BC-5800 Liquid System

Hematology Global Technical Support Department


Service@mindray.com
Version : 1.0
NO.:MXQ-12024-BC-5800

© 2011 Mindray Confidential


OTY-12002 ( 3.0 )
Liquid System

 Components
 Counting Principle
 Counting Sequence
 Troubleshooting

2 © 2011 Mindray Confidential


Components

© 2011 Mindray Confidential


Components
Solenoid valve 22 3-way Mindray valves, 26 2-way Mindray valves, 4
2-way HI-Pressure valve,
Valve
1 SMC valve ( 2-way valve)
Pinch valve 3 small tube, 6 big tube
Check valve 6 CHV

Metering Pump Metering pump 1 0.32mL, 2 0.52mL, 2 1ml, 1 1.5ml,


1 2mL, 1 2.6mL
Shear valve , Syringe: 100uL, 2.5mL
Sampling Parts Sample probe, Pierce needle, Wipe block
RBC Pre-mixing bath, RBC/PLT bath
Pre-heat bath, WBC Reaction bath, Flow Cell bath, FCM filter
Chamber/Bath HGB bath
3 Reagent chamber, 3 Waste chamber, Buffer chamber

4 © 2011 Mindray Confidential


SV: Solenoid Valve

New

New New

5 © 2011 Mindray Confidential


Application of Valves

New

6 © 2011 Mindray Confidential


CHV: Check Valve

7 © 2011 Mindray Confidential


Application of Valves

8 © 2011 Mindray Confidential


SMC valve ( 2-way valve)

New

Note: Pay attention to the direction of tubing connection

9 © 2011 Mindray Confidential


Application of Valves

New

10 © 2011 Mindray Confidential


PV: Pinch Valve

11 © 2011 Mindray Confidential


Application of Valves

12 © 2011 Mindray Confidential


DP: Diaphragm pump/Metering Pump

1 metering pump + 1 GV= 1 motor + 1 syringe

New

13 © 2011 Mindray Confidential


Application of Metering Pump
LEO(II)
Diff Bath

LEO(I)
Preheat 5

LBA
Preheat 7

14 © 2011 Mindray Confidential


Application of Metering Pump
WC2-2 WC2-1

HGB Bath

Flow Cell

FCM

Diff & Baso

15 © 2011 Mindray Confidential


Application of Metering Pump

SRV

DIL

DIL
New SRV

16 © 2011 Mindray Confidential


Application of Metering Pump

SRV DIL

New

17 © 2011 Mindray Confidential


Application of Metering Pump

DP1 DP2, 3 DP4 DP5,6 DP7 DP8

Specification 0.32mL 0.52mL 1.5mL 1mL 2mL 2.6mL

DIFF DIFF DIFF/BASO HGB HGB RBC


Usage
LEO(II) LEO(I) sample LH Diluent Diluent

BASO RBC
LBA Premixing

PS: DP2 & DP3, DP5 & DP6 are using different membrane inside, so could not exchange the position with
each other.

18 © 2011 Mindray Confidential


Notice : when using metering pump

 Liquid port should be toward up in case of some air bubble remain in t


he pump.

Liquid port

Gas port

19 © 2011 Mindray Confidential


Sampling Parts

120ul Blood (WB)


Sample Probe
2.5ml Syringe SRV 120ul Dilution (PD)
Pierce Needle 180ul Blood (WB)

20 © 2011 Mindray Confidential


Sample Probe

21 © 2011 Mindray Confidential


Pierce Needle

Sample aspirate

internal needle clean (Gas hole)

external needle clean Blow-dry

Sample aspirate Clean procedure (Sample hole & Gas hole)

22 © 2011 Mindray Confidential


Syringe

Syringe

Syringe motor Syringe sensor

23 © 2011 Mindray Confidential


Application of Syringe
FCM

Flow
l
Cel

BASO&DIFF

24 © 2011 Mindray Confidential


Application of Syringe

 Sample Aspiration
 SRV with Diluent Cleaning

DIL

25 © 2011 Mindray Confidential


Shear Valve

26 © 2011 Mindray Confidential


Shear Valve

Assembly Chart

27 © 2011 Mindray Confidential


Manufacture of SRV

 Materials: special alumina ceramic

 Advantage:
1. wearable, stability of chemical character, anti-rust, good temper
ature character
2. dispensing fast

 Disadvantage: frangible, hard to machining

28 © 2011 Mindray Confidential


Shear Valve Components

Inside
New

Sampling valve Rotary valve Rear fixed valve

29 © 2011 Mindray Confidential


Shear Valve Components
Difference
BC5500 SRV, BC5800 SRV,
DIFF&BASO distribution hole=1*20ul DIFF&BASO distribution hole=2*10ul
RBC distribution hole=8.3ul RBC distribution hole=6ul
One position fixing hole Two position fixing hole: UP & DOWN position

