Professional Documents
Culture Documents
BGE-800 Series
User Manual
BIOBASE GROUP
Version 2020.08
Content
1. General introduction.........................................................................................................................1
1.1 Intended use..................................................................................................................................1
1.2 Model family................................................................................................................................ 1
1.3 Product features............................................................................................................................1
1.4 Main parts.....................................................................................................................................2
1.4.1 General overview..................................................................................................................2
1.4.2 Sensor system....................................................................................................................... 4
1.4.3 Barcode scanner....................................................................................................................4
1.4.4 Reagent cassette....................................................................................................................5
1.4.5 Requested circumstance........................................................................................................6
2. Operating and safety information........................................................................................................6
2.1 Oerator’s qualifications................................................................................................................ 6
2.2 Operating environment.................................................................................................................6
2.3 Prevention of EMI........................................................................................................................ 7
2.4 Electric shock............................................................................................................................... 7
2.5 Personnel injuries......................................................................................................................... 7
2.6 Biological hazards........................................................................................................................ 7
2.7 Chemical hazards......................................................................................................................... 8
2.8 Waste hazards...............................................................................................................................8
2.9 Cleaning and sterilization.............................................................................................................8
2.10 Disposal...................................................................................................................................... 8
2.10.1 Disposal of instrument........................................................................................................8
2.10.2 Disposal of battery..............................................................................................................9
2.10.3 Disposal of sample port, electrode, pump tubes, Cal pack.................................................9
2.11 Operation.................................................................................................................................... 9
2.12 Maintenance............................................................................................................................... 9
2.13 Sample........................................................................................................................................ 9
2.13.1 Sample sources................................................................................................................. 10
2.13.2 Sample requirements........................................................................................................ 10
2.13.3 Anticoagulant.................................................................................................................... 11
2.13.4 Sample collection devices.................................................................................................11
2.13.5 Sample storage..................................................................................................................12
2.14 Reagents, calibrators, and controls...........................................................................................12
3. Flow path...........................................................................................................................................13
3.1 Flow path of flush.......................................................................................................................14
3.2 Flow path of Cal A..................................................................................................................... 15
3.3 Flow path of Cal B..................................................................................................................... 16
3.3 Flow path of sample test.............................................................................................................16
4. Specifications and parameters...........................................................................................................17
4.1 Parameters.................................................................................................................................. 17
4.1.1 Measuring parameters.........................................................................................................17
4.1.2 Calculated parameters.........................................................................................................18
4.1.3 Input parameters................................................................................................................. 18
4.1.4 Electrode Parameters.......................................................................................................... 18
4.2 Specifications............................................................................................................................. 18
4.3 Sample type................................................................................................................................ 20
4.4 The speed of analysis................................................................................................................. 20
4.5 The volume of sample................................................................................................................ 20
4.6 Output ways................................................................................................................................20
4.7 Dimensions.................................................................................................................................20
4.8 Fuse protector............................................................................................................................. 20
4.9 Memory...................................................................................................................................... 20
4.10 Printer....................................................................................................................................... 20
4.11 Battery...................................................................................................................................... 20
5. Installation.........................................................................................................................................21
5.1 Operating environment...............................................................................................................21
5.2 Unpacking and check................................................................................................................. 21
5.3 Electrodes installation................................................................................................................ 22
5.3.1 Electrodes preparation........................................................................................................ 22
5.3.2 Electrodes introduction.......................................................................................................22
5.3.3 Electrodes installation.........................................................................................................25
5.4 Barcode scanner installation.......................................................................................................26
5.5 Pump tube installation................................................................................................................ 27
5.6 Cal pack installation................................................................................................................... 29
5.7 Auto QC installation(optional)...................................................................................................31
5.8 Printer paper installation............................................................................................................ 32
5.9 Install the battery........................................................................................................................ 33
5.10 Check for correct installation................................................................................................... 34
5.11 Open up.................................................................................................................................... 35
5.12 Self test..................................................................................................................................... 35
5.13 Shut down.................................................................................................................................36
5.13.1 Within 24 hour..................................................................................................................... 36
5.13.2 Beyond 24 hour....................................................................................................................36
6 Menu introduction............................................................................................................................37
6.1 Main menu................................................................................................................................37
6.1.1 General introduction........................................................................................................... 37
6.1.2 Basic operation................................................................................................................... 38
6.1.3 Menu structure.................................................................................................................... 38
6.2 Test menu................................................................................................................................... 39
6.3 1 Point calibration and 2 point calibration menu..................................................................... 41
6.4 Flush menu............................................................................................................................... 41
6.5 QC menu...................................................................................................................................42
6.5.1 QC lot management............................................................................................................ 42
6.5.2 QC test................................................................................................................................ 44
6.6 Data management menu............................................................................................................. 44
6.7 Maintenance menu..................................................................................................................... 45
6.8 System Setup menu.................................................................................................................... 45
6.9 Expendable menu....................................................................................................................... 46
7 Calibration..........................................................................................................................................47
7.1 General information................................................................................................................... 47
7.2 1 Point Calibration......................................................................................................................47
7.3 2 Point Calibration......................................................................................................................49
7.4 Calibration intervals................................................................................................................... 51
7.5 pO2 zero calibration....................................................................................................................52
7.6 Calibration data.......................................................................................................................... 52
8 Quality Control.................................................................................................................................. 54
8.1 General information................................................................................................................... 54
8.2 Preparing QC solutions.............................................................................................................. 54
8.3 Lot number management............................................................................................................54
8.4 QC test........................................................................................................................................57
8.5 QC plot data................................................................................................................................58
8.6 QC plot....................................................................................................................................... 59
9 Patient sample analysis...................................................................................................................... 59
9.1 Collect samples...........................................................................................................................60
9.1.1 Collect syringe samples...................................................................................................... 60
9.1.2 Collect capillary samples....................................................................................................61
9.2 Analyze the sample.................................................................................................................... 61
9.3 Check the test result................................................................................................................... 63
10 Setup system.................................................................................................................................... 65
10.1 Correlation................................................................................................................................65
10.2 Reference Range.......................................................................................................................66
10.3 Unit...........................................................................................................................................67
10.4 Test parameters.........................................................................................................................67
10.5 Calibration intervals................................................................................................................. 68
10.6 Date and time............................................................................................................................68
10.7 Language Setting......................................................................................................................69
10.8 Version..................................................................................................................................... 69
10.9 Replace Sample port.................................................................................................................70
10.10 LIS interface........................................................................................................................... 70
10.11 Printer..................................................................................................................................... 72
10.12 Self test................................................................................................................................... 73
10.13 Power Off............................................................................................................................... 73
10.14 Other utilities..........................................................................................................................74
11 Maintenance..................................................................................................................................... 74
11.1 Maintenance functions..............................................................................................................75
11.1.1 Clean................................................................................................................................. 75
11.1.2 Deproteinize......................................................................................................................76
11.1.3 Condition.......................................................................................................................... 77
11.2 Sterilization...............................................................................................................................78
11.3 Daily maintenance.................................................................................................................... 78
11.4 Weekly maintenance.................................................................................................................79
11.5 Monthly maintenance............................................................................................................... 79
11.6 Every 3 months.........................................................................................................................79
11.7 Every 6 months.........................................................................................................................80
11.8 As necessary............................................................................................................................. 80
12 Troubleshooting............................................................................................................................... 80
12.1 Liquid pathway problem.......................................................................................................... 80
12.2 ISE Electrodes problem............................................................................................................81
12.2.1 Only one electrode............................................................................................................81
12.2.2 More than two electrodes..................................................................................................81
12.3 Blood gas electrodes problem.................................................................................................. 82
12.4 Correlation coefficients......................................................................................................... 83
A Appendix........................................................................................................................................... 84
A.1 Potential range for electrodes.................................................................................................... 84
A.2 Interference substances..............................................................................................................85
A.3 Reference Range for adult......................................................................................................... 87
A.4 Table of Critical values..............................................................................................................87
A.5 Calculations............................................................................................................................... 88
A.5.1 Temperature corrections.....................................................................................................88
A.5.2 Calculate parameters..........................................................................................................89
A.6 Working principle...................................................................................................................... 91
A.6.1 ISE electrodes.................................................................................................................... 92
A.6.2 pCO2 electrode...................................................................................................................93
A.6.3 pO2 electrode......................................................................................................................93
A.6.4 Hct electrode...................................................................................................................... 94
A.6.5 Glu/Lac electrode...............................................................................................................94
A.7 References................................................................................................................................. 94
A.8 Warranty.................................................................................................................................... 95
A.8.1 Warranty guidelines........................................................................................................... 95
A.8.2 Warranty limitation............................................................................................................ 96
B Appendix Barcode scanner initialization........................................................................................96
B.1 Apply for LV880(32B)...............................................................................................................97
B.2 Apply for LV880........................................................................................................................98
Index.......................................................................................................................................................100
Preface
1. About this manual
Thank you very much for purchasing our products.
