You are on page 1of 9

Ann. Occup. Hyg., Vol. 55, No. 8, pp.

922–930, 2011
Ó The Author 2011. Published by Oxford University Press
on behalf of the British Occupational Hygiene Society
doi:10.1093/annhyg/mer016

Modelling Infrared Radiation Exposure by Black


body-Like Sources
RENATA SISTO1*, FRANCESCO FRIGERIO2, ANDREA MILITELLO1,

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


MASSIMO BORRA1, DANILO COTTICA2 and ELENA GRIGNANI2
1
Department of Occupational Hygiene, National Research Centre for Occupational Safety and
Prevention, Italian Workers’ Compensation Authority, Monte Porzio Catone, Rome 00040, Italy;
2
Environmental Research Center, Fondazione Salvatore Maugeri IRCCS, Scientific Institute of Pavia,
Pavia 27100, Italy
Received 26 November 2010; in final form 11 February 2011; published online 7 April 2011

In this work, a method previously proposed in the literature (Sisto R, Pinto I, Stacchini N et al.
Infrared radiation exposure in traditional glass factories. AIHAJ 2000; 61: 5–10) to evaluate
the exposure to infrared (IR) radiation when the source can be approximated as a black body
is implemented in a mathematical code developed in Matlab. Some practical situations are
discussed. A comparison between the results obtained by a spectroradiometric technique
and that obtained by using a broadband radiometer and the modelling of the source is shown.
The IR radiation exposure evaluations in a cement industry and in a steel forge are shown and
compared to the exposure limit values.
Keywords: black body source; cataract; infrared radiation

INTRODUCTION best technology available for field measurements is


based on detectors whose sensitivity does not extend
Evaluating worker’s exposure to optical artificial much .1100 nm. For thermal hazards not involving
radiation, following the guidelines of International the retina, the integration must be extended to 3000
Commission on Non Ionizing Radiation Protection nm but, spectral weighting being not required,
(ICNIRP) requires the calculation of a number of a simple integrated measure instead of a spectral
integrals over the wavelength of the spectrum of one is adequate.
the incident radiation (ICNIRP, 1997). The knowl- The IR radiation exposure is known to cause eye
edge of the spectrum can be obtained from the manu- damage, in particular through thermal cataractogenesis
facturer of the source or, more often, from (Vos and van Norren, 2004). In particular, the damage
measurements. mechanisms due to IR-A and IR-B radiation are well
At the moment, a single instrument designed to known, reproduced in animal studies and specific expo-
perform the measurements required for optical radi- sure limits are given by ICNIRP in these spectral ranges.
ation evaluation in all the wavelength ranges of As regards IR-C, the role of radiation in
hygienistic interest is not available. this spectral range is not clarified in the aetiopatho-
For infrared (IR) sources, a serious problem in genesis of cataract and no specific limits have been
measurements is that retinal damage risk must be defined for radiation from 3000 nm to 1 mm.
evaluated from weighted integrals between 380 Cataractogenesis in workers who work in proximity
or 780 and 1400 nm, depending on the intensity of of overheated materials is known to be due to
the stimulus in the visible range. Unfortunately, the electromagnetic radiation in a spectral region ran-
*Author to whom correspondence should be addressed.
ging from the visible to the IR-C band (Okuno, 1994).
Tel: +39 06 94181552; fax: +39 06 9419453; The visible and IR-A radiation is absorbed by
e-mail: renata.sisto@ispesl.it the iris and the thermal energy is transmitted by

922
Modelling IR radiation exposure 923

conduction to the crystalline lens. The IR-B and IR- METHODS


C radiation is absorbed by the cornea and the heat is
When the source of radiation is an overheated
transmitted to the lens by thermal conduction from
material and no screens or other dispersive media
the nearby tissues. In both cases, visible and IR-A
are present between the source and the receiving
or IR-B and IR-C radiation, the cataractogenesis is
surface, the spectrum can be modelled as a black
due to the rise of temperature in the lens.
body.
In an epidemiological study conducted in
The flux of energy emitted in a spectral band by
Sweden’s iron and steel industries, Lydahl and
a black body at the absolute temperature T is given
Philipson (1984) found some evidence of the correl-
by the Planck’s law (e.g. Greiner, 2000):
ation between IR radiation exposure and thermal

