
NOTE:
The difference between 2 values
must be equal by at least 5%. If
not, the datum is not accepted.

If the tube can rotate at high


speed, press the key ORG1 to
24 display the values of the
maximum large focus load at a
rotation of 9000 turns/min.

The FIP display is illustrated in


the figure. The maximum load
value in function of the exposure
25 time with a voltage value (kV =
100) must be calculated as
described on paragraph 5.5.2.

With the mA increase and kV


increase keys, move the cursor
and position it in correspondence
with the field to be changed. The
26 key mA increase moves the
cursor from right to left, the key
kV increase moves the cursor
from left to right.

GENIUS HF 9-18 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The AEC fine adjustment keys


change the value of the field on
which the cursor has been
positioned. If we associate the
lighting of one of the leds relative
to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
27 variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.
NOTE:
The difference between 2 values
must be equal by at least 5%. If
not, the datum is not accepted.

Press the side projection selection


key. The fluoroscopy DISPLAYS
will display the values described
28 here below.

The fluoroscopy kV DISPLAY will


display the switch-on time for
fine focus expressed in secs.
Press the key 5R to decrease the
29 value and the fluoroscopy time
reset key to increase it.

The fluoroscopy mA DISPLAY will


display the switch-on time for
large focus expressed in secs.
Press the key DSI1 to decrease
30 the value and the key DSI2 to
increase it.

The value of the switch-on


"Boost" for both filaments are
displayed on the fluoroscopy min.
DISPLAY expressed in msecs.
31 This value gives a rapid switch-on
of the emission value set. An
excessive value can generate an
incorrect initial emission. The
normal value set is 300.

(Rev. 3) 9-19 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

To end the configuration routine


press the key M1.
32

The FIP is displayed as indicated


in the figure, requesting the
confirmation or cancellation of
the settings made during the
33 above phases.

If the key ORG2 is pressed, the


settings will be confirmed. A
buzzer will sound at the end of
34 the memorization phase
signalling that the operation has
been successfully completed.

If the key ORG4 is pressed, the


settings made during the above
35 phases will be cancelled and the
memorized data will not be
changed.

Press the sturdy build key to quit


the configuration routine.
36

Turn the switch located on the


front of the MPU card to the left
37 to return to normal operating
conditions.

GENIUS HF 9-20 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.4. Radiography calibration

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG3 (indicated in


bold type in point 4) to access the
5 tube manual calibration program.

The FIP display is illustrated in


the figure.

Press the mA increase key


(indicated in bold type in point 6)
7 to access the calibration routine.

The FIP display is illustrated in


the figure.

(Rev. 3) 9-21 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the mA increase key to


select the tube to be calibrated
9 (tube 1, 2, 3).

Press the Reset key and select the


focus to calibrate (fine focus,
10 large focus).

Press the key M1 to proceed with


the calibration routine.
11
The FIP display is illustrated in
the figure. The first number on
the left indicates the minimum
radiography switch-on value, the
second indicates the maximum
12 switch-on value and the third
indicates the stand-by switch-on
time.
At the same time the fluoroscopy
DISPLAY displays the parameters
described in points 16 and 17.

The mA increase and decrease


keys change the minimum
filament switch-on value . If we
associate the lighting of one of
the leds relative to the AEC
dominants keys with the pressing
13 of the increase or decrease keys,
each time the key is pressed we
will obtain a variation as
described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

The kV increase and decrease


keys change the maximum
filament switch-on value . If we
associate the lighting of one of
the leds relative to the AEC
dominants keys with the pressing
14 of the increase or decrease keys,
each time the key is pressed we
will obtain a variation as
described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

GENIUS HF 9-22 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The sec. increase and decrease


keys change the filament stand-
by switch-on value . If we
associate the lighting of one of
the leds relative to the AEC
dominants keys with the pressing
15 of the increase or decrease keys,
each time the key is pressed we
will obtain a variation as
described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

The fluoroscopy kV DISPLAY


displays the multiplication
coefficient for the mA relative to
the calibration stored. The
number 1000 corresponds to a
multiplying value of 1, then the
16 basic calibration will be applied.
Numbers over 1000 will increase
the mA value, numbers less than
1000 will decrease the value.
Press the key 5R to decrease the
value and the fluoroscopy time
reset key to increase it.

The fluoroscopy mA DISPLAY


displays the data relative to the
activation of the automatic mA
correction function during each
radiography, for the coefficient
set at point 16. When the key
17 DSI1 is pressed, the datum 1 that
corresponds to the activation of
the correction is displayed. When
the key DS12 is pressed, the
datum 0 that corresponds to the
deactivation of the correction is
displayed.

Press the key M1 to proceed with


the calibration routine. If the
switch-on calibration values have
18 not been changed, the FIP will
appear as in point 22. If the data
have been changed, the routine
continues has described.

(Rev. 3) 9-23 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

The FIP display is illustrated in


the figure.

19

If the kV increase key is pressed,


the changes will be confirmed. A
buzzer will sound at the end of
20 the memorization phase
signalling that the operation has
been successfully completed.

If the RESET key is pressed, the


settings made during the above
21 phases will be cancelled and the
data stored in memory will not be
changed.

The FIP is displayed as in the


figure, AND SHOWING the first
manual calibration "STEP".

22

Press the key ORG1 and the key


mA to increase or decrease the
23 number of the displayed step.

Once the desired step is


displayed, press the key M1 to
24 enter into the calibration mode of
the set step.

The FIP displays the parameters


of:
• the kV with which to perform
the radiography
• the msec with which to
25 perform the radiography
• the safety mA beyond which
the breaker will be activated
together with the typical switch-
on and mA parameters obtained
at the end of the radiography.

GENIUS HF 9-24 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The ORG3 and Reset keys


increase or decrease the value of
the kV used to realize an
emission curve within a step. The
26 values given are 40, 60, 80, 100,
120, 140. 3 emissions with
different kV values are sufficient
for each step.

The mA increase and decrease


keys set the mA value that will
guarantee the intervention of the
27 breaker if during emission the
value reaches levels which are
dangerous to the tube.

The kV increase and decrease


keys set a different kV value with
respect to that proposed by the
28 step. This possibility is used
when the kV value is considered
too high.

The sec. increase and decrease


keys set a different radiography
29 time value with respect to that
proposed. The typical value
recommended is 100 msec.

Make a radiography when the


settings described above have
30 been completed.

If the radiography was not made


correctly or a safety device
intervened, the fluoroscopy min.
DISPLAY will display an alarm
31 message. Consult chapter 10.1
for a description of the displayed
alarm. Code 171 indicates that
the radiography was made
successfully.

Press the Reset key to reset the


alarm condition.
32

(Rev. 3) 9-25 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Change the step on which to


perform the calibration using the
ORG1 and mA increase keys.
Repeat the operations starting
33 from point 23. Once the
calibration of all the steps has
been completed proceed to point
34.

Press the sturdy build key to quit


the calibration routine.
34

Turn the switch located on the


front of the MPU card to the left
35 to return to normal operating
conditions.

GENIUS HF 9-26 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.5. Fluoroscopy calibration

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG3 (indicated in


bold type in point 4) to access the
5 fluoroscopy calibration program.

The FIP display is illustrated in


the figure.

Press the kV increase key


(indicated in bold type in point 6)
7 to access the fluoroscopy
calibration routine.

The FIP display is illustrated in


the figure.

(Rev. 3) 9-27 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the mA increase key to


select the tube to be calibrated
9 (tube 1, 2, 3).

Press the key M1 to proceed with


the calibration routine.
10

The fluoroscopy kV DISPLAY


displays the maximum kV value
reachable in fluoroscopy. Press
the key 5R to decrease the value
11 within the range of 90 ÷ 120 and
the fluoroscopy time reset key to
increase it.

The fluoroscopy mA DISPLAY


displays the maximum switch-on
value for fine focus. Press the key
DSI1 to decrease the value and
12 the key DSI2 to increase it.

Press the key M1 to proceed with


the calibration routine. A buzzer
informs the user that calibration
13 can start.

The FIP display is illustrated in


the figure.
14

Press the fluoroscopy pedal to


start emission.
15
The fluoroscopy kV DISPLAY
indicates the kV values set
automatically during the entire
16 calibration cycle. The values
which will be successively
displayed are 50, 80 and the
value set at point 11.