This manual provides the information and procedures necessary to operate and
maintain the BGE-800 series Blood gas/Electrolyte. It's designed to meet the needs of
medical personnel who use the system on a daily basis and perform routine maintenance
and troubleshooting. All the BGE-800 series products are very similar in basic operation
and maintenance, most of the information and procedures provided here also applied to
other BGE-800 series models.
2. Conventions used in this manual
The following text and symbol conventions are used throughout this manual:
Italic To indicate a document reference.
CAPITAL This indicates a displayed menu.
Underline This indicates a command that appears under a menu.
“message” This indicates a quoted message displayed on the LCD.
3. Understand the symbols
This section describes the symbols that may appear on the exterior of the system. These
symbols provide you with either important information or warning for proper operations.
Biohazard Warning
Caution to alter the user to possible personnel injury or damage to the analyzer.
Place upward
European representative
Manufactured date
Manufacturer
1. General introduction
1.1 Intended use
1.The BGE-800B analyzer system is an automated analyzer for determining pO2, pCO2,
potassium(K+), sodium(Na+), chloride(Cl-), ionized Ca(Ca2+), pH, Glucose(Glu),
Lactate(Lac) and tHb in the arterial, venous and capillary whole blood sample.
2.This analyzer is intended for using by trained technologists, nurses, physicians and
therapists in a laboratory environment,near patient or point-of-care setting.
Page 2 of 106
Part Function
Touch screen Communication interface between the operater and the
analyzer
Measurement module All the samples are tested here
Pump Provide pressure to move the fluid in the tube
Reagent cassette Store reagent and waste
Sampler To aspirate samples and all kinds of test reagent
Flow path board An intergrated borad for controlling fluid flow tubes
Part Function
Power cord socket For connecting the power cord
Battery switch For turning the backup battery on and off
RJ45 Ethernet network Ethernet interface connection
port
RS232 serial port For tranferring data
RS232 serial port For connecting scanner
Nameplate Describe the basic information and configuration of the
analyzer
USB slave To connect to PC
USB host To connect to U disk,keyboard and mouse
Page 3 of 106
Printer Thermal printer to print data and plots
Fan For dissipating heat
Battery cassette The internal battery provides power for the analyzer when
power off
The barcode contains the lot number and use life of the product, and the barcode
scanner is used to scan the barcode so the system can control the consumables effectively.
The scanner just can be used to input barcode, please don't use it in any
other processes to avoid interfere.
Page 4 of 106
Specifications
Light source Visible laser diode, wave length 650mm
Scanning speed 48±2 per second
Supported port RS232,PS2 keyboard,USB
Dimension 95mm×70mm×160mm
Weight 128g
Voltage 5V
Working current 85mA
Static current 36mA
Laser security Comply with the national laser security standard
Working temperature 0℃-45℃
Storage temperature -20℃-60℃
Humidity Relative Humidity:5%-95%(no condensation)
The reagent cassette provide the neccesary solution for calibrating the pH, pCO2,pO2, Na+,
K+, Ca2+, Cl-, Hct, Glu, Lac including three calibration solution。
Page 5 of 106
The reagent cassette may have biohazard risk when the solution is run out,
please dispose it according to the related biohazard regulation in your
country.
Devices which make big noise may interfere this instrument and this instrument
should be kept away from noisy surrounding.
1. The electromagnetic devices such as mobile or radio transmitter should be turned off in the
room where the instrument is installed.
2. CRT display is not recommended to be used with this instrument.
3. Other medical instruments should be avoided to be operated nearby.
4. The electromagnetic wave generated by this instrument may interfere with other
instruments close to this system and cause them fail to function.
1. None authorized personnel are permitted to open the cover while power is connected.
2. Spilled reagent or specimen can cause system failure or electric shock when they enter into
the interior of the analyzer. Reagent or specimen shall not be placed on the analyzer. Turn
off the analyzer immediately if spill of reagent or specimen is observed.
3. There are two delayed fuse inside the power module and their specification is F3.15A
250VAC. The replacement of the fuse should be performed by authorized personnel only. It
is totally prohibited for the operator to do so.
4. The power supply should be in accordance with the requirement of 100-240V~ 50/60Hz.
Any power supply other than this requirement can cause system failure.
5. Earth grounding is essential to the safety and operation of the analyzer. The socket shall
ground to earth to protect the operator from shocks and prevent electric interferences.
Never put your hand or fingers into any opened component or touch the probe
when the analyzer is under operation.
2.10 Disposal
Page 8 of 106
2.10.2 Disposal of battery
Sealed lead batteries contain heavy metals such as lead which can contaminate the
environment when batteries are improperly disposed of. Please refer to local wasted
regulation and disposal for more details.
2.11 Operation
Please follow the instructions listed in this manual to operate the instrument. Improper
operation may result in unreliable results or even system failures or injuries.
2.12 Maintenance
1. Please follow the instructions in this manual to maintain this instrument. Improper
maintenance may result in unreliable results or even system failures or injuries.
2. Use soft cloth and water to clean the surface. A small quantity of soap is permitted to be
added into the water where appropriate.
3. Organic solvent such as ethanol is totally prohibited. Wipe dry the surface after cleaning.
4. The power supply should be cut off before cleaning. Necessary precautions should be
taken to prevent water from spilling into inside the instrument.
5. Check pump tube periodically and replace pump tube when necessary. The tubing is
recommended to be replaced for every 6 months to prevent them from aging or blockage.
2.13 Sample
This section describes sample requirement, collection and handling techniques for pH,
blood gas, and electrolyte analysis. There is complete sample handling and storage
information in the standard Clinical Chemistry procedures published by CLSI. For detailed
information about sample collection, please refer to CLSI document C27-A, Blood Gas
Preanalytical Considerations: Specimen Collection, Calibration, and Controls.
Page 9 of 106
Wear disposal gloves and other appropriate precautions to avoid potential
infections during these operations
Page 10 of 106
2.13.3 Anticoagulant
1. Calcium titrated(balanced) heparin and lithium heparin are the only acceptable
anticoagulant.
2. Do not use anticoagulants such as EDTA, citrate, oxalate and fluoride since they may
significantly alter pH, ionized calcium, chloride and metabolites.
3. Heparin final concentration should not exceed the limits previously reported. The
recommended concentration is 15-20U/mL. Higher heparin concentration can affect
potassium and sodium.
