Professional Documents
Culture Documents
All contents in this manual were strictly compiled according to related laws
and regulations in China, as well as the specific condition of URIT BH-5390
5-Part-Diff Auto Hematology Analyzer, covering all the updated information
before printing. URIT Medical Electronic Co., Ltd. is fully responsible for the
revision and explanation of the manual, and reserves the right to renovate the
relevant contents without separate notification. Some of the demonstration
pictures are for reference and subject to real object if any differences.
All the information included is protected by copyright. No part of this
document may be reproduced, stored or transmitted in any form or by any
means unless written authorization by URIT Medical Electronic Co., Ltd.
All instructions must be followed strictly in operation. In no event should
URIT Medical Electronic Co., Ltd. be responsible for failures, errors and other
liabilities resulting from user's noncompliance with the procedures and
precautions outlined herein.
Limited Responsibility for Quality Warranty
CAUTION
THE ANALYZER IS FOR PROFESSIONAL AND PRESCRIPTION USE
ONLY.
Technical service and troubleshooting are provided by URIT Customer
Support Center. Professional technician and sale representative will be sent to
offer you timely service when necessary.
Tel: +86(773)2288586
Fax: +86(773)2288560
Web: www.urit.com
Email: service@uritest.com
Version: 06/2019-C3.
II
Content
I
Content
II
Content
III
Content
IV
Content
V
Content
VI
Content
VII
Content
VIII
Chapter 1 Introduction
1.1 Overview
Welcome to read the URIT BH-5390 5-Part-Diff Auto Hematology Analyzer’s
operation manual. This manual includes instrument operation, maintenance
instructions and matters needing attention. In order to keep the instrument in a
good performance, operation and maintenance should be done according to
this manual.
NOTE
If the user does not operate the instrument according to this manual,
misuse will cause misdiagnose and delayed treatment due to inaccurate
measurement, or harms to operator himself, even instrument damage.
The operator must strictly follow the procedures in the instruction manual.
1
Chapter 1 Introduction
operators. Please read it thoroughly at the first use. You can use contents to
quickly find the required information in daily use. All related personnel should
read this manual.
1.3 Guidance
Operator can find the information needed according to the table.
Information Reference
2
Chapter 1 Introduction
Diluent: CBC,CBC+5DIFF.
Test Mode Peripheral blood: CBC,CBC+5DIFF.
3
Chapter 1 Introduction
Input
External printer with Chinese and English report format can print
Information
histogram, abnormal conditions (warning symbols) and reference
Output
range. Printed items, format and template is editable.
4
Chapter 1 Introduction
It can connect to web and transmit data via USB port and
Network interface. Supporting two-way LIS/HIS system makes information
in each department networking.
Power 80VA-250VA
Weight 76.5kg
5
Chapter 2 Safety Information for Operation
2.1 Overview
In addition to the safety use information, the general matters of operators in
terms of security are also shown in this chapter. Please read this chapter
carefully before operation.
Only allow to use the reagents and detergents mentioned in this manual.
Operating requirements also include regular cleaning and maintenance.
Symbols Meaning
6
Chapter 2 Safety Information For Operation
Beware
Warning of electric Biohazard
shock
On Alternating
Validity
(power) current
In vitro
Off diagnosti Protect
(power) c medical from heat
device
Environme
Serial
nt-friendly Fuse
number
use period
Dateof
manufactu Fragile Ethernet
re interface
Electronic
products Laser Stacking
recycling hazard level limit
logo
Consult
instruction Upward CE Logo
s for use
No Watch
Protect
turning your finger
from water
over and hand
Read the operation manual before using. Understand all the important
signs. Please keep manual for future reference.
7
Chapter 2 Safety Information For Operation
Start the analyzer following the manual instructions, otherwise its function
loss might be caused due to accidental mechanical damage and
undesirable environment.
Keep long hair, fingers and clothes away from rotating parts.
Turn off the power and unplug the power cord immediately if the analyzer
gives off odor or smoking, otherwise it causes fire, electric shock or injury.
If this happened, please contact the after-sale service department.
Do not spill the samples or reagent and do not make other things falling
into the analyzer, otherwise it causes short circuit. If this happens, turn off
the power, unplug the power cord and contact the after-sale service
department immediately.
Do not touch the circuit, especially by wet hand, which causes electric
shock.
Make sure to connect the analyzer with correct voltage and grounding.
Avoid damaging the power cord. Do not put any devices upon the power
cord, and do not pull the power cord.
Turn off the power before connecting other devices (host computer,
printer).
Data conversion errors and incorrect results are caused due to strong
electromagnetic interference and poor grounding.
Declaration
8
Chapter 2 Safety Information For Operation
2.5 Installation
The analyzer must be installed in dry and dust-free place. Do not place it in
a wet, dirty, airless or salt and sulfur place. Shell material is ABS + PC, it is
corrupted if being placed in a high pH environment.
Do not expose it to the place with large temperature difference and direct
sunlight.
Avoid vibration. Put it into a box with foam to prevent damages during
storage and transportation. Improper package may lead to abnormal
operation of the analyzer.
Installation site must be far away from strong ionizing radiation, such as
X-ray and γ-ray. This may cause test results errors.
9
Chapter 2 Safety Information For Operation
Do not contact the waste and its components with free hands.
2.7 Reagent
Check marks on the package.
Avoid direct contacting with reagents, since the reagents may irritate eyes,
skin and mucous membranes.
If eyes contact with the reagents, wash with water and seek medical
advice immediately. Please read reagent safety instruction.
Protect the reagents from being polluted by dust, dirt and germs.
Do not make the reagents spilt. If it happens, wipe away with a cloth.
Diluent is a kind of good conductor, if being spilt next to the wire or device,
it may cause electric shock. Please turn off the power, unplug the plug and
clean the diluent.
10
Chapter 2 Safety Information For Operation
The probe detergent is strongly alkaline cleaner. Do not let it contact the
skin or clothes. If that happens, rinse the skin and clothes with plenty of
water immediately.
Ensure that the reagents keep the same level with the analyzer or lower.
Do not put reagents on the top of the analyzer.
2.8 Maintenance
As a precision electro-optical analyzer, maintenance is necessary for
normal operation. The test data may have small deviations without regular
cleaning. In rare case, operator might be infected due to poor cleaning.
To prevent infection, electric shock and burn, operator must wear rubber
gloves in maintenance work. Wash hands with disinfectant after work.
If the analyzer is not used for a long time, empty and rinse the flow system
according to the procedure before disuse. Ensure the analyzer is in a
good working condition before reuse.
2.9 Laser
11
Chapter 2 Safety Information For Operation
2.10 Consumables
The disposal of residual reagents, cleaning agent and all waste must comply
with local laws and regulations. Used samples and reagents should be
separated from ordinary wastes, or they may cause environmental pollution.
Pollutants may also make the equipment unable to work.
2.11 Operators
This medical analyzer must be operated by well-trained personnel
exclusively. If being operated incorrectly by non-skilled staff,
misemployment will lead to inaccurate measurement and cause
misdiagnosing, delaying patient’s treatment or doing harm to the operator
himself, even damaging the analyzer.
Invalid hardware / software would affect the accuracy of test results. The
operator needs to contact the after-sale service personnel as soon as
possible.
WARNING
WARNING
12
Chapter 2 Safety Information For Operation
2.Do not install other application programs except virus checker program.
4.Do not download any file which has nothing to do with the software
program.
6.Do not use U disk or other storage media on the computer to prevent
them bringing virus to the computer.
13
Chapter 3 System and Function
3.1 Overview
BH-5390 5-Part-Diff Auto Hematology Analyzer, which is a vitro diagnostic
medical device, is used for blood cell count, WBC five part differential,
hemoglobin concentration measurement, and body fluid cell count in clinical
tests. The analyzer produces accurate test data of human venous blood, which
provides necessary reference for clinical diagnosis.
3.2 Parameters
The analyzer can analyze and arrange the sample data automatically and give
blood cell count and white blood cellfive part differential count respectively. It
also generates the three-dimensional plot and scatter diagram of white blood
cells and histogram of red blood cells and platelet.The BH-5390 generates the
following test parameters in Table 3-1(including 35 test parameters and 14
research parameters).
14
Chapter 3 System and Function
Immature Reticulocyte
IRF %
Fraction Percent
Mean Corpuscular
MCH pg
Hemoglobin
Mean Corpuscular
MCHC g/L
Hemoglobin Concentration
PCT Plateletcrit %
15
Chapter 3 System and Function
Abnormal lymphocyte
InR‰ ‰
count Infected RBC Permillage
Abnormal lymphocyte
ALY# 109 /L
count
16
Chapter 3 System and Function
Note: PCT and PDW are the inferred parameters, which are provided for
laboratory use only. Research parameters also includes ALY%, ALY#, LIC%,
LIC#, NRBC%, NRBC#, BLAST%, BLAST#, InR#, InR‰, Eos-BF%, Eos-BF#,
Neu-BF% and Neu-BF#.
3.3 Structure
WARNING
Take out the analyzer and then check whether the appearance is intact.
Ensure there is no damage during transport.
17
Chapter 3 System and Function
3------- Buttons
18
Chapter 3 System and Function
2--------Valve
3--------Sampling Mechanism
4--------Mixing Mechanism
5--------Syringe
7--------Card Reader
19
Chapter 3 System and Function
1-----------Left Door
2-----------Lock
20
Chapter 3 System and Function
2-----------Gasholder
3-----------Valve Module
4-----------Sample Cup
6-----------Syringe
7-----------Air Pump
8-----------Sampling Mechanism
21
Chapter 3 System and Function
1-----------Right Door
2-----------Lock
22
Chapter 3 System and Function
3-----------Switch Power
23
Chapter 3 System and Function
1-----------Net Port
2-----------Power Switch
4-----------Power Socket
5-----------Diluent inlet
7-----------Lyse inlet
8-----------Detergent inlet
9----------Sheath Inlet
24
Chapter 3 System and Function
Semiconductor Laser is above the analyzer. Do not open the upper cover
for your safety, only the personnel authorized by URIT can open it.
