You are on page 1of 12

MD-30

3. Descriptions of Electrical Power


Supply
Fuse-Equipped
Component
Fuse
No.
Blown-Fuse
Detection Circuit
Target for Protection

Operations F14 B141 Solenoid valve : SVA1/SVA2/SVA3/SVA4


DRV17A board
F15 B151 Clutch: CLA1/CLA2/CLA3/CLA4
3.1 Power Supply Voltage Output Detection F1 B161 Pulse motor: MC1
Function F2 B171 Pulse motor: MC2
To facilitate early recovery from a failure, the RU incorporates the power supply F3 B181 Pulse motor: MC3
voltage output detection function, which offers the following features:
F4 B191 Pulse motor: MC4
- Each load circuit incorporates a fuse that serves to isolate a fault.
F5 B201 DC motor: MZ4
- The isolation points are provided with a current detection circuit, which constantly +24V (B)
monitors the power supply voltage output. F6 B211 Pulse motor: ME2
DRV17B board
- The detection point can be identified by checking the combination of current F7 B221 Pulse motor: MB1
detection circuits (points) in which an error has occurred. For the lamps, however, F8 B241 Pulse motor: MB3
their point is detected as either the INV17A and/or INV17B board.
F9 B231 Pulse motor: MB2
 Fuses and Detection Points(Machines with Serial Numbers F10 B251 Solenoid: SOLB1
20001 or Those Later than 20001) F11 B261 Solenoid: SOLE1

Power Fuse-Equipped Fuse Blown-Fuse F12 B281 Fan: FAN (1-6)


Target for Protection
Supply Component No. Detection Circuit
F18 D011 DC motor: MZ2
PSU17A power supply A011 Panel: LCD17A board +15V (D) DRV17A board
F19 D021 DC motor: MZ3
F1 A021 Sensor
F1 - Lamp: LAMP1
F2 A031 LED
F2 - Lamp: LAMP2
+5V (A) SNS17B board INV17A board
F3 A041 Barcode reader F3 - Lamp: LAMP3
+24V (E)
F4 A051 INV17A/B board F4 - Fan: FANE1/FANE2/FANE3
SCN17A/SCT17B board F5 A061 IP leading-edge sensor: SED17A F1 - Lamp: LAMP4
INV17B board
F1 B011 Pulse motor: MA1 F2 - Lamp: LAMP5

F2 B021 Pulse motor: MA2 F1 G041 Laser: LDD12A board

F3 B031 Pulse motor: MA3 Start-point sensor: SYN12A board


F3 G011
+15V (G) Photomultiplier (front): PMT17A board
F4 B041 Pulse motor: MA4
F8 G031 +5VAS protection
F5 B051 Solenoid: SOLA1
F7 G051 Photomultiplier (back): PMR17B board
F6 B061 Solenoid: SOLA2
SCN17A/SCT17B Photomultiplier (front): PMT17A board
+24V (B) DRV17A board F7 B071 Solenoid: SOLA3 board Photomultiplier (back): PMR17B board
-15V (H) F2 H011
Laser: LDD12A board
F8 B081 Solenoid: SOLA4 Start-point sensor: SYN12A board
F9 B091 DC motor: MD1 +24V (J) F6 J011 Polygon: POL17B board
F10 B101 DC motor: MD2 F4 G021 Start-point sensor: SYN12A board
F11 B111 Pulse motor: ME1 +5V (A) Start-point sensor: SYN12A board
F5 G061
IP leading-edge sensor: SED17A
F12 B121 Pulse motor: ME3

F13 B131 Pump: PA1/PA2/PA3/PA4

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-30
MD-31
 Fuses and Detection Points(Machines with Serial Numbers Power Fuse-Equipped Fuse Blown-Fuse
Target for Protection
50001 or Those Later than 50001) Supply Component No. Detection Circuit
B161 Pulse motor: MC1
Power Fuse-Equipped Fuse Blown-Fuse Target for Protection
Supply Component No. Detection Circuit B171 Pulse motor: MC2
(1.6A) - B181 Pulse motor: MC3

