Professional Documents
Culture Documents
Marc Kachelrieß
z x
GE LightSpeed
Toshiba Aquilion
What does CT Measure?
• Polychromatic Radon transform
− dL µ ( r , E ) µ (r , E )
p( L ) = − ln dE w( E ) e
p ( L ) ≈ dL µ ( r , Eeff )
– 2D filtered backprojection
– Spiral z-interpolation 1998 0.5 s 4×1 mm 4×1 mm, 30 s 124
– ASSR and EPBP (cone-beam recon.)
2002 0.4 s 16×0.75 mm 16×0.75 mm, 12 s 604
– Phase-correlated CT (e.g. cardiac CT)
• Image quality and dose 2004 0.33 s 64×0.5 mm 64×0.5 mm, 3 s 2404
– Spatial resolution (PSF, SSP, MTF) 2010 0.2 s 512×0.5 mm 512×0.5 mm, 0.2 s 2500
– Relation of noise, dose and resolution
– Dose values (CTDI, patient dose) 1 assuming a breath-hold limit of 30 s
– Dose reduction techniques 2 factor 4 converts from head FOM to full body FOM
3 assuming p = 1, otherwise Seff is increased
4 assuming p = 1.5 since IQ is independent of pitch for MSCT
Seite 1
1
Fan-Beam Geometry
(transaxial / in-plane / x-y-plane)
p.a. →
x-ray tube
Acquisition
180°
x lateral →
Reconstruction
field of measurement
(FOM) and object
Object, Image
detector (typ. 1000 channels)
y
a.p. →
Sinogram, Rawdata
x
(illustration without quarter detector offset)
Data Completeness
y y
x x
Basic Parameters
(best-of values typical for modern scanners) Demands on the Mechanical Design
• In-plane resolution: 0.4 … 0.7 mm
• Continuous data acquisition in spiral scanning mode
• Nominal slice thickness: S = 0.5 … 1.5 mm
• Able to withstand very fast rotation
• Effective slice thickness: Seff = 0.5 … 10 mm – Centrifugal force at 550 mm with 0.5 s: F = 9 g
• Tube (max. values): 100 kW, 140 kV, 800 mA – with 0.4 s: F = 14 g
– with 0.3 s: F = 25 g
• Effective tube current: mAseff = 10 mAs … 1000 mAs
• Mechanical accuracy better than 0.1 mm
• Rotation time: Trot = 0.33 … 0.5 s
• Compact and robust design
• Simultaneously acquired slices: M = 4 … 64
• Short installation times
• Table increment per rotation: d = 2 … 50 mm
• Long service intervals
• Pitch value: p = 0.3 … 1.5
• Low cost
• Scan speed: up to 16 cm/s
• Temporal resolution: 50 … 250 ms
Seite 2
2
Demands on X-Ray Sources Tube Technology
anode
anode
• High cooling rates (typ. >1 MHU/minute) C B
anode
cathode anode
∂z
tan φ =
∂RF Demands on CT Detector Technology
B
C
* in the order of 105 counts per reading and 103 readings per second
Straton Tube
40 mm
64 × 0.625 mm
Philips β
64 / 0.4 s / 3.8°
19 mm
16 channels
2⋅2⋅32 × 0.6 mm Siemens (of 103) shown
24 × 1.2 mm 2⋅64 / 0.33 s / 1.9°
32 mm
Toshiba
64 × 0.5 mm
M = 64 / 0.4 s / 3.2°
z
Data courtesy of Siemens Medical Solutions, Forchheim, Germany
Number of simultaneously acquired slices M / Rotation time trot / Cone-angle Γ
Seite 3
3
Rows vs. Slices Scan Trajectories
z z z
FOM
CT Basics
From Single-Slice to Cone-Beam Spiral CT
• Technology
– Basic parameters
– Detector concepts, tube technology
– Scan trajectories, scan modes
• Algorithms
– 2D filtered backprojection
– Spiral z-interpolation
– ASSR and EPBP (cone-beam recon)
– Phase-correlated CT (e.g. cardiac CT)
• Image quality and dose
– Spatial resolution (PSF, SSP, MTF)
– Relation of noise, dose and resolution
Animation by Udo Buhl, Aachen – Dose values (CTDI, patient dose)
– Dose reduction techniques
Seite 4
4
Fan-beam geometry Parallel-beam geometry
transaxial
rebinning
(β ,α ) (ξ ,ϑ )
y β y y
RF
α ϑ ξ ϑ ξ
x x x
Measurement:
(β ,α ) (ξ ,ϑ ) p(ϑ , ξ ) = Rf (ϑ , ξ ) = dx dy f ( x, y ) δ (x cos ϑ + y sin ϑ − ξ )
π ∞
2π iu ( x cos ϑ + y sin ϑ )
Inversion: f ( x, y ) = dϑ du u P2 (ϑ , u ) e
0 −∞ p(ϑ , ξ )
π
= dϑ p(ϑ , ξ ) ∗ k (ξ )
0 ξ = x cos ϑ + y sin ϑ Add ray value to each pixel in the “vicinity“ of the ray.
