Professional Documents
Culture Documents
H50P - Operator's Manual - V7.0 - EN PDF
H50P - Operator's Manual - V7.0 - EN PDF
Operator’s Manual
© 2015-2016 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issued Date is 2016-12.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel.
the electrical installation of the relevant room complies with the applicable national and
local requirements; and
I
WARNING
It is important for the hospital or organization that employs this equipment to carry
out a reasonable service/maintenance plan. Neglect of this may result in machine
breakdown or personal injury.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will be
unreliable, which would damage the analyzer components and cause personal
injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
Malfunction of the instrument or part whose serial number is not legible enough.
III
Customer Service Department
Tel: 0049-40-2513175
Fax: 0049-40-255726
IV
Table of Contents
1 Using This Manual ............................................................................................... 1-1
1.1 Introduction ............................................................................................................ 1-1
1.2 Who Should Read This Manual ............................................................................. 1-2
1.3 How to Find Information ........................................................................................ 1-3
1.4 Safety Information ................................................................................................. 1-4
1.4.1 Protection from Biohazards ........................................................................... 1-4
1.4.2 Protection from Other Hazards ..................................................................... 1-5
1.4.3 Using the System .......................................................................................... 1-5
1.4.4 Maintaining the System ................................................................................. 1-6
1.4.5 Using Reagents ............................................................................................. 1-6
1.4.6 Disposing of Waste........................................................................................ 1-7
1.5 Symbols ................................................................................................................. 1-8
1
Table of Contents
2
Table of Contents
3
Table of Contents
Appendices ...........................................................................................................................A-1
A Index ...................................................................................................................... A-1
B Specification .......................................................................................................... B-1
C Communication Protocol .......................................................................................C-1
4
1 Using This Manual
1.1 Introduction
This chapter explains how to use your H50P operator’s manual, which is shipped with your
H50P Automated Glycohemoglobin Analyzer (hereinafter referred to “the analyzer") and
contains reference information about the analyzer and procedures for operating,
troubleshooting and maintaining the analyzer. Read this manual carefully before operating
your H50P analyzer and operate your H50P Automated Glycohemoglobin Analyzer strictly as
instructed in this manual.
1-1
Using This Manual
1-2
Using This Manual
1-3
Using This Manual
Symbols: It means...
read the statement below the symbol . The statement is
alerting you to a potentially biohazardous condition.
All the samples, reagents, calibrators, controls, wastes and areas contacted by
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
All the analyzer components and surfaces are potentially infectious, so take
proper protective measures for operation and maintenance.
When the analyzer is with error and need servicing, it is recommended to put
warnings on the analyzer to prevent people touching it and avoid biohazard and
other hazards.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the sharp sample probe when working around it.
WARNING
If the reagents accidentally spill on your skin, wash them off with plenty of water
and if necessary, go see a doctor; if the reagents accidentally spill into your eyes,
wash them off with plenty of water and immediately go see a doctor.
1-4
Using This Manual
WARNING
If any abnormal smell or fog are observed, power off the system and disconnect
the power plug immediately, or else fire, electric shock or other hazards may
occur.
Do not touch the power cords inside the system, and make sure your hands are
dry when operating the system to avoid electric shock.
Please check the firmness of all the doors, covers and boards before running the
system, and make sure they will not get open or loose during operation.
Make sure all the safety measurements are adopted. It is prohibited to disable any
safety device or sensor.
Do not touch the moving parts. Keep your clothes, hairs and hands away from the
moving parts to avoid injury.
CAUTION
Please use the analyzer strictly as instructed by this manual.
Keep blood, reagent and any other metal substances from getting inside the
system, or fog may be generated due to short circuit.
Using pinboard may bring the electrical interference and the analysis results may
be unreliable. Please place the analyzer near the electrical outlet to avoid using
pinboard.
Do not damage the power cord, put any heavy object upon it or pull it forcefully.
Power off the system before connecting it to external devices (e.g., computer), or
else fire or electric shock may occur.
If any of the pipes or fluidic components are worn out, stop using the analyzer and
contact Mindray customer service department immediately for inspection or
replacement.
The user should ensure the data safety of the USB devices connecting to the
analyzer.
1-5
Using This Manual
WARNING
It is important for the hospital or organization that employs this system to carry out
a reasonable service/maintenance plan. Neglect of this may result in machine
breakdown or injury of human health.
Be sure to wear rubber gloves and use specified tools and accessories when
inspecting, servicing or maintaining the system. Wash your hands with disinfectant
after finishing the operations.
NOTE
If the system stop running due to errors, the operator may deal with the errors as
instructed by the Operator's Manual; in case of errors not specified in the manual,
contact Mindray customer service department.
CAUTION
Use the reagents specified by the manufacturer only. Store and use the reagents
as instructed by instructions for use of the reagents.
Check if the reagent tubings are properly connected before using the analyzer.
After installing a new bag of reagent, keep it still for a while before use.
Do not use the reagents that have been contaminated or affected by other factors.
Replace it with normal one.
The test may be interfered by the following factors: expired or invalid reagents;
reagents being contaminated by dust in the air; unclean samples; mix or
combined use with reagent produced by other factories; mix use of residual
reagent in the bag with newly opened reagent; and improper use under
undesirable conditions.
1-6
Using This Manual
WARNING
Discard the system according to government regulations.
1-7
Using This Manual
1.5 Symbols
Symbols used in this manual:
Symbols: It means...
read the statement below the symbol . The statement is
alerting you to a potentially biohazardous condition.
CAUTION
During the daily use of the system, especially the cleaning process, the operator
shall ensure the intactness of the labels.
Symbols: It means...
BIOLOGICAL RISK
PROTECTIVE EARTH
EARTH (GROUND)
ALTERNATING CURRENT
1-8
Using This Manual
SERIAL NUMBER
DATE OF MANUFACTURE
NETWORK INTERFACE
PRICKING DANGER
1-9
Using This Manual
1.
Warning
1. Connect only to a properly earth grounded outlet.
2. To avoid electric shock, disconnect power cord prior to troubleshooting.
3. Replace fuses only with the type and rating specified.
2.
Biohazard
1-10
Using This Manual
Warning
The probe is sharp and may contain biohazardous material exercise caution when
working around the probe!
1-11
Using This Manual
To avoid personal injury, do not put your hand under the syringe or inside the slot!
1-12
2 Understanding Your Analyzer
2.1 Introduction
The H50P Automated Glycohemoglobin Analyzer is used for in vitro quantitative
determination of the percentage (%) and substance content (mmol/mol) of hemoglobin A1c
(HbA1c) in human blood samples.
The analyzer uses an analytical column to separate various hemoglobin components and
provides the analysis results of HbA1c (%, NGSP) and HbA1c (mmol/mol, IFCC), as well as a
chromatogram displaying various components of HbA1a, HbA1b, HbF, LA1c, SA1c (HbA1c)
and A0; in the mean while, the estimated average blood glucose (eAG) is calculated with a
formula produced by ADA, EASD and IDF (Diabetes Care 2008;31:1-6).
This chapter introduces the components, functions and user interface of H50P.
NOTE
The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that require
additional studies.
The software screenshots, images and certain details related to the product are
for illustration purposes only, and may not exactly match your real product.
2-1
Understanding Your Analyzer
2.2 Mode
Loading Mode
The H50P provides two loading modes: the autoloading mode and closed-tube mode.
Sample Mode
The H50P supports two sample modes: the whole blood mode and the prediluted mode.
2-2
Understanding Your Analyzer
Touch screen
The touch screen locates on the front side of the main unit, which can be used to operate the
analyzer and display information.
2-3
Understanding Your Analyzer
CAUTION
Do not use anything sharp on the touch screen or strike on it.
Clean the touch screen with clean and soft cloth, as well as neutral cleanser or
ethanol rather than chemical solution, acid or alkali solution.
Status indicator
The indicator locates on the top of the touch screen; and it tells you about the status of the
instrument including ready, running, error, standby and on/off, etc. See below table for details:
2-4
Understanding Your Analyzer
Waste connector
The waste connector locates at the left side on the back of analyzer. It connects the waste
tube to discharge the waste liquid. You can select to use a waste container or discharge the
waste directly. For the setup of waste discharge method, please refer to 7.6 Auxillary (“Setup”>
“Auxillary”).
.
2-5
Understanding Your Analyzer
2-6
Understanding Your Analyzer
2.3.3 Peripherals
Barcode reader
Recorder (optional)
2-7
Understanding Your Analyzer
2.3.4 Accessories
Analytical column
The analytical column is an important measurement component of the analyzer. It is installed
in the column temperature-control box.
The WB tube rack (see below) has a 3-digit number marking (for example 005) on both sides
of the tube rack, and a 2-digit number marking (for example 05) between the No. 9 and No.
10 tube positions.
2-8
Understanding Your Analyzer
The PD tube rack (see below) has the “PD” marking both on two sides of the tube rack and
between the No. 9 and No. 10 tube positions.
The CRL tube rack (see below) has the “CRL” marking both on two sides of the tube rack and
between the No. 9 and No. 10 tube positions.
The CAL tube rack (see below) has the “CAL” marking both on two sides of the tube rack and
between the No. 9 and No. 10 tube positions.
2-9
Understanding Your Analyzer
Adapters
3 types of adapters are used with the analyzer: φ13×75(mm) adapter for closed-tube mode;
and 12 × 27(mm) micro sample cup adapter and 1.5ml centrigual tube adapter for
autoloading mode. For more information regarding the replacing of the adapters, refer to
5.6.3 and 5.7.5 Replacing the Adapter.
2-10
Understanding Your Analyzer
2-11
Understanding Your Analyzer
2.4.3 Analysis
The analyzer supports the data collection (acquisition), storage, analysis and output (report)
functions under both whole blood and prediluted modes and outputs analysis results for
several parameters.
2.4.5 Calibration
The analyzer supports calibration function; and users may define the calibration factors by
themselves.
2.4.6 Maintenance
The analyzer supports maintenance time reminding, maintenance recording and procedure
prompting functions.
2-12
Understanding Your Analyzer
2.4.9 Communication
The analyzer supports LIS communication function.
2.4.10 Logs
The analyzer logs user operations and errors. The user may save and export such logs.
2.4.11 Printing
The user may print analysis reports.
2-13
Understanding Your Analyzer
Client area
Error
Auxiliary information System time
information area
area
Menu
Tap the menu icon on the upper left of the screen to display the system menu. Tap a
menu item to display the sub-menu or enter the corresponding menu screen or dialog box.
2-14
Understanding Your Analyzer
Menu Tree
No Level 1 Level 2 Level 3
1. Sample Analysis / /
2. Table Review / /
Setup /
3. QC Graph /
Table /
Manual (for administrators only) /
Calibrator (for administrators /
4. Calibrate only)
Touch Screen (for /
administrators only)
Date/Time (for administrators /
only)
User Management /
Parameter (for administrators /
only)
Shortcut Code /
Auxillary (for administrators /
only)
5. Setup Maintenance (for administrators /
only)
Barcode Setup (for /
administrators only)
Print Setup /
Communication /
Gains (for administrators only) /
Rack Type (for administrators /
only)
Consumables Reagents
Column (for
administrators only)
Filter (for administrators
6. Maintenance only)
Reagents /
Fluidics (for administrators only) Packup (for
administrators only)
Logs (for administrators only) /
Temp.&Pressure /
7. Status
Version Info. /
8. Logout / /
Shut Down the / /
9.
Analyzer
2-15
Understanding Your Analyzer
Shortcut Buttons
Client area
Displays related information of the screen. For example, the "Sample Analysis" screen
displays "Next Sample ID" and "Presentation Mode".
System time
depending on the contents to be entered into the text box), see the following figure. Tap
to close the pop-up keyboard.
2-16
Understanding Your Analyzer
Table
You will see tables on the screens like "Review". Tap the buttons on the right and below the
table to view information in the table. These buttons are: (previous page), (next
You can shift the cursor in the text box by pressing [Tab] and [Shift]+[Tab].
2-17
Understanding Your Analyzer
Check box
Tap the check box, a "√" mark appears in the frame, indicating the option is chosen.
