You are on page 1of 6

Specification: V3

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2020 V 1.0
Ventilator
V3

Technical Specification TV (Tidal Volume): Adult: 100 - 2200 mL (increments


Physical Characteristics of 10 mL)
Dimensions 1365±10 mm × 526±5 mm × 544±5 Pediatric: 15 - 300 mL (increments
mm (Height × Width × Depth) of 1 mL)
(including trolley); Respiratory Rate (RR): 1 - 100 bpm (increments of 1 bpm)
354±5 mm × 315±5 mm × 249±5 fSIMV (Ventilation 1 - 60 bpm (increments of 1 bpm).
mm (Height × Width × Depth) frequency in SIMV
(excluding trolley) mode):
Weight 30±0.5 kg (including trolley) I:E range: 4:1~1:10.
10±0.5 kg (excluding trolley) Tinsp (Inspiratory time): 0.10 - 10 s (increments of 0.05 s).
Screen Tslope (Time of Pressure 0 - 2.00 s (increments of 0.05 s).
Screen Size: 12.1” TFT touch screen Rising):
Resolution 1280 × 800 High Pressure Time 0.2 - 30 s (increments of 0.1 s)
Brightness: Adjustable (Thigh)]:
Ventilator Specification Tlow (Low Pressure 0.2 - 30 s (increments of 0.1 s)
Ventilation mode V-A/C (Volume assist/control) Time):
P-A/C (Pressure assist/control) Max inspiratory Time 0.20 ~ 15.00 s
V-SIMV (Volume - Synchronized (Timax):
Intermittent Mandatory Tpause: 5 % - 60 % (increments of 5 %), Off
Ventilation) ΔPinsp(Inspiratory 5 - 85 cmH₂O (increments of 1
P-SIMV (Pressure - Synchronized pressure): cmH₂O)
Intermittent Mandatory ΔPsupp: 0 - 85 cmH₂O (increments of 1
Ventilation) cmH₂O)
CPAP/PSV, Phigh (High Pressure 0 - 85 cmH₂O (increments of 1
DuoVent, Level): cmH₂O)
APRV, Plow (Low Pressure 0 - 80 cmH₂O (increments of 1
PRVC, PRVC-SIMV, Level): cmH₂O)
VS PEEP: 1 - 50 cmH₂O (increments of 1
PSV-S/T cmH₂O), Off
Controlled parameter ranges Flow trigger 0.5 -15 L/min (increments of 0.1
O₂%: 21 - 100% (increments of 1 %) L/min), Off
Pressure trigger -10 to - 0.5 cmH₂O (increments of Manual Breath
0.5 cmH₂O), Off Expiration Hold
Exp % (Expiration 10 - 85% (increments of 5%), Auto Inspiration Hold
termination/trigger Nebulizer
level) O2↑(O2 enrichment)
Apnea Ventilation Sputum Suction
TVapnea Adult: 100 - 2200 mL (increments Paw - Volume, Flow - Volume, Paw
Loops
of 10 mL) Pediatric: 20 - 300 mL - Flow
(increments of 1 mL) Smart Sync IntelliSynTec
ΔPapnea 5 - 85 cmH₂O (increments of 1 O2 Therapy
cmH₂O) Control accuracy
RRapnea (Apnea 1 - 80 bpm (increments of 1 bpm) O₂ % ± (3 vol.% +1 % of setting)
Respiratory Rate) TV ± (10 mL + 10% of the set value)
Apnea Tinsp 0.20 - 10 s (increments of 0.05 s) Tinsp ± 0.1 s or ± 10 % of setting,
Automatic Tube Resistance Compensation whichever is greater
Tube Type endotracheal intubation and I: E 1:4~2:1: ± 10% of the set value;
tracheotomy tube Other range: ± 15% of the set
Tube I.D. Adult: 5.0 - 12.0 mm (increments of value.
0.5 mm) Pediatric: 2.5 - 8.0 mm RR ±1 bpm
(increments of 0.5 mm) fSIMV ±1 bpm
Compensate 0 -100 % (increments of 1 %) Tslope (Rising Time) ± (0.2s + 20% of the set value)
Expiration Compensation ON, Off Phigh ± (2 cmH2O + 5% of the set value)
Switch Plow ± (2 cmH2O + 5% of the set value)
Monitoring Thigh ±0.2s or ± 10% of the set value,
Airway pressure range Ppeak, Pplat, Pmean (Range -20 - whichever is larger
120 cmH₂O) Tlow ±0.2s or ± 10% of the set value,
PEEP 0~120 cmH2O whichever is larger
Tidal volume range: 0~4000 mL Pressure Trigger ± (1 cmH2O + ± 10% of the set
Respiratory Rate ftotal, fmand, fspn (Range 0 - 200 value)
bpm) Flow Trigger ± (1 L/min + 10% of the set value)
Minute volume range MV, MVspn, MVleak (Range 0 - 100 Δint.PEEP ± (2 cmH2O + 5% of the set value)
L/min) Exp % ±10% (absolute error)
Resistance Rinsp, Rexp (0 - 600 cmH₂O/L/s) Fapnea (Apnea ±1bpm
Compliance Cstat, Cdyn (0 - 300 mL/cmH₂O) Frequency)
Inspired Oxygen (FiO₂) 15 - 100 % ΔPapnea ± (2 cmH2O + 5% of the set value)
WOB (Work of 0 - 100 J/min TVapnea ± (10 mL + 10% of the set value)
Breathing) Apnea Tinsp ±0.1s or ± 10% of the set value,
RCexp (Expiratory Time 0 - 10 s whichever is larger
Constant) Monitoring Accuracy
Waveforms Airway pressure - time, Flow - time, Airway pressure (Ppeak, Within the range of 0 cmH2O~120
