You are on page 1of 12

Laboratory S. M.

Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,


Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 1 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

MECHANICAL CALIBRATION

I. DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)


$
1. Caliper (Dial/Digital) 0 to 300 mm 8.0 µm Using Slip Gauges,
L.C.: 0.01 mm 0 to 450 mm 10.0 µm Length Bars
0 to 600 mm 12.0 µm Length Measuring
0 to 1000 mm 15.0 µm Machine (LMM) By
0 to 1500 mm 20.0 µm Comparison Method
As per IS: 3651
L.C.: 0.02 mm 0 to 2000 mm 25.0 µm Part-I, II & III and
NABL-122-01
$
2. Depth Vernier 0 to 300 mm 10.0 µm Using Slip Gauges &
L.C.: 0.01 mm 0 to 600 mm 12.0 µm Length Measuring
Machine (LMM) By
Comparison Method
As per IS : 4213 and
NABL-122-01
$
3. External Micrometer 0 to 25 mm 1.5 µm Using Micrometer
L.C.: 0.001 mm 25 mm to 100 mm 2.0 µm Checking Set & Length
100 mm to 200 mm 3.0 µm Bars By Comparison
200 mm to 600 mm 5.1 µm Method As per IS: 2967
600 mm to 1000 mm 6.0 µm and NABL-122-01

4. Hole Test 3 mm to 200 mm / 0.001 3 µm Using Plain Ring


Micrometer/ Three Gauges Set By
$
Point Micrometer Comparison Method
$
5. Inside Micrometer Micrometer Head Using Length Measuring
(Stick Micrometer) Up to 50 mm 2 µm Machine (L.M.M.) By
L.C.: 0.001 mm Up to 250 mm 3 µm Comparison Method
As per I.S. 2966 &
NABL-122-01

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 2 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Extension Rod / Up to 300 mm 3.5 µm Using Length Measuring


$
Overall Length Up to 1000 mm 6.0 µm Machine (L.M.M.) By
Comparison Method
$
6. Dial Gauge 0 to 1 mm 1 µm Using Electronics Dial
(Plunger Type) 0 to 10 mm 1.5 µm Tester By Comparison
L.C.: 0.0005/ 0 to 25 mm 2.8 µm Method As per IS 2092
0.001 mm & NABL-122-01

0 to 50 mm 3.0 µm Using Length Measuring


0 to 100 / 200 mm 3.0 µm Machine
$
7. Dial Gauge 0 to 0.2 mm / 0.001 2.7 µm Using Electronics Dial
(Lever Type) 0 to 2.0 mm / 0.01 4.0 µm Tester By Comparison
Method As per IS: 11498
& NABL-122-01

8. Micrometer Up to 100 mm 1.2 µm Using Slip Gauges,


Setting Bar / 100 mm to 300 mm 3.0 µm Electronic Comparator,
$
Setting Masters 300 mm to 1000 mm 7.0 µm Length Bars, Profile
Projector etc. By
Comparison Method

9. Plain Plug Gauge / Up to 25 mm 1.1 µm Using Slip Gauges,


$
Taper Plug Gauge 25 mm to 100 mm 2.5 µm Electronic Comparator,
100 mm to 300 mm 1.5 µm L.M.M By Comparison
Half Included angle 20 sec of arc. Method As per IS: 3455 ,
Up to 60 ° IS: 2251, IS: 9475
IS: 9529 & NABL-122-01
$
10. Snap Gauge 3 mm to 25 mm 0.8 µm Using Slip Gauges &
25 mm to 100 mm 1.2 µm L.M.M & By Comparison
100 mm to 750 mm 5.4 µm Method As per IS: 3455
& NABL-122-01

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 3 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

11. Plain Ring Gauge / 3 mm to 25 mm 1.5 µm Using Setting Ring


$
Taper Ring Gauge 25 mm to 100 mm 2.0 µm Gauges & Length
100 mm to 300 mm 2.5 µm Measuring Machine By
Half Included angle 30 sec of arc. Comparison Method As
Up to 60 ° per IS: 3455,
IS: 2251, IS:9475 &
NABL-122-01
$
12. Measuring Pins Up to 20 mm 1.2 µm Using Length Measuring
Machine By Comparison
Method As per
IS :11103 &
NABL-122-01

