You are on page 1of 143

BS-200/BS-220 Chemistry Analyzer

Service Manual
© 2011-2015 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights
Reserved.

For this Service Manual, the issued Date is 2015-09 (Version: 5.0).

Intellectual Property Statement


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this
manual. This manual may refer to information protected by copyrights or patents and
does not convey any license under the patent rights of Mindray, nor the rights of
others. Mindray does not assume any liability arising out of any infringements of
patents or other rights of third parties.

Mindray intends to maintain the contents of this manual as confidential information.


Disclosure of the information in this manual in any manner whatsoever without the
written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rent, adaption and translation of this


manual in any manner whatsoever without the written permission of Mindray is
strictly forbidden.

, , , , , are the
registered trademarks or trademarks owned by Mindray in China and other countries.
All other trademarks that appear in this manual are used only for editorial purposes
without the intention of improperly using them. They are the property of their
respective owners.

Responsibility on the Manufacturer Party


Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not
be liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.

Mindray is responsible for safety, reliability and performance of this product only in
the condition that:

 all installation operations, expansions, changes, modifications and repairs of


this product are conducted by Mindray authorized personnel;

 the electrical installation of the relevant room complies with the applicable
national and local requirements;

 the product is used in accordance with the instructions for use.

i
Upon request, Mindray may provide, with compensation, necessary circuit diagrams,
calibration illustration list and other information to help qualified technician to
maintain and repair some parts, which Mindray may define as user serviceable.

WARNING:
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or injury of human
health.

NOTE:
This equipment is to be operated only by medical professionals trained
and authorized by Mindray or Mindray-authorized distributors.

Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions
Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
use of parts or accessories not approved by Mindray or repairs by people other than
Mindray authorized personnel.

This warranty shall not extend to:

 any Mindray product which has been subjected to misuse, negligence or


accident;

 any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;

 any product of any other manufacturer.

ii
Return Policy

Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:

1. Obtain return authorization: Contact the Mindray Service Department and obtain
a Customer Service Authorization (Mindray) number. The Mindray number must
appear on the outside of the shipping container. Returned shipments will not be
accepted if the Mindray number is not clearly visible. Please provide the model
number, serial number, and a brief description of the reason for return.

2. Freight policy: The customer is responsible for freight charges when this
product is shipped to Mindray for service (this includes customs charges).

3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department

Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, ShenZhen 518057, P.R.China,
Tel: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500

iii
Preface

Who Should Read This Manual


This manual is written for service personnel authorized by Mindray.

Conventions Used in This Manual

Safety Symbols

In this manual, the signal words BIOHAZARD, WARNING, CAUTION


and NOTE are used regarding safety and other important instructions. The signal
words and their meanings are defined as follows. Please understand their meanings
clearly before reading this manual.

When you see … Then …


WARNING: Read the statement following the symbol. The
statement is alerting you to an operating hazard
that can cause personal injury.
Read the statement following the symbol. The
BIOHAZARD: statement is alerting you to a potentially
biohazardous condition.

CAUTION: Read the statement following the symbol. The


statement is alerting you to a possibility of
system damage or unreliable results.
Read the statement following the symbol. The
NOTE: statement is alerting you to information that
requires your attention.

Graphics
All graphics, including screens and printout, are for illustration purpose only and
must not be used for any other purposes.

EC Representative
Name: Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestrasse 80 D-20537 Hamburg Germany
Tel: +49 40 2513174
Fax: +49 40 255726

iv
Contents

Intellectual Property Statement................................................................................................... i

Responsibility on the Manufacturer Party ................................................................................. i

Warranty ....................................................................................................................................... ii
Exemptions.............................................................................................................................. ii

Return Policy .............................................................................................................................. iii


Return Procedure ................................................................................................................... iii
Company Contact................................................................................................................... iii

Preface ......................................................................................................................................... iv
Who Should Read This Manual ............................................................................................. iv
Conventions Used in This Manual ......................................................................................... iv

1 Specifications.................................................................................................................... 1-1
1.1 General Features ................................................................................................. 1-1
1.2 Features of Loading System................................................................................ 1-1
1.3 Features of Photometric System ......................................................................... 1-2
1.4 Operating Interface .............................................................................................. 1-2

2 System Installation ........................................................................................................... 2-1


2.1 Space and Accessibility Requirements ................................................................ 2-1
2.2 Power Requirements ........................................................................................... 2-1
2.3 Environmental Requirements .............................................................................. 2-2
2.3.1 Installation Environment Requirements................................................ 2-2
2.3.2 Temperature and Humidity Requirements ............................................ 2-2
2.4 Installation Procedure .......................................................................................... 2-3
2.4.1 Preparation for Installation ................................................................... 2-3
2.4.2 Beforehand Check ................................................................................ 2-3
2.4.3 Hardware Installation ............................................................................ 2-4
2.4.4 Software Installation ........................................................................... 2-12
2.4.5 Turning on the analyzer ...................................................................... 2-12
2.4.6 Adjusting ............................................................................................. 2-14
2.4.7 Setup and Test.................................................................................... 2-16
2.4.8 Training ............................................................................................... 2-16

3 System Description .......................................................................................................... 3-1


3.1 General ................................................................................................................ 3-1
3.1.1 System Structure .................................................................................. 3-1
3.1.2 Work Flow............................................................................................. 3-2
3.2 Major Components .............................................................................................. 3-3
3.2.1 Dispensing System ............................................................................... 3-3
3.2.2 Temperature Control System................................................................ 3-5
3.2.3 Photometric System ............................................................................. 3-8
3.2.4 Fluid System ......................................................................................... 3-9
3.2.5 Bar Code Module ............................................................................... 3-11
3.2.6 ISE Unit (optional) .............................................................................. 3-13

4 Functions of Boards ......................................................................................................... 4-1

1
Contents

4.1 Hardware System ................................................................................................ 4-1


4.1.1 Control Framework ............................................................................... 4-1
4.1.2 Main Board ........................................................................................... 4-2
4.1.3 Drive Board........................................................................................... 4-3
4.1.4 Pre-amp Board ..................................................................................... 4-4
4.1.5 AD Conversion Board ........................................................................... 4-4
4.1.6 Reagent Refrigeration Board................................................................ 4-5
4.1.7 Level Detection Board .......................................................................... 4-5
4.1.8 Reaction Disk Temperature Sampling Board ....................................... 4-6
4.1.9 Three Probes Connection Board .......................................................... 4-6
4.1.10 ISE Power Board .................................................................................. 4-6
4.1.11 Heater Voltage Selecting Board ........................................................... 4-6
4.2 On Board LED Indication ..................................................................................... 4-6
4.3 Power Supply Unit ............................................................................................... 4-8
4.3.1 Features of Power Supply Unit............................................................. 4-9
4.3.2 Protective Function of Power Supply Unit .......................................... 4-10
4.3.3 Block Diagram .................................................................................... 4-11

5 Software Introduction ....................................................................................................... 5-1


5.1 Software Structure and Response Mechanism ................................................... 5-1
5.1.1 Operating Software .............................................................................. 5-1
5.1.2 Control Software ................................................................................... 5-7
5.2 Software Installation ............................................................................................ 5-8

6 Maintenance and Service ................................................................................................. 6-1


6.1 Cleaning and Washing......................................................................................... 6-1
6.1.1 Cleaning Probe ..................................................................................... 6-1
6.1.2 Cleaning Mixing Bar ............................................................................. 6-2
6.1.3 Cleaning Sample/Reagent Compartment ............................................ 6-2
6.1.4 Cleaning Wash Well of Probe............................................................... 6-2
6.1.5 Cleaning Wash Well of Mixing Bar ....................................................... 6-3
6.1.6 Cleaning Panel of Analyzing Unit ......................................................... 6-3
6.1.7 Washing Deionized Water Tank ........................................................... 6-3
6.1.8 Washing Waste Tank ............................................................................ 6-4
6.1.9 Washing Dust Screens ......................................................................... 6-4
6.2 Replacing ............................................................................................................. 6-5
6.2.1 Unclogging Probe ................................................................................. 6-5
6.2.2 Replacing Probe ................................................................................... 6-9
6.2.3 Replacing Mixing Bar ........................................................................... 6-9
6.2.4 Replacing Plunger Assembly of Syringe ............................................ 6-11
6.2.5 Replacing Lamp ................................................................................. 6-13
6.2.6 Replacing the Optical Filter ................................................................ 6-15
6.2.7 Replacing the Optical Fiber ................................................................ 6-18
6.2.8 Replacing Filter Assembly .................................................................. 6-20
6.3 Repairing the Bar Code Module ........................................................................ 6-23
6.3.1 Replacing the Bar Code Reader ........................................................ 6-23
6.3.2 Installing the Anti-fogging Heater Assembly ....................................... 6-23
6.4 Bar Code Reader Adjustment ............................................................................ 6-24
6.4.1 Adjusting Sample Bar Code Reader .................................................. 6-24
6.4.2 Adjusting Reagent Bar Code Reader ................................................. 6-27
6.5 ISE Unit (optional) Maintenance ........................................................................ 6-28
6.5.1 Replacing tubings ............................................................................... 6-28
6.5.2 Replacing ISE module ........................................................................ 6-32
6.5.3 Replacing ISE Components (optional) ............................................... 6-33
6.6 Adjusting ............................................................................................................ 6-35
6.6.1 Adjusting Probe and Disks ................................................................. 6-35
6.6.2 Adjusting Signal Gain of Photoelectric Unit ........................................ 6-36
6.6.3 Adjusting Photoelectric Collection Position ........................................ 6-36
6.7 Checking ............................................................................................................ 6-38

2
Contents

6.7.1 Checking Reagent Refrigeration Board ............................................. 6-38


6.7.2 Checking Optical Couplers ................................................................. 6-39

7 Maintenance and Test Software ...................................................................................... 7-1


7.1 Basic Operations ................................................................................................. 7-1
7.1.1 Installation ............................................................................................ 7-1
7.1.2 Overview............................................................................................... 7-1
7.2 Operating Commands ......................................................................................... 7-3
7.2.1 Main Unit .............................................................................................. 7-3
7.2.2 Mixing Unit ............................................................................................ 7-3
7.2.3 Reagent Unit......................................................................................... 7-4
7.2.4 Sample Unit .......................................................................................... 7-5
7.2.5 Reaction Unit ........................................................................................ 7-6
7.2.6 ISE Module ........................................................................................... 7-7
7.2.7 Temperature Unit .................................................................................. 7-9
7.3 Parameter .......................................................................................................... 7-11
7.3.1 The Precondition for Parameter Configuration ................................... 7-12
7.3.2 Detailed Operations ............................................................................ 7-14
7.4 Temperature....................................................................................................... 7-16
7.4.1 Functions ............................................................................................ 7-16
7.4.2 Operation Details ................................................................................ 7-16
7.5 Photoelectric ...................................................................................................... 7-17
7.5.1 Filter Offset ......................................................................................... 7-17
7.5.2 Photoelectric Gain .............................................................................. 7-18
7.5.3 Light Source Background ................................................................... 7-19
7.5.4 Dark Current Test ............................................................................... 7-20
7.5.5 Other Functions of the Photoelectric Test .......................................... 7-21
7.6 Macro Instructions ............................................................................................. 7-21
7.6.1 Function .............................................................................................. 7-21
7.6.2 Detailed Operations ............................................................................ 7-21

Appendix A Diagrams .................................................................................................... A-1

Appendix B Spare Parts List ......................................................................................... B-1

3
Contents

4
1 Specifications

1.1 General Features


1 Dimension: 860mm×700mm×625mm (W×D×H, optional lower cabinet
excluded); 860mm×700mm×1160mm (W×D×H, optional lower cabinet
included)
2 Weight: 116kg (optional lower cabinet excluded; the optional lower cabinet
weighs 51Kg)
3 Power supply: AC 110-130V/200-240V, 50/60Hz
4 Input power: 1000VA
5 System: random selection of tests, multi-channel and multi-test.
6 Theory of analysis: Colorimetry, turbidity, latex agglutination, and
homogeneous immunity
7 Test method: Endpoint, Kinetic and Fixed-time. Supporting double-reagent
and double-wavelength, Ion Selective Electrode module(optional)
8 Test option: set up one by one on the operating software, profiles and
calculations provided.
9 Calibration rule: Linear (one-point, two-point and multi-point), Logit-Log 4P,
Logit-Log 5P, Exponential 5P, Polynomial 5P, Parabola and Spline
10 
QC rule: X -R, L-J, Westgard multi-rule, Cumulative sum check, Twin-plot
11 Tests analyzed simultaneously: 19/38 (double/single-reagent) or
18/36(double/single-reagent) with ISE Unit
12 Throughput: 200tests/h or 330tests/h with ISE unit

1.2 Features of Loading System


1 Sample/reagent disk: Round style with samples and reagents on the disk
2 Sample volume:
2-45μl with precision of 0.1μl
3 Sample position: 40
4 Reagent volume:
10-450μL with precision of 1μl
5 Reagent position: 40. Each position is available for the Mindray bottles only
6 Pre-dilution: Dilution ratio 4-150, taking the reaction cuvettes as containers
7 Probe: With a built-in level detector, equipped with auto safeguard, capable
of tracking level
8 Washing function: Automatically washing interior and exterior of the probe,
with carryover no more than 0.1%
9 Mixing bar: For single-reagent tests, it functions immediately (within the
same period) after sample dispensing; for double-reagent tests, it functions

1-1
Specifications

immediately (within the same period) after dispensing of R2

1.3 Features of Photometric System


1 Light source: 12V/20VA tungsten-halogen lamp
2 Light splitting mode: Splitting by optical fiber, filtering by an interference filter
wheel
4 Wavelength: 340, 405, 450, 510, 546, 578, 630 and 670; 700 and
800nm(optional)
5 Absorbance range: 0-4Abs, 10mm optical path conversion
6 Reaction cuvette: 5×6×25mm, optical path: 5mm, disposable, volume: 900μl
7 Volume of reaction mixture: 180-500l
8 Max. reaction time: 10 minutes
9 Reaction temperature: 37±0.3℃, with fluctuation no greater than ±0.1℃
10 ISE Unit(optional):
Sample type: serum, plasma and urine
Sample volume:
 Serum: 70uL,
 Diluted urine: 140uL(1 part of urine sample and 9 parts of urine diluent)
Test range(mmol/L):
 Serum: Na: 100.0-200.0, K:1.00-8.00, Cl:50.0-150.0
 Urine: Na: 10-500, K: 5-200 Cl: 15-400

1.4 Operating Interface


1 Operating system: Windows XP Home, Windows Vista Business
2 CPU: Above 1.8GHz
3 Memory: Above 1G
4 Hard disk: Above 40G
5 Display: Resolution 1024×768, color ≥ 16 bit
6 System interface: RS-232
7 Printer: any model

1-2
2 System Installation

2.1 Space and Accessibility Requirements

2.2 Power Requirements


 Power supply: AC 110-130V/200-240V, 50/60Hz, three-wire power cord and
properly grounded.
 The system should be connected to a properly grounded power socket. If
possible, the system should be connected to a power line specially designed for
medical devices.
 The distance between the power socket and the system should be less than 3
meters.

WARNING:
Make sure the power socket is grounded correctly. Improper
grounding may lead to electric shock and/or equipment damage.
Be sure to connect the system to a power socket that meets the
above-mentioned requirements and has a proper fuse installed.

NOTE:
It is recommended for hospitals to prepare a 2000VA online UPS.

2-1
System Installation

2.3 Environmental Requirements

2.3.1 Installation Environment Requirements


 This system is for indoor use only.
 The bearing platform (or ground) should be level (gradient less than 1/200).
 The bearing platform (or ground) should be able to bear 120Kg weight.
 The installation site should be well ventilated.

CAUTION:
The system radiates heat when operating. A well-ventilated
environment helps keep the room temperature stable. Use ventilation
equipment if necessary. But if so, be sure not to expose the system to
the direct draft that may lead to unreliable results.

 The site should be free of dust as much as possible.


 The site should not be in direct sun.
 The site should not be near a heat or draft source.
 The site should be free of corrosive gas and flammable gas.
 The bearing platform (or ground) should be free of vibration.
 The site should not be disturbed by large noise or power supply.
 The system should not be placed near brush-type motors and electrical contacts
that are frequently turned on and off.
 Do not use such devices as mobile phones or radio transmitters near the system.
Electromagnetic waves generated by those devices may interfere with operation
of the system.
 The altitude height of the site should be lower than 2000 meters.

WARNING:
When the forward or backward gradient is more than 8 degrees, the
system may tip over. Be sure to take proper measures when
transporting and storing it.

2.3.2 Temperature and Humidity Requirements

Storage temperature and humidity


 The system should be stored in a 0℃-40℃ environment with fluctuation less than
±2℃/H.
 The relative humidity should be between 30%RH-80%RH and with no
condensation.

2-2
System Installation

CAUTION:
Storing the system in an environment other than the specified may
damage the system.

Operating temperature and humidity


 When the system is running, be sure to fix the ambient temperature between
15℃ -30℃ with fluctuation less than ±2℃ /H.
 The ambient humidity should be between 35%RH-80%RH and with no
condensation.

CAUTION:
Operating the system in an environment other than the specified may
lead to unreliable test results.
If the temperature or relative humidity does not meet the
above-mentioned requirements, be sure to use air-conditioning
equipment.

2.4 Installation Procedure

2.4.1 Preparation for Installation


Ensure available installation fields in hospital: enough space, electrical environment,
room temperature and humidity and other conditions as specified.

PC, monitor and printer are prepared and can be installed with the operating
software and Maintenance and Test software at any moment.

NOTE:
Be sure the digital and date format of the Windows are same as them
of the English version Windows.

The reagent, calibrator, the tool kit, lubricating grease are prepared (engineer also
can take these parts by himself).

2.4.2 Beforehand Check


Go to the installation site and do the following beforehand check wok.

1 Check the delivery list for acceptance, and then unpack the wood package.
2 Make one copy of the original parameter configuration list that the engineer
will take back for archiving. The original copy is to be kept by the hospital
carefully.

2-3
System Installation

3 Hold the probe arm by hand, and move it up and down to its full travel for 8
times respectively at the washing position, reaction disk position, inner and
outer circles of the sample / reagent disk, so as to check whether the moving
resistances are uniform at all directions.
Hold the mixing bar arm by hand, and move it up and down to its full travel
for 8 times respectively at the washing position and reaction disk position, so
as to check whether the moving resistances are uniform at all directions.
Move the probes up and down to their full travel in turns to check whether
their moving resistances are uniform. Then record the results in the following
table.
Item Result Remark
Moving resistances of reagent □ Uniform □ Different
/sample probe driver assembly
at all directions
Moving resistances of mixing □ Uniform □ Different
bar driver assembly at all
directions
If the moving resistances are different, lubricate the driver assemblies, and
record the results after lubrication in the Remark column.
4 Use the syringe to inject water in the filter, and connect the filter to the cap
assembly of the deionized water tank. Avoid introducing foreign matters to
the filter or the tube when installing the filter.

