Version 1.1
User Manual
IDS Geo
Legal Notices
Use of the Manual
This Manual is intended to be used solely by authorized user of IDS GeoRadar Systems. It is specific for internal business purpose only.
If you are not an authorized user, you are NOT allowed to read this Manual and you must return it to IDS GeoRadar as soon as possible.
Access to and use of this Manual is also provided subject to the acceptance of the Disclaimer set out below.
Disclaimer
This Manual has been drafted by IDS GeoRadar with the due care in order to ensure the accuracy and completeness. However, please note that no warranty is provided about the fact that the
Manual will be free from error or omission and about the accuracy and/or fitness to the purpose of any information contained herein.
Any representation and/or statement contained herein has been written in good faith and on the basis that, due to negligence, lack of care or otherwise, IDS GeoRadar is not liable for any loss,
including direct, indirect or consequential losses, in contract or in tort for any misstatement or omission contained herein which has occurred or may occur with reference to any representation
and/or statement contained in the Manual.
User shall also acknowledge that some information contained in the Manual may not be up to date. IDS GeoRadar may amend the information contained in the Manual from time to time and IDS
GeoRadar shall not be deemed liable in the event that damages arise with reference to changed circumstances or other information which may affect the accuracy of any information contained in
this version of the Manual.
All products designed, produced and distributed by IDS GeoRadar, its Subsidiaries, Affiliated and authorized Distributors (Products) shall not be used, in the design, development, production or use
of chemical, biological, nuclear ballistic weapons. Buyer/ User will defend, indemnify and hold IDS GeoRadar harmless against any liability (including attorney’s fees) for non-compliance with the
terms of this paragraph.
If IDS GeoRadar requires that Buyer/User shall carry out a training with reference to some Product categories, no operation or use of the Products shall be started before its designated Operator/s
has got the User Certificate, as defined by IDS GeoRadar specific procedure which the Buyer confirms to know and accept .
For Products which include specific “Operational” software with automatic data processing and analysis “Tools”, e.g. RockSpot, IBIS Products; Hydra Products, User shall be aware that the results
provided by these “Tools” may be not error free. User that completely relies on the outcomes provided by these Tools only, does it at his own risk.
In no event IDS GeoRadar shall be liable for special, direct, indirect, incidental, exemplary, punitive or consequential damages including, but not limited to, loss of profits or revenue, caused by the
missed or incomplete observance of the instructions and prescriptions for the use of the Products, either separately or in combination with other products.
This paragraph provides only a non-exhaustive summary of the full Product disclaimer. Please see the full Product disclaimer included into the relevant section of this Manual for detailed information.
Confidentiality
This Manual is strictly confidential and may be subject to legal and other privilege. IDS GeoRadar specifically reserves its rights in relation to this Manual.
None of these rights or any other rights subsisting in the Manual are waived if you have received this Manual in error or without permission.
No parts of this Manual may be copied, translated or reduced to any electronic medium, or machine-readable form, nor photographed or recorded in any way, without the prior written consent of
IDS GeoRadar.
4.1 System features ............................................................................ 12 7.1.5 How to regulate the Levelling Kit ......................................... 47
4.2 System operating principles ......................................................... 13 7.1.6 How to install the HYDRA on the Levelling Kit...................... 48
4.2.1 Radar..................................................................................... 13
8 HOW TO CONNECT AN EXTERNAL PC (REMOTE DESKTOP) TO THE
4.2.2 FMCW (Frequency-Modulated Continuous-Wave) .............. 13
HYDRA PC ...................................................................................................... 49
4.2.3 Interferometry ...................................................................... 13
9 HOW TO DISASSEMBLE A HYDRA SYSTEM ........................................... 51
4.2.4 ArcSAR Techniques ............................................................... 13
9.1 Dismantling the measurement site .............................................. 51
4.3 System composition ..................................................................... 13 10 GENERAL DESCRIPTION OF THE SW INTERFACE ................................... 52
5 SYSTEM BREAKDOWN .......................................................................... 15
10.1 Wizard interface ........................................................................... 52
5.1 How system is delivered ............................................................... 15 10.2 Settings interface .......................................................................... 53
5.2 HYDRA Radar Sensor .................................................................... 18 10.3 Session interface ........................................................................... 54
5.3 Antennas ....................................................................................... 19 11 HOW TO GET STARTED ......................................................................... 55
5.4 Positioning Unit ............................................................................ 20
11.1 Installation Procedure .................................................................. 55 17.2 Procedure to edit radar position .................................................. 74
11.2 Suggested Preliminary Operations on Settings ............................ 57 17.2.1 Edit Hydra-X position from the Session Wizard .................... 74
11.2.1 General Settings ................................................................... 57
17.2.2 Edit Hydra-X position from an Active Session ...................... 74
11.2.2 Session Settings .................................................................... 57
18 HOW TO MANAGE WARNINGS AND ALARMS ...................................... 75
11.2.3 System Settings..................................................................... 58
18.1 Description of the function ........................................................... 77
11.2.4 Data management ................................................................ 58 18.2 Procedure to hide the pop-up ...................................................... 77
18.3 Procedure to check the system error ........................................... 78
Guardian operating mode ............................................................................ 59
18.4 Procedure to export the system log ............................................. 78
11.2.5 Georeferencing ..................................................................... 60 19 HOW TO CHECK SYSTEM CURRENT CONSUMPTION ............................ 79
12 HOW TO SET UP GEOREFERENCING PARAMETERS .............................. 61 19.1 Description of the function ........................................................... 79
19.2 Procedure to verify current consumption .................................... 79
12.1 Description of the function........................................................... 61
20 HOW TO CHECK SYSTEM COMPONENTS CONNECTION STATUS.......... 80
12.2 Procedure to load the local coordinate projection ...................... 61
12.3 Procedure to load the DTM from Settings ................................... 62 20.1 Description of the function ........................................................... 80
12.4 Procedure to load the DTM from Session Wizard ........................ 63 20.2 Procedure to test Hydra Unit connection..................................... 80
13 HOW TO SET UP SESSION OPTIONS...................................................... 63 20.3 Procedure to test Supply Control Unit connection ...................... 81
20.4 Procedure to visualize System components status ...................... 82
13.1 Description of the function........................................................... 65
20.5 Procedure to Switch On/Off system components ........................ 82
13.2 Procedure to set up Session root folder ....................................... 65
21 HOW TO CHECK THE CAMERA FRAMING ............................................. 83
13.3 Procedure to edit default session name ...................................... 66
14 HOW TO SET UP DATA MANAGEMENT ................................................ 67 22 HOW TO CHECK SESSION STATUS......................................................... 84
14.1 Description of the function........................................................... 67 23 HOW TO CHECK THE HYDRA-X COVERAGE ........................................... 86
14.2 Procedure to set up Guardian location ........................................ 67
24 HOW TO CHECK HYDRA LICENSES ........................................................ 86
14.3 Procedure to set up data transfer ................................................ 67
15 HOW TO RUN A MONITORING SESSION............................................... 67 25 CARE AND TRANSPORT ......................................................................... 87
15.1 Description of the function........................................................... 67 25.1 Cleaning Information .................................................................... 87
15.2 Procedure to run a Session with Quick Start Wizard ................... 68 25.2 Proper use of the system .............................................................. 87
15.3 Procedure to run a Session with Quick Start Wizard disabled ..... 70 26 ERROR MESSAGE AND TROUBLESHOOTING......................................... 88
16 HOW TO LOAD A SESSION .................................................................... 73
APPENDIX A - Technical Specification........................................................... 89
16.1 Description of the function........................................................... 73
APPENDIX B – Disclaimer .............................................................................. 91
16.2 Procedure to load a Session ......................................................... 73
17 HOW TO EDIT THE HYDRA POSITION.................................................... 73 PPENDIX C - Radio Frequency exposure compliance.................................... 94
17.1 Description of the function........................................................... 73 APPENDIX D – Customer Support ................................................................. 94
1 INTRODUCTION improve safety in open pits, strip coal mines, tailing dams,
and industrial minerals operations.