(1.6A) Sensors B191 Pulse motor: MC4


+5V (A-1) SNS17C board
(1.6A) LED board (x5) : LED17A (x5) B201 DC motor: MZ4

(1.6A) BCR(220mA/max x5) B211 Pulse motor: ME2


+24V (B-2) DRV17B board
B221 Pulse motor: MB1
B011 Pulse motor: MA1
B231 Pulse motor: MB2
B021 Pulse motor: MA2
B241 Pulse motor: MB3
B031 Pulse motor: MA3
B251 Solenoid: SOLB1
B041 Pulse motor: MA4
B261 Solenoid: SOLE1
B051 Solenoid: SOLA1
B281 Fan: FAN (1-6)
B061 Solenoid: SOLA2
SCN17A/SCT17B
+24V (B-3) (1.6A) Polygon driver board : POL17A
B071 Solenoid: SOLA3 board

B081 Solenoid: SOLA4 D011 DC motor: MZ2


+24V (B-2) DRV17A board +15V (D) DRV17A board
B091 DC motor: MD1 D021 DC motor: MZ3
(1.6A) Photomultiplier board : PMR17B (SCT17B only)
B101 DC motor: MD2

B111 Pulse motor: ME1 (1.6A) LD Driver board : LDD12A

B121 Pulse motor: ME3 (0.5A) +5VAS protection


+15V (G)
B131 Pump: PA1/PA2/PA3/PA4 Photomultiplier board : PMT17A
(1.6A)
Start-piont detection board : SYN12A
SCN17A/SCT17B
Solenoid valve : SVA1/SVA2/SVA3/
B141 board (1.6A) Start-piont detection board : SYN12A
SVA4
(1.6A) Start-piont detection board : SYN12A
B151 Clutch: CLA1/CLA2/CLA3/CLA4
Photomultiplier board : PMR17B (SCT17B only)
LD Driver board : LDD12A
-15V (H) (1.6A)
Photomultiplier board : PMT17A
Start-piont detection board : SYN12A

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-31
MD-32
 Fuses and Detection Points(Machines with Serial Numbers Power Fuse-Equipped Fuse Blown-Fuse
Target for Protection
Supply Component No. Detection Circuit
60001 or Those Later than 60001)
B161 Pulse motor: MC1
Power Fuse-Equipped Fuse Blown-Fuse Target for Protection B171 Pulse motor: MC2
Supply Component No. Detection Circuit

(1.6A) - B181 Pulse motor: MC3

(1.6A) Sensors B191 Pulse motor: MC4


+5V (A-1) SNS17E board
B201 DC motor: MZ4
(1.6A) LED board (x5) : LED17A (x5)
B211 Pulse motor: ME2
(1.6A) BCR(220mA/max x5) +24V (B-2) DRV17B board
B221 Pulse motor: MB1
B011 Pulse motor: MA1
B231 Pulse motor: MB2
B021 Pulse motor: MA2
B241 Pulse motor: MB3
B031 Pulse motor: MA3
B251 Solenoid: SOLB1
B041 Pulse motor: MA4
B261 Solenoid: SOLE1
B051 Solenoid: SOLA1
B281 Fan: FAN (1-6)
B061 Solenoid: SOLA2 SCT17P/SCN17P
+24V (B-3) (1.6A) Polygon driver board : POL17C
B071 Solenoid: SOLA3 board
D011 DC motor: MZ2
B081 Solenoid: SOLA4 +15V (D) DRV17A board
+24V (B-2) DRV17A board
D021 DC motor: MZ3
B091 DC motor: MD1
(1.6A) Photomultiplier board : PMR27B
B101 DC motor: MD2
(1.6A) LD Driver board : LDD12A
B111 Pulse motor: ME1
(0.5A) +5VAS protection
B121 Pulse motor: ME3
+15V (G)
Photomultiplier board : PMT27B
B131 Pump: PA1/PA2/PA3/PA4 (1.6A)
Start-piont detection board : SYN12A
SCT17P/SCN17P
Solenoid valve : SVA1/SVA2/SVA3/ board
B141 (1.6A) Start-piont detection board : SYN12A
SVA4
(1.6A) Start-piont detection board : SYN12A
B151 Clutch: CLA1/CLA2/CLA3/CLA4
Photomultiplier board : PMR27B
LD Driver board : LDD12A
-15V (H) (1.6A)
Photomultiplier board : PMT27B
Start-piont detection board : SYN12A

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-32
MD-33
 Overcurrent Protection Circuit Block Diagram(Machines with
Serial Numbers 20001 or Those Later than 20001)

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-33
MD-34

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-34
MD-35
 Overcurrent Protection Circuit Block Diagram(Machines with
Serial Numbers 50001 or Those Later than 50001)

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-35
MD-36

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-36
MD-37
 Overcurrent Protection Circuit Block Diagram(Machines with
Serial Numbers 60001 or Those Later than 60001)
Power supply board
PSU29A