3.4.1 Introduction
In addition to released version No. and complete Version No., other version
numbers are independent subroutine version number, standing for one of
25
Chapter 3 System and Function
Main board MCU function: host’s general control and communication with PC;
Memory:4GB or above;
Software has to been installed and equipped well, and pass strict inspection
before leaving factory. Software installation and update must be performed by
engineers with URIT technical service qualification.
26
Chapter 3 System and Function
Control
Data Status Panel
Area Area
27
Chapter 3 System and Function
3.Data Area
4.Status Area
6.Control Panel
28
Chapter 3 System and Function
Icon Area
Instrument Status
Sample No.
Buttons
Control Area
Buttons
29
Chapter 3 System and Function
Icon Area:
Instrument Status:
Sample No.:
Buttons:
Control Area:
You can choose various test modes in blood routine mode, RETIC mode and
Body Fluid mode.
NOTE
The blank test should be done after the replacement of diluent, detergent,
sheath and detergent to ensure it is within the normal range.
The reagent inlet tubes have a cap attached that minimizes evaporation
and contamination during shipping. The tubes can only insert reagent to
right connections. Please close the cap tightly.
30
Chapter 3 System and Function
3.6.1 Diluent
Diluent which is tasteless transparent isotonic fluid can be used for blood cells
counting and classification. It has the following functions.
Storage condition:5℃-35℃
Once opened (connected to the analyzer), the product shelf life is only 60
days.
3.6.2 Sheath
Sheath is used to keep the original ecology of blood cells and bleach RBC to
eliminate the scattering of laser.
WBC, which is diluted in sheath, maintains the cell structure closest to its
original state. Basophilic granule is soluble in water, so the structure of
Basophil has minor changes. RBC osmotic pressure is higher than that of
sheath, so RBC is changed by sheath. The hemoglobin of RBC diffuses from
the cells, and moisture content of sheath diffuses into cells. Although the cell
membrane remains in a good shape, the RBC and sheath have the same
refractive index, which makes the RBC invisible under the laser.
Storage condition:5℃-35℃
Once opened (connected to the analyzer), the product shelf life is only 60
days.
3.6.3 Lyse
Lyse which doesn’t contain azide and cyanide can achieve the following
requirements.
(1) Quickly dissolve the RBC and generate less ground substance
31
Chapter 3 System and Function
complex.
(2) Change the WBC cell membrane and make its cytoplasm diffused
gradually. Meanwhile, the cell membrane surrounds the cell nucleus and
contracts, making the WBC become granular.
(4) Cyanide free. Avoid harm to your body and the environment caused by
the cyanide .
Storage condition:5℃-35℃
Once opened (connected to the analyzer), the product shelf life is only 60
days.
3.6.4 Detergent
Detergent contains the active enzyme which can clean the agglomerated
protein in the cups and fluid system of WBC and RBC. It prevents plugging
holes.
Storage condition:5℃-35℃
Once opened (connected to the analyzer), the product shelf life is only 60
days.
Probe detergent contains the effective oxides which can clear away the protein
stain, so as to solve problems of WBC and RBC probe clogging.
WARNING
2.If skin contacts with the reagents, wash with water immediately.
3.If eyes contacts with the reagents, wash with water and seek medical
advice immediately.
32
Chapter 3 System and Function
Control materials and calibrators are for analyzer quality test and calibration.
The "control material" and "calibrator" mentioned in this Manual refers to the
special control material and calibrator assigned by URIT. Users can purchase
from URIT or agents designated by URIT.
33
Chapter 4 Installation
4.1 Overview
CAUTION
Environment requirements:
Place the analyzer on a smooth and big enough platform which is easy to
operate. Away from direct sunlight.
WARNING
This instrument has been tested strictly before delivery. It is carefully packed
before transport to avoid damage. Please carefully check the packaging as
receiving. If finding any damages, please immediately contact the after-sale
service department of URIT or local agent.
34
Chapter 4 Installation
(3) 100 cm of space above to either side of the analyzer for service
access.
(4) sufficient space under the analyzer for placing reagents and waste
containers.
AC 100V-240V 50/60 Hz
WARNING
35
Chapter 4 Installation
(7) Away from communication equipment which may interfere the analyzer
by producing high frequency electric wave.
WARNING
It is prohibited to pour the waste into the sewer directly. The waste must be
processed by biological or chemical methods before pouring into the
sewer. Hospitals and laboratories have the obligation to comply with the
relevant provisions of environmental protection department of local
government.
36
Chapter 4 Installation
CAUTION
Please ensure that the equipped computer is only used for the analyzer.
Installing other software, using removable storage devices such as U disk,
playing games or surfing the Internet on the computer may cause
computer virus, system damage or other failure.
There are five liquid interfaces on the rear panel, which are LYSE,
DILUENT,DETERGENT, SHEATH, and WASTE. Each of them is wrapped with
a faucet to avoid contamination before shipment. Take the faucet off for the
first time installation. Please store it carefully.
NOTE
37
Chapter 4 Installation
base or lower.
Take out the lyse inlet tube with red faucet from the accessories box, and inset
it to the LYSE interface on the rear panel. Place the other end of the tube into
the lyse container and twist the cap tightly.
Take out the diluent inlet tube with blue faucet from the accessories box, and
inset it to the DILUENT interface on the rear panel. Place the other end of the
tube into the diluent container and twist the cap tightly.
Take out the detergent inlet tube with green faucet from the accessories box,
and inset it to the DETERGENT interface on the rear panel. Place the other
end of the tube into the detergent container and twist the cap tightly.
Take out the sheath inlet tube with yellow faucet from the accessories box, and
inset it to the SHEATH interface on the rear panel. Place the other end of the
tube into the sheath container and twist the cap tightly.
Take out the waste outlet tube with faucet from the accessories box, and inset
it to the interface on the rear panel. Inset BNC plug to the SENSOR connector
on the rear panel. Tightly twist the tube’s cap clockwise onto the waste
container. Place the waster container on the level at least 50cm lower than the
analyzer.
Installation steps:
(1) Place the printer in an appropriate location adjacent to the analyzer for
easy operation.
(6) Connect the power cord to the printer, and connect it to the grounding
38
Chapter 4 Installation
plug.
(7) Confirm that the printer and computer are properly connected.
(8) Install the ink cartridges and paper according to the instructions,
ensure the printer is adjusted to the correct receiving size.
(9) Connect the power cord to a grounded outlet and turn the power on.
NOTE
39
Chapter 5 Principles of Operation
5.1 Overview
BH-5390 tests the amount and volume distribution of white blood cells, red
blood cells and platelets by the electrical impedance method (also known as
Coulter theory), measures hemoglobin concentration by colorimetry, analyzes
five part differential of white blood cells by the 4-angle laser scattered method.
Three separated channels are used for getting the blood cells counting results
respectively.
(1) WBC and five part differential data of are detected by laser in sheath flow
regulator.
(2) HGB and total amount of WBC is detected by electrical impedance method
and colorimetry in WBC counting chamber.
(3) The data of RBC and PLT is detected by electrical impedance method in
RBC counting chamber.
Test Mode:
40
Chapter 5 Principles of Operation
the WBC counting chamber, RBC counting chamber, WOC cup. Then finally
results of WBC count/HGB measurement, WBC/PLT count and WBC five part
differential are produced.
41
Chapter 5 Principles of Operation
The counting procedure is the same as whole blood auto sampling mode. It is
mainly used to interrupt the current batch test and insert emergency sample
test.
42
Chapter 5 Principles of Operation
43
Chapter 5 Principles of Operation
The whole blood samples are diluted with an appropriate proportion of sheath.
WBC remains its original state. Make cells in a single arrangement flow by
FCM (flow cytometry). The scattering density can be measured from four
different angles through the laser beam detection zone.
(1) 00: Forward Angle Light Scatter (10~30), roughly measure size of cells,
(2) 100: Narrow-Angle Light Scatter (70 ~ 110), measures cell structure and
relative characteristic of complexity.
(3) 900 D:Depolarized Light Scatter (700 ~1100),separates the oxyphil cells
from neutrophil cells and other cells, based on the characteristics that
particles can depolarize a laser with a vertical polarization angle.
(4) 900: Ninety-Degree Light Scatter (700~1100), mainly measures the cell
component and internal particle.
44
Chapter 5 Principles of Operation
The light source shown in the above figure is a vertical direction laser with
wavelength of 638nm, and the power is 7-10mw. Laser beam goes through a
cylindrical lens which can change the shape of beam spot from circle to oval. It
is shaped into a spot with a 80um-wide cell through an imaging lens and focus
on the cells in quartz sheath flow.
The horizontal forward angle light directly scatters to the punch hole through
the convergent lens. The light of 0° pass through the hole to the
siliconphotodiode detective unit of 0°. 10° scattering light reaches to the 10°
silicon photodiode detection unit by reflector.
Vertical scattered light is collected by the condenser lens group. After the
scattered light which contains cell information passing through the condenser
lens group, the vertical scattered light will be divided into two parts by a beam
splitter mirror. A part of light directly scatters to the 90° photomultiplier tube.
The remaining scattered light will go through the line polarizer, and only the
depolarizing scattered light can reach 90°D depolarizing photomultiplier tube.
45
Chapter 5 Principles of Operation
13
1 33
12
8 11
3
2 14
5 9
6 10
7
2—Reflector Unit
8—Microscope Unit
11—90° PMT
12—90°D PMT
13—Laser
46
Chapter 5 Principles of Operation
The gray area in left side figure is the ghost cells. It’s the reflection as RBC
dissolving into pieces. The green area is the lymphocyte group, the pink area
is the monocyte group, the blue area is the neutrophil, the white area is the
basophilgroup, and the red area is the eosinophil group.The blue part in the
right scatter diagram is the neutrophil group, and the red part is the eosinophil
group.
