Mars1417XF
User Manual
Before operating, please read this user manual and pay attention to all safety precautions. Please
ensure that user manual is properly maintained so that it can be accessed at any time. Please use
correctly on the basis of full understanding of content.
To Customers
Congratulations on your purchase of Mars1417XF wireless digital flat panel
(hereinafter referred to as Mars1417XF) which is manufactured by iRay Technology
Co. Ltd. (Hereinafter referred to as iRay).
At iRay, we strive to not only make the world-class product that delivers the best
value to our customers but also offer the highest quality of service and customer
care. Please take time to read user guide in order to utilize product effectively. We
hope you enjoy experience with Mars1417XF.
If you have any questions or suggestions, please feel free to contact us.
Service Office
E-mail: service@iraychina.com
(2) This device must accept any interference received, including interference that may cause
undesired operation.
Attention that changes or modification not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Note: This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This product generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this product does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
—Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
About SAR
After the laboratory measurement, the max SAR value is 0.058mW/g for 2.4G WLAN and max
SAR value is 0.277mW/g for 5G WLAN which satisfies the RF exposure requirement.
4. Use only computers and image display monitors complying with IEC 60601-1 or
IEC 60950-1. For details, consult our sales representative or local iRay dealer.
5. Use dedicated cables. Do not use cables other than those supplied with product.
7. Request your sales representative or local iRay dealer to install this product.
Disclaimer
• iRay shall not be liable to any damage, loss, or injury incurred to purchaser
and the third parties as a result of fire, earthquake, any accident, misuse or
abuse of product.
• iRay shall not be liable to any damage, loss, or injury arising from unauthorized
modifications, repairs, or alterations or failure to strictly comply with iRay’s
operation and maintenance instruction.
• iRay shall not be liable to any damage or loss arising from use of any option or
consumer goods other than those dedicated as Original iRay product.
• It is the responsibilities of users and physicians for maintaining the privacy of
image data and providing medical care service. iRay shall not be responsible
for legality of image processing, reading and storage nor shall it be responsible
for loss of image data for any reason.
• Information regarding specification, components, and appearance of product is
subject to change without prior notice.。
Copyright
All rights reserved
No part of this publication may be reproduced in any form or by any means without
the written permission of iRay. The information included is designed only for use
with Mars1417XF.
Trademarks
The iRay name and iRay logo are registered trademarks of iRay Technology Co.
Ltd.
Marking Description
Numerical Order
Year
Date
Month
Version
Product Code
Stand-by.
Non-ionizing radiation
Contents
1 Safety Information ...................................................................................................................... 11
2.1 Scope 17
2.2 Model 17
2.3 Characteristic 18
3 Installation ..................................................................................................................................28
4 Operation ....................................................................................................................................40
4.9 Software 67
4.10 IT-network 78
8.2 Repairment 93
1 Safety Information
WARNING
Installation and environment Do not use or store device near flammable chemical such as
alcohol, thinner, benzene, etc.
of use
If chemical is spilled or evaporated, it may result in fire or electric
shock through contact with electric parts inside device. Also,
some disinfectant is flammable. Be sure to take care when using
them.
Power supply Do not operate using any type of power supply other than
indicated on rated label.
Otherwise, it may result in fire or electric shock.
WARNING
Handling Never disassemble or modify device. No modification is allowed.
Doing so may result in fire or electric shock. Also, since device
contains components that may cause electric shock and other
hazardous parts, touching them may cause death or serious
injury.
Have the patient take a fixed posture and do not let patient touch
unnecessary parts.
If patients touch connectors or switches, it may result in electric
shock or malfunction.
When problem occurs Should any of the following occurs, immediately unplug the
power cord, and contact sales representative or local iRay
dealer:
When there is smoke, odd smell or abnormal sound.
When liquid has been spilled inside or metal object has entered
through an opening.
When the device has been dropped and damaged.
Maintenance and inspection Please turn off the power of device and unplug power cord of
adaptor before cleaning.