GENIUS HF 9-28 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The fluoroscopy mA DISPLAY


indicates the mA value set
automatically during the entire
17 calibration cycle. The value
increases automatically from 0.3
to 5 mA for all the kV steps.

Once the maximum kV value has


been reached with 5 mA,
emission is automatically stopped
and the FIP displays the message
18 "END". At this point release the
pedal.
If an alarm state occurs during
the calibration cycle, exposure
will be stopped.

The alarm code of the alarm


condition that occurred during
the calibration process will be
19 displayed on the fluoroscopy min.
DISPLAY. Turn to paragraph 10.1
for a description of the alarm
code.

Press the sturdy build key to quit


the calibration routine.
20

Turn the switch located on the


front of the MPU card to the left
21 to return to normal operating
conditions.

(Rev. 3) 9-29 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 9-30 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.6. Fluoroscopy ABC parameters configurations

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG1 (indicated in


bold type in point 4) to access the
5 workstation configuration
program.

Select the work station to which


to activated the fluoroscopy
6 function. The leds associated
÷ with the relative workstation keys
will flash.

Press the key M1 to proceed with


the configuration routine.
7

The FIP display is illustrated in


the figure.

Press the kV increase key


(indicated in bold type in the
9 figure in point 8) to display the
ABC data setting screen.

(Rev. 3) 9-31 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

The FIP is displayed as indicated


in the figure. The "T" data
indicate the adjustment
channels, while the "C" and "B"
10 parameters are the analogic
signal adjustment values. A
detailed explanation of these data
are given in paragraphs 4.10.2
and 7.4.

The keys ORG2 and ORG5 select


the mode of the adjustment
channels for continuous and
11 pulsed fluoroscopy. The typical
value to set is T = 1 for both
the continuous and pulsed
fluoroscopy mode.

The mA increase and decrease


keys change the value of dose
parity for continuous fluoroscopy
(see paragraphs 4.10.2 and 7.4).
The typical value to set is
C=+703. If we associate the
lighting of one of the leds relative
12 to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

The kV increase and decrease


keys change the value of the
proportional intervention window
for continuous fluoroscopy (see
paragraphs 4.10.2 and 7.4). The
typical value to set is B=1003.
If we associate the lighting of one
of the leds relative to the AEC
13 dominants keys with the pressing
of the increase or decrease keys,
each time the key is pressed we
will obtain a variation as
described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

GENIUS HF 9-32 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

The sec. increase and decrease


keys change the value of the dose
parity level for pulsed fluoroscopy
(see paragraphs 4.10.2 and 7.4).
The typical value to set is
C=+703. If we associate the
lighting of one of the leds relative
14 to the AEC dominants keys with
the pressing of the increase or
decrease keys, each time the key
is pressed we will obtain a
variation as described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

The ORG increase and decrease


keys change the value of the
proportional intervention window
for pulsed fluoroscopy (see
paragraphs 4.10.2 and 7.4). The
typical value to set is B=1003.
If we associate the lighting of one
of the leds relative to the AEC
15 dominants keys with the pressing
of the increase or decrease keys,
each time the key is pressed we
will obtain a variation as
described here below:
AEC left = units
AEC center = tens
AEC right = hundreds.

Press the key M1 to end the


configuration routine.
16

The FIP is displayed as indicated


in the figure, requesting the
confirmation or cancellation of
the settings made during the
17 above phases.

If the key ORG2 is pressed, the


settings will be confirmed. A
buzzer will sound at the end of
18 the memorization phase
signalling that the operation has
been successfully completed.

(Rev. 3) 9-33 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

If the key ORG4 is pressed, the


settings made during the above
19 phases will be cancelled and the
memorized data will not be
changed.

Press the sturdy build key to quit


the configuration routine.
20

Turn the switch located on the


front of the MPU card to the left
21 to return to normal operating
conditions.

GENIUS HF 9-34 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.7. AEC configuration

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG1 (indicated in


bold type in point 4) to access the
5 workstation configuration
program.

Select the workstation to which to


associate the measuring chamber
6 to be configured. The led
÷ associated to the workstation
keys will flash.

Press the keys relative to the


dominants to be activated. The
7 lighted led relative to the key
pressed confirms the selection
made.

Press the key M1 key to proceed


with the configuration routine.
8

(Rev. 3) 9-35 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

The FIP display is illustrated in


the figure.

Press the mA increase key


(indicated in bold type in point 9)
10 to display the AEC setting screen.

The FIP display is illustrated in


the figure and show the
parameters relative to the AEC
calibration.
11

With the mA increase and


decrease keys select the number
12 of the chamber to be configured.

Select the film-screen couple to


calibrate.
13

Change the amplification value


with the kV increase and
14 decrease keys. For further details
on these functions see paragraph
5.5.4.

Change the compensation value


for radiographies with values of
15 kV<80 with the ORG4 and Reset
keys. For further details on these
functions see paragraph 5.5.4.

Change the compensation value


for radiographies with values of
16 kV>80 with the ORG5 and M
keys. For further details on these
functions see paragraph 5.5.4.

GENIUS HF 9-36 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

Repeat the operations described


in points 14, 15, 16 for the other
17 2 film-screen couples.

The emission protection


threshold value expressed in mV
is displayed on the fluoroscopy
18 kV DISPLAY. For further details
on these functions see paragraph
5.5.4.

Decrease the protection threshold


value 1 unit at a time with the
19 key 5R. The value is increased
with the fluoroscopy time reset
key.

Decrease the protection threshold


value 100 units at a time with the
20 key DSI1. The value is increased
with the DS2 key.

Press the key M1 to end the


configuration routine.
21

The FIP is displayed as indicated


in the figure, requesting the
confirmation or cancellation of
the settings made during the
22 above phases.

If the key ORG2 is pressed, the


settings will be confirmed. A
buzzer will sound at the end of
23 the memorization phase
signalling that the operation has
been successfully completed.

If the key ORG4 is pressed, the


settings made during the above
24 phases will be cancelled and the
memorized data will not be
changed.

(Rev. 3) 9-37 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the sturdy build key to quit


the configuration routine.
25

Turn the switch located on the


front of the MPU card to the left
26 to return to normal operating
conditions.

GENIUS HF 9-38 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.8. Date and Time configuration

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG5 (indicated in


bold type in point 4) to access the
5 date and time configuration
program.

The FIP display is illustrated in


the figure.

Change the day with the mA


increase and decrease keys.
7

Change the month with the kV


increase and decrease keys.
8

Change the year with the sec.


increase and decrease keys.
9

(Rev. 3) 9-39 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Change the hour with the Reset


and M keys.
10

Change the minutes with the


ORG increase and decrease keys.
11

Press the key AUX to memorize


the setting.
12

Press the key M1 to cancel the


setting and return to point 4.
13

Press the sturdy build key to quit


the configuration routine.
14

Turn the switch located on the


front of the MPU card to the left
15 to return to normal operating
conditions.

GENIUS HF 9-40 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.9. Historical errors display

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG2 (indicated in


bold type in point 4) to access the
5 historical alarm display program.

The FIP display is illustrated in


the figure.
The following information is
displayed:
• the alarm code
6 • the date and time during
which the alarm was
generated.
Consult paragraph 10.2 for a
detail of the alarm codes.

Press the ORG1 to display the


less recent alarm in the list.
7

Press the ORG5 to display the


most recent alarm in the list.
8

(Rev. 3) 9-41 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the ORG2 key to scroll


backward through the list.
9

Press the ORG4 key to scroll


forward through the list.
10

Press the key AUX to return to


point 4.
11

Press the sturdy build key to quit


the display routine.
12

Turn the switch located on the


front of the MPU card to the left
13 to return to normal operating
conditions.

GENIUS HF 9-42 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.10. Calibration copy

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG3 (indicated in


bold type in point 4) to access the
5 copy program.

The FIP display is illustrated in


the figure.

Press the Reset key proceed with


the calibration copy routine.
7

The FIP display is illustrated in


the figure.
Source indicates the tube from
8 which the calibration is taken
and destination is the tube where
the calibration will be copied.

Press the kV increase key to


change the number of the tube
9 from which to take the
calibration.

(Rev. 3) 9-43 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the Reset key to change the


calibration destination tube.
10

Press the M1 key to start the


copy of the calibration. A buzzer
informs the user that the copy
11 operation has been completed
successfully.

Press the sturdy build key to quit


the configuration routine.
12

Turn the switch located on the


front of the MPU card to the left
13 to return to normal operating
conditions.

