Professional Documents
Culture Documents
VERIFICATION PROCEDURE
CONTENTS Page
1. SCOPE .................................................. 2
2. FIELD OF APPLICATION ................................... 2
3. REFERENCES ............................................. 2
4. DEFINITIONS ............................................ 3
5. TEST CONDITIONS ........................................ 7
6. SOUND CALIBRATOR ....................................... 8
7. MICROPHONE ............................................. 11
8. ELECTRICAL TEST OF THE SOUND LEVEL METER ............... 13
9. REPORT ................................................. 30
Appendix A: Tests of the overload detection characteristics. 31
Appendix B: Theoretical response to tone bursts............ 33
0. FOREWORD
The test method assumes that the design and construction of the sound
level meters, etc, are such that an apparatus without any defects will
fulfil the relevant IEC-standards.
1. SCOPE
2. FIELD OF APPLICATION
3. REFERENCES
British Calibration Service (NPL) No. 302: General criteria for labo-
ratory approval, Electrical measurements.
- 3 -
4. DEFINITIONS
The crest factor of a signal is the ratio of the peak value to the
rms-value measured over a specified time interval, the instantaneous
values of the signal being measured with respect to the arithmetic
mean value. The relation between crest factor and pulse duty factor
for sequences of rectangular pulses and for tone bursts is given in
Appendix A.
The linearity range is defined for the integrating sound level meter
as the difference, in decibels (dB), between the upper and lower rms-
levels of continuous sinusoidal signals applied to the input within
which the linearity requirements given in Sub-clause 8.3.1 are met.
The indicator range is defined for the integrating sound level meter
- 5 -
where:
LAeq,y is the equivalent continuous A-weighted sound pressure level
re 20 µPa, determined over a time interval T = t2 - 61
PA(t) is the instantaneous A-weighted sound pressure of the sound
signal
PO is the reference sound pressure of 20 µPa
Note - When, optionally, a frequency weighting other than A is used,
the frequency weighting used shall be included explicitly in
the title and the formula of the quantity, for example equi-
valent continuous C-weighted sound pressure level:
where:
is the A-weighted sound exposure level, re 4x10-10Pa2*s, deter-
LEA,T
mined over a time interval T = t2 - tl
PO is equal to 20 µPa
To is equal to 1 s
t2 2
Note - The expression PA(t)dt is the A-weighted sound exposure,
s t1
EAT*
2
The quantity (po*To) is the reference sound exposure and is
10
equal to 4x10- Pa2*s.
The burst duty factor of the test signal of Sub-clause 8.3.2 is the
ratio of the duration of the tone burst to the duration of a complete
cycle at the repetition frequency.
5. TEST CONDITIONS
5.1 Temperature
5.2 Humidity
It is essential that the laboratory area be kept above the dew point
at all times; the RH should be kept between 35% and 70%.
6. SOUND CALIBRATOR
6.1.1 Tolerance
6.1.2 Stability
Calibrator Class 0 1 2
Tolerance (dB) ±0.15 ±0.3 ±0.5
Stability (dB) ±0.05 ±0.l ±0.2
6.2 Frequency
6.2.1 Tolerance
The frequency of the sound generated by the sound calibrator under the
reference ambient conditions, averaged over 20 s of operation and after
any stabilizing time specified by the manufacturer, shall not deviate
from the corresponding nominal value by more than the tolerances
specified in Table 2.
6.2.2 Stability
Calibrator Class 0 1 2
Tolerance (%) ±l ±2 ±4
6.4.1 Timing
The accuracy of this check shall be such that it relates to less than
half the tolerances in table 1 for the correction procedure for ambient
pressure or temperature.
7. MICROPHONE
Note 3 - The normal free field correction curves imply that the
microphone is mounted on an infinitely long cylindrical
body of the same diameter as the microphone. If this con-
dition is not satisfied with sufficient accuracy for the
actual use of the microphone, other appropriate correction
curves should be used.
- 12 -
7.2 Sensitivity
Note - Normally the sound level meter is calibrated by its user be-
fore and after each measurement with a sound level calibra-
tor. The sensitivity of the microphone is therefore in
principle without any interest. A significant change in the
sensitivity may, however, still be important as it may be a
sign of instability in the microphone.
If the sound level meter is normally calibrated by its user
with an internal reference voltage, it is essential that the
microphone sensitivity is known and stable.
The total acoustical and electrical noise from the microphone and
the sound level meter shall be at least 5 dB below the specified
minimum A-weighted sound pressure level.