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


cataract development. One of the most relevant 2phc2
causes of IR exposure of workers is the processing MT ðkÞdk 5 5
dk: ð1Þ
k ðehc=kKT  1Þ
of high temperature materials. Such sources can
be modelled by black body radiation spectrum, as The quantity MT (k) is the black body spectral
given by Planck in 1910. Calculations of retinal, emittance, representing the spectral flux of energy
corneal and lens hazards from black body radiation emitted per unit surface by the source.
have been developed since 1980 (Sliney and For the emittance to be expressed in watts per
Wolbarsht,1980). The use of a black body model to eval- square metre, the following values of the physical
uate exposure of workers has been already described in constants must be used
the literature for glass industry (Sisto et al., 2000). 8
In a 2006 statement, ICNIRP discusses the < h 5 6:62  10  34 J  s;
applicability of this assumption to a number of c 5 2:99  108 ms  1 ; ð2Þ
:
materials showing that assuming black body emission K 5 1:38  10  23 J  K  1 :
can lead to an overestimate of the energy emitted; Assuming that no dispersive media are present,
from the point of view of protection, this is not a big only a geometrical factor defines the proportionality
problem, moreover, increasing the temperature, the between the radiance of the source and the measured
spectrum actually approaches a black body curve. irradiance due to source itself at distance r.
When the radiation source present in working envi- Integrating equation (1) gives the Stefan–Boltzmann’s
ronment can be approximately modelled as a simple law, the power emitted per unit area of the surface of
black body, we will show that it is possible to evaluate a black body is directly proportional to the fourth power
the exposure in terms of irradiance in any range of of its absolute temperature. That is,
hygienistic relevance.
To achieve this, knowledge is needed of the j 5 r  T 4; ð3Þ
geometrical factor relating the source radiance to
the irradiance measured at the level of the eye of where r is the Stefan–Boltzmann constant
exposed workers and of the spectral sensitivity of r 5 5:67  108 W  m2 K4 :
the detector. The total power emitted by the source is given
In this work, the physical bases of this method are by the equation (3) times the surface of emission.
discussed and practical application tools for the In the case in which the emission is isotropic, the
calculation are presented. irradiance measured at a distance r from the source
The simplest one is a spreadsheet developed in MS can be easily evaluated as the ratio between the total
Excel; the relative simplicity of the formulas used power emitted by the source and the area of the
makes it usable with most of the available open source spherical surface of radius r :
spreadsheets too. The second one is a more versatile
r  T4  S
computation code developed in Matlab ((c) 1984– Er 5 : ð4Þ
2007; The MathWorks, Inc.), which allows to modify 4p  r2
the black body curve in order to consider also emis- In special conditions, the evaluation is therefore
sivity corrections, the effect of goggles or other possible even without any irradiance measurement;
shielding and even to consider the different IR the temperature, indeed, is one of the best-controlled
weighting factors proposed by American Conference parameters of the industrial process.
of Government Industrial Hygienist and ICNIRP. In many practical situations, the geometrical
An application of the method is shown for two factor relating the emittance of the source to the
cases in which verification measurements on field in irradiance measured on a particular surface can be diffi-
Italy were possible. cult to evaluate. In addition, the positions occupied by
924 R. Sisto et al.

the worker during the working day and, in particular, the Z 1400
orientation of the surface tangent to the eyes of the ex- 6  106
LR 5 LðkÞRðkÞdk , ðWRðkÞ m  2 sr  1 Þ:
posed worker with respect to the direction orthogonal 780 Ca
to the emitting surface can be highly variable. ð7Þ
It is so necessary to perform direct measurements
of the irradiance at the different positions occupied For extended sources, Ca is the angle subtended by
by the eyes of the exposed worker. the source, expressed in mrad.
As no dispersive media between the source and the R(k) is the dimensionless weighting factor
receiving surface are assumed to be present, the tabulated in the ICNIRP (1997) guidelines and
following equation holds: further on called RICNIRP(k) to avoid any confusion

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


with the weighting factor tabulated by the ACGIH
  R ky
Ekx  ky r dk9wðk9ÞMk ðk9Þ for thermal retinal damage risk evaluation. The latter
5 Rkkx 2 : ð5Þ will be called RACGIH(k). A comparison between
Emeas ðrÞ
k1 dk9sðk9ÞMk ðk9Þ RICNIRP(k) and RACGIH(k) is shown in Fig. 1.
The weighting factor w(k) can be the appropriate In the case of corneal burn and cataractogenesis,
weighting factor for the considered biological the physical quantity that defines the exposure limits
damage effect, when a comparison with exposure limit is the irradiance evaluated at the level of the eyes of
values is required, or it can be identically 5 1 in other the exposed workers.
cases; sðkÞ represents the detector sensitivity curve. In In the case of short exposure duration t , 1000 s,
this paper, the exposure limit values, EL’s, for retinal the following limit holds:
damage, stated by the ICNIRP (1997) guidelines and Z 3000  
 34 2
transposed into the Commission of the European EIR 5 EðkÞdk , 18 000  t Wm : ð8Þ
Communities (2006) are considered. In particular, the 780

