You are on page 1of 3

2011 8th International Multi-Conference on Systems, Signals & Devices

A POSTURAL STABILITY ANALYSIS


BY USING PLANTAR PRESSURE MEASUREMENTS

Samir Boukhenous and Mokhtar Attari

Laboratory of Instrumentation, Faculty of Electronics and Computers


USTHB, BP. 32, Bab-Ezzouar, Alger, Algeria, http://www.lins.usthb.dz
Email: sboukhenous@gmail.com, attari.mo@gmail.com
ABSTRACT transducers. However, foot-to-floor interaction is
commonly studied using force and pressure measurement
The foot and ankle contribute significantly to postural systems [3], [4]. Force platforms are widely used both in
control by providing a contact area to the support surface. posture and gait analysis. They reliably deliver all the
Persons with visual and neuromuscular disorders have components of the resulting ground reaction and the free
deficits in response to perturbations of balance and often moment; they also furnish the displacement of the
require assistive devices. This work investigated to instantaneous center of pressure but they do not return
posture analysis by using plantar pressure measurements any information related to foot location on the platform
coupled with an EMG analog signal processor. The or the local distribution of loads under the foot. Several
output signals from the plantar pressure and the EMG empirical solutions have been proposed in the literature
analog processor are feeding to a high resolution data to obtain this information essential for gait and posture
acquisition system of National Instrument data analysis [5]. Foot location is frequently determined using
acquisition card with PCI interfacing to a PC. a force platform in conjunction with an optoelectronic
Experimental tests have been carried out concerning the system, which detects a variable number of markers
postural stability conducted by subjects with eyes open placed in correspondence to anatomical points [6].
and closed. However, an analog circuit has been realized Simpler solutions are also adopted, especially in
for recording the EMG signal of Soleus muscle during posturographic analysis, in which markers placed on the
the experimentation tests. subject’s feet match reference lines on the platform.
Sometimes foot placement is determined from footprints
Index Terms— Foot and ankle, postural stability, left by the subject on the force platform [7]. The
plantar pressure, EMG, Soleus muscle, neuromuscular. evaluation of local contribution to the reaction force
resultant is more difficult, and, when mandatory, it is
sometimes obtained by superimposing smaller dedicated
force platforms. All the above requirements could be met
1. INTRODUCTION by a matrix of sensors capable of detecting the normal
stresses. A variety of different types of sensors have been
The postural control system coordinates information from used, including resistive strain gauges, piezoelectric film,
the visual and neuromuscular systems to maintain a stable infrared displacement sensors capacitive, sensors
standing posture. The integrity of these systems can be detecting conductance and ultrasonic sensors.
damaged by diseases, accident or as well as aging [1]. In The present paper illustrates a prototype of new
some cases, the performance of the postural control is plantar pressure measurements for a postural stability
affected in a way that results in an increased risk of analysis. The instrument is based on the use of low cost
falling. Objective methods of diagnosis may be useful in Hall device, magnet and a polymer elastic materiel.
identifying persons at risk. Furthermore, the ability to Basically, the signal is acquired, amplified and feed to a
assess postural performance can help to monitor the high resolution data acquisition card (DaqBoard 1005).
progression of diseases affecting the postural control The 16 bits digital signal is analyzed and processed by a
system and to evaluate effeteness of rehabilitation compatible PC. A preliminary experimental test has been
treatments [2]. Stability can be defined as “the property carried out concerning the postural stability with one foot
of a body that causes it when disturbed from a condition conducted by subjects. The effect of visual input is
of equilibrium or steady motion to develop forces or studied by performing the test with eyes open and fixed
moments that restore the original condition”. Nowadays, visual surround and with eyes closed. However, an
various methods have been devised to quantify postural analog circuit has been realized for recording the EMG
stability. The most widely used method for both static signal of Soleus muscle during the control of postural
and dynamic posturography is based on measurement of stability.
ground reaction forces and center of pressure as the
subject stands on a platform equipped with force

978-1-4577-0411-6/11/$26.00 ©2011 IEEE


2. EXPERIMENTAL MATERIALS

2.1. Principle of plantar pressure measurements

The plantar pressure measurement is based of six tactile


sensors locations [8]. The considered sensor element is
constructed by placing a magnet which produces a
constant magnetic field nearby the selected Hall sensor
[9]. Fig.1 shows the realized sensors were located within Figure 2. EMG conditioning circuit
predetermined anatomic areas under the calcaneus
(CAL), medial (MMF) and lateral (LMF) midfoot, medial 3. EXPERIMENTATION TESTS
(MMH) and lateral (LMH) metatarsal heads, and hallux
(HAL). These six discrete anatomic areas were identified For dynamic experimentation test, two subjects are
as useful in analyzing the paediatric foot during participating for postural stability with one foot by using
ambulation [10]. These locations have also been shown the plantar pressure measurement and EMG conditioning
to present pressure gradients associated with circuit. Fig.3 shows the principle of the experimentation
biomechanical variations and foot pathology in other test. For acquisition signals by National Instrument card
studies. Sensor sites within the defined plantar anatomic (DaqBoard 1005), one channel is selected to the EMG
areas were determined through a combination of clinical signal according to the Soleus muscle and six channels
examination. The objective was to determine the precise are selected to six tactile sensors. Fig.4 and Fig.5 shows
coordinates of highest load centre within each anatomic the response corresponding to six tactile sensors
area. The outputs signals for each sensor element are distributed on the insole surface during the experiment test
carried onto a low level instrumentation amplifier with eyes-open and eyes-closed. The ratio of eyes-closed
(AD622) with low offset voltage, low noise and high to eyes-open measure referred to as the Romberg quotient,
CMRR. After the conditioning, the signals are included estimates the impairment of the perception and vestibular
in the data acquisition card. systems by eliminating inputs from the visual system.