Note - This test gives information about the noise from both the
sound level meter and the microphone.
Besides testing the sound level meter, the result of the test
is of value for practical low level noise measurements as the
noise floor of the complete apparatus is measured.
All tests shall be performed after the warm-up period for the in-
strument being tested and at the lowest adequate supply (battery)
voltage stated by the manufacturer.
The frequency weighting with tolerances are given in Tables 3 and ft.
For sound level meters of a given type the tolerances are identical
for all weighting characteristics.
Nominal frequencies are as specified in IS0 Standard 266. Exact frequencies are given above to four significant figures
and are equal to 1 000 . 10jril(), where n is a positive or negative integer.
- 15 -
Nominal frequency
(Hz) Type 1 Type 2
10 +3; - ∞ +5; - ∞
12.5 +3; - ∞ +5; - ∞
16 +3; - ∞ ‘+5; - ∞
20 ±3 ±3
25 ±2 ±3
31.5 ±1.5 ±3
40 ±1.5 ±2
50 ±1.5 ±2
63 ±1.5 ±2
80 ±1.5 ±2
100 ±1 ±1.5
125 ±1 ±1.5
160 ±1 ±1.5
200 ±1 ±1.5
250 ±l ±1.5
315 ±1 ±1.5
400 ±1 ±1.5
500 ±l ±1.5
630 ±l ±1.5
800 ±1 ±1.5
1 000 ±1 ±1.5
1 250 ±1 ±1.5
1 600 ±l ±2
2 000 ±l ±2
2 500 ±1 ±2.5
3 150 ±1 ±2.5
4 000 ±l ±3
5 000 ±1.5 ±3.5
6 300 +1.5; - 2 ±4.5
8 000 +1.5; - 3 ±5
10 000 +2; - 4 +5; - ∞
12 500 +3; - 6 +5; - ∞
16 000 +3; - ∞ +5; - ∞
20 000 +3; - ∞ +5; - ∞
Frequency
Type 1 Type 2
(Hz)
The residual noise of the sound level meter shall be measured accord-
ing to Sub-clause 7.3.
Crest factor
<3 G5 < 10
1 I ±1 ±1.5
±O.5 - - - - - - - - - - - - - - - - - - - - - -- - - - - - - - - - - ------ - - - - - - - -
1 - -
2I ±1 -
±1 ._-__-_-___------------------.----------------------------
2 - -
Note - If only one of the tests is performed, the tone burst test
is to be preferred.
The rectangular test pulses shall have durations of 200 µs and rise
times less than 10 µs. The tone burst signal shall consist of an
integral number of sine waves starting and ending at zero crossing.
The repetition frequency shall be 40 Hz.
The test signals shall be such that when compared with the reference
sinusoidal signal, using a device that introduces frequency weight-
ing corresponding to that in the sound level meter being tested whit-
in the tolerances given in Table 4 page 15, the rms-values shall be
equal. The C-weighting or Lin characteristic, if any, shall be used.
If the instrument has only A- or B-weighting, then only the tone
burst test shall be performed.
The test signal is fed to the electrical input of the sound level
meter, and the test is performed for the S detector-indicator cha-
racteristic or for the F characteristic, if S is not available in
the sound level meter being tested.
- 18 -
Maximum response to
Tolerances on maximum response
Duration of test test tone burst referred
for each instrument type
Detector-indicator tone burst to response to
characteristic continuous signal (dB)
(see Appendix C)
1 2
(ms) (dB)
Continuous 0
+l
F 200 - 1.0 +1
-2
S 500 - 4.1 ±1 ±2
-
Detector-indicator
Type 1 Type 2
characteristic
1.1
1.6
- 19 -
When the range of the indicator is more than 20 dB, the tests of
onset transient characteristics and overshoot shall be conducted
using signals that step up (i.e. suddenly increase) in amplitudes
by 20 dB. The tests shall be performed at 4 dB below the upper limit
of the primary indicator range and at intervals of 20 dB below this
level for all signals that produce an indication.
Continuous 0
20 - 3.6 ±1.5 ±2
5 - 8.8 ±2 ±3
2 - 12.6 ±2 No test
Continuous 0
100 -2.7
20 -7.6
2 -8.8
The indication of Leq, if any, shall not differ from the indica-
tion in the S-mode (F-mode, if S-mode is not available) by more than
± 0.3 dB for Type 1 and ± 0.5 dB for Type 2.