Seite 5
5
Spiral CT Scanning Principle
Filtered Backprojection (FBP) Start of Scan
trajectory
spiral scan
1. Filter projection data with the reconstruction kernel.
2. Backproject the filtered data into the image:
0 z
Direction of 0 t
continuous
Reconstruction kernels balance between patient
spatial resolution and image noise. transport
1996: 1998: 2002: 2004:
1× 5 mm, 0.75 s 4× 1 mm, 0.5 s 16× 0.75 mm, 0.42 s 2⋅⋅32×0.6 mm, 0.33 s
Kalender et al., Radiology 173(P):414 (1989) and 176:181-183 (1990)
z = zR
M=4
z
120 cm in 40 s
Seite 6
6
The Cone-Beam Problem
Animation by Siemens
Kachelrieß M, Schaller S, Kalender WA. Med Phys 2000; 27(4):754-772 Kachelrieß M, Schaller S, Kalender WA. Med Phys 2000; 27(4):754-772
Seite 7
7
Comparison to Other Approximate Algorithms Patient Images
180°LI d=1.5mm Π d=64mm MFR d=64mm ASSR d=64mm
with ASSR
• High image quality
• High performance
• Use of available
2D reconstruction
hardware
• 100% detector usage
• Arbitrary pitch
• Sensation 16 at
• 0.5 s rotation
• 16×× 0.75 mm collimation
• pitch 1.0
• 70 cm in 29 s
• 1.4 GB rawdata
• 1400 images
Bruder H, Kachelrieß M, Schaller S. SPIE Med. Imag. Conf. Proc., 3979, 2000
CT-Angiography
Sensation 64 spiral scan with 2⋅⋅32×0.6 mm and 0.375 s
CTA, Sensation 16 at
Seite 8
8
z Kymo
β The complicated
pattern of overlapping
3-fold
C data …
longitudinally
rebinned 4-fold … will become even
detector C+B more complicated with
phase-correlation.
5-fold
C Individual voxel-by-
voxel weighting and
normalization.
• Spiral • Spiral
• EPBP Std • EPBP Std
The 180°Condition • p = 0.375 • p = 1.0
y
ϑ
dϑ w(ϑ ) = π 180°in 3 segments
and r
x
w(ϑ + kπ ) = 1 • Spiral
k
• ASSR Std
• p = 1.0
Seite 9
9
Motion Artifacts of the Heart Cardiac CT
• Periodic motion
• Synchronisation needed (ECG, Kymogram, others)
• Prospective Gating
• Phase-correlated reconstruction = Retrospective Gating
– Single-phase (partial scan) approaches, e.g. 180°MCD
– Bi-phase approaches, e.g. ACV (Flohr et al.)
– Multi-phase Cardio Interpolation methods, e.g. 180°MCI (gold-standard)
• Generations
– Single-slice spiral CT: 180°CD, 180°CI (introduced 1996*)
– Multi-slice spiral CT: 180°MCD, 180°MCI (introduced 1998*)
– Cone-beam spiral CT: ASSR CD, ASSR CI (introduced 2000*)
– Wide cone-beam CT: EPBP (introduced 2002*)
*Med. Phys. 25(12) 1998, Med. Phys. 27(8) 2000, Proc. Fully 3D 2001, Med. Phys. 31(6) 2004
Sync-Signal
ECG, Kymogram, ... • E.g. trot = 0.5 s and fH = 60 bpm implies p < 0.5
R R R R • The smaller the pitch value the more segments can
0 trot 2trot 3trot 4trot 5trot 6trot t be combined
Width, and thus teff, corresponds to the FWTM of the phase contribution profile.
