TSX-A-03052
There are report of IGBT broken caused by loosen screw of fuse in the IGBT UNIT.
[Phenomenon]
Screws inside IGBT UNIT are loosen by quakes of transport.
At the result, IGBT break when Xray operation.
[affected model]
All HighVoltage Generator will be installed near future.
Reported device is (X)UD150B-30/L-30 only , but the cause is quakes of transport.
So, all generator is applicable.
However, this problem is already measured after shipping July,2003.
***Contents***
The screw at fuse of IGBT is changed from “M6/Length10” to “M6/Length12/washer”.
[treatment method]
Before installation, tighten screws inside of IGBT UNIT in tightly and check the fixing.