4. A blank syringe must be heparinized first before collecting blood sample. Exclude the air
from the disinfected syringe, then draw the anticoagulant of 1000 unit/mL heparin, pull
the plunger to make the anticoagulant wet full wall, and push it back to discharge the
excess. Remaining heparin in the needle and syringe dead space is enough for the anti-
coagulation of 2mL whole blood.
Page 11 of 106
2.13.5 Sample storage
1. For optimal results, samples should be analyzed within five minutes.
2. If a sample cannot be measured within 15 minutes after drawing, it must be placed in a
bath containing ice and water to slow down the metabolic process. Ice water stored
samples are stable for up to 30 minutes.
3. The pO2 value may be affected by the time interval between sampling and analysis.
4. It should be noted that potassium levels are generally affected by icing. Potassium
elevations of several mmol/L have been observed after only a few minutes on ice.
Adequate precautions should be taken to avoid cell hemolysis, resulting in falsely elevated
potassium values.
Page 12 of 106
Conditioner 0.8mL*5 Condition electrode Stored at 5-25℃.
Do not freeze.
Auto QC cartridge H/M/L Auto Quality control Stored at 5-25℃.
Do not freeze.
Quality controls H*10/M*10/L*10 Manual Quality Stored at 5-25℃.
control Do not freeze.
3. Flow path
The following chart shows the flow of fluid through this system.
Valve:
V1/V2: Air valve
V3/V11: Flush valve
V5: Clean valve
V7/V6//V4: Cal A/Cal B/Cal E
V6: flush
Page 13 of 106
Pump
p1/p2/p3: Waste/Flush/Aspiration
Liquid sensor
J1/j2/j3: Reagent Cassette sensor/ Measurement module front sensor/ Measurement
module back sensor
Page 14 of 106
3.2 Flow path of Cal A
Page 15 of 106
Flow path of A-2
Page 16 of 106
Flow path of sample test-2
Page 17 of 106
Hct % 12~65%
B.P. mmHg 500~800 mmHg
Glu mmol/L (1.1~66.7)mmol/L
Lac mmol/L (0.4~30.0)mmol/L
4.2 Specifications
Page 18 of 106
This section provides the requirements, specifications and typical performance of the
analyzer.
Dimension
Size Length Width Height
Value(mm) 400 574 344
Weight
Main unit 22.3kg
Power requirements and consumption
Power supply AC100~240V
Rating Consumption 100VA
Fuse 2×F3.15AL 250VAC
Built-in Battery 2×12V 2.3AH Sealed lead-acid battery
Operation environment
Ambient temperature 5~40℃
Relative humidity Up to 80% noncondensing
Atmosphere pressure 86Kpa~106Kpa
Storage and transportation conditions:
Temperature -20℃~+60℃
Humidity Up to 95% noncondensing
Sample volume
Typical 200µL
Sample type
Type Arterial, Venous, Dialysate, Capillary
Screen
Type 10” TFT Touch screen
Interface
Interface Descriptions
RS 232 Serial port
USB
TCP/IP Internet/Ethernet
Barcode Barcode reader
Printer
Page 19 of 106
Type Thermal printer
Resolution 240×128 pixel
Full graphics 8 dots/mm
Printing speed 15mm/s
Paper 80 mm(W) × 30 mm(D)
4.7 Dimensions
Length×Width×Height: 344mm×400mm×574
Net weight of the analyzer:25kg
4.9 Memory
RAM;64Mb flash:256Mb
4.10 Printer
Auto thermal printer:
Paper width:80mm(W)×50mm(D)
Speed:15mm/s
Resolution:240×128 pixel
4.11 Battery
Page 20 of 106
Internal battery: 2×12V 2.3AH
Charging time: 6 hours to get full charge
Maximum using time: 30
5. Installation
5.1 Operating environment
The installation of the analyzer should be proceed in such environment:
1 Temperature : 15-30℃.
2 Relative humidity :≤80% (without condensate).
3 Power supply :100-220V~50Hz/60Hz.
4 Atmosphere pressure: 86~106KPa
5 The earth of the socket shall be well grounded and keep way the analyzers from
possible electromagnetic interference.
6 Working area: L*W (1.5*0.6) and at least 0.5m far from other analyzers.
7 Others: Avoid sunlight irradiation, erosive gas, great temperature change and dust.
Front cover shall be kept close when operation. Open it by skilled or trained
medical personal only where appropriate.
Never put your fingers into reagent pack housing at the bottom. The sharp pin
inside can cause bodily injury.
Page 21 of 106
5.3 Electrodes installation
Page 22 of 106
pO2 and pCO2.electrode are independent electrodes. Their performance is
limited to electrode itself only and irrelevant to other electrodes.
ISE electrodes
Reference electrode
1. Take out the reference electrode.
2. Remove the tape on both sides that covers the sample hole.
3. Install O ring on both sides.
4. Screw out the cap on the right cavity.
Page 24 of 106
5. Use a syringe to aspirate refill solution for reference(20mL), then inject into
reference electrode. The solution surface should reach at least 1/3 height of right
cavity.
6. Flip the bottom if there is any air bubbles above membrane area on the left cavity.
Refill solution for ISE is totally different from refill solution for REF. Never
mix use those two solutions under any circumstances.
Page 25 of 106
Exclude air bubble above membrane area before installation of Gas electrodes
and ISE electrodes.
Measuring chamber and electrode surface should be cleaned before installation.
The following picture shows you the specific position of all the electrodes, you can
take it as reference:
1. Take out the scanner and connect the series bus of the scanner to the RS232 port on the rear
panel of the analyzer.
Page 26 of 106
5.5 Pump tube installation
There are 3 pumps on this system just below the measuring chamber. These pumps have
different functions. From left to right they are : Waste pump, Reagent pump and aspiration
pump.
Waste pump: Pump waste dripped from sample port
Flush pump: Aspirate flush solution.
Aspiration pump: Aspirate sample and other solutions from measuring chamber.
Please install the pump tube according to the following steps:
1. Click “Expendable” in the main menu to enter the expendable management
interface.
2. Take out the pump tube and use the barcode scanner to scan the barcode on
package.
3. Take down the Tygon tube connected to the ends of the old pump tube.
(Skip to next step if it's your first time to install pump tubes)
4. Take the old pump tubes out slowly by rotating the pump wheel anticlockwise.
(Skip to next step if it's your first time to install pump tubes)
Page 27 of 106
5. Put the one end of the new pump rube to the pump slot.
(Please take the pump tubes out when it comes to long term transportation)
6. Rotate the pump wheel the pimp tube will be squeezed into the gap, put the
the other end of the tube in the slot when all the tube is squeezed into the gap.
7. Rotate the pump wheel clockwise and anticlockwise until the pump tube is
placed in the gap perfectly.
Page 28 of 106
8. Then insert the Tygon tube into the steel tube of the pump tube, and ensure the
joint length is no less than 5mm.
9. The table below shows how to connect the pump tube to the Tygon tubes correctly:
Never put your hand into the tube slot to avoid possible injuries.
Page 29 of 106
2. Take out the cal pack and scan the barcode on it.
3. Tear off the protection membrane on the new cal pack and pull out glass plug on the pack.
3. Hold the both ends of the pack, slowly insert the pack into the reagent cavity at the bottom
of the analyzer.
Page 30 of 106
On board life for cal pack is one monthly only. Extended use of cal pack can
lead to wrong test results.
There is possibility of waste leakage from cal pack when operation. Special
cautions shall be always taken while handling cal pack or operating analyzer.
Used Cal pack which may contain or contact with biohazard materials should
be disposed in accordance with local regulations of government
Connecting pin inside the reagent housing is sharp and dangerous. Never stretch
your fingers into the housing to avoid possible injuries.
Cal pack on this system is for single use only. Reuse of Cal pack is not
permitted once it is pulled out.