2.4.3 Hardware Installation

Install the probe


1 Take out the probe from the box, and remember there’s a little tip also.
2 Take off the arm cover: Grab the lower parts of the arm cover with two hands
and pull them slightly outwards and remove the cover upward from the arm
base.

2-4
System Installation

3 Remove the retaining screw with a screwdriver and take away the spring.
Remember to move the arm to the highest position for convenient operation.

4 Insert the probe into the probe arm.

5 Sleeve the spring over the pole in the probe arm, and screw the retaining
screw back onto the pole.

6 Insert the connector of the probe lead into corresponding jack on the level
detection board. The probe arm is delicate. Exercise caution when inserting
the connector. Excessive force may damage the connector and/or the level
detection board.

2-5
System Installation

7 Screw the probe’s fluid connector back to the tubing connector. Note that
excessive force may bend the probe.

8 Then push the probe upward and then release the probe to check if it can
move freely. If not, find the reason and solve the problem.

9 Installing the arm cover.

2-6
System Installation

10 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if it can move freely. If not, remove
the arm cover and re-install it.

Install the mixing bar


1 Rotate the bar arm to the highest position for convenient operation.
2 Take out the mixing bar and retaining nut from the box. Note that pinch the
knurled part of the bar and take out it.

3 Pinch the new mixing bar by the knurled part and align the it to the bigger
hole end of the retaining nut and gently screw it into the nut until the end of
the bar is in line with the smaller hole end of the nut.

2-7
System Installation

4 Pinch the mixing bar by the knurled part 2 and align the hole of the nut 1 to
the axis on the bar arm, then push the bar upward in the direction of the axis
until it can’t proceed.

5 Tighten the nut by screwing clockwise with the other hand.


6 Check whether the bar is vertical to the bar arm.
If not, remove the bar and re-install it.
If so, proceed to the next step.
7 Pull the bar arm to its highest point and rotate it to move the bar to a position
above its wash well.

Install the sample/reagent disk


1 Carry the sample/reagent disk. Keep the handle at the vertical position and
put the disk in proper position.

2-8
System Installation

2 Then rotate the disk nightly, until the point can be seen from the hole.

3 Put the handle back to horizontal position.

Connect the deionized water tank

CAUTION:
When placing the deionized water tank, ensure the top of the tank is
lower than the bottom of the upper cabinet.
Ensure the deionized water pickup tube is not blocked, bent, or
twisted.

1 Place the Power to OFF.


2 Connect the filter with the pickup tubes.
3 Put the pickup tubes and the sensor into the deionized water tank, and then
turn the cap of the deionized water tank clockwise.

CAUTION:
Make sure that the filter sink smoothly to the tank bottom and does not
twist with the floater connecting rod.

4 Put the deionized water tank on an appropriate place. Do not tip it.
5 Plug the red and the green connectors to their counterparts marked
DEIONIZED WATER on the rear side of the analyzing unit and turn the
connectors clockwise until secure.

2-9
System Installation

6 Plug the sensor connector to its counterpart marked D-SENSOR on the rear
side of the analyzing unit and turn it clockwise until secure.

Connect the waste tank


1 Put the waste tube and the sensor into the waste tank, and then turn the cap
of the waste tank clockwise.
2 Put the waste tank on an appropriate place.
3 Keep pressing the pin on the waste connector marked WASTE on the rear
side of the analyzing unit and grab the waste tap and insert it to the
connector.
4 Plug the sensor connector to its counterpart marked W-SENSOR on the rear
side of the analyzing unit and turn it clockwise until secure.

5 Be careful about the position of the waste tube, because the waste drive by
its own gravity, so we have to make sure all of the waste tube is above the
waste tank and below the analyzer.

Waste Tank

2-10
System Installation

Waste Tank

Connect the PC to the Analyzing Unit


1 Connect the PC to the Analyzing Unit by serial communication cable.

Serial Port on the


Analyzing Unit

Serial Port on the PC

2-11
System Installation

Connect power cable and grounding wire of the analyzer

2.4.4 Software Installation

Install the operating software


Before installing the operating software, check the operating system and settings of
the PC and make sure:

 Operating system: Windows XP Home/Professional, Windows Vista Business or


above
 Memory: Above 1G
 Resolution: 1024×768

1 Power on the PC and start the Windows.

2 Double click the Setup.exe file at the directory of the installation CD.

3 Choose the language, click the Next button.

4 Click the Next button.

5 Enter the user name, company name and the serial number. The serial
number is “11111111”.
6 Click the Change button to change the installation directory of the software.

7 Click the Install button.

8 Click the Finish button when the installation is finished.

Install the driver of the printer

Shut down the PC

2.4.5 Turning on the analyzer


1 Fill the deionized water tank with deionized water.

2-12
System Installation

2 Load detergent to position 39 and distilled water to position 40 on the


reagent disk.

NOTE:
Mindray has specified the following detergents:
Acid: 0.1mol/l hydrochloric acid;
Alkaline: 0.5% (V/V) javel water.
Do not mix the acid detergent with the alkaline one.
Be sure to use the detergent recommended by Mindray.
Otherwise, proper result may not be obtained.
Mindray recommends the acid and alkaline detergents be used
alternately.

3 Switch on the analyzer as follows:


MAIN POWER → Power → monitor of PC → PC → printer
4 After logging on Windows, double-click the shortcut icon of the operating
software on the desktop.
5 Enter user name “Admin” and Password “Admin”, and then press “OK” to log
on the operating software.

6 The system automatically completes the self-check.


After that, the system asks you to unload the first segment of cuvettes to
measure background (as shown in the figure below). Because it is the first
time the analyzer runs, there is no cuvette in the reaction disk and you
press ”OK” to go to the next step directly.

2-13
System Installation

7 Add 8 new segments of cuvettes into the reaction disk to warm up. Check
whether the surfaces of the cuvettes are smooth before loading. Do not
touch the light transmission part of the cuvette in which the colorimetric
reading is taken.
8 Operate according to the screen prompt until the main screen of the
analyzer software is displayed.
9 If this is the first time the analyzer is installed, please follow the instructions
indicated in 5.5.2 Replacing Filter Assemby (step 4) to expel air ffrom the
filter assembly.

2.4.6 Adjusting
1 After starting the analyzer, adjust the resistance VR1 of the level detection
board (in the probe arm) until the output of TP2 is 4.5V (reference test point
is TP4).

Level Detection Board

2 Select MaintenanceDaily Maint, and then observe the system status and
record it in the table below:
Reaction
temperature
Preheat
temperature
Light Intensity
Waste tank Abnormal (full) Normal (not full)
Deionized water Abnormal (empty) Normal (available)
Printer No printer Normal
Main unit Unconnected Idle Running
Reaction unit Unconnected
Sample/reagent Unconnected
unit
Mixing unit Unconnected
Temperature unit Unconnected

Wavelength (nm) Dark Current Background


340
405
450

2-14
System Installation

510
546
578
630
670
Reference
3 On the Maintenance screen, select the Alignment tab, and implement all
sub-steps of each unit to see whether they are normal. In case of any
exception, adjust it.

4 Wash the interior and exterior of the probe and mixing bar for several times
to make all the tubing filled.

5 Check the top of the syringe for air bubles.

Check here for air


bubbles

6 If air bubbles exist, remove them according to the instructions in "5.6.5


Removing Air Bubbles" of the Operator's Manual. Exercise caution not to
damage the syringe.

2-15
System Installation

2.4.7 Setup and Test


1 Finish test, reagent and calibrator settings on the Parameters, Reagent and
Calibration screens, and then conduct a double-reagent blank.
2 Request for calibration and samples and then debug the results.
3 After debugging the results, fill them in the table below:
Test ALT CREA BUN
Target value
2sd range
Test value 1
Test value 2
Test value 3
Test value 4
Test value 5
Test value 6
Test value 7
Test value 8
Test value 9
Test value 10

2.4.8 Training
Can the customer complete daily tests? Yes □ No □
Is the customer familiar with the analytical methods such as Yes □ No □
kinetic, two-point, endpoint?
Is the customer familiar with the daily, weekly and monthly Yes □ No □
maintenance and relevant maintenance methods?
Is the customer familiar with replacing the filter assembly? Yes □ No □
Is the customer skilled in washing dust screens? Yes □ No □
Is the customer skilled in cleaning and replacing the probe and Yes □ No □
mixing bar?
Is the customer skilled in replacing the plunger assembly of Yes □ No □
syringe?
Is the customer skilled in replacing the lamp? Yes □ No □
Does the customer know the positions, roles and preparation Yes □ No □
methods of diluents and acid and alkaline detergents?

2-16
3 System Description

3.1 General

3.1.1 System Structure


The system consists of the analyzing unit, operation unit and output unit.

3-1
System Description

Figure 3-1 Front view

C
o
v
e
r

3.1.2 Work Flow


The measurement flow of the system is as follows:

Figure 3-2 Work flow of the system

Incubation
t test R2 Mix
Load ag e n
Startup le-re
new Cuvette Doub Output Replace
Start and R1 Sample Mix Absorbance
cuvette blank Single-reagent test results cuvettes
initialize
s

Test Procedures:

1 Power on the analyzer, then start up and initialize it. The initialization
includes resetting mechanical assemblies, switching on the lamp and
incubating by the temperature unit. You can’t start testing unless the lamp
becomes steady and temperature of the reaction disk reaches 37 ℃
constantly. This process may take about 15-20 minutes.
2 At the beginning of test, the probe and the mixing bar are washed.
3 The probe dispenses R1 to the cuvettes. It may skip to the next one when
encountering a dirty cuvette.

3-2
System Description

4 The probe dispenses samples to the cuvettes. When the incubation time
set for double-reagent tests is up, R2 is dispensed.
5 After reagents and samples are dispensed for a batch of tests, the probe
and the mixing bar will be washed.
6 When the test is finished, a prompt will appear to inform you to replace the
cuvettes manually.
7 When finishing all the tests, exit the operating software and the analyzer
will perform the shutdown procedure: washing the probe and bar with
detergents, resetting the mechanical assemblies.

3.2 Major Components


The analyzing unit of the system consists of the dispensing system, temperature
control system, photometric system, fluid system and etc.

3.2.1 Dispensing System


The dispensing system consists of the probe module, mixing bar module,
sample/reagent disk assembly and reaction disk assembly.

3.2.1.1 Probe Module and Bar Module

The mixing bar module is the same as the probe module, except that the φ12 shaft
is 30mm shorter.

3-3
System Description

Figure 3-3 Drive Assembly

The drive assemblies of the probe module and the bar module have a horizontal and
vertical photoelectric switch, which are used for defining horizontal and vertical initial
positions of the probe and bar. The horizontal and vertical step motors precisely
control the horizontal and vertical movements of the probe and bar. The
synchronous belts serve as the gearing.

3.2.1.2 Sample/reagent disk, reaction disk assemblies

The sample/reagent disk assembly consists of the sample/reagent disk and drive
support. The reaction disk assembly consists of the reaction disk and drive support.
The two drive support are different in their upper connection block and their coders.
The gaps of the coder define the positions where the disk should stop at. Each coder
has two transducers for inducing the rotation and initial position of the disk.

3-4
System Description

Figure 3-4 Side view of the sample/reagent disk assembly

Figure 3-5 Coder of reaction disk

The step motors control the disk assemblies, and the synchronous belts serve as the
gearing.

3.2.2 Temperature Control System

3.2.2.1 Temperature Control Assembly of Reaction Disk

The temperature control assembly of reaction disk consists of temperature-controlled


pot, bottom heater, heat insulator, bottom sponge, inside sponge, outside sponge,
cover board, top heater, press board, cover sponge,reaction disk/cuvettes,
photoelectric seat, temperature transducer, fans and control circuit.

3-5
System Description

Figure 3-6 Structure of the temperature control assembly

1 2 3 4 5 6 7 8 9 10 11 12 13

1-Temperature transducer and support 2-Fans

3- Outside sponge 4- Photoelectric seat 5- Cover board

6- Top heater 7- Press board 8- Cover sponge

9- Bottom sponge 10- Heat insulator 11-Temperature-controlled


chamber

12- Bottom heater 13- Inside sponge

 Temperature-controlled chamber
The temperature-controlled pot, photoelectric seat, outside sponge, inside sponge,
bottom sponge, cover board, and the cover sponge constitute an enclosed cavity, in
which the reaction disk rotates. The heat-insulator, with good heat insulation,
protects the temperature-controlled chamber from being interfered by external
environment. There is a gap on the cover, which is used for loading/unloading
reaction cuvettes manually.

 Heater
Top heater: square in shape, 110/220VAC, 125W

Bottom heater: ring in shape, 110/220VAC, 350W

Total power: 475W.

The heaters compensate the heat for incubating the reagents and maintaining the
temperature of the temperature-controlled chamber.

 Fans
Fans are used in series in the temperature-controlled chamber, so as to circulate the
air in the chamber and enhance the convective heat exchange. There are four fans
in the chamber. All of them have alarm function.

 Temperature transducer
It feeds back the air temperature at the position near the reaction cuvette. With an
overheat protection switch, the temperature transducer can disconnect the pot from
the power supply to avoid overheat and fire when the temperature inside the pot is
up to about 55°C. The switch can also be reset automatically when the temperature
inside the pot decreases to about 35°C.

3-6
System Description

 Control
PID control. PID parameters can be self-adjusted at any time.

 Safety precaution
The power of the two heaters is 110/220V, and the total power is nearly 0.5kW, so
the following precautions are considered: grounding protection of the
temperature-controlled pot, temperature switch protection against controller failure
and fan protection.

3.2.2.2 Reagent Preheating

The preheating assembly consists of two aluminum plates, a Teflon tube, heating
components, transducer, temperature switch, thermal conductive colloid, a section of
tubing and the probe. The transducer detects the temperature, and then transmits it
to the temperature controller. Thus the heating assembly is controlled, and the
temperature is controlled.

The temperature of the thermal source of the preheater is 45°C. The initial
temperature of the reagent is 2-8°C when it is taken out of the refrigerator. When the
reagent passes the heater, its temperature increases quickly. Then the reagent is
added into the reaction cuvette and the preheating process is finished.

Figure 3-7 Reagent preheating assembly

Preheating
Module

3.2.2.3 Reagent Refrigeration

The refrigeration module consists of refrigeration chamber 8, heat-insulator 9, cover,


reagent disk, rotor sleeve 11, temperature transducer, circuit and two refrigeration
assemblies 2. The refrigeration module is as shown in the figure below.

3-7
System Description

Figure 3-8 Reagent refrigeration module

The key components of refrigeration assembly include refrigeration accumulation


block 5, connector 4, Peltier 3 and radiator 6, as shown in the figure below. Each
analyzer has two such refrigeration assemblies.

Figure 3-9 Refrigeration assembly

If you want to replace any part of the refrigeration assembly, screw the eight screws
12 (as shown in the figure below) of the reagent chamber and replace it.

Figure 3-10 Reagent chamber (bottom view)

3.2.3 Photometric System


The lamp gives out the lights,and the light is transmitted to optical measurement
assembly by fiber. lights go into the cuvettes of the reaction disk, transmit through
the cuvettes and optical filters of specific wavelengths, and then go to the

3-8
System Description

photoelectric conversion board. The electric signal, converted from the optical signal,
is amplified and converted on the A/D conversion board, and then sent to the main
control board, which will submit the absorbance to the PC for calculation. The
multiple monochromatic wavelengths in the pototometer system are obtained by
utilizing the filter wheel. Rotate the filter of a specific wavelength to the light path
while performing the colormetric measurement.

Figure 3-11 Photometric principles

3.2.4 Fluid System


The fluid system washes the probe/bar and dispenses reagents/samples.

3.2.4.1 Washing
The fluid system of the system consists of deionized water tank, filter, wash pump,
solenoid valve, syringe, probe, wash well and waste tank.

3-9
System Description

Figure 3-12 Fluid system

As shown in the figure above, the fluid system consists of interior washing and
exterior washing parts.

The interior washing circuit begins from the deionized water tank, passes filter C12
and CV1 (Check Valve), and enters P1 (Interior Pump). Then the circuit goes
through SV (solenoid valve), syringe and the probe. Finally, the waste water goes
into the probe wash well after washing the probe.

The exterior washing circuit begins from the deionized water tank, goes through filter
C12 and CV2 (Check Valve), and then enters P2 (Exterior Pump). When the pump is
running, the detergent goes through J16 (three-way) and finally enters the probe
wash well and mixing bar wash well.

The waste water from probe wash well and mixing bar wash well converges with the
fluid of the reagent refrigeration module, and finally enters the waste tank.

3-10
System Description

3.2.4.2 Dispensing

The syringe assembly controls the aspiration/dispensing volume by controlling the


travel of the syringe. It is the key part of the fluid system.

Figure 3-13 Syringe assembly

The step motor, through the synchronous belt, controls the ball screw. The
photoelectric switch on the side controls the travel of the syringe.

3.2.5 Bar Code Module


The bar code module is composed of the bracket assembly, anti-fogging heater
assembly, bar code reader, and small bracket.

3-11
System Description

Figure 3-14 Syringe assembly


Anti-fogging heater assembly

Small bracket

Bar code reader

Bracket assembly

Figure 3-15 Syringe assembly

Mounting plate of
the anti-fogging
device Anti-fogging glass Heater Switch plate Dust shield Sponge window

Silicone pad
Heat-conducting Overheat
aluminum plate protection switch

3-12
System Description

3.2.6 ISE Unit (optional)

3.2.6.1 Components

Figure 3-16 Components of ISE unit

The ISE unit consists of the ISE module, peristaltic pump assembly and reagent
pack.

 ISE module: Composed of the sample entry port, ion selective electrodes and
circuit boards. It is located in a metallic shielding box.

Figure 3-17 ISE module

 Peristaltic pump assembly: Composed of three peristaltic pumps (A, B and W). It
is designed to control dispensing and discharging of calibrators and samples.

3-13
System Description

Figure 3-18 Peristaltic pump assembly

 Reagent pack: Composed of a reagent package.

Figure 3-19 Reagent pack

3.2.6.2 Fluidic System of ISE Unit

Figure 3-20 Fluidic system of ISE unit

Calibrator Calibrator
B A
Pump Pump Pump ISE
Waste container B A W

The reagent pack consists of calibrator A, calibrator B and a waste container. During
measurement, calibrator A and B are dispensed into the sample entry port while
being driven by pump A and B respectively, and the waste is discharged into the
waste container by pump W.

3-14
System Description

3.2.6.3 Notes for Calculation of Reagent Pack Inventory

Reagent pack as one part of the ISE module should be installed as instructed by this
manual. You should install reagent pack while the analyzer is powered off, and then
check the reagent inventory and prime the ISE module with Calibrator A and B. If the
reagent pack is not properly installed, error may be produced in inventory calculation,
and the actual volume may differ from that displayed on the screen.

The following operations may result in the actual inventory of the reagent pack
less than the displayed one.