47
Chapter 5 Principles of Operation
The analyzer divides the WBC into basophil, eosinophil, monocyte, neutrophil
and lymphocyte via four-angle scatter analysis as the WBC going through the
sheath flow regulator. The default unit of cell amounts is 10^9/L.
WBC Count
The WOC and WIC is obtained by electrical impedance method and laser,
and finally gets the total numbers of WBC.
Lymphocyte Percent
Lym% = Lym#/WBC
Monocyte Percent
Neutrophil Percent
Neu%=Neu#/WBC
Eosinophil Percent
Eos%=Eos#/WBC
Basophil Percent
Baso%=Baso#/WBC
After adding lyse into the diluted sample in WBC counting chamber, the RBC
dissolves and the hemoglobin is released. The hemoglobin combines with lyse
to form hemoglobin mixture which is illuminated by the LED light-emitting diode
with a 540nm-wavelength monochromatic light at one end of the WBC
counting chamber. Using the optical tube to receive the transmitted light at the
48
Chapter 5 Principles of Operation
other end, amplifying the light intensity signal and convert it to the voltage
signal. Compare it with the voltage generated by the transmission light
intensity before adding the sample into the colorimetry chamber (only with
diluent), the hemoglobin concentration is achieved. Hemoglobin concentration
is proportional to the sample absorbance in 540nm wavelength. The process
of measurement and calculation is done automatically by the analyzer,
relevant results is displayed in the analysis results area.
E
HGB K Ln B
ES
K is a constant.
As the pulses amount corresponds to the amount of cells passing through the
pores, and the pulse amplitude corresponds to the cells volume, the analyzer
measures each cell and classifies it according to its volume. The analyzer
automatically divides the cells into RBC, WBC, PLT and other groups in
accordance with pre-set volume classification procedure.
49
Chapter 5 Principles of Operation
The volumetric metering unit controls the sample size passing through the
pore during counting to obtain the exact counting results in quantitative
samples. The volumetric metering unit includes metering tube and two
photoelectric sensors.
As shown in Figure 5-7, empty the metering tube before counting. The liquid
level of metering tube declines slowly as the sample passing through the pore.
When the liquid level passes through the start detector, one electrical signal
generates, and the analyzer starts counting. When the liquid level reaches the
stop detector, it also generates an electrical signal, then the counting finishes.
If there were bubbles or other abnormal stream in the flow system, "bubble" or
"clog" alarm pops up. Please refer to Chapter 11 Troubleshooting.
50
Chapter 5 Principles of Operation
● RBC Count
RBC = n ×1012/L
● Mean Corpuscular Volume
The mean corpuscular volume (MCV) is the average volume of each RBC. The
MCV is derived from the RBC size distribution data. The unit is fL.
●HTC
51
Chapter 5 Principles of Operation
52
Chapter 5 Principles of Operation
●Platelet Count
PLT=n×109/L
The mean platelet volume (MPV) is calculated by PLT histogram. The unit is fL.
The platelet distribution width (PDW) is gotten from the PLT histogram. It is the
geometric standard deviation of PLT volume distribution (10 GSD).
●Thrombocytocrit
53
Chapter 5 Principles of Operation
During data acquisition, 10 degree and 90 degree scatter are collected for up
to 300,000 signals. The 0 degree threshold is set high enough to exclude most
platelets. Histogram data are used to differentiate reticulocytes, mature RBCs,
platelet clumps, and nucleated cells. Reticulocytes have 10 degree scatter that
are similar to the scatter for mature RBCs, but differ from them by exhibiting
greater 90 degree scatter. Reticulocytes are reported in percent. The analyzer
will automatically calculate the reticulocyte Absolute value if an RBC count is
entered. The RBC value may be obtained from the Standard Hematology Data
Log, or it may be entered by the operator directly on screen.
Immature reticulocytes contain more RNA and absorb more stain than mature
reticulocytes, therefore, they exhibit greater 90 degree scatter. On BH-5390,
immature reticulocytes are classified as the population of reticulocytes that
exceed a predetermined scatter threshold. Consequently, it is possible to
determine the Immature Reticulocyte Fraction (IRF) from the scatter
measurements.
The IRF was initially designated as the Reticulocyte Maturation Index (RMI),
and defined by NCCLS H44-A as a quantitative expression of the relative
maturation of the reticulocytes in the observed reticulum in New Methylene
blue-stained preparations. However, these quantitative visual measurements
of reticulocyte maturation have been little used due to the subjectivity and
imprecision of the manual analysis. Since Auto reticulocyte methods allow the
enumeration of immature reticulocytes as a subfraction of the total reticulocyte
population, the preferred nomenclature is Immature Reticulocyte Fraction
(IRF). The immature reticulocytes are then reported as a fraction (or percent)
of the reticulocytes.
Monitor bone marrow toxic insults from drugs (for example, AZT)
Classify anemia
54
Chapter 6 Settings
6.1 Overview
Initialization setting of BH-5390 has been done before delivery. Setting
interface at the first boot is default. To meet the different needs, some
parameters can be reset.
When changes are completed, click “Save”. Click “Default” to recovery the
changed settings.
6.2.1 Alarm
55
Chapter 6 Settings
Adding in the box in front of“Turn on alarm tone”, the failure warning is
enabled. Adding in the box in front of “Turn on waste Alarm”, the analyzer
will raise a alarm when waste is full.
Various alarm threshold can be set. When a parameter is less than the set
threshold,the corresponding alarm prompt appears on instrument. For
example, if the alarm threshold of "Leukopenia" is set to be less than 2.5, when
the value of leukopenia is less than 2.5, the instrument will give an alarm, as
shown in Figure 6-2.
Suspected Alarm:
Adding in the box in front of “Turn on alarm tone”, this function is enabled.
Unchecked means no prompt.
After the alarm is turned on, the alarm sensitivity of each item can be selected.
As a median, 50 is default. If sensitivity is not sensitive enough, you can raise
the numerical value until 100, as Figure 6-3.
56
Chapter 6 Settings
6.2.2 Display
General:
57
Chapter 6 Settings
When one item is selected, you can also adjust its position by click or
58
Chapter 6 Settings
Parameters:
Parameter language, unit and range order can be set in “Param”. Clicking
to display more options. Operators can choose them according to their own
habits and local regulations, as shown in Figure 6-6.
Selecting all is default in “Patient info input”. You can cancel the in front of
the item that not need to be displayed. This item can not be edited, as shown
in Figure 6-7.
Blood Type:
By default, all blood types are selected. When data is entered, all blood types
can be selected. If a certain blood type is canceled, the type is not selectable.
Cancel the in front of the item to not display it and this item is not
selectable, as shown in Figure 6-8.
59
Chapter 6 Settings
6.2.3 Sampler
Single Sampling
“Diluent mode prompt” can be set in “Single”. If it’s selected, a prompt appears
when you test in diluent mode. Cancel to cancel the reminder. See Figure
6-10.
60
Chapter 6 Settings
Auto Sampling:
You can set the reminder function in auto sampling. If it’s selected, a prompt
appears when you test in auto sampling mode. Cancel to cancel the
reminder.
Handheld Scanner:
When the user needs to connect the external handheld scanner, please
calibrate here, and verify the new scanner first. (Provided by clients)
After connection, add in front of “Enable the scanner” and click “Start
Cal”. Then you can scan barcode. Scanning data is displayed at Scan results.
61
Chapter 6 Settings
Scanner:
62
Chapter 6 Settings
Matching rules:
Match by serial No.: create a new work sheet and enter serial No. manually.
When serial No. is detected, test results will be matched automatically.
Match in order: after the new sample is created, the test results will be
matched from the first one orderly in main interface.
Add in front of “Use 24 hour format” to display 24-hour format. Cancel the
63
Chapter 6 Settings
You can set whether the system taskbar is displayed when maximizing window.
Adding in front of “Auto run when system start”, software will start
automatically after computer is turned on, as shown in Figure 6-16.
64
Chapter 6 Settings
Select the enumerated description and click “Edit” to modify it. Rules name,
rule code, Sex, Upper age limit and lower age limitcan be modified.
Click Setting and select one of the parameter rules to re-edit it. Click “Save” ,
as shown in Figure 6-18.
65
Chapter 6 Settings
FLAG rules can be edited in Setting interface. Select the items you need and
click “OK”, as shown in Figure 6-19.
66
Chapter 6 Settings
Parameter rules are mainly the requirements of parameter. FLAG rules are
mainly the requirements of cases. The relationship between these two can be
“AND” or “OR”. Also you can only use parameter rules but not FLAG rules,
selecting in drop-down box, as shown in Figure 6-20.
67
Chapter 6 Settings
Click “New” to increase the number of rules. The process is the same as
editing.
6.2.8 Language
68
Chapter 6 Settings
6.3 Maintenance
Click “Maint”in Setting interface, as Figure 6-22.
Click “Save” to save your settings after every modification. If you want to
recovery settings, please click “Default”.
General:
69
Chapter 6 Settings
Serial No.:
“Auto-sample reset value” are reset automatically when staring up. Cancel ,
and the serial No. will not reset but increase.
When single-sample reset value is set as 1000, the serial No. will increase
from 1000. When auto-sample reset value is set as 1, the serial No. will
70
Chapter 6 Settings
increase from 1. Users can change according to their own habits, as shown in
Figure 6-24.
6.4 Print
Click “Print” in Setting interface, as shown in Figure 6-25.
Click “Save” to save your settings after every modification. If you want to
recovery settings, please click “Default”.
6.4.1 General
You can set hospital name at “Print title”, print histogram and print alarm.
