Professional Documents
Culture Documents
Section V NDE
III.NONDESTRUCTIVE EXAMINATION
ASME Section V, Nondestructive Examination
NOTE: The examination will cover ONLY the main body of
each referenced Article, except as noted.
Section V
T – 130 Equipment
T – 150 Procedure
Systems (API Exam Preparation Training)
The Manufacturer, fabricator, or installer shall be
responsible for all required Records/Documentation.
a. required marking
b. type, selection, number, and placement of IQI’s,
c. allowable density
d. control of backscatter radiation
e. location markers
3. Records
Section 15 Haward Technology Middle East 10
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
Article 2 Radiographic
Examination
Section V
T – 210 Scope
IQIs shall be either the hole type or the wire type. Hole-type
IQIs shall be manufactured and identified in accordance with the
requirements or alternates allowed in SE-1025. Wire-type IQIs
shall be manufactured and identified in accordance with the
requirements or alternates allowed in SE-747, except that the
largest wire number or the identity number may be omitted. ASME
standard IQIs shall consist of those in Table T-233.1 for hole type
and those in Table T-233.2 for wire type.
However this rule does not hold for the smaller IQI Numbers 5 and
7 which are both identical to Number 10. To say it another way,
Numbers 5, 7 and, 10 all have the same diameter of 1T, 2T and
4T holes. Take a look!
Section 15 Haward Technology Middle East 20
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
For Example:
of Location Markers
T – 276.2 Size
T – 276.2 Size
T – 276.2 Size
(1) The IQI on the source side of the separate block shall
be placed no closer to the film than the source side of the
part being radiographed.
(2) The separate block shall be placed as close as
possible to the part being radiographed.
(3) The block dimensions shall exceed the IQI
dimensions such that the outline of at least three sides of
the IQI image shall be visible on the radiograph.
(b) Film-Side IQI(s). Where inaccessibility prevents hand
placing the IQI(s) on the source side, the IQI(s) shall be
placed on the film side in contact with the part being
examined. A lead letter “F” shall be placed adjacent to or
on the IQI(s), but shall not mask the essential hole where
hole
Section 15 IQIs are used. Haward Technology Middle East 41
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
(c) IQI Placement for Welds — Hole IQI s. The IQI (s) may
be placed adjacent to or on the weld. The identification
number (s) and, when used, the lead letter “F,” shall not be
in the area of interest, except when geometric configuration
makes it impractical.
(d) IQI Placement for Welds — Wire IQI s. The IQI (s) shall
be placed on the weld so that the length of the wires is
perpendicular to the length of the weld.
Hole IQIs
T – 600 Scope
T – 620 General
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
T – 651 Techniques
T – 651 Techniques
Systems (API Exam Preparation Training)
Temperatures
T – 653 Techniques for Nonstandard Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
Temperatures
Removal
T – 675 Developing
T – 675.3
Article 6 Appendix II
II-610 Scope
This Appendix contains requirements for
the control of contaminant content for all
liquid penetrant materials used on nickel
base alloys, austenitic stainless steels,
and titanium.
Section V
T – 720 General
Table T – 721
T- 741.1 Preparation
(a) Satisfactory results are usually obtained when the surfaces
are in the as-welded, as-rolled, as-cast, or as-forged
conditions. However, surface preparation by grinding or
machining may be necessary where surface irregularities
could mask indications due to discontinuities.
(b) Prior to magnetic particle examination, the surface to be
examined and all adjacent areas within at least 1 in………..
T – 741.1 Preparation
T- 751 Techniques
Equipment
(a) Frequency. Each piece of magnetizing equipment with an
ammeter shall be calibrated at least once a year, or
whenever the equipment has been subjected to major
electric repair, periodic overhaul, or damage. If equipment
has not been in use for a year or more, calibration shall be
done prior to first use.
(b) Procedure. The accuracy of the unit’s meter shall be
verified annually by equipment traceable to a national
standard. Comparative readings shall be taken for at least
three different current output levels encompassing the
usable range.
(c) Tolerance. The unit’s meter reading shall not deviate by
more than ±10% of full scale, relative to the actual current
value as shown by the test meter.