Tap the option again, the “√” disappeared, it means the option is not chosen as shown
below.
Tab screen
Tab screen displays one page of the multipage information.
2-18
Understanding Your Analyzer
2.6 Reagents
You must only use the Mindray specified reagents in order to prevent analyzer damage and
achieve optimal system performance. All references related to reagents in this manual refer
to the reagents specifically formulated for this instrument.
Each reagent package must be examined before use. Product integrity may be compromised
in damaged packages. Inspect the package for signs of leakage or moisture. If there is
evidence of leakage or improper handling, do not use the reagent.
NOTE
Store and use the reagents as instructed by instructions for use of the reagents.
Pay attention to the expiration dates and open-bag stability days of all the
reagents. Be sure not to use expired reagents.
After installing a new bag of reagent, keep it still for a while before use.
2.6.1 Reagents
Eluent A and Eluent B (for HbA1c analysis)
Used with the HbA1c analysis system to determine the HbA1c percentage and substance
content in the blood samples.
Hemolysis Solution
Used for the pre-process of blood sample before the HbA1c analysis.
2-19
3 System Principles
3.1 Introduction
H50P uses the ion-exchange-based high-performance liquid chromatography (HPLC)
method to build the Hb chromatogram and calculate relevant parameters. With the HPLC
method, the blood sample containing various types of hemoglobin is loaded to an analytical
column. The HbA1c, which hardly has any positive charge, is the first being eluted by eluent
of the pre-defined low ionic strength and pH conditions; and the positive charged HbA0 is
eluted by the eluent of high ionic strength. In this way, the analyzer builds up the Hb
chromatogram, and calculates the ratio of HbA1c peak value to the total Hb.
3-1
System Principles
3-2
System Principles
3-3
4 Installing Your Analyzer
4.1 Introduction
WARNING
Installation by personnel not authorized or trained by Mindray may cause personal
injury or damage your analyzer. Do not install your system without the presence of
personnel authorized by Mindray.
The analyzer was carefully packaged and labels were placed to tell the carrier how to treat
this analyzer. When you receive your analyzer, carefully inspect the carton. If you see any
signs of mishandling or damage, contact Mindray customer service department or your local
distributor immediately.
4-1
Installing Your Analyzer
WARNING
Make sure the analyzer is properly grounded.
Before turning on the analyzer, make sure the input voltage meets the
requirements.
CAUTION
Using pinboard may bring the electrical interference and the analysis results may
be unreliable. Please place the analyzer near the electrical outlet to avoid using
pinboard.
Use the power cord provided by the manufacturer. Using the power cord other
than provided by the manufacturer may lead to system damage or unqualified
smear output.
NOTE
Only store and use this system in specified environmental conditions.
4-2
Installing Your Analyzer
The environment shall be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
It is advisable to evaluate the electromagnetic environment prior to operation of this
analyzer.
Do not use this analyzer in close proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF sources), as these may interfere with the proper
operation.
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
The environment shall be well ventilated.
Do not place the analyzer on a slope.
Connect only to a properly earth grounded outlet.
Only use this analyzer indoors.
WARNING
Installation by personnel not authorized or trained by Mindray may cause personal
injury or damage your analyzer. Do not install your analyzer without the presence
of Mindray-authorized personnel.
4-3
Installing Your Analyzer
Make sure that the top of the waste container is lower than the analyzer desk. If
the waste is discharged directly, make sure the waste pump is at a lower position
than the waste outlet on the analyzer.
4-4
Installing Your Analyzer
4-5
Installing Your Analyzer
4-6
5 Operating Your Analyzer
5.1 Introduction
This chapter provides step-by-step procedures for operating H50P analyzer on a daily basis.
A flow chart indicating the common daily operating process is presented below.
Initial Checks
Power On
Sample Preparation
See 5.5 Sample Preparation
Sample Analysis
Closed Tube Mode Autoloading Mode
See 5.6 Sample Analysis Under See 5.7 Sample Analysis Under
Closed Tube Mode Autoloading Mode
Shutdown
See 5.9 Shut Down the Analyzer
5-1
Operating Your Analyzer
All the samples, reagents, calibrators, controls, wastes and areas contacted by
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc. according to
government regulations.
If the reagents accidentally spill on your skin, wash them off with plenty of water
and if necessary, go see a doctor; if the reagents accidentally spill into your eyes,
wash them off with plenty of water and immediately go see a doctor.
Do not touch the moving parts. Keep your clothes, hairs and hands away from the
moving parts to avoid injury.
Check whether there are enough reagents for the test of the day. If not, replace the reagents.
If the reagents run out during analysis, the analyzer will pause working automatically and
prompt the operator to replace the reagents. Please follow the instruction to replace the
reagents.
Check and make sure the reagent, waste and pneumatic unit tubes are properly connected
and not bent.
Check and make sure the power cord of the analyzer is properly plugged into the power
outlet.
5-2
Operating Your Analyzer
5.3 Startup
5.3.1 Power Switch
Press the [Start] button on the analyzer.
NOTE
You should remove all the errors before proceeding with any operation.
When the analyzer has been idle for more than 72 hours, the analyzer self-test will
take longer time.
Check the status area at the lower part of the screen, and make sure all the reagents are
sufficient; and the analytical column does not reaches their maximum uses.
5-3
Operating Your Analyzer
5-4
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted them
are potentially biohazardous. Wear proper personal protective equipment (e.g.
gloves, lab coat, etc.) and follow safe laboratory procedures when handling them
in the laboratory.
WARNING
Do not contact the patients' sample blood directly.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
NOTE
Be sure to use clean anticoagulant collection tubes (recommended
anticoagulants: K2EDTA, K3EDTA, heparin lithium salt, sodium citrate, potassium
oxalate monohydrate/sodium fluoride), fused silica glass/plastic test tubes,
centrifugal tubes and borosilicate glass capillary tubes.
CAUTION
To ensure accurate analysis results, the sample volume should be no less than
1mL under whole blood mode.
5-5
Operating Your Analyzer
The samples stored under refrigeration condition (2℃-8℃) must be kept at room
temperature for at least 30 minutes before analysis.
When a sample has been kept for a while after preparation, be sure the mix it
again before running it.
In case of insufficient sample volume or very low hemoglobin concentration, the analyzer may
report “Chromatogram area too small”. Predilute the sample as required, and run the
sample using predilute mode.
When the analyzer reports “Chromatogram area too small” for a sample, predilute the
sample as required, and run the sample again using predilute mode.
Add certain amount of sample blood (recommended: 10ul) into a 1.5ml centrifugal tube
1. (or micro sample cup);
If the small area flag is reported because of sample volume is insufficient, use a
2. pipette to dispense lyse into the sample at the ratio of 1:100 (lyse to blood sample) and
well mix it;
If the small area flag is reported because the HGB of the whole blood sample is
too low, use a pipette to dispense lyse into the sample at the ratio of 1:50 (lyse to
blood sample) and well mix it.
The analyzer supports auto dispensing of Hemolysis Solution. For more information, please
refer to 5.10 Auto Dispensing Hemolysis Solution.
CAUTION
Take methods to prevent the diluent from dust; otherwise the results may be
unreliable.
Be sure to run the prediluted samples within 30 minutes after the dilution;
otherwise the results may be unreliable.
Be sure to mix any sample that has been prepared for a while before running it.
To ensure accurate analysis results, the sample volume should be no less than
0.5mL under predilute mode.
5-6
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted them
are potentially biohazardous. Wear proper personal protective equipment (e.g.
gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in
the laboratory.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
Repeatedly piercing the cap of the evacuated blood collection tube may damage
the cap, and the scraps produced may result in inaccurate analysis results. It is
recommended that each tube is pierced for no more than 3 times.
To ensure the correctness of analysis results, do not run prediluted samples under
the Closed Tube mode.
NOTE
Proper reference range shall be entered on the "Setup"- “Parameter” screen before
analysis. Otherwise, the results may be flagged erroneously.
If the 2-way LIS/HIS mode is selected, then after the sample ID is entered/scanned
and saved, all the corresponding information will be obtained from the LIS/HIS, and
then the analyzer starts running per the obtained information. Once the running is
finished, the result, graph and sample/patient information will be uploaded to the
LIS/HIS.
5-7
Operating Your Analyzer
When it is ready to run a sample (i.e. the analyzer indicator stays in green), shake the
1. tube of whole blood sample as instructed by the picture below to mix the sample
thoroughly.
Press the [Open] key on the analyzer to open the sample compartment door. Install
2. theФ13×75 (mm) adapter into the sample compartment. Place the sample into the
adapter.
Tap the "Mode" button on the "Sample Analysis" screen, and the "Mode Setup"
4. screen will display.
Check the "CT" radio button for "Mode" and enter the "Start Sample ID".
5. If necessary, check "Auto Increase", and the subsequent sample IDs will
automatically increase by 1 based on the previous one.
6. Tap "OK" on the dialog box or simply press [Enter] on the keyboard to finish setup.
Tap the "Start" button to start the analysis. After the analyzer finishes sample
7. aspiration, the sample compartment door opens automatically, and the analyzer
indicator flickers in green.
When the analysis completes, the analyzer indicator returns to "Ready" status (stay in
8. green).
5-8
Operating Your Analyzer
Parameter flags
If the parameter is marked by a high/low flag ("H"/"L"), it means the analysis result has
exceeded the upper or lower limit of the reference range, but still within the display range.
If a result is indicated by “****”, it means the result is invalid or out of the display range.
If a result is indicated by “@”, it means the result is out of the linear range.
Chromatogram flags
Messages Meanings
Analysis not completed Analysis not completed
Abnormal signal The analyzer signals may become abnormal
The analyzer does not aspirate sample properly; or the
Chromatogram area too small
hemoglobin concentration of the sample may be too low
The hemoglobin concentration of the sample may be too
Chromatogram area too large
high
Abnormal peak number Abnormal variant may be present
SA1c peak not properly
The measured HbA1c% value may be inaccurate
separated
HbA0 peak abnormal The measured HbA1c% value may be inaccurate
Chromatogram abnormal The measured HbA1c% value may be inaccurate
Shorter SA1c retention time The analyzer may be in abnormal status
Longer SA1c retention time The analyzer may be in abnormal status
Shorter HbA0 retention time The analyzer may be in abnormal status
Longer HbA0 retention time The analyzer may be in abnormal status
5-9
Operating Your Analyzer
5-10
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted them
are potentially biohazardous. Wear proper personal protective equipment (e.g.
gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in
the laboratory.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
When a sample has been kept for a while after preparation, be sure the mix it again
before running it.
Repeatedly piercing the cap of the evacuated blood collection tube may damage
the cap, and the scraps produced may result in inaccurate analysis results. It is
recommended that each tube is pierced for no more than 3 times.
Be sure that the entered sample ID, rack No., tube No. and the analysis mode are
strictly in accordance with the sample to be run.
NOTE
Proper reference range shall be entered on the "Setup"-“Parameter” screen before
analysis. Otherwise, the results may be flagged erroneously.
If the 2-way LIS/HIS mode is selected, then after the sample ID is entered/scanned
and saved, all the corresponding information will be obtained from the LIS/HIS, and
then the analyzer starts running per the obtained information. Once the running is
finished, the result, graph and sample/patient information will be uploaded to the
LIS/HIS.
5-11
Operating Your Analyzer
When it is ready to run a sample (i.e. the analyzer indicator stays in green), shake
1. the tube of whole blood sample as instructed by the picture below to mix the
sample thoroughly.
2. Tap the "Mode" button on the "Sample Analysis" screen, and the "Mode Setup"
dialog box will display.
4. Enter the "Start Sample ID" (skip the step if you are using the whole blood mode
and have selected "Auto Scan ").
5. To run a whole blood sample, directly place the prepared sample tube into the
whole blood sample rack.
To run a prediluted sample, fit the prepared sample tube (or micro sample cup)
uncapped into a corresponding adapter (clip the tube cap into the adapter notch)
and place the tube together with the adapter into an even-numbered position of the
PD tube rack.
6. Place the rack with tubes on the right tray of the autoloader, with the side with
openings facing the analyzer.