Never use alcohol, ether and other flammable cleaning agent for
safety. Never use methanol, benzene, acid and base because
they will erode device.
Don’t dip the device into liquid.
Please make sure that the device’s surface & plug is dry before
turning on.
Otherwise, it may result in fire or electric shock.
CAUTION
Installation and environment Do not install device in any of locations listed below. Doing so
may result in failure, malfunction, falling, fire or injury.
of use
Close to facilities where water is used
Where it will be exposed to sunlight directly
Close to air outlet of air-conditioner or ventilation equipment
Close to heat source such as heater
Where power supply is unstable
In dusty environment
In saline or sulfurous environment
Where temperature or humidity is high
Where there is freezing or condensation
In areas prone to vibration
On incline or unstable area
Take care that cables do not become tangled during use. Also,
be careful not to get your feet caught by cable.
Otherwise, it may cause malfunction of device or injury of user
due to tripping over cable.
Handling Do not spill liquid or chemicals onto device. In case the patient
is injured, iti is not allowed to contact with blood or other body
fluids.
Doing so may result in fire or electric shock.
In such situation, protect device with a disposable cover as
necessary.
Turn off power and pull out plug to each piece of for safety when
not used.
CAUTION
CAUTION
Do not close to fire, do not use in high temperature
Do not invert positive and negative pole
Do not contact with metal in case of short circuit
Do not insert sharp objects into battery
Do not beat battery
Do not stand on battery
Do not use battery out of rules
Do not dispose battery or change internal structure
Do not submerge battery in water, please keep dry in storage
and do not contact with water in use
Please charge battery with charger provided by iRay
Do not mix battery with ones not provided by iRay
Do not charge battery with broken charger .
Charge battery regularly to avoid over discharge failure.
Do not use battery when it is severe ballooning.
Before exposure
Be sure to check device daily and confirm that it works properly.
Be sure there is battery installed on Mars1417XF to avoid power off suddenly
Sudden heating of room in cold areas will cause condensation to form on device.
In this case, wait until condensation evaporates before performing an exposure.
If it is used while condensation is formed, problems may occur in the quality of
captured images. When air-conditioner is used, be sure to raise/lower the
temperature slowly so that difference of temperature in room and device does
not occur, to prevent condensation.
The detector should be warmed more than 20 minutes before exposure or
updating gain or defect template.
During exposure
Do not move power cable during exposure, or it may cause image noise or
artifacts, even incorrect images.
Do not use equipment near detector generating strong magnetic field. Otherwise,
it may cause image noise, artifacts or even incorrect images.
Do not exposure in 60s after 4 times of full range exposure. Otherwise, image
would be incorrect. Neither do exposure in 30s after a full range exposure. The
larger dose using, the longer time should be waited before next exposure.
During image acquisition, Mars1417XF should not be influenced by physical or
electrical way.
If detector receives dose which is close to full range, clipping data lines appear in
full range area.
After exposure
If detector would not be used in 5 days, it is required to take out battery. If battery
would not be used at long time, it must be charged to 30%~50% every 3 month
or 50%~70% every 6 month.
2 General Description
Mars1417XF is a cassette-size wireless X-ray flat panel detector based on amorphous silicon
thin-film transistor technologies. It is developed to provide the highest quality of radiographic
image, which contains an active matrix of 2336×2836 with 150um pixel pitch. Detectors’
scintillator has two options which are GOS (Gadolinium Sulfoxylate) and CsI (CaesiumIodide).
However, the greatest improvement is wireless communication between detectors and PC.
Meanwhile it can be powered with battery for portable panel use.
2.1 Scope
This manual contains information about Mars1417XF. All operators must read and understand
this manual before using. All information in this manual, including illustrations, is based on
equipment prototype. If your configuration does not have any of these items, information about
these items in the manual does not apply to your detector.
Mars1417XF
2.2 Model
Mars □ □- □
XF
Mars series
2.3 Characteristic
Wireless static Flat Panel Detector used for general radiography.
