NSX-2100 series
29 November 2017
th
,
Skytech Systems, Thane , India
Mitsubishi Chemical Analytech Co., LTD.
Euro6 Specs Fuel
Almost All the Sulphur Analyser in the market based on EDXRF technique will
go obsolete due to non complaince of methods for <10 ppm. UVF can easily meet
the need of Euro6 specs Fuel and more Accurately and Precisely backed with an
international method and offer the following advantages
No matrix effect – only 1 calibration line using Iso-octane or toulene can be used for measurement of
all samples (ie. Petrol, Diesel, Base oil, Lube etc..)
No films or samples cups consumables after every analysis. Very very low running cost
Gas
Trap
column
filling heater
Separation of NOx
: SO2
: NO
Trap and Release Unit is consist of a trap column x
NOx
SO2
Combustion&Trap Release&Sampling
Feature 2: Trap & Release
Obtain higher signal/noise ratio
50ppb 200ul
with Trap& Release unit
50ppb 200ul
without Trap& Release unit
Effect of TRU
Peak profile by changing amount of sample
Non –trap Sulfur peak BIG peak by trap
95
50ng/ml 25,50,100,150,200ul
50ng/ml 25,50,100,150,200ul 50ng/ml 50,100,200ul
43
43
37
30
37
gas
: SO2
Separation
: NO x
Gas
Concentration
Application (Ultra Trace Sulfur)
Conc.
Sample Injection (ml ) Repeat RSD (%)
( V/V)
Desulfurized
100 3 30ppb 10.8
Natural gas
NSX-2100V (SD)+GI-220+TRU-200
Reference, NSX-2100V+TRU+GI220
customer Sample purpose
120.00
Sulfur conc (ppb V/V)
100.00
80.00
60.00
40.00
20.00
0.00
Gas injector GI-220
Auto syringe injection
Up to 200mL
Variable injection amount
High level safety for
flammable sample handling
TRU-210
ASC-250L GI-220
liquid gas/pump •TRU unit is for Vertical, UVFL sulfur.
•Auto sample injector
•ASC-250L (liquid)
•GI-220 (gas)
SD-210 VF-210
Detectors Furnace
Ultra trace with Trap & Release
system injector LOQ injection High amount injection/low
concentration is realized
0.05
NSX-2100V
GI-220 ug/L
200ml by NSX-2100V+TRU-
+TRU-200
200+GI-220
Option 2: STC-210L
2.59 7.59
4.71 7.62
3.56 7.42
3.43 7.53
3.67 7.52
3.59 21.05 % 7.54 1.02 %
tray=20 C,syringe = 15 C
STC-210L
Peltier cooling for both tray and syringe
Condensation considered
Cooling and/or Heating
Cool:
High volatile sample
Tray: 15 C
Syringe: 15C
Heat:
Viscous sample
Tray: 60 C
Syringe 60 C
Feature
Modular system – customized configuration
ABC-210
GI-210
GI-230
GI-220
Pyrolysis Tube
CONTROLLER
Filter
Mixer
Exhaust
Tube Dryer
Furnace
Electric
OZONE
SCRUBBER
N Cell
OZONIZER
S cell
(exhaust)
Chlorine
Ag Working electrodes Constant rate injection
Ag Detection electrode
Reference electrode
Valve
Mass flow controller
Heater furnace
Combustion tube
20
18
Concentration (ug/ml)
16
14
12
10
8
6
4
2
0
0 5 10 15 20 25 30
Number of Measurements
Vertical Furnace
Open/Close furnace.
Higher Volume > 200 uL can be
injected
Samples is injected at 1ul /sec and
forms a drop and falls at the same point
again and again.
Very high precision
Better controlled combustion
Connection to gas line is done via ball
joint and clip.
Incomplaince with International method
Coulometry MCD-210
Only Vertical System compliant to ASTM D7457 (ie.
LDL of 70 ppb)
High sensitive measurement on Cl (S)
Cl- 0.05ppm, S- 0.10ppm
Stable cell, Efficient dehydration, No acid mist.
One cell for all over range.
No need to have sets of cell for each range.
Larger volume dehydration for automation (20ml,
vertical ) .
Software NSX-2100
Full-automation
from “Heater on”
to “Gas shut
down”.
Simultaneous N & S
Recent trends from Indian Petroleum Industry
…. NSX 2100 with TRU (UOP987 Part B)
UOP Gudgaon
Borl
Ril Dahej
Essar Jamnagar
We though that only Private Refineries and Petroleum institutes can go for this ….
BPCL Mahul
Convertible KF
moisture KF-31
Potentiometric Titrator
GT-200
Training for
Petroleum/PetroChemicals
Elemental Analysis – OEA
Trace N/S/Cl Analysis
Vertical Furnace – higher volume injection > 200 ul – better lower detection limit
NSX 2100 V with TRU – 2 ppb LOD and has method compliance (UOP 987 part B) from 10
ppb onwards – only instrument in market for method compliance for < 500 ppb
NSX 2100 V N – 30 ppb with method compliance (ie. UOP 981) – only instrument in market
for method compliance for < 100 ppb
NSX 2100 V Cl – 50 ppb and has method complaince (ASTM D7457) from 70 ppb –
Only vertical instrument in market for method compliance for < 1000 ppb
India’s customers preference - NSX 2100 V with TRU for N, S, Cl analysis – only instrument
with lowest detection limit with International Method compliance
Incase of interest for Carbon analysis, opt for ASTM D5291 complaint instrument. Separate
instrument for C determination will ensure your all important trace level NSCl analysis (ie. ppb
levels) is not contaminated
Training and application infrastructure of Skytech in addition to the expertise of >20 years