7. Tap "OK" on the dialog box or simply press [Enter] on the keyboard to finish setup.
5-12
Operating Your Analyzer
8. Tap "Start" and the analyzer will automatically start to run in order from the preset
starting position. During this process, the analyzer indicator is flickering in green.
9. After every analysis cycle, the results will be saved to the sample database
10. When the analyses are finished, a statistical result dialog box will display.
11. Tap "Return" to return to the “Sample Analysis” screen, the analyzer indicator will
return to green.
12. When the autoloading analyses are finished, all the tube racks are moved to the
left tray of the autoloader automatically, and then you can remove the racks of
tubes safely.
NOTE
If more racks are needed when running the samples, you should load the rack from
the right of the autoloader while removing the completed rack from the left of the
autoloader in time.
Stop Running
During the analysis process, the "Start" button will turn into "End" button. Tap the "End"
button, the analyzer will stop running after the current sample is analyzed, and the rack of the
5-13
Operating Your Analyzer
sample will move to the unloading tray one the left of the autoloader.
STAT
Use the function to insert STAT samples during the analysis process.
1. Press the [Open] key on the analyzer to open the sample compartment door, and
place the uncapped STAT sample into the compartment. Close the compartment
door.
2. Wait for the analyzer to complete the analysis of current sample and convert to the
STAT mode. You will know that the STAT mode is ready when a “Cancel STAT”
button appears.
3. Tap “Mode” to enter the sample ID. Tap “OK” to close the dialog box.
Cancel STAT
After the running is finished, you can tap the “Cancel STAT” button to cancel the STAT, and
then the analyzer will judge whether to switch to the analysis mode before the inserted STAT
accordingly, and then proceed with autoloading.
Parameter flags
If the parameter is marked by a high/low flag ("H"/"L"), it means the analysis result has
5-14
Operating Your Analyzer
exceeded the upper or lower limit of the reference range, but still within the display range.
If the parameter is marked by the suspect flag ("?"), it means the analysis result is suspicious.
If a result is indicated by “***”, it means the result is invalid or out of the display range.
If a result is indicated by “@”, it means the result is out of the linear range.
Chromatogram flags
Messages Meanings
Analysis not completed Analysis not completed
Abnormal Signal The analyzer signals may become abnormal
The analyzer does not aspirate sample properly; or the
Chromatogram area too small
hemoglobin concentration of the sample may be too low
The hemoglobin concentration of the sample may be too
Chromatogram area too large
high
Abnormal peak number Abnormal variant may be present
SA1c peak not properly
The measured HbA1c% value may be inaccurate
separated
HbA0 peak abnormal The measured HbA1c% value may be inaccurate
Chromatogram abnormal The measured HbA1c% value may be inaccurate
Shorter SA1c retention time The analyzer may be in abnormal status
Longer SA1c retention time The analyzer may be in abnormal status
Shorter HbA0 retention time The analyzer may be in abnormal status
Longer HbA0 retention time The analyzer may be in abnormal status
CAUTION
The following errors may cause misreading of barcodes:
To ensure good readability of the barcode, you must place the label right on the region
marked by X as shown in Figure 5-1, and place the label correctly as shown in Figure 5-2.
5-15
Operating Your Analyzer
NOTE
If several labels are stuck to one tube, or the label is misplaced, peeled or
wrinkled, it may cause autoloading error. To avoid such error, the notes below
shall be followed:
5-16
Operating Your Analyzer
5-17
Operating Your Analyzer
5.8 Standby
When the time for which the analyzer is free from fluidic operations reaches standby waiting
time, a dialog box will display, prompting " Backing up data. Please wait…". When data
backup completes, the analyzer prompts “Entering standby…” and enters the standby
status. The screen blacks out and prompts "Analyzer in standby status. Tap the screen to
exit."
NOTE
If it is time for standby, current operations will pause. After the analyzer exits the
standby status, the operations will be continued.
5-18
Operating Your Analyzer
5-19
Operating Your Analyzer
1. Tap the button, and the following dialog box pops up.
Tap “OK” on the screen, and the following dialog box pops up. In the meanwhile, the
3. analyzer automatically transports the WB tube rack, until its first tube position is just
under the sample probe.
Tap “Add Hemolysis Solution”, and the analyzer will dispense 1mL of Hemolysis
4. Solution into the tube.
When the screens displays the message box “Added 1ml, maximum: 5ml”, tap “Add
5. Hemolysis Solution” again to collect additional Hemolysis Solution, or tap “Completed”
to exit the function. When you tap “Completed”, the analyzer will push out the WB tube
rack, and you may remove the tube safely.
5-20
Operating Your Analyzer
NOTE
The analyzer dispenses 1mL of Hemolysis Solution each time; and one tube can
only collect 5mL of Hemolysis Solution at most.
Use a pipette to aspirate needed volume of Hemolysis Solution from the tube and
add it to the blood samples, controls or calibrators for dilution. Follow the
standard procedure of your laboratory to mix the Hemolysis Solution and sample
properly
For the preparation of pre-dilute samples, please refer to 5.5.2 Prediluted Sample;
For the preparation of controls, please refer to the Instruction for Use of the
controls;
For the preparation of calibrators, please refer to the Instruction for Use of the
calibrators.
5-21
6 Sample Review
6.1 Introduction
After every analysis cycle, the analyzer will save the analysis result automatically including
sample information, parameter results, flag messages as well as the chromatograms. The
analyzer can store up to 50,000 analysis results.
You can review all the sample information, parameter results, flag messages as well as the
chromatograms in table or graph modes.
NOTE
The sample result data must have proper backup in case of data lost caused by
hardware or software error.
6-1
Sample Review
The table displays a list of analyzed samples with basic sample information like sequence no.,
sample ID, analysis mode, analysis date and time, tube position, status and the test panel.
Above the sample result table, you may find the following function buttons: "Print", "Search",
"Comm." and "Export"; below the sample result table is the function button area, where you
may find the following buttons: (from left to right) "Graph ", "Edit", "Validate", "Cancel Val."
and "Delete". The "Validate", "Cancel Val." and "Delete" buttons are only available for
administrators. You can tap the buttons to perform corresponding functions.
6-2
Sample Review
Edit information
NOTE
Make sure you enter the patient information carefully and correctly.
You can edit the information displayed on the "Table Review" screen.
Tap the desired record you want to edit, and the selected record will be highlighted. Tap the
"Edit" button, below dialog box will display. You do not need to enter the patient age, the
validator and operator.
NOTE
You can enter letters, digits and all other characters on the keyboard (including
special characters) for sample ID.
The sample ID must end with a digit; and it cannot only consist of "0".
6-3
Sample Review
NOTE
When you enter the patient birth date, the system will automatically calculate the
patient age using the entered "Birth Date" and current "System Date" and display
the result in the "Age" box. The "Age" box will then be greyed, and will become
editable again when the "Date of Birth" is cleared.
The patient birth date should be no later than current system date.
OK
When you have finished entering the sample information, tap "OK" to save the information
and return to the "Table Review" screen.
Cancel
If you do not want to save the entered sample information, tap "Cancel" to return the "Table
Review" screen without saving the changes.
icon would appear at the “Status” cell box of the record. Tap "Cancel Val." to cancel
the validation.
6-4
Sample Review
Delete (administrator)
Users of the administrator level may delete desired records by tapping the "Delete" button.
Tap the "Delete" button, below dialog box will display:
Select "Selected Records" or "All Records" and tap "OK", below dialog box will display. Tap
"Yes" to perform deletion, or tap "No" to cancel deletion.
NOTE
If you tap the "Delete" button when the system is communicating or printing certain
records, a dialog box will display prompting "System busy. Please try again later!".
Select
You can select one or more sample records to perform certain operations like "Comm." or
"Print".
6-5
Sample Review
2. Enter "Sample ID", “Patient ID”, patient “Name”, "Date", and Sample "No.", and check
the needed “Sample Status” (“Not Vali.”, “Not Pri.” and “Not Com.”) as necessary to
specify the searching conditions.
3. Check "Select sample(s) found" if necessary, then the samples found will be selected.
If no results are found, a dialog box will display prompting "No sample found!". Tap
"OK" to close the dialog box.
On the found results screen, all the buttons are the same as of the "Table Review" screen,
only the original "Search" button is replaced by the "Return" button. Tap the "Return" button to
return to the "Table Review" screen.
6-6
Sample Review
Print
Select (a) sample record(s) and tap "Print", the following dialog box will display.
You can "Print Selected" or "Print All" records. Up to 700 records can be printed at the same
time. Tap "OK" to start print. During the printing process, the print icon flickers. Tap "Cancel"
to cancel the print.
Communicate
You can transmit selected sample results to an eternal computer.
Select one or more samples and tap the "Comm." button, below dialog box will display.
You can transmit selected or all records. Tap the "OK" button to start communication. During
the communication process, the communication icon flickers. Tap "Cancel" to close the dialog
box.
Export
You can export selected sample data to a USB card.
NOTE
The USB port locates at the right side of the analyzer. Please refer to Figure 2-4
Right side of the analyzer for reference.
Do not pull out the USB before the export process completes.
6-7
Sample Review
You can choose to export the selected records or all records. Besides, check “Chromatogram”
for “Export Content” if you want to export the chromatograms. Tap "OK" to start export; tap
"Cancel" to cancel the export. When the export process completes, the system will prompt
whether the operation completes successfully or fails; and a log will created if you select to
export "Sample Data".
6-8
Sample Review
Sample
information
Sample
results
Chromatogram Flags
Peak
information
6-9
Sample Review
Interpretation of chromatogram
The chromatogram shows the peaks of various hemoglobin components in order of the
sequence they are eluted from a sample (from left to right: A1a, A1b, F, LA1c, SA1c (HbA1c),
and A0). The peak identified as SA1c(HbA1c) is shaded. The x-axis shows the retention
times (the elapsed time between the start of the run and the apex of the peaks) for these
components. The unit is in seconds.
The table below the chromatogram displays detailed information for each hemoglobin
component peak.
Name
The name of the hemoglobin component.
Time
Retention time (in seconds), the elapsed time between the start of the run and the apex of the
peaks.
Area
The area of a peak corresponds to the volume of the corresponding component. The unit is
OD.s.
6-10
Sample Review
Total Area
This is the sum area for the peaks of all hemoglobin components. The acceptable range is
30~300. In case of insufficient sample volume or very low hemoglobin concentration, the total
area may drop under 30, and the analyzer will report “Chromatogram area too small”.
Predilute the sample as required, and run the sample using predilute mode.
%
The percentage is the ratio of a peak area against the total area.
6-11
7 Customizing Your Analyzer
Software
7.1 Introduction
The analyzer is a flexible laboratory instrument that can be tailed to your work environment.
You can customize the software options as introduced in this chapter.
When you finishes the setup, tap other menu options or other icon buttons. A dialog box will
display asking if you want to save the new settings. Tap "Yes" to save the settings.
7-1
Customizing Your Analyzer Software
Set time
You can select any of the 6 date formats: YYYY-MM-DD, YYYY/MM/DD, MM-DD-YYYY,
MM/DD/YYYY, DD-MM-YYYY and DD/MM/YYYY. Select the desired date format from the
pull-down list.
7-2
Customizing Your Analyzer Software
7-3
Customizing Your Analyzer Software
1. Tap the "New" button on the "User Management" screen, and the following text box
pops out.
2. Enter the "User ID" (required) and "Name" (if needed); set a "Password" and confirm
it in the "Confirm Password" text box. At last, select the "Access Level" as "General
User" or "Administrator".
3. When you complete the information, tap "OK" to save the new record (if the setting is
valid). A new user account will be added to the user list.
NOTE
You must enter a User ID, otherwise the new user account cannot be saved.
1. Select a user record on the "User Management" screen and tap "Change Password".
Below dialog box pops up.
7-4
Customizing Your Analyzer Software
2. Enter your old password first, then set a new one and confirm it. Tap "OK" to save the
new password (if it is valid).
1. Select the user to be deleted (current user cannot be deleted) on the "User
Management" screen and tap "Delete". The following dialog box pops out.
7-5
Customizing Your Analyzer Software
You can set the reference range and unit for each parameter on the screen, and select to
apply the parameters or not. Only the selected parameters will display on the "Sample
Analysis" and "Sample Review" screens and be available for print.
7-6
Customizing Your Analyzer Software
You can set up shortcut code for departments to facilitate quick information input.
Tap the "Add" button, a blank row will become available for edit. Enter the name of the
department and the desired shortcut code accordingly.
Select the shortcut codes you want to delete in the table and tap "Delete". Below dialog box
will display.
7-7
Customizing Your Analyzer Software
When “Start analysis after startup” is checked, place the tube racks on the autoloader, and
enter the “First Sample ID”; the analyzer will start the analysis automatically when the
initialization process is completed.
Set how to handle a sample if 2-way LIS inquiry for it fails (when 2-way LIS is used)
When using the 2-way LIS, set up whether to “Continue Analysis” or “Skip the Sample” when
the 2-way LIS inquiry for the sample fails.
You can set whether to “Continue Analysis” or “Stop Analysis” when QC results fall out the
acceptable range.
7-8
Customizing Your Analyzer Software
You can either choose to use a “Waste Container” or choose “Direct Discharge”.
NOTE
When you select “Waste Container”, you need to prepare the waste containers by
yourself.
When you select to use “Waste Container”, the analyzer will give warning when the
container is full.
When “Monitor Reagent Exp. Date” is checked, the analyzer will give warnings when the
reagents are expired.
NOTE
When “Monitor Reagent Exp. Date” is not checked, the analyzer will not give
warnings when the reagents are expired. Examine the reagent status regularly and
do not use expired reagents. For more information about reagent management and
replacement, refer to 10.5.2 Replacing Reagents.
7-9
Customizing Your Analyzer Software
1. Select the check box of the desired code system to enable the code system.
2. Select the number of digits used on site in the “Digits” area or select "No length
limit".
For the length limits and check bit of different code system, please see Appendix B.15
Barcode Specifications.
NOTE
The code systems and length limits set on the analyzer shall be those used in your
laboratory. Do not select the code systems that are not used, which may increase
the rate of misreading.
7-10
Customizing Your Analyzer Software
Standby Setup
Tap the "Wait" text box and enter the waiting time before entering the standby status. The
range allowed is [1, 30] minutes.
7-11
Customizing Your Analyzer Software
Print Setup
Enter the title you want to display on the printed reports to the "Report Title" text box.
Select the print template direction as "Landscape" or "Portrait".
Printing Content
Check/uncheck the content to be printed in the "Printing Content" area.
Auto Print
Set the auto print function to:
Off
Auto Print after Analysis
or
Auto Print after Validation
7-12
Customizing Your Analyzer Software
Protocol Setup
Tap the text boxes to enter the correct "IP Address", "Subnet Mask", "Gateway" and "Mac
Address".
Transmission Mode
Select the chromatogram transmission method from the pull down list:
Not to be transmitted
Bitmap
Data
You can select the functions based on your needs by tapping on the check boxes.
2-Way LIS (when it is checked, you can enter the “Overtime” value for 2-Way LIS)
Auto Retransmit (this function can only be activated when "ACK Synchronous
Transmission" is checked. If you enable auto retransmit, the analyzer will continue
to retransmit data until it receives response from ACK.)
Auto Communication (when this function is checked, the sample results will be
transmitted to LIS automatically)
7-13
Customizing Your Analyzer Software
You can adjust certain gain values as necessary. When the value entered is out of allowed
range or any gain is left empty, a dialog box will display prompting "Invalid entry", and the
gains will be restored to the original ones before edit.
7-14
Customizing Your Analyzer Software
The analyzer automatically identifes tube rack types by scanning tube rack information; but
when it fails to identify a rack type, you can also select to specify the rack type yourself. Tap
“User Specify” and select the correct rack type.
7-15
8 Using the QC Program
8.1 Introduction
Quality Control (QC) consists of strategies and procedures that measure the precision and
stability of the analyzer. The results imply the reliability of the sample results.
Analysis of the results with statistical methods allows the inference that sample results are
reliable. Mindray recommends you run the QC program daily with low and high level controls.
A new lot of controls should be analyzed in parallel with the current lot prior to their expiration
dates.
8-1
Using the QC Program
General users may only review the QC file settings. Administrators may select the QC file to
be used by checking the "In Use" check box of it. The QC results will then be saved to the
selected file.
Edit a QC file
Select the QC file you want to edit in the QC file list and tap "Edit" to enter the "QC File Setup"
screen.
8-2
Using the QC Program
NOTE
Refer to the Instruction for Use of the control for its lot no., expiration date,
parameter reference values and limits.
The entered expiration date should be either the expiration date printed on the
labeling or the open-bag expiration date, whichever is earlier. The open-bag
expiration date is calculated as follows: the date that bag is opened + the
open-bag stability days.
8-3
Using the QC Program
NOTE
Only valid parameter targets and limits can be saved.
8.2.2 QC Analysis
When a QC file is selected as "In Use" (only one file may be selected for each level), all the
QC results will be saved to the file.
All the samples, controls, calibrators, reagents, wastes and areas contacted them
are potentially biohazardous. Wear proper personal protective equipment (e.g.
gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in
the laboratory.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the sharp sample probe when working around it.
The sample may spill from the uncapped collection tubes and cause biohazard.
Exercise caution to the uncapped collection tubes.
8-4
Using the QC Program
WARNING
Collection tubes broken may cause personal injury and/or biohazard. Exercise
caution when loading the collection tubes to the rack or getting the collection tubes
from the rack, be sure not to break the tubes.
Do not touch the moving parts. Keep your clothes, hairs and hands away from the
moving parts to avoid injury.
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of water
and if necessary, go see a doctor; if the reagents accidentally spill into your eyes,
wash them off with plenty of water and immediately go see a doctor.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
Gently shake and mix the prepared control to avoid bubbles. Do not use a
mechanical mixer.
NOTE
Use the controls and reagents specified by the manufacturer only. Store and use
the controls and reagents as instructed by their instructions for use.
Prepare a control of CRL-1 level and another of CRL-2 level as instructed by instructions
1. for use of the controls.
NOTE
The analyzer supports auto dispensing of Hemolysis Solution. You can use the
function to dilute and prepare controls. For more information, refer to 5.10
Auto Dispensing Hemolysis Solution.
When the controls are completely dissolved, take 200ul of each level of control into two
2. specified containers (1.5ml centrifugal tube or the micro sample cup).
NOTE
The remaining controls can be stored at 2~8℃ for 14 days. When you
intend to use them, take the controls out from the low temperature
environment and place them at room temperature for 15 minutes.
8-5
Using the QC Program
Mark the open-bag date of every vial of control on the vial. Enter the
parameter target value, lot no. and expiration date of the control to the
analyzer software when a vial of control is newly opened.
When the analyzer is ready to run the QC program, place the tube or micro sample cup
3. into the specified positions of the CRL tube rack (CRL-1 should be place at the 2
th
nd
tube
position, and CRL-2 at the 4 tube position), and put the tube rack onto the autoloader.
NOTE
You must place the controls to the control tube rack to run the QC
program; otherwise the analyzer cannot recognize them as controls.
Enter the "Sample Analysis" screen and tap the "Start" button. The analyzer will
4. automatically recognize and run the controls. The QC results will be saved to the "In
Use" control files of corresponding levels.
QC graph review
Tap "Graph" on the "QC Setup" screen to enter the graph screen of the latest used QC file.
8-6
Using the QC Program
Select the file number from the "File No." pull down list and switch the screen to the QC graph
of the selected QC file.
NOTE
When there are less than 3 QC results, the Mean, SD and CV values on the right
side of the QC graph will not display.
The x-axis of the graph represents the number of QC results, and the y-axis represents
the QC results.
The three figures on the left side of the graph corresponds to the three cross lines on the
graph, which from bottom to up demonstrate the lower limit, target and upper limit of the
parameter results.
x i
Mean = i1
n
SD
X i Mean
2
n 1
SD
CV% 100
Mean
Among which, n indicates QC runs, Xi indicates the number i measurement result for the
parameter.
8-7
Using the QC Program
The QC results are displayed in order of their saved time: from left to right, the earliest to the
latest. Tap the arrow buttons on the lower part of the graph to review the results.
Exit
QC table review
Tap "Table" on the "QC Setup" screen to enter the table screen of the latest used QC file.
Select the file number from the "File No." pull down list and switch the screen to the QC table
of the selected QC file.
You can perform the following operations on the QC Table screen:
You can tap the arrow buttons on the right side to review the QC results. The QC results are
displayed in order of save times. The latest is displayed at the top row.
Communicate
8-8
Using the QC Program
Export
Tap the "Export" button to export the QC information in current QC table to a USB connected
to the analyzer.
NOTE
The USB port locates at the right side of the analyzer. Please refer to Figure 2-4
Right side of the analyzer for reference.
Do not pull out the USB before the export process completes.
Delete (administrator)
Tap "Selected Records" or "All Records", and then tap "OK" to delete. The operation will be
logged in the system log. Tap "Cancel" to cancel the delete.
Exit
When QC fails
When any QC result falls out the controlled range, the QC button on the top of the screen
lights in red:
8-9
Using the QC Program
NOTE
If you have selected “Stop Analysis” when QC fails on “Setup”-“Auxillary” screen,
the analyzer will not analyze following samples. For more information, refer to 7.6
Auxillary (“Setup”> “Auxillary”).
If any QC point falls out of the controlled range, follow below steps to solve the problem. If the
problem cannot be solved, contact Mindray Customer Service Department.
1. Check the error information area on the screen. If any error exists, see Chapter 11
Troubleshooting for solutions.
2. Check the settings of QC file and correct any wrong information that may exist.
3. Run the QC analysis again.
4. Run the QC analysis with a new vial of control.
5. When the QC results are still out of range, re-calibrate the analyzer.
8-10
9 Calibrate Your Analyzer
9.1 Introduction
Calibration is a procedure to standardize the analyzer by determining its deviation under
certain specified conditions. In order to get accurate sample analysis results, you should
calibrate the analyzer per the procedure below when necessary.
There are two calibration programs available on this analyzer: manual calibration and auto
calibration using calibrators.
NOTE
Calibration procedures can only be performed by users of the administrator-level.
Use the calibrators and reagents specified by the manufacturer only. Store and
use the calibrators and reagents as instructed by their instructions for use.
9-1
Calibrate Your Analyzer
you are going to use this analyzer for the first time (usually done by a
Mindray-authorized representative when installing the analyzer).
a major component (for example, the analytical column) has been changed.
NOTE
All of the measured parameters must be calibrated before readings of this analyzer
can be used as valid analysis results.
9-2
Calibrate Your Analyzer
It is recommended that you create a log table for your analyzer. This log table should contain
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are: calibration date, supplier of calibrator, lot number,
expected results and limits.
NOTE
Be sure to use the evacuated collection tubes recommended in the Appendix.
9-3
Calibrate Your Analyzer
Select the suitable reference method (“NGSP”, “IFCC” or “Mono-S”) in accordance with your
laboratory procedure from the pull-down list.
Enter values for “a” and “b” as needed.
In the formula Y=aX+b, Y means the results finally shown on the screen, Y means the raw
result measured by the analyzer.
Tap “Factory Setup” and set the calibration factors back to factory setup.
Lot No.
Enter the lot No. of both the “CAL-1” and “CAL-2” level calibrators into the "Lot No." boxes.
9-4
Calibrate Your Analyzer
Expiration Date
Runs
Select a number from the "Runs" pull down list. The analyzer will run the calibrators for the
number of times you selected. For example, if you select "2" runs, then the calibrators of both
levels will be run twice each.
Reference Method
Select the suitable reference method (“NGSP” or “IFCC”) in accordance with your laboratory
procedure from the pull-down list.
Parameter Target
Enter the targets of the CAL-1 and CAL-2 calibrators into the corresponding "Target" boxes
NOTE
See the instruction for use and the target sheet of the calibrators for the lot No.,
expiration dates and the targets.
The entered expiration date should be either the expiration date printed on the
labeling or the open-bag expiration date, whichever is earlier. The open-bag
expiration date is calculated as follows: the date that bag is opened + the open-
bag stability days.
If you start the calibration analysis before you complete the calibrator information,
a dialog box will display prompting "Invalid entry!". Tap "OK" and enter complete
calibrator information to start calibration analysis.
Samples, controls, calibrators and waste are potentially infectious. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them and the contacted areas in the
laboratory.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
9-5
Calibrate Your Analyzer
NOTE
Only Mindray-specified calibrators shall be used. Mindray will not be responsible
for any erroneous result caused by using other calibrators.
Refer to the instructions for use of the calibrators for their use and storage.
Use clean 1.5mL centrigual tube or the micro sample cup to run the calibration.
Use the CAL tube rack and corresponding tube or sample cup adapters.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
NOTE
The analyzer supports auto dispensing of Hemolysis Solution. You can
use the function to dilute and prepare controls. For more information, refer
to 5.10
2. Place the well-mixed calibrators into the specified calibrator tube rack (CAL-1 level
calibrator should be placed at the 2
nd
tube position, and CAL-2 level calibrator at the
th
4 tube position), and put the tube rack loaded with the calibrators onto the
autoloader.
3. Tap “Calibration”-“Calibrator” screen, and make sure the calibration information and
parameter targets are correct.
4. Tap “Start”, and the analyzer will run the both levels of calibrators for the “Runs” you
specified.
5. After the analysis is done, the calibration results will display on the screen.
NOTE
If the ambient temperature is out of the specified operating range, the analyzer will
9-6
Calibrate Your Analyzer
alarm you for abnormal ambient temperature and the analysis results may be
unreliable. When temperature errors are reported in the error information area
after analysis, see Chapter 11 Troubleshooting Your Analyzer for solutions.
if any of the results is invalid, a dialog box will display prompting the results are invalid.
if the results are valid, the analyzer will display the new calibration factors automatically.
when you switch to another screen, a dialog box will display asking if you want to save
the calibration results. Tap “OK” to save the calibration results, or tap “Cancel” to give
up.
When calibration has been performed for specified times, the analyzer will calculate and
display the CV value and calibration factors (a and b) automatically.
The calculated calibration factors calculated under different reference systems shall be within
the ranges as below:
When the calculated factors are out of above ranges, a dialog box will display prompting the
results are invalid. Check if there is any error. If you cannot solve the problem, contact
Mindray Customer Service Department.
Other functions
Print
9-7
Calibrate Your Analyzer
From the menu, tap “Calibration”-“Touch Screen” to enter the touch screen calibration screen.
CAUTION
Do not use anything sharp on the touch screen or strike on it.
If the touch screen does not correctly respond to the positions you touched, perform the
procedure to calibrate the touch screen.
Do as follows:
Tap the black plus sign at the upper left corner of the screen as instructed by the screen
2. display to start the calibration.
Then tap the black plus successively displayed at the lower left corner, lower right
3. corner, upper right corner and in the centre of the screen.
When the calibration is finished, the screen displays “Calibration succeeded.” Tap any
4. other menu option to exit touch screen calibration.
9-8
10 Servicing Your Analyzer
10.1 Introduction
To ensure the good operating condition of the system, operators shall follow this chapter to
maintain and troubleshoot the system.
10-1
Servicing Your Analyzer
10.2 Maintenance
10.2.1 Consumables ("Maintenance">"Consumables")
Reagents
When a reagent runs out or reaches expiration date, tap “Maintenance” > “Consumables” >
“Reagents” or tap the icon on the menu bar directly to enter the reagent
management screen.
Tap the “Replace” button for the reagent to be replaced. The following dialog box pops up:
10-2
Servicing Your Analyzer
Follow the instructions on the screen to install a new bag of reagent (for more details, please
refer to 10.5.2 Replacing Reagents), and tap “OK”. The reagent information will be
automatically loaded and displayed.
Column (Administrators)
Users at the administor level may tap “Maintenance” > “Consumables” > “Column” to enter
the Column Information screen and check relevant information of the analytical column.
When the analytical column has reached its expiration date, you will be prompted to install a
new one. Tap “Replace” to install a new column.
For detailed instructions, refer to 10.5.4 Replacing Analytical Column in this manual.
"Fluidics")
Pack-up
If the analyzer is not to be used for over 2 weeks, you should perform this procedure.
Do as follows to pack up:
1. Tap the “Packup” button, and then a message box will display:
10-3
Servicing Your Analyzer
2. Remove the analytical column from the column box (refer to 10.5.4 Replacing
Analytical Column for instructions), and install a 2-way fitting instead. Fasten the two
screws on both sides of the fitting, then tap “OK”. A dialog box will display:
NOTE
The 2-way fitting was given to you when the analyzer was first installed by
our service people. Please restore it properly.
2-way fitting
3. Place all reagent bag cap assemblies into a clean and empty container; then tap
“OK”. The analyzer will start to empty fluidics:
10-4
Servicing Your Analyzer
5. Place all reagent bag cap assemblies into the pure water, and then tap "OK" to start
cleaning.
7. Take out the cap assemblies and place them into a cleam and empty container again
as instructed, and then tap "OK" to drain the fluidics.
8. After the emptying is complete, a message box will display. Tap “OK” and turn off the
power switch according to the prompt displayed on the screen.
10-5
Servicing Your Analyzer
The log records the key operations performed on the analyzer. It provides the operators an
access to review the operating history and service personnel the facilitation of
troubleshooting.
The analyzer can save logs of the recent two years. If number of logs exceeds the upper limit,
the latest log will overwrite the oldest one. You can browse and print logs, but cannot delete
them.
10-6
Servicing Your Analyzer
1.
To export the logs of a selected time range, enter the starting and ending number in the
When the "Export succeeded" message box pops up, tap "OK" to close it and return to
4. the log screen.
10-7
Servicing Your Analyzer
10.4 Cleaning
10.4.1 When, Why and Tools Needed
Procedure When to replace Why to Clean Tools needed
Cover cleaning As needed Remove contaminators on Disinfectant (see 10.4.2
the cover Cover for types of
disinfectant)
Mindray does not claim the validity of the listed chemicals in infection control. For effective
control of infection, please consult the Infection Prevention Department of the hospital or the
epidemic professionals.
The cleaning may damage the system to some extent. It is recommended to perform the
cleaning only when necessary according to your laboratory protocol.
10-8
Servicing Your Analyzer
10.5 Replacing
Samples and waste are potentially infectious. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when
handling them and the contacted areas in the laboratory.
WARNING
Do not touch the moving parts. Keep your clothes, hairs and hands away from the
moving parts to avoid injury.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc. according to
government regulations.
If the reagents accidentally spill on your skin, wash them off with plenty of water
and if necessary, go see a doctor; if the reagents accidentally spill into your eyes,
wash them off with plenty of water and immediately go see a doctor.
10-9
Servicing Your Analyzer
NOTE
After replacing the reagent bag, check the tubing connected to the cap assembly
and make sure it is not bent over.
The reagent shelf is separated into 3 compartments for different reagents. There are marks
on the right side of the main unit to indicate the connection of different reagents.
2. Open the reagent replacing window on the left side of the analyzer.
3. Turn the cap of the old bag counterclockwise, and then take out the cap
assembly with caution. Remove the old reagent bag.
4. Place the new reagent bag on the corresponding compartment, the side with the
RFID card facing the front.
5. Insert the pickup tube of the cap assembly into the new bag, and then turn the
cap clockwise until it is secured.
6. Cap the old bag with the cap of the new one, and then dispose of the waste
properly.
7. When the reagent bag is installed, the analyzer will automatically scan the
reagent bag and enter the reagent information to the system.
10-10
Servicing Your Analyzer
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc. according to
government regulations.
Remove the waste container cap and replace the waste container only when the
power indicator is not flickering, in order not to make the waste overflow from the
container.
1. Get an empty waste container, remove the cap and place it next to the one to be
replaced.
2. Turn the cap counterclockwise and remove the cap assembly from the old container
with caution.
3. Insert the old cap assembly into the new container as vertically as possible, and secure
the cap by turning it clockwise.
4. Cap the old container with the cap of the new one, and then dispose of the waste
properly.
NOTE
When installing the analyzer, make sure the direction of the arrow on the column
is the same as that on the column box.
10-11
Servicing Your Analyzer
2) Tap the "Replace" button on the screen; the analyzer prepares for
replacement.
When the analyzer is ready, a dialog box will display prompting about how to
2. remove the old column:
10-12
Servicing Your Analyzer
2) Lift the latch on the inner side on the top of the box, the locking ring on the
outer side will be released, and the column box will be opened.
3) Pull the locking ring to its back side, and place it as well as the latch on the
top of the box.
4) Place a piece of tissue under the column.
3. You can see the column inside the box. Take it out from its slot.
10-13
Servicing Your Analyzer
4. 1) Loosen the fittings and the tubings on both inlet and outlet sides of the
column successively, and detach the column.
2) Push the tubes in both the inlet and outlet fittings, until they protrude at
least 3mm out of the fittings.
10-14
Servicing Your Analyzer
NOTE
When you remove the column, there may be a small amount of
liquid dropping down from the fixing position on the column. Put a
piece of tissue under the column to absorb the liquid.
3) After you remove the old column, tap “OK” on the screen, and a dialog box
will display prompting about how to connect the inlet end of the column.
10-15
Servicing Your Analyzer
2) Screw the inlet fitting loosely into the column inlet for about 3 rounds. Do
not tighten it tightly.
3) Use a piece of tissue to wrap the connecting position of the tube fitting and
column inlet; then tap “OK”.
10-16
Servicing Your Analyzer
5) When there is liquid flowing out the the column inlet, tap "Stop", and screw
the fitting tightly into the column inlet.
6) Remove the tissue wrapping the inlet, and wipe the liquid from the
connection.
NOTE
If there is no liquid flowing from the connection, loosen the fitting a little,
and repeat steps 4) to 6).
7) Tap “OK”, and a dialog box will display prompting about how to install the
outlet side of the column.
10-17
Servicing Your Analyzer
1) Screw the other fitting tightly into the outlet of the column.
6.
10-18
Servicing Your Analyzer
NOTE
The ring form on the column should be aligned with the groove in
the center of the slot.
1) Tap the "Prime" button on the screen; and wait for 1 minute to tap the
7. "Stop" button (a timer will display on the screen);
2) Make sure there is no liquid leaking from either the inlet or outlet of the
column.
10-19
Servicing Your Analyzer
NOTE
When you see liquid leakage, tap the “Stop” button, and make sure
the two fittings on the outlet and inlet of the column are tightly
fastened. Then repeat above steps.
8. 1)
2)
Remove the tissue under the column, and wipe any liquid away.
4) Tap “OK” on the screen, and the “Load Column Information” screen will
display.
9. Take out the new RFID card of the new column from its package box, and
present the card to the card reader located near the analyzer autoloader. The
analyzer will automatically scan the RFID card and load the information.
10-20
Servicing Your Analyzer
10. Make sure the column information displayed on the screen is correct. Tap “OK”,
and the analyzer starts to prime the column tubings. A progress bar will show.
11. After priming, a dialog box will display prompting “Column replacement
succeeded”. Make sure to close the analyzer front cover and re-calibrate the
analyzer. Tap “OK” to return to the “Load Column Information” screen.
NOTE
When there is any error during the column replacement, tap “Cancel” to end the
replacement procedure. Restore the tubes and the column to their original status, and
close the analyzer front cover; remove the error and try again. When you cannot solve
the problem, contact Mindray Customer Service Department.
10-21
Servicing Your Analyzer
NOTE
The filter Wear gloves when you replace the filter.
1. Preparation
1) Tap Menu -"Maintenance" > “Consumables” > “Filter” to enter the filter
reaplcement screen.
2) Tap the "Replace" button on the screen; the analyzer prepares for
replacement.
10-22
Servicing Your Analyzer
3) Remove the lower part of the filter by rotating it clockwisely; then remove the
10-23
Servicing Your Analyzer
filter element.
4) Push the tube in the fitting, until it protrude at least 3mm out of the fitting.
After you remove the old filter and tap “OK” on the screen. The following screen
displays.
10-24
Servicing Your Analyzer
1) Place a new filter element into the lower part of the filter.
2) Tighlty screw up the lower part of the filter to the upper part of the filter by
roating the lower part counter-clockwisely;then tightly screw up the tube fitting
to the lower part of the filter.
10-25
Servicing Your Analyzer
4. Leakage self-test
When you have installed a new filter element and tapped “OK” on the screen, the
following screen displays.
NOTE
If you observe liquid leakage, tap “Stop” on the screen. Maker sure the fitting
is tightly screwed , and repeat the step.
5. Complete replacement
10-26
Servicing Your Analyzer
1) Wipe off the leaked liquid with a tissue, and close the two front covers of the
analyzer.
10-27
Servicing Your Analyzer
Temp.&Pressure
Version Info.
You cannot modify but only browse the status information at the Status Screen. To check the
status information at the Status Screen is significant for you to locate and remove errors of
the analyzer.
"Temp.&Pressure")
The screen displays the current value of each item and the corresponding normal range, and
if the out-of-range results are highlighted by the red background, which is significant for you
to judge and locate errors of the analyzer.
10-28
Servicing Your Analyzer
You can check the current version information of the analyzer software and hardware.
10-29
11 Troubleshooting
11.1 Introduction
This chapter contains information that is helpful in locating and correcting problems that may
occur during operation of your analyzer.
NOTE
This chapter is not a complete service manual and is limited to problems that are
readily diagnosed and/or corrected by the user of the analyzer.
11-1
Troubleshooting
11-5
Troubleshooting
11-6
Annex I Operating with Extended
Program
I-1 Applicable Area
This annex provides the instruction for operating with the Extended Program of the
Automated Glycohemoglobin Analyzer (hereafter the "analyzer"). The Extended program is
an advanced and optional feature of the H50P analyzer.
For the operations with Standard Program, the setup and maintenance of the analyzer, as
well as other information, check the main part of the Operator's Manual.
For the configuration of the Extended Program, contact your sales representative.
1
Operating with Extended Program
2
Operating with Extended Program
3
Operating with Extended Program
1. Tap the menu icon -"Setup"-"Analysis Program" to enter the "Analysis Program"
screen.
From the "Analysis Program" pull-down list, select "Extended (E)" program.
2.
NOTE
If you want to go back to the standard program, select "Standard (S)"
on this screen.
4
Operating with Extended Program
Tap another icon on the menu bar. A dialog box displays asking if you want to save
3. the change.
NOTE
To ensure the accuracy and reliability of sample analysis results, every time you
have switched the analysis program, be sure to perform the QC procedure with
both CRL-1 and CRL-2.
5
Operating with Extended Program
Initial Checks
Power On
Sample Preparation
See I-4.5 Sample Preparation
Sample Analysis
Closed Tube Mode Autoloading Mode
See I-4.6 Sample Analysis Under See I-4.7 Sample Analysis Under
Closed Tube Mode Autoloading Mode
Shutdown
See I-4.10 Shut Down the
Analyzer
6
Operating with Extended Program
All the samples, reagents, calibrators, controls, wastes and areas contacted by
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc. according to
government regulations.
If the reagents accidentally spill on your skin, wash them off with plenty of water
and if necessary, go see a doctor; if the reagents accidentally spill into your
eyes, wash them off with plenty of water and immediately go see a doctor.
Do not touch the moving parts. Keep your clothes, hairs and hands away from
the moving parts to avoid injury.
Check whether there are enough reagents for the tests of the day. If not, replace the
reagents.
If the reagents run out during analysis, the analyzer will pause working automatically and
prompt the operator to replace the reagents. Please follow the instruction to replace the
reagents.
Check and make sure the reagent, waste and pneumatic unit tubes are properly connected
and not bent.
Check and make sure the power cord of the analyzer is properly plugged into the power
outlet.
7
Operating with Extended Program
I-4.3 Startup
I-4.3.1 Power Switch
Press the [Start] button on the analyzer.
NOTE
You should remove all the errors before proceeding with any operation.
When the analyzer has been idle for more than 72 hours, the analyzer self-test
will take longer time.
8
Operating with Extended Program
Make sure the Extended program is selected. When the "Extended" program is selected, the
"Sample Analysis" icon on the menu bar will display an "E".
Check the status area at the lower part of the screen, and make sure all the reagents are
sufficient; and the analytical column does not reaches their maximum uses.
NOTE
If you are using the Extended Program, be sure to perform the quality control
under the same program.
I-4.4.1 Introduction
Quality Control (QC) consists of strategies and procedures that measure the precision and
stability of the analyzer. The results imply the reliability of the sample results.
Analysis of the results with statistical methods allows the inference that sample results are
reliable. Mindray recommends you to run the QC program daily with both CRL-1 and CRL-2
levels of controls. A new lot of controls should be analyzed in parallel with the current lot prior
to their expiration dates.
This may be accomplished by running the new lot of controls twice a day for five days using
any empty QC file. The QC files calculate the mean, standard deviation and coefficient of
variation for each selected parameter.
9
Operating with Extended Program
General users may only review the QC file settings. Administrators may select the QC file to
be used by checking the "In Use" check box of it. The QC results will then be saved to the
selected file.
Edit a QC file
Select the QC file you want to edit in the QC file list and tap "Edit" to enter the "QC File Setup"
screen.
10
Operating with Extended Program
Enter the lot number into the "Lot No." text box.
Select "CRL-1" or " CRL-2" control level from the "Level" pull down list.
You can enter the expiration date of the control to the "Exp. Date" text box in the format you
set in 7.2 Date/Time.
NOTE
Refer to the Instruction for Use of the control for its lot no., expiration date,
parameter reference values and limits.
The entered expiration date should be either the expiration date printed on the
labeling or the open-container expiration date, whichever is earlier. The
open-container expiration date is calculated as follows: the date that container is
opened + the open-container stability days.
Enter the target and limit into the "Target" and "Limit" text boxes.
Select suitable reference method (NGSP or IFCC) based on the trace system used in your
laboratory.
11
Operating with Extended Program
Select "Absolute Value" or "Percentage" from the "Limit Format" pull down list.
NOTE
Only valid parameter targets and limits can be saved.
When you complete the settings, tap "OK" to save the settings and return to the "QC Setup"
screen.
QC analysis
When a QC file is selected as "In Use" (only one file may be selected for each level), all the
QC results will be saved to the file.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the sharp sample probe when working
around it.
The sample may spill from the uncapped collection tubes and cause biohazard.
Exercise caution to the uncapped collection tubes.
12
Operating with Extended Program
WARNING
Collection tubes broken may cause personal injury and/or biohazard. Exercise
caution when loading the collection tubes to the rack or getting the collection
tubes from the rack, be sure not to break the tubes.
Do not touch the moving parts. Keep your clothes, hairs and hands away from
the moving parts to avoid injury.
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
Gently shake and mix the prepared control to avoid bubbles. Do not use a
mechanical mixer.
NOTE
Use the controls and reagents specified by the manufacturer only. Store and use
the controls and reagents as instructed by their instructions for use.
Prepare a control of CRL-1 level and another of CRL-2 level as instructed by instructions
1. for use of the controls.
NOTE
The H50P analyzer supports auto dispensing of Hemolysis Solution. You
can use the function to dilute and prepare controls. For more information,
refer to 5.10 Auto Dispensing Hemolysis Solution.
When the controls are completely dissolved, take 200ul of each level of control into two
2. specified containers (1.5ml centrifugal tube or the micro sample cup (included)).
NOTE
The remaining controls can be stored at 2~8℃ for 14 days. When you
intend to use them, take the controls out from the low temperature
13
Operating with Extended Program
Mark the open-container date of every vial of control on the vial. Enter the
parameter target value, batch no. and expiration date of the control to the
analyzer software when a vial of control is newly opened.
When the analyzer is ready to run the QC program, place the tube or micro sample cup
3. into the specified positions of the CRL tube rack (CRL-1 should be place at the 2nd tube
position, and CRL-2 at the 4th tube position), and put the tube rack onto the autoloader.
NOTE
You must place the controls to the control tube rack to run the QC
program; otherwise the analyzer cannot recognize them as controls.
Enter the "Sample Analysis" screen and tap the "Start" button. The analyzer will
4. automatically recognize and run the controls. The QC results will be saved to the "In
Use" control files of corresponding levels.
Reviewing QC Results
QC Graph
QC Table
QC graph review
Tap "Graph" on the "QC Setup" screen to enter the graph screen of the latest used QC file.
14
Operating with Extended Program
Select the file number from the "File No." pull down list and switch the screen to the QC graph
of the selected QC file.
NOTE
When there are less than 3 QC results, the Mean, SD and CV values on the right
side of the QC graph will not display.
The x-axis of the graph represents the number of QC results, and the y-axis represents
the QC results.
The three figures on the left side of the graph corresponds to the three cross lines on the
graph, which from bottom to up demonstrate the lower limit, target and upper limit of the
parameter results.
x i
Mean = i1
n
SD
X i Mean
2
n 1
SD
CV% 100
Mean
Among which, n indicates QC runs, Xi indicates the number i measurement result for the
parameter.
The QC results are displayed in order of save times: from left to right, the earliest to the latest.
Tap the "Next Page" button on the lower part of the graph to review the results.
Exit
QC table review
Tap "Table" on the "QC Setup" screen to enter the table screen of the latest used QC file.
16
Operating with Extended Program
Select the file number from the "File No." pull down list and switch the screen to the QC table
of the selected QC file.
You can perform the following operations on the QC Table screen:
You can tap the "Next Page" button on the right side to review the QC results. The QC results
are displayed in order of save times. The latest is displayed at the top row.
Communicate
Export
Tap the "Export" button to export the QC information in current QC table to a USB connected
to the analyzer.
17
Operating with Extended Program
NOTE
The USB port locates at the right side of the analyzer. Refer to Figure 2-4 Right
side of the analyzer for reference.
Do not pull out the USB before the export process completes.
Delete (administrators)
Tap "Selected Records" or "All Records", and then tap "OK" to delete. The operation will be
logged in the system log. Tap "Cancel" to cancel the delete.
Exit
When QC fails
When any QC result falls out the controlled range, the QC button on the top of the screen
lights in red:
18
Operating with Extended Program
NOTE
If you have selected “Stop Analysis” when QC fails on “Setup”-“Axillary” screen,
the analyzer will not analyze following samples. For more information, refer to
7.6 Auxillary (“Setup”> “Auxillary”)
If any QC point falls out of the controlled range, follow below steps to solve the problem. If the
problem cannot be solved, contact Mindray Customer Service Department.
1. Check the error information area on the screen. If any error exists, see Chapter 11
Troubleshooting for solutions.
2. Check the settings of QC setup and correct any wrong information that may exist.
3. Run the QC analysis again.
4. When the QC results are still out of range, run the QC analysis with a new vial of control.
5. When the QC results are still out of range, re-calibrate the analyzer.
19
Operating with Extended Program
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
WARNING
Do not contact the patients' sample blood directly.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
NOTE
Be sure to use clean anticoagulant collection tubes (recommended
anticoagulants: K2EDTA, K3EDTA, heparin lithium salt, sodium citrate,
potassium oxalate monohydrate/sodium fluoride), fused silica glass/plastic test
tubes, centrifugal tubes and borosilicate glass capillary tubes.
CAUTION
To ensure accurate analysis results, the sample volume should be no less than
1mL under whole blood mode.
The samples stored under refrigeration condition (2℃-8℃) must be kept at room
temperature for at least 30 minutes before analysis.
When a sample has been kept for a while after preparation, be sure the mix it
again before running it
1. Add certain amount of sample blood (10ul) into a 1.5ml centrifugal tube (or micro
sample cup) .
If the small area flag is reported because of sample volume is insufficient,, use a
2. pipette to dispense lyse into the sample at the ratio of1:100 (lyse to blood sample)
and well mix it;
If the small area flag is reported because the HGB of the whole blood sample is
too low, use a pipette to dispense lyse into the sample at the ratio of1:50 (lyse to
blood sample) and well mix it.
The H50P analyzer supports auto dispensing of Hemolysis Solution. You can use the function
to dilute and prepare controls. For more information, refer to 5.10 Auto Dispensing Hemolysis
Solution.
CAUTION
Take methods to prevent the diluent from dust; otherwise the results may be
unreliable.
Be sure to run the prediluted samples within 30 minutes after the dilution;
otherwise the results may be unreliable.
Be sure to mix any sample that has been prepared for a while before running it.
To ensure accurate analysis results, the sample volume should be no less than
0.5mL under prediluted mode.
21
Operating with Extended Program
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
Repeated piercing the cap of the evacuated blood collection tube may damage
the cap, and the scraps produced may result in inaccurate analysis results. It is
recommended that each tube is pierced for no more than 3 times.
NOTE
Proper reference range shall be entered on the "Setup"- “Parameter” screen
before analysis. Otherwise, the results may be flagged erroneously.
When it is ready to run a sample (i.e. the analyzer indicator stays in green), shake the
1. tube of whole blood sample as instructed by the picture below to mix the sample
thoroughly.
22
Operating with Extended Program
Press the [Open] key on the analyzer to open the sample compartment door. Install
2. theФ13×75 (mm) adapter into the sample compartment. Place the sample into the
adapter.
Tap the "Mode" button on the "Sample Analysis" screen, and the "Mode Setup"
4. screen will display.
Check the "CT" radio button for "Presentation Mode" and enter the "Start Sample ID".
5. If necessary, check "Auto Increase", and the subsequent sample IDs will
automatically increase by 1 based on the previous one.
6. Tap "OK" on the dialog box or simply press [Enter] on the keyboard to finish setup.
Tap the "Start" button to start the analysis. After the sample probe finishes sample
7. aspiration, the sample compartment door opens automatically, and the analyzer
indicator flickers in green.
When the analysis completes, the analyzer indicator returns to "Ready" status (stay in
8. green).
23
Operating with Extended Program
24
Operating with Extended Program
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling
them in the laboratory.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
When a sample has been kept for a while after preparation, be sure the mix it
again before running it.
Repeated piercing the cap of the evacuated blood collection tube may damage
the cap, and the scraps produced may result in inaccurate analysis results. It is
recommended that each tube is pierced for no more than 3 times.
Be sure that the entered sample ID, rack No., tube No. and the analysis mode
are strictly in accordance with the sample to be run.
NOTE
Proper reference range shall be entered on the "Setup"-“Parameter” screen
before analysis. Otherwise, the results may be flagged erroneously.
1. When it is ready to run a sample (i.e. the analyzer indicator stays in green), shake
the tube of whole blood sample as instructed by the picture below to mix the
sample thoroughly.
25
Operating with Extended Program
2. Tap the "Mode" button on the "Sample Analysis" screen, and the "Mode Setup"
dialog box will display.
4. Enter the "Start Sample ID" (skip the step if you are using the whole blood mode
and have selected "Auto Scan ").
5. To run a whole blood sample, directly place the prepared sample tube into the
whole blood sample rack.
To run a prediluted sample, fit the prepared sample tube (or micro sample cup)
uncapped into a corresponding adapter (clip the tube cap into the adapter notch)
and place the tube together with the adapter into an even-numbered position of the
PD tube rack.
6. Place the racks with tubes on the right tray of the autoloader, with the side opening
facing the analyzer.
7. Tap "OK" on the dialog box or simply press [Enter] on the keyboard to finish setup.
8. Tap "Start" and the analyzer will automatically start to run in order from the preset
starting position. During this process, the analyzer indicator is flickering in green.
9. After every analysis cycle, the results will be saved to the sample database
10. When the analyses are finished, a statistical result dialog box will display.
26
Operating with Extended Program
11. Tap "Return" to return to the “Sample Analysis” screen, the analyzer indicator will
return to green.
12. When the autoloading analyses are finished, all the tube racks are moved to the
left tray of the autoloader automatically, and then you can remove the racks of
tubes safely.
NOTE
If more racks are needed when running the samples, you should load the rack
from the right of the autoloader while removing the completed rack from the left
of the autoloader in time.
Stop Running
During the analysis process, the "Start" button will turn into "End" button. Tap the "End"
button, the analyzer will stop running after the current sample is analyzed, and the rack of the
sample will move to the unloading tray one the left of the autoloader.
STAT
Use the function to insert STAT samples during the analysis process.
27
Operating with Extended Program
1. Press the [Open] key on the analyzer to open the sample compartment door, and
place the uncapped STAT sample into the compartment. Close the compartment
door.
2. Wait for the analyzer to complete the analysis of current sample and convert to the
STAT mode. You will know that the STAT mode is ready when a “Cancel STAT”
button appears.
3. Tap “Mode” to enter the sample ID. Tap “OK” to close the dialog box.
5. When the compartment door opens, it means the analyzer aspirates enough
sample. You can remove the sample then; but it will still take a little time for the
analyzer finishes the analysis.
NOTE
The STAT mode only supports whole blood samples.
Cancel STAT
After the running is finished, you can tap the “Cancel STAT” button to cancel the STAT, and
then the analyzer will judge whether to switch to the analysis mode before the inserted STAT
accordingly, and then proceed with autoloading.
CAUTION
The following errors may cause misreading of barcodes:
28
Operating with Extended Program
To ensure good readability of the barcode, you must place the label right on the region
marked by X as shown in Figure 5-1, and place the label correctly as shown in Figure 5-2.
NOTE
If several labels are stuck to one tube, or the label is misplaced, peeled or
wrinkled, it may cause autoloading error. To avoid such error, the notes below
shall be followed:
29
Operating with Extended Program
If the parameter is marked by the suspect flag ("?"), it means the analysis result is suspicious.
If a result is indicated by “***”, it means the result is invalid or out of the display range.
If a result is indicated by “@”, it means the result is out of the linear range.
30
Operating with Extended Program
Messages Meanings
The analyzer does not aspirate sample properly; or
Chromatogram area too
the hemoglobin concentration of the sample may be
small
too low
Chromatogram area too The hemoglobin concentration of the sample may
large be too high
Abnormal peak number Abnormal variant may be present
SA1c peak not properly
The measured HbA1c% value may be inaccurate
separated
HbA0 peak abnormal The measured HbA1c% value may be inaccurate
Chromatography
The measured HbA1c% value may be inaccurate
abnormal
Shorter SA1c retention
The analyzer may be in abnormal status
time
Longer SA1c retention
The analyzer may be in abnormal status
time
Shorter HbA0 retention
The analyzer may be in abnormal status
time
Longer HbA0 retention
The analyzer may be in abnormal status
time
Suspected HbE HbE variant may be present
I-4.9.1 Introduction
After every analysis cycle, the analyzer will save the analysis result automatically including
sample information, parameter results, flag messages as well as the chromatographies. The
analyzer can store up to 50,000 analysis results.
You can review all the sample information, parameter results, flag messages as well as the
chromatographies either in table or graph mode.
NOTE
The sample result data must have proper backup in case of data lost caused by
hardware or software error.
31
Operating with Extended Program
Table Review
Tap the "Table Review" button on the menu to enter the "Table Review" screen.
The table displays a list of analyzed samples with basic sample information like sequence no.,
sample ID, analysis mode, analysis date and time, tube position, status and the test panel.
Above the sample result table, you may find the following function buttons: "Print", "Search",
"Comm." and "Export"; below the sample result table is the function button area, where you
may find the following buttons: (from left to right) "Graph ", "Edit", "Validate", "Cancel Val."
and "Delete". The "Validate", "Cancel Val." and "Delete" buttons are only available for
administrators. You can tap the buttons to perform corresponding functions. The "Validate",
"Cancel Val." and "Delete" buttons are only available for administrators. You can tap the
buttons to perform corresponding functions.
Tap the "Graph Review" button to switch to the "Graph Review" screen. For detailed
information about the "Graph Review" screen, see 6.2.2 Graph Review.
32
Operating with Extended Program
Edit information
NOTE
Make sure you enter the patient information carefully and correctly.
You can edit the information displayed on the "Table Review" screen.
Tap the desired record you want to edit, and the selected record will be highlighted. Tap the
"Edit" button, below dialog box will display. You do not need to enter the patient age, the
validator and operator.
NOTE
You can enter letters, digits and all other characters on the keyboard (including
special characters) for sample ID.
The sample ID must end with a digit; and it cannot only consist of "0".
33
Operating with Extended Program
You can either enter the department name in the "Dept." box, or select the desired
department from the "Dept." pull-down list (if there are previous entries saved in the list). At
most 30 entries can be automatically saved to the "Dept." pull-down list. At most 30 entries
can be automatically saved to the "Clinician" pull-down list.
Select the desired item (Male, Female, or null) from the "Gender" pull-down list. The default
option is "null".
NOTE
The system will automatically calculate the patient age using the entered "Birth
Date" and current "System Date" and display the result in the "Age" box. The
"Age" box will then be greyed, and will become editable again when the "Date of
Birth" is cleared.
The patient birth date should be no later than current system date.
OK
When you have finished entering the sample information, tap "OK" to save the information
and return to the "Table Review" screen.
Cancel
If you do not want to save the entered sample information, tap "Cancel" to return the "Table
Review" screen without saving the changes.
34
Operating with Extended Program
Users of the administrator level may tap "Validate" to validate current or selected record. The
icon would appear at the “Status” cell box of the record. Tap "Cancel Val." to cancel the
validation.
Delete (administrators)
Users of the administrator level may delete desired records by tapping the "Delete" button.
Tap the "Delete" button, below dialog box will pop out:
Select "Selected Records" or "All Records" and tap "OK", below dialog box will pop out. Tap
"Yes" to perform deletion, or tap "No" to cancel deletion.
NOTE
If you tap the "Delete" button when the system is communicating or printing
certain records, a dialog box will pop out prompting "System busy. Please try
again later!".
Select
You can select one or more sample records to perform certain operations like "Comm." or
"Print".
35
Operating with Extended Program
Tap any cell in a sample record to select it. The selected record will be highlighted and a "*"
will be marked next to the sequence no. of the record. Tap again to deselect the record.
You can search for records based on certain conditions in the sample database.
2. Enter "Sample ID", “Patient ID”, patient “Name”, "Date", and Sample "No.", and check
the needed “Sample Status” (“Not Vali.”, “Not Pri.” and “Not Com.”) as necessary to
specify the searching conditions.
3. Check "Select sample(s) found" if necessary, then the samples found will be selected.
If no results are found, a dialog box will pop out prompting "No sample found!". Tap
"OK" to close the dialog box.
On the found results screen, all the buttons are the same as of the "Table Review" screen,
only the original "Search" button is replaced by the "Return" button. Tap the "Return" button to
return to the "Table Review" screen.
36
Operating with Extended Program
Select (a) sample record(s) and tap "Print", the following dialog box will display.
You can "Print Selected" or "Print All" records. Up to 700 records can be printed at the same
time. Tap "OK" to start print. A progress bar will display. Tap "Cancel" to cancel the print.
Communicate
Select to transmit "Selected" or "All" records. Tap the "OK" button to start communication. A
progress bar will display. Tap "Cancel" to close the dialog box.
Export
NOTE
The USB port locates at the right side of the analyzer. Please refer to Figure 2-4
Right side of the analyzer for reference.
Do not pull out the USB before the export process completes.
37
Operating with Extended Program
You can choose to export the selected records or all records. Besides, check “Chromatogram”
for “Export Content” if you want to export the chromatograms. Tap "OK" to start export; tap
"Cancel" to cancel the export. When the export process completes, the system will prompt
whether the operation completes successfully or fails; and a log will created if you select to
export "Sample Data".
38
Operating with Extended Program
Sample
Information
Sample
results
Chromatogram Chromatogram
flag message
Peak
information
Chromatogram
39
Operating with Extended Program
For a sample with the HbE variant, you may see the characteristic peak of HbE between the
SA1c peak and the A0 peak.
The table below the chromatogram displays detailed information for each hemoglobin
component peak.
Name
The name of the hemoglobin component.
Time
Retention time (in seconds), the elapsed time between the start of the run and the apex of the
peaks.
Area
The area of a peak corresponds to the volume of the corresponding component. The unit is
OD.s.
Total Area
This is the sum area for the peaks of all hemoglobin components. The acceptable range is
30~300. In case of insufficient sample volume or very low hemoglobin concentration, the
analyzer may report “Chromatogram area too small”. Predilute the sample as required, and
run the sample using predilute mode.
40
Operating with Extended Program
%
The percentage is the ratio of a peak area against the total area.
You can tap the "Return" button on the upper part of the screen to return to the "Table
Review" screen.
Users of the administrator level may tap "Validate" to validate current record. Tap "Cancel
Val." to cancel the validation.
1. Tap "Logout" from the menu, a dialog box will pop out.
41
Operating with Extended Program
2. After the cleaning, follow the screen instructions to turn off the analyzer.
42
Operating with Extended Program
I-5.1 Introduction
Calibration is a procedure to standardize the analyzer by determining its deviation under
certain specified conditions. In order to get accurate sample analysis results, you should
calibrate the analyzer per the procedure below when necessary.
There are two calibration programs available on this analyzer: manual calibration and auto
calibration using calibrators.
NOTE
Calibration procedures can only be performed by users of the administrator-level.
Use the calibrators and reagents specified by the manufacturer only. Store and
use the calibrators and reagents as instructed by their instructions for use.
43
Operating with Extended Program
you are going to use this analyzer for the first time (usually done by a Mindray-authorized
representative when installing the analyzer).
a major component (for example, the analytical column) has been changed.
NOTE
All of the measured parameters must be calibrated before readings of this analyzer
can be used as valid analysis results.
44
Operating with Extended Program
NOTE
Repeated piercing the cap of the evacuated blood collection tube may damage the
cap, and the scraps produced may result in inaccurate analysis results. It is
recommended that each tube is pierced for no more than 3 times.
NOTE
Be sure to use the evacuated collection tubes recommended in the Appendix.
Tap the menu icon -"Calibrate"-“Manual” to enter the manual calibration screen.
1.
45
Operating with Extended Program
NOTE
If you want to calibrate the Standard program, select "Standard (S)" for
the "Analysis Program".
Select the suitable reference method (“NGSP”, “IFCC” or “Mono-S”) in accordance with your
laboratory procedure from the pull-down list.
Enter values for “a” and “b” as needed.
In the formula Y=aX+b, Y means the results finally shown on the screen, X means the raw
result measured by the analyzer.
Tap “Factory Setup” and set the calibration factors back to factory setup.
46
Operating with Extended Program
From the “Menu” tab, tap “Calibration”--"Calibrator" to enter the “Calibration with Calibrators”
screen.
Lot No.
Enter the lot No. of both the “CAL-1” and “CAL-2” level calibrators into the "Lot No." boxes.
Expiration Date
Runs
Select a number from the "Runs" pull down list. The analyzer will run the calibrators for the
number of times you selected. For example, if you select "2" runs, then the calibrators of both
levels will be run twice each.
Reference Method
Select the suitable reference method (“NGSP” or “IFCC”) in accordance with your laboratory
procedure from the pull-down list.
Parameter Target
Enter the targets of the CAL-1 and CAL-2 calibrators into the corresponding "Target" boxes.
47
Operating with Extended Program
NOTE
See the instruction for use and the target sheet of the calibrators for the lot No.,
expiration dates and the targets.
The entered expiration date should be either the expiration date printed on the
labeling or the open-container expiration date, whichever is earlier. The
open-container expiration date is calculated as follows: the date that container is
opened + the open-container stability days.
If you start the calibration analysis before you complete the calibrator information,
a dialog box will pop up prompting "Invalid entry!". Tap "OK" and enter complete
calibrator information to start calibration analysis.
After you complete the calibration settings, prepare the calibrators for the calibration.
Samples, controls, calibrators and waste are potentially infectious. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them and the contacted areas in the
laboratory.
The sample probe tip is sharp and may contain biohazardous materials. Exercise
caution to avoid contact with the probe when working around it.
NOTE
Only Mindray-specified calibrators shall be used. Mindray will not be responsible
for any erroneous result caused by using other calibrators.
Refer to the instructions for use of the calibrators for their use and storage.
Use clean 1.5mL centrigual tube or the micro sample cup to run the calibration.
Use the CAL tube rack and corresponding tube or sample cup adapters.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes, capillary
tubes and so on.
Make sure the analyzer indicator shows that the analyzer is ready for analysis. Do as follows
to calibrate the analyzer with calibrators.
48
Operating with Extended Program
NOTE
The H50P analyzer supports auto dispensing of Hemolysis Solution. You
can use the function to dilute and prepare controls. For more information,
refer to 5.7 Sample Analysis Under Autoloading Mode.
2. Place the well-mixed calibrators into the specified calibrator tube rack (CAL-1 level
calibrator should be placed at the 2nd tube position, and CAL-2 level calibrator at the
4th tube position), and put the tube rack loaded with the calibrators onto the
autoloader.
3. Tap “Calibration”-“Calibrator” screen, and make sure the calibration information and
parameter targets are correct.
4. Tap the "Start" button, the analyzer will run the calibrator of each level for the number
of times you set.
5. After the analysis is done, the calibration results will display on the screen.
NOTE
If the ambient temperature is out of the specified operating range, the analyzer will
alarm you for abnormal ambient temperature and the analysis results may be
unreliable. When temperature errors are reported in the error information area
after analysis, see Chapter 11 Troubleshooting Your Analyzer for solutions.
After the analysis, the analyzer will have different responses to different analysis results:
if any of the results is invalid, a dialog box will pop up prompting the results are invalid.
if the results are valid, the analyzer will display the new calibration factors automatically.
When you switch to another screen, a dialog box will display asking if you want to save the
calibration results. Tap “OK” to save the calibration results, or tap “Cancel” to give up.
When calibration has been performed for specified times, the analyzer will calculate and
display the CV value and calibration factor automatically.
The calculated calibration factors calculated under different reference systems shall be within
the ranges as below:
49
Operating with Extended Program
When the calculated factors are out of above ranges, a dialog box will display prompting the
results are invalid. Check if there is any error. If you cannot solve the problem, contact
Mindray Customer Service Department.
After calibration, run the remaining calibrators of both levels again using the “AL-PT” mode,
and check the analysis results against the target sheet of the calibrator to make sure they fall
in the acceptable ranges. For more information of the analysis under “AL-PT” mode, refer
to5.7.1 Running the Samples.
If the results falls out of the acceptable ranges, re-calibrate the analyzer. If the problem
cannot be solved, contact Mindray Customer Service Department.
Other Functions
50
Appendices
A Index
accessories, 2-8 linearity range, B-2
analyzer components, 2-3 repeatability, B-2
barcode specifications, B-5 stability, B-2
calibrate, 9-1 peripherals, 2-7
calibration power requirements, B-4
auto calibration, 9-4 Reagent, B-1
manual calibration, 9-3 reagents, 2-19
classification, B-1 replacing
collection tube types, B-1 replacing waste container, 10-11
communication protocol, C-1 safety information, 1-4
connecting the system, 4-4 sample analysis under autoloading
contraindication, B-8 mode, 5-11
Dimensions and Weight, B-5 sample analysis under closed tube mode,
EMC description, B-4 5-7
environment requirements, B-4 sample collection and storage
error information and handling, 11-2 requirements, 3-3
Extended Program, 1, 2 sample preparation, 5-5
fluidics maintenance, 10-3 sample review, 6-1
HbE sampling features, B-2
HbE variant, 2 analysis rate, B-2
Hemoglobin E, 2 minimum sample volume, B-2
how to find information, 1-3 sample volumes required for each
input/output devices, B-3 analysis, B-2
inspection before startup, 5-2 setting up the analyzer, 7-1
installation requirements, 4-2 shut down the analyzer, 5-19
environment requirements, 4-2 software interfaces and operations, 2-14
power requirements, 4-2 sound pressure, B-4
space requirements, 4-2 standby, 5-18
interfaces, B-3 startup, 5-3
limitations, B-8 status
major components and modules, 2-3 temperature & pressure, 10-27
moving and installing the analyzer, 4-3 version information, 10-28
operating your analyzer, 5-1 storage environment requirements, B-5
parameter Symbols:, 1-8
loading modes, 2-2 system functions, 2-12
sample mode, 2-2 system principles, 3-1
parameters, 2-2, B-1 system status, 10-27
performance specifications, B-2 touch screen calibration, 9-8
accuracy, B-2 using the QC program, 8-1
carryover, B-2 when to calibrate, 9-2
A-1
B Specification
B.1 Classification
According to the CE classification, the H50P automated glycohemoglobin analyzer belongs to
In vitro diagnostic medical devices other than those covered by Annex II and devices for
performance evaluation.
Electric Safety Classification
Level of transient overvoltage: Category II.
Rated pollution degree: 2
B.2 Reagents
Name Model
Hemolysis Solution /
Eluent A /
Eluent B /
B.3 Units
Fetal Hemoglobin (HbF), represented in %;
Hemoglobin A1c (A1c), represented in % and mmol/mol;
Total Hemoglobin (HbA1), represented in %;
Estimated Average Blood Glucose (eAG), represented in mg/dL and mmol/L.
B-1
Specification
B.6.2 Repeatability
Running an HbA1c sample with the concentration of 4.0% ~ 6.5% (20.2mmol/mol ~
47.5mmol/mol) for several times, the coefficient variation (CV) between different measured
results should not be higher than 1.0%.
B.6.4 Carryover
The HbA1c carryover of the analyzer should not be higher than 1.5%.
B.6.5 Stability
During the 8 hours after analyzer normal startup, the relative variation between the measured
HbA1c results of the same normal sample should be no higher than ±3%.
B-2
Specification
NOTE
Be sure to use the specified devices only.
B.7.2 Indicator
The indicator indicates analyzer the status of the analyzer, including on/off, running or stanby.
B.7.4 Mouse
USB mouse.
B.7.7 Buzzer
The buzzer sounds when there is an error. Tap the touch screen or when the error is removed,
the buzz alarm will be silent automatically,
B.8 Interfaces
1 10M/100M network interface
4 USB 2.0 interfaces
B-3
Specification
This equipment complies with the emission and immunity requirements of the EN
61326-1:2013 and EN 61326-2-6:2013.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take
measures to mitigate the interference.
NOTE
It is the manufacturer's responsibility to provide equipment electromagnetic
compatibility information to the customer or user.
B-4
Specification
NOTE
Only store and use this system in specified environmental conditions.
Height
Depth
Width
Height(mm) ≤530
Depth(mm) ≤572
Weight(Kg) ≤65
B-5
Specification
Supported Barcodes
B-6
Specification
B-7
Specification
B.16 Contraindication
None
B.17 Limitations
NOTE
Following information is based on the tests running with Standard Program.
The criteria used to determine whether or not a method shows interference that is clinically
significant (indicated by “Yes”) is >±7% at 6 and/or 9% HbA1c.
It is verified against the criteria that the following elements within the stated ranges won’t
present significant interference to the HbA1c results of the H50P analyzer:
Lipemia≤33g/L;
Unconjugated/conjugated bilirubin≤20 mg/dL;
Glucose≤1000mg/dL;
Aldehyde sulfoxylate≤25 mg/dL;
Aspirin≤65 mg/dL;
Sodium cyanate≤5 mg/dL
In accordance with the above mentioned criteria, various hemoglobin variants may interfere
the HbA1c results of the H50P analyzer.
B-8
Specification
Note1: The “Extended Program” of H50P may exclude the interference of hemoglobin variant
HbE to the HbA1c results.
Note2: The Guideline for Glycosylated Hemoglobin Laboratory Assay of China: China J
Diabetes, August 2013, Vol. 21, No. 8.
B-9
C Communication Protocol
The LIS/HIS communication function of H50P enabled the communication between the
analyzer and the PC in laboratory through Ethernet. For details about the connection control,
and the introduction, message definition and examples, please contact Mindray Customer
Service Department or your local distributor.
C-1
P/N: 046-007132-00 (7.0)