This document describes the HYDRA-X system (Hyper Definition Radar), the
tactical and compact safety monitoring radar provided by IDS GeoRadar. The
1.3 Authorization for use – national restriction
system is basically composed by the radar (and all its equipment) and the
HYDRA-X Controller software that manages the radar system. HYDRA-X The intended reader of this manual should be the technician in charge of
Controller allows to set up and start a Monitoring Session in a very easy and using the system that has undergone the IDS GeoRadar training for HYDRA-
quick way. The direct access to the Guardian software permits also to have X.
the on-site processing and alarming.
Authorization for use – national restriction
This document refers to the concepts the user should learn before initiating
the utilization of this device. Therefore, we recommend reading the entire The use of HYDRA-X system is subject to authorization by the Competent
document before starting to work with the system. Ministry of the country where the system will be used.
1.1 Purpose
This manual gets you up and running quickly with HYDRA-X system. It is a
collection of how-to chapters. Each chapter contains
the basic understanding you will need and a list of step-by-step actions
to follow. Almost everything you need to know to:
• Install
• Configure
• Use
• Maintain
the system is covered by this manual. For more details with reference to the restriction, please refer to the
following website:
1.2 Application field
https://www.efis.dk/views2/search-general.jsp
The main audience of Hydra-X is anyone who needs to set up and start a
monitoring session with Hydra-X system that is designed to
Tip text
• Quarries Person responsible for the product: the person responsible for the product
has the following duties:
3.3 Reasonably Foreseeable Misuse
The following actions are considered misuse of the system: • To understand the safety instructions on the product and the
instructions in the user manual.
• Use of the product without instruction. • To ensure that it is used in accordance with the instructions.
• Use outside of the intended use and limits. • To be familiar with local regulations relating to safety and
accident prevention.
• Opening the product using tools, for example screwdriver, unless
• To inform IDS GeoRadar s.r.l. immediately if the product and
this is permitted for certain functions. the application becomes unsafe.
• Modification or conversion of the product. • To ensure that the national laws, regulations and conditions for
• Use after misappropriation. the operation of electromagnetics transmitters are respected.
3.6 Hazard of use NOTICE: If the accessories used with the product are not
properly secured and the product is subjected to mechanical
NOTICE: Watch out for erroneous measurement results if the shock, for example blows or falling, the product may be
product has been dropped or has been misused, modified, damaged.
stored for long periods or transported.
Precautions:
Precautions: When setting-up the product, make sure that the accessories
Periodically carry out test measurements and perform the field are correctly adapted, fitted, secured, and locked in position.
adjustments indicated in the user manual, particularly after Avoid subjecting the product to mechanical stress
the product has been subjected to abnormal use and before
and after of important measurements.
DANGER: If the product is used with accessories, you may
increase the risk of being struck by lightning.
Precautions:
During operations, the user of the product must be fully aware
of the existing dangers.
DANGER: Inadequate securing of the working site can lead to
dangerous situations, for example in traffic, on building sites,
and at industrial installations.
DANGER: Because of the risk of electrocution, it is dangerous
to use poles and extensions near electrical installations such as Precautions:
power cables or electrical railways. Always ensure that the working site is adequately secured.
Adhere to the regulations governing safety and accident
Precautions: prevention and road traffic.
Keep at a safe distance from electrical installations. If it is
essential to work in this environment, first contact the safety
authorities responsible for the electrical installations and
follow their instructions.
Moving parts. Be aware of the risk of collision with the Moving parts. The movement of the Acquisition Unit can cause
Acquisition Unit during its movement. loss of stability.
Precautions: Precautions:
Do not stand inside the Area of Operation during the It is suggested to screw the tripod to the ground. An
functioning of the system. An emergency stop button is emergency stop button is installed on the Power Supply Unit
installed on the Power Supply Unit to immediately stop the to immediately stop the movement in case of hazard (Fig. 2).
movement in case of hazard (Fig. 1).
NOTICE: If the product is improperly disposed of, the following During dynamic applications, for example stakeout procedures
can happen: there is a danger of accidents occurring if the user does not pay
• If polymer parts are burnt, poisonous gases are produced attention to the environmental conditions around, for
which may impair health. example obstacles, excavations or traffic.
• If batteries are damaged or are heated strongly, they can
explode and cause poisoning, burning, corrosion or
environmental contamination. HYDRA emits non-ionizing radiations that can cause
• By disposing of the product irresponsibly you may enable
interference with implanted electrical or ferromagnetic
unauthorized persons to use it in contravention of the
devices (such as a pacemaker).
regulations, exposing themselves and third parties to the risk of
severe injury and rendering the environment liable to
contamination.
CAUTION: High mechanical stress, high ambient temperatures
or immersion into fluids can cause leakage, fire or explosions
of the batteries.
RECYCLING
Precautions:
The crossed out wheeled bin symbol shown on the equipment Protect the batteries from mechanical influences and high
indicates that the product must be recycled separately from ambient temperatures.
other waste at the end of its useful life.
Separate waste disposal of this product at the end of its useful Do not drop or immerse batteries into fluids
life will be organised and managed by IDS GeoRadar. When you
decide to dispose of the equipment, contact IDS GeoRadar and
follow the system that IDS GeoRadar has set up to permit the
separate collection of the apparatus at its life end.
Adequate separate collection for its subsequent recycling, NOTICE: Only IDS GeoRadar authorized technical service are
treatment and environmental friendly disposal contribute entitled to repair this product.
towards avoiding any unnecessary effects on the environment
and to health and favour the reuse or recycling of the materials
that make up the equipment. Unauthorised disposal of this
product as unsorted waste by its possessor will lead to an
administrative penalty foreseen by national regulations.
DANGER: During the transport, shipping or disposal of Always remove the system from the monitoring area, when
batteries it is possible for inappropriate mechanical influences any hardware defect or measuring error occur.
to constitute a fire hazard.
Precautions:
Before shipping the product or disposing of it, discharge the
batteries by running the product until they are flat.
When transporting or shipping batteries, the person in charge
of the product must ensure that the applicable national and
international rules and regulations are observed. Before
transportation or shipping contact your local passenger or
freight transport company.
4 HW GENERAL DESCRIPTION • Permits the possibility to manage separate campaign thanks to the
repositioning kit (optional).