EL’s for thermal retinal hazard in the visible and near In the case of long exposure duration t . 1000 s,
IR or in the near IR only, if the visual stimulus is weak, the exposure limit is independent from t:
are kept into account, as reported in the following:
Z 1400 Z
2:8  107
3000    
LR 5 LðkÞRðkÞdk , ðWRðkÞ m  2 sr  1 Þ EIR 5 E k dk , 100 Wm  2 : ð9Þ
380 Ca 780

ð6Þ On the basis of equation (5), it is possible to


evaluate the irradiance in the given hygienistic range
valid for exposures with a duration t . 10 s, and

Fig. 1. Comparison between weighting factors for thermal retinal hazard given by ACGIH and tabulated in the ICNIRP 1997
guidelines and in the European Directive.
Modelling IR radiation exposure 925
 
of interest kx  ky by simply measuring the irradi- Application without measurements
ance in the spectral range ðk1  k2 Þ in which the avail-
In special cases of simple geometry of the
able detector is sensitive and than correcting the results
source, an evaluation is possible even without any
by the correction factor on the right side of equation (5).
measurement.
The possibility of evaluating the correction
In the case of steel bolts forging, the worker is
factors, by setting the hygienistic range of concern,
requested to observe each piece just after induction
the range of sensitivity of the detector, and by
heating to the temperature of 927°C, from a minimum
importing the sensitivity curve of the available
50 cm distance.
detector was implemented in a mathematical
The overheated bolt can be assumed to have an
programme developed in Matlab. The programme

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


irradiating surface of 28 cm2.
also permits to set the emitting source temperature.
The total power irradiated is found multiplying
In Fig. 2, the black body emittance of a source at
equation (3) by the surface S of irradiation:
1400°C is shown as calculated by the Matlab code
when a radiometer with flat sensitivity between  
e 5 r  T 4  S 5 329 W : ð10Þ
200 and 6000 nm is considered as detector.
In the first plot, the code itself compares the The irradiance at the eye of the worker in this case
calculated emittance at the chosen temperature with can simply be found considering that, when the bolt
the same curve multiplied by the RICNIRP(k) is observed from a 50 cm distance, it can be assumed
weighting factor and the one multiplied for the as a point-like source irradiating on a spherical
Commission Internationale de l’Eclairage curve, thus surface.
suggesting the relative importance of the visible The irradiance at 50 cm is then given by
range. e
The total energy emitted by the source, repre- E50 cm 5 5 104:8 Wm  2 : ð11Þ
4p  r2
sented by equation (3), is in fact given by the area
under the whole curve between 200 and 10 000 nm, while the total radiance can be estimated as:
the fractions of the energy in other intervals are
r2
given by the areas subtended by the corresponding L 5 E50 cm  ffi 9:36  103 Wm  2 sr  1 ð12Þ
curves. S
It is accordingly easy to see that, at the tempera- with a source apparent diameter Ca  120 mrad.
tures typical of industrial processes, most of the At 1200 K, it can be shown using the code that the
energy is emitted in the IR range. ratio between the integral of black body spectrum in

Fig. 2. First graphical output of the Matlab code, the solid line represents the portions of the block body curve that is detected by
the radiometer.
926 R. Sisto et al.

the range 780–3000 nm and the total emittance The measurements obtained by the power meter
represented by (3) is 0.5. Multiplying the result were corrected by the suitable correction factors
of (11) by this value yields a total radiance estimate calculated by equation (5) to quantify the irradiance
below the limit value for IR cataractogenesis given in the ranges of protectionistic interest for cataracto-
by equation (9). genesis from 780 to 3000 nm.
As regards the retinal hazards, similar consider- In order to verify the method in the optical IR
ations should be applied considering also that the range, the same measurements were performed also
spectral radiance must be multiplied by the weight- by using a spectroradiometer Ocean Optics HR4000,
ing function RICNIRP(k). As the visual stimulus is shown in Fig. 4.
weak, the limit for the weighted radiance given by As the spectroradiometer is sensitive in the