+5V Driver board


5A DRV17B
(A-1)
+5V +5V B161 Pulse motor: MC1
SLOT1 5A
(A-2)
B171 Pulse motor: MC2
+5V
3A B181 Pulse motor: MC3
(A-3)
B191 Pulse motor: MC4
+24V Driver board
5A +24V(E) B201 DC motor: MZ4
DRV17A
B211 Pulse motor: ME2
+24V B011 Pulse motor: MA1
5A B221 Pulse motor: MB1
SLOT2 (B-1) B021 Pulse motor: MA2
+24V B231 Pulse motor: MB2
+24V B031 Pulse motor: MA3
5A
(B-2) B241 Pulse motor: MB3
B041 Pulse motor: MA4
+24V B251 Solenoid: SOLB1
1A (B-3) B051 Solenoid: SOLA1
B261 Solenoid: SOLE1
B061 Solenoid: SOLA2
2A +15V(G) B281 Fan: FAN (1-6)
B071 Solenoid: SOLA3
SLOT3
B081 Solenoid: SOLA4 TP1
+15V
5A +15V(D)
B091 DC motor: MD1
B101 DC motor: MD2
3A +15V(F)
B111 Pulse motor: ME1
-15V B121 Pulse motor: ME3
1.5A -15V(H)
B131 Pump: PA1/PA2/PA3/PA4
AUX12V +12V
+5V(S) B141 Solenoid valve: SVA1/SVA2/SVA3/SVA4
B151 Clutch: CLA1/CLA2/CLA3/CLA4

D011 DC motor: MZ2


D021 DC motor: MZ3 : Blown-Fuse Detection Circuit

TP1 FR9H1304_E.ai

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-37
MD-38

Power supply board


PSU29A

5A +5V Sensors board


(A-1) SNS17D
+5V +5V TP1 : Blown-Fuse
SLOT1 5A Erasure board Sensors board Detection Circuit
(A-2)
ERS23A SNS17E : DC-DC converter
+5V Inverter board : INV26A (15V → +5VAS)
3A 1.6A
(A-3)
Inverter board : INV26A
1.6A Sensors
+24V Inverter board : INV26A
5A +24V(E) 1.6A LED board (x5) : LED17A (x5)
Inverter board : INV26A
1.6A BCR(220mA/max x5)
5A +24V
SLOT2 (B-1)
+24V +24V
5A PNL board : PNL29C Scanner board
(B-2) SCT17P/SCN17P
+24V 1.6A Polygon driver board : POL17C
1A (B-3)

1.6A Photomultiplier board : PMR27B


2A +15V(G)
SLOT3
1.6A LD Driver board : LDD12A
+15V 5PM:CRD5107P
5A +15V(D) +5VAS
0.5A Photomultiplier board : PMT27B

3A +15V(F) 1.6A

1.6A Start-piont
-15V detection board
1.5A -15V(H)
Image 1.6A SYN12A
processing board
AUX12V +12V IMG17A/B
+5V(S) 1.6A
TP1
CPU board
CPU91A
1.6A IP leading-edge detection board : SED23A
FR9H1305.ai

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-38
MD-39
3.2 Erasure Unit Control 3.2.2 Components of the Erasure Unit Control (#20001 or later)
- The control circuit for the erasure lamp unit is located on the SNS17B board. A
3.2.1 Overview of the Erasure Unit Control control signal from the SNS17B board causes the relay circuit and lamp inverter
circuit of the driver board (INV17A/B) to operate, thereby turning ON or OFF the
(#20001 or later) lamps and fans.
The erasure unit provides idle illumination control, IP erasure control, and fan control. - To enable discrimination between failures of the control board (SNS17A) and driver
board (INV17A/B), LEDs are provided on the SNS17B board to indicate the status
Feature: Erasure unit initialization is effected after verifying that the machine is of the control signal.
emptied of IPs.
{MD:3.3.3_Correspondence between LEDs and Erasure Control Signals and
 Idle illumination control FFM Control Signals}
To obtain an adequate lamp light intensity for IP erasure after RU power ON, lamp
(LAMP1 - LAMP5) illumination and erasure unit cooling fan (FANE1) operation are Power supply unit
controlled so as to maintain a constant temperature inside the erasure unit. PSU17A
+5V(A) +24V(E)
CN1 CN2
 IP erasure control
To effect IP erasure, the erasure unit pass time is controlled in accordance with the Sensor board CN20
dose received by a processed IP. SNS17B FANE3 FANE2 LAMP2
Safety CN5
TSWE1 FANE1 LAMP1 LAMP3
 Fan control thermostat
control circuit CN9 CN7 CN6 CN3 CN4 CN5
Control is exercised so as to operate the erasure unit cooling fan (FANE1) when the
temperature inside the erasure unit is high. LED F4 F1 F2 F3
 Temperature adjustment control Lamp
CN2 Relay CN1
- The heat generated during lamp (LAMP1 - LAMP5) illumination is detected by the inverter
Lamp circuit
temperature thermistor (THE1) and used to exercise control of the erasure unit CN16 circuit
cooling fan (FANE1) operation and lamp illumination. temperature
control circuit Inverter board
- While the machine is idle and processing no IP, control is exercised to automatically INV17A CN8
illuminate the lamps for temperature adjustment purposes. Unlit lamp
detection
- Before the lamps are illuminated, they are subjected to preheating. circuit CN6
Inverter board
Temperature INV17B
control fan Lamp
(FANE1) CN2 Relay CN1
circuit inverter
control circuit circuit F1 F2