The HYDRA system (Fig. 3) is designed to provide real time displacement
information with an accuracy of a tenth of a millimeter. The HYDRA system
is particularly suitable for a range of applications depending on the
configuration, and the HYDRA-X is especially suitable to detect and measure
both slow and fast movements in different environments, like for example
Quarries, Landslides and Cut-slopes.
For the system complete operation, the following components, which are
not provided by IDS GeoRadar, are also required:
5 SYSTEM BREAKDOWN • Supply & Control Unit: it is a stand-alone object (Fig. 6).
This section of the manual provides a detailed description of the components
making up HYDRA system.
Fig. 5 – Delivered packaging: (a) Power Supply Unit (b) 3 trolley bags.
Fig. 8 – Trolley bag A including: Tablet (1), hydra sensor (2), levelling kit (3), antenna kit 101
and 102 (4), camera (6), and siren (7).
Fig. 7 – Trolley bag A including: Tablet (1), hydra sensor (2), levelling kit (3), antenna kit 101
and 102 (4), soft bag with cables (5). Fig. 9 – Soft bag with cables: 5 in Fig. 7
• Trolley Bag B (Fig. 10): • Trolley Bag C: tripod suitcase (Fig. 11)
1. Pan tilt unit (see Par. 5.3)
2. Camera PTZ (see Par. 5.7);
3. GPS unit (see Par.5.13 )
4. Laser kit (optional, see Par. 5.9)
Fig. 10 - Trolley bag B including: Pan Tilt unit (1), camera (2), GPS kit (3), Laser kit (4).
-3 dB 16° 16°
-10 dB 29° 27°
Tab. 1 – Main lobes width of HYDRA-ANT101 antennas at -3 dB and -10 dB
-3 dB 27° 32°
-10 dB 53° 55°
Tab. 2 – Main lobes width of HYDRA-ANT102 antennas at -3 dB and -10 dB (optional)
5.3 Antennas
The HYDRA system is provided with a couple of HYDRA-ANT 101 (Fig. 17)
and/or a couple of HYDRA-ANT102 (optional, Fig. 16) antennas (depending
on the configuration) operating in horizontal polarization and characterized
by a maximum gain of 21 dBi (for ANT 101) or 16 dBi (for ANT 102). The Fig. 16 – HYDRA-ANT 102 (optional)
amplitude characteristics of the antenna main lobe at -3 dB and -10 dB are
provided in Tab. 2 and Tab. 2.
The Radar Sensor is fitted with two waveguides, each with four threaded
holes, for the installation of the pair of antennas (Fig. 13).
HYDRA antennas are protected by a Radome protective cover (Fig. 12). Fig. 18 – Pan-Tilt Unit
5.4 Positioning Unit The elevation pointing system is fixed on the Arc SAR arm through 2 cam
levers and 2 knobs. The cam levers are the rotation fulcrums of the pointing
The Positioning Unit (Fig. 18) consists of: system, while the knobs permit to select different Radar Sensor bearings,
between -30° and +30°.
• The Pan-Tilt main body, made in aluminum and plastic, with
dimensions 32x20x21 cm and weight <10 kg, capable of moving the
Radar Sensor in azimuth and elevation;
• A Triggering Magnetic Sensor to synchronize the Pan-Tilt rotation and
the radar acquisition;
• The Arc SAR arm 50 cm long holding the payload items;
On the Elevation Pointing System there are 3 fixing points to hold the sensor
Fig. 19 – Elevation Pointing System
in place, two of them are composed by a pull-tab, while the third one uses a
Two screw holes under the elevation pointing system permit the installation knob (see Fig. 15).
of an IR Laser (optional) under the Elevation Pointing System (see Fig. 20).
The Radar Sensor must be connected to the PTU with the Sensor Cable,
attached to the 19-pin connector (Fig. 39).
The PTU must be connected to the Supply & Control Unit (SCU) with the Main
Cable, attached to the 32-pin connector (Fig. 21):
Fig. 21 – 32-pin connector for the Main Cable Fig. 22 – Magnetic triggering sensor
The PTU is used to move the Radar Sensor in azimuth for the radar
Pinch point. Beware of the risk of finger pinch between the
monitoring. During the scan the PTU moves in azimuth at a fixed elevation
magnetic sensor and the magnetic trigger during rotation of
from -60° to +60° clockwise, and then returns to -60° (counterclockwise). the system (Fig. 22).
5.5 Tripod
A tripod is provided for the installation of the PTU, with an interface
that allows easy fastening (Fig. 23).
Fig. 23 – Tripod
Fig. 24 – Power Supply Unit
5.6 Supply & Control Unit (SCU)
The SCU has 7 interfaces (Fig. 25), from left to right :
The Supply & Control Unit (or SCU) supplies electric power to the system. Its
function is to: • Tablet Power;
• IR Power;
• distribute power and signals to the PTU, Radar Sensor and other
peripherals included in the system; • Siren Alarm;
• receive power from mains supply and charge the pair of batteries • Power Input (supports 100-240 V AC or 12-24 V DC);
required to guarantee a continuous power supply for 2 hours after a • WiFi antenna;
blackout event;
• PTU Main Cable;
The SCU is a box with dimensions of 600 x 400 x 385 mm and a weight of
25 kg, with 2 handles plus an additional pull out handle, two wheels and • Local Network;
fasteners fitted with plug locks (Fig. 24). • Camera Ethernet cable.
The SCU has an emergency stop button accessible on its external surface (Fig.
26) that, if pressed, immediately stops the movement of the PTU. Once
pressed, the button switches off the supply of the PTU.
Rotate the button in order to rearm it and switch the PTU on again and
restart the acquisition.
Fig. 26– Emergency Stop Button
Opening the SCU gives access to 3 buttons:
• ON button (Fig. 27);
• OFF button (Fig. 27);
• Alarm Mute button (Fig. 28).
The SCU has inside also the HYDRA computer (Fig. 29): that is an industrial
PC supplied with the following software:
IDS GeoRadar takes no responsibility for bad functioning if there is a As optional a 4 MP IR Fixed Dome Network Camera (Fig. 32) could be
functional conflict between its software and any software installed provided.
on the HYDRA computer by the user. IDS GeoRadar does not The 4 MP IR Fixed Dome Network Camera (optional) is a dome camera with
guarantee that the performance of its equipment will be maintained integrated IR illuminator and auto switch between day and night vision. Both
using a different configuration from the recommended one types of camera are installed on the fixed part of the Positioning Unit, on the
plate of connection between PTU and tripod.
Fig. 32 – 4 MP IR Fixed Dome Network Camera Fig. 33 – Radar map and Camera display for Hydra- X
For the 4 MP IR Fixed Dome Network Camera there is also an elevation Both models of the Camera are connected to the SCU though an Ethernet
pointing system that permits the orientation towards the area to be cable.
observed. An inclinometer is stuck on the elevation pointing system to
indicate camera bearing.
