Professional Documents
Culture Documents
Complete Form 7 to verify the operational qualification for the ACQUITY UPLC:
• Column manager-A
• Column manager-Aux
• Column manager
• Column heater-A
• Column heater
• HT column heater
• Column heater/cooler
• 30-cm column heater-A
• 30-cm column heater
For information about any steps in this form, see page 4.
Owner’s System ID
Complete the applicable sections of this form for the column heating/cooling device(s) in your system.
For column device(s) not in your system, enter N/A on each serial number line.
Owner’s System ID
Complete the applicable sections of this form for the column heating/cooling device(s) in your system.
For column device(s) not in your system, enter N/A on each serial number line.
Temperature Accuracy Test for Column Manager, Column Heater-A, Column Heater, HT Column Heater,
Column Heater/Cooler, 30-cm Column Heater-A, or 30-cm Column Heater/Cooler
Comments
Certification
The undersigned performer attests that the test results recorded above are accurate and complete.
Performer Date
Signature
The undersigned reviewer accepts that the test results recorded above are accurate and complete.
Reviewer Date
Signature
2011 Waters Corporation
Recommended materials
• Calibrated thermometer probe, with readout device
• Tape, to hold the thermometer probe in place, such as Kapton tape (part number
410001630)
4. Ensure that the thermometer probe tip touches the inside of the column compartment.
Requirement: Use tape to hold the tip against the compartment.
5. Close the front panel of the column manager-A or column manager-Aux.
6. In the ACQUITY Console window, specify one set-point value between 4 °C and 90 °C
(inclusive), and record this value in Form 7.
Restriction: Use no value that is within ±5 °C of ambient temperature. Most users choose a
temperature that corresponds to the setting they most frequently set or a temperature at the
extreme end of the range: for example, 4 °C or 90 °C.
7. Allow the column manager-A or column manager-Aux temperature to reach and stabilize at
the set temperature.
8. In Form 7, record the temperature you observe from the thermometer probe.
Tip: The time required to reach the set temperature varies depending on the difference
between the set and ambient temperatures.
9. Repeat step 3 through step 7 for the lower trough in the column manager-A or column
manager-Aux.
10. Return the column manager-A or column manager-Aux temperature to 40 °C.
11. Complete Form 7.
Recommended Materials
• Calibrated thermometer probe, with readout device
• Kapton® tape, (part number 410001630) to hold the thermometer probe in place
Note:
• This procedure is not for the ACQUITY UPLC Column Heater module, which holds only
one column. See “Testing the ACQUITY UPLC Column Heater Temperature Accuracy”
on page 7 if you have the ACQUITY UPLC Column Heater module.
• This procedure is not for the ACQUITY UPLC High Temperature Column Heater
module, which holds only one column. See “Testing the ACQUITY UPLC High
Temperature (HT) Column Heater Temperature Accuracy” on page 8 if you have the
ACQUITY UPLC Column Heater module.
Recommended materials
• Calibrated thermometer probe, with readout device
• Tape to hold the thermometer probe in place, such as Kapton tape (part number
410001630)
4. Ensure that the thermometer probe tip touches the inside of the column tray.
Requirement: Use tape to hold the tip against the tray.
5. Close the front panel of the column heater-A.
6. In the ACQUITY Console window, specify a set-point value from 5 °C above ambient
temperature to 90 °C, and record this value in Form 7.
7. Allow the column heater-A temperature to reach and stabilize at the set temperature. In
Form 7, record the temperature you observe from the thermometer probe.
Tip: The time required to reach the set temperature varies depending on the difference
between the set and ambient temperatures.
Recommended Materials
• Calibrated thermometer probe with readout device
• Kapton tape, (part number 410001630) to hold the thermometer probe in place
Note:
• This procedure is not for the ACQUITY UPLC Column Manager or ACQUITY UPLC
Column Heater/Cooler that holds four columns. See “Testing the ACQUITY UPLC
Column Manager or Column Heater/Cooler Temperature Accuracy” on page 5 if you
have that module.
• This procedure is not for the ACQUITY UPLC High Temperature Column Heater
module, which holds only one column. See “Testing the ACQUITY UPLC High
Temperature (HT) Column Heater Temperature Accuracy” on page 8 if you have the
ACQUITY UPLC Column Heater module.