Sensor board CN3 CN4


SNS17A
CN1 THE1 LAMP4 LAMP5
CN3
Image
processing board FANE1: Erasure unit cooling fan
IMG17A/B FANE2: Inverter board cooling fan
FANE3: Inverter board cooling fan
CPU board TSWE1: Thermal switch
CPU91A THE1: Temperature thermistor
Main CPU
FR9H1085.EPS

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-39
MD-40
3.2.3 Overview of the Erasure Unit Control  Control regarding service life of the lamp
(#50001 or later) - When the illumination count of the erasure lamp approaches the assumed machine
working hours, a warning “It’s about time to replace” is issued.
 Temperature adjustment control - When the illumination count of the erasure lamp reaches the assumed machine
Temperature adjustment control is exercised to keep the inside of the erasure unit at a working hours, a warning “The fluorescent lamp reaches the end of life. Contact the
service for replacement” is issued.
constant temperature by controlling the operation of the erasure cooling fan (FANB1)
in orderto obtain adequate light intensity required for IP erasure after the RU power is - When the illumination count of the erasure lamp exceeds the assumed machine
turned ON. working hours, a warning “The erasure cannot be guaranteed” is issued.

 IP erasure control
Erasure unit pass time is controlled in accordance with the dose received by a read IP,
and the IP is erased.

 Control in case of error


When an error occurs due to a burnt-out lamp during bootup or routine processing, the
machine transfers to “erasure time extension mode” or “erasure degeneration” mode
after the detection of the error. The relation of the conditions of the erasure unit lamps,
the mode to transfer, and the RU operation are as follows:

Erasure unit Mode to RU operation


condition transfer
One or two lamps Erasure time Conveyance speed is delayed (erasure time is
have burnt out. extension extended) since the light intensity needed for the
IP erasure is insufficient with the conveyance
speed for routine processing, so as to erase the
image exposed on the IP.
Three or more lamps Erasure The image is read, but the IP is erased for the
have burned out. degeneration same period of time as the erasure time extension
mode.
The message mentioning that the IP erasure
cannot be guaranteed is displayed on the RU
operation panel.

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-40
MD-41
3.2.4 Components of the Erasure Unit Control
(#50001 or later)
The control circuit for the erasure lamp unit is located on the ERS23A board. A control
signal from the ERS23A board causes lamp inverter circuit to operate, thereby turning
ON or OFF the lamps.

Erasure board SNS17C


Power supply ERS23A board
board Temperature
Temperature Safety Unlit lamp Lit lamp LED
PSU23A control fan
management thermostat detection control
sensor circuit (FANB1) monitor circuit
control circuit circuit
control circuit
CNB61
Relay Main CPU FPGA
+24V
circuit
+5V RS232C RS232C
CN11 CN13 CN17 CN19
CN12 CN14 CN18 CN20
CN31 CN33 CN32
Inverter board Inverter board
INV26A INV26A
Lamp Lamp
inverter inverter
THB1 circuit circuit

LAMPB4, 6 LAMPB3, 5

Inverter board Inverter board


THB1: Lamp temperature sensor INV26A INV26A
Lamp Lamp
inverter inverter
circuit circuit

LAMPB2, 8 LAMPB1, 7 FRCH103008.ai

011-201-11E
06.29.2012 FM6046
CR-IR 362 Service Manual MD-41