Professional Documents
Culture Documents
ACQUISITION INFORMATION
The Gain Linearity test determines that the detector has a linear response within the detector 's gain range. The
sample set injects one concentration from the same vial at five different gain settings of 1, 2, 5, 10, and 20. There
are two custom fields used to determine the gain linearity : Area_Gain (which is area/amount), and Height _Gain
(which is height /amount). The test passes if the % RSD of these 2 fields is < or = 4.0.
ACQUISITION INFORMATION
1000.00
900.00
800.00
700.00
600.00
Energy
500.00
400.00
300.00
200.00
100.00
0.00
0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00
Minutes
Gain 1
Gain 2
Gain 5
Gain 10
Gain 20
ACQUISITION INFORMATION
Name: anthracene
Name RT Area Amount Area_Gain Result Id
Mean 1113980.072
% RSD 2.76
Name: anthracene
Height
Name RT Amount Height_Gain Result Id
(µV)
1 anthracene 1.003 425588 1 425588.367 32608
Mean 441645.365
% RSD 2.12