GENIUS HF 9-44 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

9.11. Changing the kV transfer tables

Turn the switch located on the


front part of the MPU card to the
1 right (access to service routine).

Press one of the 7 vertical


anatomical selection keys.
2

Press the child key to enter the


configuration routine.
3

The FIP display is illustrated in


the figure.

Press the key ORG1 (indicated in


bold type in point 4) to access the
5 workstation configuration
program.

Select the workstation on which


to activate the 0-point technique.
6 The leds associated to the keys
÷ relative to the workstations will
flash.

The FIP display is illustrated in


the figure.

Press the sec. increase key


(indicated in bold type in point 7)
to activate the 0-point technique.
8 The 00 value indicates that the
function is deactivated. Values
other than 00 indicate that the
transfer is activated.

(Rev. 3) 9-45 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the key M1 to proceed with


the program.
9

If the 0-point technique has been


activated, the message indicated
10 in the figure is illustrated on the
display.

Select the transfer table to be


configured.
11

The value of the fluoroscopy kV


which will be used as reference
will be displayed on the kV
display.
12 At the same the FIP will display
the kV, mA and sec. values (see
point 14) that will be
automatically set for that
particular fluoroscopy value.

Change the reference kV in steps


of 5 kV within the range 40 to
13 115 kV with the 5R and
fluoroscopy time reset keys.

As shown in the figure, the FIP


displays the kV, mA and time
values which will be used for the
14 radiography for each step of the
fluoroscopy kV (see point 12).

Change the radiography mA value


with the mA increase and
15 decrease keys.

Change the radiography kV value


with the kV increase and
16 decrease keys.

Change the radiography time


value with the sec. increase and
17 decrease keys.

GENIUS HF 9-46 (Rev. 3)


SERVICE MANUAL
Synoptic tables for equipment set-up

Press the M1 key to memorize the


set table. The program will return
18 to point 10 ready for the insertion
of a new table.

Press the key AUX to cancel the


settings and return to point 6.
19

Press the sturdy build key to quit


the configuration routine.
20

Turn the switch located on the


front of the MPU card to the left
21 to return to normal operating
conditions.

(Rev. 3) 9-47 GENIUS HF


SERVICE MANUAL
Synoptic tables for equipment set-up

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 9-48 (Rev. 3)


SERVICE MANUAL
Alarm messages

10. ALARM MESSAGES

10.1. Alarm messages in calibration

Alarm Description

More than 20 seconds have elapsed from the request for


129 preparation without radiography being made

During preparation the microprocessor requests the heating of the


131 filament (olf) or (osf) and checks that the reply from the Filament
Inverter Driver is correct (ilf) or (isf).
If the current value obtained is different from the value requested,
the microprocessor halts the preparation and enters into alarm
state.
Check:
• the filament heating signal output from the microprocessor
(olf) or (osf). If incorrect, check the route:
PCB 9999102700 MPU è PCB 9999102600 inverter filament
driver
• for malfunction in the PCB 9999102600 (inverter filament
driver)
• for malfunction in the transmission of the feedback signal (ilf)
or (isf) PCB 9999102600 inverter filament driver è PCB
9999102700 MPU.

During preparation [INS] (or [IHS] if starter HS and high speed is


133 selected) has not been activated by the ANODE's starter PCB
9999103200, PCB 9999103300 or PCB 5810 dual speed starter.
Check:
• problem with the starter
• if the anode rotates check the route in [INS] (or [IHS] if high
speed): PCB 9999103200 or PCB 9999103300 or PCB
9999103400 (starter) è PCB 9999102700 MPU.

Signals the failed closure of the main contactor. The


134 microprocessor checks the [IPW] to verify closure, i.e. the
presence of 540 Vdc. If [IPW] does not activate check:
• that the mains line contactor located on the front of block 4
(inverter control board and line supply interfacing) is closed
and that there is voltage on the DC-RAIL (capacitors on the
back of block 4) If this is correct check the [IPW] signal and
its route: PCB 5811 snubber and service è PCB 9999102500
CCB
• if [IPW] is not correct, the CCB is not correctly supplied, so
check: PCB 9999102500 CCB è PCB 9999102700 MPU.

(Rev. 3) 10-1 GENIUS HF


SERVICE MANUAL
Alarm messages

Alarm Description

During preparation, the inverter filament card does not make the
135 [ICF] signal active. This means that the selected filament has not
exceeded the 2.5 amps for heating. Therefore:
• there is a problem in the inverter filament driver PCB
9999102600
• there is a problem in the [ICF] route through to the
microprocessor: PCB 9999102600 (inv. fil. driver) è PCB
9999102700 MPU.

During radiography, the microprocessor sees the [IKV] signal


136 become active. This means that the general safety mode is
activated. DO NOT RESET the alarm and check that the red led is
lit on PCB 9999102500 CCB.
This condition means that:
• 166 kV have been exceeded
• the current has exceeded 1300 mA
• the current on bridge IGBT has exceeded 850 mA
• the clock on the inverter control card PCB 9999102500 CCB
does not function correctly.

The [IUB] signal is not active in preparation. This means that the
137 IGBT bridge is not balanced. Check the following:
• the snubbers
• the IGBT
• the circuit that performs this check
• a problem in the [IUB] route through to the microprocessor
PCB 5800 IGBT driver è PCB 9999102500 è PCB
9999102700 MPU.

The [IST-1], [IST-2] or [IST-3], in relation to the selected tube, is


138 not active. Check:
• the commutator switch magnet is not charged
• if charged the [IST-1] or [IST-2] signal does not arrive in the
microprocessor, check the course: GTB 400 è PCB
9999102700 MPU.

During radiography the safety door input [IDS] has been opened.
139 PCB 9999102500 CCB è PCB 9999102700 MPU.

If during radiography there is a fall of the kV values and this is


140 detected by the microprocessor, the [IRM] signal becomes inactive.
This is usually caused by a discharge of high voltage. Check:
• whether there has been a discharge of high voltage
• that the [IRM] is transmitted correctly PCB 9999102500 CCB
è 9999102700 MPU.

GENIUS HF 10-2 (Rev. 3)


SERVICE MANUAL
Alarm messages

Alarm Description

The real kV value during calibration radiography does not


150 correspond to the value set at the console: Check:
• the voltage value on the DC-BAR
• whether the breaker value, shown on the radiography mA
display has been set correctly.

NOTE:
Every direct digital signal or digital signal from the inverter has a
led that indicates its status.
Every direct analogic signal or signal from the inverter has a test
point so that it may be monitored by an oscilloscope.

The nicroprocessor did not interrupt radiography. The


151 microprocessor checks the [IRM] signal to see if radiography has
been completed. If this signal is interrupted before an established
maximum period of time, this means that radiography has been
halted. If this does not occur this means that there is still high
voltage and current circulating inside the tube. Check:
• that the signal stop exposure [OHV] is active at CCB, check
the course: PCB 9999102700 MPU è PCB 9999102500 CCB
• the course of [IRM] signal, high voltage on: PCB 9999102500
CCB è PCB 9999102700 MPU.

NOTE:
If the load is very low (e.g. the filament is not heated), high voltage
and mA persist for a long time and so activate this safety mode.

The kV-mA couples are not sufficient for the duty-cycle


170 calculation.
Other couples must be input.

RADIOGRAPHY COMPLETED CORRECTLY.


171

(Rev. 3) 10-3 GENIUS HF


SERVICE MANUAL
Alarm messages

10.2. Alarm messages in anatomical technique

The alarm messages inform the user of a function anomaly.


After the alarm has been recognized and the necessary actions taken, it
must be reset in order to continue.

Alarm Description

128 The anatomical technique selected does not contain any


NOT EXISTANT data.

129 This alarm occurs when 20 seconds have elapsed from


TIME-OUT PREP the request for preparation and the operator has not
requested exposure although he retains preparation
active.
In seriograph or digital mode, this alarm may also be
caused by the external system that does not proceed with
the sequence.

130 The tomography time input are wrong.


TOMO PREP SAF

131 This alarm is generated when the filament activation


HEAT RADIO? value calculated by the microprocessor on the base of the
kV-mA data, exceeds the limits given by the
manufacturer. Usually, this happens when high mA
values are set with low kV values. Reduce the mA value
or increase the kV value to overcome this problem.