14 × 17 inch
AED trigger
Battery recycling
IPX4
The detector could be used for general X ray diagnosis of usual body part. It is not intended for
mammography, dental applications, neonatal and fluoroscopy. More care should be taken when
diagnosis with allergic constitution. Not only this but also prohibited for pregnant women and
children. Shield of none inspection body part is necessary during x ray exposure. There is no
contraindication.
According to Mars1417XF intended use and result of risk management, essential performance is
identified and described as following:
To maintain data transmission, Mars1417XF shall be not influenced to data and signal
transmission.
Item Description
Control Panel
B Reserved Reserved
D Reserved Reserved
2.6.2Battery Pack
D E
B
A Battery Label /
C Guide block /
2.6.3Battery Charger
Item Specification
Mars1417XF-CSI (CsI)
Greyscales 16 bit
14dBm(Typ.) @802.11b
13dBm(Typ.) @802.11g
11a/g/n: OFDM(BPSK,QPSK,16QAM,64QAM)
5.19GHz≤40MHz
5.8GHz≤40MHz
Note: Mars1417XF-GSI active area and defect calculation area 2336*2836, TFT module size is
Mars1417XF-CSI defect calculation area is 2320*2818, active area is 2336*2836, TFT module size
is 2342*2840
2.7.2Battery
Item Specifications
Model Battery-KX
2.7.3Battery Charger
Item Specifications
Model Charger-KX
Item Description
Memory 8G DDR3
2.7.5Environment
Storage
10%~90% 700~1060hPa <10kp/min
(without -20~50℃ <1k/min
RH (1kp=1.0197E-5Pa)
battery)
Mars1417XF detectors should operate at altitude specified not more than 3000m, the requirement is
only for detector.
If storage with battery, temperature should be in the range of -20℃~45℃ when expected time is less
than 3 months. Meanwhile, -20℃~25℃ for 12 months.
3 Installation
Mars1417XF could be powered by both battery pack and DC power. Once battery pack is
inserted or DC power is connected, Detectors would be turned on immediately. If neither battery
nor DC power is connected, Mars1417XF would power off. Please see below for battery
installation.
3.1.2Booting Up
Power indicator
Status
Power Indicator Lighting Status
Battery Capacity DC Input Description
>15% NO
Green ON Detector is turned on
NO YES
Link indicator:
Detector Initialization
Blue blinking
Wireless configuration reset
Mode indicator:
Blue ON Default
Status indicator:
3.1.3Adapter
In order to meet safety and function requirements of detector, the standard component is
recommended:
LXCP120-0240625(C005)
Wireless setup
Configure 2.4GHz wireless network SSID: XF-SN-2G (No conflict in the air)
Security: WPA2-PSK
Password: 12345678
Configure 5GHz wireless network SSID: XF-SN-5G (No conflict in the air)
Security: WPA2-PSK
Password: 12345678
LAN setup
Open iDetector
Click “Detector”
Click “Add”
Click “Apply”
Wireless setting would be activated immediately. Default SSID and password is configured,
detector would connect to wireless router after powered on next time.
If users don’t know wireless configuration of detector. There provides a way to reset configuration
Select with”Mars1417XF_AP”
Note: If users don’t want to use default IP address of workstation to build the connection, it is
necessary to close iDetector before changing IP address of workstation and iDetector configuration.
4 Operation
Mars1417XF provides users SDK for integration into DR system. Additionally, it also provides
application demonstration, i.e. iDetector.
Workstation means host PC device installed with iDetector or SDK. Chapter 3.3 has described
how to establish connection between detectors and workstation. In Software mode, workstation
does not control X-ray generator, users decide when to shoot X-ray.
Click “Acquire”
Detector
2 3 10 11
Workstation
1 4 9 12
Image received would be preview image, preview image is image without much correction which
causes them some stripes, it cannot be used for final diagnosis.
Detector would do another dark image acquisition for offset correction. If Hardware Post offset
and Hardware calibration is selected, detector uploads processed image to workstation finally
after offset, gain and defect calibration.
If Software Post offset and Software calibration is selected, corrected image is shown on screen
after workstation finishes offset, gain and defect calibration.