 When the analyzer is turned on and the ISE module is in failure status, the
reagent pack is changed, without recovering the module failure or checking the
inventory immediately.
Operation explanation: When the ISE module is in Stopped status, the software will
stop writing volume information into the reagent pack chip. However, the analyzer
will automatically execute the sip action of the ISE module during startup to consume
certain amount of reagent. Therefore, when the reagent pack is replaced, but the
ISE module is not restored or the screen display refreshed, the software will not
calculate the reagent consumption volume until reading the new reagent pack. At
this moment, the actual reagent volume = displayed volume - consumption of Sip
after replacing the reagent pack, which means that the actual reagent volume is less
than the displayed volume on the screen.

 When the analyzing unit is powered on but the software is exited or the PC is shut
down, the ISE reagent pack is changed.
Operation explanation: During shutdown, the analyzer will automatically execute the
sip action of the ISE module to consume certain amount of reagent. When the
analyzer is started up, the software will detect a new reagent pack and zero the
consumption that is not written into the reagent pack chip, and then display the
volume read from the reagent pack chip. At this moment, the actual reagent volume
= displayed volume - consumption of Sip after replacing the reagent pack, which
means that the actual reagent volume is less than the displayed volume on the
screen.

 When the analyzing unit is powered on but the software is exited or the PC is shut
down, the ISE reagent pack is removed to be used on another analyzer and
install back on the initial analyzer after its usage.
Operation explanation: During shutdown process, the analyzer cannot detect if the
reagent pack is normal. If the reagent pack is installed on the current instrument after
being used on another one, the analyzer, when started up, will detect that the
recorded reagent volume on the chip is less than the original one, and the software
will zero the reagent consumption of Sip during shutdown and display the read
volume on the screen. At this moment, the actual reagent volume = displayed
volume - consumption of Sip during shutdown (with reagent pack installed), which
means that the actual reagent volume is less than the displayed volume on the
screen.

 The ISE module is de-configured, but the reagent pack is still on the instrument.
Operation explanation: Though the ISE module is de-configured, the original reagent
pack information is not cleared from the database. When the ISE module is
configured again with the same reagent pack, the Sip period without ISE module will
be included in calculation of the reagent consumption.

Note

 In condition 1, restore the ISE module and check the reagent inventory
immediately to minimize or eliminate the calculation error.

3-15
System Description

 In condition 2 and 3, before running the operating software to start up the


analyzer, remove the reagent pack and then run the operating software. After the
startup procedure is complete, re-install the reagent pack and inquire the reagent
volume immediately to minimize or eliminate the calculation error.
 In condition 4, remove the reagent pack and then de-configure the ISE module.
The following operations may result in the actual inventory of the reagent pack
more than the displayed one.

 When the analyzer is turned on and the ISE reagent pack is changed, ISE
inventory is not inquired immediately.
Operation explanation: The analyzer will automatically refresh the ISE reagent
consumption, by subtracting the consumption of calibrator A or B from the known
volume (of the original reagent pack) and writing it into the reagent pack chip, when
calibrator A or B is consumed for 1%. The written volume is based on the original
reagent pack before replacement, and the volume of a used reagent pack is often
less than that of a new one. So the volume recorded on the chip of the new reagent
pack is less than the actual volume, and the actual volume is greater than the
displayed volume on the screen.

 When the analyzing unit is powered on but the software is exited or the PC is shut
down, the ISE reagent pack is removed but not used on another analyzer and
then it is installed back before the analyzer is turned on.
Operation explanation: During shutdown process, the analyzer cannot detect if the
reagent pack is normal. If the reagent pack is loaded before the analyzer is started
up, the analyzer, when started up, will detect that the reagent volume recorded on
the chip is same as that recorded by the software, but the software will confirm by
mistake the reagent consumption of sip during shutdown. (Actually, no reagent is
consumed.) Therefore, after calculating the reagent consumption, the software will
write the new volume information onto the chip and display it on the screen. At this
moment, the actual reagent volume = displayed volume + consumption of Sip during
shutdown (with reagent pack installed), which means that the actual reagent volume
is greater than the displayed volume on the screen.

 After the operating software is exited, the analyzer is powered off and then it is
powered on before the operating software is run.
Operation explanation: When closed normally, the operating software cannot detect
if the ISE module is working. When powered on before the software is run, the
analyzer, when started up, will detect that the reagent volume recorded on the chip is
same as that recorded by the software, but the software will confirm by mistake the
reagent consumption of sip during shutdown. (Actually, no reagent is consumed
because the ISE module is not working while the analyzer is powered off.) Therefore,
after calculating the reagent consumption, the software will write the new volume
information onto the chip and display it on the screen. At this moment, the actual
reagent volume = displayed volume + consumption of Sip during shutdown (with
reagent pack installed), which means that the actual reagent volume is greater than
the displayed volume on the screen.

Note

 In condition 1, inquire the ISE reagent volume immediately to minimize or


eliminate the calculation error.

 In condition 2 and 3, before running the operating software to start up the


analyzer, remove the reagent pack and then run the operating software.
After the startup procedure is complete, re-install the reagent pack and
inquire the reagent volume immediately to minimize or eliminate the
calculation error.

3-16
System Description

3.2.6.4 Calculation of Reagent Pack Inventory

The reagent volume decreases as ISE instructions are executed, and it is updated
timely on the software. The table below shows reagent volumes consumed for
different instructions:

Module instructions Cal A volume (μl) Cal B volume (μl)


Calibration 360 360
Serum test 180 /
Urine test 100 230
Wash 180 /
Purge A 100 /
Purge B / 100
Pump check 475 175
Air bubble detector calibration 75 /
Sip period 130 130

The software calculates the consumption volume (μl) of calibrator A and B by


multiplying the total execution times of each ISE instruction and the reagent volume
per time.

When started up, the software writes the volume information into the reagent pack
as the calibrator A and B are consumed for 1%. When they are consumed for 100%,
the software marks reagent exhausted on the reagent pack chip.

The Sip period is special. The instrument equipped with the ISE module should not
be powered off, because the Sip period instruction is executed automatically for
every 30 minutes even the ISE module is in standby status (without running any
tests). During biochemistry tests, the Sip period instruction will be also executed as
long as the ISE module is continuously standby for over 30 minutes. Thus the
software will record the standby time of the instrument and the shutdown/startup time
of itself. For the shutdown period, the times of Sip period are calculated by dividing
the total time with 30 minutes. For the running period, the times of Sip period are
recorded based on the ISE module's continuous standby periods that are over 30
minutes. Sum of the two numbers is the final Sip period number, and it can be used
to determine the reagent consumption of the Sip period instruction.

3-17
System Description

3-18
4 Functions of Boards

4.1 Hardware System


The BS-200/220 consists of the following three units: the analyzing unit (main unit),
the operation unit (PC) and the output unit (printer).

The analyzing unit (main unit) consists of the following units: the temperature control
system, the reaction system (include ISE), the photometric system, the sample and
reagent delivery system, the mixing system and etc.

Figure 4-1 shows the control framework of the BS-200/220. The function of the
hardware system is as follows.

 Communicating with the PC through the RS232 to send commands, reply data
and test results.
 Controlling the data collecting process of optical system.
 Controlling the movement of the moving units and collecting the status signal.
 Controlling the temperature adjustment system and collecting the temperature
status signal.

Figure 4-1 Control Framework

4.1.1 Control Framework


The BS-200/220 consists of the following three units: the analyzing unit (main unit),
the operation unit (PC) and the output unit (printer).

The analyzing unit (main unit) consists of the following units: the temperature control
system, the reaction system (include ISE), the photometric system, the sample and
reagent delivery system, the mixing system and etc.

4-1
Functions of Boards

Figure 4-2 shows the control framework of the BS-200/220. The function of the
hardware system is as follows.

 Communicating with the PC through the RS232 to send commands, reply data
and test results.
 Controlling the data collecting process of optical system.
 Controlling the movement of the moving units and collecting the status signal.
 Controlling the temperature adjustment system and collecting the temperature
status signal.

Figure 4-2 Control Framework

4.1.2 Main Board


The function of the main board:

 Transmitting the data and commands with PC through RS232.


 Communicating with the sub-units(including ISE), transmitting the data and
commands through the extended RS232.
 Adjusting the numeric adjustable resistor on the AD conversion board and
collecting the photoelectric data.
 Controlling the rotation of step motors and driving moving units to accurate
position.
 Switching pumps and valves to control the flow of liquid.
 Detecting the level of liquid and obstruction signal.
 Detecting the signal of position sensors.
 Controlling the heater for heating.
 Detecting the signal of the water surface.
 Controlling the ISE module.
Figure 4-3 shows the function framework of the main board.

4-2
Functions of Boards

Figure 4-3 Function of the Main Board


PC
Control Moule 串口

Main Unit Sensors

Bus
Sensors

Sample probe BUS


Unit Sensors

Temperature BUS
Control Unit
Module of
Reaction Disk BUS ISE
Unit FPGA

Auto Washing BUS


Unit

Mixing Unit

Control Control
Signal Signal

Driving Module

Step Motor Heator Relay Pump Valve DC motor

4.1.3 Drive Board


The drive board is to receive the control signals from the main control board and
control the drive components, such as the reaction disk, the reagent\sample disk, the
sampling probe, the mixer and the filter wheel. It also controls the switches of two
pumps, the valves, the lamp and the temperature controlling system. The detaied
functions are:

 Controlling the movement components


 Controlling the pumps and the valves
 Controlling lamp
 Controlling heater
Figure 4-4 shows the function framework of the drive board.

4-3
Functions of Boards

Figure 4-4 Function of the Drive Board

Drive Board

Main Board Signal input Power input Power supply

Motor of reaction disk Switch of light

Motor of sample/reagent Others drive reaction disk heating

6 steper motor
Motor of mixing bar reagent preheating
name as
59D1300
Motor of sample probe
vertical movement

2 Motor of auto washing


Outside washing pump for
syring
Logic sample probe

Motor of sample probe Inside washing pump for


horizontal movement sample probe
Drive of pump
Inside washing vlave for
Motor of sample syring
sample probe
5 steper motor
Motor of filter wheel name as 4 valve for reserved
42D2120
Motor of auto washing
header vertical movement Drive of DC
Mixing DC motor
motor
Motor of mixing bar
horizontal movement

4.1.4 Pre-amp Board


The Pre-amp circuit board can converse the light signals into electrical analog
signals by the photoelectric diode. The photoelectric diode converses the light signal
into the current signal in different wavelengths and the Pre-amplifier turns the current
signal to the voltage signal and transmit this signal to the AD conversion board.

4.1.5 AD Conversion Board


This unit can process the analog signals from the pre-amp board and convert the
analog signals into the digital signals. The photoelectric diode converts the
monochromatic light signal into the voltage signal and the signal arrives the AD input.
The AD conversion board filters and amplifies it; at the same time, the AD collects
the signals indicated by the signal from the main board and then sends the AD value
to the main board for further process. The AD conversion board also provides power
to the Pre-amp board.

Figure 4-5 shows the function framework of the AD conversion board.

Figure 4-5 AD Conversion Board

AD conversion board
Photoelectric signal Photoelectric
Gain data
Pre-amp AD Main
Power regulating Power
board collection board
circuit Control signal
Controlling signal of gain
regulation

4-4
Functions of Boards

4.1.6 Reagent Refrigeration Board


Compared with other circuit boards, the reagent refrigeration board is an
independent unit; it can control the cooler chip on or off and then make the reagents
cool; it can adjust the temperature in the reagent carousel; also, it can drive the fans
of the whole system and feedback the signal of the working status of the funs to the
main board; the detailed functions are:

 Controlling the refrigeration


 Controlling the indicating LED of the refrigeration
 Controlling the fans and defogging the code scan’s windows (reversed)
 Feedbacking the sigual of the working status of the fans for lamp
Figure 4-6 shows the function framework of the reagent refrigeration board.

Figure 4-6 Reagent Refrigeration Board


Reagent Refrigeration Board Temperature
transducer

Temperature Refrigeration 2 pieces of


indication Control peltiers

2 fans for the


peltiers

12V Voltage 5 reserved fans


conversion
12VFA 1 fan for out
N
4 fans for reaction
Fog proof temperature control

(reserved) 1 fan for drive board

Main Photo Fan feedback


board 2 fans for light
coupler circuit

4.1.7 Level Detection Board


The board detects the liquid level, the detailed functions are:

 Detecting the reagent level and sample level with high reliability
 Transferring the level detection signal to the main board when the probe touches
the liquid level
 Protecting the probe from collision vertically and generating the detection signal
which is sent to the main board.
Figure 4-7 shows the function framework of the level detection board.

4-5
Functions of Boards

Figure 4-7 Level Detection Board Function

Oscillator, freq.- Obstruction


PLL Filter/amp. circuit
division circuit detection

probe Comparator Main Board

4.1.8 Reaction Disk Temperature Sampling Board


This board collects signals from the reaction disk temperature sensors and then
converts them into digital signals. This board provides an SPI interface, and is
connected to the main board.

4.1.9 Three Probes Connection Board


This board transfers the control signal of the reagent preheating, mixer motor pulse
and level detection board.

4.1.10 ISE Power Board


The board supplies the transformation of the voltage from 12V to 24V which provides
power to ISE module.

ISE module should be on power all the time, and the input 12V is provided by
5V/12V Board.

4.1.11 Heater Voltage Selecting Board


The board supplies two connection type for the 220V AC power and the 110V AC
power separately. If the AC power is 220V, please use the socket for 220V; if the
power is 110V AC power, please use the socket for 110V. The heater will provide the
same power consumption so as to ensure the effect of the heating in the different AC
power environment.

4.2 On Board LED Indication

Table 4-8 On-board LED Indication List


LED Mark LED On LED Off Remark
Main board(051-000485-00)
The pre-heater is The pre-heater is not
REAG
heating heating
The heater of
The heater of reaction
TRAY reaction disk is
disk is not heating
heating
ACID reserved

4-6
Functions of Boards

LED Mark LED On LED Off Remark


ALK reserved
For debug use only,
TEMP / /
can be ignored
WATER reserved
Indication of 5V
5V Exception of 5V power
power
Indication of 12V
12V Exception of 12V power
power
Indication of 3.3V
3.3V Exception of 3.3V power
power
For debug use only,
MAIN / /
can be ignored
For debug use only,
REAC / /
can be ignored
FPGA is being Configuration has been
CFG Off after power is on
configured finished
For debug use only,
FILL / /
can be ignored
Flickering indicates
FPGA
FPGA is running
For debug use only,
WASH / /
can be ignored
For debug use only,
STIR / /
can be ignored
Drive board(051-000486-00)
24V works
D1-24V 24V off or erroneously
normally
D2-5V 5V works normally 5V off or erroneously
12V works
D3-12V 12V off or erroneously
normally
D16 Inside pump on Inside pump off
D17 outside valve on outside valve off
D18 outside pump on outside pump off
AD conversion board(BA10-30-77759)
+15V works
+15V +15V off or erroneously
normally
-15V works
-15V -15V off or erroneously
normally
+5V works
+5V +5V off or erroneously
normally
Level detection board(BA20-30-75263)
The liquid surface
signal also occurs
Detecting the when the electric
LED1 No liquid surface signal
liquid surface things, such as hands
or metal, contact the
probe,
Reagent refrigeration board(BA20-30-75227)
12V works
D3-12VLED 12V off or erroneously
normally
Reagent
refrigeration Reagent refrigeration
D4-RED_LED temperature temperature T in other
status
T>10℃

4-7
Functions of Boards

LED Mark LED On LED Off Remark


Reagent
refrigeration Reagent refrigeration
D5-YELLOW_LED temperature T in other
temperature
status
T<0℃
Reagent
refrigeration Reagent refrigeration
D6-GREEN_LED temperature T in other
temperature 0℃
status
<T<10℃
Reagent Reagent refrigeration
D7-PELTIER_LED refrigeration power is off or
power is on. erroneously.
The J73 fan block
The J73 fan works
D8-FAN3_LED up, please check
normally.
the fan.
The J71 plug
D9-FAN1_LED doesn’t connect or normal
is broken.
The J74 fan block
The J74 fan works
D14-FAN4_LED up, please check
normally.
the fan.
12VFAN works 12VFAN off or
D15-12VFAN_LED
normally erroneously
The J72 plug
D16-FAN2_LED doesn’t connect or normal
is broken.
Reaction disk temperature sampling board(BA10-30-78268)
12V works
+12V 12V off or erroneously
normally
PFC board(BA33-30-35078)
Controlled by
LED1 A12V on A12V off or erroneously
software
24V board(BA33-30-35082)
24V on 24V off or erroneously
12V&5V board(BA33-30-35080)
LED350 B12V on B12V off or erroneously
LED350 C12V on C12V off or erroneously
LED350 5V on 5V off or erroneously

4.3 Power Supply Unit


 The power supply unit consists of three circuit boards: PFC board, 24V board and
12V&5V board.
 PFC Board:

①. EMI filter network;

②.Supplying the PFC circuit which conforms to the IEC61000-3-2 standard.


Supplying the +390V to drive the 24V board and 12V&5V Board;

③.Controlling the 24V, 12V (4.7A) and 5V outputs through the switch of the
analyzing unit;

4-8
Functions of Boards

④.Supplying stable 12V (4.5A) for the lamp, controlled by the system signal;

⑤.Supplying 14 V VDD voltage to drive the 24V board and 12V&5V board;

 24V Board:

①.Converting the 390V into 24V.The 24V of this board is output under the
control of the Power of the analyzing unit;

②. Transferring AC to the heater and controlling the heater;

 12V Board:
①. B12V/10A out;

②. C12V/4.7A out;

③. The 5V/12V board converts the 390VDC from the PFC board to 12V
(10A), 12V (4.7A) and 5V. The 12V (4.7A) and 5V are controlled by the switch
of the analyzing unit;

 The structure of the power supply unit make it manufactured and maintained
easily;
 The power supply unit is cooled by fans;

4.3.1 Features of Power Supply Unit

4.3.1.1 Input

 Input AC voltage : 100~240V(±10%)


 Frequency: 50/60±3HZ
 Input Power: 800VA

4.3.1.2 AC Output

 AC output: 90~264V
 Frequency: 50/60±3HZ
 Output power:470VA

4.3.1.3 DC Output: 5V-for Circuit Boards

 Output current:: 4.5A


 Output voltage: 4.85~5.25V
 Output ripple noise: 100mVp-p

4.3.1.4 DC Output: A12V-for Lamp

 Output voltage: 11.75~12.25V

4-9
Functions of Boards

 Output ripple noise: 100mVp-p


 Control request: A12V is controlled by an analog signal from the system
 Output current: 2A
 Insulation: A12V have not electrical connection with other power

4.3.1.5 DC Output: B12V-for Refrigeration and Fans

 Output current:: 10A, Inrush current 14A, continuing 300ms


 Output voltage : 11.8~12.8V
 Output ripple noise: 140mVp-p

4.3.1.6 DC Output: C12V-for Pumps, Valves and Boards

 Output current:: 4.7A


 Output voltage: 11.8~12.8V
 Output ripple noise: 140mVp-p

4.3.1.7 DC Output: 24V-for Motors and Boards

 Output current:: 5 A
 Output voltage: 23.5~25V
 Output ripple noise: 140mVp-p

4.3.2 Protective Function of Power Supply Unit

4.3.2.1 Over-current Protection


 All the output current have over-current protection
 The over-current protection point is in the range of 105~140% of the normal
output current except the A12V&B12V. Make sure the A12V&B12V are safe and
reliable before the current is over the the over-current protection point
 The status after protection: The statuses after protection are board-locked except
A12V which is the system-locked protection.

