LWD6500 FAQ

LW ICS Dept.
V1.1 Shawn Cai
6/27/23
Content
Reagent Consumption per test?

Blood consumption?

Life of Laser Diode (lamp)?

How many valves?

Why LWD6500 don’t use detergent?

How to read scattergram?

How many results can be stored?

Time for “start up” and “shutdown”?

Landwind Customer Service Department


Reagent Consumption per test?

Mode Whole Blood Prediluted

Reagent CBC+Diff CBC CBC+Diff CBC

Diluent 41.018 29.148 40.97 29.1

Lyse-LH 0.5 0.5 0.5 0.5

Lyse-I 1.7 0 1.7 0

Lyse-II 0.09 0 0.09 0

Landwind Customer Service Department 3


Blood consumption?

Mode Whole Blood Prediluted


CBC 7 µL
20 µL
CBC+Diff 16 µL

Landwind Customer Service Department 4


Life of Laser Diode (lamp)?
From 10,000 to 20,000 hours.

Low Angle Scattered light

Middle Angle
Laser Diode Scattered light
(λ=635nm)

Flowcell

Landwind Customer Service Department 5


How many valves?
29 valves.

Picture No. Quantity Name

/ 16 Two-way valve

/ 11 Three-way valve

Three-way valve
4 1
(large flow)

21 1 Pressure off valve

Landwind Customer Service Department 6


Why LWD6500 don’t use detergent?
LWD6500 using probe clean and diluent to clean hydraulic system, it can reduce
blockage rate.
Using Detergent will consume more reagent, it’s not economic!
Using Detergent will cause problems, will affect performance for long-term use!
The following is an example from competitor of Landwind:

1. ASCO valve break;


2. HGB background voltage changing;
3. Stability of detergent is poor;
4. Detergent is not easy to clean,
residue affects performance.

Using detergent would increase


leakage of wash block.
Landwind Customer Service Department 7
How to read scattergram?
Neu

Mon Eso

Lymph Bas

Note: Reference of Bas% is from 0% to 1%, so we usually see the Lymph, Mon, Neu and Eso on the
Scattergram;

Landwind Customer Service Department 8


How many results can be stored?
Maximum storage is 100,000 . We suggest doctor backup data every month.

Landwind Customer Service Department 9


Time for “start up” and “shutdown”?
Start up time is about 4 mins 50 s.
Shutdown time is about 9 mins 20 s.

Landwind Customer Service Department 10