Page 31 of 106
Connecting pin inside the reagent housing is sharp and dangerous. Never stretch
your fingers into the housing to avoid possible injuries.
Cal pack on this system is for single use only. Reuse of Cal pack is not
permitted once it is pulled out.
2. Load a new roll of paper inside the compartment, with the free end of the paper coming
forward off the roll from the bottom. Ensure the free end of the paper extends beyond the
lip of the printer cover.
Page 32 of 106
3. Then close the printer cover, get rid of the redundant paper outside, the installation is
complete.
The inside of the printer might be hot. To reduce the risk of injury from a hot
component, allow the surface to cool before touching it.
Page 33 of 106
2. Plug in the battery.
3. Push the battery into the unit of the analyzer and tighten the 4 screws.
Please pay attention to the positive and negative pole of the battery, you can
install it as the label shown on the battery box.
Page 34 of 106
5.11 Open up
1. Turn off the switch on power socket
2. Insert the power cord into the socket of the analyzer.
3. Plug another end of power cord into a grounded outlet.
4. Turn on the switch on power socket and start up the analyzer.
5. Turn on the switch on the rear panel of the analyzer to enable UPS support.
1.An ungrounded power supply can lead to drift problem of all electrodes.
2.High-power analyzer with the same outlet can interfere with the analyzer.
Power switch on the rear panel is for UPS control only. Power outlet should be
equipped with a switch to control power supply.
Battery of UPS should be recharged before use. Open UPS switch and keep
power on to recharge battery.
Page 35 of 106
If self test is failed for a second time, power off the analyzer and start trouble
shooting. Please refer to the section of trouble shooting.
For new electrodes, if the problem of drift is happened, feed fresh serum to
activate electrode for 30 minutes, then check again.
Page 36 of 106
4 Release pump tube.
5 Turn off the switch on the rear panel to disable UPS.
6 Unplug power cord to switch off the analyzer.
7 Open battery cover to remove battery.
8 Clean the surface of analyzer and put it into carton box
6 Menu introduction
6.1 Main menu
1 Point Calibration, 2 Point Calibration and Flush are shortcut menus. Calibration or Flush
can be directly activated by clicking the menu.
Temp is the temperature of measuring chamber.
Baro pressure and temperature are displayed after date and time.
All test can be found under Test menu.
Page 37 of 106
6.1.2 Basic operation
1. Menu activation
All menus can be activated or selected by direct click on the menu. Click Main Menu to
return to home if it is displayed on the top of window.
2. Activate command
Click command icon directly to activate an operation. Click OK to confirm or save, Click
Cancel or Return to exit.
3. Select option
Click the marquee before an option to select or cancel it.
4. Data input
A. The system support virtual keypad to edit or enter data. Click data field to pop up virtual
keypad and click OK to close after finished.
B. There are two kinds of keypad. One is for digital input only. Another one is for letter,
symbol and number.
Click to switch number input, click to input symbol, click to input letter.
Page 38 of 106
6.2 Test menu
1. Click “Test” on the main menu, the test interface appears as below:
Page 39 of 106
2. Select the sample type, then select parameters(You cannot select the one not selected in
“Test Parameters” menu).
3. Put the syringe at the sample port, click “Go” , then probe starts to aspirate
the sample. A sign saying “Please remove sample” shows on bottom of the screen when the
aspiration is finished, the system starts to test the sample.
4. We can see a series motion of the machine part when the system starts to test. Meanwhile,
the user can type into the corresponding patient information such as “Patient ID” ,
“P.Name”(Patient name), “Age”(Patient age), “Temp”(Patient temperature), “Hb”(Total
hemoglobin value measured from other analyzer) and “FIO2”(Fraction of inspired oxygen).
Page 40 of 106
6.3 1 Point calibration and 2 point calibration menu
Click “1 Point Calibration” or “2 Point Calibration” on the main menu, the system
conducts the corresponding calibration, the result of the calibration displayed on the screen
automactically.( Please see more details about calibration in chapter 7)
The status bar located at the left top of the screen, and the status changes along with the
process.
For more detailed information, please refer to chapter 7.
Page 42 of 106
Icon Function
More To check and edit the data of the test results
Add To add a new lot number
Delete Delete the data base of a lot number
Empty DB Empty the data base of all the lot number
2. Input the lot number and the expired data, a dialog box appears when the lot number has
already existed, the user can edit the old lot number according to the actual condition.
3. Select the QC level.
Page 43 of 106
4. Edit the mean and the 2SD values. Please ensure the effective lot number and expired date
have been input before edit or the edit cannot be done. Click “Save” when all the data are
input, a dialog box saying “Save current changes and exit?” appear, click “OK” to save.
5. The interface of Level 1 to level 3 are very similar, the status bar which level you are editing.
Here we take level 1 as reference to describe the interface.
6.5.2 QC test
Click”QC”, select an effective lot number as the current lot number, then click”Go” the test
interface appears on the screen. The status bar at the left corner shows the current status of
the system.
The QC pack should be stored in room temperature.
Don't shake the QC solution before openning, making the
solution keep stable.
Push off two drops of the solution form the syringe before
feeding the it to the analyzer.
Page 44 of 106
6.7 Maintenance menu
Click “Maintenance button, users can perform the “Clean” “Deproteinize” and “Condition”
according to the note.
There's no need to aspirate clean solution when conduct clean because specific clean
solution existed in the reagent pack.
But users need to feed solution to the sample port when conducting ”Deproteinize” and
“condition”. Users can also set the maintenance period.
Page 45 of 106
For more detailed information, please refer to chapter 10.
2. The barcode need to be scanned when the expendable is expired, click “Enter Barcode” to
scan the barcode of new expendable or you can input the barcode manually. The barcode of
every reagent pack just corresponding to one client code, the reagent pack which doesn't
match the client code cannot be used, and the related test cannot proceed.
3. Click “List” the detailed information of the expendable appears on the screen.
Page 46 of 106
7 Calibration
7.1 General information
Sensor calibration is the process of relating sensor electrical outputs to know analyte
values. Calibration establishes a relationship between the electrical output of a sensor and
the concentration of the analyte measured by the sensor. The relationship between the
sensor signal and the concentration of a measured analyte is linear. Once the relationship is
established, concentration of sample can be calculated once its response is measured.
mV
Vb
Va
Ca Cb mmol/L
Hct just needs one point calibration, and Hct should be separated from others.
Page 48 of 106
There are 4 possible results of 1 point calibration:
Pass Calibration is passed
The mV difference from two consecutive tests exceeds the limit. It means
Drift
electrode is unstable.
Unstable Unstable mV reading within limited time.
Over Range mV exceeds the limit range of electrodes.
“Calibrating A1”: Status area. Here “A1” means the first trial of 1 point calibration.
If calibration A1 is failed, the system proceeds calibration A2 automactically.
“00:00”. Calibration timer.
“Voltage” : Electrode response in mV format.
“Concentration” : Concentration from Cal A in the format of mmol/L for
K/Na/Cl/Ca/Glu/Lac and mmHg (or kPa) for pCO2 and pO2.
“Status”: Calibration Status. When calibration is failed, electrode problem is displayed.
Page 49 of 106
2. Cal A, Cal E and Cal B are aspirated respectively for test. Before and after each test, flush
solution is aspirated to wash flow path.
3. If calibration A1 failed, the system proceeds calibration A2. Each test will try 3 times until
it is passed. “2 point calibration finished” is present after 3 trials regardless of the
calibration result.
4. If both Cal A, Cal E and Cal B is passed, Return button is displayed. The result windows
will be present for 2 minutes if without user interruption.
Page 50 of 106
Abnormal Slope(mVb-mVa) is out of limit range. It means a narrow linearity range.
