Pre-use Test – AC Failure Alarm Test – Make sure that power off failure alarm occurs, it has the following
characteristics: Alarm sound; “AC Failure!” message displays on the top screen.
Pre-use Test – AC Failure Alarm Test – (Connect power supply cable again) Make sure the alarm is
eliminated.
Pre-use Test – Alarm Test – Perform the Continuous high pressure alarm test.
Pre-use Test – Alarm Test – Perform the Low tidal minute volume test.
Pre-use Test – Alarm Test – Perform the Low oxygen concentration alarm test.
Pre-use Test – Alarm Test – Perform the Circuit Disconnect alarm test.
Pre-use Test – Alarm Test – Perform the Apnea alarm test.
Pre-use Test – Breathing System Test – Check the work state of the ventilator.
Pre-use Test – Breathing System Test – Check tidal volume – Check the tidal volume monitor is at a
range of 400mL ±15%.
Pre-use Test – Breathing System Test – Check the SIMV ventilation mode.
Pre-use Test – Breathing System Test – Sigh – (standard work state) Record the tidal volume.
Pre-use Test – Breathing System Test – Sigh – (Sigh mode) Check that the tidal volume data is 1.5 time as
normal the second time respiration takes place.
Pre-use Test – Breathing System Test – Check the SPONT ventilation mode.