Professional Documents
Culture Documents
APPENDIX 6
KEY CLINICAL EQUIPMENT DESIGN DATA
[Drafting Note: This design data is exemplar information only provided to enable Participants to evaluate typical requirements. It is the Authority’s intention to refresh and clarify its approach to these as soon
as possible in the Dialogue Period, as procurements progress.]
IMG9025A Digital R/F (XRAY- FLUOROSCOPY) Fluoroscopy/General X-Ray Room 3 ph. 400 50Hz 10 to 35 °C 20% to 75% Exam Room: 7kW
System, with Wall Stand, complete unit. x1 Minimum: 18kVA Temperature Control Room: 3kW
Peak, short duration: 160kVa should not
vary more than
5°C/hr
IMG141A IMAGING, complete Digital Radiography General Purpose X-Ray Room x 1 1. 380/400/415/440/460/480 VAC 3-Phase and 15 to 35 °C 20% to 50% 9kW Needs ventilation for cooling and possible
Unit, with Wall stand and Table. earth without neutral Temperature patient comfort.
2. Daily Voltage variations +/- 10% (VAC) any de- should not
rating in accuracy. vary more than
3. Nominal line frequency (Hz) 50 Hz 10°C/hr
4. Daily frequency variation (Hz) +/- 3%
5. 4.3 kW Continuous Power
6. 97kW Peak Power (Duration is 22 seconds
maximum)
IMG081 IMAGER, MAGNETIC RESONANCE MRI Scanner 1.5 Tesla x 2 1. Voltage: TN-S 400 V + 10% / -10%, 3 ~ 18 to 22 °C 40 to 60 % Examination room: ≤ 4 kW RF Shielding
IMAGING (MRI), 1.5 Tesla. 2. 50 Hz ± 1 Hz Control Room: ≤ 3 kW An RF-shielding (faraday cage) is required for
3. Line to line unbalanced: max. 2 % Equipment Room: ≤ 2 kW the MR-examination room. This shielding
4. Internal line impedance: < 110 mΩ protects the environment from RF interference
5. Power requirement: 84 kVA An air handling system is and conversely protects the MR system from
6. Momentary power: 91 kVA required to meet the specified external interference.
7. Measurement sequences < 3 s conditions. A separate air Required attenuation: >90 dB over the
8. Rated fuse current:160A (NH-Type) handling system for the frequency range 15 to 128 MHz (>100 dB at Co-
equipment room is not Siting).
necessary if an air exchange, These values must be certified by measuring
e.g. natural air convection, is before the MR system is installed.
guaranteed.
RF Doors
All RF-doors leading into the examination room
have to be equipped with a door switch for
indicating the closed / open position of the RF-
door! It must be possible to lock the RF-doors
from the outside. In addition to that it must
always be possible to open the RF-door without
Flooring
An antistatic floor covering is necessary. The
floor in the vicinity of the magnet and patient
table (2.6 x 4.8 m) must be levelled to within
max. ± 2 mm.
The loading capacity of the flooring must be
designed with the weight of the respective
system components in mind.
Cryogens
Liquid helium (He) and also helium gas are
required for operation of the superconducting
magnet. The transport of these liquid gases to
the examination room requires the use of
special vessels. The size and weight of the
vessel should be checked with the local cryogen
supplier.
Quench Pipe
A thermally insulated tube (quench pipe) made
of non-magnetic metal must be fitted from the
super-conducting magnet to the highest point
of the building in order to vent the vaporizing
helium gas.
.