1. Enter System setup and select Calibration intervals. The system default interval for one
time calibration is 60 minutes, the two point calibration interval is always four times of the
on point calibration interval.
2. There are four options available: 60 minutes, 120 minutes, 180 minutes and 240 minutes.
Click the circle before option to select one option and re-click to deselect.
3. Press OK to save changes.
Page 51 of 106
2 point Calibration will be performed after 3 times of 1 point calibration in auto
calibration mode.
1 In Test window,make sure pO2 Calibration is passed,prepare anaerobic water and aspirate it
with syringe.
2 Insert syringe to sample port,test it twice and put down the Voltage(mV) of pO2.
3 Select pO2 zero calibration under Other Utilities,enter the average of two results.
4 After calibration, the results window is shown as below.
Calibration data are automatically kept by the system. This function provides user a useful
method to learn about performance and status of electrodes especially when there is an
electrode problem.
Page 52 of 106
1 Select Cal Results under Data Manager.
2 Enter date range for searching. Press Search to start search. Available calibration data are
displayed on the right window. Click function extended more function icons will appear,
click the corresponding icon to perform functions.
3 Select the Cal you need and click more to get more information, slope=Cal B - Cal A.
Click print to print the Cal result.
Page 53 of 106
Cal type Description
A 1,2 or 3 First ,second or third trials of Cal A
E 1,2 or 3 First ,second or third trials of Cal E
B 1,2 or 3 First ,second or third trials of Cal B
8 Quality Control
8.1 General information
Quality control is essential to monitor the accuracy and precision of the complete
analytical system to detect immediate errors due to system failure, adverse environment
conditions, and operator performance.
Quality control includes 3 levels of solutions to covers the entire clinically significant
range: low(Acidosis), Middle(Normal) and high(Alkalosis). Quality control must be
performed before any patient sample testing and they are required for every 8 hours of
patient testing. Each lab should establish its own QC program. 2 level of QC are required
for every 4 hours and a third for 24 hours. 3 levels of aqueous Blood gas controls and 2
levels of hematocrit controls are required to be performed weekly on this system.
Additional quality control should be run after any troubleshooting or maintenance which
might alter performance.
Page 54 of 106
subsequently selection.
1. Click “Add” on the lot number management interface, the interface for inputting lot number
and related figures shown as below:
Page 55 of 106
2. Input the lot number and expired date, if the input number is the same with an exist one, a
warning will pop up to remind you to edit the existed lot number according to the real
condition.
4. Edit the “Mean” and “2SD” figure, please make sure an valid lot number and expired date
have been input already. Click “OK” to save it.
Page 56 of 106
Range = 2SD. Normally it should be equal to Mean - Min or Max - Mean
8.4 QC test
1. Select the lot number and the QC level, click “Go”, a dialog box saying “Place qc sample in
the sample port. Press ok to continue!” appears , put the syringe containing QC solution at
the sample port.
2. Remove the syringe when the aspiration is finished. Then the system starts to test, the
picture shown as below, the status of the test shown at the left top of the interface.
3. The system starts to clean the flow path after test, Click “Print” to print the result of the QC
test.
Page 57 of 106
8.5 QC plot data
This function can manage QC data stored on this system. QC data can be searched by
specifying a data range. Listed records can be further viewed by specifying levels.
1. Select QC plot Data under Data manager. Input data range and select the level ,the related
QC plot data appears.
2. Select an item, click “More” to see the detailed information. Click “print” to print the
result out.
Page 58 of 106
8.6 QC plot
The system can plot one recent QC result of one day stored in the current QC file for each
parameter and level, and this system can plot the recent 30 days QC results in total. Click
“ Data Manager” “ QC Plot” to check the QC plot, the picture shown as below:
Page 60 of 106
During storage blood cells tend to settle. If complete mixing is not
achieved before analysis, results may be significantly different from
actual values.
Expel bubbles out from the syringe.
2. Prepare the sample and insert the syringe device into the sample port firstly, click ”Go”, the
system starts to aspirates the sample.
Page 61 of 106
3. If the sampler lever detects no sample, the system will give out continuous beeps reminding
you to put the sample at the sample port. Click return to start again.
4. Remove the sample device when aspiration finished, then you can see a series motion of the
system. Meanwhile you can input the corresponding sample information of the patient, click
“Confirm input”.
Page 62 of 106
5. The system perform washing at the end of the analysis and display the related data on the
screen. When the test is finished the system will return back to test interface after 3 minutes.
If you don't remove the sample after aspiration, the washing for sample
port would pollute the sample.
Aspiration for any samples on the system is automatic. Never inject
samples under any circumstances.
2 Click More to get the detailed information of the result, click print to print the result
through the internal thermal. And if other formats (A4 or A5) of the result is needed please
connect the analyzer to an external printer to print.
Page 63 of 106
3 Click “Analysis Report”, an analysis report of the patient appears for your reference.
Page 64 of 106
10 Setup system
This chapter shows useful information about system setup. Use these function to configure
and control your system. Most of these functions are required to be set up on first start-up
and not required by normal operation once they have been set up. Change system setup
only when necessary.
10.1 Correlation
When test results is other than target value or methodology, it can be corrected by using
correlation coefficients through slope and intercept by using a model of y=ax+b where a is
slope, b is intercept, x is test results and y is target value. Only mathematical calculation is
involved with correlation. The slope and intercept are got from a regression analysis of test
results against results from reference system.
1 Click “Correlation” input the pass word “77”, the picture appears as below, user s can also
edit the data and click “OK” to save.
2 The first column is Slope (default 1.0), and the second column shows Intercept (default
0.0).
3 Click data field to activate virtual keypad. Change to numeric layout and input correct
numbers. Press OK when finished.
4 Click another data field to finish another input.
5 Press OK to save and press Return to exit.
Page 65 of 106
Correlation coefficients must be re-calculated if new electrode is replaced.
Page 66 of 106
Each lab should establish its own reference range for the evaluation of patent
results.
↑↓will be printed on test reports if the value exceeds reference range limit.
10.3 Unit
Click "Unit" to select unit (mmHg or kPa)for the values according to your preference.
Click “Edit” and input the password “77”, user can set the Hb calculation value.(Hb value
ranging from 0.10 to 0. 99).
10.8 Version
The version of software appears as below:
Page 69 of 106
10.9 Replace Sample port
Click “Replace Sample port”, the probe will go backward, which is convenient to take
down the sample port. Click OK to check if the sample port is replaced ok.
Page 70 of 106
1. Click “HL7 Server Setting” to set server IP and server port, select “Enabled”and click
“save&Link” .
2. Click “Local Network setting”, enter the IP address to establish a connection to LIS.
Page 71 of 106
10.11 Printer
This menu controls the setup of printer.
1 Select Printer.
2 There are two options to control the printer and test report.
1. Click “Select Item” to choose the parameters needed to print(for USB printer only).
Page 72 of 106
10.12 Self test
Click “Self Test” the system will start to perform self test.
Page 73 of 106
10.14 Other utilities
Click “other utilities” to enter this interface, click the corresponding icon to set the
related value.
11 Maintenance
This section includes the recommended maintenance procedures for the analyzer. The
frequency of preventive maintenance operations is based on average workload of 20 to 30
samples per day with a normal analyzer use. Facilities with heavier workloads should
schedule maintenance operations more frequently.
Page 74 of 106
Wear disposable gloves to avoid contact with potentially infectious materials
while maintaining the analyzer.
11.1.1 Clean
Replace the Glu and Lac electrode with Maintenance electrode C before cleaning the
sample path. Reinstall them within 2 hours.
Clean can remove fat built up in the sample pathway. It can also be used to maintain
electrodes. Clean solution from Cal pack is used in this procedure.