Adding in front of “Auto Print”, the test results will be printed automatically
after test. See Figure 6-26.
If users want to print other logos on report, please click “Browse” and select file.
71
Chapter 6 Settings
Click“Open” and the logo is checked. If the logo need not print, click
“Cancel”.See Figure 6-27.
6.4.2 Printer
The printer can be selected after printer drive program is installed. If there is
not corresponding model, please click “Refresh” and to choose, as shown
in Figure 6-28.
72
Chapter 6 Settings
6.5 Transmit
Click “Transmit” in Setting interface, as Figure 6-30.
Click “Save” to save your settings after every modification. If you want to
recovery settings, please click “Default”.
73
Chapter 6 Settings
Lis server can be connected and data can be transmitted, as shown in Figure
6-32.
74
Chapter 6 Settings
Lis Server:
Computer can be connected with Lis server. You can click “Connect” or
“Disconnect” as needed, as shown in Figure 6-33.
Data:
After Lis server connection, you can set whether to transmit histograms and
scatter diagram or not. “Auto transmit” and various transmit protocol can be
chosen, as shown Figure 6-34.
75
Chapter 6 Settings
NOTE
6.6 Limit
Click “Limit” in Setting interface, as shown in Figure 6-35.
In Limit interface, users can use default limit or enter proper parameter limit
manually.
Click “Save” to save your settings after every modification. If you want to
recovery settings, please click “Default”.
General:
To monitor blood sample’s abnormal test parameters, users can establish and
set their own range of normal parameters according to laboratory and clinical
76
Chapter 6 Settings
requirements. A prompt or sign will appear when the sample is out of range.
For example, when the sample test result exceeds the parameter limit range,
there is a “H” or “L” next to the results. (“H” means the result exceeds the
upper limit; “L” means the result exceeds the lower limit.)
NOTE
Click “Save” to save your settings after every modification. If you want to
recovery settings, please click “Default”.
77
Chapter 6 Settings
6.7.1 General
Counting time of whole blood sample and diluted sample, and their time range
are displayed here. The unit is S. See Figure 6-38.
The upper limit of whole blood RBC counting time is 14 seconds, which means
“Clog” alarm will appear in case of exceeding time; the lower limit is 12
seconds, which means “Bubble” alarm will appear when counting time is less
than 12 seconds. Notice that the counting time has already been set before it
leaves factory, so you’d better not change at will lest false alarm appears.
Please contact with URIT after-service department if you need to change.
78
Chapter 6 Settings
6.7.2 RRBC
RRBC reaction time has already been set before it leaves factory. Generally
speaking, it’s not necessary to change it lest the test results are affected.
6.8 Users
Click “User” in Setting interface, as shown in Figure 6-40.
Doctors or operators should login the system with identity to operate the
routine check, therefore, it is necessary to set up doctors’ information. Only the
administrator can make user settings.
79
Chapter 6 Settings
Click “Add” and enter user name, group and password. The serial No. and
mnemonic symbol are generated automatically. Click “OK” to complete adding
new user.
Select one user and click “Modify” to modify user’s information. Click “Delete”
to delete this user. See Figure 6-41.
80
Chapter 6 Settings
Permission
In order to ensure the proper use, it is necessary for the administrator to only
give partial permissions to other operators, such as query and count data. Do
not give access to delete data.
Select “Group”, and you can select the permission that you want to give, then
click “Save”, as shown in Figure 6-42.
81
Chapter 6 Settings
Reagent’s manufacture date, activated date and Lot No. are recorded here.
When a reagent is activated, its information is recorded automatically.
If a prompt of insufficient reagent appears, you need to replace it. Clicking the
82
Chapter 6 Settings
“Replace” button next to the corresponding reagent, the following dialog box
pops up, as shown in Figure 6-45.
Take the reagent card in the container close to the analyzer’s active device
until there are double beeps, as shown in Figure 6-47.
83
Chapter 6 Settings
84
Chapter 7 Daily Operation
7.1 Overview
This chapter introduces the whole procedures of daily operation, focusing on
the process of different modes of sample analysis in detail.
Preparations
Startup
Quality Control
Sample Preparation
Data Input
Sample Count
Statistical Analysis
Shutoff
85
Chapter 7 Daily Operation
NOTE
7.2 Preparations
Check the analyzer as the following steps before startup.
The waste should be disposed properly and cleaned up before startup every
day.
Ensure diluent, detergent, lyse, and sheath meet the test requirements.
Ensure the tubes of reagents and waste connects well and without bending.
Ensure the plugs of analyzer and computer are inserted into socket safely.
Ensure the power is on and the cable has been connected with the analyzer
and the computer properly.
Ensure the mouse and the keyboard are connected with the computer
properly.
7.3 Startup
Turn on the start button on the right panel, the status indicator on the front
panel turns orange and then turns blue several seconds later. After the
software is opened, you can input username and password to enter blood cells
86
Chapter 7 Daily Operation
test interface. The default username and password are admin.The software
connects to analyzer automatically and fluid system initialization starts. During
self-check and initialization, running condition of all are checked. Then, it
rinses the fluid system and do blank test. Now the boot process is completed,
as shown in Figure 7-1.
Fi
Figure7-1 Test Interface
After startup, blank test should be done before sample test. Operator can set
to run it automatically after startup, see Chapter 6 Settings for details. The
acceptable range of blank test is listed in Table 7-1.
WBC ≤0.2×109/L
RBC ≤0.02×1012/L
HGB ≤1g/L
PLT ≤10×109/L
HCT ≤0.5%
WBC-BF ≤0.001×109/L
RBC-BF ≤0.003×1012/L
87
Chapter 7 Daily Operation
If the blank test result is out of this range, please repeat the above procedures
until it is acceptable. If the result cannot reach the above range requirements
after five times testing, please refer to Section 11.4.2 of Chapter 11
Troubleshooting.
All the clinical specimens, control materials and calibrators may contain
human blood or serum and have potentially infectious hazards, so
operators should comply with the safe operation provisions in laboratory
and wear personal protective equipment (lab coats, glasses, gloves etc.)
when handling these materials.
NOTE
Venous blood can only be stored for 4 hours at room temperature. If it’s not
used up in a short period, it’s recommended to keep the blood sample at
the temperature between 2℃~8℃.
The indicator in different color means different work status of analyzer. Please
see the Table 7-2 for details. If the indicator turns red, please troubleshoot
according to prompts.
88
Chapter 7 Daily Operation
1 Green Standby
2 Orange Working
1. Heparin
2. Trisodium citrate
Trisodium citrate is a kind of liquid which is filled in the tube and diluted to
10/11 of the original. This anticoagulant is used for agglutination, it also can be
used when suspected thrombocytopenia from EDTA.
4. EDTA
In the salt of EDTA, use EDTA K2(United States and Japan)and EDTA K3
(United States and Europe),sometimes NA2EDTA. And EDTA K2, EDTA K3
are most widely used in the blood test in the world. It’s recommended to use by
ISCH in 1993. But other EDTA salts can also be used. EDTA could lead to
Pseudothrombocytopenia through Platelet aggregation. (Incidence is about
1/800)
5. Monofluoride
89
Chapter 7 Daily Operation
Use it before using EDTA. No adverse reactions have been found according to
the investigation.
Single Draining
Set the current test mode to “Diluent” in “Test” interface, and rotate the
sample holder of emergency bin to the position of disposable plastic tube.
Put a clean disposable plastic tube in the Emergency bin and close it.
Press “Drain” in the shell or click in pop-up dialog box. Click “Drain”
button, and the analyzer automatically drain 480μL diluent into the plastic tube.
Draining times is displayed, as shown in Figure 7-2.
Take out the tube after draining, and inject the collected 20μL blood into tube
with diluent, then mix it well.
Multiple Draining
2.Put the disposable plastic tubes in the test tube rack and placethe rack
at the right side of sample loader. Starting from the first position on the left, put
90
Chapter 7 Daily Operation
3.Click , and draining interface pops up. Select “Multiple” , input the
needed times and click “Drain”, as shown in Figure 7-4. Multiple sampling
starts. 480μL diluent will be drain into the prepared tubes orderly. The number
of draining must be consistent with the plastic tubes.
4.Take out the tube after multiple draining is completed. Inject the
collected 20μL blood into the tube filled with diluent and mix it well.
91
Chapter 7 Daily Operation
NOTE
Avoid dust falling into the prepared diluent, otherwise error may be
caused.
Let the reacted peripheral blood and diluent still for 3 minutes, and mix it
well again before analysis.
Please analyze the sample within 30 minutes after dilution, otherwise the
analysis results are not reliable.
CAUTION
Avoid contact with skin and clothing. The ingredient New methylene blue
stains skin, clothing and other objects.
Click “Print” to print test results, and click “Transmit” to transmit results to
computer. Click “Page up” and “Page down” to input information for the last
92
Chapter 7 Daily Operation
93
Chapter 7 Daily Operation
If group is not selected, the reference values are offered according to Table
7-3.
Table 7- 3 Groups
94
Chapter 7 Daily Operation
NOTE
The SN 0 is the special one for blank test. Please do not input 0 in sample
test.
CAUTION
7.7 Worksheet
A new sample can be created in worksheet in advance. Input sample
information and then analyze it.
Click “Worksheet” and “New” to create a new sample. Users can create
several samples as needed. After creation, sample information can be input
below, as shown in Figure 7-6.
Click “Save” to save your input. Click “Today” to return to main interface to test
it.
Only when you click “Save” after inputting information, can the next new
sample be created.
95
Chapter 7 Daily Operation
7.8 Review
To view today’s test results, you can click “Review”, then “Query”, as shown in
Figure 7-7.
After testing new samples in main interface, you need to click the “Query”
button in Review interface to display the new sample.
96
Chapter 7 Daily Operation
For the tested samples before today, you can choose a exact date. Also, you
can query by case ID, name, department, auditor and mode.