5.9 Laser (optional)
The Laser (Fig. 34), is mounted on the elevation pointing system, below the
Radar Sensor and, when moved by the PTU, act as a laser scanner, enabling
The Camera, integrated into a Hydra installation, provides several additional the surface reconstruction of the monitored scenario.
features:
The laser has good measuring performance on a rocky surface up to 50 m
• it is used to correlate the picture scenario with a given Guardian radar far.
map, thus enabling the visual recognition of moving areas (Fig. 33);
The Laser is connected directly to the Radar Sensor.
• it can stream a real time video of the areas of interest.
• Power Cable Adapter (Fig. 38), this cable is different based on the
country.
• Sensor Cable (Fig. 39);
• Local Network Cable (Fig. 40);
• DC Power Cable (Fig. 41);
• Tablet Battery Cable (Fig. 42).
The PTU Main Cable (Fig. 35) is used to connect the Pan-Tilt to the SCU, it
provides power to the PTU and it exchange data between the HYDRA
computer embedded in the SCU and the PTU.
Fig. 34 – Laser
The laser is rated Class 1M. Do not look at the laser with
magnifying optics such as microscopes and telescopes.
5.10 Cables
The Hydra system is provided with a Cable Kit composed by:
The Camera Ethernet Cable (Fig. 36) is used to connect the Camera to the The AC Power Cable (Fig. 37) is used to connect the SCU to the main power
SCU; it provides power to the Camera and it exchange data between the line; it can be used in conjunction with the Power Cable Adapter (Fig. 38).
Camera and the HYDRA computer embedded in the SCU.
The Sensor Cable (Fig. 39) is used to connect the PTU to the Sensor. The
curved end needs to be connected to the Sensor (Fig. 14).
The Local Network Cable (Fig. 40) is an Ethernet cable used to connect the The DC Power Cable (Fig. 41) is used to connect the SCU to a direct current
SCU to an external network. power source (12-24V).
The Tablet Battery Cable (Fig. 42) is used to recharge the Tablet through the 5.11 Siren Alarm (optional)
SCU.
A Siren Alarm (Fig. 43) can be connected to the SCU to provide Early Warning
based on the detected movements in the scanned area. It provides both a
visual and auditory alarm.
The Mounting kit (Fig. 44) provides a stable measurement platform, that 1. Ground Mount Kit (Fig. 45)
allows to acquire campaign of data with an installation precision of 0.5mm. 2. Ground Mount Kit screws (1
The complete Mounting kit is composed by two main parts: the Ground
3. Fig. 46)
mount kit and Levelling kit.
4. Allen key (2
Ground mount kit and levelling kit are two separate articles. 5. Fig. 46)
The Levelling kit can be used also fixed on a concrete block or
any stable surface, while for the use of the Ground mount kit 6. Levelling Kit screws (3-4
the installation of levelling kit on top of it, is mandatory. 7. Fig. 46)
8. Levelling Kit base support (5
9. Fig. 46)
10. Levelling Kit (6
11. Fig. 46)
The Mounting Kit can be used with any Hydra configuration and it is
composed by the following parts (Fig. 45;
This is a good solution when a DTM is provided and the operator needs to
know the position and orientation of the Hydra to properly visualize the
illuminated area on the DTM.
The supply GPS cable must be connected to the radar sensor, as shown in
the pictures below (Fig. 48; Fig. 49)
Fig. 46 - Mount Kits components: 1 - Ground Mount Kit screws; 2 - Allen Key; 3-4 –
Levelling Kit Base Support; 6 - Levelling Kit.
6 HOW TO INSTALL THE HYDRA-X SYSTEM Below some considerations when choosing an installation site:
The installation of the HYDRA-X system requires about 15 minutes of work. • Installation site should have a clear view of the area to be monitored.
The procedure can be broken down into following parts: Shadowed areas or other areas that cannot be seen will not be present
in the data;
1. preliminary operations;
• Installation site should be a place with enough room to permit free
2. installation of the tripod; movement of the PTU both in pan and tilt, avoiding collisions of the
3. installation of the positioning unit on the tripod; Arc-SAR metallic arm with other obstacles;
4. installation of the HYDRA Radar Sensor on the positioning unit; • HYDRA should not be placed at a distance less than 2 m from the area
of interest for the monitoring, since targets nearer than 2 m will not
5. installation of the Camera;
be visible;
6. installation of the SCU.
• HYDRA should be placed far enough away from the wall so that it can
The procedure for the disassembling of the equipment from the monitor the desired scenario. The system will monitor a cone shaped
measurement site is given in Section 9 after the description of the area extending from the installation point. Refer to Par. 4.1 for the
abovementioned six steps. maximum distance and beam width of the various configurations;
6.1 Preliminary operations • HYDRA measures the line of sight displacement. This means that it
measures the displacement that occurs along a line that extends from
These are the preliminary operations suggested before the actual installation the radar to the monitored surface. Movements which are
of a HYDRA system: perpendicular to the radar’s line of sight can’t be seen. For this reason,
avoid monitoring walls which are parallel to the radar’s line of sight;
• choose installation site;
• Avoid area with vegetation;
• verify in advance that the contents of the transport cases were
properly loaded to avoid any damage during transportation.
• arrange a suitable means of transportation, capable of containing all
the needed material.
6.2 HYDRA Installation procedure 2. Level the Tripod using the three legs and checking bubble level for a
correct installation (Fig. 51). It is suggested to screw the tripod to the
1. Pull out the Tripod from the transport bag and place it in the ground using the special holes on the tripod’s feet.
installation point (Fig. 50);
3. Pull out the PTU from the transport bag; 5. Rotate the Positioning Unit on the Tripod until it is possible to access the
fixing screw in the Tripod’s cylindrical housing;
4. Mount the PTU on the Tripod by accommodating the metallic cylinder
bottom of the PTU) in the Tripod’s cylindrical housing (Fig. 52); 6. Use the hex key with the red handle, provide with the Tripod, to screw
on the positioning Unit to the Tripod (Fig. 53);
7. Pull out the Camera (already mounted on its mechanical holding system)
from the transport bag;
8. Fix the Camera to the metallic plate of the PTU using the two knobs (Fig. 11. Mount the Radar Sensor on the PTU using the knob and the two fixing
54); levers (Fig. 55);
9. Use the Camera lateral knobs to point the camera at 0° (only for the
model 4 MP WDR Dome Network Camera with IR);
12. Use the 2 quarter turn screws to point the Radar Sensor at 0° (Fig. 56); 13. Connect the Sensor Cable, the Triggering Sensor Cable, and the
GPS/Laser Cable, to the back of the Radar Sensor (Fig. 57);
Check the gap between the magnetic sensor and the magnetic trigger;
the gap needs to be 0.1÷3mm for correct functioning
14. Place the SCU in a convenient location around 1.5m-2m from the Tripod; 16. Connect Camera Ethernet Cable to the SCU and to the Camera (Fig. 59);
15. Connect the PTU main cable to the Pan-Tilt and to the Power Supply Unit
(Fig. 58);
20. Switch on the tablet or PC and then connect to the Hydra PC, as
Fig. 60 – Connect the Wi-Fi Antenna described in Section 8.