4.3.2.2 Over-volt Protection

 All the output volts have the locked over-volt protection.


 The protection point of the over-volt is 115-140% of normal volt.
 The status after over-volt protection is just the same as the status after
over-current protection.

4-10
Functions of Boards

4.3.3 Block Diagram


shows the framework of the power unit.

Figure 4-8 Framework of Power Supply Unit

4-11
Functions of Boards

4-12
5 Software Introduction

5.1 Software Structure and Response Mechanism


The system software is composed of the operating software and the control
software.

5.1.1 Operating Software

Figure 5-1 Operating software interface

The operating software manages the test and sample information entered by you,
generates a test procedure based on the requested tests, and sends test instructions
to the analyzing unit. These instructions inform the analyzing unit to start testing and
then output and save the results.

The operating software controls the analyzing unit through the control software. They
communicate with each other through the serial port upon the specific protocols.

The operating software is composed of 14 modules: calculation, startup, shutdown,


database, interface, serial port receiving/sending, processing received data, error
processing, testing, commissioning, failure restore, replacing cuvettes, printing and
protocol encapsulation.

5-1
Software Introduction

5.1.1.1 Calculation

This module calculates the calibration parameters and the concentration of the
reaction mixture. The calculation procedure includes:

 AD → absorbance
 Absorbance → response
 Response → calibration parameters
 Calibration parameters, response → concentration of the mixture

5.1.1.2 Startup

This module initializes the operation unit and analyzing unit to prepare for testing.

The startup process includes: checking and initialization of the operation unit,
communication with and initialization of analyzing unit.

The operation unit is checked and initialized before the main screen is displayed.

1 To check the operating system of the PC


The operating software must run under Windows XP or Windows Vista.
Otherwise, a prompt will appear indicating that the operating software
cannot run under the operating system other than the specified, and then
the operating software will be closed automatically.
2 To check the screen resolution of the current operating system
The operating software must run under the resolution of 1024×768.
Otherwise, the software can’t be displayed in full screen and a warning will
be given. You can set the resolution immediately and check again, or exit
the software and start it again after setting the resolution.
3 To check the color of the current operating system
If the color is less than 8 bits, the software can’t be displayed normally and a
prompt will be given to inform you of wrong color. You can set the color
immediately and check again, or exit the software and start it again after
setting the color.
4 To disable the screen saver
The operating software must keep displayed while running. To prevent the
screen saver from disturbing you in operating and observing, the screen
saver should be disabled. A warning will be given for disabling failure, you
can try it again or neglect or exit the software.
5 To initialize the database
The current database is compressed and restored and will backed up to the
historical database if the year is changed. When operation is failed, a
warning will be given to ask you to restore the database. You can exit the
software after confirming.
6 To check the printer
Check whether a printer has been set and connected to the PC.
Startup will not be affected when there is no printer. When you need to print,
the system will check the printer again.

5-2
Software Introduction

7 Log on
If the username and password are not correct, the user must not proceed to
the next step.
The user can choose to cancel logging on the system.

After the operation unit is checked and initialized, the startup screen will be
displayed. The operation unit begins communicating with and initializing the
analyzing unit. In case that it is the first startup or the last startup was failed, you can
select a serial port and then click Start to start the analyzing unit. Otherwise the
system starts the analyzing unit automatically.

Communication with and initialization of the analyzing unit is finished during startup.

1 To set and initialize the communication port


Set the parameters regarding the serial port: baud rate, data bit, start bit,
stop bit, parity bit, sending/receiving buffer, control protocol, and start the
serial port receiving/sending thread and received data processing thread.
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
2 To shake hand with the analyzing unit
A handshake instruction is sent to the analyzing unit. If the analyzing unit
responds to it by sending a handshake instruction back to the PC, it
indicates the PC handshakes with the analyzing unit successfully.
Otherwise, another one will be sent again for 3 times at a specific interval
(2s).
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
3 To check the self-check results of the analyzing unit
An instruction is sent to check the self-check result of each unit. The units
have been checked when the analyzing unit is started, and the results are
stored in the main unit.
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
4 To download parameters to the analyzing unit
An instruction is sent to download parameters to the analyzing unit. If there
is no response within 30s, the instruction will be sent again for 3 times.
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
5 To initialize the ISE unit and bar code reader
The instruction <ISE?> is sent to the ISE unit to check if it is connected. If
error occurs, ISE measurement can not be started.
ISE initialization: includes cleaning and calibration. If auto cleaning or
calibrator is configured for startup, ISE measurement will start if the
calibration is passed; otherwise another calibration is needed before
measurement.
Bar code reader initialization: includes query of bar code parameters. <H> is
sent to turn on the bar code reader, and <I> to turn off it.

5-3
Software Introduction

6 To reset the mechanical parts


An instruction is sent to reset mechanical parts of each unit.
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
7 To switch off the lamp and test the dark current
If the dark current is beyond the specified range, a warning will be given but
will not affect the startup and testing.
8 To switch on the lamp and test the background
A dialog box is displayed to ask you to unload the first cuvette segment, so
as to test the dark current and background. After unloading, the system
starts testing the background.
If unloading is failed, a prompt will be given and the startup is halted. You
can take actions immediately and try again, or exit the startup process.
If the background approaches to the dark current, a prompt will be given
indicating lamp on failed, lamp failure or brightness too low, and the startup
is halted. You can exit the startup process.
If the background is less than the specified threshold, a prompt will be given
indicating brightness too low, which will affect the test results. You can
neglect it or exit the startup process.
If the background approaches to the threshold, the startup process will not
be affected. However, when the main screen is displayed, a prompt will
appear indicating that the brightness approaches to the threshold and the
lamp should be replaced.
9 To test cuvette blank
A dialog box is displayed to ask you to replace all the cuvette segments with
new ones. After confirmation, the analyzer will start testing the cuvette
absorbance and mark the used cuvettes.
If loading is failed, a prompt will be given and the startup is halted. You can
take actions immediately and try again, or exit the startup process.
10 To reset the mechanical parts and wash the probe and bar
An instruction is sent to reset all the mechanical parts and wash the interior,
exterior of the probe and the mixing bar.
If failed, a prompt will be given and the startup is halted. You can take
actions immediately and try again, or exit the startup process.
11 To check the temperature and finish startup process
If the temperature becomes steady and is within specified range, the
analyzer will enter standby status, otherwise the incubation status.

5.1.1.3 Shutdown

This module restores the operation unit and analyzing unit to the default status.

During this process, the analyzing unit and the operation unit will be shut down.

You can start the shutdown process by selecting Exit on the main screen or Shut
Down on the Maintenance → Alignment → System screen. Then the system will
display a dialog box and start the shutdown automatically.

The following operations are conducted to shut down the analyzing unit.

5-4
Software Introduction

1 To switch off the lamp


If failed, the system will enter the failure status and the shutdown is ended.
2 To unload all the used cuvettes
When a prompt is displayed to ask you to unload all the used cuvettes, you
should unload them and confirm. Then the system will continue the
shutdown. If failed, the system will enter the failure status and the shutdown
is ended.
3 To clean the ISE unit
This step will be performed only if auto cleaning is configured for shutdown.
4 To wash the probe and the bar
When a prompt is displayed to ask you to prepare acid or alkaline detergent,
you should prepare it and confirm. Then the system will wash the probe and
the bar in enhanced washing mode. If failed, the system will enter the failure
status and the shutdown is ended.
5 To reset the mechanical parts
Each mechanical part will be reset. If failed, the system will enter the failure
status and the shutdown is ended.

The following operations are conducted to shut down the operation unit.

1 To stop all the threads


2 To stop the timer
3 To close the database

5.1.1.4 Database

This module saves the settings and test data, such as system, patient, sample
information and test data.

5.1.1.5 Interface

This module provides an interface for you to view each status and result, make
settings, request tests, start and stop testing, etc.

5.1.1.6 Serial Port Receiving/Sending

This module establishes the communication between the operating software and
control software. It sends the instructions in the sending buffer to the main unit,
through which the instructions are transferred to the specified units. Then the echo,
results, test data and warnings received from the main unit are saved in the received
data buffer.

5.1.1.7 Processing Received Data

This module processes the received echo, results, data, warnings, running reports
and requests. It gets the frame information from the received data buffer, sets
corresponding commands, records the measured photoelectric data and warnings,
and calls the error processing module when an error occurs.

5-5
Software Introduction

5.1.1.8 Error Processing

When an error occurs, this module finds the causes and evaluates the level and type
of the error. Additionally, it records the error logs, reminds you of error, and refreshes
the system status.

5.1.1.9 Testing

The testing module is to:

 Check the reaction temperature periodically.


 Form a test list according to the requests and divide the testing into several
periods, which are converted to instructions and saved in the buffer.
 Maintain the test list according to the results from the data processing module.
 Inform the database and interface modules to update relevant information.

5.1.1.10 Commissioning

This module sends relevant commissioning instructions and processes the


commissioning results.

It sends instructions regarding simple operations, commissioning, parameter


downloading, temperature control switch, and etc, and processes the corresponding
results.

5.1.1.11 Failure Restore

This module restores the failed parts as required by you.

The failed parts can be restored actively or passively. The active restoring means
when a failure occurs during startup, shutdown or testing, the system will try to
execute the instruction again in terms of settings or reset the mechanical parts of
corresponding units and wash the probe/bar. While the passive restoring means the
system restores the failed parts only when you sends the reset instruction.

5.1.1.12 Replacing Cuvettes

This module assists you in unloading used cuvettes and loading new ones.

Cuvettes can be replaced during startup, shutdown or under normal condition.

Normal condition refers to incubation, standby and failure status. Cuvettes are
replaced in same procedures during startup, shutdown and under normal condition.

After replacing, the system will test the cuvette blank automatically and mark the
used cuvettes.

5.1.1.13 Printing

This module prints out in proper format the test results, calibration results, QC results
or statistic results.

5-6
Software Introduction

5.1.1.14 Protocol Encapsulation

This module provides encapsulation functions for all the instructions in accordance
with the communication protocol.

5.1.2 Control Software


The control software receives the instructions from the operating software, analyzes
them and drives the corresponding units of the analyzing unit to execute them. And it
also transmits the data from the analyzing unit to the operating software.

The units of the analyzing unit can be functionally divided into the following ones:
main unit, photoelectric unit, reaction unit, sample/reagent unit, mixing unit and
temperature unit. The functions of these units are listed in the following table.

Unit Function
Main unit Receives macroinstructions from the PC,
decomposes them into a series of action instructions,
and then delivers them to destination units (in specific
order) at a certain interval.
Monitors the status of each unit, and transmits the
data converted by the photoelectric unit to the PC.
Photoelectric unit Detects photoelectrical data of the solution in reaction
cuvettes, and amplifies and converts it. And stores
data at the dual port FIFO for the main unit to read
and transmit to the PC.
Reaction unit Contains 80 cuvettes.
Rotates specified cuvette, following the instruction
sent from the main unit, to the reagent/sample
dispensing position, mixing position and photoelectric
detection position at specified wavelength.
Sample/Reagent unit Contains 40 reagent positions in the inner circle and
40 sample positions in the outer circle.
Rotates specified reagent bottles, following the
instruction sent from the main unit, to the reagent
aspirating position.
Rotates specified sample tubes, following the
instruction sent from the main unit, to the sample
aspirating position.
Mixing unit Receives instructions from the PC, and controls the
mixing bar to mix the solution in the cuvette where the
reagent and sample has been dispensed and that has
been rotated to the mixing position.
Moves the mixing bar to the wash well, and washes it
after each mixing to avoid carryover.
Temperature unit Controls the reaction temperature, reagent preheating
temperature and refrigeration temperature.

5-7
Software Introduction

5.2 Software Installation


Check the operating system and relevant settings, and ensure that a Window XP or
Windows Vista is used, with memory of above 1G, resolution of 1024*768, and color
qulaity of no less than 16 bit.

Check if the computer is connected to the analyzing unit and the analyzing unit is
powered on.

For the serial number of the operating software, contact Mindray customer service
department.

5-8
6 Maintenance and Service

6.1 Cleaning and Washing


WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.

BIOHAZARD:
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal and consult the manufacturer
or distributor of the reagents for details.

6.1.1 Cleaning Probe


1 Place the Power to OFF.
2 Remove the cover from the sample/reagent disk.
3 Remove the sample/reagent disk.
4 Pull the probe arm to the highest point by hand. Rotate the probe arm to
move the probe to a position above the sample/reagent compartment and
convenient to operate.
5
CAUTION:
The tweezers may scratch the probe. Exercise caution when
using it to clean the probe. Avoid direct contact between the
tweezers and the probe. Do not use excessive force when
cleaning the probe. Otherwise it may bend.

NOTE:
Mindray recommends the acid and alkaline detergents be used
alternately for this purpose. For instance, if the acid detergent
has been used for last maintenance, the alkaline detergent had
better be used for this time.

Pinch acid or alkaline detergent-soaked gauze with tweezers and gently


clean the exterior of the probe until it is clean and smooth.
6 Pinch deionized water-soaked gauze to clean the probe.
7 After cleaning, gently pull the probe arm to its highest point and rotate the
probe arm to move the probe to a position above the wash well.
8 Load the sample/reagent disk.
9 Close the cover.

6-1
Maintenance and Service

6.1.2 Cleaning Mixing Bar


1 Place the Power to OFF.
2 Pull the mixing bar arm to the highest point by hand. Rotate the bar arm to
move the bar to a position convenient to operate.
3
CAUTION:
The tweezers can scratch the bar. Exercise caution when using
the tweezers to clean the bar. Avoid direct contact between the
tweezers and the bar. Do not use excessive force when cleaning
the bar. Otherwise it may bend.

NOTE:
Mindray recommends the acid and alkaline detergents be used
alternately for this purpose. For instance, if the acid detergent
has been used for last maintenance, the alkaline detergent had
better be used for this time.

Pinch acid or alkaline detergent-soaked gauze with tweezers and gently


clean the exterior of the mixing bar until it is clean and smooth.
4 Pinch deionized water-soaked gauze to clean the mixing bar.
5 After cleaning, gently pull the bar arm to its highest point and rotate the bar
arm to move the bar to a position above the wash well.

6.1.3 Cleaning Sample/Reagent Compartment


1 Place the Power to OFF.
2 Remove the cover from the sample/reagent disk.
3 Take out all calibrators, controls, samples, reagents, distilled water and
detergent from the sample/reagent disk.
4 Remove the sample/reagent disk.
5 Wash the disk with clean water and wipe it dry with clean gauze.
6 Use clean gauze (water or disinfector-dipped gauze if necessary) to clean
the inside of the compartment.
7 Load the sample/reagent disk.
8 Close the cover.

6.1.4 Cleaning Wash Well of Probe


1 Place the Power to OFF.
2 Pull the probe arm to its highest point. Rotate the arm to move the probe
away from the wash well.
3 Clean the inside of and the place around the wash well with cotton swabs.
4 Pull the probe arm to its highest point and rotate it to move the probe to a
position above the wash well.

6-2
Maintenance and Service

6.1.5 Cleaning Wash Well of Mixing Bar


1 Place the Power to OFF.
2 Pull the mixing bar arm to its highest point. Rotate the bar arm to move the
bar away from the wash well.
3 Clean the inside of and the place around the wash well with cotton swabs.
4 Pull the mixing bar arm to its highest point and rotate it to move the bar to a
position above the wash well.

6.1.6 Cleaning Panel of Analyzing Unit


1 Place the Power to OFF.
2 Wipe the panel of the analyzing unit with clean gauze (water or
disinfector-dipped gauze if necessary).

6.1.7 Washing Deionized Water Tank

CAUTION:
The deionized water to be used on the analyzer must meet the CAP
Type II water requirements.

1 Place the Power to OFF.


2
CAUTION:
After removing the cap of the deionized water tank (together
with the pickup tube and sensor), place it on a clean table.

Unscrew (counter-clockwise) the cap (together with the deionized water


pickup tube and the sensor).
3 Wash the tank interior with deionized water. Use a clean brush to clean the
interior if necessary.
4 Wash the pickup tube and the sensor with deionized water. Use clean gauze
to wash them if necessary.
5 Wipe water off the tank exterior, pickup tube and sensor cable with clean
gauze.
6 Add deionized water into the tank.
7 Screw (clockwise) the cap (together with the pickup tube and sensor) back
onto the tank until secure.

CAUTION:
When placing the deionized water tank, ensure the top of the tank is
lower than the bottom of the upper cabinet.
Ensure the deionized water pickup tube is not blocked, bent, or
twisted.

6-3
Maintenance and Service

6.1.8 Washing Waste Tank

BIOHAZARD:
Exercise caution and do not spill the waste onto other people or
things.

1 Place the Power to OFF.


2
BIOHAZARD:
After removing the cap of the waste tank (together with the
waste tube and sensor), place it on an appropriate place to
avoid biohazard contamination.

Unscrew (counter-clockwise) the cap (together with the waste tube and the
sensor).
3 Empty the waste tank.
4 Wash the tank interior with clean water. Soak the tank with disinfector if
necessary.
5 Wash the waste tube and the sensor with clean water.
6 Wipe water off the tank exterior, waste tube and sensor cable with clean
gauze.
7 Screw (clockwise) the cap (together with the waste tube and sensor) back
onto the tank until secure.

CAUTION:
When placing the waste tank, ensure the top of the tank is lower than
the bottom of the upper cabinet.
Ensure the waste tube is over the tank and not blocked, bent, or
twisted. A blocked, bent or twisted waste tube may lead to wastewater
overflow that may damage the analyzer.

6.1.9 Washing Dust Screens


1 Place the MAIN POWER to OFF.
2 Use a screwdriver to unscrew the screws on the left and right plates and
remove the two plates.

6-4
Maintenance and Service

3 Remove the dust screens from the plates. The dust screen is as shown in
the figure below.

4 Wash the screens with clean water and dry them by airing.
5 Install the screens back to the left and right plates.
6 Install the left and right plates with screws.

6.2 Replacing
WARNING:
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.

BIOHAZARD:
Wear gloves and lab coat and, if necessary, goggles.

CAUTION:
Please use Mindray-recommended consumables. Other consumables
may decrease the system performance.

6.2.1 Unclogging Probe


When the probe is clogged, the fluid flow will become abnormal. Follow the steps
given below to remove, unclog and install the probe.

6.2.1.1 Removing Probe

1 Place the Power to OFF.


2 Remove the cover from the sample/reagent disk.

6-5
Maintenance and Service

3 Pull the probe arm to its highest point. Rotate the probe arm to move the
probe to a position above the sample/reagent compartment and convenient
to operate.