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


equation (7) holds. Considering the apparent range 200–1000 nm, an extrapolation is needed to
diameter of 120 mrad, a limit of 0.5105 W m2 the spectral range 1000–1400 nm, corresponding to
sr1 was found for the RICNIRP-weighted radiance. the IR radiation focused by the crystalline lens.
On the other hand, the ratio between the RICNIRP- The simplest and conservative way to obtain the
weighted spectral radiance integrated over the IR-A shape of the spectrum outside the spectroradiometer
range and the black body total radiance in the power range is to linearly extrapolate the spectrum to 1400
meter sensitive range is 3.3103. Multiplying the nm with the same slope measured between 900 and
radiance of equation (12) by this last factor results 1000 nm. The measured and extrapolated spectra
in a negligible exposure for thermal retinal hazard. were finally convoluted with the RICNIRP-weighting
function. In the spectral region between 1000 and
Verification measurements 1400 nm, the relative importance of the weighted
With the aim at evaluating the applicability of the irradiance with respect to the weighted irradiance
method in working environments, measurements in the region 380–1000 nm is negligible.
were performed in two different working situations: To prove this, each measured spectrum has been
set, in the range from 1100 to 1400 nm, to the value
1. the forging of high diameter steel valves (about
measured at 1100 nm; this approximation was
0.5–3 m in diameter and 1 m length),
defined as flat spectrum extrapolation.
2. the inspection window of a cement production
In Fig. 5, a spectrum measured in front of an
kiln.
inspection window of the cement production kiln is
To measure the irradiance, a laser power meter shown. In the same figure, the linearly and flat
Ophir Nova II with thermopile detector 3-A-P- extrapolated spectra are also shown to better clarify
DIF-V, shown in Fig. 3, was used. the two different techniques.
This device, designed for laser measurements, has All screens of the window were removed in order
a flat response between 200 and 6000 nm and a power to measure the black body spectrum of the source,
dynamic range from 3 lW to 3 W. cement burning at temperature ranging from 900 to
The detector is calibrated for incident power and, 1400°C depending on the position of the material
knowing the surface, the irradiance can be obtained. in the furnace.

Fig. 3. Detector used for irradiance measurements. Fig. 4. Spectrodiameter used for verification.
Modelling IR radiation exposure 927

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


Fig. 5. Extrapolation of spectrum to 1400 nm; the source is cement burning at 1200 °C.

Table 1. Effect of the extrapolation method on the RICNIRP-weighted irradiance measured by the spectroradiometer
Source Spectral Linearly extrapolated Flat extrapolated % Difference
range (nm) irradiance (W m2) irradiance (W m2)
Cement kiln 380–1400 5.77 5.46 5.4
780–1400 4.60 4.29 6.7
Cement kiln 380–1400 29.72 28.46 4.2
780–1400 22.05 20.79 5.7
Steel forging 380–1400 1.73 1.58 8.9
780–1400 1.42 1.27 10.7
Steel forging 380–1400 2.77 2.50 9.8
780–1400 0.57 0.31 45.6

Considering that the spectral range of interest this region ranges from 0.2 to 0.02 and the influence of
starts at 380 nm, measured spectra were also set to this part of the integral must be compared to that in the
zero for k , 350 nm. The presence of false signals visible region where the weighting factor increases up
at the extreme of the sensitivity band of the spectror- to 10, as can be seen from Figs 1 and 2.
adiometer, indeed, is a typical problem in measure, Nonetheless, considering that the real spectrum
as described in the literature (Frigerio, 2010). has to be modelled with the black body curve, the
linear extrapolation is the most accurate.
The results obtained by the spectroradiometer can
RESULTS
now be compared to those obtained by integrated ra-
In Table 1, the RICNIRP-weighted integrals of the diometric measurements. These last have to be mul-
measured spectra obtained with the two diffe- tiplied by the suitable correction factor, evaluated by
rent extrapolation technique are compared: the flat equation (5) with sðkÞ 5 1, as the detector has a flat
extrapolation gives a value ,10% lower than the response in its sensitivity range.
linear extrapolation, except in the last measurement. In Table 2, the results of this comparison are
In this particular case, the measurements were presented in the ranges 380–1400 nm and 780–
made in condition of high visible background due 1400 nm. The comparison between the weighted
to the illumination of the workplace; the conse- irradiance from spectroradiometric measures and
quence of this condition will be discussed further. the estimated weighted irradiance by radiometric
A change in the extrapolation method is thus proven to measurements gives a quite good agreement between
be negligible, mainly because the RICNIRP(k) curve in the two different techniques. In only one case, the
928 R. Sisto et al.

