3. Pull out the handle in one motion until it is locked, and then remove
your hand (from the handle). Wait
until the sampling time (approx., 30
seconds) has elapsed.
*If the handle does NOT return to the original position and
remains 5mm. (or more) away from it, the pump is NOT airtight
and measurements would NOT yield correct results! (See the
section about Repair and Maintenance in the instruction manual).
Gastec Pump and Tubes
Measuring
5. Insert a detector tube tip to the tip holder.
Turn the detector tube and make a scratch
with the file in the tip holder. Snap the
detector tube and cut off the tip of the
detector tube.
6. Then slip the sleeve-like rubber inlet cover over the top of the pump (the
end of the pump with the arrow [G⇒]). For detector tubes 2EL, 2EH the
mark is red; for tube 31E it is blue.
When the
demarcation of the
colour change layer is
When the end of the pale, the mean value
colour change layer is between the dark
When the end of the slanted, read the and the pale layer
colour change layer is value in the middle of ends is taken.
flat, simply read the the slant.
value at the end.