You are on page 1of 5

Original article 295

Continuous mixed venous and central venous oxygen


saturation in cardiac surgery with cardiopulmonary bypass
Pierre-Yves Lequeux, Yves Bouckaert, Hicham Sekkat, Philippe Van der Linden, Constantin Stefanidis,
Chi-Hoang Huynh, Gilbert Bejjani and Philippe Bredas

Background and objective Replacing mixed venous oxygen agreement of 13.6 and R22.5%, respectively. Trends of SvO2
saturation (SvO2) monitoring by central venous oxygen and ScvO2 values followed very different patterns for some
saturation (ScvO2) monitoring in order to avoid the use of a patients. Surgery, cardiopulmonary bypass and anaesthesia
pulmonary artery catheter and its related complications is drugs did not influence the relationship between the two
still controversial in the setting of cardiac surgery. The influence methods.
of surgery, cardiopulmonary bypass and anaesthesia drugs on Conclusion Because of the large interindividual variability in the
the relationship between SvO2 and ScvO2 has never been difference between SvO2 and ScvO2, the measure of ScvO2
studied. should not replace the measure of SvO2 with a pulmonary artery
Methods Fifteen patients scheduled for cardiac surgery with catheter for the management of patients undergoing cardiac
cardiopulmonary bypass were included in the study. SvO2 (from surgery with cardiopulmonary bypass.
the pulmonary artery) and ScvO2 (from the superior vena cava) Eur J Anaesthesiol 2010;27:295–299
were continuously measured with fibre-optic catheters from Keywords: cardiac, central venous oxygen saturation, mixed venous oxygen
induction of anaesthesia to 24 h postoperatively. saturation, oximetry, surgery
Results A total of 9267 pairs of measurements were recorded. Received 12 May 2009 Revised 27 July 2009
Mean bias between SvO2 and ScvO2 was 4.4% with limits of Accepted 29 July 2009

Introduction operative period, which is characterized by rapid and


Mixed venous oxygen saturation (SvO2) is routinely profound haemodynamic changes, has never been studied.
used to assess the oxygen supply/uptake balance in In addition, the effect of CPB on the SvO2 –ScvO2 corre-
patients undergoing cardiac surgery with cardiopulmon- lation remains largely unexplored.
ary bypass (CPB).1 However, SvO2 measurements can
Most of the published studies compared paired measure-
only be performed with a pulmonary artery catheter
ments of SvO2 and ScvO2 obtained from blood samples,
which also measures pulmonary pressures and cardiac
but only three of them used fibre-optic measure-
output.2 These catheters are associated with potential
ments.14,15,18 Using continuous fibre-optic measure-
severe complications and their usefulness in the manage-
ments14,18 is more similar to clinical practice and allows
ment of haemodynamically unstable patients is still
for a large number of comparisons.
debated.3–5 Cardiac output measurements can be per-
formed without pulmonary artery catheters with less Factors that influence the correlation between the two
invasive monitoring.6 Central venous oxygen saturation measurements remain largely undefined. The haemato-
(ScvO2), however, can be obtained from a central venous crit level,14 the blood pH,14 the patient’s temperature14
catheter that is easier to place and has fewer compli- and even the cardiac index17 do not seem to influence this
cations than a pulmonary artery catheter. The difference relationship. During surgical procedures, other factors
between both measurements is that ScvO2 analyses blood may influence the correlation between SvO2 and ScvO2,
only from the superior vena cava, whereas SvO2 also takes such as the use of anaesthetic drugs. Indeed, anaesthesia
account of blood from the inferior vena cava and the induces a decrease in cerebral oxygen consumption and
coronary sinus. to a lesser extent in cerebral blood flow, resulting in a
positive balance between oxygen supply and oxygen
Several studies investigated the correlation between SvO2
demand from the brain compared with the awake state,20
and ScvO27–19 with contradictory results. Some of these
which may influence the correlation between SvO2 and
studies were performed in cardiac surgery patients7–11 but
ScvO2. This effect of anaesthetic drugs on the SvO2 –
only in the preoperative (at initial catheter placement)11 or
ScvO2 relationship has never been studied.
in the postoperative period.7–10 Surprisingly, the intra-
The goal of this study was to evaluate the correlation
From the Department of Anesthesiology (P-YL, PB), Intensive Care Unit, CHU- and agreement between continuous fibre-optic SvO2
Tivoli, La Louviere (YB), Department of Anesthesiology, CHU-Brugmann, Brussels
(HS, PVdL), Department of Cardiac Surgery, CHU-Tivoli, La Louviere (CS, C-HH) and ScvO2 measurements in order to evaluate their
and Department of Anesthesiology, Erasme Hospital, ULB, Brussels (GB), interchangeability during cardiac surgery with CPB
Belgium
and to assess the still unexplored effects of anaesthesia,
Correspondence to Dr Pierre-Yves Lequeux, MD, Department of Anesthesiology,
CHU-Tivoli, La Louviere, Belgium
surgery and CPB on the relationship between the two
Tel: +32 64276111; fax: +32 64276529; e-mail: pilequeu@ulb.ac.be methods.
0265-0215 ß 2010 Copyright European Society of Anaesthesiology DOI:10.1097/EJA.0b013e3283315ad0