30 © 2011 Mindray Confidential


Shear Valve Components

Sketch Map

Sampling valve Rotary valve Rear fixed valve

31 © 2011 Mindray Confidential


Function of SRV

 Use flow paths, take the certain length and section to get or measure
the sample
 Ration and shear the blood sample

32 © 2011 Mindray Confidential


Function of SRV

SRV New
DP1

DP2

DP3

DP8

DP8 DP6

DP7

DP5

33 © 2011 Mindray Confidential


Tubing Connection of SRV

34 © 2011 Mindray Confidential


Work Principle of SRV (1)
Open-vial mode

The handle on the


sampling valve is
turned to the upper
position. 2.5ml syringe
aspirates sample into
the shear valve.

35 © 2011 Mindray Confidential


Work Principle of SRV (2)
Open-vial mode

The handle of the sampling


valve is turned to the lower
position. The sampling valve
divides the sample into 4
segments and sends them
into different baths for
analysis.

36 © 2011 Mindray Confidential


Work Principle of SRV (3)
Open-vial mode

The handle of the


sampling valve is on the
lower position. Certain
amount of primarily
diluted sample is
aspirated from the
premix bath.

37 © 2011 Mindray Confidential


Work Principle of SRV (4)
Open-vial mode

The handle of the


sampling valve is turned to
the upper position. Certain
amount of primarily
diluted RBC sample is
injected into the RBC bath.

38 © 2011 Mindray Confidential


Work Principle of SRV (1)
Autoloader/close tube mode

The handle on the


sampling valve is
turned to the lower
position. 2.5ml syringe
aspirates sample into
the shear valve.

39 © 2011 Mindray Confidential


Work Principle of SRV (2)
Autoloader/close tube mode

The handle of the sampling valve


is turned to the upper position.
The sampling valve divides
the sample into 4 segments
and sends them into different
baths for analysis.

40 © 2011 Mindray Confidential


Work Principle of SRV (3)
Autoloader/close tube mode

The handle of the sampling


valve is on the upper
position. Certain amount of
primarily diluted sample is
aspirated from the premix
bath.

41 © 2011 Mindray Confidential


Work Principle of SRV (4)
Autoloader/close tube mode

The handle of the sampling


valve is turned to the lower
position. Certain amount of
primarily diluted RBC
sample is injected into the
RBC bath.

42 © 2011 Mindray Confidential


RBC Pre-mixing Bath & RBC/PLT Bath

43 © 2011 Mindray Confidential


Application of RBC Pre-mixing Bath

RBC Sample In
First Dilution
Out

44 © 2011 Mindray Confidential


Application of RBC/PLT Bath

-0.04Pa
vacuum

New

45 © 2011 Mindray Confidential


Volumetric Metering

This analyzer uses a volumetric metering unit to control the coun


t cycle and to ensure that a precise volume of sample is analyzed.

Uppersensor Uppersensor

Lowersensor Lowersensor

1Em
ptywhenstart 2Liquidsurfacefaldownthroughthe
volum
eritctube

Uppersensor Uppersensor

Lowersensor Lowersensor

3Countingstartwhenthe 4Countingendwhentheliquid
liquidsurfacepasstheupper surfacepassthelowersensor
sensor.

46 © 2011 Mindray Confidential


HGB Bath

47 © 2011 Mindray Confidential


Application of HGB Bath

Lyse HGB sample from SRV

Bubble mixing

48 © 2011 Mindray Confidential


Pre-heat Bath

49 © 2011 Mindray Confidential


Pre-heat Bath

50 © 2011 Mindray Confidential


Application of Pre-heat Bath

SRV-Diff
LEO(I)
LBA
DIL

51 © 2011 Mindray Confidential


WBC Reaction Bath

52 © 2011 Mindray Confidential


WBC Reaction Bath

53 © 2011 Mindray Confidential


Application of WBC Reaction Bath
FCM

LEO1
LBA
LEO2

Flow Cell

WC1

54 © 2011 Mindray Confidential


Flow Cell Bath

55 © 2011 Mindray Confidential


Application of Flow Cell Bath

56 © 2011 Mindray Confidential


Application of FCM filter

New

Function:
Filter impurity of FCM, protect flow cell
Shelf life : 2 year
Attention:
In the installation, inlet port must be upward

57 © 2011 Mindray Confidential


Reagent, Waste and Buffer Chamber

DIL FCM

WC2

New
Buffer bath
Cleanser

WC1
WC3

58 © 2011 Mindray Confidential


Reagent Chamber

59 © 2011 Mindray Confidential


Waste and Buffer Chamber

60 © 2011 Mindray Confidential


Application of DIL&FCM Chamber

Cleanser Container

DIL Container

Soak the liquid path Generate the sheath Sample dilute and cleaning
when shut down fluid in flow cell

61 © 2011 Mindray Confidential


Application of WC1

WC3

Waste Container

62 © 2011 Mindray Confidential


Application of WC2

(ASCO Valve Protection)

WC1

63 © 2011 Mindray Confidential


Application of WC3

WC1
64 © 2011 Mindray Confidential
FAQ1

 What are the 8 metering pumps used for?