For the same signal the indication of the Peak mode shall be 3 dB
higher than the S-mode (F-mode, if S-mode is not available) within
tolerances of ± 0.5 dB for Types 1 and 11 and ± 0.7 dB for Type 2.
Note 1 - When the range of the indicator is more than 20 dB, the
test may be performed at 5 dB divisions only. This also
applies to the lower part of the indicator range if the
resolution of this part of the scale is poor.
would exceed the limits of the indicator range. The duration of the
tone burst shall be at least 1 ms, and the burst duty factor shall
not lie outside the range of values of Table 14. Alternatively, if
the linearity range extends below or above the indicator range, the
linearity range may be determined using suitable signals and test
points within the integrating sound level meter.
The numerical value of the linearity range shall be not more than
3 dB below the numerical value of the pulse range. It may, however,
be equal to or exceed the pulse range.
Tolerances in decibels
Burst duty R.M.S. amplitude of an individual tone burst relative
factor of test to the r.m.s. amplitude of the continuous signal
signal expressed in decibels Type Type
1 2
Continuous 0 - -
l/10 10 ± 0.5 ± 1.0
l/102 20 ± 0.5 ± 1.0
l/1043 30 ± 1.0 ± 1.5
l/l0 40 ± 1.0 -
The time averaging test compares the reading indicated on the re-
ference range for continuous sinusoidal signals with that obtained
from a sinusoidal tone burst sequence having the same equivalent
continuous level.
- 25 -
Continuous 15 0 11.8
1/10 10 ms 15 0 11.8
l/102 100 ms 15 0 11.8
l/103 1 s 3ob 0 14.8
l/104 10 s 6OC 0 17.8
Note - The tolerances for the pulse range refer to the deviations
from the theoretical value.
The test shall be performed with a peak level of the tone burst
which exceeds the rms-level of the continuous signal by the number
of decibels specified in Table 16.
At no peak level of the tone bursts during this test shall the in-
dicated value deviate from the theoretical value of the equivalent
continuous level of the test signal by more than the tolerances
specified in Table 16.
1 ms 20 10.4 *k
10ms 30 20
Tone burst duration
100ms i 40 30
1s 50 40
The overload indicator shall be checked at the same time as the per-
formance of the pulse test described in Sub-clause 8.3.3.
This test applies only to instruments that have the frequency weight-
ing characteristic A. The instrument shall be set to A-weighting and
the microphone cartridge replaced by an electrical impedance equal
to that of the microphone. A sinusoidal signal of frequency 1000 Hz
shall be applied to the instrument through this impedance with an
amplitude which gives a reading of 5 dB below that of the maximum
A-weighted sound pressure level which the instrument is designed to
measure. When applicable, dual independently adjustable level range
controls shall be set according to the manufacturer's instructions.
The frequency of the input signal shall then be lowered in steps to
20 Hz while the amplitude is simultaneously raised by multiples
corresponding to the inverse of the A-weighting characteristic given
in Table 3 page 14. If, at any frequency, the indication of the
instrument deviates from its initial value at 1000 Hz by more than
the tolerance (in practice, the lower tolerance) given in Table 4
page 15, for the corresponding frequency, a clear overload indica-
tion shall occur.
- 30 -
9. REPORT
Appendix A
* The zero of the amplitude scale is taken as the arithmetic mean value of the waveform.
Apply the 2000 Hz sinusoidal signal to the instrument under test and
simultaneously to a reference system having a true r.m.s. response
and a frequency weighting network N, corresponding to that in the
sound level meter being tested. Note the indication of the reference
meter.
Apply the rectangular pulse sequence and adjust its amplitude to give
an indication on the reference r.m.s. meter identical to that for the
sinusoidal signal. The instrument under test shall then give an indi-
cation within the tolerances specified in this standard.
- 32 -
For the rectangular pulse shown, the relation between crest factor
(G/u) and pulse duty factor (ti/T) is given by:
where:
ti is the time during which the siganl is at its peak value XT.
(l&h> = \/ 2T / ti
where:
6 u and T are as defined above, and where ti is the time during which
the signal has a non-zero value.
- 33 -
Appendix B
The values given in Tables 7, 9 and 10, pages 18, and 20, are obtained
using the following formulae:
1 - exp Gt;/d dB
∆ L = 10 log
10
1 - exp (-Y/-r)
where:
Exponential
Exponential
Peak detector;
averaging circuit Indicator
Input Squaring decay time
N one pole; I H calibrated
circuit constant
time constant in decibels
1 500 ms
35 ms