1. Select Clean under Maintenance.
2. Clean solution is then aspirated to clean the pathway.
3. Clean procedure lasts around 1 minute. Press Return for an early exit.
4. 2-PT calibration is performed automatically after performing Clean.
Page 75 of 106
11.1.2 Deproteinize
Replace the Glu and Lac electrode with Maintenance electrode C before cleaning the
sample path. Reinstall them within 2 hours.
Use Deproteinize to remove protein in sample path. It can also be used to maintain
electrodes.
1. Select Deproteinize under Maintenance.
2. Prepare the Deproteinizer
Take out one pair of enzyme and dilutor. Add the dilutor into the enzyme. Shake the vial
for several times then wait for 2 minutes until the enzyme power is completely dissolved. It
should be a clear solution.
3. Aspirate Deproteinizer with a syringe and insert the syringe into sample port.
4. Press OK to aspirate.
5. The whole procedure lasts 30 minutes. Press Return for an early exit.
6. 2-PT calibration is performed automatically after exiting Deproteinize.
Page 76 of 106
Feed fresh serum instead of deproteinizer when activating new electrodes.
Manual rotate reagent pump until serum is moved out of PCO2 and PO2 area
since those two electrodes do not need activation.
11.1.3 Condition
Replace the Glu and Lac electrode with Maintenance electrode C before cleaning the
sample path. Reinstall them within 2 hours.
Conditioning is effective to Na+ and pH electrodes. Perform it only when Na+ or pH has
problem.
1. Select Condition under Maintenance.
2. Take out one piece of conditioner and transfer condtioner to a syringe. Then insert it into
sample port. Press OK to start.
3. The condition procedure lasts 5 minutes. Press NO to stop if an early exit is desired. 2 PT
calibration is automatically performed after exiting Condition.
Page 77 of 106
11.2 Sterilization
The purpose of sterilization is to minimize the danger of infection when contacting with
blood-contaminated parts.
The sterilization should be performed routinely.
The operator is recommended to comply with sterilization procedures and special
requirement of lab.
Only use liquid disinfector such as 2% hydrogen peroxide. Never use organic
solution or alcohol to clean or sterilize the surface.
Do not pour any liquid such as the disinfector directly on the surface, or it will
cause electrical short circuit.
Careful cautions should be taken when sterilizing probe to avoid injury and
potential infections.
For K+, Na+, Cl-, Ca2+ and pH electrodes, empty the remaining refill solution
before filling new.
2. Take out reference electrode, remove crystal if there is too much. Refill if the solution is
not enough.
3. Clean all the pathway to avoid blockage.
4. Perform Hct calibration.
5. Perform pO2 calibration.
Page 79 of 106
iv. Remove sample port from housing part and replace with a new one.
v. Install housing part and resume probe and cover.
vi. Enter Other Utilities, select Self test to start self test.
Reinstall sample port and probe if there is problem with reagent test.
11.8 As necessary
The performance of electrodes will decrease with the increasing tests of samples. It needs
routine maintenance to ensure its accuracy and prolong its life time.
1. Clean
Perform this cycle to remove fat colt inside sample path and electrodes. Every 10 samples
to clean once.
2. Deproteinize
Every 30 samples to deproteinize electrodes. Also perform it when K+, Cl-, Ca2+ has
problem.
3. Condition
Perform this cycle when Na or pH has problem.
12 Troubleshooting
This section covers the trouble shooting procedures of the analyzer.
The analyzer can perform self test when power up. It will detect most of the problem
except electrodes.
INTERCEPT=YH-SLOPE*XH Or INTERCEPT=YL-SLOPE*XL
Where: YH, YL is the target value of the High level, Low level QC
XH, XL is the average test value excluding the highest and lowest test value.
The program in the System Setup>>Correlation is used to change slope and intercept
values.
The following procedure is performed on the basis of Quality control materials.
Prepare the Quality Control Materials of 3 levels from same manufacturer.
1. Enter System Setup>>Correlation. Reset slope and intercept as 1.00 and 0.00. Press OK
to save.
2. Open a vial of QC level 1, transfer it to a syringe. Cap the syringe after excluding air
bubble.
Page 83 of 106
3. Insert syringe into sample port.
4. Enter QC test and select level 1 to start test.
5. Open another vial and repeat QC level 1 test for at least 4 times.
6. Repeat step 2-5 for level 2/3.
7. New an excel sheet, input test results and target values for each parameters.
8. Use linearity regression functions provided by excel to calculate slope and intercept.
K_Test K_Target y=ax+b
5.72 5.7 SLOPE INTERCEPT CORRELATION FACTORS
5.71 5.7 1.096444 -0.57389 0.999978
5.72 5.7
5.73 5.7
Note:
3.90 3.7
Function available in EXCEL
3.89 3.7
SLOPE(array y,array x) INTERCEPT(array y,array x)
3.90 3.7
3.90 3.7
3.90 3.7
9. Enter System Setup>>Correlation. Enter respective slope and intercept, press OK to save.
10. Enter QC test and select level 2, test QC level 2 and compare with target value.
11. If the result is still a little low or high, just fine adjust intercept accordingly. After
adjustment, recheck with QC level 2.
Input results from reference system instead of target vaue when comparing
with a reference system.
A Appendix
A.1 Potential range for electrodes
Range limit (mV) mV(CAL B) – mV(CAL A) difference limit
K 45 ~ 140 12~ 21.0
Na 45 ~ 120 -4.2 ~ -7.3
Cl 50 ~ 120 5.4 ~ 10.8
Ca 35 ~ 100 6.6 ~ 10.5
pH 70 ~ 170 16 ~ 28
Li 50 ~ 150 5.0 ~ 9.0
pCO2 500 ~ 2500 -30 ~ 75 mV/Dec
pO2 0 ~ 500 +1.5 ~ 10 pA/mmHg
Page 84 of 106
When response potential of the electrode is outside the normal range, the system prompts
OR error.
When potential difference of Cal B-Cal A falls beyond above limit, the system prompts
abnormal error.
Page 85 of 106
Contamination in container Increase (↑)
Procaine Decrease(↓)
Bromide Increase (↑)
Salicylate Decrease(↓)
Ammonium carbonate Decrease(↓)
Nystatin Decrease(↓)
Amphotericin Decrease(↓)
Lidocaine Decrease(↓)
iCa2+
Thiocyanate Decrease(↓)
Perchlorate Decrease(↓)
Benzalkonium chloride Increase (↑)
Thiopental sodium Decrease(↓)
Acetylcysteine Decrease(↓)
Acetaminophen Decrease(↓)
Magnesium Increase (↑)
Lactate Decrease(↓)
Heparin lithiumanticoagulant Increase (↑)
Li+
Contamination in container Increase (↑)
Thiopental sodium Decrease(↓)
pH
Halothane Decrease(↓)
Thiopental sodium Increase (↑)
pCO2
Halothane Increase (↑)
pO2 Halothane Decrease(↓)
Crystal dilution Decrease(↓)
Hct
Colloid dilution Decrease(↓)
Acetaminophen Increase (↑)
Acetylcysteine Decrease(↓)
Sodium fluoride Decrease(↓)
Potassium oxalate Increase (↑)
Mannose Increase (↑)
Glu
Thiocyanate Increase (↑)
Galactose Increase (↑)
2-deoxy D-Glucose Increase (↑)
Ascorbic acid Increase (↑)
Uric acid Increase (↑)
Lac Glycolic acid Increase (↑)
Page 86 of 106
Thiocyanate Increase (↑)
Acetoacetate Increase (↑)
D-Glucose Decrease(↓)
Sodium fluoride Decrease(↓)
Ascorbic acid Increase (↑)
Uric acid Increase (↑)
Other than above factors, electromagnetic interference, irradiation of strong light, degraded
calibrate, additives and preservatives in the QC materials, imperfection of the grounding,
unstable power supply, dry out of electrode refill solution, chloride layer falling off the
silver stick of the electrode or corrosion and moist of grounding points are all the
inducements to imperfection in calibration and testing.
