The Excitation Wavelength Accuracy test uses the Rayleigh scattering process of the emission wavelength to verify
accuracy of the detector.
Emission wavelength is set to 700 nm and the excitation wavelength is changed in 1nm increments from 343 nm
to 357 nm to measure the second order Rayleigh scattering band.
The Excitation Wavelength Accuracy test passes only when all 15 expected peaks are identified, and the
wavelength associated with the highest peak is 350 nm + or - 6 nm.
Waters Project Name and Version: SQT_E_2695_2475 v4.00 © 2007 - 2017 Waters Corporation. All rights reserved.
Date: Monday, February 20, 2017
Printed: 11:05:50 AM US/Eastern Page: 1 of 2
EXCITATION WAVELENGTH ACCURACY REPORT
A C Q U I S I TI O N I N F O R M A TI O N
350.00
352nm - 6.979
351nm - 6.557
353nm - 7.582
350nm - 5.825
354nm - 8.322
349nm - 5.317
348nm - 4.717
300.00
355nm - 8.742
347nm - 4.134
346nm - 3.531
345nm - 2.946
356nm - 9.340
250.00
357nm - 9.944
344nm - 2.394
200.00
343nm - 1.891
EU
150.00
100.00
50.00
0.00