Copyright © European Society of Anaesthesiology. Unauthorized reproduction of this article is prohibited.


296 Lequeux et al.

Methods monitors were calibrated after initial catheter placement,


Patient population after CPB and upon arrival in the ICU with concomitant
After approval from institutional review board and blood samples (1 ml) from the pulmonary artery and the
patients’ written and informed consent, 15 patients central venous catheters after withdrawal of dead space
scheduled for cardiac surgery with CPB were enrolled flushing fluid. Blood samples were analysed with the
in this experiment. Bayer Blood Gas Analyzer Rapidlab 865 (Bayer Health-
care LLC, East Walpole, Massachusetts, USA) allowing
Exclusion criteria included contraindication to pulmon-
direct measurements of blood oxygen saturation. CPB
ary artery catheter placement or age less than 18 years.
was performed under normothermia with 1800 ml mixed
crystalloid/colloid solution and myocardial preservation
Protocol with 800 ml cold blood cardioplegia (Kalium 30 mEq/l).
Midazolam 5 mg was given orally 30 min before induction Transfusion triggers were classically haematocrit of
of anaesthesia. Upon arrival in the operating room, rou- 21 during surgery, 18 during CPB and 27 in the post-
tine monitoring for cardiac anaesthesia was placed includ- operative period. These were not the only triggers used as
ing five electrodes – electrocardiography, pulse oximetry, the decision to carry out transfusion was also based on
noninvasive blood pressure cuff, capnography, bispectral other parameters such as SvO2 or lactate level. Postopera-
index (BIS) monitoring, 18-G femoral artery catheter for tive sedation was performed with propofol and remifenta-
invasive blood pressure monitoring (Arrow International nil. Remifentanil infusion was gradually decreased and
Inc., Reading, Pennsylvania, USA), urine output monitor- shifted towards a morphine infusion titrated according to
ing, rectal temperature and transoesophageal echocardio- patients’ needs. Propofol infusion was stopped when
graphy. A 16-G catheter (BD Venflon Pro, Becton Dick- patients were deemed ready to recover from anaesthesia.
inson, Helsinborg, Sweden) was inserted in a left forearm Time at spontaneous eyes opening was recorded in
peripheral vein. A spinal anaesthesia was performed with the ICU.
an intrathecal injection of 20 mg sufentanil and 0.5 mg
The intraoperative period was defined as the period from
morphine. Anaesthesia was induced and maintained with
induction of anaesthesia to the end of surgery and the
a target-controlled infusion (TCI) of propofol and remi-
postoperative period as that from arrival to discharge from
fentanil in order to keep BIS values between 40 and
the ICU. The effect of surgery was analysed by compar-
60. Neuromuscular block was achieved with cisatra-
ing intraoperative and postoperative measurements; the
curium. The trachea was then intubated and the lungs
effect of CPB was analysed by comparing prebypass and
were ventilated with a mixture of air and oxygen. A 7F,
postbypass measurements; and the effect of the anaes-
20 cm long three-lumen central venous catheter (Arrow
thetic drugs was analysed by comparing measurements
International Inc.) and a 7.5F pulmonary artery catheter
obtained before patients opened their eyes with those
(Swan-Ganz CCO/SvO2; Edwards Lifesciences, Irvine,
obtained after patients opened their eyes in the ICU.
California, USA) through a 8.5F introducer (Intro-Flex;
Edwards Lifesciences) were then inserted through the
Statistics
right internal jugular vein. Correct tip placement of both
Correlation and agreement between SvO2 and ScvO2
central venous (1–2 cm above the right atrium) and
measurements were calculated with the method
pulmonary artery catheters was confirmed with the ultra-
described by Bland and Altman,21 with limits of agree-
sonography microbubbles test and postoperative chest
ment as the mean difference  1.96 SD.
radiography. The ultrasonography microbubbles test con-
sisted of injecting a solution of saline mixed with air Mean bias and SD between SvO2 and ScvO2 values for
(obtained by rapidly transferring saline from a syringe to each period (intraoperative and postoperative, pre-CPB
another syringe filled with air) through the catheter in and post-CPB, anaesthetized and awake) were calculated
order to visualize its tip. The pulmonary artery catheter for each patient from the Bland and Altman results
was connected to the Vigilance monitor (Edwards Life- (SD ¼ limit of agreement  bias/1.96). Data obtained
sciences) and a CeVOX optic fibre (Pulsion Medical during the different periods were compared with a paired
Systems, Munich, Germany) was inserted through the Student’s t-test.
distal lumen of the central venous catheter and connected
to the CeVOX monitor (Pulsion Medical Systems). SvO2 Results
and ScvO2 values were recorded every minute for 24 h The central venous catheter was misplaced for one patient
except for the CPB period with the Multi Data Logger as confirmed by the intraoperative ultrasonography micro-
software (Edwards Lifesciences) for the SvO2 values and bubbles test and postoperative chest radiography. Results
with the CeVOX Win software (Pulsion Medical Sys- from this patient were not analysed. Results for the pre-
tems) for the ScvO2 values, both running on Windows CPB period from one patient and from the ICU period for
XP Professional (Microsoft Corporation, Redmond, another could not be recorded for technical reasons. One
Washington, USA). The Vigilance (Edwards Life- patient died intraoperatively; therefore, only intraopera-
sciences) and CeVOX (Pulsion Medical Systems) tive measurements could be recorded.
European Journal of Anaesthesiology 2010, Vol 27 No 3

