Professional Documents
Culture Documents
922–930, 2011
Ó The Author 2011. Published by Oxford University Press
on behalf of the British Occupational Hygiene Society
doi:10.1093/annhyg/mer016
In this work, a method previously proposed in the literature (Sisto R, Pinto I, Stacchini N et al.
Infrared radiation exposure in traditional glass factories. AIHAJ 2000; 61: 5–10) to evaluate
the exposure to infrared (IR) radiation when the source can be approximated as a black body
is implemented in a mathematical code developed in Matlab. Some practical situations are
discussed. A comparison between the results obtained by a spectroradiometric technique
and that obtained by using a broadband radiometer and the modelling of the source is shown.
The IR radiation exposure evaluations in a cement industry and in a steel forge are shown and
compared to the exposure limit values.
Keywords: black body source; cataract; infrared radiation
922
Modelling IR radiation exposure 923
the worker during the working day and, in particular, the Z 1400
orientation of the surface tangent to the eyes of the ex- 6 106
LR 5 LðkÞRðkÞdk , ðWRðkÞ m 2 sr 1 Þ:
posed worker with respect to the direction orthogonal 780 Ca
to the emitting surface can be highly variable. ð7Þ
It is so necessary to perform direct measurements
of the irradiance at the different positions occupied For extended sources, Ca is the angle subtended by
by the eyes of the exposed worker. the source, expressed in mrad.
As no dispersive media between the source and the R(k) is the dimensionless weighting factor
receiving surface are assumed to be present, the tabulated in the ICNIRP (1997) guidelines and
following equation holds: further on called RICNIRP(k) to avoid any confusion
EL’s for thermal retinal hazard in the visible and near In the case of long exposure duration t . 1000 s,
IR or in the near IR only, if the visual stimulus is weak, the exposure limit is independent from t:
are kept into account, as reported in the following:
Z 1400 Z
2:8 107
3000
LR 5 LðkÞRðkÞdk , ðWRðkÞ m 2 sr 1 Þ EIR 5 E k dk , 100 Wm 2 : ð9Þ
380 Ca 780
Fig. 1. Comparison between weighting factors for thermal retinal hazard given by ACGIH and tabulated in the ICNIRP 1997
guidelines and in the European Directive.
Modelling IR radiation exposure 925
of interest kx ky by simply measuring the irradi- Application without measurements
ance in the spectral range ðk1 k2 Þ in which the avail-
In special cases of simple geometry of the
able detector is sensitive and than correcting the results
source, an evaluation is possible even without any
by the correction factor on the right side of equation (5).
measurement.
The possibility of evaluating the correction
In the case of steel bolts forging, the worker is
factors, by setting the hygienistic range of concern,
requested to observe each piece just after induction
the range of sensitivity of the detector, and by
heating to the temperature of 927°C, from a minimum
importing the sensitivity curve of the available
50 cm distance.
detector was implemented in a mathematical
The overheated bolt can be assumed to have an
programme developed in Matlab. The programme
Fig. 2. First graphical output of the Matlab code, the solid line represents the portions of the block body curve that is detected by
the radiometer.
926 R. Sisto et al.
the range 780–3000 nm and the total emittance The measurements obtained by the power meter
represented by (3) is 0.5. Multiplying the result were corrected by the suitable correction factors
of (11) by this value yields a total radiance estimate calculated by equation (5) to quantify the irradiance
below the limit value for IR cataractogenesis given in the ranges of protectionistic interest for cataracto-
by equation (9). genesis from 780 to 3000 nm.
As regards the retinal hazards, similar consider- In order to verify the method in the optical IR
ations should be applied considering also that the range, the same measurements were performed also
spectral radiance must be multiplied by the weight- by using a spectroradiometer Ocean Optics HR4000,
ing function RICNIRP(k). As the visual stimulus is shown in Fig. 4.
weak, the limit for the weighted radiance given by As the spectroradiometer is sensitive in the
Fig. 3. Detector used for irradiance measurements. Fig. 4. Spectrodiameter used for verification.
Modelling IR radiation exposure 927
Table 1. Effect of the extrapolation method on the RICNIRP-weighted irradiance measured by the spectroradiometer
Source Spectral Linearly extrapolated Flat extrapolated % Difference
range (nm) irradiance (W m2) irradiance (W m2)
Cement kiln 380–1400 5.77 5.46 5.4
780–1400 4.60 4.29 6.7
Cement kiln 380–1400 29.72 28.46 4.2
780–1400 22.05 20.79 5.7
Steel forging 380–1400 1.73 1.58 8.9
780–1400 1.42 1.27 10.7
Steel forging 380–1400 2.77 2.50 9.8
780–1400 0.57 0.31 45.6
Considering that the spectral range of interest this region ranges from 0.2 to 0.02 and the influence of
starts at 380 nm, measured spectra were also set to this part of the integral must be compared to that in the
zero for k , 350 nm. The presence of false signals visible region where the weighting factor increases up
at the extreme of the sensitivity band of the spectror- to 10, as can be seen from Figs 1 and 2.
adiometer, indeed, is a typical problem in measure, Nonetheless, considering that the real spectrum
as described in the literature (Frigerio, 2010). has to be modelled with the black body curve, the
linear extrapolation is the most accurate.