Volume - time Pplat, Pmean, PEEP) cmH2O, ± (2 cmH2O + 4% of the
Weaning indicator actual reading)
RSBI (Rapid Shallow
0 - 999 1/(L•min)
Breathing Index)
Special Function
Tidal Volume (TVi, TVe, Within the range of 0 mL~100 mL, Lower alarm limit Adult: OFF,
TVe/IBW, TVe spn) ± (10 mL + 3% of the actual 50~3995 mL
reading); Pediatric: OFF, 10~595 mL
Within the range of 100 mL~4000 Minute Volume Upper alarm limit: Adult: 0.2~100.0
mL, ± (3 mL + 10% of the actual L/min
reading) Pediatric: 0.2~60.0 L/min
Minute Volume (MV, Within the range of 0.0 L/min~ Lower alarm limit: Adult: 0.1~50.0
MVspn, MVleak) 100.0 L/min, ± (0.2 L/min + 10% of L/min
the actual reading) Pediatric: 0.1~30.0 L/min
Frequency (ftotal, Within the range of 0 bpm~200 Airway pressure Upper alarm limit: 10~90 cmH2O.
fmand, fspn) bpm, ±1 bpm or ±5% of the actual Lower alarm limit: OFF,5~ (upper
reading, whichever is larger alarm limit -5) cmH2O
Inspired Oxygen (FiO₂) Within the range of 15 vol.%~100 Frequency (Respiratory Upper alarm limit: 2~160 bpm, OFF
vol.%, ± (2.5 vol. % + 2.5% of the Rate) Lower alarm limit: OFF,1~159 bpm
actual reading). Inspired oxygen (FiO₂) Upper alarm limit: 20 vol.%~100
Resistance Within the range of 0 vol.%
cmH2O/(L/s) ~20 cmH2O/(L/s), Lower alarm limit: 18 vol.%~ 98
±10 cmH2O/(L/s); vol.%
Within the range of 20 Apnea alarm time 5~60s, in nasal positive airway
cmH2O/(L/s) ~600 cmH2O/(L/s), pressure ventilation mode, it can
±50% of the actual reading). be set to OFF
Compliance Within the range of 0 mL/ Trend
cmH2O~300 mL/ cmH2O, ± (2 mL/ Type Tabular, Graphic
cmH2O + 20% of the actual Length 72 hours
reading). Content Monitor Parameters, Setting
RSBI Within the range of 0 /(min·L)~ Parameters (Setting Ventilation
999 /(min·L) , ± (3 /(min·L)+15% mode and Parameters) includes
of the actual reading). parameter alarm events and
WOB Within the range of 0.0 J/min~ parameter waveforms related to
100.0 J/min, ± (1 J/min + 15% of the alarm time
the actual reading). Data Review
NIF Within the range of -45.0 cmH2O~ Event logs Up to 8000 event logs can be
0.0 cmH2O, ± (2 cmH2O + 4% of saved, including alarm logs and
the actual reading) operation logs. The alarm log
P0.1 Within the range of -20.0 cmH2O~ includes parameter alarm events
0.0 cmH2O, ± (2 cmH2O + 4% of and parameter waveforms related
the actual reading). to the alarm time.
RCexp Within the range of 0.0s~10.0s, Freeze the waveform Freeze the waveform of the
± (0.2s + 20 % of the actual review interface at the current time and
reading). use the knob to review the data.
Alarm Settings When freezing, 30 most recent
Tidal Volume Upper alarm limit Adult: historical waveforms can be
110~4000 mL, OFF reviewed by sliding the screen or
Pediatric: 25~600 mL, OFF rotating the knob.
Freeze the loop review Up to 5 reference loops can be Regulatory compliance YY1040.1-2003
saved.
EN ISO5356-1:2004
O2 Therapy
System compliance and resistance
O₂ % 21 - 100 % (increments of 1 %) ± (3
vol.% +1 % of setting) Compliance Adult disposable circuit (including
Flow 2 - 60 L/min ± (2 L/min +10 % of inspiratory safety valve, adult
setting) (BTPS)
disposable breathing tube, water
Gas Circuit Specification
collection cup, expiratory valve): ≤
Gas type Air, O2
Gas source requirement Medical compressed oxygen and 4 mL/cmH2O;
medical compressed air Adult reusable circuit (including
High-pressure O2 source inspiratory safety valve, adult
Gas source pressure 280~600 kPa reusable breathing tube, water
range
collection cup, expiratory valve, Y-
Rated flow rate 120 L/min
joint): ≤ 2 mL/cmH2O;
requirement
Input connector NIST (ISO 5356-1) or DISS (CGA Pediatric disposable circuit
1240) (including inspiratory safety valve,
Standards compliant YY/T 0799-2010 pediatric disposable breathing
EN ISO5359:2008
tube, water collection cup,
Low-pressure O2 source
expiratory valve): ≤ 2 mL/cmH2O;
Input pressure range < 100 kPa
Pediatric reusable circuit (including
Maximum flow rate 15 L/min
Input connector CPC quick coupling inspiratory safety valve, pediatric