Copyright © European Society of Anaesthesiology. Unauthorized reproduction of this article is prohibited.


SvO2 vs. ScvO2 during cardiac surgery 297

Table 1 Patients characteristics Fig. 2

Weight (kg) 83  22
Height (cm) 168  8
Age (years) 66  16
Sex (male/female) 10/5
CPB duration (min) 120  46
Aortic clamping duration (min) 80  34
Surgery duration (min) 245  71

Data are expressed as mean  SD or ratio (sex). CPB, cardiopulmonary bypass.

Patient characteristics are presented in Table 1.


A total of 9267 pairs of measurements were recorded
(354 for the intraoperative pre-CPB period, 768 for the
intraoperative post-CPB period, 2834 for the postopera- Per patient SvO2 –ScvO2 mean bias and 95% limits of agreement.
tive anaesthetized period and 5311 for the postoperative SvO2 –ScvO2 mean bias and 95% limits of agreement were calculated
by the methods described by Bland and Altman.
awake period). The mean bias between SvO2 and ScvO2
(for the 9267 pairs of measurement) was 4.4% and the
limits of agreement were 13.6 and þ22.5%, respectively
(Fig. 1). The results of the Bland and Altman tests for
thermodilution cardiac output. But the placement of
each patient are displayed in Fig. 2. The mean bias
these catheters is associated with potential severe com-
between the two methods was >3% in 7284 out of
plications,3 accounting for the popularity of less invasive
9267 measurements (79%). There was no statistical
methods of cardiac output monitoring which measure
difference for SvO2 –ScvO2 mean bias and SD between
cardiac output without a pulmonary artery catheter.6
the intraoperative and postoperative period, the pre-CPB
Because of this, it has been proposed to replace SvO2
and post-CPB period or the anaesthetized and awake
measurements by central venous oxygen saturation
state period (Table 2).
(ScvO2), which can be obtained from a central venous
catheter that is easier to place and has fewer compli-
Discussion cations. Animal studies suggest that ScvO2 correlates well
Continuous SvO2 measurement is a classical method of with SvO2 and could reliably replace it.22,23
monitoring the oxygen supply/uptake balance during Human studies, however, are more contradictory. ScvO2
cardiac surgery with CPB.1 Such measurements are seems to correlate with postoperative complications24,25
performed with a fibre-optic catheter inserted in the and its use for early goal-directed therapy has been shown
pulmonary artery.2 These catheters may also measure to reduce mortality in septic shock26 or to reduce post-
operative complications and length of hospital stay when
Fig. 1
used perioperatively.27 These studies, however, did not
assess interchangeability of both methods but only the
usefulness of ScvO2. Studies comparing ScvO2 and SvO2
are quite contradictory; some report an acceptable bias
between the two methods,15,17,19 whereas others do
not.8–13,16,18 An acceptable bias should be a change in
SvO2 values less than that which would induce a thera-
peutic change. Bendjelid et al.9 considered 3% to be the
maximal acceptable bias. But other authors consider the
bias between the two methods of minor importance,
arguing that ScvO2 could replace SvO2, provided that