Page 87 of 106
appropriate action is possible. It is the laboratory’s responsibility to communicate these
values immediately and flawlessly to the responsible clinicians. Each laboratory should
determine its own critical values. The above values are stated for reference purposes only.
A.5 Calculations
A.5.1 Temperature corrections
In order to obtain a more accurate reflection of in vivo conditions, pH, pCO2 and pO2
values are often corrected according to the patient body temperature. The system
automatically performs these calculations according to the experimental relationship
proposed by NCCLS:
pH
-0.0147 0.0065 (7.4 pH)
T
Where pH is that measured at 37℃
Page 88 of 106
log10 pCO 2
=0.019
T
log10 pO2 5.49 10 -11 pO 23.88 +0.071
=
T 9.72 10 -9 pO 23.88 +2.30
3. Total dissolved carbon dioxide is calculated from the measured pCO2 value according to
the equation proposed by NCCLS:
TCO 2 HCO3 0.0307 pCO 2
4. Blood base excess or “in vitro” base excess is defined as the titratable base of blood and
may be experimentally determined by titration with strong acid or base to a plasma pH of
7.400 with pCO2=40 mmHg and at 37.0℃. The analyzer derives this parameter from the
measured pH and pCO2 according to the following equation:
BE b (1 0.014Hb ) (HCO 3 24.8 (1.43Hb 7.7) ( pH 7.4))
The hemoglobin value employed in this calculation is the Hemoglobin value of blood
under examination which is known and entered by the operator via keyboard as the input
parameters of the patient prior to measuring.
5. Base excess of extracellular fluid (or in vivo base excess or standard base excess) is a
measure of the metabolic component of the acid-base balance in a patient.
This parameter may be determined experimentally titrating a model of extracellular fluid to
a pH of 7.400 with pCO2=40 mmHg and at 37.0℃. It is calculated according to the
equation proposed by NCCLS:
BEecf HCO3 24.8 16.2 (pH 7.4)
Page 89 of 106
6. Standard bicarbonate of blood is defined as the bicarbonate concentration in plasma
obtained from blood which has been equilibrated at 37.0℃ with a gas of pCO2 =40 mmHg
and pO2 higher than 100 mmHg. It is calculated according to the following equation:
SBC 25 0.78BE b 0.002Hb (sO 2% 100)
7. The Anion Gap is the difference in the measured cation and the measured anion in serum,
plasma, or urine. It is calculated according to the following equation :
AG Na K (Cl HCO 3 - )
The anion gap is normally 8-16 mmol/L . It provides information about the level of anions
(including sulfate, phosphate, organic acids such as ketone bodies and lactic acid, and
proteins) that are not routinely measured in laboratory tests.
8 Oxygen saturation measures the percent of hemoglobin which is fully combined with
oxygen. It is calculated from the measured value of pO2, pH and pCO2 according to the
equation proposed by. NCCLS :
( pO2 )3 150 pO2
sO2 %= 100
( pO2 )3 150 pO2 23400
Where:
pO2 pO2 e 2.3026(0.48( pH 7.4)0.0013(HCO 3 25))
Normally, oxygen saturation on room air is in excess of 95%. With deep or rapid breathing,
this can be increased to 98-99%. While breathing oxygen-enriched air (40% - 100%), the
oxygen saturation can be pushed to 100%.
8. The Alveolar-arterial oxygen gradient (A-a gradient) is a measure of the difference
between the alveolar concentration of oxygen and the arterial concentration of oxygen. It is
used in diagnosing the source of hypoxia.
AaDO2 (mmHg ) A pO2 (T )
A normal A-a gradient for a young adult non-smoker breathing air, is between 5-10 mmHg.
Normally, the A-a gradient increases with age. For every decade a person has lived, their
A-a gradient is expected to increase by 1 mmHg.
Page 90 of 106
9. Respiratory index (RI) is a measure of the oxygenation function of lung, and an increase in
RI reflects the presence of pulmonary shunting in a variety of conditions including
atelectasis, pulmonary contusion, and pulmonary thromboembolism. The RI was calculated
from arterial blood gas assay as follows:
AaDO2
RI
pO2
where AaDO2 is the alveolar-arterial oxygen gradient.
10. P50 is defined as the partial pressure of oxygen in a hemoglobin solution having an oxygen
saturation of 50%. For oxygen saturation between 40% to 90%, the relation between PO2
and sO2 is approximately quite well by the Hill equation.
1 sO2
log P50 log pO2 log
2.7 sO2
11. Calcium correction
Ionized calcium values are dependent on sample pH. The calcium value adjusted to pH of
7.40 reflects the true ionized calcium concentration of blood normalized to pH 7.40.
Calcium is corrected according to the following equation.
Ca ++ (7.4) Ca ++ measured 10 0.19(7.4 pHmeasured )
The adjusted calcium value is valid only when pH is between 7.2 and 7.6.
Page 91 of 106
A.6.1 ISE electrodes
Ion measurement is performed using two separated electrodes: ISE electrode and a
reference electrode. Each electrode represents a half-cell in which an electrical potential is
developed. ISE electrode is a silver-silver chloride electrode surrounded by a solution of
constant ion concentration and enclosed by a membrane sensitive to specific ion. Reference
electrode provides a stable potential regardless of sample ion. It is a silver-silver chloride
electrode filled with saturated solution potassium chloride. A leaky membrane permits
current flow from reference electrode through sample. Both electrodes are connected to the
measuring chamber and bridged by sample to form a close circuit when a voltmeter is
connected
Potential response of electrode is corresponding to ion concentration in the sample. It can
be expressed by the following equations:
Ecell = Em - Eref -Ej
Where
Ecell is electrode potential response
Em is measured electrode potential
Eref is measured electrode potential
Ej is liquid junction potential
pH electrode
The relationship between ion activity and potential is expressed by Nernst equation
2.303RT
E E0 logC xf x
nF
Where:
E= electric potential of ion-selective electrode in the solution being measured
E0= standard electrode potential of ion-selective electrode
n =electrovalence of the ion being measured
R=gas constant (8.314 J/K.mol)
Page 92 of 106
T=absolute temperature (273+t℃)
F=Faraday constant (96487 c/mol)
Cx=concentration of the ion being measured
fx=activity coefficient of the ion being measured
In given conditions such as room temperature, Nernst equation shows that electrode
potential of ion-selective electrode is linear to the logarithm of the activity (or
concentration) of the ion being measured.
The output of this modified pH electrode is proportional to the pCO2 present in the sample.
pCO2 electrode
Page 93 of 106
The circuit is completed when silver is oxidized at the anode:
The current developed by these reactions is directly proportional to the pO2 in the sample.
pO2 eletrode
A.7 References
The section lists the references for this manual.