The results obtained by the spectroradiometer can
RESULTS
now be compared to those obtained by integrated ra-
In Table 1, the RICNIRP-weighted integrals of the diometric measurements. These last have to be mul-
measured spectra obtained with the two diffe- tiplied by the suitable correction factor, evaluated by
rent extrapolation technique are compared: the flat equation (5) with sðkÞ 5 1, as the detector has a flat
extrapolation gives a value ,10% lower than the response in its sensitivity range.
linear extrapolation, except in the last measurement. In Table 2, the results of this comparison are
In this particular case, the measurements were presented in the ranges 380–1400 nm and 780–
made in condition of high visible background due 1400 nm. The comparison between the weighted
to the illumination of the workplace; the conse- irradiance from spectroradiometric measures and
quence of this condition will be discussed further. the estimated weighted irradiance by radiometric
A change in the extrapolation method is thus proven to measurements gives a quite good agreement between
be negligible, mainly because the RICNIRP(k) curve in the two different techniques. In only one case, the
928 R. Sisto et al.
Table 2. Comparison between radiometric and spectroradiometric RICNIRP-weighted irradiance evaluation in visible plus optical
IR range (380–1400 nm) and optical IR range (780–1400 nm). The RICNIRP weighting factor is that given by ICNIRP (1997) and
tabulated in the Commission of the European Communities (2006)
Source T (°C) Spectral range RICNIRP-weighted Weighting factor RICNIRP-weighted
irradiance from implementing estimated irradiance
spectroradiometric w(k) 5 RICNIRP(k) by radiometric
measures (WR(l)m2) measurements
(WR(l)m2)
Cement kiln 1200 380–1400 nm 5.77 0.0126 5.1
780–1400 nm 4.60 0.0112 4.54
Table 3. RACGIH-weighted irradiance evaluation in visible plus optical IR range (380–1400 nm) and optical IR range (780–1400
nm). The absolute irradiance in the 780–3000 nm range is also shown
Source T(°C) Total irrad. Spectral Weighting Weighting factor RACGIH-weighted Estimated
(W m2) in range factor from implementing estimated irradiance absolute
the range equation (12) w(k) RACGIH(k) by radiometric irradiance by
200–6000 nm with w(k) 5 1 measurements radiometric
(WR(l) m2) measurements
(W m2)
Cement 1200 405 380–1400 nm 0.0884 0.021 8.5 35.8
kiln 780–1400 nm 0.0868 0.0196 7.94 35.2
780–3000 nm 0.6241 252.8
Cement 1300 771 380–1400 nm 0.1145 0.0283 21.82 88.3
kiln 780–1400 nm 0.1116 0.0258 19.89 86
780–3000 nm 0.6609 509.6
Steel 1033 233 380–1400 nm 0.0516 0.0116 2.7 12
forging 780–1400 nm 0.0511 0.0112 2.6 11.9
780–3000 nm 0.5504 — — 128.2
Steel forging 960 229 380–1400 nm 0.0386 0.0085 1.95 8.8
780–1400 nm 0.0383 0.0082 1.88 8.8
780–3000 nm 0.5131 117.5
Bold values are those exceeding the TLV.
difference between the results is .100%; this mea- 780–3000 nm is also compared to the exposure limit
sure shows that several experimental conditions values for cataractogenesis of equation (9).
can affect field measurements. In this particular case, As it can be easily read from the same table, the
it was not possible to make the measure with the exposure limit values for cataractogenesis are al-
spectroradiometer within a short time interval from ways exceeded, both at the inspection of the cement
the measure with the radiometer. kiln and in the steel forging activity. The exposure
In Table 3, the RACGIH-weighted irradiance limit value of equation (9) is exceeded by a factor
evaluation by power meter measurements in the 5 in the case in which the temperature is 1300°C.
ranges 380–1400 nm and 780–1400 nm are shown, Among the various conditions that can affect
together with the suitable correction factors calcu- measurements, it is worth quoting the following
lated from equation (5). The limit exposure values ones:
for thermal retinal hazards largely exceed the values
found if also the RACGIH weighting factor is applied, (a) uncertainty in the positioning of the detector,
which adds a penalty in the IR-A range with respect specially critical when an instrument like that
to the RICNIRP. The absolute irradiance in the range presented in Fig. 4 has to be used on field,
Modelling IR radiation exposure 929
(b) uncertainty in extrapolation of spectra, For the same reason, the method does not apply to
(c) acceptance of detector: the spectroradiometer, as outdoor workplaces where sunlight contribution,
usual for broad beam instruments, has a cosine- even in cloudy conditions, would overwhelm black
corrected angular acceptance, while the radio- body spectrum; fortunately, for the worker and the
meter, designed for laser beams, has a simple consultant, outdoor working on incandescent mater-
cylindrical aperture as shown in Fig. 3. ials is not very common in western countries.
This last, in particular, introduces some inaccu-
racy, which is important when the evaluations are CONCLUSIONS
based on radiometric measures only.
FUNDING