Table 2 Effect of surgery, cardiopulmonary bypass and


anaesthesia drugs on SvO2 –ScvO2 bias and SD
Mean bias  SD (%) P

Intraoperative period 3.6  6.0


Postoperative period 4.6  6.5 0.40
Total Bland and Altman SvO2 –ScvO2 comparison. Bland and Altman Pre-CPB period 3.6  3.1
comparison between continuous ScvO2 from a central venous fibre- Post-CPB period 2.0  4.9 0.45
optic catheter and continuous SvO2 from a pulmonary artery fibre-optic Anaesthetized state period 8,2  3,7
catheter. The solid line represents the difference between both Awake state period 3.7  4,6 0.18
measurements (mean bias) and the dotted line represents precision
(95% limits of agreement). Awake state period was defined as beginning at spontaneous eye opening in the
ICU. CPB, cardiopulmonary bypass.

European Journal of Anaesthesiology 2010, Vol 27 No 3

Copyright © European Society of Anaesthesiology. Unauthorized reproduction of this article is prohibited.


298 Lequeux et al.

Fig. 3

SvO2 –ScvO2 24 h time course: example of two patients. (A) Example of trends of SvO2 and ScvO2 values showing a similar pattern with a large
bias (11.2%). (B) Example of trends of SvO2 and ScvO2 values showing different patterns with a small bias (2.8%). (a) Intraoperative
precardiopulmonary bypass period. (b) Intraoperative postcardiopulmonary bypass period. (c) Postoperative anaesthetized period. (d) Postoperative
awake period.

the trends of both methods show a similar pattern what- Indeed, our study was designed to compare ScvO2 and
ever the bias.13,14 SvO2 values but not to evaluate the efficacy of ScvO2
monitoring to guide haemodynamic therapeutics along
Our results show a mean bias of 4.4 and 95% limits of with other parameters (preload parameters, cardiac output,
agreement of 13.6 and þ22.5%, and a mean bias of over blood pH, haemoglobin, lactate levels, etc.). Our results
3% for 79% of the time, thus far above the maximal suggest that SvO2 and ScvO2 values are not equivalent, but
tolerable difference described by Bendjelid et al.9 In a randomized trial comparing cardiac surgery patients
addition, our results reveal that some patients show, monitored with either SvO2 or ScvO2 measurements is
although sometimes with a very large mean bias, a required to evaluate the impact of both types of monitoring
remarkably comparable pattern in the trends of the on clinical outcomes and to better establish their respect-
two methods (Fig. 3a), which suggests that ScvO2 can ive roles in the management of cardiac surgery patients.
be used to adequately replace SvO2, whereas other
patients show trends of SvO2 and ScvO2 of very different Several other studies have been performed in cardiac
patterns. In the example presented in Fig. 3b, ScvO2 surgery patients. Most of them compared SvO2 and
remains very stable, around 70% during 24 h, whereas ScvO2 during the postoperative period7–10 and one during
SvO2 varies widely (down to 45%) over the same period, the preoperative period only (at catheter placement).11 All
which calls into question the reliability of ScvO2 in concluded that both methods are not interchangeable.
detecting changes in tissue oxygenation in the setting However, none of them studied the intraoperative period,
of cardiac surgery at least for some patients. which is characterized by wide haemodynamic changes
that could influence the relationship between SvO2 and
The present study also demonstrates a large interindivi- ScvO2. Our results, however, suggest that neither surgery
dual variability in the bias between SvO2 and ScvO2 that nor CPB significantly influences the relationship between
has never been emphasized in previous studies. Indeed, SvO2 and ScvO2.
mean bias between SvO2 and ScvO2 for each patient
varies from 7.1 to þ16.8%, with large 95% limits of Anaesthesia is another factor that could influence this
agreement for each patient (Table 2). Because the bias SvO2 –ScvO2 relationship. Indeed, most anaesthetic drugs
varies widely from one patient to another and even for a induce a greater decrease in cerebral oxygen consump-
single patient over the course of time, it is almost impos- tion than in cerebral blood flow, resulting in a positive
sible to predict a SvO2 value from a ScvO2 value even if balance between brain oxygen supply and oxygen uptake
the mean difference between the two methods is known. compared with the awake state.20 The higher oxygen
saturation in the blood from the superior vena cava in
However, the usefulness of ScvO2 monitoring in the anaesthetized patients than in awake patients could have
setting of cardiac surgery still remains to be investigated. an effect on the difference between ScvO2 and SvO2, but

