You are on page 1of 1

137

Error 137 means a communication between microcontroller that check the weight sensor and the
one who save the memory of the card.

In this case it could be that the problem is the UGSG card, or a short circuit of power on weight
sensor.

If you remove the 2 wires that powered the weight sensor and the error disappear, the problem is
the sensor,

if the error will be there in any case, is necessary replace the UGSG card.