Replace the Oxygen sensor.

Replace the Filter element.

1 year

Replace the Ni-MH battery.

Replace the Screwed pipe.

1500 cycles

Replace the Sampling silicone pipe.

Replace the Silicone pipe Ф3x1.5.

6 months

Record the number of operating cycles.

Replace the Air inlet filter sponge.

Calibration – Perform the Pressure Sensor Calibration.

Calibration – Perform the O2 Sensor Calibration.

Calibration – Perform the Flow Sensor Calibration.

Calibration – Perform the Expiratory Valve Calibration.

Calibration – Perform the Altitude Calibration.

Pre-use Test – AC Failure Alarm Test – Make sure that power off failure alarm occurs, it has the following
characteristics: Alarm sound; “AC Failure!” message displays on the top screen.

Pre-use Test – AC Failure Alarm Test – (Connect power supply cable again) Make sure the alarm is
eliminated.

Pre-use Test – Alarm Test – Perform the High-pressure alarm test.

Pre-use Test – Alarm Test – Perform the Continuous high pressure alarm test.

Pre-use Test – Alarm Test – Perform the Low tidal minute volume test.

Pre-use Test – Alarm Test – Perform the Low oxygen concentration alarm test.

Pre-use Test – Alarm Test – Perform the Circuit Disconnect alarm test.
Pre-use Test – Alarm Test – Perform the Apnea alarm test.

Pre-use Test – Breathing System Test – Check the work state of the ventilator.

Pre-use Test – Breathing System Test – Check tidal volume – Check the tidal volume monitor is at a
range of 400mL ±15%.

Pre-use Test – Breathing System Test – Test trigger pressure sensitivity.

Pre-use Test – Breathing System Test – Check the SIMV ventilation mode.

Pre-use Test – Breathing System Test – Sigh – (standard work state) Record the tidal volume.

Pre-use Test – Breathing System Test – Sigh – (Sigh mode) Check that the tidal volume data is 1.5 time as
normal the second time respiration takes place.

Pre-use Test – Breathing System Test – Check the SPONT ventilation mode.