Table 2. Comparison between radiometric and spectroradiometric RICNIRP-weighted irradiance evaluation in visible plus optical
IR range (380–1400 nm) and optical IR range (780–1400 nm). The RICNIRP weighting factor is that given by ICNIRP (1997) and
tabulated in the Commission of the European Communities (2006)
Source T (°C) Spectral range RICNIRP-weighted Weighting factor RICNIRP-weighted
irradiance from implementing estimated irradiance
spectroradiometric w(k) 5 RICNIRP(k) by radiometric
measures (WR(l)m2) measurements
(WR(l)m2)
Cement kiln 1200 380–1400 nm 5.77 0.0126 5.1
780–1400 nm 4.60 0.0112 4.54

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


Cement kiln 1300 380–1400 nm 29.72 0.0181 13.96
780–1400 nm 22.05 0.0155 11.95
Steel forging 1033 380–1400 nm 1.73 0.0061 1.42
780–1400 nm 1.42 0.0056 1.3
Steel forging 960 380–1400 nm 2.77 0.0041 0.94
780–1400 nm 0.57 0.0039 0.89

Table 3. RACGIH-weighted irradiance evaluation in visible plus optical IR range (380–1400 nm) and optical IR range (780–1400
nm). The absolute irradiance in the 780–3000 nm range is also shown
Source T(°C) Total irrad. Spectral Weighting Weighting factor RACGIH-weighted Estimated
(W m2) in range factor from implementing estimated irradiance absolute
the range equation (12) w(k) RACGIH(k) by radiometric irradiance by
200–6000 nm with w(k) 5 1 measurements radiometric
(WR(l) m2) measurements
(W m2)
Cement 1200 405 380–1400 nm 0.0884 0.021 8.5 35.8
kiln 780–1400 nm 0.0868 0.0196 7.94 35.2
780–3000 nm 0.6241 252.8
Cement 1300 771 380–1400 nm 0.1145 0.0283 21.82 88.3
kiln 780–1400 nm 0.1116 0.0258 19.89 86
780–3000 nm 0.6609 509.6
Steel 1033 233 380–1400 nm 0.0516 0.0116 2.7 12
forging 780–1400 nm 0.0511 0.0112 2.6 11.9
780–3000 nm 0.5504 — — 128.2
Steel forging 960 229 380–1400 nm 0.0386 0.0085 1.95 8.8
780–1400 nm 0.0383 0.0082 1.88 8.8
780–3000 nm 0.5131 117.5
Bold values are those exceeding the TLV.

difference between the results is .100%; this mea- 780–3000 nm is also compared to the exposure limit
sure shows that several experimental conditions values for cataractogenesis of equation (9).
can affect field measurements. In this particular case, As it can be easily read from the same table, the
it was not possible to make the measure with the exposure limit values for cataractogenesis are al-
spectroradiometer within a short time interval from ways exceeded, both at the inspection of the cement
the measure with the radiometer. kiln and in the steel forging activity. The exposure
In Table 3, the RACGIH-weighted irradiance limit value of equation (9) is exceeded by a factor
evaluation by power meter measurements in the 5 in the case in which the temperature is 1300°C.
ranges 380–1400 nm and 780–1400 nm are shown, Among the various conditions that can affect
together with the suitable correction factors calcu- measurements, it is worth quoting the following
lated from equation (5). The limit exposure values ones:
for thermal retinal hazards largely exceed the values
found if also the RACGIH weighting factor is applied, (a) uncertainty in the positioning of the detector,
which adds a penalty in the IR-A range with respect specially critical when an instrument like that
to the RICNIRP. The absolute irradiance in the range presented in Fig. 4 has to be used on field,
Modelling IR radiation exposure 929

(b) uncertainty in extrapolation of spectra, For the same reason, the method does not apply to
(c) acceptance of detector: the spectroradiometer, as outdoor workplaces where sunlight contribution,
usual for broad beam instruments, has a cosine- even in cloudy conditions, would overwhelm black
corrected angular acceptance, while the radio- body spectrum; fortunately, for the worker and the
meter, designed for laser beams, has a simple consultant, outdoor working on incandescent mater-
cylindrical aperture as shown in Fig. 3. ials is not very common in western countries.
This last, in particular, introduces some inaccu-
racy, which is important when the evaluations are CONCLUSIONS
based on radiometric measures only.