4 Loose the screws on the arm cover and remove the cover upward from the
arm base. After you remove the cover, the inside structure of the probe arm
is as shown in the figure below.

Probe’s Circuit Connector


Circuit Board Tubing Connector
Probe’s Fluid Connecor
Retaining Screw

Spring

5 Hold the probe’s fluid connector with one hand and the tubing connector with
the other. Rotate the tubing connector counter-clockwise until it disconnects
from the probe. Remove the tubing from the probe.
6 Press the circuit board with one hand and disconnect the probe’s circuit
connector from the board with the other hand.

CAUTION:
Exercise caution when disconnecting the connector. Excessive
force may damage the connector and/or the circuit board.

7 Use a small screwdriver to remove the retaining screw on the probe and take
out the spring.
8
WARNING:
Store the removed probe in a safe place where it will neither
endanger people working around the area nor be damaged.

NOTE:
Exercise caution when pulling the probe away from the arm.

Slowly pull the probe away from the probe arm. Exercise caution so that the
gasket inside the probe does not drop out and if it does, store it in a clean
place for later installation.

6-6
Maintenance and Service

6.2.1.2 Unclogging Probe

BIOHAZARD:
Dispose of the used needle in accordance with your local or national
guidelines for biohazard waste disposal.

1 Place a container like beaker below the sample probe.


2 Connect the unclogging device to the tubing connector.
3 Use a single-use syringe to aspirate 3ml CD80 wash solution, remove the
needle and connect the syringe to the other end of the unclogging device.
Push the syringe plunger slowly until there comes liquid out of the sample
probe tip.

4 If no liquid comes out of the sample probe tip, insert a needle into the
sample probe tip and push the syringe plunger.
5 Leave the sample probe soaked with wash solution for about 10 minutes.
Push and pull the syringe plunger for several times until liquid comes out
of the probe tip evenly.
6 Use the syringe to aspirate deionized water and rinse the sample probe for
at least 3 times.
7 Remove the unclogging device and the syringe, and then connect the
tubing connector.
8 Remove the container.
9 Check the level detection function.

NOTE
A bent or damaged probe will lead to unreliable test results and should
be replaced immediately.

6.2.1.3 Installing Probe

1 Place the Power to OFF.


2 Insert the probe back into the hole on probe arm, and align the hole on probe
plate to the rotor inside the arm.
3 Sleeve the spring on the rotor and screw the retaining screws to secure.
4 Pinch the probe by the part near the probe arm. Gently push the probe
upward and then release the probe to see if the spring can move freely.
If so, proceed to the next step.
If not, check for errors and try again after removing the errors.

6-7
Maintenance and Service

5 Connect the probe’s circuit connector back to the circuit board.


6 Ensure the gasket is inside the probe.
7
CAUTION:
The fluid tube inside the probe arm should be bent into a
circle when being installed.

Exercise caution when connecting the probe. Excessive


force may bend the probe.

Screw (clockwise) the probe’s fluid connector back to the tubing connector.
8 Place the ANALYZING UNIT POWER to ON while ensuring that the sample
probe is not attaching any conducting .object, such as hands.
9 Calibrate the sample probe manually. Check if indicator D2 (yellow) is
lightened within 2 seconds when the ANALYZING UNIT POWER to ON. Press
the switch S2 on the level detection board and then release it, then check if
indicator D2 is first extinguished and then lightened. If it is, that means the
calibration succeeds.
Exercise caution to prevent the sample probe from attaching any conducting
object during the calibration.

6-8
Maintenance and Service

10 Add deionized water to a clean cup. Immerse the probe tip into the water by
2-3mm and indicator D5 on the circuit board should be lighted. Take the probe
tip out of water, and the indicator should go out. If the test succeeds, proceed
to the next step; if not, please contact our Customer Service Department or
the distributor.

11 Install the cover back to the probe arm and tighten the screws on the cover.
Ensure that the square end is facing the probe and the round head is facing
the rotor.
12 Pull the probe arm to its highest point and rotate it to move the probe to a
position above the wash well.
13 Install the sample/reagent disk.

6.2.2 Replacing Probe


If the probe is bent or damaged, it must be replaced immediately. Follow the
procedure given below to replace the damaged or bent probe.

1 Remove the bent or damaged probe as instructed by 6.2.1.1 Removing


Probe.

BIOHAZARD:
Dispose of the bent or damaged probe in accordance with your
local or national guidelines for biohazard waste disposal.

2 Install a new probe as instructed by 6.2.1.3 Installing Probe.

6.2.3 Replacing Mixing Bar


If the mixing bar is damaged, it must be replaced immediately. Follow the procedure
given below to replace the damaged mixing bar.

WARNING:
When replacing the bar, pinch the bar only by the knurled part and do
not touch the other part of the bar. Protect the flat part of the bar from
been scratched.

BIOHAZARD:
Dispose of the damaged mixing bar in accordance with your local or
national guidelines for biohazard waste disposal.

6-9
Maintenance and Service

1 Place the Power to OFF.


2 Gently pull the bar arm to its highest point and rotate it to move the bar to a
position convenient to operate.
3
CAUTION:
When trying to pull out the bar, concentrate your force in the
direction of the axis on the bar arm. Biased force may damage
the bar and/or the axis.

Pinch the bar by the knurled part with one hand and unscrew
(counter-clockwise) the retaining nut with the other hand until the mixing bar
looses. Pull the bar downward to remove it and remove the nut.

Retaining Nut
Knurled Part

4 Align the new mixing bar to the bigger hole end of the retaining nut and
gently screw it into the nut until the end of the bar is in line with the smaller
hole end of the nut.
Retaining Nut
Knurled Part

5 Pinch the mixing bar by the knurled part and align the hole of the nut to the
axis on the bar arm, then push the bar upward in the direction of the axis
until it can’t proceed. Tighten the nut by screwing clockwise with the other
hand.

CAUTION:
When trying to push the bar, concentrate your force in the
direction of the axis on the bar arm. Biased force may damage
the bar and/or the axis.
Ensure the bar is all the way pushed to the end.

6 After replacing the bar, visually check whether the bar is vertical to the bar
arm.
If not, remove the bar and re-install it.
If so, proceed to the next step.
7 Pull the bar arm to its highest point and rotate it to move the bar to a position
above its wash well.

6-10
Maintenance and Service

6.2.4 Replacing Plunger Assembly of Syringe


The old plunger assembly of syringe should be replaced with a new one when

 The old one has served for three months; or


 The old one has been used for over 100,000 tests; or
 The old one is apparently damaged.

CAUTION:
Exercise caution when installing the assembly. Excessive force may
crack the syringe.

1 Place the Power to OFF.


2 Unscrew the screws on the syringe cover and remove the cover.
The structure of the syringe is as shown in the figure below.

Tee
Syringe Connector
Retaining Screw
Retaining
Screw Space Bar
Syringe
Retaining Screw
Retaining Space Bar
Screw
Plunger Guide Cap

Holder

Plunger Button

Lower Retaining Screw

3 Prepare a new plunger assembly (shown in the figure below) and soak the
plunger tip in deionized water to eliminate bubbles.

Plunger Tip Plunger Rod Plunger Button


Plunger Guide Cap

4 Unscrew (counter-clockwise) the lower retaining screw.


5 Unscrew (counter-clockwise) the four retaining screws, remove the screws
and space bars, and remove the syringe from the holder.

6-11
Maintenance and Service

6
CAUTION:
There may be residual water in the syringe connector. Do not
drop water onto the analyzing unit.

Grab the Tee with one hand and the syringe connector with the other hand
and unscrew (counter-clockwise) the syringe.
7
CAUTION:
There may be residual water in the syringe. Do not drop water
onto the analyzing unit.

Unscrew (counter-clockwise) the plunger guide cap and pinch the plunger
button to gently pull the plunger assembly from the syringe.

Syringe Connector Syringe Plunger Guide Cap


Plunger Rod Plunger Button

8 Pinch the new plunger assembly by the plunger button and carefully insert
the plunger tip into the syringe and push it all the way to the end. Screw
(clockwise) the plunger guide cap until secure.
9 Immerse the syringe connector into deionized water. Pinch the plunger
button, pull it to aspirate half syringe of deionized water and then push it to
expel the deionized water and the air from the syringe.
10 Grab the Tee with one hand and the syringe connector with the other hand.
Screw (clockwise) the syringe into the Tee until secure.
11 Place the syringe on the holder. Install space bars and fix retaining screws.

NOTE:
The upper edge of the upper space bar must reach the seventh
line of the scale on the syringe.
When fixing retaining screws, be sure to tighten them
alternately with equilibrium force.

12 Screw (clockwise) the lower retaining screw until secure.


13 Place the Power back to ON.
14 Enter the Alignment screen of the operating software and set the Vol.(R.
Syringe) to 450. Click R. Syringe Aspirate. After the syringe finishes the
motion, click R. Syringe Dispense. You may repeat this action several
times.
Pay attention to bubbles during the aspiration/dispensing process.
If there are bubbles observed during the process, they may be caused by the
air leak between the syringe and the Tee. Uninstall the syringe and re-install
it.
If the bubbles are found again, there is something wrong with the fluid path.

6-12
Maintenance and Service

6.2.5 Replacing Lamp


Replace the lamp with a new one when the system reminds to do so, or the service
time of the lamp has added up to 2,000 hours.

CAUTION:
Do not touch the light entrance of the lamp. In case the entrance is
dirty, clean it with ethanol-soaked defatted cotton.

1 Place the MAIN POWER to OFF. Wait at least 15 minutes for the lamp and its
housing to cool down.

WARNING:
After working for a while, the lamp and its housing are usually
hot enough to burn you. Do not proceed with this procedure
until they have cooled down.

2 Remove the right cover of the analyzing unit.

3 Unscrew the two cable connectors, and then remove the cables.

6-13
Maintenance and Service

Remove the retaining screw on the lamp.


4

Remove the lamp.


5

6 Install the new lamp in the reversed steps.

7
Install the right cover of the analyzing unit.

8 Place the Power back to ON.


9 Log on the test and maintenance software. Username: bs200, Password:
Bs200@Mindray!

6-14
Maintenance and Service

10 Wait 10 minutes until the light source is stable, and then enter the
Maintenance screen and click New Lamp. Follow the software instruction
to refresh the air blank AD value (Background value).

11 Check the air blank AD (Background value) after executing New Lamp.The
replacement is successful when the value of all channels is below 65535.If
the background is above 65535, the background overflows and it is
necessary to adjust the gain.

6.2.6 Replacing the Optical Filter


It is necessary to replace the optical filter when it is damaged or aged. The steps of
replacing the optical filter are as follows.

1 Place the MAIN POWER to OFF.


2 Remove the right plate and right top plate of the analyzing unit.

3 Pull out the sensor and the motor cable socket.

6-15
Maintenance and Service

4 Unscrew the two screws fixing the dustproof cover and remove the plate.

Dustproof cover

5 Unscrew the two screws fixing the filter wheel driver assembly to remove it.
NOTE: While removing the screws, hold the filter drive assembly to prevent it
from falling off and being damaged.

6-16
Maintenance and Service

6 Use the fixture BA10-J12-01 to cover the filters and then unscrew the 8 screws
fixing the filter by using the cross-head screwdriver. It is necessary to use the
fixture to protect the filters from being scratched. If you can not find the fixture
on the spot, handle carefully.

7 Push the filter out with cotton swabs..


The corresponding relationship between the filter and the filter wheel installation
hole is :
Wavelength
340 405 546 670 450 510 578 630
(nm)
Hole
1 2 3 4 5 6 7 8
Position
Pay attention to the direction of the filter wheel during installation. The arrow on
the filter wheel indicates the direction of the light path, or put the end surface
with chamfer angle to the bottom of the installation hole.

Self-attack Screw Gasket

Filter

Fliter Wheel

8 Install the filter to the filter wheel hole with the right order and direction. Cover
the filter with fixture BA10-J12-01 and tighten the 8 screws on the filter. If you
can not find the fixture on the spot, handle carefully.
9 Install the sensor and the motor cable socket.

6-17
Maintenance and Service

10 {<}0{>Install dustproof cover.<0}


11 Install the right plate and the right top panel

NOTE:
1. Wear white clear cotton gloves while operating.
2. Do not get the filter damaged while screwing the screws.
3. Press the filter gently into the installation hole on the filter
wheel.
4. If there is dirty on the surface of the filter, clean it with
ethanol-soaked defatted cotton.
5. Check the rotation of the filter wheel.

6.2.7 Replacing the Optical Fiber


1 Place the MAIN POWER to OFF.
2 Unscrew the four screws on the reaction disk cover.

6-18
Maintenance and Service

3 Open the reaction disk cover. Pay attention to the power cord of the
heater.

Power cord
of the heater

4 Remove the three M3 socket screws and take out the reaction disk.

5 Use a hex wrench to loosen the screw that is used to fasten the optical
fiber on the optical base.

6 Take out the optical fiber, and fix the optical component in the optical base
by fastening the screws slightly.

6-19
Maintenance and Service

7 Nip off the cable tie and draw out the optical fiber from the bottom of the
reaction compartment.

8 Loosen the M3 screw (used to retain optical fiber) on the lamp housing,
and then draw out the optical fiber.

9 Put the new optical fiber into the reaction disk through its bottom, loosen
corresponding screw, insert the optical fiber to the end, and then fasten the
screw.
10 Loosen the screw of the lamp housing, insert the optical fiber, and then
fasten the screw and fix the fiber on the shore use cable tie.
11 Put back the reaction disk and fix it.
12 Put back the reaction disk cover and fix it.

NOTE:
When replacing the optical fiber, ensure that its bending radius is no
less than 20cm, or the fiber will be damaged.

6.2.8 Replacing Filter Assembly


When used for a long period, the filter may get clogged with the result of decreased
wash flow and degraded system performance. The filter should be replaced every
six months. The filter assembly must no be canceled, if it is missing, the pumps and
valves can be blocked, putting the system into failure status.

1 Remove cap assembly from the DI water tank and place it on a clean
desktop. Carefully remove the filter assembly from the cap assembly.

6-20
Maintenance and Service

2 Inject water into the new filter assembly through the big adapter by using the
syringe. When water wells up from the small adapter, the injection is
completed. Purpose: increase the weight of the filter assembly to make it
sink to the bottom of the water tank.

3 Connect the new filter assembly to the tubes of the cap assembly, the small
adapter to the thin tube and the big adapter to the thick tube. Keep the two
tubes parallel and straight to avoid tangling.

4 Expel air from the filter assembly


1) Put the cap assembly into the deionized water tank and tighten the cap.
2) Select Fluidic System on the Maintenance > Alignment tab page, select
Exterior, and then select Clean R. Probe to clean the reagent probe exterior.
Repeat this operation for 10 times to remove the air from the filter assembly
and the wash tube.

6-21
Maintenance and Service

3) During the washing process, check for air bubbles entering the wash tube
from the filter. If you find air bubbles, repeat the previous step; otherwise,
proceed to the next step.

4) Check the top of the syringe for air bubles.

Check here for air


bubbles

5) If air bubbles exist, remove them according to the instructions in "5.6.5


Removing Air Bubbles" of the Operator's Manual. Exercise caution not to
damage the syringe.

6-22
Maintenance and Service

5 The replacement is completed. Avoid introducing foreign matters to the filter


or the tube when replacing the filter assembly.

6.3 Repairing the Bar Code Module

6.3.1 Replacing the Bar Code Reader


Installation procedure:

1. Install the small bracket on the bracket assembly by using two M3×8 hexagon
socket screws with spring and flat washers.

2. Install the bar code reader on the small bracket by using two M3×10 hexagon
socket screws with spring and flat washers.

Precautions:

1. During installation, keep the glass surface of the reader clean to avoid influencing
the scanning effect.

2. Do not tighten the screws on the bar code reader until you finish the alignment.

Anti-fogging heater assembly

Small bracket

Bar code reader

Bracket assembly

6.3.2 Installing the Anti-fogging Heater Assembly


1. Place the silicone pad in the slot of the anti-fogging device mounting plate.

2. Place the anti-fogging glass on the silicone pad.

6-23
Maintenance and Service

3. Lay the heat-conducting aluminum plate with the groove side facing upwards, and
use six M2×8 cross pan head screws to fix it on the mounting plate.

4. Stick a heater to the groove on the aluminum plate.

5. Daub some thermal paste on the bottom side of the overheat protection switch
and place the switch on the heater. Use an M3×12 hexagon screw to fix the
protection switch with the switch plate.

6. Use four M3×6 pan head screws to fix the dust shield on the mounting plate.

7. Stick a sponge cushion to the dust shield.

Precautions:

1. Align the window of the silicone pad with that of the mounting plate.

2. Ensure that the glass is not cracked or dirty.

3. Lay the cables of the heater and overheat protection switch downwards as shown
in the figure below for the purpose of convenient connection.

4. Align the window of the sponge cushion with that of the dust shield in order to
ensure that the scanning light comes smoothly.

Mounting plate of
the anti-fogging
device Anti-fogging glass Heater Switch plate Dust shield Sponge window

Silicone pad
Heat-conducting Overheat
aluminum plate protection switch

6.4 Bar Code Reader Adjustment

NOTE:
Do not stare into the lamp or the laser beam from the bar code reader
when the analyzer is in operation.

6.4.1 Adjusting Sample Bar Code Reader


Requirements:

The light beam of sample bar code reader is about 1mm wide and is centered to
around 0.5mm from the left of the scanning hole.

6-24
Maintenance and Service

Checking:

1) Place a bar-coded sample tube on the sample/reagent disk.

2) Start scanning the bar code label on the sample tube and turn on the laser.

3) Check the position of light beam according to the scanning hole.

6-25
Maintenance and Service

Adjustment:

1) Turn on

2) If deviation in direction A (see figure above) occurs, loosen the screw A and
adjust the bar code reader until the position of light beam meets the
requirements.

3) If deviation in direction B (see figure above) occurs, loosen the screw A and
adjust the bar code reader until the position of light beam meets the
requirements.

6-26
Maintenance and Service

6.4.2 Adjusting Reagent Bar Code Reader


Requirements:

The light beam of reagent bar code reader locates within 1mm from the center line
of the reagent bottle.

Movement
direction

Light beam

1mm

Checking:

1) Place a bar-coded reagent bottle on the sample/reagent disk.

2) Start scanning the bar code label on the reagent bottle and turn on the laser.

3) Check the position of light beam on the reagent bottle.

Adjustment:

1) First check if light beam of sample bar coder reader is adjusted properly. If not,
adjust it and then check the light beam position of the reagent bar code reader.

2) If the light beam of sample bar code reader is adjusted properly but the position
of reagent bar code reader light beam does not meet the requirements, please
adjust the rotation deviation of sample/reagent disk while considering the
reagent aspiration position and bar code scanning position.