European Journal of Anaesthesiology 2010, Vol 27 No 3

Copyright © European Society of Anaesthesiology. Unauthorized reproduction of this article is prohibited.


SvO2 vs. ScvO2 during cardiac surgery 299

our results do not demonstrate it. Factors influencing the 6 De Wilde RB, Schreuder JJ, van den Berg PC, et al. An evaluation of
cardiac output by five arterial pulse contour techniques during cardiac
correlation between SvO2 and ScvO2 thus remain largely surgery. Anaesthesia 2007; 62:760–768.
unknown, as haematocrit, blood pH, patient temperature14 7 Yazigi A, El Khouri C, Jebara S, et al. Comparison of central venous to mixed
and cardiac index17 did not seem to affect this relation- venous oxygen saturation in patients with low cardiac index and filling
pressures after coronary artery surgery. J Cardiothorac Vasc Anesth 2008;
ship either. 22:77–83.
8 Chawla LS, Zia H, Gutierrez G, et al. Lack of equivalence between central
A limit of the present study is the small number of patients and mixed venous oxygen saturation. Chest 2004; 126:1891–1896.
included. However, continuous fibre-optic measurements 9 Bendjelid K, Treggiari MM, Suter PM, et al. Continuous SvO2
measurements and co-oximetry are not interchangeable immediately after
allowed for the collection of large amounts of data per cardiopulmonary bypass. Can J Anaesth 2004; 51:610–615.
patient and a large number of paired SvO2 –ScvO2 10 Pieri M, Brandi LS, Berolini R, et al. Comparison of bench central and mixed
measurements (9267). In addition, the sample was suffi- pulmonary venous oxygen saturation in critically ill postsurgical patients.
Minerva Anestesiol 1995; 61:285–291.
cient to demonstrate a large interindividual variability in 11 Turnaoglu S, Tugrul M, Camci E, et al. Clinical applicability of the
the bias between SvO2 and ScvO2. substitution of mixed venous oxygen saturation with central venous oxygen
saturation. J Cardiothorac Vasc Anesth 2001; 15:574–579.
In conclusion, our study demonstrates that the bias 12 Varpula M, Karlsson S, Ruokonen E, et al. Mixed venous oxygen saturation
between SvO2 and ScvO2 varies widely from one patient cannot be estimated by central venous oxygen saturation in septic shock.
Intensive Care Med 2006; 32:1336–1343.
to another and for a single patient over the course of time 13 Dueck MH, Klimek M, Appenrodt S, et al. Trends but not individual values of
and also that the trends in the two methods follow very central venous oxygen saturation agree with mixed venous oxygen
saturation during varying hemodynamic conditions. Anesthesiology 2005;
different patterns for some patients. In addition, as factors 103:249–257.
such as anaesthesia, surgery and CPB as well as haema- 14 Reinhart K, Kuhn HJ, Hartog C, et al. Continuous central venous and
tocrit, blood pH, patient temperature and cardiac index pulmonary artery oxygen saturation monitoring in the critically ill. Intensive
Care Med 2004; 30:1572–1578.
do not seem to influence the relationship between SvO2 15 Ladakis C, Myrianthefs P, Karabinis A, et al. Central venous and mixed
and ScvO2, the difference observed between the two venous oxygen saturation in critically ill patients. Respiration 2001;
methods still remains unexplained. Therefore, in patients 68:279–285.
16 Edwards JD, Mayall RM. Importance of the sampling site for measurement
undergoing cardiac surgery with CPB, the value of ScvO2 of mixed venous oxygen saturation in shock. Crit Care Med 1998;
obtained from a central venous catheter should not be used 26:1356–1360.