18. Connect the AC Power Cable to the SCU and to the main power line,
eventually using the Power Cable Adapter (Fig. 37 and Fig. 38);
19. Open the SCU and switch on the system pushing and holding for 2
seconds the ON button inside the SCU (Fig. 61);
Fig. 65 – Levelling Kit components, in red square Levelling Kit Support Base screws
Fig. 67 – Levelling Kit support base fixing operation, from the bottom view
Fig. 69– Levelling Kit components, levelling kit screws in red square
Follow the procedure described in Par. 6 to install the HYDRA Radar Sensor,
Fig. 74 – PTU insertion
Camera and SCU.
8 HOW TO CONNECT AN EXTERNAL PC (REMOTE 2. In the new window, click on CHANGE ADAPTER OPTIONS (Fig. 77, STEP
3)
DESKTOP) TO THE HYDRA PC
In order to remote the Hydra PC, every tablet or PC with a WI-FI connection
available can be used as a remote desktop.
Follow the steps below for the access to the HYDRA PC:
1. Click on the icon of the connection on the bottom left of the screen
(STEP1), then click Network and Internet settings (STEP2), as in Fig. 76 :
2
1 Fig. 77- Status window, step3
3. In the Network connection list, right click on wi-fi connection and select
WIFI “properties” (Fig. 78, STEP 4).
Fig. 76 – Network view, step 1 and 2. 4. Click on INTERNET PROTOCOL VERSION 4 (TCP/iPv4) and select
PROPERTIES in the tab below (Fig. 78, STEP 5).
The user is now connected with the HYDRA computer inside the SCU. Run
7 the software HYDRA Controller (see chapters from 11 to 27).
10 GENERAL DESCRIPTION OF THE SW INTERFACE The following paragraphs briefly describe the main three sections of the SW:
The Hydra Controller home page is visualized as in the following picture: • Wizard
• Settings
• Session
At the end of the Session Wizard, Hydra Controller directly opens the related
Fig. 80 – Hydra Controller home page.
Guardian project (if Guardian runs “on-site”, as selectable in the settings
From the main page, the user can access the following features: section) in order to allow the set-up of alarms and the visualization of the
monitoring results on-site.
• Set up and start a Monitoring Session through a guided Wizard
The Session Wizard is available in two operating modes:
• Load a previous Session and proceed with the Monitoring
• Quick Start mode OFF (Fig. 81)
• Set up the Hydra Controller settings • Quick Start mode ON (Fig. 82)
By default, the Quick Start mode is enabled (that is it’s set to ON), skipping
• Get information about the Hydra-X system the visualization of the Georeferencing page of the Wizard (the
georeferencing is automatically done, see Section 15.2). The Quick Start
mode can be disabled through a dedicated button in the first step of the
• Minimize to icon Wizard.
Note that during a Session Wizard you can adopt, for that
particular Session, a different set of configuration parameters
Fig. 81 - Wizard: quick start mode OFF.
(i.e. Session name, Session root folder and DTM).
The Settings interface is always accessible both during a running session and
when the session is stopped/paused.
The supported languages are: The interface is composed of 3 main parts (Fig. 84):
a. Run CLM d. Verify that all the needed licenses are enabled in the table. If
the licenses are correct click on Activate new licenses
b. Click on Activate new licenses
TIME FORMAT 24 H 24 H, 12 H
• Session Root Folder: consists in the path of the directory where all
the session's data are saved. The default root folder is
D:\HYDRA_SESSION.
For each item, if the connection test is successful, a green flag will
appear, while failing makes a red cross appearing instead.
Data Management ➢ On Site: the data backup will operate on GBD files only since
PSV and JPG files are moved within the Guardian project
If the Data Management switch is enabled the user can choose folder
➢ On Remote Computer: the data backup will operate on PSV,
between two options, Archive or Backup.
GBD and JPG files.
11.2.5 Georeferencing
From the Georeferencing tab it is possible to access to the following panels:
Fig. 95 – Data management mode If the EPSG code/WKT or DC file is not loaded before
entering the Wizard, the GPS data will be not converted
• Archive (move): If this option is selected the data are moved from in local coordinates and the georeferencing step is not
the Hydra Controller session folder to the selected archive folder. completed
According to the Guardian Operating mode different data types
can be archived: • Local Coordinate DTM: to load the DTM in local coordinates.
➢ On Site: the data archive will operate on GBD files only since Hydra Controller accepts two different file formats:
PSV and JPG files are moved to the Guardian project folder ➢ .dxf
➢ On Remote Computer: the data archive will operate on PSV, ➢ .txt
GBD and JPG files.
both in International and U.S. customary unit system.
12 HOW TO SET UP GEOREFERENCING PARAMETERS 3. To make the georeferencing working, it is mandatory to insert the
local coordinate projection method that converts the geographic
12.1 Description of the function coordinates (WGS84) taken from the embedded GPS into the local
To assure the correct georeferencing, the DTM and the radar position must one (XYZ). Click on the icon Local Coordinate Projection
be expressed in the same coordinates reference system. The DTM is defined
in the local coordinates (XYZ), the embedded GPS instead in the geographical
coordinates (according to WGS84 standard, i.e. latitude, longitude and
elevation). The following procedure describes the steps to set up the
parameters to project GPS WGS84 (latitude, longitude and elevation)
coordinates into XYZ local coordinates and to load the DTM.
4. Select from the field Projection Info the file format for the
projection if it is different from the EPSG code (EPSG code is the
5. Browse the DTM file clicking on the Select File field (Fig. 101);
1. Click on the Load DTM icon, the Load DTM window will appear (Fig.
101)
Fig. 104 – Preview of the DTM Acquired by Hydra 2. Click on the Load DTM icon and browse the DTM file clicking on the
Select File field;
The coordinates of the DTM will be in WGS84 in case a conversion file has
not been loaded (see paragraph 12.2 for reference). If instead a conversion
file has been loaded the acquired DTM will be in local coordinates. The supported DTM formats are: .txt and .dxf
The new DTM will not be automatically Click the OK button in the Local Coordinate DTM window (Fig. 100) to
transferred to the Guardian project. confirm the DTM loading.
12.7 Procedure to acquire the DTM from an active session The following paragraphs describe how to change the session root folder
path and how to edit the session name.
1. From the main session interface (Fig. 84) click on the DTM tab. A new
window will appear (Fig. 119). 13.2 Procedure to set up Session root folder
The Root Folder path cannot be changed while a session is opened, it can be
2. Click on the Acquire DTM icon to start the surface reconstruction.
edited only from the Hydra Controller home page (Fig. 80):
This operation lasts 5 minutes (Fig. 102) .
1. Click on the geared icon on the top right side of the window
3. The Preview of the acquired DTM will appear, click on Apply to load
it within the Controller session (Fig. 104). 2. Open the Session tab (Fig. 106).
Click the OK button in the Local Coordinate DTM window (Fig. 100) to
confirm the DTM loading.
Only those sessions located under the new path will be listed
in the Load Session list. Therefore, in case a session folder is
located under a different path with respect to the one set in
the Session Root Folder field, it is necessary to cut and paste it
under the new Session Root Folder path.
