SITE ANALYZER

Bird SA 2500 Ex

Site analyzer 1
Bird Electronics SA 2500 Ex
Selection Buttons

Display
PWR
CHRG

7 8 9
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 2
Bird Electronics SA 2500 Ex

Top View

Printer Port
Charging Port

RS232 Port
RF Antenna Port

Site analyzer 3
VSWR Measurement

Bird Electronics -Site Analyzer

Model No. 2500 Ex

PWR
CHRG
Operating Frequency: 780 – 2500 MHz

7 8 9
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 4
VSWR Measurement
Screen displays the mode in which it was switched off at latest use.

Fault location
1.5
Save /Recall

Autoscale

VSWR
Limit Test
OFF
PWR
Hold CHRG

1.0
0 Distance(M) 450
Print

7 8 9
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 5
VSWR Measurement
After pressing Mode, Screen displays the Mode Options .

Mode
Measure Match

Fault Location

Measure Power

PWR
CHRG

Utility

7 8 9
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 6
VSWR Measurement

Measure Match mode


Save /Recall
1.5

Autoscale

VSWR
Limit Test
OFF

PWR
Hold
CHRG
1.0

Print 780 Frequency(MHz) 2500


Calibration Full

Set the freq


7 8 9 range
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 7
VSWR Measurement

Measure Match mode


Freq

Limit

Start 780 MHz

PWR

Stop MHz CHRG


2500

7 8 9
ESC
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 8
VSWR Measurement

Measure Match mode


Freq

Limit

Start 890 MHz

PWR

Stop MHz CHRG


960

Press to
7 8 9 select start
ESC
freq
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 9
VSWR Measurement

Measure Match mode


Freq

Limit

Start 890 MHz

PWR

Stop MHz CHRG


960

Press to
7 8 9 select stop
ESC
freq
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 10
VSWR Measurement
Freq range is changed but calibration is not done.

Measure Match mode


Save /Recall
1.5

Autoscale

VSWR
Limit Test
OFF

PWR
Hold
CHRG
1.0

Print 890 Frequency(MHz) 960


Calibration off

Calibration
must be full
7 8 9 before
ESC
measurement
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 11
VSWR Measurement Connect
open
connector to
RF ant port

Calibration
Open
1.5
Short

VSWR
Load

PWR
CHRG
1.0
890 Frequency(MHz) 960
Calibration off

7 8 9
ESC
4 5 6
1 2 3
Enter
-
0 .
/+

Site analyzer 12
VSWR Measurement

Calibration
Open Done
1.5
Short

VSWR
Load

PWR
CHRG
1.0
890 Frequency(MHz) 960
Calibration off

7 8 9
ESC
4 5 6
1 2 3
Enter
-
0 .
/+

Site analyzer 13
VSWR Measurement Connect
short
connector to
RF ant port

Calibration
Open Done
1.5
Short

VSWR
Load

PWR
CHRG
1.0
890 Frequency(MHz) 960
Calibration off

7 8 9
ESC
4 5 6
1 2 3
Enter
-
0 .
/+

Site analyzer 14
VSWR Measurement

Calibration
Open Done
1.5
Short Done

VSWR
Load

PWR
CHRG
1.0
890 Frequency(MHz) 960
Calibration off

7 8 9
ESC
4 5 6
1 2 3
Enter
-
0 .
/+

Site analyzer 15
VSWR Measurement Connect
Load
connector to
RF ant port

Calibration
Open Done
1.5
Short Done

VSWR
Load

PWR
CHRG
1.0
890 Frequency(MHz) 960
Calibration off

7 8 9
ESC
4 5 6
1 2 3
Enter
-
0 .
/+

Site analyzer 16
VSWR Measurement
Now Site Analyzer is ready for VSWR measurement.

Measure Match mode


Save /Recall
1.5

Autoscale

VSWR
Limit Test
OFF

PWR
Hold
CHRG
1.0

Print 890 Frequency(MHz) 960


Calibration Full

Calibration is
7 8 9 changed to
ESC full
4 5 6
1 2 3
Enter
0 . -/+

Site analyzer 17
VSWR Measurement

Jumper
Connect Adopter

Cable
and than jumper
cable to RF ant
port

Measure Match mode


Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 890 Frequency(MHz) 960
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 18
VSWR Measurement

Jumper
Cable
To read the accurate value set the marker.

Measure Match mode


Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 890 Frequency(MHz) 9600
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 19
VSWR Measurement

Jumper
Cable
To read the accurate value set the marker.

Activate Marker
M1
1.5
OFF

VSWR
Triangular

PWR
CHRG
1.0
780 Frequency(MHz) 2500
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 20
VSWR Measurement

Jumper
Cable
Measure Match mode
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 890 Frequency(MHz) 960
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 21
VSWR Measurement

Jumper
Cable
Marker reads high VSWR. To locate the fault go to fault
location mode.

Measure Match mode


Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 890 Frequency(MHz) 960
M1 1.47 940 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 22
VSWR Measurement

Jumper
Cable
Measure Match mode
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 890 Frequency(MHz) 960
M1 1.47 940 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 23
VSWR Measurement

Jumper
Cable
Mode
Measure Match
1.5

Fault Location

VSWR
Measure Power

PWR
CHRG
1.0
Utility 890 Frequency(MHz) 960
M1 1.47 940 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 24
VSWR Measurement

Jumper
Cable
Fault Location
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 0 Meters 450
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 25
VSWR Measurement

Jumper
Cable
Fault Location
Distance
1.5
Scale

VSWR
PWR
CHRG
Min 0 1.0
0 Meters 450
Max 450 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 26
VSWR Measurement

Jumper
Cable
Fault Location
Distance
1.5
Scale

VSWR
PWR
CHRG
Min 0 1.0
0 Meters 450
Max 45 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 27
VSWR Measurement

Jumper
Cable
Fault Location
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 0 Meters 45
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 28
VSWR Measurement

Jumper
Cable
Activate Marker
M1 1.5

OFF

VSWR
Triangular

PWR
CHRG
1.0
0 Meters 45
Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 29
VSWR Measurement

Jumper
Cable
Fault Location
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 0 Meters 45
M1 5 1.24 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 30
VSWR Measurement

Jumper
Cable
Fault Location
Save /Recall
1.5
Autoscale

VSWR
Limit Test
OFF
PWR
Hold
CHRG
1.0
Print 0 Meters 45
M1 25 1.47 Calibration Full

7 8 9
ESC
4 5 6
1 2 3
- Enter
0 .
/+

Site analyzer 31