132 The I.M.A. signal AEC ready is not active.


AEC NOT READY

133 During preparation [INS] (or [IHS] if starter HS and high


Anode safety speed is selected) the STARTER has not been activated
(PCB 9999103200, PCB 9999103300 or PCB 5810 dual
speed starter).
Check that:
• the starter works correctly
• the anode rotates
• the route of [INS] (or [IHS] if high speed) is correct:
PCB 9999103200 or PCB 9999103300 or PCB
9999103400 (starter) è PCB 9999102700 MPU.

134 The principal inverter contactor was not activated and


Sicurez. PWS therefore there is no voltage on the DC-RAIL.
Check:
• the contactor and its closing command [OON]
• the reply towards the microprocessor [IPW]
• all the components of block 4 that enable the
DC-RAIL to reach 540 V.

GENIUS HF 10-4 (Rev. 3)


SERVICE MANUAL
Alarm messages

Alarm Description

135 The selected filament is not lit or is only partially lit. The
Dig. Fil safety signal [ICF] is not active. Check:
• the selected filament or the PCB 9999102600 inverter
filament driver
• the signal route: PCB 9999102600 inverter filament
driver è PCB 9999102700 MPU.

136 Different conditions may generate this alarm.


General safety Before resetting the alarm, check on the PCB
9999102500 CCB which of the red leds is lit:
• led20 [CLK]: malfunction in the CCB clock system
• led14 [IPO]: the primary circuit current has exceeded
the maximum limit
• led13 [mAx]: the current in the tube has exceeded the
maximum limit
• led19 [kVx]: the kV values in the tube have exceeded
the maximum limit
• led18 [Pwx]: no power supply at the IGBT gates
• led17 [Scx]: a short circuit occurred during X-ray
• led15: the positive and negative branches of the high
voltage were not balanced during X-ray.

To cut out the alarms that generate the [IBO] signal apply
the appropriate jumpers at the following points:
• JP11 = > 166 kV control
• JP12 = > 1150 kV control
• JP14 = > 950 Amp in jumper control
• JP15 = short circuit control
• JP16 = supply on PCB 5800 control.

137 The inverter is in a state of alarm. Reset and try again


Inverter safety with a new preparation. If the problem persists, check the
[IUB] signal, relative to the balancing of the jumper.
Check:
• the inverter and in particular the state of the IGBT
and the snubber system
• the signal route: è PCB 5800 IGBT driver è PCB
9999102500 CCB è PCB 9999102700 MPU.

139 The door safety input has been opened.


Door safety This safety mode interrupts the operations in course.

(Rev. 3) 10-5 GENIUS HF


SERVICE MANUAL
Alarm messages

Alarm Description

140 During the first few hundred microseconds of the first


Low kv safety part of radiography the high voltage has not exceeded
18 kV and so radiography is interrupted by the
microprocessor.
The [IRM] signal not active.
Check:
• the start/stop signal to the inverter [OHV] and its
route:
PCB 9999102700 MPU è PCB 9999102500 CCB
• The [IRM] signal route: PCB 9999102500 CCB è
PCB 9999102700 MPU.

141 The incident radiation on the chamber is lower than the


AEC dose safety lowest limit value input during the exposure meter
calibration phase. Check that the radiographic
parameters are correct and the correct selection of the
dominant.

142 Radiography was not halted by A.E.C. but by the


AEC safety release microprocessor's time-out safety device or the operator
released exposure before the A.E.C interrupted the
exposure. Check the exposure meter circuit.

143 AEC did not stop the exposure, thus reaching the time
AEC TIME-OUT limit.

144 Preparation command has been released during


TOMO REL SAF tomography.

145 Radiography command has been released during


TOMO REL SAF tomography.

146 Tomography did not stop X-ray before the time limit.
TOMO TIME-OUT

147 kV value set for tomography is too high.


MAX TOMO SAF

148 kV value set for tomography is too low.


MIN TOMO SAF

GENIUS HF 10-6 (Rev. 3)


SERVICE MANUAL
Alarm messages

Alarm Description

149 During fluoroscopy, the microprocessor detected a


Fluoro. mA safety difference between the real and the set mA that cannot be
corrected.
Check:
• the measuring system (IFC). Check the mA regulation
graph to locate possible error points
• that the terminals of the H.V. cables are clamped
correctly, clean and with an adequate layer of silicon,
and the pins provide a good contact
• noise on the current signal (ICF)

150 The real value of kV set during radiography adjustment is


kV RADIO SAF not corresponding to the relevant value set on the
console.

151 The microprocessor has stopped the radiography. The


RAY OFF SAFETY microprocessor senses a signal to understand if
radiography is terminated. If this signal is interrupted
before a given time, it means that the radiography is
interrupted, otherwise it means that high voltage and
current are still present in the X-ray tube.

152 State of alarm generated by an accumulation of the


Thermal overload quantity of heat in the anode that exceeds the limit
established by the manufacturer of the tube.
Wait a few minutes to enable the heat to disperse and
then continue.

153 Alarm generated when the filament heating value in


Radiography fluoroscopy exceeds the maximum limit established by
heat.? the manufacturer.

154 Fluoro kV value is out of range.


kV FLUORO SAF

155 During fluoroscopy, the microprocessor detected a


Fluoro mA safety difference between the real and the set mA that cannot be
corrected.
Check:
• the measuring system (IFC). Check the mA regulation
graph to locate possible error points
• that the terminals of the H.V. cables are clamped
correctly, clean and with an adequate layer of silicon,
and the pins provide a good contact
• noise on the current signal (ICF).

(Rev. 3) 10-7 GENIUS HF


SERVICE MANUAL
Alarm messages

Alarm Description

156 This alarm is generated only during the fluoroscopy


Fluoroscopy time phase.
It indicates that 5 minutes of fluoroscopy irradiation have
elapsed. Reset to proceed.

157 The thermal safety of X-ray tube 1 has been triggered.


TU-1 THERM SAF

158 The thermal safety of X-ray tube 2 has been triggered.


TU-2 THERM SAF

159 The thermal safety of X-ray tube 3 has been triggered.


TU-2 THERM SAF

162 The mAs value exceeds the limit threshold.


mAs SAFETY

GENIUS HF 10-8 (Rev. 3)


SERVICE MANUAL
Alarm messages

10.3. Alarms and equipment status of PCB Supervisor


(PCB 99991030)

10.3.1. Status LED’s

• ACTIVE SENSING LED (yellow)


Blinking (frequency 2Hz) indicates the status of waiting commands
from the operator. If LED stops blinking, the generator must be
switched off and then switched on again (put switch Si1 in OFF
position, wait for 10 seconds then Si1 in ON position).

• GENERATOR ON LED (yellow)


Lit when generator is ON, receiving data, X-ray emission in progress.
This LED becomes off after pressing key ‘O’ on the command console,
after 1 hour of X-ray inactivity with jumper JP8 removed, or forced
from the generator (check the activation of the alarm LED’s).

• X-RAY ON LED (yellow)


Lit when generator is in X-ray on status.

• RESERVED #1 LED (yellow)

• RESERVED #2 LED (yellow)

(Rev. 3) 10-9 GENIUS HF


SERVICE MANUAL
Alarm messages

10.3.2. Alarm LED’s

• FILAMENT ERROR LED (red)


– Origin of the control: FILAMENT INVERTER DRIVER board,
OBF signal
– Status: NORMALLY OFF
– Disabling of the control: jumper JP12
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates the presence of a problem on the Filament inverter
driver board (problem detected during the last activation session).
– BLINKING IN GENERATOR ON STATUS
Indicates that a problem on the Inverter filament driver has been
detected during the previous activation session.

• WATCH DOG ERROR LED (red)


– Origin of the control: MASTER PROCESSING UNIT board,
OSM signal
– Status: NORMALLY OFF
– Disabling of the control: jumper JP13
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates the presence of a problem on the MPUA board detected
during the last activation session.
– BLINKING IN GENERATOR ON STATUS
Indicates that a problem on the MPUA board has been detected
during the previous activation session.

GENIUS HF 10-10 (Rev. 3)


SERVICE MANUAL
Alarm messages

• MPU SWITCH OFF REQUEST LED (red)


– Origin of the control: MASTER PROCESSING UNIT board,
OAB signal
– Status: NORMALLY OFF
– Disabling of the control: jumper JP13
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates that the MPU board has forced the switching off,
detected during the last activation session.
– BLINKING IN GENERATOR ON STATUS
Indicates that the MPU board has forced the switching off,
detected during the previous activation session.