14.1 Description of the function ➢ On Site: the data archive will operate on GBD files only since
PSV and JPG files are moved within the Guardian project
The current paragraph shows how to manage the data collected by Hydra-X. folder
➢ On Remote Computer: the data archive will operate on PSV,
Hydra Controller can acquire new data only if there is enough space on the GBD and JPG files.
disk where the session folder is located.
• Backup (copy): If this option is selected the data will be copied
As soon as the residual space on the disk is below 100 GB, from the Hydra Controller session folder on the selected backup
Hydra Controller starts to delete old files: GBD, PSV and JPEG folder. According to the Guardian Operating mode different data
in order to free up space and let new data to get stored. Once type can be backed up:
the residual space reaches the threshold of 10 GB the
acquisition is stopped and Hydra cannot perform any further ➢ On Site: the data backup will operate on GBD files only since
acquisition. PSV and JPG files are moved within the Guardian project
folder
➢ On Remote Computer: the data backup will operate on PSV,
According to the Guardian operating mode the data can be sent from GBD and JPG files.
Controller to Guardian in two different mode:
14.2 Procedure to set up Guardian location
1. On Site: Guardian running on the Hydra industrial PC. Guardian get
The procedure is described in the paragraph 11.2.4.
data directly from the Hydra Controller session folder. In this case
data will be transferred through FTP protocol. 14.3 Procedure to set up data transfer
2. On Remote Computer: Guardian running on a Remote PC. TCP or The procedure is described in the paragraph 11.2.4.
FTP protocol can be set.
If the Data Management tab is enabled the user can set up the data 15 HOW TO RUN A MONITORING SESSION
archive option:
15.1 Description of the function
• Archive (move): If this option is selected the data are moved from As described in the paragraph Wizard interface, the Hydra Controller Wizard
the Hydra Controller session folder to the selected archive folder. allows a rapid deployment of the system for critical monitoring applications.
Prior to start a Session Wizard, close the Guardian software, HYDRA Controller assigns a project name in the format
in case Guardian is running on site. If a Guardian project is SystemName_YYYYMMDD in case the field “Default Session Name”
opened, a new project will be not automatically opened at was enabled during the Preliminary Operations phase (Paragraph
the end of the session wizard.
11.2). The maximum length of the Session Name allowed is 24
It is recommended to close Guardian even if it is running on
characters. No special characters are allowed.
a remote PC in order not to mix-up data coming from
different sessions.
A description can be added and can be reviewed during the
reloading of the session.
The Quick Start wizard mode is enabled by default. It skips the visualization
of the georeferencing step (Fig. 82) of the wizard and it automatically.
By default the Wizard has the Quick Start mode ON
When the Quick Start mode is not enabled (Fig. 81), the wizard goes through
all the configuration steps.
From the Home page (Fig. 80) click on the icon New Session to
create a new monitoring session;
Fig. 107 – Session name window Clicking on the Edit icon two new icons will appear (Fig. 109):
Click on the bottom right arrow to proceed to the next Wizard step.
2. System Bearing: the camera streaming shows the area that Hydra is
able to monitor. You can adjust the vertical bearing/field of view of
Hydra by clicking on the up/down buttons in the range of +/-30°. The
horizontal orientation must be done manually, by adjusting the
tripod position.
If the camera is not recognized, or is not working properly, a Fig. 109 – Edit mask icons
new window will appear allowing the user to insert manually
the camera tilt. Modifications of the radar mask can affect
the reliability of the monitoring session, only
expert users should go through this
procedure.
3. Coverage: the Coverage section shows the amplitude map of the
monitored area draped on the DTM, in case a DTM has been
Forced Area: Click on Forced Area in case an area needs to be forced
loaded, or on a flat DTM, in case the DTM has not been loaded.
into the radar mask, once the icon is pressed the user can create the
Excluded Area: Click on Excluded Area in case an area needs to be If the Radar Sensor or the Pan Tilt Unit are disconnected,
excluded from the radar mask, once the icon is pressed the user can it is not possible to create a new session
create the polygon of the excluded area. Once completed click again
on the Excluded Area icon to save the area. 1. Session Name: fill the name in the field. A new session folder with
the selected name will be created in the session root folder.
Click on the right button arrow, the pop-up shown in Fig. 110 will
appear, click Yes to start the monitoring session. HYDRA Controller assigns a project name in the format
SystemName_YYYYMMDD in case the field Default Session Name
was enabled during the Preliminary Operations phase (Paragraph
11.2). The maximum length of the Session Name allowed is 24
characters. No special characters are allowed.
Fig. 110 - Session Start Pop-up By default the Wizard has the Quick Start mode ON
In case the Guardian operation mode is set on On Site mode (see the
paragraph Procedure to set up Guardian location), a new Guardian
project will be automatically created under the Guardian project 2. System Bearing: the camera streaming shows the area that Hydra is
folder, the project name is the same of Hydra session. The Guardian able to monitor. You can adjust the vertical bearing/field of view of
project will be automatically opened. Hydra by clicking on the up/down buttons in the range of +/-30°.
The horizontal orientation can be done manually adjusting the tripod
position. Once the Tilt has been adjusted the PTU moves to the
15.3 Procedure to run a Session with Quick Start Wizard extreme left position to acquire WGS84 position and orientation by
disabled means of the embedded GPS (Fig. 111).
Clicking on the right bottom arrow Hydra will perform 4 scans used
to create the Selection Mask. During this phase, Controller software
will automatically select all good quality pixels. The pixels discarded
are not processed, although they are still present in all raw data files.
Clicking on the Edit icon two new icons will appear (Fig. 109):
2. Click on the Edit Position button located on the top right position
3. Edit the Hydra position manually or click on Get GPS to acquire the
position using the embedded GPS (Fig. 118). More satellites are
Fig. 118 – Edit position panel
detected by the GPS higher will be the number of bars on the GPS
receiver icon. In case the loaded DTM uses Local Coordinates it is
necessary to load a conversion file to get the GPS
coordinates in Local Coordinates . See paragraph
HOW TO SET UP GEOREFERENCING PARAMETERS to
get more details on how to load a conversion file. If a
If the Hydra Position falls outside from the DTM conversion file is not loaded the GPS coordinates will
boundaries, the position can’t be loaded. The be displayed in WGS84 and the georeferencing step
message Position out of DTM bounds will appears. is not completed.
3. Click on the Edit Position icon, the panel shown in Fig. 118 will open. 18.1.1 Edit the radar mask from the session wizard (Both for Quick start
From here it is possible to edit the position manually or clicking on On and Quick Start Off mode wizard)
the Get GPS icon to acquire the position using the embedded GPS. 1. Follow the session wizard until the Coverage window appears, from
this window it is possible to edit the radar mask.
The coordinates shown in the Edit Position window
opened from an active session can have slightly 2. Click on the Edit button located at the top right position, two new
different values compared to the ones acquired icons will appear (Fig. 109).
during the session wizard. The reason is that the GPS
coordinates are averaged on the last 24 hours of data
acquired by the embedded GPS. So the position will
be more accurate.
Fig. 121 – Edit Mask: Excluded areas are represented in light blue while Forced areas in green
2. Click on the Update icon, Hydra Controller will generate a new mask
file based on the last ten acquired GBD files.