Time Time
Heartbeat 1 Heartbeat 2 Heartbeat 3 Heartbeat 1 Heartbeat 2 Heartbeat 3
Kachelrieß, Ulzheimer, Kalender, Med. Phys. 27(8):1881-1902 (2000) Kachelrieß, Ulzheimer, Kalender, Med. Phys. 27(8):1881-1902 (2000)
Seite 10
10
Multi-Threaded CT, Dual Source CT
Volume Zoom, 4 × 2.5 mm, 0.5 s, 1998 Sensation 64, 2⋅⋅32 × 0.6 mm, 0.33 s, 2004
Multi-segment 180°MCI reconstruction, 90 bpm Data courtesy of Stephan Achenbach
compact
bone What is Displayed?
CT Basics
From Single-Slice to Cone-Beam Spiral CT
1000
80
800
• Technology 70
liver
600 blood
– Basic parameters spong. 60
400 bone
– Detector concepts, tube technology pancreas
CT-value / HU
50
– Scan trajectories, scan modes 200 kidney
water 40
0 fat
• Algorithms 30
– 2D filtered backprojection -200
20
– Spiral z-interpolation -400 lungs ImpactX.ocx V 1.1.0.0
10
– ASSR and EPBP (cone-beam recon.) -600
0
– Phase-correlated CT (e.g. cardiac CT) -800
air
• Image quality and dose -1000
Spatial Resolution 1
out
In-plane resolution z-resolution
0.5 mm
0 1 in 0.4 mm
0.4 mm
out
y y z
0 1 in
out x x x
in
0 1 Std. scan, x/y UHR scan, x/y Std. or UHR scan, x/z
(0, 5000) (0, 1000) (-750, 1000)
Seite 11
11
Spatial Resolution 3
Spatial Resolution 2 Point Spread Function (PSF), Slice Sensitivity Profile (SSP)
In-plane resolution z-resolution
Standard (no zFFS)
FWHM = 1.3 S
0.3 mm
Double z-
ImpactX.ocx V 1.1.0.0 sampling (zFFS)
FWHM = 1.0 S
0.4 mm
0.5 mm FWHM
0.6 mm
FWTM
Std. scan, x/y UHR scan, x/y Std. or UHR scan, x/z z
FWHM = Seff = effective slice thickness = freely selectable parameter during image recon.
Dose Calculator
Patient Dose in CT
Typical Values for 16-Slice Scanners
Head Thorax Abdomen Pelvis
Seite 12
12
Strategies for Dose Reduction Standard Display
Potential reasons for an increase:
• Higher volume coverage
• Multiphasic examinations
• More examinations
• Higher spatial resolution
• New special applications
Constant
5 000
(attenuation: 50)
total mAs! Conventional scan: 327 mAs Online current modulation: 166 mAs
53% dose reduction on average for the shoulder region
Modulated tube current: Low, homogeneous noise. 49% dose reduction in this case
2 2
σ pixel = const. ⋅ σ projection ,n Kalender WA et al. Med Phys 1999; 26(11):2248-2253
n
Seite 13
13
Automatic Exposure Control (AEC) Multidimensional transaxial filtering
(z-dependent + angular dependent tube current modulation) Adaptive Filtering
a)
(MAF)
Standard CT AEC • Rawdata based
• Local smoothing of noisy data
rel. tube current
Standard 180°MFI
Summary
• CT technology is further evolving towards
Thank You!
– more slices
– faster rotation times
– higher spatial resolution
• CT algorithms
– reconstruct cone-beam data for any trajectory
– perform phase-correlated imaging (4D)
– significantly reduce artifacts (beam hardening, truncation …)
• CT dose
– is becoming more and more an important issue (also in the US)
– is being reduced by manufacturers‘ efforts (e.g. MAF, AEC)
3d .org
– can be most significantly reduced by user training www.fu lly
Seite 14
14