• POWER LINE TOO LOW LED (red)


– Origin of the control: SAFETY SUPERVISOR board, LINE
signal
– Status: NORMALLY OFF
– Disabling of the control: not applicable
– LIT IN GENERATOR OFF STATUS
Indicates that line voltage value sensed during the last activation
session was dangerously low.
– BLINKING IN GENERATOR ON STATUS
indicates that the line voltage value sensed during the previous
activation session was dangerously low.

• BREAKER INTERNAL ERROR LED (red)


– Origin of the control: SAFETY SUPERVISOR board, BRKDR
signals
– Status: NORMALLY OFF
– Disabling of the control: jumper JP14
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates a malfunction of the breaker reference generating
circuit during the last activation session of the generator.
– BLINKING IN GENERATOR ON STATUS
Indicates a malfunction of the breaker reference generating
circuit during the previous activation session of the generator.

(Rev. 3) 10-11 GENIUS HF


SERVICE MANUAL
Alarm messages

• HIGH VOLTAGE CURRENT TOO HIGH LED (red)


– Origin of the control: CONVERTER CONTROL BOARD,
MASTER PROCESSING UNIT boards,
icl, ich, OTC, OTD, ICK signals
– Status: NORMALLY OFF
– Disabling of the control: jumper JP14
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates that a too high current value of the high voltage has
been detected during the last exposure, compared to the value
driven by the MPUA board.
– BLINKING IN GENERATOR ON STATUS
Indicates that a too high current value of the high voltage has
been detected during the last exposure, compared to the value
driven by the MPUA board during the previous activation session.

• HIGH VOLTAGE CURRENT TAIL TOO HIGH LED (red)


– Origin of the control: CONVERTER CONTROL BOARD,
MASTER PROCESSING UNIT boards,
icl, ich, OTC, OTD, ICK signals
– Status: NORMALLY OFF
– Disabling of the control: jumper JP14
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates that a current value of the high voltage higher than
expected has been detected following the de-activation of X-rays.
– BLINKING IN GENERATOR ON STATUS
Indicates that a current value of the high voltage higher than the
one expected during the previous activation session has been
detected following the de-activation of X-rays.

GENIUS HF 10-12 (Rev. 3)


SERVICE MANUAL
Alarm messages

• LED X-RAY TIME TOO LONG (rosso)


– Origine del controllo: scheda MASTER PROCESSING UNIT
segnali ORS, OFT
– Stato: NORMALMENTE SPENTO
– Disabilitazione controllo: jumper JP15
– BLINKING IN GENERATOR OFF STATUS
Indicates that this control has been disabled by means of the
jumper.
– LIT IN GENERATOR OFF STATUS
Indicates that, during the last exposure, an exposure time longer
than the one driven by the MPUA board has been detected.
– BLINKING IN GENERATOR ON STATUS
Indicates that, during the last exposure, an exposure time longer
than the one driven by the MPUA board during the previous
activation session has been detected.

(Rev. 3) 10-13 GENIUS HF


SERVICE MANUAL
Alarm messages

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 10-14 (Rev. 3)


SERVICE MANUAL
Maintenance

11. MAINTENANCE
In order to ensure the safety of the patient, the operator and of other
parties that could come into contact with the equipment, we recommend
an inspection of the machine to be made at least every 12 months.

Two types of tests are required:


• maintenance performed by the operator
• maintenance performed by the servicing department, i.e. by
specialized personnel having the appropriate instruments.

VILLA SISTEMI MEDICALI decline all responsibility for malfunctions


of the equipment if:
• the installation, calibrations and repairs are not performed by
qualified personnel or personnel not authorized by the manufacturer
• original spare parts have not been used as listed in the Spare Parts
List (chapter 13)
• the electrical system in the room where the equipment is lodged does
not comply with the provisions of chapters 2, 3 and 4 of this manual
• incorrect use of the equipment not in conformance with the
instructions of this Manual and in the User Manual.

(Rev. 3) 11-1 GENIUS HF


SERVICE MANUAL
Maintenance

11.1. Operator maintenance

11.1.1. Cleaning

WARNING:
1 The appliance must be switched off before any cleaning action is
undertaken.

The equipment must be cleaned using cotton cloths and non-abrasive


detergents. Do not use organic solvents or agents containing solvents of
any kind.
Do not use sprays. Under no circumstances must detergent penetrate
inside the equipment.
To disinfect the surface, water solutions based on aldehyde are
recommended. We recommend especially aggressive agents based on
alcohol not to be used.

11.1.2. Daily tests

WARNING:
1 Protective clothing must be worn when performing any tests that require
the emission of radiation. Adequate distance must be kept from the
source and, if necessary, the personal dosimeter must be used.

Before starting the day's work, a visual inspection of the machine must
be made, and any damage reported to VILLA SISTEMI MEDICALI.
During the tests check that the yellow fluoroscopy and radiography rays
indicator lights up.

11.1.3. Weekly tests

Check that the cables and electrical connections are not damaged in any
way.
No unusual noise should issue from the inverter.

GENIUS HF 11-2 (Rev. 3)


SERVICE MANUAL
Maintenance

11.1.4. Monthly tests

Check that the AEC and ABC devices function as follows.

A.E.C. test
1. Completely close the jaws of the collimator.
2. Activate the AEC device.
3. Make one exposure. The emission must be blocked immediately.
4. Open the collimator and carefully place an object of medium density
on the patient resting surface (in aluminum or iron with small
thickness).
5. Make one exposure. In this case the radiation indicator must remain
ON for a time depending upon the density of the object and the set
kV value.

A.B.C. test
1. Completely close the jaws of the collimator.
2. Make one fluoroscopy exposure with the A.B.C. device inserted. The
yellow fluoroscopy indicator must remain lit during the entire time of
exposure and the dose, given as a kV-mA value on the display, must
rise to the maximum value.
3. Stop fluoroscopy.
4. Open the collimator and make a new fluoroscopy. In this case, the
radiation indicator must remain ON and the dose must fall to the
lowest level.

(Rev. 3) 11-3 GENIUS HF


SERVICE MANUAL
Maintenance

11.2. Technical servicing maintenance

The following program must be followed for a correct maintenance


activity of the appliance:

Frequency of the
Descriptions of the activities
inspections

Every 6 months or every The correctness of the radiography


time a card with a parameters must be verified:
regulation function and/or • the peak value of the kV
control of the radiography • the average value of the mA during
parameters is substituted. radiography
• the accuracy of the mA values.
These verifications can be made by comparing
the value set with that read on the display on
the MPU or using non-invasive reading
systems.
In case one or more of the above parameters
are not within the established limits, perform
a new calibration (for the mAs) or a regulation
of the trimmers associated with the relative
signals on the PCB 9999102500 (paragraph
12.1).

Verify the good working order of the exposure


meter, both the correctness of the times and
the intervention of the safety function in the
case of "low dose".

Perform all tests required by the laws of the


country where the equipment is installed.

Every 12 months. Verification of the operations and clean state


of the equipment as described in the
procedures of this paragraph.

Every 6 years or every time Substitution of the RAM buffer battery located
it is deemed necessary. in the RAM socket.
This operation requires the loading of the
data in RAM from EEprom as previously
described.

GENIUS HF 11-4 (Rev. 3)


SERVICE MANUAL
Maintenance

The following procedures should be followed to ensure a correct


maintenance:
• a check to ensure the cables, insulation and covers preventing
contact with live parts are in good condition and not damaged in any
way
• verification of the correct functioning of the machine (test in previous
paragraph)
• measurement of current leakage to ground
• measurement of the resistance between the ground and all metallic
parts of the equipment
• removal of dust from all electronic and non-electronic parts
contained in the power rack
• substitution of the insulating silicone of the high voltage cable
terminals
• check and cleaning of the relay contacts and contactors
• check of the oil level in the high voltage transformer
• check of the fuses
• check of the alarm states in the form of lighted leds on the electronic
cards
• check of the radiography parameters.

We recommend these tests to be performed at least once a year.


More frequent tests will make the use of the generator safer.

(Rev. 3) 11-5 GENIUS HF


SERVICE MANUAL
Maintenance

11.2.1. Condition of the cables and connections

Switch off the generator. Open the power rack and make a visual
inspection of the connections:
• cables: search for cuts and loss of insulation. Check that the cable
clamps are well fixed.
• electronic cards: check that the connectors on the printed circuits
are connected correctly.

