You are on page 1of 1

Uncertainty Analysis

Parameter: Method: Pins and Plugs Diamter Measure Direct Standards: ATL1584 A7177 Date: UUT: Range: 3/11/2008 54-0150 0-10"

This anlaysis is based on Type B analysis of the parameter in question. Any historical Type A or pooled uncertainties that may apply will also be included and noted. Uncertainty Model: Lu Su Ru Eu Sm Lu=Su+Ru+Eu+Sm Length Uncertainty Standard Uncertainty Resolution Uncertainty Environmental Uncertainty * Stability of Measurements Measurement Model: La Li Length applied Length andicated La=Li

Uncertainty Analysis Table Symbol Xi Sm Su Ru Eu uc U Inches Observations:


1 2 3 4 5 6 7 8 9 10

11.6 uIn per In/deg C , Lab Env is 20 deg +/- 1 deg C * Max 1 degree shift from Standards cal temp. Type Contribution Units Value Probability Distribution Divisor Sensitivity coefficient ci 1 1 1 1 Uncertainty contribution ui(y)

Estimate xi 0.E+00 0 0 1

A B B B

In Normal 1.000 1.26E-05 In 2.32E-05 Rec 1.732 In 2.0E-05 Rec 1.732 In 1.16E-05 U 1.414 Combined Standard Uncertainty (1 Sigma) in Inches Expanded Uncertainty (k=2) in Inches Uncertainty: Welch-Satterwaite 100.612607 Uc: 2.91671E-19 Ua: 2.56E-20 Ub1: 3.21929E-20 Ub2: 1.77799E-20 Ub3: 4.52933E-21

1.26E-05 1.34E-05 1.15E-05 8.20E-06 2.3239E-05 4.6479E-05

10.00

4.65E-05 Inches

9.99998 10.00000 9.99998 10.00000 10.00000 10.00002 10.00000 10.00000 9.99998 10.00000 10.000 1.2649E-05 9

Stundent's T: Mean: Std. Dev Degree of Freedom:


Su= 4.2 +1.9L uIn for Gage blocks

Infinite

2.00

at 95.45% CL

Atmospheric Pressure/humidity noted sources of uncertainty but found to be negligible in this analysis User error taken into account on stability of measurement, Where different technicians found deviations of 20 uIn. Flatness/Parellelism/Bilateral deviation verified with optical flats and found to be of negligible values to include in budget.