6-27
Maintenance and Service

6.5 ISE Unit (optional) Maintenance

6.5.1 Replacing tubings

Note

Before replacing tubings, please follow the steps below:

1. Take off the ISE Unit gate

2. Take off the joint of Cal A tubing of the wand from that of the pump tubing

3. Lift up the Cal A tubing of the wand to make the solution is the tubing flow back
to the reagent pack.

4. Do the same operation to Cal B tubing.

6-28
Maintenance and Service

Note

Do not take off the waste tubing of the wand

5. Use the maintenance and test software to purge A and B for 25 times, then
replace the tubings. After the replacement, insert the tubings of the wand which
has been taken off before back to their original position. Then purge A and B for
another 25 times to ensure the replacement is successful.

6.5.1.1 Replacing Calibration tubings

1. Take off the tubing joint from the pump tubing, and another end from the ISE
sample port

2. Take off the joint from the replaced tubing.

6-29
Maintenance and Service

3. Insert the joint into the new tubing and make it connected to the ISE sample port
and the pump tubing.

6-30
Maintenance and Service

6.5.1.2 Replacing pump tubings

1. Take off the tubings which are connected to the pump tubing.

2. Hold on one side of the pump tubing, and lift up a little then take it off from the
pump shelf.

3. Take another side of the pump tubing from the shelf.

6-31
Maintenance and Service

4. Install the new pump tubing to the pump shelf and insert the tubings which have
been taken off in step 1 back to the new tubing.

6.5.1.3 Replacing the wand


After emptying the solution in the tubing, take off the waste pump of the wand and
change a new one.

Note

Do not insert the tubings of the wand into the wrong place of the pump tubings,
otherwise it will cause function failure.

6.5.2 Replacing ISE module


When the ISE module is damaged, it needs to be taken out of the analyzer with the
shielding box and then be replaced.

1. Disconnect the communication lines in the ISE module, and take off the Cal A
and Cal B tubings.

2. Release the screws which fix the shielding box to the main board.

3. Take off the shielding box from the analyzer

6-32
Maintenance and Service

4. Release the four screws on top of the shielding box to replace the ISE module

5. After the replacement, put the shielding box back to the analyzer. Tight the
screws and insert the Cal A and Cal B tubings back to the sample port.

6.5.3 Replacing ISE Components (optional)

6.5.3.1 Replacing Reagent Pack

1 Place the POWER to OFF.


2 Open the ISE unit door.
3 Remove and install a new reagent module.
4 Enter the ISE screen of the Maintenance of the system software.
5 Select the Daily Maint tab.
6 Enter digit “25” in the blank on the right of the button Purge Comb, then
click it.
7 Click Purge A and Purge B button to check whether the initialization of
the Reagent Pack is finished. If no error occurs during the process, the
Reagent Pack is replaced successfully.

6.5.3.2 Replacing Electrodes


If you run no more than 100 samples requested for the ISE tests a day, replace the
electrodes according to the following recommended schedule:

6-33
Maintenance and Service

Na+ Electrode Every 6 months 10000 samples


+
K Electrode Every 6 months 10000 samples
-
Cl Electrode Every 6 months 10000 samples
Reference Electrode Every 6 months 10000 samples

Note:

Since the electrodes must be installed sequentially, you have to take out the
electrode to be replaced and those (or that) over it from above to below.

1 Enter the ISE screen of the Maintenance of the system software.


2 Select the Daily Maint tab.
3 Click the Maintenance button.
4 Replace the electrodes.
5 Click Purge A button and if no error occurs during the process, it means
the electrode is replaced successfully.

6.5.3.3 ISE Unit Storage (optional)

Caution:

 The maintenance is necessary to be performed when the ISE unit (optional) is


connected.
 The ISE unit (optional) should be on power all the time. In some cases that the
POWER will be shut down for a long time more than half an hour, the following
steps should be performed.
 Use the consumables recommended by our company. Other consumables may
degrade system performance.
1 Enter the ISE screen of the Maintenance of the system software.
2 Select the Daily Maint tab.
3 Click the Clean button to clean the ISE module.
4 Pull out the joint A and joint B of the wand tubing which has been inserted
into the adapters of the pump tubing. Hold them on for a few seconds until
the solution in the wand tubing flows back to Reagent Pack.

Joint

Adapter of
pump tubing

5 Enter digit “25” in the edit box to the right of the Purge Comb button, then
click the Purge Comb button.

6-34
Maintenance and Service

6 Click the Maintenance button.


7 Remove the electrodes.
8 Remove the Reagent Pack.
9 Put the reference, Na+ and Cl- electrodes into their individual sealed bags.
10 Aspirate a small amount of Calibrant A from the port of the reagent module
with a syringe and inject it into the lumens of the K+ electrode until the
lumens are full.
Cover both ends of the lumens with tapes to prevent the Calibrant A flows
from the lumens.
Put the K+ electrode into their individual sealed bags.

6.6 Adjusting

6.6.1 Adjusting Probe and Disks

NOTE:
Adjust the position of the probe according to the position of the cuvette
first, and then adjust other positions.
Before adjusting parameters at the PARA and Speed tab of the
Maintenance and Test Software, remember to click the Enable
modifying parameters button at the Command → Main Unit tab and
use a jumper to shortcut the 3.3V of J1 of the main control board; after
finishing adjusting, remember to use a jumper to shortcut GND of J1 of
the main control board.

1 Adjust the position of the probe.

NOTE:
To minimize cumulative errors, ensure that the probe returns
to the initial position before each adjustment.

(1) Adjust the dispensing position of the probe. Ensure the probe is in
the center of the cuvette and the probe tip is 2-3mm away from the
top of the cuvette.
(2) Adjust the position of the initial-position transducer of the
sample/reagent disk. Ensure the reagent bottles and sample tubes
fit the disk cover well.
(3) Adjust the washing position of the probe. Ensure the probe is in the
center of the wash well and the probe tip is 5mm away from the
bottom of the wash well. If necessary, adjust the position of the wash
well.
(4) Adjust the probe’s position on the outer circle of the disk. Ensure the
probe is in the center of the sample tube.
(5) Adjust the probe’s position on the inner circle of the disk. Ensure the
probe is in the center of the reagent bottle.

6-35
Maintenance and Service

(6) Adjust the probe’s position on the ISE sample port. Ensure the port
is with the circle of diameter of 1 mm around the center of the port.
2 Adjust the position of the mixing bar.
(1) Adjust the mixing position of the bar. Ensure the bar is in the center
of the cuvette on the reaction disk.
(2) Adjust the vertical position of the bar in the cuvette. Ensure the bar
tip is 1-2mm away from the bottom of the cuvette.
(3) Adjust the washing position of the mixing bar. Ensure the bar is in
the center of the wash well and about 5mm away from the bottom of
the wash well. If necessary, adjust the position of the wash well.
3 In the Maintenance and Test Software, select corresponding command
to rotate the reaction disk and move the probe and the mixing bar to see if
they clash with any reaction cuvette. If the mixing bar knocks the side or
the bottom of any cuvette, repeat Step 2.

6.6.2 Adjusting Signal Gain of Photoelectric Unit


1 Start the Maintenance and Test Software.
2 Select Command → Main Unit tab.
3 Click the Enable modifying parameters button.
4 Select Command → Reaction Unit tab.
5 Click the Turn on light button to turn on the lamp of photoelectric unit.
6 Select PARA and Speed tab.
7 Select the Reaction Unit from the pulldown list to the right of Unit Name.
8 Click the Query button to display the photoelectric gain of each wavelength.
9 Ensure there is no cuvette on the reaction disk.
10 Adjust and configure the photoelectric gain of each wavelength (except
700nm) at the PARA and Speed screen until the voltage of each wavelength
before A/D conversion is between 4.70V and 4.90V. You can measure the
voltage of each wavelength on the A/D conversion board with a multimeter.

NOTE:
Before adjusting, remember to use a jumper to shortcut the
3.3V of J1 of the main control board.
After finishing adjusting, remember to use a jumper to
shortcut GND of J1 of the main control board.

The following table displays the corresponding test points on the A/D
conversion board (the reference test point is TP12).
Wavelength 340 405 450 510 546 578 630 670 REF
Test Point TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8 TP10

6.6.3 Adjusting Photoelectric Collection Position


1 Start the Maintenance and Test Software.
2 Select Command → Main Unit.

6-36
Maintenance and Service

3 Click the Enable modifying parameters button.


4 Click the Mechanical Reset button to reset all the mechanical parts of the
analyzer.
5 Place 8 segments of new cuvettes in the reaction disk.
6 Measure the TP1 and TP11 with a dual trace oscillograph. The reference test
point is TP12. TP11 is the A/D Conversion Start Signal.
7 Select Command → Reaction Unit.
8 Click the Rotate and Measure button and observe the positions of the
pulses relative to the waveform from the oscillograph.
All pulses(the A/D Conversion Start Signal) should be generated at the flat
parts of the waveform and most of them should be at a little right of the
center of the flat parts, as shown in the figure below.

pulse
9 If the positions of the pulses do not meet the above requirements, you can
adjust the initial position of photoelectric collection according to the following
steps.
(1) Select PARA and Speed tab of the Maintenance and Test Software.
(2) Select Reaction Unit from the pulldown list to the right of Unit Name.
(3) Click the Query button.
(4) Adjust and configure the parameter of Initial position of photoelectric
collection and go back to the step 2 (skip step 3).

NOTE:
Before adjusting, remember to use a jumper to shortcut the
3.3V of J1 of the main control board.
After finishing adjusting, remember to use a jumper to shortcut
GND of J1 of the main control board.
10 Repeat through step 2 to step 9 (except step3) to measure the waveform of
other wavelengths whose test points are from TP2 to TP10 (except TP9).

6-37
Maintenance and Service

NOTE:
(1) If the waveform is oblique, as shown in the figure below, loosen the
screw that is used to fasten the optical fiber on the colorimetric rack or
the reference light support and rotate the fiber slightly (avoiding
breaking the fiber) until the waveform is not oblique. After adjusting,
fasten the screw.

oblique

not oblique
(2) The parameter of Initial position of photoelectric collection
cannot be less than 0. If it is 0 and the pulses still cannot meet
requirements, you need to adjust the optical coupler of the reaction
unit that is used to detect the position of the reaction disk.

6.7 Checking

6.7.1 Checking Reagent Refrigeration Board


 Test Device
1. A multimeter

2. A constant power supply

3. A resistor

4. 4 jumpers

5. A slotted screwdriver (2mm)

 Connection Diagram

Constant
Multimeter Resistor
power supply

Reagent refrigeration board

 Checking Procedure
1. To test the power supply

Connect the constant power supply to J86 (PIN 1, 5, 6 to 12V; PIN 2, 3, 4 to GND)
and J87 (PIN3 to +5V; PIN1 to GND). Use the multimeter to check the constant
power supply (12V±0.6V and 5V±0.3V). Power on the constant power supply and
check the power supply as following:

6-38
Maintenance and Service

1 9V: 9±0.45V; test point: the voltage at two terminals of the C2 capacitor.

2 5FAN: 5±0.25V; test point: the voltage at two terminals of the C14 capacitor.

3 5V: 5±0.25V; test point: the voltage at two terminals of the C16 capacitor.

4 2.5V: 5±0.25V; test point: the voltage at two terminals of the C7 capacitor.

2. To adjust the temperature control point

Adjust VR1 to make the TP1 (TP3: GND): 1.75V±0.01V.

Adjust VR2 to make the TP2 (TP3: GND): 1.84V±0.01V.

After adjustment, fix the resistance of VR1 and VR2 with a little glue or other
materials.

3. To adjust the temperature control

Connect a resistor to the slot of the J88 transducer and adjust the resistance to
simulate the variation of temperature.

Adjust the resistor to make TP5 (TP3: GND) supply switch between 1.85V (reference
resistance of resistor is 12.5K) and 1.75V (reference resistance of resistor is 10.2K)
and measure the voltage at 1 and 2 terminals of the J89.

1) If the TP5 supply is more than 1.84V, the output is 0V±0.6V.

2) If the TP5 supply decreases from 1.84V, the output doesn’t change. If it is no more
than 1.75V, the output is 12V±0.6V.

3) If the TP5 supply increases from 1.75V, the output doesn’t change. If it is no less
than 1.84V, the output is 0V±0.6V.

4. Test the feedback signals of fans

When no fan is connected to J71-J74, LED1-LED4 light and the U6 PIN4 (TP3: GND)
has a high level (>4V).

When PIN1 and PIN2 of any of the J71-J74 are short circuited with a jumper, the
corresponding LEDs are off and the U6 PIN 4 has a high level (>4V).

When PIN1 and PIN2 of the J71-J74 are short circuited with jumpers, all LEDs are
off and the U6 PIN 4 has a low level (<0.8V).

Tempe 2 3 4 5 6 7 8 9 10 11 12
rature
(℃)
Voltag 1.88 1.86 1.84 1.82 1.79 1.77 1.75 1.73 1.7 1.68 1.66
e (V)

6.7.2 Checking Optical Couplers


This procedure is applicable to all four-pin optical couplers used on the analyzer.

1 There are two labels E and S on the optical coupler. E stands for emitter and
S stands for sensor.

6-39
Maintenance and Service

2 A constant DC 5V or above is needed.

3 Usually the two pins on the E terminal are red and black or red and blue. The
red end is for high level.

The E terminal that is powered by 5V is in serial connection with a 220Ω


resistor.

4 Usually the two pins on the S terminal are white and green. The white end is
for high level.

The S terminal that is powered by 5V is in serial connection with a 5.1KΩ


resistor.

5 Voltage of the white end can be measured with a multimeter.

When the optical coupler is not shielded, the white end has a low level which
does not exceed 0.6V; otherwise it has a high level which exceeds 4.4V.

If the measured voltage is within the above thresholds, the optical coupler
runs normally.

6-40
Maintenance and Test Software

7 Maintenance and Test Software

7.1 Basic Operations


7.1.1 Installation
It is not necessary to install the test and maintenance software. Just copy the
software files to the PC and click the NewDebug.exe icon to start the program.
Parameters can be downloaded automatically when software is running. The
communication frame is shown on the system interface when downloading the
parameters. After downloading finished, you can use the functions of the software.

7.1.2 Overview
7.1.2.1 Screen Layout
The main screen of the test and maintenance software consists of two parts. See
Figure 7-1.

Figure 7-1 Screen layout of Test and Maintenance Software

The upper area provides various function buttons to test each unit.

The lower area displays the communication data associated with the main unit/subunits.
The lower-right area provides options and buttons to control the communication frame.

7-1
Maintenance and Test Software

7.1.2.2 Function Distribution


Click Command on the higher-left area as shown in Figure 7-2. At the Command
screen, you can select different units and then you can see the debugging buttons of the
corresponding units.

Similarly, you can choose PARA and Speed, Temperature, Photoelectric on the
higher-left area to run different debugging functions.The following is an introduction of the
Command, PARA and Speed, Temperature, Photoelectric.

The function buttons for serial port setup, version query are located on the higher-right
area of the main interface.

Figure 7-2 Screen Layout of Test and Maintenance Software

7.1.2.3 Language Switch


The software support English and Chinese display. After opening the software, click
setup on the higher-left area and input “bs200” in textbox of the pupped-up dialog box
(see Figure 7-3), and then click ok. As long as the Language button is selectable, click it
to change the language.

7-2
Maintenance and Test Software

Figure 7-3 Input Serial Port Password

7.2 Operating Commands


Click Command on the higher-left as shown in Figure 7-2, and then you can see the tabs
for each unit.

7.2.1 Main Unit


The Main Unit screen includes the following commands:

 Download Parameters: Download parameters for all of units.

 Mechanical reset: Select this command to reset the mechanical parts of subunits.

 Enable modifying parameters: Select this command to enable modifying the


parameters of all units.

 Disable modifying parameters: Select this command to disable modifying the


parameters of all units.

 Shake Hands With Machine: Select this command to communicate with the main
unit.

NOTE
Enable modify parameters and Disable modify parameters can
be used only when the parameter write protection of the main
board is disabled.The main board will be at the status of parameter
write protection in normal situation. Please refer to 7.3.1 for details
about how to cancel parameter write protection.

 Send original command: Select this command to send the instruction to the main
unit directly. Please note that this function can be performed only by debug users.

7.2.2 Mixing Unit


Click Mixing Unit on the higher-left area and then you can see the fuction buttons of the
mixing unit.

 Mixing bar to the top of wash pool——Move the mixing bar to the position above
the wash pool

 Mixing bar into wash pool——Move the mixing bar into the wash pool to wash

 Mixing bar to the top of reaction tray—— move the mixing bar to the position
above the reaction tray

7-3
Maintenance and Test Software

 Mixing bar into reaction tray——Move the mixing bar down into the reaction tray

 Mixing bar stirs for given time——Turn on the motor and the mixing bar stirs for
given time

 Mixing unit reset——Execute mechanical resetting to the mixing unit. The mixing
bar moves to the top of the reaction tray from the current position and then back to
the wash pool to wash.

Figure 7-4 Mixing Unit Debugging Interface

7.2.3 Reagent Unit


Click Reagent Unit on the higher-left area and then you can see the function buttons of
the reagent unit. For Reagent tray rotate, input the number of circles to rotate in the first
edit box, and the specified position to stop in the second edit box. The specified position
means specified bottle to aspirate.

The Reagent probe postion correct and Stop are reserved functions, but not available
in the current version.

Like other units, the actual working postion of the sample unit and reagent unit is
determined by their respective position parameters. Please refer to 8.3 for details about
parameters.

7-4
Maintenance and Test Software

Figure 7-5 Reagent Unit Debugging Interface

7.2.4 Sample Unit


Since the sample unit and the reagent unit share the same probe and sample/reagent
disk, the debugging functions for the sample unit and reagent unit are the same.

7-5
Maintenance and Test Software

Figure 7-6 Sample Unit Debugging Interface

7.2.5 Reaction Unit


Click Reaction Unit on the higher-left area and then you can see the fuction buttons of
the reaction unit.

 Reaction tray rotate——Input the number of the circles to rotate in the first edit box
and the position to stop in the second edit box, which means to stop the specified
cuvette to the 1# dispensing position of the reaction disk.

When you input “29” into the second edit box, the first cuvette segment will stop at the
replacement position; input “34”, the second cuvette segment will stop at the replacement
position; input “39”, the third cuvette segment will stop at the replacement position; input
“4”, the fourth cuvette segment will stop at the replacement position…and the rest cuvette
segment can be deduced by analogy.

 Reaction tray rotate some cups——The number of cuvettes the reaction disk
passes while rotating.

 Turn on light

 Turn off light

 Rotate and measure—— Collect photoelectric data of all the cuvettes on the
reation disk at all wavelengths.

7-6
Maintenance and Test Software

Figure 7-7 Reaction Unit Debugging Interface

7.2.6 ISE Module


Click ISE Module on the higher-left area and then you can see the fuction buttons of the
ISE module.

7-7
Maintenance and Test Software

Figure 7-8 ISE Module

The following is the introduction of the sending instruction and the returning instruction.