In Review interface, you can perform delete, print, preview, print stub, audit,
transmit and compare.
NOTE
Be aware that once the data are deleted, it can NOT be recovered. Please
operate with caution.
The query results can be sorted by various conditions, but by time is default.
Click “Query” to confirm after selecting a certain condition, as shown in Figure
7-8.
97
Chapter 7 Daily Operation
Briefing, Detail and data mode are offered for choice to display query results.
Select a certain mode and click “Query” to confirm, as shown in Figure 7-9.
The display effect of these three modes are shown in Figure 7-10-1、7-10-2、
7-10-3.
98
Chapter 7 Daily Operation
Operator can view result comparison chart of selected samples. Press and
hold Ctrl key, and click samples need to compare at the same time. Release
Ctrl key and click “Compare” to open data comparison window. Now you can
view results comparison of selected samples. See Figure 7-11-1、7-11-2.
99
Chapter 7 Daily Operation
100
Chapter 7 Daily Operation
For venous whole blood test, there are various measurement methods,
including single sampling mode and auto sampling mode. You can choose
CBC, CBC+5DIFF, CBC+5DIFF+RRBC to test in single sampling mode, and
CBC, CBC+5DIFF test in auto sampling mode, as shown in Figure 7-13.
Only single sampling can be selected in peripheral whole blood mode, choose
CBC or CBC+5DIFF to test, as shown in Figure 7-14.
101
Chapter 7 Daily Operation
For diluent test, there is only single sampling mode, with CBC and CBC+5DIFF
for choice, as shown in Figure 7-15.
NOTE
102
Chapter 7 Daily Operation
RBC and PLT. “CBC+5Diff " counts and classifies the WBC.
Open emergency bin and put the tube for whole blood or diluent test to the
corresponding hole position. The specification is shown in Figure 7-16. The
hold position is rotatable. Rotate the position to the front.
Close the door and press the “Count” button or click to start test, as shown
in Figure 7-17.
103
Chapter 7 Daily Operation
Put the whole blood tube into hole position of test tube rack. Place the rack to
the right of auto sample loader. 5 rows of rack can be placed at most each time.
Do not place it at the innermost position but leave a blank position for adding
rack during test. See Figure 7-18.
Press “Start” button or click to start test. See Figure 7-19. Holding the
tubes, analyzer mixes and then tests them automatically. In the process of test,
104
Chapter 7 Daily Operation
mode. After the completion of the current blood sample test, the dialog box
appears as shown in Figure 7-20. At this time, open the emergency bin, put the
sample tube in, and close the door. Press the “Start” button on the front
housing or click to start test.
In the process of auto sampling test, if you want to stop, click to reset auto
sample. After the completion of current samples, the test tube rack resets and
the auto sampling test is completed.
WARNING
105
Chapter 7 Daily Operation
CAUTION
NOTE
1.Add 20µL blood sample into staining solution for hematology analysis
(4mL). Shake the tube up and down 15 times, and place it into an incubator
in which the temperature is 35℃ for 15 minutes.
2.Take out the staining solution and shake it 15 times to mix it well. For
the tests after the first time, it’s not necessary to shake again. You can test it
directly. Please finish the test in 15 minutes after incubation and mixture. Do
not mix it after incubation unless tested immediately.
3. It’s best to test the sample in 2 hours if it’s not mixed after incubation.
The preparation process is the same as above.
Only the first background test is necessary at the same day. For the next
RETIC test, you don’t have to test background again.
106
Chapter 7 Daily Operation
When the background test is completed, select RETIC mode. See Figure 7-24.
107
Chapter 7 Daily Operation
Place the staining solution into emergency bin and close the door. Click the
Test button, and input RBC result of blood routine test in the popup dialog. See
Figure 7-25.
Verify the RBC value and click Test button to start test. RETIC results and
diagram are produced, as shown in Figure 7-26.
108
Chapter 7 Daily Operation
Three basic parameters, two scatter plots, two three-dimensional diagram, and
two histograms (S0 degree, S90 degree distribution histograms) are provided
for RET test in the above figure.
In the scatter diagram above, the red is RBC, the white is WBC, and green and
pink are RET.
109
Chapter 7 Daily Operation
7-26-4 Histogram
In the figure above, S0 is RBC volume distribution histogram, and S90 is RBC
complexity distribution histogram.
If you need merged display and print of RETIC results and blood routine
results, click Serial No. with the right mouse button. The prompt box pops up.
See Figure 7-27.
110
Chapter 7 Daily Operation
Click RETIC, the selected Serial No. is displayed in the dialog box, and the
RBC value is displayed automatically. See Figure 7-28.
Place staining solution for hematology analysis into emergency bin and close it.
Click to start test. RETIC results and blood routine results are displayed in
screen.
111
Chapter 7 Daily Operation
Only single sampling applies to body fluid mode. Fluid system and sample
cups will be cleaned automatically when body fluid mode is selected, as shown
in Figure 7-30.
Place the test tube with body fluid into emergency bin after cleaning. Close the
emergency bin and press Count button to start body fluid test. The test result is
shown as Figure 7-31.
Whole blood anticoagulant tubes can be used for testing body fluid samples.
112
Chapter 7 Daily Operation
The figure above shows 7 test parameters and 4 research parameters for body
fluids. At the same time, 2 scatter plots,2 three-dimensional diagram and 2
histograms (WBC-BF and RBC-BF distribution histograms) are provided.
In the scatter diagram above, the pink is MN, the blue is PMN, and red is EOS
(research parameter).
113
Chapter 7 Daily Operation
Figure 7-34 shows the volume distribution histogram of body fluid WBC and
body fluid RBC.
7.12 Shutdown
Please do the shutdown procedure before power off, which is to clean the
counting chambers and related tubes. At the end of one day test or in the case
of continuous use, shutting off procedure should be performed at least once
every 24 hours.
Procedures of shutdown:
114
Chapter 7 Daily Operation
4.After the rinse procedure finished, click “OK” in the popup dialog box,
and the computer is turned off automatically;
5.Finally, turn off the power switch on the rear panel of the instrument.
NOTE
Data loss and abnormal boot may be caused if the shutoff procedures are
not performed.
7.13 Log
Click “Log” to enter Log interface, as shown in Figure 7-35.
Count:
Params:
115
Chapter 7 Daily Operation
Faults:
Others:
All logs:
7.14 Status
Click “Status”to enter Status interface, see Figure 7-36.
116
Chapter 8 Quality Control
8.1 Overview
There are probably some test errors in the process of long term use of
hematology analyzer. In order to maintain accurate count and analysis and
eliminate system errors, it’s necessary to perform quality control (QC).
It’s recommended to use low, normal and high control materials to perform QC
respectively every day. At least run the normal control in daily use. When using
control materials of new batch, please use it together with the existing control
materials for 5 days, twice per day, and the results should be within the range
of parameters of the control instruction.
It suggests use URIT control materials and run QC in the following conditions.
After calibration
All the clinical specimens, control materials and calibrators may contain
human blood or serum and have potentially infectious hazards, so
operators should comply with the safe operation provisions in laboratory
and wear personal protective equipment (lab coats, glasses, gloves etc.)
when handling these materials.
117
Chapter 8 Quality Control
NOTE
3. And gently mix the test tube upside down for a dozen times.
1. L-J QC
__ X i
X i 1
n
1 n __
2
SD Xi X
n 1 i 1
SD
CV __
100%
X
118
Chapter 8 Quality Control
2. X-R QC
3. X QC
X QC is the variation of X-R QC. They have the same basic principle. The
difference is that the control dot in X graph indicates the mean value of two
__
values other than one value. On this foundation, it calculates the X , SD and
CV.
4. X-B QC
QC Mode Selection
119
Chapter 8 Quality Control
For blood routine QC, there are four modes: L-J, X-B, X-R and X.
8.3 L-J QC
Click “QC” to enter QC interface, see Figure 8-2. L-J QC is the default mode.
Click “New” in default “L-J QC” interface. Input Lot NO., expiry data, level
(high, normal,low) and rule(3SD、2SD、1SD) according to the manual, see
Figure 8-3.
Click “Forward” after inputting to enter the interface as shown in Figure 8-4.
Select the QC items you need in this interface by checking them. 6 items are
default. Then input corresponding parameters’ reference value and SD value.
Finally, click “Done”, and the QC new creation is finished.
120
Chapter 8 Quality Control
121
Chapter 8 Quality Control
122
Chapter 8 Quality Control
If you want to modify reference value for certain parameters or SD, click “Edit” ,
and the following dialog pops up, see Figure 8-6.
Move mouse to the box to modify it directly. Click “OK” complete modification,
and click “Cancel” to discard changes.
Select whole blood single sampling mode and click button. The interface
after counting is shown as Figure 8-7.
123
Chapter 8 Quality Control
8.4 X-B QC
X-B QC is different to others, only three parameters are edited, which are
MCV, MCH and MCHC. It is a QC without control materials and a means of
monitoring analyzer like controls, but they can’t replace each other.
Please run X-B QC when the quantity of samples is more than 100 everyday.
X-B QC is for the use of random sample, not for classified samples. Observed
the trend of QC result in reference range which is made up of reference, upper
124
Chapter 8 Quality Control
Click “QC” and “X-B QC” to enter corresponding interface, as shown in Figure
8-9.
125
Chapter 8 Quality Control
When finish QC edit, click “Count” to operate quality control in main menu. The
system automatically operates a QC rest after analyzing 20~200 samples
(according to your setting), and get a QC point that corresponds with each
reference of X-B QC and save it in X-B QC graph and X-B QC list.
Operator can review QC results of three parameters through graphs. After the
count of group samples completed, the results of MCV, MCH and MCHC
depict a dot on the graph. For example, the “Open X-B” is chosen and
“Records/Group” is 20. After the subsequent 20 counts, the system calculates
a X-B QC value and a corresponding control dot which displays on the graph.
