A C Q U I S I TI O N I N F O R M A TI O N
TE S T D E S C R I P TI O N
System precision is determined by running six injections of a caffeine solution at the same
concentration and injection volume. The standard deviation of the retention time and the %RSD
for Area and Height will be measured and should be minimal in a well-functioning system.
The %RSD of the area and height gives the injection precision, and the standard deviation
of the retention time is the system precision. System precision is a good indicator of many
factors including pump flow rate precision, solvent proportioning precision and temperature
precision.
The qualification passes if the Area %RSD is < or = to 1.0 and the Height %RSD is < or = 1.5,
and the Retention Time SD is < or = to 1.0.
0.90
0.80
0.70
0.60
0.50
AU
0.40
0.30
Caffeine - 0.728
0.20
0.10
0.00
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00
Minutes
Sample Name: Caffeine; Injection: 1
Sample Name: Caffeine; Injection: 2
Sample Name: Caffeine; Injection: 3
Sample Name: Caffeine; Injection: 4
Sample Name: Caffeine; Injection: 5
Sample Name: Caffeine; Injection: 6