2. Click on Hide.
The mask update can be performed while the acquisition is running. It is
necessary have at least ten GBD files created within the session folder.
20 HOW TO CHECK SYSTEM CURRENT CONSUMPTION 3. Click on the Supply Info icon, the Supply chart will open (Fig. 129).
➢ If the SCU revision is < 6: the Aux chart shows the current
consumption of the Pan Tilt Unit
➢ If the SCU revision is ≥ 6: the AUX chart shows the current
consumption of the Hydra batteries.
These logs are acquired by the Power Control board. At the top of the graph
the buttons Day, Week, Month, All allow the analysis of different time
periods.
• Acquisition Unit: Composed by the Radar sensor, the Pan Tilt and
the Tripod;
2. Click on the geared icon of the Hydra Unit field, a new window will
open (Fig. 132).
PAN TILT: Ensure that the 32 pin cable is well plugged from
the Hydra SCU to the Pan tilt.
Ensure that the white ethernet cable (the one with the label
ETH PTU) of the PTU is well plugged to the Router board.
Ensure that the emergency stop button is fully released.
Read the following tips in case the test connection to one or more devices
fails:
CAMERA: Ensure that the ethernet cable is well plugged to the 3. Click on the Test button to test the connection between the
SCU as well as to the Hydra Camera.
industrial PC and the Supply Control Unit. A green flag will appear in
GPS: Ensure that both 32 and 19 pin cable are well plugged on case the test is successful, otherwise a red cross will appear.
the Hydra SCU as well as on the Hydra acquisition Unit.
Ensure that the connector plugged to the GPS is well fastened.
Green flag means that there’s communication between the Hydra PC and the
relevant device. Red Cross means that there’s no communication between
the Hydra PC and the relevant device.
More details for each device are shown in the right panel.
Clicking on the Export icon, on the bottom right side, it is possible to export
all the devices information in a .txt file.
22 HOW TO CHECK PAN TILT FUNCTIONING PAN TILT: Ensure that the 32 pin cable is well plugged from
the Hydra SCU to the Pan tilt.
22.1 Description of the function Ensure that the white ethernet cable (the one with the
label ETH PTU) of the PTU is well plugged to the Router
The following procedure explains how to check the Pan Tilt functioning.
board.
22.2 Procedure to test Pan Tilt Unit connection Ensure that the emergency stop button is fully released.
1. iTo access to the Hydra Unit Setting tab, click on the top right geared 22.3 Procedure to test Pan Tilt movement
icon and select the System tab.
1. Before to check the Pant Tilt movement it is necessary to stop the
2. Click on the geared icon of the Hydra Unit field, a new window will running acquisition, clicking on the Stop icon located on the bottom
open (Fig. 136). right side.
2. On an active session, open the System tab and open the Pan Tilt
Panel (Fig. 137)
3. Check the connection between the Pan Tilt and the industrial PC by
clicking on the Test button. A green flag will appear in case the test
is successful, otherwise a red cross will appear.
Read the following tips in case the test connection fails: Fig. 137 – Pan-Tilt Control Panel
4. Send the Pan tilt arm to the extreme left and to the extreme right on
the Pan axis. In the Pan [deg] field write -168 and then press Go To,
then write + 168 and press Go To. Ensure that the arm movement is
smooth.
5. Send the Pan tilt arm to the extreme left and to the extreme right in
on the Tilt axis. In the Tilt [deg] field write -16 and then press Go To,
then write + 30 and press Go To. Ensure that the arm movement is
smooth.
2. The scheduling panel will open, click on the + icon to add a new 25 HOW TO CHECK SESSION STATUS
Schedule (Fig. 140). 1. On an active session, select the session tab
3. From the Add Schedule panel it is possible to select schedule date
and time. In the “Name” field it is possible to insert the name for the 2. Open the Info panel (Fig. 141). This panel provides a rough analysis
just created schedule. Click Ok to save it. of the acquisition statistics and useful information on the volume of
4. Click on the bin icon delete a selected Schedule. data created and archived.
5. Click on the pencil icon to change the planned date and time of a
specific Schedule.
6. Click on the Calendar icon to load the Schedules configured in other
sessions.
The aspect of this graph should usually be an almost flat line, 27 HOW TO CHECK HYDRA LICENSES
since the number of acquisitions should be constant if no See paragraph 11.1 for the License activation procedure.
configuration change happens. A sudden drop in the graph may
in fact be due to a hardware failure or a power outage. To check the status of the Hydra Controller License click on the question
Mark located on the top right side of the interface, the License Info panel will
The Data Statistic panel shows the following fields: appear (Fig. 142)
• Radar data per day: Corresponds to the space occupied by
the total number of GBD and PSV files created in one day.
• Total Acquired Data: Size of the Session folder.
• Total video data: Size of the Video folder.
• Total time of acquired data: It corresponds to the total time
of acquired data.
The Hard Disk field gives real time information of the memory
occupation of the disks and estimated autonomy in terms of
time. The following fields are shown:
From this panel is it possible to check the expiration date of the support plan, • Periodically check if the PTU has acquired some play, that is if the
and the type of license currently active on the Hydra Controller software. clearance between different angular positions has increased. You
can test it with the stopped PTU trying gently to move it around its
steady position. The movement allowed should be minimum.
In case the Support Plan is expired it won’t be possible to
upgrade the software to newer versions.
• Lubricate periodically the cylindrical interface between PTU FLIR
and Moog Tripod, use silicone grease or silicone spray lubricant,
see Fig. 52.
28 CARE AND TRANSPORT
• After 5 years of use replace the batteries inside the SCU.
28.1 Cleaning Information
In the present section information for HYDRA system maintenance are
28.2 Proper use of the system
provided. For further details please contact IDS GeoRadar Customer
Care. The following procedure should be followed in order to avoid short-circuiting
and overloads:
All maintenance operations need to be performed with the
system stopped and disconnected from the power source. • The User must not disconnect and connect components and system
cables when the Power Supply Unit is turn on.
PROBLEM SUGGESTION
The SCU does not turn on. The backup batteries could be
low. Remember they ensure
continuous operation only for 2
hours.
The SCU is powered on but If the system is disconnected
some components of the from the main supply, make
system seems to be powered sure to keep the on-button
off (e. g. the pressed for at least 7 seconds in
camera/PTU/radar are off). order to start the SCU switching
on all the peripherals of the
system.
The Wifi network “Hydra Wifi” Verify that the Wifi antenna is
for the remote connection of properly connected to the SCU
the Tablet with HYDRA (see Fig. 60).
computer is not listed among
the Tablet’s Wireless Network
Connection list.