There are three graphs of MCV, MCH and MCHC. The graphs updates at once
after each QC count. QC results are arranged in graphs according to storage
time. The latest is on the left side and its serial number is 1.
QC Graph Instruction
126
Chapter 8 Quality Control
3. The above line of every parameter graph means Reference plus SD.
4. The below line of every parameter graph means Reference value subtract
SD.
Significance of QC point:
If the QC point is between upper and lower limit of the corresponding graph, it
means that it’s under control range, If not, the QC point is not under control
range.(see Figure 8-11)
8.5 X-R QC
X-R QC is a quality control mode with control materials. Control material is
sucked in instrument for QC test. If running a blank test, the system defaults
the QC results invalid.
1. Click “QC” in main interface, then click “X-R QC”-“New” to enter X-R QC
creation interface. (See Figure 8-12).
3. Input “Lot No.”, and select “Expiry date” according to Operation Manual.
127
Chapter 8 Quality Control
4. Click “Forward” and tick items you need, as shown in Figure 8-13.
128
Chapter 8 Quality Control
After QC edit, operator can select blood routine in whole blood single sampling
After QC count, click “X-R QC”to enter QC analysis interface, see Figure 8-14.
Unlike L-J QC,each dot at X-R QC Graph indicates mean value or range of two
times QC results. The system cannot display low, normal and high control
graphs simultaneously in one interface, please select Level to change.
129
Chapter 8 Quality Control
In X-R QC interface, there are X graph and R graph. X graph displays the
mean value dot while the R graph displays the range dot.
If operator selects “Low” and do QC test twice, the dot is within X graph
corresponding with low level. It also fits for the QC of other groups—the dot
correspond with range is within corresponding R graph.
b) middle line —— X
R Graph Instruction
b) middle line —— R
130
Chapter 8 Quality Control
Significance of QC dot:
If the control dot falls in the area between above and below lines, it means the
dot is under control range. If not, the dot is not under control range.
8.6 X QC
X-R QC is a quality control mode with control materials. Control material is
sucked in instrument for QC test. Operator can do QC test for multiple
parameters. Three groups of files are offered to facilitate QC parameters and
results saving.
8.6.1 X QC Edit
1. Click “QC” in main interface, then click “XQC”-“New” to enter XQC creation
interface. (See Figure 8-13).
3. Input control material’s “Lot No.”, and select “Expiry date” according to
Operation Manual.
6. Click “Forward” button and tick items you need, as shown in Figure 8-16.
131
Chapter 8 Quality Control
132
Chapter 8 Quality Control
8.6.2 X QC Running
After QC edit, operator can select blood routine in whole blood single sampling
mode to enter QC running interface, as shown in Figure 8-17.
133
Chapter 8 Quality Control
If running a blank test, the system defaults the QC results invalid, and displays
0.
The same as L-J, the QC results are produced after QC running. See Figure
8-18.
134
Chapter 8 Quality Control
Unlike L-J QC, each dot at XQC Graph indicates mean value of two times of
QC results. If operator selects “Low” and do QC count twice, the dot
corresponding to mean value appears in graph area. It also fits for other types
of QC count.
QC Graph Instruction
135
Chapter 8 Quality Control
Significance of QC dot
If the control dot falls in the area between above and below lines, it means the
dot is under control range. If not, the dot is not under control range.
136
Chapter 9 Calibration
9.1 Overview
Analyzer is detected and calibrated before delivery. For some reasons the
result may be a little out of the range. Calibration is to insure the accuracy of
results. Calibration is a process to standardize the analyzer by its deviation of
value and parameter, calibration factor.
The analyzer provides blood routine calibration mode. Blood routine calibration
mode includes “Standard Cal”, “Blood Cal” and “Manual Cal”. L-J, X, X-R and
X-B QC graph can be drawn in blood routine mode.
WARNING
Make sure the calibrators are brought to room temperature and well mixed
slowly before use.
Never apply to the laboratory or clinic use unless all the parameters are
accurately calibrated.
NOTE
137
Chapter 9 Calibration
MCV and HCT are relative parameters to each other, thus one can be obtained
from given value of the other. Only MCV can be calibrated by the analyzer.
Usually the manufacturer gives the reference value of MCV.
All the clinical specimens, control materials and calibrators may contain
human blood or serum and have potentially infectious hazards, so
operators should comply with the safe operation provisions in laboratory
and wear personal protective equipment (lab coats, glasses, gloves etc.)
when handling these materials.
9.3 Preparation
Before calibration, inspect the analyzer as the following requirements:
(1) Ensure the reagents are in the shelf life, adequate and uncontaminated.
(2) Run a blank test and make sure the results are accordance with the
following ranges.
Parameter Range
WBC ≤0.2×109/L
RBC ≤0.02×1012/L
HGB ≤1g/L
PLT ≤10×109/L
HCT ≤0.5%
WBC-BF ≤0.001×109/L
138
Chapter 9 Calibration
RBC-BF ≤0.003×1012/L
(4) Verify the accuracy of precision. Test with mid-value control material or
human blood 11 times continuously.Take the results from the second to the
eleventh, and check the precision in “Data” interface. Make sure the CVs
are accordance with the range in Table 9-2 and 9-3.
Precision of whole
Parameter Measurement Range blood sample
(CV/d※)
(CV/SD)
139
Chapter 9 Calibration
(5) Conduct the test in the way described in Section 6.7 of YY/T 0653-2017.
Take a sample or calibrator whose concentration is in the high
concentration range in Table 9-4. Mix it well and test it 3 times continuously.
The test values are H1, H2 and H3. Take another sample or calibrator
whose concentration is in the low concentration range in Table 9-4. Test it
3 times continuously. Test values are H1,H2 and H3. Their results should
be in accordance with those in Table 9-5.
L1 L3
CR 100 %
H 3 L3
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
PLT ≤1.0%
HCT ≤0.5%
140
Chapter 9 Calibration
1.Set the test mode as whole blood single sampling mode and do
calibrator count at least three times to obtain mean.
2.Click “Cal” in main interface. Enter Manual mode by default. See Figure
9-1.
141
Chapter 9 Calibration
(1) Test the calibrators three times at least, and check whether the results are
within the allowed range.
(2) Test control material in “High”, “Normal” and “Low” value, and each
concentration should be tested at least three times. Check whether the
results are within the allowed range.
(3) Analyze three normal fresh blood samples, three times for each at least.
And check whether the results are within the allowed range.
2.Select the parameters you need and tick them, then click “Next”.
4.Select whole blood single sampling mode, and press “Count” to start
calibration. The analyzer could count 11 times at most to get mean. It’s
recommended to count 3 to 5 times at least. See Figure 9-5.
142
Chapter 9 Calibration
143
Chapter 9 Calibration
144
Chapter 9 Calibration
(1) Test the calibrators three times at least, and check whether the results are
within the allowed range.
(2) Test control material in “High”, “Normal” and “Low” value, and each
concentration should be tested at least three times. Check whether the
results are within the allowed range.
(3) Analyze three normal fresh blood samples, three times for each at least.
And check whether the results are within the allowed range.
Mean value=(value1+value2+value3+value4)/times
For example, the PLT reference value of the calibrator is 220, current
calibration value is 103% and test results is 230, thus the new calibration
value=103%×220/230=98.52%
NOTE
Click “Cal” in main interface to enter calibration interface. Select “Blood Cal”.
See Figure 9-7.
145
Chapter 9 Calibration
(3) Test each of the prepared samples at least three times to get the
mean and take the mean value as the reference value, or test and
calculate according to the reference method, taking the results as the
reference value.
(4) Select whole blood single injection mode, and press “Count” to
start calibration. The analyzer could count 11 times at most to get mean.
It’s recommended to count 3 to 5 times at least. See Figure 9-9.
Note: WIC refers to the WBC results produced by impedance method. WOC
refers to WBC results produced by optical method.
146
Chapter 9 Calibration
147
Chapter 9 Calibration
NOTE
Click “Cal” in main interface to enter calibration interface. Select “Cal His”. See
Figure 9-10.
Former calibration coefficients are recorded in History. User can view the
calibration time and calibration coefficient last time. If the calibration coefficient
need to be recovered to the former, select it and click “Replace” on the top left
corner. See Figure 9-11.
148
Chapter 9 Calibration
Calibration only applies to routine mode, but not RETIC or body fluid mode.
Clicking calibration button in these two modes, it will prompt that the current
mode do not need calibration, as shown in Figure 9-12.
149
Chapter 10 Service
10.1 Overview
Routine care and regular maintenance are essential to keep the best status
and precision, and to minimize system problems and extend its life.
Procedures and instruction for preventive maintenance are discussed in this
chapter. More information is available at URIT after-sale department.
150
Chapter 10 Service
151
Chapter 10 Service
The analyzer offers the following maintenance operations for fluid system.
NOTE
152
Chapter 10 Service
Operation Procedures:
Operation Procedures:
Operation Procedures:
153
Chapter 10 Service
Operation Procedures:
10.3.5 Flush
Operation Procedures:
Operation Procedures:
3. Click “OK”, and analyzer starts to soak. The remaining time is played
in screen.
154
Chapter 10 Service
NOTE
Operation Procedures:
1. Take out the diluent inlet tube connecting with the “DELUENT” on
the side panel from container, and put it into deionized water.
2. Take out the lyse inlet tube connecting with the “LYSE” on the side
panel from container, and put it into deionized water.
3. Take out the detergent inlet tube connecting with the “DETERGENT”
on the side panel from the container, and put it into deionized water.
4. Take out the sheath inlet tube connecting with the “SHEATH” on the
side panel from container, and put it into deionized water.
5. Screw the caps of the remaining reagents’ containers and store them
according to instructions. Operator should establish and confirm to the
effective storage measures to prevent reagent from deterioration, misusage or
misdrinking. The reagent should be away from temperature extremes.