13. Thread Measuring 0.17 mm to 6.35 mm 1.2 µm Using Length Measuring


$
Wires Machine By Comparison
Method As per IS: 6311
& NABL-122-01
$
14. Feeler Gauge Up to 2 mm 1.2 µm Using Length Measuring
Machine By Comparison
Method As per IS : 3179
& NABL-122-01
$
15. Thickness Gauge 0 to 50 mm / 0.01 3.0 µm Using Slip Gauge Set &
50 mm to 100 mm / 0.01 7.0 µm By Comparison Method
$
16. Pistol Caliper 0 to 100 mm / 0.1 mm 58.0 µm Using Slip Gauge Set
By Comparison Method
$
17. Thickness Foils 0 to 2 mm 1.2 µm Using Electronic
Comparator By
Comparison Method

18. Coating Thickness 0 to 2 mm 6.0 µm Using Thickness Foils


$
Gauge By Comparison Method

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 4 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

$
19. Measuring Scale Up to 2 mtr (150*√2L) µm Using Length Measuring
For each additional L in meters Machine By Comparison
meter Method As per IS 3481
& NABL -122-01

20. Measuring Tape / 0 to 50 mtr (150*√2L) µm Using Length Measuring


$
PI Tape For each additional For any length, Machine By Comparison
0.5 Mtr L in m or first 0.5 m Method As per IS1269 &
(150*√2L) µm NABL -122-01
$
21. Height Gauge 0 to 300 mm 8 µm Using Slip Gauges &
(Dial / Digital/ Analog) 0 to 600 mm 10 µm Length Bars By
L.C.: 0.01 mm 0 to 1000 mm 15 µm Comparison Method
As per IS 2921 &
NABL-122-01

22. Dial Calibration 0 to 25 mm / 0.0001 mm 2 µm Using Slip Gauge,


$
Tester Electronic Probe By
Comparison Method
As per NABL-122-01
$
23. Thread Plug Gauge Up to 300 mm 3 µm Using Length Measuring
Machine

24. Taper Thread Plug Half included angle 3√2 µm By Comparison Method
$
Gauges Up to 60º As per I.S.:4218 &
NABL-122-01
$
25. Thread Ring Gauge Up to 300 mm 4 µm Using Length Measuring
Machine

26. Taper Thread Ring Half included angle upto 4√2 µm By Comparison Method
$
Gauge 60º As per I.S.:4218 &
NABL-122-01

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 5 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

$
27. Bore Gauge Only 2 mm 2 µm Using Length Measuring
L.C.: 0.001 / 0.01 mm Transmission Machine By Comparison
Mechanism Only Method As per
NABL-122-01

28. Angle Gauge 0º to 90º 3 min. Using Sine Setting


Bevel Protractor 0º to 360º Device Angle Gauge
$
Inclinometer Blocks As per IS 4239 &
NABL-122-01

29. Angle Gauge / 0 to 90º 20 sec. Using Sine Setting


$
Setting Gauge 0º to 180º Device Angle Gauge
Blocks & Slip Gauge &
Profile Projector As per
IS 4239 & NABL-122-01

30. Combination Set 0 to 360º 35 min. Using Slip Gauge Set


$
Angle Protractor Profile Projector
As per IS 4239 &
NABL-122-01

$
31. Radius Gauge Up to 100 mm 10 µm Using Profile Projector
By Comparison Method.
As per IS 5237 &
NABL-122-01
$
32. Thread Pitch Gauge 55, 60º 10 min Using Profile Projector
Pitch 0.3 mm to 6 mm 10 µm By Comparison Method.
As per I.S.4211 &
NABL-122-01
$
33. Comparator Stand Up to 300 mm 5.05 µm Using Plunger Dial
By Comparison Method.
As per I.S. 7599

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 6 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

