Photometric Report 12/01/2023 15:43:59

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......
Standard Table
Sample ID Type Ex Conc WL251,0 Wgt.Factor Comments
1 5ppm Standard 5.000 0.841 1.000
2 4ppm Standard 4.000 1.106 1.000
3 3pmm Standard 3.000 0.547 1.000
4 2ppm Standard 2.000 0.366 1.000
5 1ppm Standard 1.000 0.191 1.000
6 0ppm Standard 0.000 0.000 1.000
7

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
1 IV.3.151 Unknown 4.929 0.848 25UL+4ML ETOH
2 IV.3.30 Unknown 1.092 0.203 50UL+2ML ETOH
3 OR.3.15 Unknown 1.190 0.219 25UL+1ML ETOH
4 IV.4.15 Unknown 0.084 0.033 75UL+2ML ETOH
5 IV.4.30 Unknown 1.764 0.316 25UL+1ML ETOH
6 OR.4.15 Unknown 4.345 0.750 25UL+2ML ETOH
7 OR.4.30 Unknown 0.552 0.112 50UL+2ML ETOH
8 IV.2.45 Unknown 0.056 0.028 75UL+2ML ETOH
9 IV.2.30 Unknown 0.676 0.133 25UL+1ML ETOH

Page 1 / 6
Photometric Report 12/01/2023 15:43:59

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
10 IV.2.60 Unknown 0.717 0.140 25UL+1ML ETOH
11 IV.2.15 Unknown 2.046 0.363 25UL+1ML ETOH
12 OR.3.45 Unknown 0.898 0.170 25UL+1ML ETOH
13 OR.3.30 Unknown 3.357 0.584 25UL+2ML ETOH
14 OR.3.60 Unknown 1.049 0.195 25UL+1ML ETOH
15 OR.3.90 Unknown 0.917 0.173 25UL+1ML ETOH
16 OR.4.60 Unknown 0.710 0.138 25UL+1ML ETOH
17 OR.4.45 Unknown 0.782 0.150 25UL+1ML ETOH
18 IV.4.45 Unknown 0.631 0.125 25UL+1ML ETOH

Page 2 / 6
Photometric Report 12/01/2023 15:44:00

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
19 IV.4.90 Unknown 4.804 0.827 25UL+2ML ETOH
20 OR.2.15 Unknown 0.986 0.185 25UL+1ML ETOH
21 OR.2.30 Unknown 3.035 0.529 25UL+2ML ETOH
22 OR.2.45 Unknown 0.730 0.142 25UL+1ML ETOH
23 OR.2.90 Unknown 0.979 0.184 25UL+1ML ETOH
24 OR.2.120 Unknown 0.289 0.068 50UL+1ML ETOH
25 IV.2.90 Unknown 0.894 0.169 25UL+1ML ETOH
26 IV.1.15 Unknown 0.772 0.149 25UL+1ML ETOH
27 IV.1.90 Unknown 0.972 0.182 25UL+1ML ETOH

Page 3 / 6
Photometric Report 12/01/2023 15:44:00

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
28 OR.1.90 Unknown 4.595 0.792 25UL+4ML ETOH
29 IV.2.120 Unknown 0.925 0.175 25UL+1ML ETOH
30 IV.1.45 Unknown 5.008 0.861 25UL+2ML ETOH
31 OR.2.60 Unknown 0.957 0.180 50UL+1ML ETOH
32 IV.1.30 Unknown 3.555 0.617 25UL+1ML ETOH
33 IV.1.15.2 Unknown 0.931 0.175 25UL+1ML ETOH
34 OR.1.45 Unknown 0.875 0.166 25UL+1ML ETOH
35 OR.1.15 Unknown 2.239 0.396 25UL+1ML ETOH
36 IV.4.120 Unknown 0.941 0.177 25UL+1ML ETOH

Page 4 / 6
Photometric Report 12/01/2023 15:44:00

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
37 OR.4.90 Unknown 1.194 0.220 25UL+1ML ETOH
38 OR.4.120 Unknown 1.218 0.224 25UL+1ML ETOH
39 IV.4.60 Unknown 2.584 0.454 25UL+1ML ETOH
40 IV.3.45 Unknown 1.017 0.190 25UL+1ML ETOH
41 IV.3.120 Unknown 0.303 0.070 25UL+1ML ETOH
42 OR.3.120 Unknown 0.945 0.178 25UL+1ML ETOH
43 IV.3.60 Unknown 1.155 0.213 25UL+1ML ETOH
44 IV.3.90 Unknown 2.070 0.367 25UL+1ML ETOH
45 OR.1.120 Unknown 0.903 0.171 25UL+1ML ETOH

Page 5 / 6
Photometric Report 12/01/2023 15:44:00

File Name: D:\Farmakokinetik 2022\reguler Pagi\Konversi Al-Fattah 2021\std papaverin 1-4.pho

[Wavelengths] Standard Curve


Wavelength Name: WL251,0 0,945
Wavelength: 251,00 nm

[Calibration Curve]
Column for Cal. Curve: WL251,0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
Selected Wavelength: WL251,0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled
0,500

Abs.
[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 12/01/2023 11:14:25
Comments:
Sample Preparations:

[Instrument Properties]
0,000
Instrument Type: UV-1800 Series
Measuring Mode: Absorbance -0,086
Slit Width: 1,0 nm
Light Source Change Wavelength: 340,0 nm
0,000 1,000 2,000 3,000 4,000 5,000
S/R Exchange: Normal Conc. (mg/l)
y = 0,168188 x + 0,0189706
[Attachment Properties] r2 = 0.99678
......

Sample Table
Sample ID Type Ex Conc WL251,0 Comments
46 OR.1.60 Unknown 0.936 0.176 25UL+1ML ETOH
47 IV.1.120 Unknown 1.434 0.260 25UL+1ML ETOH
48

Page 6 / 6