Perform these tests on the command console also.

11.2.2. Current leakage test

Use the instruments appropriate to this test.


Refer to the test table provided in standards I.E.C. 601-2-7.

11.2.3. Dust removal

Disconnect the generator from the mains.


Open the power rack. Using a brush clean any dusty surfaces.

WARNING:
1 DO NOT USE LIQUID DETERGENTS ON PARTS THAT ARE LIVE WHEN
THE EQUIPMENT IS SWITCHED ON.

Remove any dust from the PCBs using a dry brush, paying special
attention on those that have horizontal surfaces.
Repeat these operations for the command console.

GENIUS HF 11-6 (Rev. 3)


SERVICE MANUAL
Maintenance

11.2.4. Cleaning the high voltage cable terminals

Switch off the high voltage generator.


Remove the cable terminals from the generator and the tube. Discharge
the poles short-circuiting the pins one at a time with the ground.
Clean the high voltage cable terminals, widen the pins, test the
insulation of the positive, negative and ground poles.
Cover the cable terminals with insulating silicone or vaseline, insert the
terminals and fix to the support.
Repeat the operation for all 4 cable terminals.

11.2.5. Cleaning the contacts

Disconnect the generator and all connecting equipment from the mains.
Open the power rack and locate the relay that need cleaning (most of the
relays are on the G.T.B.).
Remove the protective cover, clean the contacts with an appropriate
spray product.
Replace the protective cover.
Disassemble the front part of the contactor, clean the contacts or change,
if necessary. Reassemble the contactor. Close the rack.

11.2.6. Oil check

Disconnect the generator from the mains.


Open the power rack and check the oil level in the high voltage
transformer as described in paragraph 3.1.2.
The oil level must be 1.0 / 1.5 cm from the inside of the generator lid.
Visually inspect the bottom of the rack to verify that there is no oil.
If there is oil, check the transformer to locate the point of origin.
If a leak is located in a critical point change the damaged component.
Inform the manufacturer.
Clean the parts stained with oil and, if necessary, fill up with oil.
Close the rack.

(Rev. 3) 11-7 GENIUS HF


SERVICE MANUAL
Maintenance

11.2.7. Alarms control

Disconnect the appliance from the mains.


Open the power rack.

11.2.7.1. Check the inverter cards CCB and QUAD DRIVER

1. Switch off the generator.


Great care must be taken because there are high capacity
electrolytic capacitors in the back part of the rack.
Avoid contact with the same when the generator is switched on with
the side panels removed, keep a suitable distance between the
capacitor terminals and parts of the body or any non-insulated item
being handled.
Before touching the terminals, ensure that the power is off and that
there is no voltage inside the capacitors.
2. Open the GTB panel. Insert jumper JP8 located on the PCB
Converter Control Board 9999102500 (paragraph 12.1). The red led
DL19 shall light up, the alarm signal "GENERAL SAFETY" shall
appear on the console's FIP and the alarm signal shall sound.
3. Reset the alarm via the relevant key. Insert jumper JP9, the red led
DL14 shall light up, the alarm signal "GENERAL SAFETY" shall
appear on the console's FIP and the alarm signal shall sound.
4. Reset the alarm via the relevant key. Insert jumper JP7, the red led
DL13 shall light up, the alarm signal "GENERAL SAFETY" shall
appear on the console's FIP and the alarm signal shall sound.
5. Reset the alarm via the relevant key.
6. Switch off the generator.

GENIUS HF 11-8 (Rev. 3)


SERVICE MANUAL
Maintenance

11.2.7.2. Inverter power supply check

1. Remove the LB contactor command located behind the GTB panel


and disconnect the Si3 switch located on the front omega element
holder.
2. Switch on the generator. The beep shall sound and the alarm
message "SAFETY PWS" shall appear on the FIP.
3. Reset the alarm, switch off the appliance, close the Si3 switch, switch
on the appliance.
4. Make an exposure at 80 kV - 100 mA - 0.100 s.
No alarm must occur during this time.

11.2.7.3. Inverter LED check

Switch on the appliance and check that the following leds are in the
correct state:
• PCB 5800 IGBT Quad Driver (paragraph 12.1):
lit: DL7 - DL8 - DL9
off: DL4 - DL6 - DL3 - DL5 - DL1 - DL2
• PCB 9999102500 CCB (paragraph 12.1):
lit: DL20 and all the green leds
off: all the red leds
Radiography/Fluoroscopy: The yellow leds LD1 - LD8 lit during
emission.

11.2.7.4. Power electronic leds check

Switch on the generator and check that the following leds are in the
correct state:
• PCB 9999102600 inverter filament driver (paragraph 12.1):
lit: DL3 - DL7 - DL8
off: DL1 - DL2
• PCB 9999103200 normal speed starter (paragraph 12.1):
Anode start-up: DL1 - DL2 lit during anode start-up.

(Rev. 3) 11-9 GENIUS HF


SERVICE MANUAL
Maintenance

11.2.8. Fuse check

Check all the fuses with an ohmmeter with the appliance off.

11.2.9. Radiography parameters check

The control of the correspondence of the set and real radiography


parameters is based on the size of the current and voltage sent to the
tube.
The current value (mA) depends upon the type of tube and so upon the
HW calibration of the cards and upon the SW calibration of the tube. The
mA value can be calibrated.
The peak kV value is independent of the type of tube installed and is a
characteristic of the generator. This value can be changed.
Another fundamental value relative to exposure is time.
With this appliance, time does not require calibration as it is calculated
by an electronic quartz based circuit.
The mA values are a consequence of the mA calibration.
The regulation of the HW circuits for the measurement of the kVp and
mA values are performed at the manufacturer's during the final test
phase.
The reference standard used for this calibration is an invasive
instrument produced by RadCal Corp. called Dynalyzer.
This instrument measures kVp, mA and time.
We therefore recommend this type of instrument or similar when a H.W.
calibration is necessary.

GENIUS HF 11-10 (Rev. 3)


SERVICE MANUAL
Maintenance

11.2.9.1. Parameters accuracy tests

Follow this procedure.


If a measuring instrument as indicated is not available, monitor the
values by connecting an oscilloscope, as described:
• kV measurement
Connect an oscilloscope to test point TP5 of the PCB 9999102500
(paragraph 12.1)
• Radiography mA measurement
Connect an oscilloscope to test point TP2 or TP9 as required.

It will be possible to verify the value of the parameters but it must be


remembered that calibration is based on a standard instrument.
Set the parameters indicated in the following table and verify whether the
results lie within the given limits.

Small focus

Set parameters Size Limits

60 kV peak kV (kVpp) 57 ÷ 63
25 mA ripple kV (kVpp) 3
100 msec ① mean mA (mA) 22 ÷ 28
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


25 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 22 ÷ 28
time (msec) 98 ÷ 102

60 kV peak kV (kVpp) 57 ÷ 63
80 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 72 ÷ 88
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


80 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 72 ÷ 88
time (msec) 98 ÷ 102

60 kV peak kV (kVpp) 57 ÷ 63
160 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 135 ÷ 165
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


160 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 144 ÷ 176
time (msec) 98 ÷ 102

① Ignore any spikes due to load of the high voltage cables and capacitors.

(Rev. 3) 11-11 GENIUS HF


SERVICE MANUAL
Maintenance

Large focus

Set parameters Size Limits

60 kV peak kV (kVpp) 57 ÷ 63
200 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 180 ÷ 220
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


200 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 180 ÷ 220
time (msec) 98 ÷ 102

60 kV peak kV (kVpp) 57 ÷ 63
400 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 360 ÷ 440
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


400 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 360 ÷ 440
time (msec) 98 ÷ 102

60 kV peak kV (kVpp) 57 ÷ 63
630 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 567 ÷ 693
time (msec) 98 ÷ 102

100 kV peak kV (kVpp) 95 ÷ 105


630 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 567 ÷ 693
time (msec) 98 ÷ 102

② 80 kV peak kV (kVpp) 76 ÷ 84
800 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 720 ÷ 880
time (msec) 98 ÷ 102

② 60 kV peak kV (kVpp) 57 ÷ 63
1000 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 900 ÷ 1100
time (msec) 98 ÷ 102

② 150 kV peak kV (kVpp) 140 ÷ 150


200 mA ripple kV (kVpp) 3
100 msec mean mA (mA) 180 ÷ 220
time (msec) 98 ÷ 102

② These radiographs are possible only with high power tubes or generators.