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


It is important to point out that also conversion The workers’ exposure to visible and near IR
from irradiance to radiance using geometrical para- radiation emitted by overheated materials can be
meters imply the assumption that the total energy evaluated with a simple radiometric measure if the
flux entering the detector comes from the source. source is modelled as a black body.
The uncertainty conditions (a) and (c) concern Evaluations without measurements are also
both the measurement method itself and the method possible in the case of particularly simple geometry.
used to verify its accuracy in the range 380–1400 The result is the more accurate the higher the
nm; the condition (b) is due to the lack of infor- temperature of the source and when other sources
mation on the spectrum .1100 nm, the filling of this in the field of view of the detector can be neglected.
gap is just the purpose of the method proposed. A simple code was developed to implement the
Considering all the uncertainty sources, a 20% method. The code needs the source temperature
accuracy in the evaluation of the single integrals re- and the detector sensitivity curve as input para-
quired for the hygienistic evaluation can be assumed. meters. The code gives as output the coefficients that
To improve the accuracy, caution must be used in permit to convert the measured irradiance in the irra-
taking integrated irradiation measurements in low diance relative to the different hygienistic ranges.
visible background conditions. The implementation of different weighting functions
In a situation like that shown in Fig. 6, a contri- for different biological damage effects is also
bution from the workplace artificial illumination is possible.
clearly present and corresponds to the last item of The method has been tested for the R-weighted
Table 2, where due to its lower temperature, less 380–1440 nm radiance evaluation showing an
energy from the source of interest enters the detector accuracy within 20%.
compared with the workplace illumination visible Further verification experiments, with direct
light. spectral measurements in the range 1100–3000 nm,
In this case, the spectrum is no longer a black body would require instrumentation at the moment not
curve and, as already noticed, also the influence of available for on field measurements.
the extrapolation technique strongly affects the Evaluations in the considered different working
result of the evaluations. environments, in the 900–1400°C temperature range
of the source of overheated material, show that the
exposure limit values given by the ICNIRP for cata-
ractogenesis are exceeded in all the working posi-
tions, while the limits provided to protect the retina
from thermal hazard are always respected.
Suitable protective glasses should be used by
workers exposed to radiation by overheated materi-
als, in particular when 1000°C temperature is ex-
ceeded, to shield the external part of the eye from
the IR-A and IR-B radiation.

FUNDING

This work was carried on with the research ordi-


nary funding of the INAIL Research, ex ISPESL
Fig. 6. Spectrum of an ovrheated steel body in presence of Department of Occupational Hygiene, Monteporzio
high optical background. Catone, ITALY, Activity programme 2009–2011.
930 R. Sisto et al.

REFERENCES Lydahl E, Philipson B. (1984) Infrared radiation and cataract I:


Epidemiologic investigation of iron and steel-workers, and
Commission of the European Communities. (2006) Directive II: Epidemiologic investigation of glass workers. Acta Oph-
2006/25/EC of the European Parliament and of the Council thalmol Scand; 62: 976–92.
of 5 april 2006. Off J Eur Union; 114: 38. Okuno T. (1994) Thermal effect of visible light and infra-red ra-
Frigerio F. (2010) La valutazione dell’esposizione dei diation (i.r.-A, i.r.-B and i.r.-C) on the eye: A study of infra-
lavoratori a radiazioni ottiche artificiali. [2010]. G Ital red cataract based on a model. Ann Occup Hyg; 38: 351–59.
Med Lav Ergon; XXXII: 40–48. (Abstract in english). Sisto R, Pinto I, Stacchini N et al. (2000) Infrared radiation
Greiner W. (2000) Quantum mechanics: an introduction. Ber- exposure in traditional glass factories. AIHAJ; 61: 5–10.
lin, Germany: Springer. Sliney D, Wolbarsht M. (1980) Safety with lasers and other op-
ICNIRP. (1997) Guidelines on limits of exposure to broad band tical sources. New York, NY: Plenum.
optical incoherent radiation (0.38 to 3 lm). Health Phys; 73: Vos JJ, van ND. (2004) Thermal cataract, from furnaces to

Downloaded from https://academic.oup.com/annweh/article/55/8/922/264342 by guest on 01 September 2022


539–53. lasers. Clin Exp Optom; 87: 372–76.

You might also like