5. Tilt the column tray upward to the closed position. Ensure that the wire from the
thermometer probe is routed through the notch on the right-hand side of the column tray.
6. Depending on your type of column heater, refit the front panel or swing it closed.
7. Use the ACQUITY Console screen for the column heater to specify a value from 5 °C above
ambient temperature to 65 °C. Record this set-point value in Form 7.
8. Allow sufficient time for the column heater’s temperature to reach and stabilize at the set
temperature. In Form 7, record the temperature you observe from the thermometer probe.
Note: The time required to reach the set temperature varies depending on the difference
between the set and ambient temperatures.
9. After completing the column heater temperature accuracy test, return the column heater
temperature to 40 °C.
10. Complete Form 7.
Recommended Materials
• Calibrated thermometer probe with readout device
• Kapton tape, (part number 410001630) to hold the thermometer probe in place
• This procedure is not for the ACQUITY UPLC Column Heater that holds one column and
heats to a maximum of 65 °C. See “Testing the ACQUITY UPLC Column Heater
Temperature Accuracy” on page 7 if you have the ACQUITY UPLC Column Heater
module.
• This procedure is not for the ACQUITY UPLC Column Manager or ACQUITY UPLC
Column Heater/Cooler that holds four columns. See “Testing the ACQUITY UPLC
Column Manager or Column Heater/Cooler Temperature Accuracy” on page 5 if you
have that module.
Requirement: The temperature accuracy of the high temperature (HT) column heater at the
temperature sensor is ±1.0 °C up to 65 °C, and ±2.0 °C above 65 °C up to 90 °C. To
compensate for typical measurement inaccuracy, the HT column heater Acceptable Range
line in Form 7 specifies a variation tolerance of ±2 °C up to 65 °C, and ±3 °C above 65 °C
up to 90 °C. If your measurement device has an inaccuracy of greater than ±1 °C, you can
adjust the acceptable range accordingly. If you do so, record the adjustment in the
Comments section.
Note: The time required to reach the set temperature varies depending on the difference
between the set and ambient temperatures.
Recommended materials
• Calibrated thermometer probe with readout device
• Tape to hold the thermometer probe in place, such as Kapton tape (part number
410001630)
Requirement: Temperature accuracy of the 30-cm column heater-A at the temperature sensor is
±1.0 °C. To compensate for typical measurement inaccuracy, the Acceptable Range line in
Form 7 specifies a variation tolerance of ±2 °C. When your measurement device has an
inaccuracy greater than ±1 °C, adjust the acceptable range accordingly. If you do so, record
the adjustment in the form’s Comments section.
Caution: To prevent damaging the probe wire, ensure it does not catch in the
closing door.
4. Close the column-heater door, being careful to avoid damaging the probe wire.
5. In the ACQUITY Console window, specify one set-point value between 20 °C and 90 °C
(inclusive), and record this set-point value in Form 7.
Restriction: Use no value that is within ±5 °C of the ambient temperature. Most users
choose a temperature that corresponds to the setting they most frequently set or a
temperature at the extreme end of the range: for example, 20 °C or 90 °C.
6. Allow the 30-cm column heater-A temperature to reach and stabilize at the set
temperature. In Form 7, record the temperature you observe from the thermometer
probe.
Tip: The time required to reach the set temperature varies depending on the difference
between the set and ambient temperatures.
7. After completing the temperature accuracy test, return the 30-cm column heater-A
temperature to 40 °C.
8. Complete Form 7.
Recommended Materials
• Calibrated thermometer probe, with readout device
• Kapton tape, (part number 410001630) to hold the thermometer probe in place
Requirement: For Empower 2 (build 2154) you must have installed feature release 3,
minimum.
Exposed
Thermocouple
Covered by a
Shroud
If your ACQUITY UPLC System includes the 30-cm column heater/cooler option, use the following
procedure to verify its proper operation.
Requirement: The temperature accuracy of the 30-cm column heater/cooler at the temperature
sensor is ±0.8 °C. To compensate for typical measurement inaccuracy, the device’s
Acceptable Range line in Form 7 specifies a variation tolerance of ±2 °C. If your measurement
device has an inaccuracy of greater than ±1 °C, you can adjust the Acceptable Range
accordingly. If you do so, record the adjustment in the Comments section.