34. Right Angle / 0 to 300 mm 11.4 µm Using V-Block, Plunger


$
Engineer’s Square 0 to 600 mm Dial, Master Cylinder,
S.P & Height Master
By Comparison Method.
As per I.S. 2103 &
NABL-122-01

35. V Block For Using Plunger Dial &


Parallelism 150 mm 5.7 µm Mandrel By Comparison
Squareness 150 mm 5.6 µm Method As per IS 2949
$
Symmetricity 150 mm 5.8 µm & NABL-122-01
$
36. Straight Edge Up to 5 mtr. 1.65 x (L+W /100) µm Using Digital Level
Meter & Slip Gauge
By Comparison Method.
As per IS 2220, IS12937
& NABL-122-01
$
37. Angle Plate Up to 500 mm 8.0 µm Using Plunger Dial &
Surface Plate
By Comparison Method.
As per I.S. 2554 IS 6232
& NABL-122-01
$
38. Glass Scale Up to 10 mm 3.0 µm Using L.M.M / By
Up to 300 mm 5.0 µm Comparison Method
& NABL-122-01
$
39. Test Sieve 20 µm to 100 mm 11.3 µm Using Profile Projector
By Comparison Method.
As per IS :460 Part.-I,
II & III & NABL-122-01

40. Electronic 0 to 25 mm 0.52 µm Using Slip Gauge Set


Comparator With By Comparison Method.
$
Probe NABL- 122- 01

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 7 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

41. Digital Height 0 - 600 mm / 0.0001 4.0 µm Using Slip Gauge Set,
#
Master 0 - 1000 mm / 0.0001 8.7 µm Length Bar By
Comparison Method
NABL- 122- 01

#
42. Surface Plate Up to 3 m x 3 m 3.0* (√2 L+W )/100 Using Electronic Level
(Flatness) ( Grade 0 / I / II ) (L & W are in mm) By Comparison Method
(Grade 0/I/II) As per IS 2285,
IS 7327 & NABL -122-01

$
43. Depth Micrometer 0 to 100 mm / 0.001 3.0 µm Using Length Measuring
L.C.: 0.001 mm 0 to 300 mm / 0.01 8.0 µm Machine Micrometer
L.C.: 0.01 mm Checking Set By
Comparison Method
As per BS : 6468 &
NABL-122-01
$
44. Surface Roughness Using Surface
a. Tester Up to 6.3 µm 8.0 % Roughness Specimen,
b. Test Block 7.5 % Reference Depth Master
(Specimens) By Comparison Method
As per IS : 3073,
IS:10707
NABL-122-01
$
45. Spherical Diameter 1 mm to 100 mm 2.8 µm Using Electronic
Comparator By
Comparison Method
As per NABL-122-01

$
46. Caliper Checker Up to 600 mm 3.0 µm Using Slip Gauge Set
Up to 1000 mm 7.41 µm and Length Bar
By Comparison Method

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 8 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

47. Spirit Level / Using Slip Gauge Set


$
Square Type By Comparison Method
L.C.: 0.02 mm / Meter Up to 300 mm 8.0 µm As per NABL-122-01

II. DIMENSION (PRECISION INSTRUMENTS)

1. Profile Projector / Up to 25 mm - Linear 2.0 µm Using Glass Scale,


#
Optical Microscope Up to 300 mm - Linear 7.6 µm Angle Gauge Set By
0 to 360 º - Angular 5 min. Comparison Method
Up to 1000 X – 0.5 % As per NABL-122-01
Magnification

III. PRESSURE INDICATING DEVICES


#
1. Hydraulic Pressure 0 to 40 bar 0.15 bar Using Digital Test
Gauge (Dial/Digital) 40 bar to 700 bar 1.1 bar Gauge & Pump
By Comparison Method
DKD-R6-1 &
NBAL-122-13

IV. WEIGHING SCALE AND BALANCE

1. Electronic 20 g to 5 kg 2g Using F-2 Class



Weighing Balance (d ≥ 1 g ) 20 g Standards Weights
200 g to 50 kg 1 mg- 200 g &
(d ≥ 10 g ) M1 Class Weights
1 to 50 kg Calibration of
class 3 Weighing
Balance and courser
as per OIML R-76