 Handshake: shake hands with ISE module.Sending instruction: <ISE?>, returning


instruction <ISE!>

 Purge A: Purge A cleaning. Sending instruction: <PUGA>

 Purge B: Purge B cleaning. Sending instruction: <PUGB>

 Bubble calibration: Sending instruction <BBCL>, returning instruction <BBC A xxx M


xxx L xxx>

 Calibration: sending instruction <CALB>, returning instruction <CAL Li xx.xx Na


xx.xx K xx.xx Cl xx.xx eeeeeeec> <CAL Li xx.xx Na xx.xx K xx.xx Cl xx.xx
eeeeeeec>

 Maintenance: Sending instruction: <MANT>, returning instruction <ISE!>

 Single step serum: Sending instruction: <SAMP>, returning instruction <ISE!>

 Single step urine 1: Sending instruction: <UWBW>, returning instruction <ISE!>

 Single step urine 2: Sending instruction: <UWBW>, returning instruction <ISE!>

 Single step wash: Sending instruction: <CLEN>, returning instruction <ISE!>

 Single step pump calibration: Sending instruction: <PMCL>, returning instruction


<ISE!>

 Single step start: Sending instruction: <STRT>, returning instruction /

 Dispense purge A: Sending instruction: <DSPA_200>, returning instruction <ISE!>

7-8
Maintenance and Test Software

 Debugging start 1: Sending instruction: <DVON>, returning instruction <ISE!>

 Debugging start 2: Sending instruction: <MVON>, returning instruction <ISE!>

 Debugging shutdown: Sending instruction: <DVFF>, returning instruction <ISE!>

 Program check: Sending instruction: <CKSM>, returning instruction <ISV cccc>

 Read electrode output voltage: Sending instruction: <RDMV>, returning instruction


<AMV Li xxx.x Na xxx.x K xxx.x Cl xxx.x>

 Read reagent chip: Sending instruction: <DLRD_pp>, returning instruction <DSN


s0s0s1s1……..s6s6s7s7 c>

 Write reagent chip: Sending instruction: <DLWR_pp_aa_dd>

7.2.7 Temperature Unit


Click Temp Unit on the higher-left area and then you can see the fuction buttons of the
reaction unit.

Select the Reaction checkbox, and then click Enable/Disable temperature control to
turn on the reaction disk temperature control. Disselect the Reaction checkbox, and then
click Enable/Disable temperature control to turn off the reaction temperature control.

Reagent PreHeat is to control the reagent preheating. The control method is the same
with Reaction.

Length of Sensor Line: orginal line- 3.5cm extra line- 51.5 cm. Both are fixed
parameters.

Orginal Data: Input 0, 37 and 100 to the T column, and corresponding resistance values
in the R column. Click Caculate to get R0, A, B.

When the reaction disk temperature sensor is replaced, the R0, A, B should be
recalculated by using the function provided by Orginal Data. The R0, A, B should be
configured in PARA and Speed-Temperature Unit.

When temperature collection board is replaced, ⊿AD should be reconfigured in PARA


and Speed-Temperature Unit.

The interface to configure temperature control parameters is shown in the Figure 7-10.

Reagent preheating sensor: keep the default.

Reaction disk sensor: input the calculated R0, A, B in the first three column; input ⊿AD
value in the fourth column ( found in the temperature collection board. If the board is not
replaced, leave it unmodified.)

Target temp. of reaction disk[0.01℃]: keep 37.8 as default.

Reagent preheating temp. control: keep 12/10/2 as default.

Target temp. for reagent preheating: keep 45 as default.

Reaction temp. control: keep 100/250/50 as default.

7-9
Maintenance and Test Software

Figure 7-9 Temperature Unit

7-10
Maintenance and Test Software

Figure 7-10 Temperature Unit Parameter Configuration

7.3 Parameter
Select PARA and Speed on the higher-left area to enter the parameter configuration
screen shown in Figure 7-11.You can configure two types of parameters: the parameters
of the subunits and the parameters of the motor speed.To configure the parameters of the
subunits, select the unit from the Unit Name drop-down list. To configure the motor
parameters, select the motor from Speed Name drop-down list.

7-11
Maintenance and Test Software

Figure 7-11 Parameter Configuration Interface

7.3.1 The Precondition for Parameter Configuration


Enable the parameter modification before configuring the parameters.

To save the modification, follow the instruction below:

Connect socket J15 on main board as shown in Figure 7-12. The parameter modification
function is then enabled.The default connection is shown in figure 8-12 (a). It is in
parameter write protection.

Figure 7-12 Connection Mode of the J15 Socket on Main Board

VDD GND VDD GND

(a)parameter write protection (b)parameter modification enabled

Select Enable modifying parameters on Command-Main Unit of the test and


maintenance software and then configure the parameters and save the modification.

After configuration, restore the main board to the status of parameter write protection.

7-12
Maintenance and Test Software

NOTE
If not necessary, please do not use this method to modify
parameters.

If really necessary, restart the analyzing unit and double check to


confirm the modification of the parameter after configuration,

After the parameter configuration is validated, restore the socket


J15 on main board to the status of parameter write protection.

If it is not necessary to validate the modified parameters perpetually, you can restore the
previous parameters after restarting. Please follow the instructions as below:

Click the Setup button on the higher-right area of the software shown in Figure 7-13,
input bs200 to the dialog box, and then click OK.

Figure 7-13 Enable Modifying Parameters 1

2) There is System instruction under each unit name. Please change the value of byte
8 from 00 to 03 in the system instruction, and then click Run to validate the
configuration. The following figure takes the reaction unit as an example.

7-13
Maintenance and Test Software

Figure 7-14 Enable Modifying Parameters 2

NOTE
1、 The configuration modification is only effective to selected
unit.If the number of the units is more than 1, modify
configuration of each unit respectively.

2、 If the main board is in the status of parameter write


protection, you can still use the method to modify parameter
configuration, but the parameters will be restored to the previous
one after restarting.

3、 If the main board is not in the status of parameter write


protection, the modification will be validated perpetually.

7.3.2 Detailed Operations


The operation units related to the parameters include Inquire, Configure, Save, Read
and Configure All. Enable the parameter modification before configuring the parameters,
The detailed operations are described in the section 7.3.1

Before performing the operation, select the target unit from the drop-down list of “Unit
Name” or “Speed Name”. If a unit from the Unit Name list and a motor from the Speed
Name list are both selected, the system can only identify the last selected one.

7-14
Maintenance and Test Software

7.3.2.1 Query
Select the target unit and click Query to inquire about the parameters of the
corresponding units and the motors.

7.3.2.2 Config
1. Inquire or read the parameters of the selected unit.

2. Modify the parameter

3. Enable the parameters to be selectable and then click Config.

4. The parameter configuration will be completed, only when the Config


command frame (the third byte of the communication frame is 01), Reply
frame (the third byte of the communication frame is 02) and the result frame
(the third byte of the communication frame is 03) show up.

5. After parameter configuration, click Query to confirm.

The parameter configuration will be successful only when the parameter modification of
the target unit is enabled. Otherwise, the warning of parameter modification protection
will show up at the bottom of the screen.

For the units in the Unit Name drop-down list, if the main board is in parameter write
protection status, the new parameters will be validated after configuration; but the system
will restore the previous parameters after restarting. If the modification function is enabled,
the new parameters will be perpetually validated after configuration.

For the motors in the Speed Name list, the configuration modification will only be
validated after configuration. After the analyzing unit is shutdown, the motor parameters
will be stored to the original ones.

7.3.2.3 Save
Click Save to back-up the current parameters of the target unit in a new file after the
operations of Inquire, Configure, Read and Configure All.

7.3.2.4 Load
1)Select a target unit.

2)Click Read and then select a parameter file in the pop-up dialog box. If the selected file
is not belong to the target unit, the system will automatically switch the current unit to the
unit specified by the parameter file.

3) Click Config All to refresh all the parameters of the target unit.

7.3.2.5 Config All


1) Select a target unit.

2) Inquire or read the parameters of the selected unit, or modify the parameter if needed.

3) Select Confiure All to complete the configuration with all current parameters to the
target unit.

7-15
Maintenance and Test Software

7.4 Temperature
7.4.1 Functions
The Temperature provides the functions of real-time monitoring on the temperature of
the reaction disk and the reagent preheating.

7.4.2 Operation Details


Figure 7-15 Reaction Disk and Reagent Preheat

This following content discribes the functions on the Temperature screen.

Value: current temperature

YRange: set up the Y coordinate range

XRange: set up the points on the X coordinate

Period: set up the interval of sampling. The unit is second.

Stop: stop temperature data collection

Freeze: stop refreshing the curve

Click Start to begin the collection of temperature data.

You can adjust the YRange during collection.

The startup temperature will be saved automatically under the running directory.

The operation to monitor the temperature curve of reagent preheating is the same as
reaction disk.

7-16
Maintenance and Test Software

7.5 Photoelectric
The following operations are performed on the photoelectric screen:

1) Filter offset test;

2) Photoelectric gain auto-configuration;

3) Light source background test;

4) Dark current test.

7.5.1 Filter Offset


1) Select Photoelectric-Filter Offset, and then click Start.

2) The process will last for 3-4 minutes until all the red curves are displayed. The green
font appears to indicate the test is finished.

3) After testing, click OK. A pink line will be at the center of the first peak.You can also
adjust the line manually by clicking the up-arrow and the down-arrow on the text box
under the curve.

4) When the line is adjusted to the center of the first peak, the offset of the filter equals to
the value in the text box minus 100.

5) Confiugure the filter offset in PARA and Speed-Reaction Unit.

7-17
Maintenance and Test Software

Figure 7-16 Filter Offset

7.5.2 Photoelectric Gain


NOTE
Before configuring the gain, make sure the filter offset has been
reasonably configured.

1) After start-up and the light source is stable, select Photoelectric-Others-Test Gain
and then click Start.

2) After the test is finished, you can view the gain value in the Gain column.The order is
340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm.

3) During the test, the software will configure the gain value to the reaction unit.

4) After the test, you can select PARA and Speed-Reaction Unit-Query to confirm the
configuration.

Note: the order of the gain on the reaction unit is opposite to the order mentioned above.
The order on the reaction unit is 670nm, 630nm, 578nm, 546nm, 510nm, 450nm, 405nm,
340nm.

5) Run background test to confirm the gain configuration.The recommended range for
background is 48000-62000.

6) If the gain values of a channel are not reasonable, adjust them manually. Change the
gain value of the specific channel on PARA and Speed-Reaction Unit. You will get low
background value by configuring large gain value. Similarly, you will get high background
value by configuring low gain value. The input range for gain value is 1-255.

7-18
Maintenance and Test Software

NOTE
During the test, the software will configure the gain parameters of
the reaction unit. Therefore, enable the parameter modification
function of the reaction disk before testing.

NOTE
Run background test after configuration to confirm the
configuration. The recommended range for background is
48000-62000.

CAUTION
It is only necessary to run gain test after replacing the lamp and
the background overflows (exceeding 65535). Do not adjust the
gain to elevate the background.

Otherwise the performance will be affected.

Figure 7-17 Gain Test

7.5.3 Light Source Background


Run background test when it is necessary to confirm the intensity of the light source or to
confirm the gain configuration.

7-19
Maintenance and Test Software

1) After start-up and the light source is stable, select Photoelectric-Others-Light Base
and then click Start.

2) When the green font above Start appears, remove the current cuvette segment and
then click Start.

3) When the test is finished, you can view the background on BackGround column. The
order is 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm.

Figure 7-18 Background Test

7.5.4 Dark Current Test


To test the dark current, select Photoelectric-Others-Dark current, and then click Start.
After the testing you can view the value of dark current in the Dark Current column. The
order is 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm and 670nm

7-20
Maintenance and Test Software

Figure 7-19 Dark Current Test Interface

7.5.5 Other Functions of the Photoelectric Test


In the Photoelectric tab, you can also test the basic performance, including absorbance
repeatability, precision, stability, drift, stray light and carryover. Since they are not
included in the maintenance procedure, we will not introduce them in this manual.

7.6 Macro Instructions


7.6.1 Function
Each unit tab of the test and maintenance software only includes the basic commands
and procedures for ordinary alignment, which do not support more complex and
comprehensive tests. However, the Macro Instruction tab allows you to create
necessary instructions or instruction set (macroinstruction) for the tests.

7.6.2 Detailed Operations


7.6.2.1 Screen Layout
Click the Setup button on the higher-right area shown in Figure 8-13, input “bs200” to the
dialog box and then click OK. You will enter the administrator interface shown in Figure
7-20.

On the right side of the screen is the Macro Instruction area, in which you can create
and edit macro instructions and execute them circularly.

On the left side of the screen is the instruction area, in which you can view the detailed
instructions of the specified unit.

7-21
Maintenance and Test Software

Figure 7-20 Macro Instruction Screen

7.6.2.2 Setting up Macro Instruction Name


1) Enter the administrator interface and click Setup, input serial port password “bs200”
and then click OK;

2) Select New.

3) Enter the name of the new macro instruction in the popup dialog box, and then
select OK. The name appears in the drop-down list box above Index and
Instruction.

7.6.2.3 Adding Single instruction


1) Select or create a macro instruction name.

2) Select an instruction.

Method: Select the desired unit and the instruction type in the Instruction area. All
qualified instructions of the unit are displayed in the lower table.

3) Enter the instruction values.

 Enter instruction value in the Value column according to the other information in
the table.

 The last two lines of the Value column can be left blank.

 The Value column for the reserved instructions can be left blank.

4) Verify the instructions.

7-22
Maintenance and Test Software

After setting up an instruction, click Send to verify the exactness in the current
environment. If the instruction is correct, the proper response and results frames will be
displayed and the available data will be uploaded; If the instruction is not correct, the
error message explained in red will be showed up in the information box at the bottom of
the screen.

5) Add an instruction to the macro instruction set.

Select Add to Macro. The selected instruction is added to the macro instruction list on
the right-hand side of the screen. Repeat steps 1) through 5) to add more instructions.

6) Select Save below the macro instruction list to save the macro instruction. The
newly-created macro instruction is temporarily stored in the memory. If not saved,
the macro instruction will disappear when you switch to other instructions or exit the
system.

7.6.2.4 Delete Macro Instruction


You can delete the created macro instruction.

To delete created macro instruction, you can choose the macro instructin in the name list,
and click Delete Instruction, and then confirm the deletion in the pop-up dialog box.

If you want to delete an instruction, select the instruction in the name list under the name
of the macro instruction set, and then click Delete Instruction.

7.6.2.5 Save Macro Instruction


Select Save to save a macro instruction to the database so as to be able to execute the
instruction after re-open the software. A new created instruction can not be saved after
exiting the software if not clicking save.

7.6.2.6 Run a Macro Instruction


Click Run to run the macro instruction. A macro instruction set can be built up with single
instruction or multiple instructions. If the macro instruction is built up with multiple
instructions, the macro instruction set will be executed by following the order of each
single instruction. When all the instructions are executed, the macro instruction finishes
one circle.

To run a macro instruction set once, select the checkbox next to the instruction name; the
instructions will be executed only once;

To run a macro instruction set circularly, deselect the checkbox; the instructions will be
executed circularly until clicking stop or pause.

7-23
Maintenance and Test Software

Figure 7-21 Run Macro Instruction

7-24
Appendix A Diagrams

Figure A-1 Connection of Boards

Power supply
unit

Reagent
refrigeration Main board Drive board
board

Three probes Reaction disk


AD conversion
connection temperature
board board sampling board

A-1
Diagrams

Figure A-2 Power Supply Unit

ISE Power Board


BA34-30-63624
ISE module
1 GND 1 1 24V
1
3 12V 2 2 GND
3 2
BA31-21-41491
J2 J1 E Ground

Power
Soket M07-00061S-
--
Reagent N
12&5V Board Refrigeration Board 24VBoard
PFC Board
BA33-30-35080 BA20-30-75227 L BA33-30-35082
BA33-30-35078
J86
P353 009-000902-00
4 12V 3 6
P1 P404
4 5 12V 2 5 1 L L 1
1 GND 6
4 1 1
P406 009-000876-00
5 2 GND 5
3 2 2 8
3 GNDFAN 4 BA20-20-75266 5 1
M26 :Upper heater
2 4
6 6 12FAN 1 3 3 N N 3 3
1 6 2 7
3 M27:Down heater
P3 7 3
Drive Board BA20-20-75267 M07-00094S- 5
2 8 4
051-000486-00 3 -- 1 M28 :009-000742-00
1
P350 009-000887-00 J17
2
1 12GND 8
3 1 4 8 3
2 GND 7
1
P405
009-000900-00
3 12V 4 J18
4 2 3 7 2 2
4 5V 3
Drive Board 2 2
051-000486-00 1 1
009-000897-00 1

051-000486-00
P6 1
J21

Drive Board
1 2
Main Board 2 2
051-000485-00 2 1
1 1 P403
P352 J2 009-000887-00 J17
1 009-000877-00 12GND 6 4 4 24V 1
3 1 GND 4 3 6 2 6
2 P5 M32:Lamp 3 3 24V 2
009-000875-00
3 12V 3 2 5 2
2 12VGND 2 1 GND 6
4 2 5V 1 1 5
4 1 4 2 GND 5
1 1 1
LAMP12V

A-2
Diagrams

Figure A-3 Main Control Board


Reaction disk temperature Reagent
BNC BNC Three probes refrigeration board
Drive Board sampling board