Figure 1. Plantar pressure measurements


and conditioning circuit

2.2. EMG Hardware realization Figure 3. Plantar pressure and EMG electrodes

The EMG signal is acquired with three Ag/AgCl From the recorded signals, two parameters are
commercial electrodes. The input signal is amplified with determined: peak pressure of plantar distribution and the
an instrumentation amplifier (AD622) by a gain 1000 time-delay of eyes-closed. The peak pressure was defined
(Fig.2). Owing to the lack of isolating amplifier, patient as the maximum vertical amplitude exerted during the
safety was increased with the use of isolated power postural instability. Time-delay was defined as the instant
source with batteries bioinstrumentation powered. After of eyes-closed to the instant of the postural instability
amplification, the signal feed a band pass filter which is with one foot. Comparing the two recorders signals, the
composed of a second order Butterworth high pass filter highest values of pressure were noted at the Calcaneus
with a cut-off frequency: 0.05Hz associated in cascade (CAL) region of the subject.2, Calcaneus (CAL) and
with a fourth order Butterworth low pass filter with cut hallux (HAL) region of the subject.1. We notice, for the
off frequency: 500Hz to remove the DC artifact and subject.2, the time-delay with eyes-closed is highest than
noise. the subject.1 result. This observation is in concordance
with the investigation of soleus muscle activity during the
phase of the control foot position. However, the ratio 4. CONCLUSION
time-delay of eyes-closed to eyes-open measure for the
subject.2 is highest than the ratio time-delay for the The plantar pressure measurements to be used for further
subject.1. This result shows the significant degree of clinical application and therapeutics treatment. The low
stability of the subject.2. cost of the system it use as a practical clinical monitoring
tool. The result obtained from this study was not
designed to investigate a specific clinical issue, but rather
PThim
as e -d
d ’a ppui
e lay
to demonstrate operation of the prototype plantar
Y
C elouse
x fe
d rm é ess
-e ye CAL pressure measurements and EMG analogue conditioning.
P h a se d’a p p u i Peak
LM F
O p e n -e ye s p h a se ’s
Y e u x o u ve rte s p re s s u re MMF In our laboratory, the system is now being used to
LM H
M TM investigate plantar distribution characteristics for foot and
Voltage (v)

HAL
ankle pathology and orthopaedic rehabilitation.
NPoh acse
o nd e nt opnh a
tac ap p ’s
se ui

5. REFERENCES

[1] B.A. McClenaghan, H.G. Williams, J. Dickerson, M.


Dowda, L. Thombs, and P. Eleazer, “Spectral characteristics of
aging postural control,” Gait & Posture, vol. 4, pp. 112-121,
1996.

[2] M. Casadio, P.G. Morasso, and V. Sanguineti, “Direct


measurement of ankle stiffness during quiet standing:
implications for control modeling and clinical application,”
Voltage (v)

Gait & Posture, vol. 21, pp. 410-424, 2005.

[3] S. Boukhenous, M. Attari, and N. Ababou, “A dynamic


study of a foot-to-floor interaction during a vertical jumping,”
AMSE journals, vol. 75, no. 1, pp. 41– 49, 2006.

[4] Edward D. Lemaire, A. Biswas, and J. Kofman, “Plantar


T im e (s )
pressure parameters for dynamic gait stability analysis,” in
Figure 4. Recorded signals of the subject.1 Proc. 28th International Conference IEEE-EMBS, NewYork
City, USA, pp. 4465-4468, Aug. 2006.

[5] S. Valerie, and G. Cobb, “Measurement of postural


P eak
CAL
stability before and after immersion in a virtual environment”
pressure
LM F Appl Ergo, vol. 30, no. 1, pp. 47-57, 1999.
MMF
LM H
M TM
HAL
[6] S.H. Scott and D.A Winter, “Talocrural and talocalcaneal
joint kinematics and kinetics during the stance phase of
Voltage (v)

NPohase
contact
de
phase’s
walking,” Journal Biomechanical, vol. 24, no.8, pp. 743-752,
non appui
1991.

[7] C. Giacomozzi, and V. Macellari, “Piezo-D ynamometric


P hase d’appui
O pen-eyes phase’s
Y eux ouvertes Tim e-delay
P hase d’appui
platform for more complete analysis of foot-to-floor
C losed-eyes
Y eux fermées interaction,” IEEE Trans. Rehab. Eng., vol. 5, no. 4, pp. 322-
330, Dec. 1999.

[8] S. Boukhenous, and M. Attari, “A plantar pressure


instrument for ankle rehabilitation,” in Proc. 4th International
Conference on Sensing Technology, ICST’2010, Lecce, Italy,
pp. 492-494, June 3-5, 2010.
Voltage (v)

[9] M. Attari, and S. Boukhenous, “A tactile sensors array


for biomedical applications,” 5th international Multi-Conference
on Systems, Signals and Devices, IEEE-SSD’08, Amman,
Jordan, July 20-23, 2008.

[10] Z.O. Abu-Faraj, G.F. Harris, and Peter A. Smith,


“Surgical rehabilitation of planovalgus foot in cerebral palsy,”
Tim e (s)
IEEE Trans. Neural. Sys. and Rehab. Eng., vol. 9, no. 2, Jun.
Figure 5. Recorded signals of the subject.2 2001.

You might also like