Section 15 Haward Technology Middle East 95
API 570: PIPING INSPECTION CODE: Inspection,
Direction
Fig. T – 764.1.1
Direction
T-764.1.2 Artificial Flaw Shims. The shim, shown in Fig. T-
764.1.2 shall be attached to the surface to be examined,
such that the…………….
Magnetic
Particles
T-910 SCOPE
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
T-910 SCOPE
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
T-910 SCOPE
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
.
Section 15 Haward Technology Middle East 118
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
T-990 DOCUMENTATION
T-1000 INTRODUCTION
T-1010 SCOPE
This Article describes methods and requirements
for the performance of leak testing.
(a) When a leak testing method or technique of
Article 10 is specified by a referencing Code
Section, the leak test method or technique shall be
used together with Article 1, General Requirements.
(b) Definition of terms used in this Article are in
Mandatory Appendix VII of this Article.
(c) The test methods or techniques of these
methods can be used for the location of leaks or the
measurement of leakage rates.
Section 15 Haward Technology Middle East 122
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
T-1030 EQUIPMENT
T-1031 Gages
(a) Gage Range. When dial indicating and recording
pressure gage (s) are used in leak testing, they should
preferably have the dial (s) graduated over a range of
approximately double the intended maximum pressure,
but in no case shall the range be less than 1-1/2 nor
more than four times that pressure. These range limits
do not apply to dial indicating and recording vacuum
gages. Range requirements for other types of gages
given in an applicable Mandatory Appendix shall be
as required by that Appendix.
T-1030
T-1031 Gages EQUIPMENT
T-1030 EQUIPMENT
T-1031 Gages
For large vessels or systems where one or more gages
are specified or required, a recording type gage is
recommended, and it may be substituted for one of
the two or more indicating type gages.
(c) When other types of gage (s) are required by an
applicable Mandatory Appendix, they may be used in
conjunction with or in place of dial indicating or
recording type gages.
I-1000 INTRODUCTION
I-1010 SCOPE
The objective of the direct pressure technique of
bubble leak testing is to locate leaks in a pressurized
component by the application of a solution or by
immersion in liquid that will form bubbles as leakage
gas passes through it.
I-1070 TEST
I-1071 Soak Time
Prior to examination the test pressure shall be held
for a minimum of 15 min.
I-1081 Leakage
Unless otherwise specified by the referencing Code
Section, the area under test is acceptable when no
continuous bubble formation is observed.
Scope
SE-797
The velocity in the material being examined is
a function of the physical properties of the material.
It is usually assumed to be a constant for a given
class of materials.
SE-797
Thickness (T), when measured by the pulse-echo
ultrasonic method, is a product of the velocity of sound
in the material and one half the transit time (round trip)
through the material.
T = Vt/2
Where:
T = Thickness
V = Velocity
t = time
SE-797
One or more reference blocks are required
having known velocity, or of the same material
to be examined, and having thicknesses
accurately measured and in the range of
thicknesses to be measured. It is generally
desirable that the thicknesses be “round
numbers” rather than miscellaneous odd
values.
One block should have a thickness value near
the maximum of the range of interest and
another block near the minimum thickness.
Section 15 Haward Technology Middle East 146
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
SE-797
7.1 Case I — Direct Contact, Single-Element
Search Unit:
7.1.1 Conditions — The display start is
synchronized to the initial pulse. All display
elements are linear. Full thickness is displayed
on CRT.
SE-797
SE-797
7.2 Case II — Delay Line Single-Element
Search
Unit:
7.2.1 Conditions — When using this search
unit, it is necessary that the equipment be
capable of correcting for the time during
which the sound passes through the delay
line so that the end of the delay can be made
to coincide with zero thickness. This requires
a so-called “delay” control in the instrument
or automatic electronic sensing of zero
thickness.
Section 15 Haward Technology Middle East 149
API 570: PIPING INSPECTION CODE: Inspection,
Repair, Alteration & Rerating of Inservice Piping
Systems (API Exam Preparation Training)
SE-797
7.2.2.1 Use at least two standardization blocks.
One should have a thickness near the maximum
of the range to be measured and the other block
near the minimum thickness. For convenience, it
is desirable that the thickness should be “round
numbers” so that the difference between them
also has a convenient round number” value.
SE-797
SE-797
SE-797
SE-797