Note: If wireless circumstance is bad, detector couldn’t send even one package in 30s. It would
stop trying sending image package. Users have to retrieve images from detector when wireless
is good enough.
Workstation means host PC device installed with iDetector or SDK. Chapter 3.3 has described
how to establish connection between detectors and workstation. In AED mode, workstation does
not control X-ray generator, user decides when to shoot X-ray.
Click “Acquire”
Detector
2 3 9 10
Workstation
1 4 11 12
2. Detector receives “clear” from workstation and begin flushing panel. Meanwhile, replies to
workstation “Exposure Prohibited”.
5. User triggers X-ray generator to initialize and do anode rotation to prepare for X-ray.
Image received would be preview image, preview image is image without much correction which
causes them some stripes, it cannot be used for final diagnosis.
Detector would do another dark image acquisition for offset correction. If Hardware Post offset
and Hardware calibration is selected, detector uploads processed image to workstation finally
after offset, gain and defect calibration.
If Software Post offset and Software calibration is selected, corrected image is shown on screen
after workstation finishes offset, gain and defect calibration.
Note: If wireless circumstance is bad, detector couldn’t send even one package in 30s. It would
stop trying sending image package. Users have to retrieve images from detector when wireless
is good enough.
Note:
1. Users must re-connect detector with different IP address when changing connection
from different net card.
2. Switching between wired and wireless connection does not need any extra operation.
3. The rule of multi-share is based on IP address. The second terminal with different IP
address is not allowed to operate when the first is connected. If there is no command
transmission between detector and workstation over 5 minutes, detector releases access
authority.
“Acquire” module:
“SDK” module:
“Detector” module:
“Calibrate” module:
Click “Start”
Note:
Yellow means image does not meet requirements, but can generate template.
Red means image does not meet requirements, cannot generate template, must be
taken again.
. Note:
Yellow means image does not meet requirements, but template can be generated.
Red means image does not meet requirements, template cannot be generated, must
take another shot.
Click “Open” /
Click “OK”
number"
iDetector provides function checking defect template. If defect template has updates, user could
add and delete defect pixel or lines.
Detector supports correction templates storage which means templates could be transmitted not
only from detector to workstation, but also from workstation to detector.
Choose “Calibrate”,
Click “Download”.
Upload templates
Choose “Calibrate”,
Detector supports two ways to do correction. Software correction defines scenario that
workstation finishes correction. If detectors do itself, that’s hardware Correction and Calibration.
Choose “Acquire”
Offset Mode:“SWPostOffset”,
Go to page “Detector”
Click “Write”
Firmware updating
Open IE browser.
Input 192.168.100.8
Password: iray
Choose “upgrade”
Choose”
Mars1417XF_IMAGE_41_ALL_XXXX
_XX_XX.ifrm”
Click “upgrade”
Note:
1. please insert battery (more than 25%) in detector, in case power is down when
upgrading. On the other hand, detector should reboot after updating.
4.9 Software
4.9.1 Main GUI
Double-click iDetector, main interface is shown on screen. See table below for detailed function
description.
SN Detector SN number
This page works for image acquisition mainly. In “operation” box, user chooses image correction
method according to requirements. “image properties” shows simple information of image
acquired. “Image list” shows lasted 5 images, if user wants to check some image, double click.
User can rotate images and do other image processing with “ROI”.
Figure 4.13.2
SWPreOffset Reserved
HWPreOffset Reserved
Offset
SWPostOffset Workstation does post offset correction
Image Properties/
Functional description
Image Process
WW Window width
WL Window level
Image Panel
A A
B D
C C
D B
A B
D C
A B
D C
The page is used to configure config.ini and set log level in real time, as shown below
4.9.5.1 Parameters
Parameter tab is activated in default. Five boxes on page are defined as following:
Zone 1: parameters
Zone 2: parameters reading from detector
Zone 3: parameters which are written into detector
Zone 4: function button
Zone 5: simple message from detector and state
SN Serial number NO
4.9.5.2 Sensor
This page includes temperature and humidity information.