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 9 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

V. VOLUME

1. Volume Glasswares 10 ml ˂ V ˂ 1000 ml 0.6 ml Using Precision Balance


(All Types) Like, & Standard Weights
Burette, Pipette, Jar By Gravimetric Method
Measuring Flask As per ISO 4784 &
$
Cylindrical Flask etc. NBAL-122-13

VI. MOBILE FORCE MEASURING SYSTEM

1. Force : Up to 500 N 0.15% Pull Mode Using Newton weights &


$
Push Pull Gauge 0.13% Push Mode Fixture As per VDI/
VDE 2624 Part 2.1 &
As per NABL-122-08

VII. TORQUE GENERATING DEVICE

1. Torque Tools / 0.2 Nm to 20 Nm 1.50 % Using Torque Calibration


Torque Wrench / 20 Nm to 200 Nm 1.40 % Systems Transducers
Screw Driver / 200 Nm to 2000 Nm 1.30 % with Indicators As per
Torque Meter IS / ISO 6789:2003
( Type I-Class NABL -122-10
B,C,D,E &
Type II – Class
$
A,B,D,E)

VIII. HARDNESS TESTING MACHINES

1. Shore A Hardness 0 to 100 Shore A 2.10 Shore A Using Rubber Hardness


Tester Tester Calibrator
Shore D Hardness 0 to 100 Shore D 2.05 Shore D ASTM D 2240-05 &
$
Tester ISO : 18898 -2006 &
As per & NABL -122-15

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 10 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

IX. ACCELERATION & SPEED

1. RPM Using Digital


Tachometer Tachometer &
Contact Type 50 rpm to 5000 rpm 05 rpm Electric Motor with
$
Non Contact Type 50 rpm to 10000 rpm 10 rpm Rotating Disc.
Tacho Generator
By Comparison Method

2. Vibration : Acceleration: Acceleration 6.40 % Using Vibration Meter


2
Vibration Meter / Up to 200 m/s Calibrator By
$
Accelerometer (Peak) Comparison Method
a) Acceleration Velocity : Velocity 6.00 % As per ISO: 16063-21
b) Velocity & Up to 200 mm/s
c) Displacement (RMS)
Displacement : Displacement 2.60 %
Up to 1500 µm
(Peak to Peak)

X. ACOUSTICS
$
1. Sound Level Meter 94 dB and 114 dB 0.75 dB Using Sound Level
Range 40- 130 dB Meter Calibrator By
Comparison Method
As per IEC: 61672,
Part-I
XI. DENSITY AND VISCOSITY
$
1. Viscosity 10 cSt to 370 cSt 3.05 cSt Using Newtonian Liquid
Ford Cups All Type : of Known Kinematic
B1 To B5 Viscosity & Timer for
Calibration of Ford cups
based on ISO 2431 by
determining the Flow
Time

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 11 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

$
2. Viscosity 370 cSt to 1800 cSt 11.66 cSt Using Newtonian Liquid
(All Types Of of Known Kinematic
Viscosity Cups) Viscosity & Timer For
Calibration of Dip Type
Viscosity cups (Zahn &
Shell) Based on
ASTM 4212-16 by
determining the Flow
Time

Ashish Kakran Avijit Das


Convenor Program Director
Laboratory S. M. Engineers, 1210/B, Sadashiv Peth, Indira Apartment, Pune,
Maharashtra
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2639 (in lieu of C-0114, C-1366) Page 12 of 12
Validity 05.02.2018 to 04.02.2020 Last Amended on 13.04.2018

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

OPTICAL CALIBRATION

I. OPTICAL
$
1. Lux Meter Up to 2000/10000 lx 3.4% Using Certified Light
Source & Reference
Illuminance Meter By
Comparison Method
$
2. Lux Meter Above 10000 to 3.4% Using Certified Light
20000 lx Source & Reference
Illuminance Meter By
Comparison Method

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Ashish Kakran Avijit Das


Convenor Program Director

You might also like