J20
connection Board BA20-30-75227 12&5V Board

C25:water-empty
051-000486-00 BA10-30-78268
BA30-30-15284

C26:trash-full
BA33-30-35080
2 4 6 8 10 12 14 16 18 20
J204 J9 J87 1 2 3

12 11

18 17

28 27

32 31

38 37

42 41
22 21

48 47
14 13

16 15

20 19

24 23

26 25

30 29

34 33

36 35

40 39

44 43

46 45

50 49
1

9
1 3 5 7 9 11 13 15 17 19
1 2 3 4

4
P352

3
1

2
16 HEAT3

18 HEAT2

20 HEAT1
1 SHIELD
2 SHIELD
4 NC
5 NC

7 NC
6 NC
2 NC
3 NC

19 +24V
1

17 +24V

1
13

14
10
12

15
11
3
4
5

6
7

8
9 NC
009-000883-00

3
009-000743-00

009-000886-00

009-000885-00

009-000877-00
009-000884-00
39 CLK_RESERVED_MOTOR3
40 DIR_ RESERVED_MOTOR3
26 CLK_RESERVED_MOTOR1

30 CLK_RESERVED_MOTOR2
31 DIR_ RESERVED_MOTOR2
27 DIR_RESERVED_MOTOR1
11 VALVE_WASH_INJECT1
12 VALVE_WASH_INJECT2

2
1
3
4
38 DIR_ SYR_MOTOR2
37 CLK_SYR_MOTOR2
36 DIR_ SYR_MOTOR1
35 CLK_SYR_MOTOR1

10 S3_TEMP
9 S2_TEMP
8 S1_TEMP
16 PWR_PV_WASH1
17 PWR_PV_WASH2
13 PREHEAT_REAG
10 PUMP_FILL_OUT

25 CLK_ WASH_UP

45 CLK_FILL_SYR
24 DIR_WASH_UP

44 DIR_ FILL _UP

5 F_REAC
9 VAVLE_FILL_IN

43 CLK_FILL_UP
32 CLK_STIR_UP

46 DIR_FILL_SYR
8 PUMP_FILL_IN

6 BUSY
3 DOUT
33 DIR_STIR_UP

2 SYNC

12 VPP
11 GND
15 LAMP_CTRL

41 CLK_FILL_R
22 CLK_STIR_R
23 DIR_ STIR_R

42 DIR_FILL_R
28 CLK_FILL_T

4 GND

7 GND
1 DIN

5 CLK
20 REAC-CLK

29 DIR_FIIL_T
4 LEVEL2

21 REAC-DIR
2 LEVEL1

14 STIR_DC

2 GND
18 RELAY_T

6 12GND
1 5V
3 GND

4 GND
1 GND

1 VCC
3 12V
34 GND

47 GND
49 VCC
1 GND

4 NC
5 NC

7 NC
6 NC
3 NC

GND
2 NC

VCC
VCC

4
50
19
NC

48
NC

NC

NC
NC

NC

J17

4
6
5
J14 10 9 8 7 6 5 4 3 2 1 1 2 3 4 1 2

12 11

18 17

28 27

32 31

38 37

42 41
22 21

48 47
2 4 6 8

14 13

16 15

20 19

24 23

26 25

30 29

34 33

36 35

40 39

44 43

46 45

50 49
1

9
J1
2 4 6 8 10 12
1 3 5 7

1
3
2
1 3 5 7 9 11
J11
J10 J2
J3

Main Board
051-000485-00

J39

J5 J6 J7
13 12 11 10 9 8 7 6 5 4 3 2 1
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 2 4 6 8 2 4 6 8 2 4 6 8 10
J12 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 J13 1 3 5 7 9 11 13 15 17 19 21 23 25 27 1 3 5 7 25 24 23 22 21 20 19 18 17 16 15 14
29 31 33 1 3 5 7 1 3 5 7 9

009-000881-00 009-000882-00
1 VCC 1 009-000879-00
+ 4 PHO_REAC_T 2 +
1 1 VCC 009-000878-00
C13:PHO_REAC_T C 2 GND 3 4 2
PHO_FILL_T 1 NC NC 1
- C
E 3 GND 4 C16:PHO_FILL_T
- 2 3 GND
3 RXD 2
PC COM
+ 1 VCC 5
E 3 4 GND 1 BA31-20-41498
4 PHO_REAC_TC 6 + 1 5 VCC 2 TXD 3 NC
C 2 GND 7 4 PHO_FILL_TC 6 2
Reserved - C 4 NC NC 4
E 3 GND 8 C15:PHO_FILL_TC
- 2 GND 7 ISE RXD232

10 DCP_CLK
1 VCC 9 3

18 AD_BUSY
GND 8 3

11 DCP_DIN
E 5 GND 5

009-000880-00
+ module

19 AD_CLK
4 PHO_FILTER_MOTOR 10 1 VCC 9

9 DCP_EN
1 VCC 1 TXD232

6 AD_DIN
+

20 AD_RC
C

22 CH_A2

23 CH_A1

24 CH_A0
21 CH_A3
4 15GND
17 15GND
C40: 4 PHO_WASH_UP 10 6 NC NC 6
- 2 GND 11 C 4

12 GND
25 GND
13 GND
1 +15V
14 +15V

5 GND
PHO_FILTER_MOTOR 2 GND 11

7 GND

8 GND
16 VCC
15 -15V
3 VCC
3 GND 12 Reserved - 2 RXD 3 8 RTS 7

2 -15V
E NC
+ 1 VCC 13 E 3 GND 12 3 TXD 2 5
4 PHO_FILL_UP 14 + 1 VCC 13 7 CTS 8
GND
C18:PHO_FILL_UP C 2 GND 15 4 PHO_WASH_SUCKSYR 14 5 GND 4
- Reserved C 2 GND 15
9 NC 6
E 3 GND 16 - 6 RTS 10 NC
+ 1 VCC 17 E 3 GND 16
7

Reagent Bar
4 PHO_FILL_R 18 + 1 VCC 17 7 CTS 6 ISP_RESE
C17:PHO_FILL_R C 4 PHO_WASH_EJECT 18
- 2 GND 19 C 8 TRIG 9 T
3 GND 20
Reserved 2 GND 19 8 14 15 16 17 18 19 20 21 22 23 24 25
E - NC P1
+ 1 VCC 21 E 3 GND 20 9 NC NC 5
4 PHO_FILL_SYR 22 + 1 VCC 21 9
C19:PHO_FILL_SYR C 2 GND 23 4 PHO_RESERVED1 22 4 NC NC 7 NC 1 2 3 4 5 6 7 8 9 10 11 12 13
- Reserved C 2 GND 23 10
E 3 GND 24 - NC 8
+ VCC 25 E 3 GND 24 NC
4 PHO_STIR_UP 26 + VCC 25 NC 11 AD conversion board
C 4 PHO_RESERVED2 26 SHIELD
C28:PHO_STIR_UP - 2 GND 27 C NC 12
BA10-30-77759
Reserved 2 GND 27
E 3 GND 28 -
1 VCC 29 3 GND 28 NC 13
+ E 1 VCC 29
C 4 PHO_STIR_R 30 + NC 14
C27:PHO_STIR_R 2 GND 31 4 PHO_RESERVED3 30
- C14:PHO_REAC_TC C 2 GND 31 NC 15
E 3 GND 32 -
33 NC E 3 GND 32
34 NC 33 NC
34 NC

A-3
Diagrams

Figure A-4 Drive Board

Three probes 12&5V Board


connection board 24V Board BA33-30-35080
BA30-30-15284 Reserved BA33-30-35082 1 P352
J207 J206 0

P403

P350
4 3 2 1
3 4
1 2 1 2
P6

3 橙

2 白

1 白
4 橙
1 2

2
1

009-000887-00
2

4 红
2 黑
1 蓝

3 黄
009-000896-00
009-000899-00
009-000898-00

1
2
009-000897-00

24V 1

24V 2

12V 4
GND 5

GND 6

GND 7
12GND

VCC
3
8
24V Board

1
4

2
2

1
1

2
BA33-30-35082
J18 009-000900-00 P405
1 2 1 2 7 5 3 1 7 5 3 1 1 2 3 4 J21 1 2 1
1 1
009-000891-00
8 6 4 2 8 6 4 2 J17 6 8 2

1
J22 J24 5 7 2 2
J7 J6
2 009-000743-00

DGND 12V
1 GND
motor

GND 24V

GND 24V

GND 5V
1 2

J8
2 NC 1 2
3
3 NC
3 4 4 NC 3 4
4

5 NC
5 6 5 6
6 NC
7 NC
009-000893-00 7 8 8 PUMP_FILL_IN 7 8
1

9 VAVLE_FILL_IN
9 4 10 PUMP_FILL_OUT 9 4
2

J13

11 VALVE_WASH_INJECT1
11 12 11 12
12 VALVE_WASH_INJECT2
motor
3

13 14 13 PREHEAT_REAG
14 STIR_DC 13 14
15 LAMP_CTRL
15 16
4

16 PWR_PV_WASH1 15 16
17 PWR_PV_WASH2
17 18 18 RELAY_T 17 18
009-000890-00 Drive Board

051-000485-00
19 GND
19 20 20 REAC-CLK 19 20 J40
1

Main Board
051-000486-00 21 22
21 REAC-DIR
22 CLK_STIR_R 21 22
2

23 DIR_ STIR_R
motor (J2-JTAG,J3、J4Reserved) J20 23 24 23 24
J7

24 DIR_WASH_UP
25 CLK_ WASH_UP
3

25 26 26 CLK_RESERVED_MOTOR1 25 26
27 DIR_RESERVED_MOTOR1
27 28 27 28
4

28 CLK_FILL_T
29 DIR_FIIL_T
29 30 30 CLK_RESERVED_MOTOR2 29 30
009-000892-00 31 DIR_ RESERVED_MOTOR2
31 32
1

32 CLK_STIR_UP 31 32

33 34 33 DIR_STIR_UP 33 34
34 GND
2

motor
J5

35 CLK_SYR_MOTOR1
35 36 35 36
36 DIR_ SYR_MOTOR1
3

37 38 37 CLK_SYR_MOTOR2 37 38
38 DIR_ SYR_MOTOR2
4

39 40 39 CLK_RESERVED_MOTOR3 39 40
40 DIR_ RESERVED_MOTOR3
41 CLK_FILL_R
41 42 42 DIR_FILL_R 41 42
1

43 CLK_FILL_UP
43 44 43 44
44 DIR_ FILL _UP
45 CLK_FILL_SYR
2

J10

45 46 45 46
46 DIR_FILL_SYR
Reserved 47 48 47 GND
47 48
3

J12 J15 J4 J14 J16 J11 48 VCC


49 VCC
49 50 49 50
4 3 2 1 4 3 2 1 4 3 2 1 4 3 2 1 4 3 2 1 4 3 2 1 50 VCC
4

Reserved
009-000889-00

009-000894-00
009-000895-00

009-000888-00

motor motor motor motor

A-4
J39
1 +15V

P1

1
14 +15V

14
2 -15V

2
15 -15V

14
3 VCC

3
16 VCC

15 16
4 15GND 3

4
17 15GND BA10-20-78126

15 16
J1
5 GND

5
18 AD_BUSY 1

17 18
6 AD_DIN

5
340nm

6
1
2
19 AD_CLK 2 Pre-amp board

17 18
7 GND
+15V

7
3 AGND BA10-30-77760
20 AD_RC

3
4

19 20
8 GND 4

8
AGND
21 CH_A3 5

19 20
AGND_SHILED

BA10-30-77759
9 DCP_EN

8
5
6

9
BA31-20-41498
6
Figure A-5 A/D Conversion Board

Main board 051-000485-00


22 CH_A2 -15V

AD Conversion board
10 DCP_CLK

9
23 CH_A1
11 DCP_DIN
24 CH_A0
12 GND
25 GND

21 22 23 24 25
13 GND

10 11 12 13
21 22 23 24 25

10 11 12 13

A-5
Diagrams
Diagrams

Figure A-6 Reagent Refrigeration Board

M42-M43:Refrigeration
fans
C35:Temp sensor BA30-21-06532
BA30-21-15175
Temp

GND
GND

GND
12V

12V

12V
Heater Protect
switch

009-000908-00
009-000907-00

009-000909-00
1 GND
1 GND

1 GND

1 GND

4 +5V
2 12V

2 12V

2 12V

3
1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 3 4
J88 J81 J79 J80 J78 J82 J83 J84 J85 J15
J89 009-000902-00
BA31-20-41456 J86 6 1
P353
1
M31:PELTIER2100-

12V
12V 1 12VFANGNDFAN
1 4 5 4 1
2
2
20-06633

GND 2 COOLER Reagent Refrigeration Board 12V GND 2 5 4 3


M30-

3 5 2
1 4
12V 12V BA20-30-75227 3 6
1 12V GND 2 5
GND 2 6 3
COOLER 1 6
2

12&5V board
BA33-30-35080
J74 J73 J72 J71 J75 J76 J77 J87
1 2 3 1 2 3 1 2 3 1 2 3 1 2 1 2 1 2 1 2 3

1 GND

1 GND

2 12V

FAN

5V
GN
2 12V

D
FAN4
12VFAN
GNDFAN

GNDFAN

GNDFAN
FAN3

FAN2
009-000905-00

12VFAN

12VFAN

GNDFAN

FAN1
12VFAN

009-000904-00

009-000885-00
009-000903-00
GND

GND

GND

GND

GND

GND
12V

12V

12V

12V

12V

12V
FAN

FAN

FAN

FAN
1 2 3
J10
M22-M25: Reaction disk fans M44: Drive board fan M39: Lamp fan Main board
BA30-21-06530 2100-20-08144 BA30-21-06532 051-000485-00

A-6
Diagrams

Figure A-7 Three Probles Connection board

C31:probe

PROBE
GND
2
1
M19: mixer motor

BA30-10-15115

BA30-10-06631
2000-10-15052

2
1
- +

J66 1 2
051-000141-00
Level detection board
J65
1 2 3 4

009-000912-00
009-000911-00

4
1

3
6 blueRP_RAM_PHO

7 black RP_LEVEL
5 green AGND

8 red 12V
2 orange
1 yellow

purple
brown
2
1

4
3

1 2 1 2 3 4 5 6 7 8

J202 J201

Three Probes Connection board


BA30-30-15284

J204
J207 J206 J205

1 2 1 2 1 2 1 2 3 4
1

4
1

009-000884-00
009-000910-00
009-000898-00

009-000899-00

4
1

2
1

J8 1 2 J10 1 2 J2 1 2 3 J11 1 2 3 4

Reaction disk
Drive Board temperature sampling Main Board
051-000486-00 board 051-000485-00
BA10-30-78268

A-7
Diagrams

Figure A-8 Fluid System

A-8
Appendix B Spare Parts List
S/N P/N Description

Shell

1. BA34-20-63652 Serial communication cable

2. 0040-10-32343 Air spring

3. BA31-20-41438 Top cover

4. BA31-20-41439 Left side cover(upper)

5. BA31-20-41440 Right side cover(upper)

6. 042-001884-00 Left side cover(lower)

7. 042-001882-00 Right side cover(lower)

8. BA31-20-41443 Front cover

9. 042-001887-00 Rear cover(upper)

10. BA20-20-75222 Little door on rear cover

11. 042-001886-00 Rear cover(lower)

12. 115-005956-00 Desktop cover(reaction section)

13. BA31-20-41656-51 Desktop cover(reagent section)

14. BA10-20-77802 Syringe compartment cover

15. BA31-30-41436 Protective cover assembly

16. 0000-10-10916 Acupuncture needle

Mechanism

17. BA31-21-56814 Step motor A(small)

18. BA31-21-56816 Step motor B(big)

19. BA20-30-75276 Reagent/sample tray assembly

Sample/Reagent compartment cover


20.
BA31-30-41523 assembly

21. BA20-30-75275 Reagent tray handle assembly

22. BA10-21-78225 Coder sensor with cable for reagent tray

Home position sensor with cable for


23.
BA10-21-78226 reagent tray

24. BA31-30-41530 Reaction tray module

25. BA30-21-15153 Coder sensor with cable for reaction tray

B-1
Spare Parts List

26. BA30-21-15154 Home sensor with cable for reaction tray

27. 115-006036-00 Sample probe drive assembly

28. 115-006037-00 Mixing bar drive assembly

29. BA10-21-78222 Vertical sensor with cable for probe

30. BA10-21-78223 Rotary sensor with cable for probe

31. BA40-30-61525 Sample probe assembly

32. BA30-21-15159 Vertical sensor with cable for mixing bar

33. BA30-21-15160 Rotary sensor with cable for mixing bar

34. BA30-10-15052 DC motor(6V)

35. BA30-20-06774 Retaining nut

36. BA31-20-41651 Mixing bar

37. BA20-20-75212 Bottle base A(white)

38. BA20-20-75212-51 Bottle base B(blue)

39. BA20-20-75214 Tube bracket1(white)

40. BA20-20-75214-51 Tube bracket2(blue)

41. BA10-21-78227 Sensor for Filter Wheel

Optical parts

42. 115-003712-00 lamp

43. BA10-10-78251 340nm filter

44. BA10-10-78252 405nm filter

45. BA10-10-78253 450nm filter

46. BA10-10-78254 510nm filter

47. BA10-10-78255 546nm filter

48. BA10-10-78256 578nm filter

49. BA10-10-78257 630nm filter

50. BA10-10-78258 670nm filter

51. BA10-30-77760 Preamplification board

52. 081-000015-00 ffiber

Liquid system parts

53. 115-006051-00 Deionized water tank head assembly

54. 115-006052-00 Waste liquid tank head assembly

55. BA30-10-06651 500ul syringe

B-2
Spare Parts List

56. 0040-10-32316 500ul plunger tip

57. 0040-10-32302 Connector p647

58. 0040-10-32303 Washer for syringe

59. 0040-10-32307 Washer for probe

60. BA30-20-06647 Reagent compartment drain connector

61. BA30-21-06470 EM valve with cable

62. BA30-21-15311 Interior-washing pump assembly

63. BA31-21-41554 Exterior-washing pump assembly

64. BA31-20-41403 3-way connector of syringe

65. BA31-30-41549 4-way connector assembly

Liquid assembly from EM valve to 3-way


66.
BA31-30-41550 connector of syringe

67. 3001-10-07066 T-connector

68. M6Y-020003--- Check valve

69. BA10-20-77752 Washing pool

70. BA31-30-41412 DI water filter assemby

Hardware

71. 051-000485-00 Main board

72. 051-000486-00 Power drive board

73. BA10-30-77759 AD converter board

74. 051-000141-00 Liquid level detection board

75. BA20-30-75227 Reagent refrigeration board

76. BA30-30-15284 Connection board of probes

77. BA33-30-35078 PFC board

78. BA33-30-35080 12V&5V board

79. BA33-30-35082 24V board

80. BA31-30-41501 Syringe sensor board

81. BA31-21-41491 Power socket and switch assembly

82. M07-00094S-- Analyzing unit switch

83. TSB1-20-20399 Fan of power module

Temperature control parts

84. BA30-10-06633 Peltier cooler

B-3
Spare Parts List

85. BA30-21-06530 Fan with cable in reaction compartment

86. BA30-21-06532 Fan with cable

87. 2100-20-08144 Fan with cable for power drive board

Temperature transducer with cable for


88.
BA30-21-15175 reagent refrigeration compartment

Temperature transducer with cable for


89.
009-000913-00 reaction compartment

Top heater with cable for reaction


90.
024-000085-00 tray(220V)

Bottom heater with cable for reaction


91.
024-000082-00 tray(220V)

Top heater with cable for reaction


92.
024-000086-00 tray(110V)

Bottom heater with cable for reaction


93.
024-000083-00 tray(110V)

94. BA31-30-41551 Reagent preheater assembly

ISE module

95. 115-006054-00 ISE module

96. BA34-30-63624 ISE power board

97. BA34-10-63661 Sample cup

98. BA34-10-63662 Bubble detector

Serial communication cable(male


99.
BA34-20-63758 connector)

100. BA34-10-63639 CL electrode

101. BA34-10-63640 Reference electrode

102. BA34-10-63641 K electrode

103. BA34-10-63642 Na electrode

104. BA34-10-63643 Spacer electrode

105. BA34-10-63644 Li electrode

Barcode scanning module

106. 0000-10-11078 Barcode scanning instrument

Antifogging preheating system for reagent


107.
BA40-30-61987 tray

108. BA40-21-61643 Heater with cable for demisting

Temperature switch with cable for


109.
BA40-21-61645 demisting

B-4
P/N: 046-001388-00(5.0)

You might also like