8. Move away waste tube, turn off the power of analyze and computer.
WARNING
All the clinical specimens, control materials and calibrators may have
potentially infectious hazards, so operators should comply with the safe
operation provisions in laboratory and wear personal protective equipment
(lab coats, glasses, gloves etc.) when handling these materials.
155
Chapter 10 Service
Sample injection mechanism After 6000 sample tests Grease, brush, cloth
Auto sampling module After 6000 sample tests Grease, brush, cloth ,
cross screwdriver
cross screwdriver
Waste filter for WOC cup After 6000 sample tests Probe detergent,
cross screwdriver
Waste filter for WIC cup After 6000 sample tests Probe detergent,
cross screwdriver
Waste filter for RBC cup After 6000 sample tests Probe detergent,
cross screwdriver
Click “Statistics” in data interface, and you can select time intervals of start
time and end time. Select “All” for query type. Click “Statistics” button and the
number of test times is shown. Users can check installation time to verify
maintenance opportunity.
156
Chapter 10 Service
cross screwdriver
cross screwdriver
cross screwdriver
Waste filter for WOC cup After 60000 sample tests Tweezers,
cross screwdriver
Waste filter for WIC cup After 60000 sample tests Tweezers,
cross screwdriver
Waste filter for RBC cup After 60000 sample tests Tweezers,
cross screwdriver
cross screwdriver
Click “Statistics” in data interface, and you can select time intervals of start
time and end time. Select “All” for query type. Click “Statistics” button and the
number of test times is shown.
All the clinical specimens, control materials and calibrators may contain
human blood or serum and have potentially infectious hazards, so
operators should comply with the safe operation provisions in laboratory
and wear personal protective equipment (lab coats, glasses, gloves etc.)
when handling these materials.
157
Chapter 11 Troubleshooting
11.1 Overview
This chapter gives instructions for identifying and troubleshooting. If the
malfunction is not solved by this guidance, or if more detailed information is
needed, please contact URIT after-sale department.
NOTE
11.2 Guidance
Troubleshooting Guidance is used to assist operator in identifying and
resolving analyzer problems. Instruction is also given for obtaining technical
assistance immediately from URIT after-sale department. The first step in the
process is to understand normal analyzer operation and preventive
Maintenance. Good experience of the analyzer is essential for identifying and
resolving operational problems.
(3) Troubleshooting
Step1Problem Confirmation
Step2Problem Classification
158
Chapter 11 Troubleshooting
Step3Troubleshooting
Engineers take appropriate action to deal with the problem. If operator can
deal with it by himself or with URIT engineer’s assistance, the efficiency of
troubleshooting can be increased.
(4) The lot number of the reagents (lyse, diluent, detergent etc.)
Common faults and disposals are in this chapter. The operator can identify the
cause according to the warning information and operate according to
Troubleshooting Guidance.
11.4 Troubleshooting
Common faults and corrective actions are listed as follows. If the problems
cannot be dealt with, and if technical assistance is needed, please contact our
after-sale department.
159
Chapter 11 Troubleshooting
160
Chapter 11 Troubleshooting
6.Motor guide rod is not 4.If the fault still exists, please contact our
lubricating enough. after-sale department.
6.Motor guide rod is not 4.If the fault still exists, please contact our
lubricating enough. after-sale department.
6.Motor guide rod is not 4.If the fault still exists, please contact our
lubricating enough. after-sale department.
161
Chapter 11 Troubleshooting
162
Chapter 11 Troubleshooting
163
Chapter 11 Troubleshooting
Host CAN bus 1.Host CAN bus and fluid 1.Please restart instrument or connect with
connecting fluid system module bus poor software again.
system module contact.
2.If the fault still exist, please contact our
164
Chapter 11 Troubleshooting
165
Chapter 11 Troubleshooting
1.Expired detergent.
Detergent service time 2. Replace detergent and activate card
Expired detergent exceeds the specified following software instruction. Instrument
time. primes detergent and alarm disappears
automatically.
166
Chapter 11 Troubleshooting
1.Expired diluent.
1.Expired sheath.
1.Expired lyse.
167
Chapter 11 Troubleshooting
is bend.
168
Chapter 11 Troubleshooting
service engineer.
1.Aperture is clogged.
1. Click "Solve", and the instrument flushes
2.Tubes are bent.
automatically.
WBC Clog 3.Reagent replacement
2.If the fault still exists, please contact
error.
service engineer.
4.Solenoid valve problem.
169
Chapter 11 Troubleshooting
1.Insufficient liquid in
sample cup front
chamber/ after chamber. 1. Click "Solve", and the instrument flushes
automatically.
WBC Bubble 2.Tubes joint is leaky.
2.If the fault still exists, please contact
3.Reagent replacement
service engineer.
error.
1.Aperture is clogged.
1. Click "Solve", and the instrument flushes
2.Tubes are bent.
automatically.
RBC Clog 3.Reagent replacement
2.If the fault still exists, please contact
error.
service engineer.
4.Solenoid valve problem.
1.Insufficient liquid in
sample cup front
chamber/ after chamber. 1. Click "Solve", and the instrument flushes
automatically.
RBC Bubble 2.Tubes joint is leaky.
2.If the fault still exists, please contact
3.Reagent replacement
service engineer.
error.
170
Chapter 11 Troubleshooting
171
Chapter 11 Troubleshooting
172
Chapter 11 Troubleshooting
173
Appendix A Specifications
A.1.1 Parameters
Immature Reticulocyte
IRF %
Fraction Percent
Mean Corpuscular
MCH pg
Hemoglobin
174
Appendix B Communication Protocol
Mean Corpuscular
MCHC g/L
Hemoglobin Concentration
PCT Plateletcrit %
175
Appendix B Communication Protocol
Abnormal lymphocyte
Permillage
Abnormal lymphocyte
ALY# 109 /L
count
A.1.3 QC Mode
L-J ;X-B
X -R ;X
A.1.4 Reagents
176
Appendix B Communication Protocol
The analyzer provides blood routine calibration mode. Blood routine calibration
mode includes “Standard Cal”, “Blood Cal” and “Manual Cal”. L-J, X, X-R and
X-B QC graph can be drawn in blood routine mode.
External Computer
External Printer
CAUTION
AC 100V-240V 50/60 Hz
177
Appendix B Communication Protocol
Temperature:15℃~35℃;
Relative humidity:30%~85%;
Atmospheric Pressure:60kPa~106kPa;
Temperature:-20℃~55℃;
Barometric Pressure:50kPa~106kPa;
A.2.5 Waste
178
Appendix B Communication Protocol
A.2.8 Diameter
WBC:100μm
RBC/PLT:68μm
A.3.1 Prevision
Precision of whole
Parameters Measurement Range blood sample
(CV/d※)
179
Appendix B Communication Protocol
0.000×1012/L~0.100×1012/L ≤±0.010×1012/L
≥0.800
or±5%
RBC-BF
0.101×1012/L~5.000×1012/L ≤±0.030×1012/L
≥0.800
or±2%
180
Appendix B Communication Protocol
The mean value of NEU, LYM, MON, EOS and BASO test results should be
≥99% upper limit or ≤99% lower limit of credible range.
A.3.4 Carryover
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
PLT ≤1.0%
WBC ≤0.2×109/L
RBC ≤0.02×1012/L
HGB ≤1g/L
PLT ≤10×109/L
HCT ≤0.5%
WBC-BF ≤0.001×109/L
RBC-BF ≤0.003×1012/L
A.3.6 Accuracy
181
Appendix B Communication Protocol
35%~50%(HCT)or
HCT or MCV ≤ ±3.0%
80fL~100fL(MCV)
WBC 0~999.99×109/L
RBC 0~99.99×1012/L
HGB 0~999g/L
HCT 0%~80%
PLT 0~9999×109/L
Name Specification
CAUTION
182
Appendix B Communication Protocol
Whole venous
21 0 7 0.4
blood CBC
Whole venous
22 16.5 8 0.4
blood CBC+5D
Switch to body
60 70 12 1
fluid mode
Power on self
70 48 27 1
test
Clean and
40 27 20 0
shutdown
Prime diluent 94 0 6 0
Prime sheath 0 44 0 0
Prime detergent 0 0 24 0
Complete soak 40 42 12 0
Wake 0 0 0 0
Note: The table above is the consumption of the latest liquid path version
V1.00.23.0330, and the consumption varies with the version of the instrument.
Subsequent changes in consumption due to upgrade of the instrument version
will not be reflected in this manual.
183
Appendix B Communication Protocol
A.6 Contraindications
None
Alarm
Interpretation Measures
information
184
Appendix B Communication Protocol
185
Appendix B Communication Protocol
DFLT:Leukocyte
categories abnormal
(NLMEB:N=NEU, L
=LYM, M=MON,
E=EOS, B=BASO)
The following
conditions may cause
DFLT alarm:
1.There is no
detection of a low trough
Check stain smear and the verified
used to distinguish
Abnormal WBC subgroup’s categorical measure by
between the two cell
differential count description box in accordance with your
populations areas.
laboratory inspection standard.
2. The number of cells
in a specific subgroup is
abnormally low. Such as
WBC increase &
decrease, neutrophil
increase & decrease,
lymphocytes increase &
decrease, eosinophils
increase & decrease,
basophils increase &
decrease
186
Appendix B Communication Protocol
inspection standard.
Height of histogram at
parting line is over 20.
Check stain smear to see if there is
Height of histogram at
Abnormal RBC abnormal RBC value or PLT morphology
parting line is over 20.
size distribution in accordance with your laboratory
Histogram between inspection standard.
parting lines distributes
double peaks.
Histogram between
parting lines distributes Check stain smear to see if there is
Bimodal RBC double peaks. abnormal RBC value or PLT morphology
distribution in accordance with your laboratory
RBC histogram has inspection standard.
two or more histograms.