GENIUS HF 11-12 (Rev. 3)


SERVICE MANUAL
Maintenance

Fluoroscopic mode

Set parameters Size Limits

40 kV peak kV (kVpp) 36 ÷ 44
0.5 mA mean mA (mA) 0.4 ÷ 0.6

80 kV peak kV (kVpp) 76 ÷ 84
2.25 mA mean mA (mA) 2 ÷ 2.5

120 kV peak kV (kVpp) 114 ÷ 126


4 mA mean mA (mA) 3.6 ÷ 4.4

(Rev. 3) 11-13 GENIUS HF


SERVICE MANUAL
Maintenance

11.2.9.2. Parameters repeatability test

Perform 10 exposures with the values indicated.


The three values (kV – mA - msec) must lie within the limits indicated for
each exposure.

Set parameters Size Limits

80 kV peak kV (kVpp) 76 ÷ 84
100 mA mean mA (mA) 90 ÷ 110
100 msec time (msec) 98 ÷ 102

• The error detected for the verified parameters must not exceed 5% for
the kV and 10% for the mA
• If significant differences are obtained for the kV values, contact VILLA
SISTEMI MEDICALI before proceeding with the calibration
• If there are significant differences in the mA values, before
commencing a new calibration check to see whether the PCB
9999102600 is correctly supplied with power, the terminal pins of
the high voltage cables are clean and wide and that the tube
filaments are heated
• kV accuracy.
If the kV value is not within the limits given on the table, appropriate
calibration must be made on the gain and feedback potentiometers
on the PCB 9999102500 Converter Control Board (paragraph 12.1).
Regulate the issued kV value via potentiometer R38.
Regulate the kV value read on the MPU and presented on the display
inside the rack via potentiometer R251
• mA accuracy.
If the mA value is not within the limits given in the table, the focus
calibration must be repeated.
This is valid if only one focus is out of limits.
If the problem lies with both focuses, calibration must be made on
the gain potentiometers on the PCB 9999102500 Converter Control
Board (paragraph 12.1).
There are three channels for measuring the mA values on this card:
– ICL fluoroscopy channel 10 V = 10 mA (potentiometer R32)
– ICM low mA values radiography channel with 10 V = 150 mA
scale (potentiometer R69)
– ICH high mA values radiography channel 10 V = 1500 mA scale
(potentiometer R54)

GENIUS HF 11-14 (Rev. 3)


SERVICE MANUAL
Maintenance

• Radiography time accuracy.


The tests must confirm the accuracy of the exposure time
measurements.
This parameter is controlled by the quartz on the MPU.
• mAs accuracy.
This depends upon the time and the value of the mA. If both these
parameters are correct, product is correct.

It must be remembered that the Genius HF generator performs an


automatic heating values correction function at the end of each
exposure.
This function, if activated, guarantees the constancy of the parameters in
time.
If this function is not activated, we recommend the calibration of the tube
to be made every year.

(Rev. 3) 11-15 GENIUS HF


SERVICE MANUAL
Maintenance

11.3. EEprom management

The EEprom located on the MPU is used to contain a copy of the data
present on the buffered RAM.
During routine operations, the microprocessor uses the data contained in
the RAM without using the contents of the EEprom.
The following data is automatically memorized on the EEprom without
any request of intervention to the operator:
• the contents of the anatomical programs available to the user
• the data related to the configuration phase
• the data related to the calibration values of the three radiogenic
tubes (six focal spots).
The EEprom has a non-volatile support function.
In this way it is possible to automatically load all the data related to an
installation onto a new MPU when this card is substituted.
We recommend a recording be made on EEprom at the end of the
installation.

11.3.1. Data recording

The back-up, i.e. the recording of the data on the EEprom, is made every
time an anatomical program or configuration or calibration program is
completed.

GENIUS HF 11-16 (Rev. 3)


SERVICE MANUAL
Maintenance

11.3.2. Loading data from the EEprom

This operation must be performed every time the MPU or buffer RAM
needs to be substituted. The average life-span of the buffer battery is five
years.
If the FIP displays the following message: Error! Call servicing technician
it may be necessary to load the data from EEprom to RAM.
The following procedure must be performed to activated this function:
1. Open the front panel of the power rack and turn the switch located
on the front of PCB 9999102700 MPU V25 to the right (paragraph
12.1).
2. Switch on.
3. After approximately three seconds, the following message appears on
the FIP:

READING EEPROM PLEASE WAIT

confirming that the EEprom data has been read. Then turn the
switch on the MPU to its original position (towards the left).

(Rev. 3) 11-17 GENIUS HF


SERVICE MANUAL
Maintenance

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 11-18 (Rev. 3)


SERVICE MANUAL
Wiring diagrams

12. WIRING DIAGRAMS

12.1. Wiring diagrams – Electronic PCBoard

Drawing N° Description Sheet Vers./Rev.

594018.5823.04 Internal connection map X-ray generator 1 1/2


594019.5827.04 Control console connection 1 1/1
9999104200 PPU peripherical processing unit (layout) 1 1/2
9999104200 Peripherical processing unit PPU (drawing) 2 1/2
9999102700 MPU master processing unit micro controller 1 1/4
V25 (layout)
9999102700 Master processing unit MPU (drawing) 8 1/4
9999104300 AEC preamplifier (layout) 1 2/2
9999104300 AEC preamp. for Comet V2 / Zhiem V3 1 2/2
(drawing)
9999102500 Converter control board (layout) 1 4/3
9999102500 Converter control board (drawing) 4 4/3
(1) 560124.5800.05 IGBT quad driver (layout) 1 1/4
(1) 560124.5800.03 IGBT quad driver (drawing) 2 1/4
(2) 560214.5879.05 IGBT quad driver (layout) 1 1/2
(2) 560214.5879.03 IGBT quad driver (drawing) 2 1/2
9999103000 Safety supervisor (layout) 1 3/4
9999103000 Safety supervisor (drawing) 6 3/4
9999103200 Anode normal speed starter (layout) 1 1/2
9999103200 Anode normal speed starter (drawing) 1 1/2
9999103300 Normal/high speed starter control (layout) 1 1/2
9999103300 Normal/high speed starter control (drawing) 5 1/2
9999103400 Normal/high speed starter power (layout) 1 1/4
9999103400 Normal/high speed starter power (drawing) 1 1/4
9999102600 Filament inverter driver (layout) 1 2/2
9999102600 Filament inverter driver (drawing) 3 2/2
9999103100 PWS filament (layout) 1 1/1
9999103100 PWS filament inverter driver (drawing) 1 1/1
9999106000 7 segments display board (layout) 1 1/2
9999106000 7 segments display board (drawing) 1 1/2

(1) for inverter code 594.027 (65kW) and 594.025 (80kW)


(2) for inverter code 594.053 (65kW) and 594.054 (80kW)

(Rev. 3) 12-1 GENIUS HF


SERVICE MANUAL
Wiring diagrams

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 12-2 (Rev. 3)


SERVICE MANUAL
Wiring diagrams

12.2. Wiring diagrams – General terminal board

Drawing N° Description Sheet Vers./Rev.

9999103600 General terminal board 100 (layout) 1 1/2


9999103600 General terminal board 100 (drawing) 1 1/2
9999103700 General terminal board 200 (layout) 1 1/2
9999103700 General terminal board 200 (drawing) 1 1/2
9999102800 General terminal board 300 (layout) 1 1/1
9999102800 General terminal board 300 (drawing) 1 1/1
9999102900 General terminal board 400 (layout) 1 3/3
9999102900 General terminal board 400 (drawing) 2 3/3

(Rev. 3) 12-107 GENIUS HF


SERVICE MANUAL
Wiring diagrams

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 12-108 (Rev. 3)


SERVICE MANUAL
Wiring diagrams

12.3. Wiring diagrams – Power and high voltage circuits

Drawing N° Description Sheet Vers./Rev.

560118.5802.05 HV tank service GTB (layout) 1 1/2


560118.5802.03 HV tank service GTB (drawing) 1 1/2
594002.5813.03 Main service transformer 2 1/2
9999104700 Power rectifier filter (layout) 1 1/2
9999104700 Power rectifier filter (drawing) 1 1/2
594001.5812.03 Power circuit 3 2/3
9999104100 HV tank 1 tube top view 1 1/1
9999103800 HV tank 3 tubes top view 1 1/1
58581111 HV tank simplified schematic 1 1/1
594020.5824.03 Filament heating circuit 1 1/1

(Rev. 3) 12-127 GENIUS HF


SERVICE MANUAL
Wiring diagrams

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 12-128 (Rev. 3)


SERVICE MANUAL
Spare parts list

13. SPARE PARTS LIST

1– MICROPROCESSOR BOARD

2– SUPERVISOR / POWER SUPPLY

3– FILAMENT DRIVER / FIL PWS

4– INVERTER FILAMENT DRIVER / FIL PWS / STARTER


NORMAL SPEDD

5– NORMAL / HIGH SPEED STARTER

6– CONSOLE

7– RESISTOR / CONVERTER DC-DC

8– SWITCH
SWITCH BREAKER / X-RAY PUSH BUTTON

9– MISCELLANEOUS ELECTRICAL MATERIAL

10 – GENERAL TERMINAL BOARD

11 – H.T. TRANSFORMER / INVERTER

12 – MISCELLANEOUS MATERIAL

(Rev. 3) 13-1 GENIUS HF


SERVICE MANUAL
Spare parts list

1 – MICROPROCESSOR BOARD

Ref. VSM code Description Notes

1 9999102100 Box for MPU PCB

2 9999102700 MPU PCB without box


(see NOTE)

NOTE:
/ If the new CPU board (Rev. 05) must be installed on a units having serial
number lower than 798k0, remove EPROM U91 from the old MPU
board and install it on the new one in place of the existing EPROM. This
action is required to match the revision level of the other software already
installed on the unit itself.
In case the new MPU board is installed on units having serial number
equal or higher than 798k0, the software mounted on the new MPU
board is compatible with the equipment and therefore no any further
action is required.

GENIUS HF 13-2 (Rev. 3)


SERVICE MANUAL
Spare parts list

(Rev. 3) 13-3 GENIUS HF


SERVICE MANUAL
Spare parts list

2 – SUPERVISOR / POWER SUPPLY

Ref. VSM code Description Notes

1 9999101900 Box for Supervisor and Power


Supply

2 9999103000 Safety Supervisor PCB

3 9999103500 Power Supply ZS

GENIUS HF 13-4 (Rev. 3)


SERVICE MANUAL
Spare parts list

(Rev. 3) 13-5 GENIUS HF


SERVICE MANUAL
Spare parts list

3 – FILAMENT DRIVER / FIL PWS

Ref. VSM code Description Notes

1 9999102000 Box for Inverter Filament Driver

2 9999102600 Inverter Filament Driver PCB

3 9999103100 Power Supply Filament

GENIUS HF 13-6 (Rev. 3)


SERVICE MANUAL
Spare parts list

(Rev. 3) 13-7 GENIUS HF


SERVICE MANUAL
Spare parts list

4 – INVERTER FILAMENT DRIVER / FIL PWS / STARTER


NORMAL SPEED

Ref. VSM code Description Notes

1 9999102000 Box for Inverter Filament Driver

2 9999102600 Inverter Filament Driver PCB

3 9999103100 Power Supply Filament

4 9999103200 Normal Speed Starter

GENIUS HF 13-8 (Rev. 3)


SERVICE MANUAL
Spare parts list

1 2

(Rev. 3) 13-9 GENIUS HF


SERVICE MANUAL
Spare parts list

5 – NORMAL / HIGH SPEED STARTER

Ref. VSM code Description Notes

1 9999101800 Box for Normal / High Speed


Starter

2 9999102200 Cable Double Shield

3 9999102300 HS Starter fuses support

4 9999103300 Normal / High Speed Starter


Control

5 9999103400 Normal / High Speed Starter


Power

GENIUS HF 13-10 (Rev. 3)


SERVICE MANUAL
Spare parts list

1 2

(Rev. 3) 13-11 GENIUS HF


SERVICE MANUAL
Spare parts list

6 – CONSOLE

Ref. VSM code Description Notes

1 9999104400 FIP

2 5468876600 Touch panel for control desk

3 9999104200 Periferical Processing Unit (PPU)


PCB

6668100100 Display X-ray lamp (for unit


S/N < 798k0)
4
9999106000 Display X-ray lamp (for unit
S/N ≥ 798k0)

8368876900 Console Genius HF Assy (with


cable)

9999104800 Connection Cable Rack-Console


(20m)

GENIUS HF 13-12 (Rev. 3)


SERVICE MANUAL
Spare parts list

2 3

(Rev. 3) 13-13 GENIUS HF


SERVICE MANUAL
Spare parts list

7 – RESISTOR / CONVERTER DC-DC

Ref. VSM code Description Notes

1 9999100700 Resistor RFC 52 ohm

2 9999100800 Resistor RFC 110 ohm

3 9999104600 Resistor 15 Kohm for Capacitor


Discharge

4 9999100100 DC/DC Converter IN 24V ,


OUT ±15V

5 9999100200 DC/DC Converter IN 24V,


OUT ±5V

GENIUS HF 13-14 (Rev. 3)


SERVICE MANUAL
Spare parts list

4 5

(Rev. 3) 13-15 GENIUS HF


SERVICE MANUAL
Spare parts list

8 – SWITCH BREAKER / X-RAY PUSH BUTTON

Ref. VSM code Description Notes

1 9999101200 Breaker NC100LH 3F 63A (S11)

2 9999101300 Breaker C50H 2F 6A (S12)

3 9999101400 Breaker C60H 3F 1A (S14)

4 9999101500 Breaker C60H 2F 1A (S15)

5 9999101600 Breaker E81N B6 (S13)

6 9999101700 X-ray Push Button

GENIUS HF 13-16 (Rev. 3)


SERVICE MANUAL
Spare parts list

1 2

3 4

5 6

(Rev. 3) 13-17 GENIUS HF


SERVICE MANUAL
Spare parts list

9 – MISCELLANEOUS ELECTRICAL MATERIAL

Ref. VSM code Description Notes

1 9999221800 Contactor 3TF 4622 220V 50Hz


(LB)

2 9999101100 Contactor 3P 63A 220/240V (LA)

3 9999100900 Fuse E1000-32 (F1/F2)

4 9999100600 Electrolytic capacitor 600uF


350Vdc (C1 ÷ C4)

5 9999100500 Display LM 16x21B

6 9999100300 Bridge diode PSD 162/12 (D1)

GENIUS HF 13-18 (Rev. 3)


SERVICE MANUAL
Spare parts list

1 2

3 4

5 6

(Rev. 3) 13-19 GENIUS HF


SERVICE MANUAL
Spare parts list

10 – GENERAL TERMINAL BOARD

Ref. VSM code Description Notes

1 9999103600 General Terminal Board 100


(GTB 100)

2 9999103700 General Terminal Board 200


(GTB 200)

3 9999102800 General Terminal Board 300


(GTB 300)

4 9999102900 General Terminal Board 400


(GTB 400)

GENIUS HF 13-20 (Rev. 3)


SERVICE MANUAL
Spare parts list

(Rev. 3) 13-21 GENIUS HF


SERVICE MANUAL
Spare parts list

11 – H.T. TRANSFORMER / INVERTER

Ref. VSM code Description Notes

1 9999103800 H.T. Transformer 2/3 Tubes

2 9999104100 H.T. Transformer 1 Tube

3 9999106200 Inverter Assy 65 kW

4 9999106300 Inverter Assy 80 kW

GENIUS HF 13-22 (Rev. 3)


SERVICE MANUAL
Spare parts list

(Rev. 3) 13-23 GENIUS HF


SERVICE MANUAL
Spare parts list

12 – MISCELLANEOUS MATERIAL

Ref. VSM code Description Notes

1 9999102400 Service Transformer

2 9999104300 AEC for Comet

3 9999102500 Converter Control Board

4 9999104700 Power Rectifier Filter PCB

GENIUS HF 13-24 (Rev. 3)


SERVICE MANUAL
Spare parts list

1 2

(Rev. 3) 13-25 GENIUS HF


SERVICE MANUAL
Spare parts list

THIS PAGE IS INTENTIONALLY LEFT BLANK

GENIUS HF 13-26 (Rev. 3)


Cod. 6968904703_Rev.3 0434

VILLA SISTEMI MEDICALI S.p.a.


Via Delle Azalee, 3
20090 Buccinasco (MI) - ITALY
Tel. (+39) 02 48859.1
Fax (+39) 02 4881844

You might also like