 How can we maintain the liquid components?
 Metering pump, Solenoid valve
 Shear valve
 DIFF/BASO bath, RBC bath, HGB bath
 Flow cell
 Reagent chamber

65 © 2011 Mindray Confidential


FAQ2: How to Check…?

 Sample volume

 Metering pump

66 © 2011 Mindray Confidential


FAQ3: How to Check…?

 Premix/RBC/HGB bath/DIL/FCM chamber

67 © 2011 Mindray Confidential


Counting Principle

© 2011 Mindray Confidential


Counting Principle

WBC channel

 First count Diff channel, get


Lym%, Mon%, Eos%, Neu%

 Then count Baso channel, get


WBC, Baso#, Baso%

 Then get
Lym#, Mon#, Eos#, Neu#

69 © 2011 Mindray Confidential


Counting Principle

HGB channel
 Colorimetry

70 © 2011 Mindray Confidential


Counting Principle

RBC/PLT channel
 Electrical Impedance Method

71 © 2011 Mindray Confidential


Counting Sequence

© 2011 Mindray Confidential


Counting Sequence

Flow Chart of Whole Blood Mode

New

73 © 2011 Mindray Confidential


Counting Sequence

Flow Chart of Prediluted Mo


de

74 © 2011 Mindray Confidential


Liquid Diagram

75 © 2011 Mindray Confidential


0~11 seconds

76 © 2011 Mindray Confidential


Aspirate sample, test HGB blank, clean & drain (0-11s)
77 © 2011 Mindray Confidential
11~18 seconds

78 © 2011 Mindray Confidential


Dispense sample & lyse, aspirate from pre-mix bath, clean & drain RBC bath
(11-18s)
79 © 2011 Mindray Confidential
18~23 seconds

80 © 2011 Mindray Confidential


Dispense into RBC bath, prepare DIFF dilution & Sheath fluid, HGB dilution stay s
till(18-23s)
81 © 2011 Mindray Confidential
23~34 seconds

82 © 2011 Mindray Confidential


RBC/PLT, DIFF counting, clean pre-mix bath, diff bath, sample probe & wipe block
(23-34s)
83 © 2011 Mindray Confidential
34~39 seconds

84 © 2011 Mindray Confidential


RBC/PLT, HGB counting, WBC channel cleaning (34-39s)
85 © 2011 Mindray Confidential
39~45 seconds

86 © 2011 Mindray Confidential


Prepare Baso& Sheath fluid, clean RBC/PLT, HGB bath, prefill pre-mix, HGB bath; ready for next samp
le (39-45s)
87 © 2011 Mindray Confidential
45~49 seconds

 Clean the WBC counting module

88 © 2011 Mindray Confidential


Clean RBC rear bath, prefill HGB, DIFF, BASO bath (4
5-49s)
89 © 2011 Mindray Confidential
49~60 seconds

90 © 2011 Mindray Confidential


Cleaning WBC channel, zap and prefill RBC bath, waste draining
(49-60s)
91 © 2011 Mindray Confidential
Maintenance Sequence

© 2011 Mindray Confidential


Cleanser Cleaning

 Cleanser Prime :
Open SV13 and corresponding valves, prime cleanser by +70kPa pressure.

 Cleanser Cleaning :
1. Open SV17, use diluent cleaning corresponding tubings by +70kPa press
ure.
2. Use Metering pump and syringe to clean counting tubing again.

93 © 2011 Mindray Confidential


Cleanser Maintenance
94 © 2011 Mindray Confidential
Probe Cleanser Cleaning

 It includes SRV, WBC bath, RBC bath, Flow Cell, HGB bath, Premix bath,
HGB bath maintenance.
 Aspiration:
Overall Maintenance: Open SV07, PV10, SV11, SV16 and SV10, use DP4
aspirate 6ml (4 times) probe cleanser from Open-Vial probe into WBC c
hannel
and flow cell.
SRV Maintenance: Use 2.5ml syringe to aspirate probe cleanser into SR
V.
 Dispensing:
Use 2.5ml syringe dispense probe cleanser into HGB, RBC, Pre-mix bath.
95
 Cleaning: it is similar to cleanser cleaning.
© 2011 Mindray Confidential
Probe Cleanser Maintenance
96 © 2011 Mindray Confidential
Troubleshooting

© 2011 Mindray Confidential


Abnormal PLT Background

 PLT: 10~40. It is caused by dirty diluent, SRV or tubing.


