Bluephase® N G4

Start-up
✓ Unpack handpiece ✓ Connect the power cord with ✓ Slide battery straight into the ✓ Place handpiece in the corresponding
1 ✓ Detach the light guide by pulling it out 2 the power supply 3 handpiece until you hear and feel it
click into place.
4 rest in the charging base  isposable protective sleeves
D
are recommended for each
If the battery is fully
discharged, the charging time is
✓ Clean them (see Instructions for Use) patient. approx. 2 hours
Attach anti-glare cone or
anti-glare shield to the light
guide.

Operation
What if …?
5 If any of these errors persist after trying their corresponding rectification, please contact your dealer or your local Service Centre.
HIGH POWER TURBO
Curing times
1,200 mW/cm2 ± 10% 2,100 mW/cm2 ± 10%
Start/Stop button
Composites Indicator Causes Rectification
• 2 mm 1)
Time selection button IPS Empress Direct / IPS Empress Direct Effect 10 seconds 5 seconds RED “X” LIGHTS UP
Program selection button Tetric N-Ceram / Tetric Prime / Tetric EvoFlow 10 seconds 5 seconds Allow the device to cool down and try again after
The device is overheated
Tetric N-Flow 10 seconds – certain time.
• 4 mm 2)
Restorative materials Tetric PowerFill 10 seconds –
Tetric PowerFlow 10 seconds 5 seconds Electronic component of the handpiece is defective. Remove and reinsert the battery.
Tetric N-Ceram Bulk Fill / Tetric N-Flow Bulk 10 seconds 5 seconds
Turbo (2,100 mW/cm2)
Compomer 3)
High Power (1,200 mW/cm2) 20 seconds 2x 5 seconds
Compoglass F
PreCure (950 mW/cm2) RED “X” AND BATTERY
Alkasites
20 seconds – SYMBOL LIGHT UP Battery empty Place the device in the charging base and charge it.
Cention N
Variolink Esthetic LC / Variolink Esthetic DC
4) 5)
per mm ceramic: per mm ceramic:
Indirect restorations / Variolink N / Variolink N LC 4) 10 seconds per surface 5 seconds per surface
cementation
materials per mm ceramic: per mm ceramic: Battery contacts dirty Remove battery and clean the battery contacts.
Multilink N 5) / Multilink Speed 5)
Curing times 20 seconds per surface 2x 5 seconds per surface
Adhesives Tetric N-Bond Universal 10 seconds 5 seconds
Telio Link 10 seconds per surface 2x 5 seconds per surface BATTERY IS NOT LOADING Remove the handpiece from the charging base and
Temporary materials Charging base does not recognise the battery
Telio Inlay/Onlay 10 second 5 seconds place it back again
Helioseal F Plus 10 seconds – Remove and reinsert the battery and place the
IPS Empress Direct Color 20 seconds 5 seconds Battery empty
Miscellaneous handpiece in the charging base
IPS Empress Direct Opaque 20 seconds 2x 5 seconds
MultiCore Flow 20 seconds 2x 5 seconds
Electronic component is defect Please contact your dealer or your local Service Centre
1) Applies to a maximum layer thickness of 2 mm and provided that 3) Applies to a maximum layer thickness of 3 mm
the Instructions for Use of the respective material do not state any other ­ 4) Applies to light-curing
recommendation (might be the case e.g. with dentin shades)
5) Applies to dual-curing
2) Applies to a maximum layer thickness of 4 mm and provided that
the Instructions for Use of the respective material do not state any other
­recommendation (might be the case e.g. with dentin shades)

The PreCure program must not


be used for conventional light-
Rx ONLY
curing! Manufacturer In USA distributed by
Ivoclar Vivadent AG Ivoclar Vivadent Inc.,
For dental use only! see 9494 Schaan/Liechtenstein 175 Pineview Drive, Date information prepared: 2021-10-12 / Rev. 1
Instructions
Made in Austria www.ivoclarvivadent.com Amherst, NY 14228, USA 754611