Parameters Description
Client
Up Move up
Others
Scan from FPD Scan SSID in air with FPD wifi module
4.9.5.4 Images
Parameters Description
Note:
4.10 IT-network
4.10.1 Purpose for IT-network
Wireless communication follows IEEE 802.11a/b/g/n protocol. It works on 2.4GHz and 5GHz.
Wireless card and detector must works on the same IP segment such as 192.168.100.XXX
15dBm(Max.) @802.11b
14dBm(Max.) @802.11g
11a/g/n: OFDM(BPSK,QPSK,16QAM,64QAM)
5.19GHz≤40MHz
5.8GHz≤40MHz
Detector sends image data acquired to workstation. Workstation sends users’ command to
detector.
Compatibility of interface;
4.10.7 Warning
connection of main unit to an IT-network that includes other equipment could result in
manufacturer of x-ray machine should identify, analyze, evaluate and control these risks.
Subsequent changes to IT-network could introduce new risks and require additional
analysis.
5 Charger Installation
Insert battery into battery charger
6 Regulatory Information
Mars1417XF safety regulatory includes safety of detector, charger and other accessory.
ADDRESS: Rm. 202, Building 7, No. 590, Ruiqing Rd., Zhangjiang East,
Pudong, Shanghai, China
TELEPHONE: +86-21-50720560
ADDRESS: Rm. 202, Building 7, No. 590, Ruiqing Rd., Zhangjiang East,
Pudong, Shanghai, China
ZIPCODE: 201201
TELEPHONE: +86-21-50720560
Flammable anesthetics Not suitable for use in situation with flammable anesthetic
mixture with air, oxygen or nitrous oxide
The detector is born with two power supply modes(power adaptor and battery pack) and a single way
of signal transmission(wireless)
EN ISO 13485:2012/EN ISO Medical devices --- Quality management systems --- Requirements
13485:2012/AC:2012 for regulatory purposes
IEC60601-1-2:2014
ANSI/AAMI ES60601-1:2005+
Medical Electrical Equipment – Part 1: General requirements for
Amendment 1:2012+
safety and essential performance
Amendment 2:2010
CAN/CSA C22.2 No. 60601- Medical Electrical Equipment – Part 1: General requirements for
Emissions
Phenomenon Compliance Electromagnetic environment
Class A
Enclosure Port
Phenomenon Basic EMC standard Immunity test levels
80MHz-2.7GHz
80% AM at 1kHz
Near fields from RF IEC 61000-4-3 Refer to table “Near fields from RF wireless
wireless communications equipment”
communications
equipment
745
780
870
930
1845
1970
5500
5785
Electrical fast ±2 kV
IEC 61000-4-4
transients/burst 100kHz repetition frequency
Surges
IEC 61000-4-5 ±0.5 kV, ±1 kV
Line-to-line
Surges
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Line-to-ground
3V, 0.15MHz-80MHz
Conducted
disturbances induced IEC 61000-4-6 6V in ISM bands between 0.15MHz and 80MHz
by RF fields
80%AM at 1kHz
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
Mars1417XF series wireless flat panel detector should not be stacked with or adjacent to other
device. If inevitable, verify detector.
Mars1417XF series wireless flat panel detector conforms to this EN60601-1-2:2007 standard on
both immunity and emission.
Accessories, transmitters and cables other than those specified by User Manual or sold by iRay
may result in increased emission or decreased immunity of detector.
SAR
IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-
acid electrolytes
Battery Label
7 Trouble shooting
Please refer service manual. If problem unsolved, turn off detector and contact iRay service
department (service@iraychina.com). We would provide the best service.
8 Product Maintenance
There is a lithium battery in detector whose lifetime is more than 5 years, battery needs to be
replaced when time is over. Contact Shanghai IRay after-sales service departments or
authorized product distributors.
8.2 Repairment
If problem cannot be solved, contact your sales representative or local iRay dealer. Please
provide following information:
Product Name:
Series Number: