Professional Documents
Culture Documents
(FKAAS)
BFC 32501
TRANSPORTATION AND ENVIRONMENTAL ENGINEERING
LABORATORY
SECTION 1
GROUP WORK
PROPOSAL
GROUP MEMBERS:
1) ALIS SYAKINA BINTI AHMAD AF160057
2) MUHAMMAD MAHATHIR BIN ABDUL SHUKUR AF160143
3) NORMADIANA BINTI NOH DF170056
4) NURUL ASHIKEN BINTI ALI AF160047
5) SITI NUR EZIANIE BINTI MOHAMAD A’ASHRI DF170121
SOURCE OF WATER
The source of water for sampling are from an artificial lake/reservoir (Kemajuan Lake), located
near G3 building in UTHM as shown in Figure 1 below. Figure 2 and Figure 3 shows the map of
the Kemajuan Lake that can be obtained from Google Maps.
PARAMETERS
According to standard methods for examination of water and wastewater, the parameters that will
be measured are:
Parameters Indicates
Biochemical Oxygen Demand (BOD) For determining how fast biological organism use
oxygen in body of water
Chemical Oxygen Demand (COD) Indicates the amount of organic pollution in water
Determination of pH value Contamination and acidification
Total Suspended Solid (TSS) Small solid particles which remain in suspension in
water as a colloid or due to motion in water
Dissolved Oxygen (DO) Measures amount of oxygen dissolved in water.
Ammonia Nitrogen Measure the amount of ammonia, a toxic pollutant
CALCULATION OF WATER QUALITY INDEX (WQI)
𝑾𝑸𝑰 = (𝟎. 𝟐𝟐∗ 𝑺𝑰𝑫𝑶) + (𝟎. 𝟏𝟗∗ 𝑺𝑰𝑩𝑶𝑫) + (𝟎. 𝟏𝟔∗ 𝑺𝑰𝑪𝑶𝑫) + (𝟎. 𝟏𝟓∗ 𝑺𝑰𝑨𝑵) +
(𝟎. 𝟏𝟔∗ 𝑺𝑰𝑺𝑺) + (𝟎. 𝟏𝟐∗ 𝑺𝑰𝒑𝑯) --------- Equation 1
where;
From calculation using equation 1, the result will be classified according to few classifications
which are:
Class Based
Pollution Status Based
Test preparation
Before starting
This test is a 5-day test. Complete all the steps carefully to make sure that the test does not have to be done again.
Prepare the dilution water with a BOD Nutrient Buffer Pillow. Make sure that the dilution water for this test does not contain
an oxygen demand or toxins. When incubated for 5 days at 20 °C, the dissolved oxygen concentration in the dilution water
must not change by more than 0.2 mg/L.
Carbonaceous BOD (CBOD) can be determined by the addition of nitrification inhibitor. A test for CBOD is recommended for
biologically-treated effluents, samples with bacterial seed, samples with biologically treated effluents and river water.
As an alternative, use the method BOD calculation-Graphical Method on page 6 to calculate the test results. The
graphical calculation method is also a tool for troubleshooting problems in BOD measurements. The graphical calculation
method is not approved for regulatory reporting.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
BOD bottle, 300 mL, glass, with glass stoppers and plastic caps 6
BOD bottle cap, plastic 6
Dilution water (refer to Prepare the dilution water on page 3) varies
Pipet, seriological, 1 mL, 5 mL and 10 mL 1
Incubator 1
Nitrification Inhibitor (for the CBOD test only) 1 bottle
Sample collection
The main consideration with sample collection is to prevent contamination of the sample
with atmospheric oxygen.
• Analyze the samples immediately. The samples cannot be preserved for later
analysis.
• Collect samples in 300 mL glass BOD bottles. Completely fill the bottles.
1
Test procedure
1. Identify five sample 2. Prepare the samples: 3. Use a pipet to add the 4. If the test is for CBOD,
volumes to use for this test. Gently stir the sample. sample volumes to five add two portions of
Refer to Select the sample 300‑mL BOD bottles. Nitrification Inhibitor
volumes on page 4. (approximately 0.16 g) to
Note: If the minimum each bottle to prevent the
sample volume is 3 mL or oxidation of nitrogen
more, determine the compounds. Record the
dissolved oxygen in the results as CBOD.
undiluted sample. Ignore
this determination if sewage
and settled effluents with a
dissolved oxygen content
near 0 mg/L are analyzed. If
industrial effluents and
disinfected samples are
analyzed, refer to
Interferences on page 8.
5. Fill each bottle with 6. Carefully insert a stopper 7. Prepare the blank: Fill 8. Use a probe or use a
prepared dilution water. in each bottle to prevent another 300‑mL BOD bottle titration procedure to
Refer to Prepare the dilution trapped air bubbles. Push with the prepared dilution measure the dissolved
water on page 3. To down on the stopper. Invert water. To prevent air oxygen concentration in
prevent air bubbles, pour the bottles several times to bubbles, pour the water each bottle. If titration is
the water down the inner mix. down the inner surface of used for the measurement,
surface of the bottle. the bottle. prepare two sets of BOD
bottles. Make sure to
measure the dissolved
oxygen of the blank.
1 Different sizes are available for different quantities of water (e.g., 3 L and 6 L).
2 Keep the raw sewage at 20 °C (68°F) and do not move for 24–36 hours before use. Pipet from the upper
portion of the sewage.
1. Keep the dilution water in clean jugs in an incubator at 20 °C (68°F). Shake the jugs
to saturate the water with air. As an alternative, loosely put the cap on the jugs and
wait a minimum of 24 hours before use.
2. (Optional) Use a small aquarium pump or an air compressor to saturate the water
with air. Make sure to use filtered air. Make sure that the filter does not grow bacteria.
3. Shake the BOD Nutrient Buffer Pillow to mix the contents.
4. Add the contents of the BOD Nutrient Buffer Pillow to the distilled water.
5. Put the cap on the jug. Shake the jug vigorously for 1 minute to dissolve the nutrients
and to saturate the water with air.
6. If the sample is known to be low in bacteria (e.g., industrial waste or sewage that has
been disinfected), immediately before the test, add 3 mL of raw sewage to each liter
of the dilution water.
Measure the BOD of the raw sewage collected. The BOD of the raw sewage will be
subtracted from the BOD of the sample.
3 Keep the phosphate buffer solution in a refrigerator to decrease the rate of biological growth.
NOTICE
Do not use the Graphical Method when the results must be given to a regulatory agency.
1. On a graph, record on the y-axis the remaining dissolved oxygen (DO) (mg/L) in each
of the prepared samples after 5 days. Record the sample volume (mL) on the x-axis.
Draw the best straight line through the plotted points. Refer to Figure 1.
Note: At least three points should be on the graph line or very near to the graph line. Ignore an
error point if seen at this time. For unseeded dilution water, the graph line should cross the
mg/L oxygen remaining scale near or at less than the oxygen saturation value for the altitude of
the laboratory as shown in Prepare the dilution water on page 3.
2. To calculate the BOD, use the equation that follows, which is mathematically the
same as the BOD equation. Refer to BOD calculation—Standard Methods on page 6.
mg/L BOD = (A × 300) – B + C
Where:
A = the slope of the graph line. The slope of the graph line is equal to the mg/L DO
consumed for each mL of sample. Select a point on the line and subtract the mg/L
Interferences
To get good BOD results, special handling is necessary to analyze chlorinated and
industrial effluents. Usually, careful experimentation with the sample shows the changes
that should be made to the test procedure. Toxins in the sample have an adverse effect
on any microorganisms in the sample, which results in lower BODs.
The substances in Table 5 interfere in the determination of oxygen demand at the given
concentrations.
Table 5 Interfering substances
Interfering substance Interference level
Chlorine Small quantities of residual chlorine—Let the sample sit for 1 to 2 hours at room temperature.
Larger quantities of chlorine—Refer to Remove chlorine from the sample on page 9.
Phenols Dilute the sample with high quality distilled water. As an alternative, acclimatize the bacterial seed
used in the dilution water to tolerate such materials. Refer to Acclimatize the bacterial seed
Heavy metals on page 9.
Cyanide
Highly buffered Less than pH 6.5 or more than pH 7.5 interfere. Adjust to pH 7.2 with acid (Sulfuric Acid, 1 N or
samples or extreme Phosphate Buffer Solution) or base (Sodium Hydroxide, 1 N).
sample pH
Cold temperature Cold samples can be supersaturated with oxygen and will have low BOD results. Fill a 1-liter (1-
quart) bottle ½ full with cold sample. Shake the bottle vigorously for 2 minutes. Let the sample
temperature increase to 20 °C (68 °F).
1. Fill a 4-liter (1-gallon) stainless steel or plastic container with domestic sewage.
2. Aerate the sewage for 24 hours.
3. Let the heavier material collect on the bottom for 1 hour.
4. Use a siphon to remove 3 liters (3 quarts) of the material from the top and discard.
5. Fill the container with a mixture of 90% sewage and 10% wastes that contain the
toxic material.
6. Aerate for 24 hours.
7. Do steps 3–5 again with more and more quantities of waste until the container holds
100% toxic waste material.
Accuracy check
Standard solution method
Use the standard solution method to validate the test procedure and the instrument.
Items to collect:
• 300-mg/L BOD Standard Solution4, Voluette Ampule, 10 mL
• Seeded dilution water
• Four 300-mL BOD bottles
• Pipet, volumetric, Class A, 1.0–4.0 mL
• TenSette Pipet
Summary of method
Biochemical oxygen demand (BOD) is a measurement of the oxygen requirements of
municipal and industrial wastewaters and sewage. The test results are used to calculate
the effect of waste discharges on the oxygen resources of the receiving waters. The BOD
test gives a limited value in the measurement of the actual oxygen demand because the
environmental factors (e.g., temperature change, biological population, water movement,
sunlight, oxygen concentration and other factors) cannot be simulated accurately in the
laboratory. The BOD test is a very important value after patterns of oxygen uptake for a
specified effluent and receiving water are identified.
For this test, a sealed wastewater sample (or a prepared dilution) is incubated for the
standard 5-day period. Then, the change in dissolved oxygen content is identified. The
BOD value is calculated from the results of the dissolved oxygen tests.
Consumables and replacement items
Required reagents
BOD Nutrient Buffer Pillow, 3 mL (for 3 L of dilution water) 1 pillow 50/pkg 1486166
Required apparatus
Recommended standards
© Hach Company/Hach Lange GmbH, 2007–2015, 2017. All rights reserved. 08/2017, Edition 10
Oxygen Demand, Chemical DOC316.53.01099
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows adapter and light shield requirements for the instruments that use them.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information for test tubes
Instrument Adapters Light shield
DR 6000, DR 5000 — —
DR 3900 — LZV849
DR 3800, DR 2800, DR 2700 — LZV646
DR 1900 9609900 (D1) —
DR 900 4846400 Cover supplied with the instrument
1 The D adapter is not available with all instrument versions.
Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started.
The reagent that is used in this test is corrosive and toxic. Use protection for eyes and skin and be prepared to flush any
spills with running water.
The reagents that are used in this test contain mercury. Collect the reacted samples for proper disposal.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Run one blank with each set of samples. Run all tests (the samples and the blank) with the same lot of vials. The lot number
is on the container label. Refer to Blanks for colorimetric determination on page 4.
Store unused (light sensitive) vials in a closed box.
If the samples contain high concentrations of chloride, refer to the Alternate reagents section.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
1
Items to collect
Description Quantity
Beaker, 250-mL 1
Blender 1
COD Digestion Reagent vials varies
DRB200 Reactor 1
Light shield or adapter (For information about sample cells, adapters or light shields, refer to
1
Instrument-specific information on page 1.)
Magnetic stirrer and stir bar 1
Opaque shipping container for storage of unused, light-sensitive reagent vials varies
Pipet, TenSette, 0.1- to 1.0-mL, with pipet tips (for use with the 200–15,000 mg/L range) 1
Pipet, volumetric, 2.00-mL 2
Pipet filler safety bulb 1
Test tube rack 2
1. Put 100 mL of sample in 2. For the 200–15,000 mg/L 3. Set the DRB200 Reactor 4. Prepare the sample:
a blender. Blend for range or to improve power to on. Preheat to Remove the cap from a vial
30 seconds or until accuracy and reproducibility 150 °C. for the selected range. Hold
homogenized. of the other ranges, pour the Refer to the DRB200 User the vial at an angle of
For samples with large homogenized sample into a Manual for selecting pre- 45 degrees. Use a clean
amounts of solids, increase 250‑mL beaker and gently programmed temperature pipet to add 2.00 mL of
the homogenization time. If stir with a magnetic stir applications. sample to the vial.
the sample does not contain plate. For 250–15,000 mg/L vials:
suspended solids, go to step Use a TenSette Pipet to add
3. 0.20 mL of sample to the
vial.
2 Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L)
5. Prepare the blank: 6. Close the vials tightly. 7. Hold the vials by the cap, 8. Put the vials in the
Remove the cap from a Rinse the vials with water over a sink. Invert gently preheated DRB200 reactor.
second vial for the selected and wipe with a clean paper several times to mix. Close the lid.
range. Hold the vial at an towel. The vials get very hot
angle of 45 degrees. Use a during mixing.
clean pipet to add 2.00 mL
of deionized water to the
vial.
For 250–15,000 mg/L vials:
Use a TenSette Pipet to add
0.20 mL of deionized water
to the vial.
9. Heat the vials for 10. Set the reactor power to 11. Invert each vial several 12. Put the vials in a tube
2 hours. off. Let the vials cool in the times while it is still warm. rack to cool to room
reactor for approximately temperature.
20 minutes to 120 °C or
less.
Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L) 3
Colorimetric procedure
Start Zero
1. Start program 431 COD 2. Clean the blank sample 3. Insert the blank into the 4. Push ZERO. The display
ULR, 430 COD LR or 435 cell. cell holder. shows 0 or 0.0 mg/L COD.
COD HR. For information
about sample cells,
adapters or light shields,
refer to Instrument-specific
information on page 1.
Note: Although the program
name can be different
between instruments, the
program number does not
change.
Read
5. Clean the prepared 6. Insert the prepared 7. Push READ. Results 8. If using High Range Plus
sample cell. sample into the cell holder. show in mg/L COD. COD digestion reagent
vials, multiply the result by
10. For the most accurate
results with samples near
1500 or 15,000 mg/L COD,
repeat the analysis with a
diluted sample.
1. Put the instrument in the absorbance mode at the applicable wavelength. Refer to
Table 3 on page 7.
2. Add 5 mL of deionized water into an empty vial.
3. Put the vial in the instrument and zero the instrument.
4. Put the blank vial that is used in the test procedure into the instrument and record the
absorbance value.
5. Keep the blank vial in the dark.
6. Prepare a new blank when the absorbance has changed by approximately
0.01 absorbance units.
4 Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L)
Interferences
Chloride is the primary interference in this test procedure. Each COD vial contains
mercuric sulfate that removes chloride interference to the level specified in Column 1 of
Table 2. Dilute samples that have higher chloride concentrations to the level given in
Column 2.
Note: For best results, use the low range and ultra-low range vials for samples that have high
chloride concentrations (near maximum concentration) and low COD concentrations.
If sample dilution causes the COD concentration to be too low for accurate
measurements, add 0.50 g of mercuric sulfate (HgSO4) to each COD vial before the
sample is added. The additional mercuric sulfate will increase the maximum chloride
concentration to the level given in Column 3.
Note: Bromide interference is not removed with mercuric sulfate.
Table 2 Chloride concentration limits in the sample
Vial range Column 1 (maximum mg/L Column 2 (mg/L Cl– for Column 3 (maximum mg/L
Cl–) diluted samples) Cl– with mercuric sulfate)
ULR1 (0.7–40.0 mg/L) 2000 1000 N/A
LR (3–150 mg/L) 2000 1000 8000
HR (20–1500 mg/L) 2000 1000 4000
HR Plus (200–15,000 mg/L) 20,000 10,000 40,000
1 The ULR is only available for spectrophotometers that can measure at a wavelength of 350 nm.
Accuracy check
Standard solution method
Items to collect:
• 1000 mg/L COD standard solution
• 100-mL volumetric flask, Class A
• Volumetric pipets, Class A and pipet filler
• Deionized water
• Potassium acid phthalate (KHP), dried overnight at 120 °C (HR Plus only)
0.7 to 40.0 mg/L ULR
1. Prepare a 30-mg/L COD standard solution as follows:
a. Use a pipet to add 3.00 mL of the 1000 mg/L standard solution into a 100-mL
volumetric flask.
b. Dilute to the mark with deionized water. Mix well.
2. Use the test procedure to measure the concentration of the standard solution.
3. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are slight
variations in the reagents or instruments.
Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L) 5
3 to 150 mg/L LR
1. Prepare a 100-mg/L COD standard solution as follows:
a. Use a pipet to add 10 mL of the 1000 mg/L standard solution into a 100-mL
volumetric flask.
b. Dilute to the mark with deionized water. Mix well.
2. Use the test procedure to measure the concentration of the standard solution.
3. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are slight
variations in the reagents or instruments.
20 to 1500 mg/L HR
1. Use the test procedure with a 300-mg/L, 800 mg/L or 1000 mg/L COD standard
solution to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are slight
variations in the reagents or instruments.
Alternate reagents
Mercury-free COD2 Reagents are available as a mercury-free alternative. These
reagents are fully compatible with test procedures and stored programs in the
instruments. Chloride and ammonia determinations are recommended for accurate
results.
NOTICE
COD2 reagents are not approved for USEPA reporting purposes. Because COD2 reagents do not
contain mercury as a masking agent, they exhibit a positive interference from chloride. More
information is available for use with specific applications.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% Confidence Interval) Sensitivity
Concentration change per 0.010 Abs change
431 (ULR) 30 mg/L COD 28.8–31.2 mg/L COD 0.5 mg/L COD
430 (LR) 80 mg/L COD 77–83 mg/L COD 3 mg/L COD
435 (HR) 800 mg/L COD 785–815 mg/L COD 23 mg/L COD
435 (HR Plus) 8000 mg/L COD 7850–8150 mg/L COD 230 mg/L COD
6 Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L)
Summary of method
The results in mg/L COD are defined as the milligrams of O2 consumed per liter of
sample under the conditions of this procedure. The sample is heated for 2 hours with
sulfuric acid and a strong oxidizing agent, potassium dichromate. Oxidizable organic
compounds react, reducing the dichromate ion (Cr2O72–) to green chromic ion (Cr3+).
When the 0.7–40.0 or the 3–150 mg/L colorimetric method is used, the amount of Cr6+
that remains is measured. When the 20–1500 mg/L or 200–15,000 mg/L colorimetric
method is used, the amount of Cr3+ that is produced is measured. The COD reagent also
contains silver and mercury ions. Silver is a catalyst, and mercury is used to complex
chloride interferences.
Test results are measured at the wavelengths that are specified in Table 3.
Table 3 Range-specific test wavelengths
Range in mg/L COD Wavelength
0.7–40.0 mg/L 350 nm (for applicable instruments)
3–150 mg/L 420 nm
20–1500 620 nm (610 nm for colorimeters)
200–15,000 mg/L 620 nm (610 nm for colorimeters)
COD, Ultra Low Range, 0.7–40 mg/L 1–2 vials 25/pkg 2415825
COD, Low Range, 3–150 mg/L 1–2 vials 25/pkg 2125825
COD, High Range, 20–1500 mg/L 1–2 vials 25/pkg 2125925
COD, High Range Plus, 200–15,000 mg/L 1–2 vials 25/pkg 2415925
Water, deionized varies 4L 27256
COD2, Low Range, 0–150 mg/L COD 1–2 vials 25/pkg 2565025
COD2, High Range, 0–1500 mg/L COD 1–2 vials 25/pkg 2565125
COD2, High Range, 0–1500 mg/L COD 1–2 vials 150/pkg 2565115
COD2, High Range Plus, 0–15,000 mg/L COD 1–2 vials 25/pkg 2834325
COD Digestion Reagent Vials, 3–150 mg/L COD 1–2 vials 150/pkg 2125815
COD Digestion Reagent Vials, 200–1500 mg/L COD 1–2 vials 150/pkg 2125915
COD Digestion Reagent Vials, ULR 0.7–40.0 mg/L 1–2 vials 150/pkg 2415815
COD Digestion Reagent Vials, HR plus, 200–25,000 mg/L 1–2 vials 150/pkg 2415915
Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L) 7
Required apparatus
8 Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L)
Optional reagents and apparatus (continued)
EZ COD™ Recycling Service with 5-gal bucket-mail back option (For US customers
each 2895405
only. 20 and 55 gallon sizes are also available. )
EZ COD™ Recycling Service with 5-gal bucket- pick up option. (For US customers
each 2895405P
only. 20 and 55 gallon sizes are also available. )
Finger cots 2/pkg 1464702
Gloves, chemical resistant, size 9–9.5 pair 24101041
Paper, for weighing, 100 x 100 mm 500/pkg 1473885
Safety goggles, vented each 2550700
Wastewater Effluent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO43–,
500 mL 2833249
COD, SO42–, TOC
1 Other sizes available
Oxygen Demand, Chemical, Dichromate Method (multi-range: 40.0, 150, 1500, 15,000 mg/L) 9
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY
In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – techhelp@hach.com FAX: (970) 669-2932
© Hach Company/Hach Lange GmbH, 1989–2014. All rights reserved. 10/2014, Edition 10
pH DOC316.53.01245
Test preparation
Before starting
Refer to the meter documentation for meter settings and operation. Refer to probe documentation for probe preparation,
maintenance and storage information.
Prepare the probe before initial use. Refer to probe documentation.
When an IntelliCAL™ probe is connected to an HQd meter, the meter automatically identifies the measurement parameter
and is prepared for use.
Condition the electrode for the best response time. To condition the electrode, soak the electrode for several minutes in a
solution that has almost the same pH and ionic strength as the sample.
Calibrate the probe before initial use. Refer to Calibration procedure on page 3.
For rugged electrodes, it may be necessary to remove the shroud before measurement and calibration.
Air bubbles under the sensor tip can cause slow response or measurement errors. To remove the bubbles, carefully shake
the probe.
To save data automatically, set the measurement mode to Press to Read or Interval. When the measurement mode is
Continuous, select Store to save data manually.
Rinse the electrode between measurements to prevent contamination.
1
Keep the electrode in a pH storage solution when not in use. Refer to the probe documentation.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
This procedure is specified for the HQd meters. The sensION+ meters can be used, but the menus and navigation will be
different.
Items to collect
Description Quantity
Sample collection
• Analyze the samples immediately. The samples cannot be preserved for later
analysis.
• Collect samples in clean glass or plastic bottles.
Test procedure
1. Rinse the probe with 2. Laboratory test: Put the 3. Push Read. A progress 4. Rinse the probe with
deionized water. Dry the probe in a beaker that bar is shown. When the deionized water. Dry the
probe with a lint-free cloth. contains the solution. Do not measurement is stable, the probe with a lint-free cloth.
let the probe touch the stir lock icon is shown.
bar, bottom or sides of the
container. Remove the air
bubbles from under the
probe tip. Stir the sample at
a slow to moderate rate.
Field test: Put the probe in
the sample. Move the probe
up and down to remove
bubbles from the electrode.
Make sure to put the
temperature sensor fully in
the sample.
1. Prepare two or three 2. Add a stir bar and put the 3. Rinse the probe with 4. Put the probe in the
fresh buffer solutions in beaker on a magnetic deionized water. Dry the solution. Do not let the
separate beakers. If two stirrer. Stir at a moderate probe with a lint-free cloth. probe touch the stir bar,
buffers are used, use a 7.0 rate. bottom or sides of the
and a 4.0 or a 7.0 and a container. Remove the air
10.0 pH buffer solution. bubbles from under the
probe tip.
5. Push Calibrate. The 6. Push Read. A progress 7. Measure the remaining 8. Push Done. A calibration
standard solution value is bar is shown. When the buffer solutions. summary is shown when the
shown. measurement is stable, the minimum number of
lock icon is shown. calibration standards are
measured.
NOTICE
Do not keep the probe in LIS samples for a long period of time because this can decrease the
probe life. Put the probe in electrode storage solution or 3 M KCl when LIS measurements are
complete.
Low ionic strength (LIS) solutions have very low buffering capacity and absorb carbon
dioxide from the air. When a sample absorbs carbon dioxide from the atmosphere,
carbonic acid forms. Carbonic acid decreases the sample pH, which causes inaccurate
Calibration accuracy
Measure the pH of a fresh buffer solution. A calibration is satisfactory when the measured
pH value agrees with the known pH value of the buffer solution.
Clean the probe
Clean the probe when:
• Drifting/inaccurate readings occur as a result of contamination on the sensing
element or incorrect storage conditions.
• Slow response time occurs as a result of contamination on the sensing element.
• The slope is out of range as a result of contamination on the sensing element.
For general contamination, complete the steps that follow.
1. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
2. Soak the glass bulb for 12 to 16 hours in Hach Electrode Cleaning Solution.
3. Rinse or soak the probe for 1 minute in deionized water.
4. Soak the probe in pH 4 buffer for up to 20 minutes, then rinse with deionized water.
5. Blot dry with a lint-free cloth.
6. If harsh contaminants are attached to the probe, polish the probe tip with a soft cloth
or cotton swab to remove the contaminants.
7. Soak for up to 20 minutes in pH 4 buffer, then rinse with deionized water.
Method performance
The accuracy of the measurements is dependent on many factors that are related with
the overall system, which includes the meter, the probe and calibration solutions. Refer to
the meter or probe documentation for more information.
Summary of method
A combination pH electrode develops a potential at the glass/liquid interface. At a
constant temperature, this potential varies linearly with the pH of the solution.
The pH is the hydrogen ion activity in a solution and is defined as –log10a(H+), where
a(H+) is the activity of the hydrogen ion. The sample pH can change when carbon dioxide
is absorbed from the atmosphere. In water that has a high conductivity, the buffer
capacity is typically high and the pH does not change much.
Recommended standards
Accessories
© Hach Company/Hach Lange GmbH, 2007–2015. All rights reserved. 03/2015, Edition 8
Suspended Solids DOC316.53.01139
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for reagent addition tests, such as
powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000 The fill line is toward the user.
DR 3900
DR 900 The orientation mark is toward the user. 2401906
Before starting
For turbidimetric methods, install the instrument cap or cover on all instruments before ZERO or READ is pushed.
Do not use the Pour-Thru Cell or sipper module (for applicable instruments) with this test.
Items to collect
Description Quantity
1
Sample collection and storage
• Collect samples in clean glass or plastic bottles.
• To preserve samples for later analysis, keep the samples at or below 6 °C (43 °F) for
up to 7 days.
• Let the sample temperature increase to room temperature before analysis.
Photometric procedure
Start
1. Start program 630 2. Blend 500 mL of sample 3. Pour the blended sample 4. Prepare the sample:
Suspended Solids. For in a blender at high speed into a 600-ml beaker. Stir the sample and
information about sample for exactly two minutes. immediately pour 10 mL of
cells, adapters or light the blended sample into a
shields, refer to Instrument- sample cell.
specific information
on page 1.
Note: Although the program
name can be different
between instruments, the
program number does not
change.
Zero
5. Prepare the blank: Fill a 6. Clean the blank sample 7. Insert the blank into the 8. Push ZERO. The display
second sample cell with cell. cell holder. shows 0 mg/L TSS.
10 mL of tap water or
deionized water.
9. Swirl the prepared 10. Clean the prepared 11. Insert the prepared 12. Push READ. Results
sample to remove any gas sample cell. sample into the cell holder. show in mg/L TSS.
bubbles and uniformly
suspend any residue.
Interferences
Samples that absorb strongly at the measurement wavelength, such as blue dyes, may
give false, high-bias readings. A user-entered calibration is advised for these samples.
Accuracy check
Standard solution method
Calibration for this test is based on the gravimetric technique with parallel sewage
samples from a municipal sewage plant. For most samples, this calibration supplies
satisfactory results. When higher accuracy is required, run parallel spectrophotometric
and gravimetric determinations with portions of the same sample. Make the new
calibration on the particular sample using a gravimetric technique as a basis.
Summary of method
This method of determining total suspended solids (TSS) is a simple, direct measurement
which does not require the filtration or ignition/weighing steps that gravimetric procedures
do. The USEPA specifies the gravimetric method for solids determinations, while this
method is often used for checking in-plant processes. The measurement wavelength is
810 nm (DR 1900: 800 nm) for spectrophotometers or 610 nm for colorimeters.
Consumables and replacement items
Required apparatus
© Hach Company/Hach Lange GmbH, 1989–2014. All rights reserved. 01/2014, Edition 9
Oxygen, Dissolved DOC316.53.01161
Test preparation
Before starting
As an alternative to the powder pillows, use standard APHA solutions for dissolved oxygen. Use 1 mL of Manganous Sulfate
Solution, 1 mL of Alkaline Iodide-Azide Reagent and 1 mL of Sulfuric Acid (concentrated) as an alternative to the powder
pillows. Add the solutions below the surface of the liquid.
The optional TitraStir Titration Stand can hold the buret and stir the sample.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
Sample collection
Good sample collection and handling techniques are important to get correct results. The
dissolved oxygen content of the sample can change with depth, turbulence, temperature,
sludge deposits, light, microbial action, mixing, travel time and other factors. A single
dissolved oxygen test frequently is not an accurate reflection of the overall condition of a
body of water. Several samples taken at different times, locations and depths are
recommended for most reliable results.
• Collect samples in clean BOD Bottles.
1
• If prompt analysis is not possible, do steps 1 through 4 of the procedure and keep the
samples protected from light at 10 to 20 °C (50 to 68 °F).
• Pour a small quantity of water into the flared lip area of the stopper to seal the bottle.
• Push a BOD bottle cap on the flared lip.
• Keep samples for a maximum of 8 hours. For analysis start with step 5.
Test procedure
1. Collect a water sample in 2. Add the contents of one 3. Immediately put the 4. Invert the bottle at least 5
a clean 300‑mL BOD bottle. Manganous Sulfate Powder stopper in the bottle. Make times to mix.
Let the sample overflow the Pillow and one Alkaline sure that no air is inside the A flocculent precipitate
bottle for 2 or 3 minutes to Iodide-Azide Reagent bottle. forms. The floc is
remove trapped air bubbles Powder Pillow to the orange/brown if oxygen is in
and to make sure that a sample. the sample or white if there
representative sample is is no oxygen.
collected. The floc forms slowly in salt
water (approximately
5 minutes more are
necessary). When the floc
formation is complete,
proceed to next step.
5. Again, invert the bottle at 6. Remove the stopper and 7. Immediately put the 8. Invert the bottle at least 5
least 5 times to mix. add the contents of one stopper in the bottle. Make times to mix.
Wait until the top half of the Sulfamic Acid Powder Pillow sure that no air is inside the The floc dissolves and a
solution is clear and the floc to the sample. bottle. yellow color develops if
collects at the bottom to oxygen is in the sample.
make sure that the reaction
of the sample and reagents
is complete. Some
suspended floc will have no
effect on accuracy.
2 Oxygen, Dissolved
9. Pour the prepared 10. Pour the contents of the 11. Fill a 25‑mL buret to the 12. Put the flask under the
sample into a 250‑mL graduated cylinder into a zero mark with 0.025 N buret. Swirl the flask. Add
graduated cylinder to the clean, 250‑mL Erlenmeyer Sodium Thiosulfate titrant until the color
200‑mL mark. flask. Standard Solution. changes yellow to a pale
yellow.
13. Add 1 mL of Starch 14. Put the flask under the 15. Calculate the
Indicator Solution. buret. Swirl the flask. Add concentration.
A dark blue color develops. titrant until the color mL of titrant = mg/L DO
changes from a dark blue to
a colorless end point.
Interferences
Nitrite interference is removed by the azide in the reagents. Other reducing or oxidizing
substances may interfere. If these are in the sample, use an alternative method, such as
the High Range Dissolved Oxygen Method (colorimetric, Method 8166) or a dissolved
oxygen electrode.
Accuracy check
Standard solution method
Use the standard solution method to validate the test procedure and the concentration of
the titrant.
Oxygen, Dissolved 3
Items to collect:
• Iodate-Iodide Standard Solution, 0.00125 N (equivalent to 10 mg/L as O2)
Summary of method
The Azide Modification of the Winkler Method is the standard test for dissolved oxygen. In
the analysis, manganous ion reacts with the dissolved oxygen in the alkaline solution to
form a manganese (IV) oxide hydroxide flocculent. Then, Azide is added to suppress
interference from nitrite, which would react with the iodide. Then, the solution is acidified
and the manganese (IV) floc is decreased by iodide to make free iodine as I3– in
proportion to the oxygen concentration. Then, the liberated iodine is titrated to the starch-
iodide end point.
Consumables and replacement items
Required reagents
Required apparatus
Recommended standards
4 Oxygen, Dissolved
Optional reagents and apparatus
APHA reagents: — —
Alkaline Iodide-Azide Reagent Solution 500 mL 27749
Manganous Sulfate Solution 500 mL 27549
Sodium Thiosulfate Standard Solution, 0.025 N 1L 35253
Sulfuric Acid, ACS 500 mL 97949
Cap, BOD Bottle Snap-over 6/pkg 241906
BOD Bottle, serialized (#1–24), 300 mL 24/pkg 2898700
Copper Sulfate-Sulfamic Acid Inhibitor Solution 100 mL 35732
Copper Sulfate-Sulfamic Acid Inhibitor Solution 500 mL 35749
Stir bar, octagonal each 2095352
TitraStir® Titration Stand, 115 VAC each 1940000
TitraStir® Titration Stand, 230 VAC each 1940010
Oxygen, Dissolved 5
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY
In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – techhelp@hach.com FAX: (970) 669-2932
© Hach Company/Hach Lange GmbH, 2007–2015. All rights reserved. 05/2015, Edition 8
Nitrogen, Ammonia DOC316.53.01078
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for specific instruments.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000 The fill line is toward the user.
DR 3900
Before starting
Hold the reagent droppers and dropper bottles vertically, not at an angle, when the reagent is added.
The reagents that are used in this test contain mercury. Collect the reacted samples for safe disposal.
If the Pour-Thru Cell is used, clean the cell periodically. To clean, add several crystals of sodium thiosulfate pentahydrate
into the cell funnel. Add deionized water to dissolve the crystals. Rinse fully with deionized water.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
1
Refer to Consumables and replacement items on page 5 for order information.
Test procedure
Start
1. Start program 380 N, 2. Prepare the sample: Fill 3. Prepare the blank: Fill a 4. Add 3 drops of Mineral
Ammonia, Ness. For a mixing cylinder to the 25- mixing cylinder to the 25‑mL Stabilizer to each mixing
information about sample mL line with sample. line with deionized water. cylinder.
cells, adapters or light
shields, refer to Instrument-
specific information
on page 1.
5. Put the stopper on the 6. Add 3 drops of Polyvinyl 7. Put the stopper on the 8. Use a pipet to add 1.0
mixing cylinders. Invert the Alcohol Dispersing Agent to mixing cylinders. Invert the mL of Nessler Reagent to
mixing cylinders several each mixing cylinder. mixing cylinders several each mixing cylinder.
times to mix. times to mix.
Zero
13. Insert the blank into the 14. Push ZERO. The 15. Pour 10 mL from the 16. Clean the prepared
cell holder. display shows 0.00 mg/L sample cylinder into a sample cell.
NH3–N. second sample cell.
Read
Interferences
Table 2 Interfering substances
Interfering substance Interference level
Chlorine Remove residual chlorine from a 250 mL sample by adding 1 drop of sodium thiosulfate for each
mg/L chlorine (Cl2). Sodium arsenite can be used instead of sodium thiosulfate. Refer to Sample
collection and storage on page 2.
Hardness A solution containing a mixture of 500 mg/L CaCO3 and 500 mg/L Mg as CaCO3 does not
interfere. If the hardness concentration is more than these concentrations, add extra Mineral
Stabilizer.
Iron Interferes at all levels by causing turbidity with Nessler Reagent.
1. Set up the distillation apparatus for general purpose distillation. Refer to the
Distillation Apparatus manual for proper assembly.
2. Measure 250 mL of sample into a 250-mL graduated cylinder.
3. Pour the sample into a 400-mL beaker. If the sample contains chlorine, add 1 drop of
0.1 N sodium thiosulfate solution for each 1 mg/L Cl2 to remove the chlorine.
4. Add 25 mL of borate buffer solution and mix. Adjust the pH to approximately 9.5 with
1 N sodium hydroxide solution. Use a pH meter to monitor the pH.
5. Pour the solution into the distillation flask.
6. Add a magnetic stir bar and 5 glass beads.
7. Use a graduated cylinder to measure 25 mL of deionized water into a 250-mL
Erlenmeyer flask. Add the contents of one Boric Acid Powder Pillow. Mix thoroughly.
8. Set the flask under the distillation apparatus drip tube. Elevate the flask so that the
end of the tube is immersed in the solution.
9. Set the stirrer power to on. Set the stir control to 5.
10. With the thermometer inserted, set the heat control to 10. The yellow pilot lamp is an
indication that the heater is on.
11. Turn on the water and adjust to maintain a steady flow through the condenser.
12. When 150 mL of distillate has been collected, turn the heater off. Immediately remove
the collection flask. Measure the distillate to make sure 150 mL was collected (total
volume = 175 mL).
13. Adjust the pH to approximately 7 with 1 N sodium hydroxide solution. Use a pH meter
to monitor the pH.
14. Quantitatively transfer the distillate into a 250-mL volumetric flask. Dilute to the mark
with deionized water. Mix well. Use the diluted distillate in the test procedure.
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• 50-mg/L Nitrogen-Ammonia Standard Solution
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 25-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
1. Use the test procedure to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are small
variations in the reagents or instruments.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% confidence interval) Sensitivity
Concentration change per 0.010 Abs change
380 1.00 mg/L NH3–N 0.99–1.01 mg/L NH3–N 0.02 mg/L NH3–N
Summary of Method
The Mineral Stabilizer complexes hardness in the sample. The Polyvinyl Alcohol
Dispersing Agent helps the color formation in the reaction of Nessler Reagent with
ammonia and certain other amines. A yellow color is formed that is proportional to the
ammonia concentration. The measurement wavelength is 425 nm.
Consumables and replacement items
Required reagents
Required apparatus
Recommended standards
© Hach Company/Hach Lange GmbH, 2007, 2010, 2012, 2014, 2017. All rights reserved. 04/2017, Edition 9