Page 94 of 106
1. Clinical and Laboratory Standard Institution. Blood Gas and pH Analysis and Related
Measurements; Approved Guideline-Second Edition. CLSI document. C46-A2 (Vol 29, No
9); 2009
2. Clinical and Laboratory Standard Institution. Definitions of Quantities and Conventions
Related to Blood pH and Gas Analysis; Approved Standard;. CLSI document. C12-A (Vol
14, No 11); 1994
3. Clinical and Laboratory Standard Institution. Standardization of Sodium and Potassium
Ion-Selective Electrode Systems to the Flame Photometric Reference Method; Approved
Standard-Second Edition. CLSI document. C29-A2 (Vol 20, No 17); 2000
4. Clinical and Laboratory Standard Institution. Ionized Calcium Determinations:
Precollection Variables, Specimen Choice, Collection,and Handling; Approved Guideline-
Second Edition. CLSI document. C31-A2 (Vol 21, No 10); 2001
5. Clinical and Laboratory Standard Institution. Protection of laboratory Workers from
analyzer Biohazards and Infectious Disease Transmitted by Blood, Body Fluids and Tissue;
Approved Guideline-Third Edition. CLSI document. M29-A3 (Vol 25, No 10); 2005
6. Clinical and Laboratory Standard Institution. Clinical laboratory Waste Management;
Approved Guideline-Second Edition. CLSI document. GP05-A2 (Vol 22, No 3); 2002
7. Carl A. Burtis PhD, David E. Bruns MD. In:Tietz NW, editor, Fundamentals of clinical
chemistry, 6th Edition. Philadelphia: WB Saunders,2007
8. Severinghaus JW, Bradley AF. Electrodes for Blood pO2 and pCO2 determination. J Appl
Physiol 1968;13:515-520
A.8 Warranty
A.8.1 Warranty guidelines
Warranty period for analyzer is 1 year from the shipping date.
Warranty period for K+/Na+/Cl-/Ca2+/pH/Hct /TH electrodes are 1 year from the shipping
date,
Warranty period for pO2/ pCO2 are 6 months from the shipping date.
BioBase undertakes to provide free repair, and parts replacement for products within the
warranty period.
BioBase undertakes to provide the supply of spare parts (including Standard solution,
electrode) and provide life-long maintenance services.
BioBase is responsible for safety, reliability and performance of this product only on the
conditions that:
Page 95 of 106
1. The electrical installation of the relevant room complies with the applicable national and
local requirements; the product is used in accordance with the operation instructions.
2. Any problems arising from the product defects or improperly packing or other
consequences in connection with the product quality.
Page 96 of 106
Please pay attention to the type of Barcode scanner, there are three types: LV880,
LV880(32B) and NLS-HR100.
If “Barcode error!” appears when scanning the barcode of consumable, please initialize the
scanner according to the following steps:
1.Make sure the scanner is well connected to analyser;
2.Please scanner the barcode as below one by one:
Steps Barcodes
1.Scan function: initialization Settings
0B
0B142
4.Scan interface - Serial port: 19200 baud
rate
Page 97 of 106
000706
6.Data Bits: 8
00081
7.Stop Bits: 1
00090
8.Parity: None
00100
3. After initialization, scan the barcode of consumable again. If it does not works,please
connect representative.
If “Barcode error!” appears when scanning the barcode of consumable, please initialize the
scanner according to the following steps:
1.Make sure the scanner is well connected to analyser;
2.Please scanner the barcode as below one by one:
Page 98 of 106
Steps Barcode
1.Scan function: initialization Settings
0B
000601
0B142
4.Scan interface - Serial port: 19200
baud rate
000706
5.Handshake protocol: None
001200
6.Data Bits: 8
00081
7.Stop Bits: 1
00090
8.Parity: None
001000
3. After initialization, scan the barcode of consumable again. If it does not works,please
connect representative.
Page 99 of 106
If “Barcode error!” appears when scanning the barcode of consumable, please initialize the
scanner according to the following steps:
1. Make sure the scanner is well connected to analyzer;
2. Please scanner the barcode as below one by one:
Steps Barcodes
1.Code Programming ON
2.Baud Rate:19200
3. After initialization, scan the barcode of consumable again. If it does not works,please
connect representative.
Index
Page 100 of 106
1 M
1 Point Calibration.......................................
Calibration 50 Maintenance
2 Every 6 months........................................82
Monthly maintenance.............................. 81
2 Point Calibration.......................................
Calibration 51
probe wiper..............................................81
A Sterilization............................................. 80
Abnormal............................................... 53, 55 Maintenance functions
Ambient temperature....................................20 Condition................................................. 79
Arterial blood............................................... 11 Mixed venous blood..................................... 11
Atmosphere pressure....................................21 Model family..................................................1
family
B O
Battery..........................................................20 Operating and safety information..................
information 7
C Operating information
Maintenance......................................
Maintenance 10, 13
Cal A............................................................ 50
Operating environment..............................7
environment
Cal B............................................................ 51
Prevention of EMI.....................................
EMI 7
Calculate parameters.................................... 91
Operation environment................................ 20
Calibration data............................................
data 54
Over Range................................ 50, 51, 52, 55
Calibration intervals.....................................53
intervals
Capillary blood.............................................11 P
Capillary tube...............................................13 Pass...............................................................50
Concentration............................................... 51 Patient sample analysis................................
analysis 62
Correlation coefficients.............................85 pO2 zero calibration.....................................54
calibration
D Power supply................................................20
Principle
Deproteinize................................................. 14
Hct electrode......................................96, 97
Dimension.................................................... 20
pO2 electrode.......................................... 96
Drift............................................50, 51, 52, 55
Printer...........................................................21
F Full graphics............................................ 21
Fuse.............................................................. 20 Paper........................................................21
G Printing speed.......................................... 21
Printer...........................................................74
Printer
Green............................................................61
Pump
I Aspiration pump...................................... 31
Installation Flush pump.............................................. 31
Auto QC installation................................33
installation Waste pump............................................. 31
Cal pack installation................................
installation 31 R
Check.......................................................36
Check
Reagent
Electrode preparation
Auto QC cartridge................................... 14
ISE electrodes..................................... 25
Cal pack...................................................13
Reference electrode............................ 26
Conditioner..............................................14
Printer paper installation.........................
installation 34
Deproteinizer........................................... 14
Intended use...................................................
use 1
Quality controls....................................... 14
Intercept....................................................... 67
Refill solution For ISE............................ 14
Interface....................................................... 21
Refill solution for REF............................ 14
Barcode....................................................21
Red............................................................... 61
RS 232..................................................... 21
Reference electrode......................................26
TCP/IP..................................................... 21
Reference Range..........................................
Range 68
USB......................................................... 21
References.................................................... 97
Interference substances................................ 87
Relative humidity
Page 101 of 106
Operation................................................. 20 Drift.................................................... 84
Storage.....................................................21 OR.......................................................84
S Single ISE
Abnormal............................................ 83
Sample sources.............................................11
Drift.................................................... 83
Sample volume.............................................21
OR.......................................................83
Screen.......................................................... 21
Self test.........................................................37
test U
Shut down....................................................
down 38 Unit...............................................................69
Unit
single use................................................33, 34 Unstable........................................... 50, 51, 52
Slope.............................................................67 V
Specifications
Venous blood................................................ 11
Calculated parameters............................. 19
virtual keypad...............................................40
Electrode Parameters...............................20
Voltage..........................................................51
Input parameters...................................... 20
Specifications..........................................
Specifications 20 W
Status............................................................51 Warnings and precautions
T Biological hazards.....................................
hazards 8
Chemical hazards......................................
hazards 9
Temperature
Cleaning and sterilization..........................9
sterilization
Operating environment............................20
Electric shock............................................
shock 8
Storage.....................................................21
Personnel injuries......................................
injuries 8
Troubleshooting
Waste hazards............................................9
hazards
Blood gas
Warranty.......................................................98
Abnormal............................................ 84
Weight.......................................................... 20
Drift.................................................... 84
Within 24 hour............................................. 38
OR.......................................................84
Working principle.........................................93
Liquid pathway problem
No Cal A............................................. 83 Y
No Cal B............................................. 83 Yellow.......................................................... 61
No flush.............................................. 83
More than one ISE
Abnormal............................................ 84