Radar Sensor
Pan-Tilt
Dimensions 32 x 20 x 21 cm
Length of the arm 50 cm
Weight 18 kg
Enclosure Class IP 65
Temperature Range -10 °C ÷ + 55 °C
Tripod
Dimensions 30 x 56.5 x 50 mm
Weight 35 g (excluding cables and
protective box)
Enclosure Class IP 65 (in the protective box)
Temperature Range 0 °C ÷ + 40 °C
Range 0
2-50 m (natural targets)
i. The User/Buyer, for setting-up and using the Products, shall consult the official
documentation provided by IDS GeoRadar s.r.l. for the Products (“Reference
Documentation”) and carefully ascertain the compliance with national laws and
requirements, which may limit or even forbid their use.
ii. For Products which shall operate by circulation in Public Areas/Roads, with or
without moving traffic, Buyer/User shall verify the approval of local authority
and/or site’s owner according to their specific procedures. IDS GeoRadar s.r.l. shall
not be liable for any direct, indirect, special, incidental or consequential damages
or injuries, including without limitation, lost revenues or lost profits, resulting by
un-authorized use of the Products in Public Areas/Roads.
Fig. 143 – Total system footprint
iii. Buyer/User further warrants:
• that these Products are not being used, in the design, development,
production or use of chemical, biological, nuclear ballistic weapons. Buyer/
User will defend, indemnify and hold IDS GeoRadar s.r.l. harmless against any
liability (including attorney’s fees) for non-compliance with the terms of this
APPENDIX B – Disclaimer article.
• That, if IDS GeoRadar s.r.l. requires that Buyer/User shall carry out a training
1. Generals. with reference to some Product categories, no operation or use of the
Products shall be started before its designated Operator/s has got the User
i. The present Disclaimer applies to all products designed, produced and distributed Certificate, as defined by IDS GeoRadar s.r.l. specific procedure which the
by IDS GeoRadar s.r.l., its Subsidiaries, Affiliated and authorized Distributors (the Buyer confirms to know and accept.
“Products”). IDS GeoRadar s.r.l. reserves full ownership and intellectual property
rights of any “Information” contained in this Disclaimer including Trade Marks and iv. For Products which include specific “Operational” software with automatic data
Graphics. No part of this Disclaimer may be used or reproduced in any forms processing and analysis “Tools”, e.g. the IBIS Products and Hydra Products, User
without the prior written agreement of IDS GeoRadar s.r.l. shall be aware that the results provided by these “Tools” may be not error free.
User that completely relies on the outcomes provided by these Tools only, does it
ii. In the event that any provision of this Disclaimer may be invalid, unlawful or at his own risk.
incapable of being enforced by a rule of law, all other provisions shall,
nonetheless, remain in full force and effect. Failure to either enforce or exercise v. In no event IDS GeoRadar s.r.l. shall be liable for special, direct, indirect, incidental,
any right, privilege, or legal remedy at any time, any provision contained in this exemplary, punitive or consequential damages including, but not limited to, loss
Disclaimer, shall not be deemed a waiver of such provisions or right, remedy, or of profits or revenue, caused by the use of the Products, either separately or in
privilege. combination with other products or relied upon the results provided by the above
“Tools”.
iii. This Disclaimer shall be interpreted, governed, construed and enforced in
accordance with the laws of Italy. User/Buyer hereby consents to the exclusive
jurisdiction of Pisa.
i. The User shall follow the instructions provided by IDS GeoRadar s.r.l. in its official
“Reference Documentation” for the Product, in particular the User’s Technical 4. Monitoring Products.
Manual which contains all the specific steps and recommendations for a correct
setting-up and use of the Product.
i. IBIS and Hydra Products (the Monitoring Products”) use a leading-edge technology
ii. In no event IDS GeoRadar s.r.l. shall be liable for special, direct, indirect, incidental,
for monitoring slope movements and has proven as an effective tool for the early
exemplary, punitive or consequential damages including, but not limited to, loss
detection of critical movements, possibly leading to slope failures. Especially in
of profits or revenue, caused by the missed or incomplete observance of the
situations where people and machinery are located or working in proximity of
instructions and prescriptions for the use of the Products, either separately or in
unstable slopes, the technology used by Monitoring Products, in addition to other
combination with other products, including but not limited to the following main
suitable monitoring technologies, maximizes the probability of detecting early
aspects:
signs of slope failures, thus permitting the prompt evacuation of people and assets
at risk.
a. Use of IDS GeoRadar s.r.l. Products outside its limitation of use, without ii. Monitoring Products have been designed and developed to cope with very harsh
proper and adequate scientific/technical knowledge or without specific environments. Should the system may become temporary unavailable due to an
training. unexpected hardware failure, monitoring data will not be available until the
instrument is repaired
b. Use of results/outcomes of the measurements performed by the Product
aimed to safety aspects without using adequate control procedures and iii. Monitoring Products are a monitoring instrument that works like other
assessment by skilled personnel. monitoring instruments and cannot cause or prevent a collapse to happen. As a
monitoring instrument, it helps professionals in providing measurements of
c. Opening of the Equipment (for HW Products) without express written
movements of the monitored area, which according to the geological context can
authorization of IDS GeoRadar s.r.l.;
be interpreted as a sign of a potential collapse. Generally, the radar is also
d. Unauthorized changes and additions to the Products. operated in conjunction with other monitoring instruments to have the largest
information and for complementing pros and cons of each technology. The
e. Use of the Products connected to suspected non-working equipment or with interpretation of data provided by the multiplicity of monitoring sensors has then
equipment (mainly PC) having characteristics not in compliance with the the purpose of detecting movements that can be interpreted as precursors of a
required specifications of IDS GeoRadar s.r.l. not expressly authorized by IDS collapse.
GeoRadar s.r.l.;
iv. Monitoring Products include specific “Operational” software with data processing
f. Poor or faulty operation of the electrical and telecommunication networks which may give outcomes/results for helping User in the monitoring of “the
not directly managed by IDS GeoRadar s.r.l. or its delegates. stability conditions” of a specific area. We remind however that the setting of the
g. Poor or faulty operation Software/Hardware of the third parties connected alarm thresholds is uniquely an expert User’s responsibility and that this setting
with IDS GeoRadar s.r.l. Equipment. must be done according to the accurate knowledge of the particular geological
context.
h. Poor or faulty operation of the Products due to Software Virus which
infected the Products after their delivery. v. Nevertheless, (even though the radar technology employed by Monitoring
Products, is deemed among the most reliable for the monitoring of instable
i. Use of the Products which have encountered suspected manumissions, slopes) IDS GeoRadar s.r.l. has the obligation to remind the Users that the
accidents, electrostatic shocks, flashes, fire, earthquake, flooding or other technology used by Monitoring Products transmits and receives an
natural disasters or unexpected events. electromagnetic signal that by its nature is influenced by many factors such as: the
parameters introduced by the User/s, the distance, the ground material, the
FCC use limits Please, feel free to contact us sending an e-mail to:
SUPPORT.GEO@IDSGEORADAR.COM OR
Use in all U.S. territories is limited to underground mines, open pit mines, SUPPORT.MINING@IDSGEORADAR.COM
and tunnels. Authorized EIRP may not exceed 48 dBm.
Or refer to:
HYDRA devices may not cause interference to and must tolerate
+39 050 8934 100 Customer Care Offices
interference from Government
IDS GeoRadar srl
Radiolocation, 76-81 GHz Band Radar, Radio Astronomy, Amateur, and
Space Research operations. Via Augusto Righi 6-6a - 8, Località Ospedaletto, 56121 PISA ITALY
Thus, this product poses no health and safety risk when operated in the
normal manner of intended use.