LEO(I)
Preheat 5
LBA
Preheat 7
HGB Bath
Flow Cell
FCM
DIL
DIL
New SRV
SRV DIL
New
BASO RBC
LBA Premixing
PS: DP2 & DP3, DP5 & DP6 are using different membrane inside, so could not
exchange the position with each other.
Liquid port
Gas port
Sample aspirate
Syringe
Application of Syringe
Flow Cell
BASO&DIFF
DIL
Sample Aspiration
SRV with Diluent Cleaning
• Advantage:
1. wearable, stability of chemical character, anti-rust,
good temperature character
2. dispensing fast
DP2
DP3
DP8
DP8 DP6
DP7
DP5
RBC Sample In
First Dilution
Out
-0.04Pa
vacuum
New
Bubble mixing
SRV-Diff
LEO(I)
LBA
DIL
LEO1
LBA
LEO2
Flow Cell
WC1
Function:
Filter impurity of FCM, protect flow
cell
Shelf life : 2 year
Attention:
In the installation, inlet port must be
upward
WC2
New
Buffer
Cleanse bath
r
WC1
WC3
Cleanser Container
DIL Container
Soak the liquid path Generate the sheath Sample dilute and cleaning
when shut down fluid in flow cell
WC3
Waste
Container
WC1
WC1
• Sample volume
• Metering pump
WBC channel
• Then get
Lym#, Mon#, Eos#, Neu#
New
• Cleanser Cleaning :
1.Open SV17, use diluent cleaning corresponding tubings by +70kPa
pressure.
2.Use Metering pump and syringe to clean counting tubing again.
• Check or decrease the RBC Gain. After that, you may recalibrate
RBC, PLT and MCV.
• Check the metal connectors of WBC reaction bath whether they are
blocked; Clean the SRV and the tubing between the SRV and WBC
reaction Bath.
Facula before
SS PD
Direct light
blocker
Out-surface dirty
Facula before
SS PD
• Check and adjust HGB Gain to make sure the HGB blank voltage is
2.0-2.4V (2.28V recommended).
• Check the HGB bath whether it is dirty or there is leakage.
• Check the HGB unit whether it is loose or dirty.
• Check whether there is enough diluent dispensed into the HGB bath
when counting. If not, check GV73, and SV19.
• Check whether there is enough LH lyse dispensed into the HGB
bath when counting. If not, check GV70, and SV28.
• Observe the mixing to see if bubble is too big or none. If yes, check
GV65 and CV6.
• Check the pressure 0.07MPa and -0.085MPa.