17 Berridge JC. Influence of cardiac output on the correlation between mixed
to assess the value of SvO2 obtained from a pulmonary venous and central venous oxygen saturation. Br J Anaesth 1992;
artery catheter. Further investigations are required to 69:409–410.
evaluate the usefulness of ScvO2 monitoring to guide 18 Martin C, Auffray JP, Badetti C, et al. Monitoring of central venous oxygen
saturation versus mixed venous oxygen saturation in critically ill patients.
the management of cardiac surgery patients. Intensive Care Med 1992; 18:101–104.
19 Tahvanainen J, Meretoja O, Nikki P. Can central venous blood replace
Acknowledgements mixed venous blood samples? Crit Care Med 1982; 10:758–761.
20 Oshima T, Karasawa F, Satoh T. Effects of propofol on cerebral blood flow
The authors thank the medical and nursing team from the Intensive and the metabolic rate of oxygen in humans. Acta Anaesthesiol Scand
Care Unit (CHU-Tivoli, La Louviere, Belgium) for their collabora- 2002; 46:831–835.
tion in collecting data and Pr C. Melot (Intensive Care Unit, Erasme 21 Bland JM, Altman DG. Statistical methods for assessing agreement
Hospital, Brussels, Belgium) for assistance in statistical analysis. between two methods of clinical measurement. Lancet 1986; 1:307–310.
22 Reinhart K, Rudolph T, Bredle DL, et al. Comparison of central-venous to
mixed-venous oxygen saturation during changes in oxygen supply/demand.
References Chest 1989; 95:1216–1221.
1 O’Connor JP, Townsend GE. Pro: perioperative continuous monitoring of 23 Shou H, Perez de Sa V, Larsson A. Central and mixed venous blood oxygen
mixed venous oxygen saturation should be routine during high-risk cardiac correlate well during acute normovolemic hemodilution in anesthetized
surgery. J Cardiothorac Anesth 1990; 4:647–650. pigs. Acta Anaesthesiol Scand 1998; 42:172–177.
2 Swan HJ, Ganz W, Forrester J, et al. Catheterization of the heart in man with 24 Collaborative Study Group on Perioperative ScvO2 Monitoring.
use of a flow-directed balloon-tipped catheter. N Engl J Med 1970; Multicentre study on peri- and postoperative central venous oxygen
283:447–451. saturation in high-risk surgical patients. Crit Care 2006; 10:R158.
3 Abreu AR, Campos MA, Krieger BP. Pulmonary artery rupture induced by a 25 Kusano C, Baba M, Takao S, et al. Oxygen delivery as a factor in the
pulmonary artery catheter: a case report and review of the literature. development of fatal postoperative complications after oesophagectomy.
J Intensive Care Med 2004; 19:291–296. Br J Surg 1997; 84:252–257.
4 Djaiani G, Karski J, Yudin M, et al. Clinical outcomes in patients undergoing 26 Rivers E, Nguyen B, Havstad S, et al., Early Goal-Directed Therapy
elective coronary artery bypass graft surgery with and without utilization of Collaborative Group. Early goal-directed therapy in the treatment of severe
pulmonary artery catheter-generated data. J Cardiothorac Vasc Anesth sepsis and septic shock. N Engl J Med 2001; 345:1368–1377.
2006; 20:307–310. 27 Pearse R, Dawson D, Fawcett J, et al. Early goal-directed therapy after
5 Vincent JL. A reappraisal for the use of pulmonary artery catheters. Crit major surgery reduces complications and duration of hospital stay.
Care 2006; 10 (Suppl 3):S1. A randomised, controlled trial. Crit Care 2005; 9:R687–R693.

European Journal of Anaesthesiology 2010, Vol 27 No 3

Copyright © European Society of Anaesthesiology. Unauthorized reproduction of this article is prohibited.

You might also like