Inspiratory module reusable breathing tube, water

Peak flow rate ≥ 210 L/min collection cup, expiratory valve, Y-


Nebulizer connector Flow rate: 6 L/min~9 L/min joint): ≤ 2 mL/cmH2O;
Safety pressure of ≤ 12.5 kPa
Neonate reusable circuit (including
respiration
inspiratory safety valve, neonate
Inspiratory-side external Coaxial 22 mm/15 mm conical
connector connector reusable breathing tube, water
Removable, sterilizable can be entirety removed quickly; collection cup, expiratory valve, Y-
and can be entirety cleaned and joint): ≤ 1 mL/cmH2O.
disinfected.
Inspiratory resistance ≤ 6 cmH2O at the flow rate of 60
Regulatory compliance YY1040.1-2003
L/min (Adult);
EN ISO5356-1:2004
≤ 6 cmH2O at the flow rate of 30
Expiratory module
Expiratory-side external L/min (Pediatric);
Coaxial 22 mm/15 mm conical
connector ≤ 6 cmH2O at the flow rate of 5
connector
L/min (Neonate).
Removable, sterilizable can be entirety removed quickly;
Expiratory resistance ≤ 6 cmH2O at the flow rate of 60
and can be entirety cleaned and
L/min (Adult);
disinfected.
≤ 6 cmH2O at the flow rate of 30 Input frequency 50/60 Hz
L/min (Pediatric); Input current 2.7 - 1.1 A
≤ 6 cmH2O at the flow rate of 5 Fuse T3.15 AH/250 V
L/min (Neonate). External DC power supply
Basic performance Input voltage 12V
Pressure monitoring -20~120 cmH2O Input current 10A
range
Internal battery
Safety pressure of In ventilation state: ≤ 125cmH2O
Number of batteries One or Two
system
In non-ventilation state or
Battery type Build-in Lithium-ion battery, 14.4
power failure or
VDC, 6A, 6700mAh
gas source failure (<0.12MPa): ≤ 14
Battery life 140 min (when a new fully charged
cmH2O
battery is used in standard
Environmental specifications operating mode)
Temperature 5 - 40 °C (operating); -20 to 60 °C 280 min (when two new fully
(storage and transport, O₂ sensor: - charged batteries are used in
20 to 50°C)
standard operating mode)
Relative Humidity 5 - 95 % (operating); 5 - 95 %
(storage and transport) I/O
Barometric Pressure 62 - 106 kPa (operating); 50 -106 Communication Rs232, Ethernet, VGA, USB port,
kPa (storage and transport) interface
Nurse call
Power Specification
Trolley MC100
External AC power supply Dimensions
Input voltage 100 - 240 V Weight 20 kg

*Notice: Specifications subject to changes without prior


notice. All rights reserved by Comen