IMG061A IMAGER, COMPUTER TOMOGRAPHY CT Scanner x 2 Voltage 380 to 480 VAC 18 to 26 °C, 30% to 60%. Exam Room: Up to 15kW (Air Ventilation with humidity control may be
(CT), 128 slice unit Capacity 150 kVA nominal 22⁰C Max rate of cooled system) required
RDT011 Frequency 50Hz ± 3 Hz Temperature change 5% per Control Room (incl. 3
IMAGER, COMPUTER TOMOGRAPHY CT Simulator Room x 2 • Maximum power demand = 150 kVA @ 0.85 PF should not hour workstations): <5kw
(CT), 64 slice unit at a selected technique of 140 kV, 715 mA. vary more than Equipment Room:
• Average (continuous) power demand at 3º C/hr Air cooled system: 3kW
maximum duty cycle = 25 kVA. Water Cooled System: Up to
The main disconnect device must provide 15kW
overcurrent protection for the system and
Key Clinical Equipment Design Data v4 Page 4 of 8 27-04-18
SCHEDULE 11
APPENDIX 6
KEY CLINICAL EQUIPMENT DESIGN DATA
[Drafting Note: This design data is exemplar information only provided to enable Participants to evaluate typical requirements. It is the Authority’s intention to refresh and clarify its approach to these as soon
as possible in the Dialogue Period, as procurements progress.]
RDT031 ORTHOVOLTAGE Unit Superficial Systems Room x 1 Main supply voltage 220VAC, +/- 10%, 47-63Hz, Ambient Humidity 6kW System ceiling mounted with free standing
Single phase** Power 6kVA temperature of limit, (non- generator cabinet.
Maximum Auxiliary supply voltage 220VAC, +/- operation 10- condensing).
10%, 47-63Hz, Single phase** Power 100VA ** 30°C Ambient 85%
Note= Derived from same phase
ROB9200 ROBOTIC PHARMACY Central Dispensing Area x 1 240 V, +/-6%. Operating 20 to 85% 6kW
Temperature: 5
Key Clinical Equipment Design Data v4 Page 5 of 8 27-04-18
SCHEDULE 11
APPENDIX 6
KEY CLINICAL EQUIPMENT DESIGN DATA
[Drafting Note: This design data is exemplar information only provided to enable Participants to evaluate typical requirements. It is the Authority’s intention to refresh and clarify its approach to these as soon
as possible in the Dialogue Period, as procurements progress.]
RDT022 Linear Accelerator Dual Photon Energies Linear Accelerator and Maze x At least 2150mm x Minimum corridor 3000mm Up to 8000Kg Minimum Treatment Pits may be required under the system
with Electrons including IMRT, EPID, 6 1250mm width 1800mm with no Room floor dimensions in the treatment room and associated
IGRT and Respiratory Gating restrictions to width at 8000mm x 12000mm system cabinets with interconnecting
turns which must not floor embedded ducts for cabling.
be greater than 90 Refer to manufacturer’s data for specific
degrees make/ model to be installed for detail..
RDT9021 Linear Accelerator Stereotactic Dual Linear Accelerator and Maze – At least 2150mm x Minimum corridor 3000mm Up to 8000Kg Minimum Treatment Pits may be required under the system
Photon Energies with Electrons including Stereotactic x 2 1250mm width 1800mm with no Room floor dimensions in the treatment room and associated
IMRT and EPID restrictions to width at 8000mm x 12000mm system cabinets with interconnecting
turns which must not floor embedded ducts for cabling.
be greater than 90 Refer to manufacturer’s data for specific
degrees make/ model to be installed for detail..
IMG9327 IMAGER SCANNER POSITRON EMISSION PET/CT Scanning Room x 1 – 2100mm x 1200mm Minimum corridor 2700mm 4000Kg 4000mm x 6500mm
TOMOGRAPHY (PET), incorporating A/V shelled space width 1800mm with no
link to control room. restrictions to width at
turns which must not
be greater than 90
degrees
IMG071-01 IMAGER GAMMA CAMERA, digital, multi SPECT CT Room x 1 2100mm x 1600mm Minimum corridor 2700mm 5000Kg 6500mm x 4000mm
detector/ SPECT CT width 1800mm with no
restrictions to width at
turns which must not
be greater than 90
degrees
IMG9025A Digital R/F (XRAY- FLUOROSCOPY) Fluoroscopy/General X-Ray Room 2200mm x 2300mm Minimum corridor 2700mm 650Kg 5450mmx 4250mm Ceiling mounted ceiling support (e.g.
System, with Wall Stand, complete unit. x1 width 2300 mm with Unistrut) grid frame system required for
no restrictions to width mounting of ceiling suspended items
at turns which must and their traversing rails.