UV SYSTEM PRECISION REPORT

A C Q U I S I TI O N I N F O R M A TI O N

System Name: 2695_2998_ISM_QDa72 Acquired By: ryan


Processing Method: MS System Precision PM, UV Sample Kit 2 Lot Number: 1245
Project Name: SQT_E_2695_2998_QDa_P_ISM_RC2 Sample Kit 2 Exp. Date: 12/22/2014 12:00:00 AM
Sample Set Id: 5089 Column Serial Number: W1412341AD
Sample Set Name: System Precision SSM Result Set Id: 5105
Sample Set Start Date: 1/6/2015 5:11:03 PM EST

TE S T D E S C R I P TI O N
System precision is determined by running six injections of a caffeine solution at the same
concentration and injection volume. The standard deviation of the retention time and the %RSD
for Area and Height will be measured and should be minimal in a well-functioning system.
The %RSD of the area and height gives the injection precision, and the standard deviation
of the retention time is the system precision. System precision is a good indicator of many
factors including pump flow rate precision, solvent proportioning precision and temperature
precision.

The qualification passes if the Area %RSD is < or = to 1.0 and the Height %RSD is < or = 1.5,
and the Retention Time SD is < or = to 1.0.

Caffeine Area and Height Results


Area Height
Area Spec Area (% RSD) Height Spec Height (% RSD)
Pass/Fail Pass/Fail
1 %RSD < or = 1.0 0.08 PASS %RSD < or = 1.5 0.99 PASS

Caffeine Retention Time Results


Retention Time Retention Time Retention Time
Spec Standard Deviation (sec) Pass/Fail
1 SD RT < or = 1.0sec 0.31 PASS

Printed 8:23:30 AM US/Eastern 1/7/2015 Page: 1 of 3


Waters Project Name and Version: SQT_E_2695_2998_QDa_P_ISM Version 1.00
UV SYSTEM PRECISION REPORT
A C Q U I S I TI O N I N F O R M A TI O N

System Name: 2695_2998_ISM_QDa72 Sample Set Id: 5089


Project Name: SQT_E_2695_2998_QDa_P_ISM_RC2 Result Set Id: 5105

Caffeine Ov erlaid Chromatograms


1.00

0.90

0.80

0.70

0.60

0.50
AU

0.40

0.30
Caffeine - 0.728

0.20

0.10

0.00

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00
Minutes
Sample Name: Caffeine; Injection: 1
Sample Name: Caffeine; Injection: 2
Sample Name: Caffeine; Injection: 3
Sample Name: Caffeine; Injection: 4
Sample Name: Caffeine; Injection: 5
Sample Name: Caffeine; Injection: 6

Printed 8:23:30 AM US/Eastern 1/7/2015 Page: 2 of 3


Waters Project Name and Version: SQT_E_2695_2998_QDa_P_ISM Version 1.00
UV SYSTEM PRECISION REPORT
A C Q U I S I TI O N I N F O R M A TI O N

System Name: 2695_2998_ISM_QDa72 Sample Set Id: 5089


Project Name: SQT_E_2695_2998_QDa_P_ISM_RC2 Result Set Id: 5105

Chromatographic Peak Results


Retention Time Height
Inj ection Name Area Peak Type Result Id
(min) (µV)
1 1 Caffeine 0.728 4493693 954164 Found 5172
2 2 Caffeine 0.728 4493119 941340 Found 5173
3 3 Caffeine 0.734 4497304 939964 Found 5174
4 4 Caffeine 0.738 4491998 928862 Found 5175
5 5 Caffeine 0.737 4500058 932942 Found 5176
6 6 Caffeine 0.739 4491354 931378 Found 5177

Caffeine Summary Results


Retention Time
Area (% RSD) Height (% RSD)
Standard Deviation (sec)
1 0.31 0.08 0.99

Printed 8:23:30 AM US/Eastern 1/7/2015 Page: 3 of 3


Waters Project Name and Version: SQT_E_2695_2998_QDa_P_ISM Version 1.00