 Cleaning by background test, probe cleanser soaking and disassemble
cleaning.

98 © 2011 Mindray Confidential


Abnormal PLT Background

 PLT: If there is interference, usually there will be a peak wave bet


ween 0-10 FL.

99 © 2011 Mindray Confidential


Abnormal PLT Background
 Check the shielding cover of RBC bath and the signal processing boar
d whether they are installed well.
 Check or decrease the RBC Gain. After that, you may recalibrate RBC,
PLT and MCV.
 Check there is leakage in the RBC rear bath.
 Check the isolation chamber under the RBC bath whether it is dirty.

100 © 2011 Mindray Confidential


RBC Bubble

RBC bath not enough diluent?


RBC rear bath leaking?
Anywhere leaking from rear bath to WC3?
Tube connectors loose?
RBC reference count time setting?
Volumetric sensors work well?

101 © 2011 Mindray Confidential


No WBC Result (about 0.00)

Leucopenia: BASO channel blocked

102 © 2011 Mindray Confidential


No WBC Result (about 0.00)

 Check the reagent of LBA. Check the liquid parts adding LBA. Such as m
etering pump DP3, gas valve GV72, liquid valve SV27 and tubing from LB
A container to SRV.

 Check the metal connectors of WBC reaction bath whether they are block
ed; Clean the SRV and the tubing between the SRV and WBC reaction Bath
.

 Check the pinch valve PV12 and gas valve GV50.

 Check the pressure 0.25MPa, 0.07MPa and -0.085MPa.

 Check the signal processing board.

103 © 2011 Mindray Confidential


Flow cell blocked

104 © 2011 Mindray Confidential


Flow cell inner-surface dirty

Facula before SS PD

105 © 2011 Mindray Confidential


Flow cell out-surface dirty

Direct light blocker

Out-surface dirty
Facula before SS PD

106 © 2011 Mindray Confidential


Hemolysis insufficient

Lyse reagents, lyse dispense volume, temperature

107 © 2011 Mindray Confidential


Lyse Resist (hemolysis insufficient)

Lyse resist & low LAS Gain

108 © 2011 Mindray Confidential


Lyse Resist (hemolysis insufficient)
 Check the ambient temperature; it should be much more than 15 centigra
de.
 Check the Gain.
 Check the temperature of WBC reaction bath and pre-mix bath.
 Check the lyse LEO(I) and LEO(II); maybe themselves.
 Check the tubing adding LEO(I) and LEO(II) and metal connectors of WBC
reaction bath;
 Clean the SRV and the tubing between the SRV and WBC reaction Bath.
 Checking the metering pump DP1 and DP2 whether the dispensing volume i
s enough. Also the gas valves (such as GV69, 71) and liquid valves (su
ch as SV25, SV26)
 Check the pressure 0.07MPa and -0.085MPa.

109 © 2011 Mindray Confidential


HGB Error or HGB Results Abnormal

 Check and adjust HGB Gain to make sure the HGB blank voltage is 2.0-2.
4V (2.28V recommended).
 Check the HGB bath whether it is dirty or there is leakage.
 Check the HGB unit whether it is loose or dirty.
 Check whether there is enough diluent dispensed into the HGB bath when
counting. If not, check GV73, and SV19.
 Check whether there is enough LH lyse dispensed into the HGB bath when
counting. If not, check GV70, and SV28.
 Observe the mixing to see if bubble is too big or none. If yes, check
GV65 and CV6.
 Check the pressure 0.07MPa and -0.085MPa.
110 © 2011 Mindray Confidential
All the stars on the scattergrams
 Check the sheath fluid in the FCM chamber whether it is enough. If not
, perform “Prime > FCM cistern” in the “Service > Maintain” screen
.
 Check the corresponding liquid parts aspirating or dispensing the dilu
ent; such as SMC valve SV53, liquid valve GV80, tube kinkled or leaked
? Make sure no air going into FCM channel.
 Check the liquid parts dispensing the sheath fluid (diluent) into the
flow cell; such as SV3.

111 © 2011 Mindray Confidential


Version Information
Version Author Date Description
1.0 Eric cheng 2012.6.18 BC-5800 Liquid System

112 © 2011 Mindray Confidential


Check Information
Version Reviewer Date Comments
1.0 Carl 2012.6.20 Initial Release

113 © 2011 Mindray Confidential


Thank you!

114 © 2011 Mindray Confidential