Injector linearity verifies that the injection volume delivered is linear within specified criteria . Injector linearity
is determined by injecting four different injection volumes of one concentration from the same vial . Empower
plots the area versus the µL injected to determine both the injector linearity and accuracy .
The coefficient of determination (R^2) is used to measure injector linearity . The injector passes the linearity
test if R^2 > or = 0.9990.
The injector accuracy is obtained by dividing the y -intercept by the slope . Accuracy passes if the result is
within the specification of < or = 0.2 µL.
ACQUISITION INFORMATION
50.00
40.00
30.00
EU
20.00
10.00
0.00
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00
Minutes
Peak : anthracene
Injection Volume
SampleName Area Level Channel Id Result Id
(ul)
1 Anthracene 2 uL 2.00 413028 1 32625 32647
2 Anthracene 3 uL 3.00 636939 2 32631 32648
3 Anthracene 5 uL 5.00 1097687 3 32637 32649
4 Anthracene 6 uL 6.00 1331446 4 32643 32650
ACQUISITION INFORMATION
I n j e c tor L i n e a r i t y R e s u l t s
1.5x106
1.0x106
Area
5.0x105
0.0