187
Appendix B Communication Protocol
188
Appendix B Communication Protocol
189
Appendix B Communication Protocol
190
Appendix B Communication Protocol
<SB>information<EB><CR>
hexadecimal 0x0B
Hexadecimal 0x1C
hexadecimal 0x0D
B. Information Grammar
1. Delimiter
~ ---Repeat Delimiter
2. Data Type
CE code element
CM composite
191
Appendix B Communication Protocol
DT data
EI entity identifier
HD hierarchic designator
FN family name
FT formatter text
NM numeric
PT processing type
PL person location
ST string
SI sequence ID
TS time stamp
TQ timing quantity
TX text data
3. Field Meaning
192
Appendix B Communication Protocol
8 Security ST 40 Security
11 Processing ID PT 3 Dispose of ID P
Product
Type
14 Retain
15 Retain
16 Retain
17 Retain
18 Encoder ST Encoding is
UNICODE
193
Appendix B Communication Protocol
MSH-10 Description
Example.
MSH|^~\&|URIT|UT-5390|LIS|PC|20100930100436||ORU^R01|0001|P|2.3.1|1
|||||UNICODE
194
Appendix B Communication Protocol
Example. PID|1|1010051|A1123145|15|Mary||19811011|M
195
Appendix B Communication Protocol
Example:
OBR|1|1010051|000001|URIT^UT-5390||20101010093000||20101010093
500||sender||| diagnosis^remark||BLD|Inspector||||||||||||verifier|
3.5. OBX
196
Appendix B Communication Protocol
6 Units CE 90 Unit
Example. OBX|1|NM|WBC||8.21|10^9/L|4.00-10.00|L|||F||
3.6. MSA
197
Appendix B Communication Protocol
2 Message Control ID ST 20
4 Expected Sequence NM 15
Number
5 Delayed ID 1
Acknowledgment
Type
198
Appendix B Communication Protocol
processing id ID
3.7. ERR
ERR-1
001 Record Test tube No. The test tube record has
already exist already existed.
002 Lis Recieved Test tube No. Lis receiving error, resending
Faild data is required.
003 Read REQ Test tube No. Fail to read request form.
error
004 Read Test tube rack Analyzer fails to read test tube
BarCode No. number.
Errer
3.8. QRD
199
Appendix B Communication Protocol
3.9. QRF
200
Appendix B Communication Protocol
3.10. QSP
1 Set ID - DSP 4 SI
2 Display Level SI 4
5 Result ID TX 20
2 Serial Number
3 Name
4 Sex
5 Birthday
6 Blood Type
7 Group
201
Appendix B Communication Protocol
8 Patient Number
9 Bed Number
10 Patient Type
11 Department
12 Sender
13 Inspector
14 Verifier
16 Clinical diagnosis
17 Remark
19 inspection time
Example
DSP|1||Mary||<CR>
4. Communication process
URIT- Lis
ORU^R0
5390 server
<SB>
MSH
PID
PV1
OBR
202
Appendix B Communication Protocol
OBX
OBX
……
<EB><CR>
For example:
<SB>
MSH|^~\&|URIT|UT-5390|LIS|PC|20110627144458||ORU^R01|0001|P|2.
3.1||||||UNICODE<CR>
PID|1||||||||<CR>
PV1|1|||<CR>
OBR|1||BAR101010101|URIT^UT-5390||||01110621143134|||||^||||||||||||||||<
CR>
OBX|1|NM|WBC||110.0|10^9/L|40.0-100.0|H|||F|||||||<CR>
OBX|2|NM|LYM||35.57|%|20.00-40.00||||F|||||||<CR>
OBX|3|NM|MON||5.84|%|3.00-8.00||||F|||||||<CR>
OBX|4|NM|NEU||57.37|%|50.00-70.00||||F|||||||<CR>
OBX|5|NM|EOS||1.14|%|0.50-5.00||||F|||||||<CR>
OBX|6|NM|BASO||0.08|%|0.00-1.00||||F|||||||<CR>
OBX|7|NM|LYM#||284.5|10^9/L|80.0-400.0||||F|||||||<CR>
OBX|8|NM|MON#||46.7|10^9/L|10.0-80.0||||F|||||||<CR>
OBX|9|NM|NEU#||458.9|10^9/L|200.0-700.0||||F|||||||<CR>
OBX|10|NM|EOS#||9.1|10^9/L|0.0-50.0||||F|||||||<CR>
OBX|11|NM|BASO#||0.6|10^9/L|0.0-10.0||||F|||||||<CR>
203
Appendix B Communication Protocol
OBX|12|NM|RBC||4.49|10^12/L|3.50-5.50||||F|||||||<CR>
OBX|13|NM|HGB||0|g/L|0-1079738368|L|||F|||||||<CR>
OBX|14|NM|HCT||26.4|%|37.0-50.0|L|||F|||||||<CR>
OBX|15|NM|MCV||59.0|fL|80.0-100.0|L|||F|||||||<CR>
OBX|16|NM|MCH||24.0|pg|27.0-31.0|L|||F|||||||<CR>
OBX|17|NM|MCHC||0|g/L|0-1081344000|H|||F|||||||<CR>
OBX|18|NM|RDW_CV||16.1|%|11.5-14.5|H|||F||||||<CR>
OBX|19|NM|RDW_SD||45.0|fL|35.0-56.0||||F||||||<CR>
OBX|20|NM|PLT||0|10^9/L|0-1079574528|H|||F|||||||<CR>
OBX|21|NM|MPV||12.3|fL|7.0-11.0|H|||F|||||||<CR>
OBX|22|NM|PDW||14.7|fL|15.0-17.0|L|||F|||||||<CR>
OBX|23|NM|PCT||0.41|%|0.10-0.28|H|||F|||||||<CR>
OBX|24|NM|P_LCR||1.37|%|0.50-1.80||||F|||||||<CR>
OBX|25NM|RBCHistogram^LeftLine||1||||||F||||||<CR>
OBX|26|NM|RBCHistogram^RightLine||118||||||F||||||<CR>
OBX|27|ED|RBCHistogram||UT5390^Histogram^512Byte^HEX^0000000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||||||F||||||<CR>
OBX|28|NM|PLTHistogram^LeftLine||8||||||F||||||<CR>
OBX|29|NM|PLTHistogram^RightLine||127||||||F||||||<CR>
OBX|30|ED|PLTHistogram||UT5390^Histogram^256Byte^HEX^00000000
05050601010203040505060708090a0b0b0b0b0b0b0a0a0a0b0b0b0b0c0c0b
0b0a0a09080706060505050506060606060505050404030303030202020202
02020202020202020202020202020201010101010101020202020203030303
02020202010101010101020202020202020202020202020203030303030303
||||||F||||||<CR>
204
Appendix B Communication Protocol
OBX|31|ED|S0_S10DIFFScattergram||UT5390^Image^BMP^Base64^Qk
32lgMAAA……<CR>
OBX|32|ED|S90_S90DDIFFScattergram||UT5390^Image^BMP^Base64^
Qk32lgMAAA……<CR>
<EB><CR
205
Appendix C License for Manufacturing
Measuring Analyzers
206
Appendix D Toxic and Harmful Substances or
Elements
Toxic and Hazardous Substances or Elements
Polybromi
Parts Polybrominated
Plumbum Mercury Cadmium Chromium nated
Diphenyl Ethers
(Pb) (Hg) (Cd) VI(Cr(VI)) Biphenyls
(PBDE)
(PBB)
Shell ○ ○ ○ ○ ○ ○
Printed
circuit
○ ○ ○ ○ ○ ○
board
Assembly
Sheet
○ ○ ○ ○ ○ ○
metal Parts
Plastic
Host ○ ○ ○ ○ ○ ○
Parts
Machining
○ ○ ○ ○ ○ ○
parts
Hardware ○ ○ ○ ○ ○ ○
Flow
System ○ ○ ○ ○ ○ ○
Parts
Cable ○ ○ ○ ○ ○ ○
Accessories ○ ○ ○ ○ ○ ○
Packaging
○ ○ ○ ○ ○ ○
Materials
207
Appendix D Toxic and Harmful Substances or Elements
×:The content of toxic or hazardous substance is exceed the acceptable range of GB/T
26572 in at least one kind of homogeneous material of the parts above.
(The circuit board used lead solder in machining process and sonme parts of the board
contain plumb;And some sheetmetal parts use chromium VI for surface )
208
Appendix E Daily Operation Procedures
(1) Make sure the power cord is properly connected. None reagent
tubes is bending or detached. Check if the waste container is full.
(2) Turn on the power of computer and analyzer.
(3) The analyzer starts to perform initialized self-checking program
automatically and rinse the fluid system, then goes to test Interface.
(4) Perform a blank test and QC to ensure the analyzer operates
normally.
(5) Analyze a group of samples or emergency sample.
(6) Query, output and print the data.
(7) Necessary maintenance should be operated according to the
situation.
2. Shutoff Procedures
(1) Click in the main interface and then “Rinse and Shutoff”.
(2) The analyzer automatically rinse the fluid system.
(3) Turn off the power switchon the rear panel of the instrument.
4. Weekly Maintenance
209
Appendix E Daily Operation Procedures
5. Monthly Maintenance
6. Other Maintenance
If the ruby aperture is clogged severely, put the tube filled with probe detergent
in emergency position and perform “Complete Soak”. The probe detergent is
injected automatically into the cup to soak counting hole. The blank test will be
done automatically after soaking to check it is still clogged or not.
210
Appendix F Key Components
INPUT: a.c.100V~240V
1 Power Supply
OUTPUT:DC+5V/+12V/+24V
211
Appendix G Accessories List
3 Reticle Piece 1
12 Grease Piece 1
13 Filter Piece 2
212