Professional Documents
Culture Documents
net/publication/284156953
CITATIONS READS
0 472
3 authors, including:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Bernhard Droste on 18 November 2015.
Abstract
Requirements and test programs for sealed radioactive sources are specified in
international standards for safety in transport and in use.
Sealed sources which are approved as special form radioactive material according to the
Transport Regulations, IAEA Safety Standards TS-R-1, must be able to withstand
mechanical (9 m drop, percussion and bending) and thermal (800°C heat) tests without
loss of radioactive content.
The International Standard ISO 2919 provides a set of tests which classifies the sources
for their safety in use. Performance tests specified in this standard are temperature (high
and low), external pressure, impact, vibration and puncture tests. Each test can be
applied at different levels of intensity depending on typical usage.
As a criterion of pass or fail, leakage testing has to be done after each test.
The poster gives an overview of BAM s comprehensive test equipment and experience
in testing sealed radioactive sources.
Introduction
Requirements and test programs for safety evaluation of sealed radioactive sources are
specified in international safety standards for transport and for use. BAM is the
competent authority for approvals for special form radioactive material and for type
testing of devices with inserted radioactive sources and possesses a comprehensive test
equipment and experience in testing for many years./1/
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
1
Poster presentations
Rolle,A., Neumeyer,T., Droste,B.
Testing of sealed radioactive sources at BAM
A different specimen may be used for each test. As a criterion of pass or fail, leakage
testing according to ISO 9978 has to be done.
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
2
Poster presentations
Rolle,A., Neumeyer,T., Droste,B.
Testing of sealed radioactive sources at BAM
For impact test of special form radioactive material the specimen should be dropped
from a height of 9 m onto a flat unyielding target so as to suffer maximum damage.
Because a definite angle of impact could be important for the damage result BAM use a
magnetic release device for a momentum free drop. A high-speed camera (image
frequency 5.400/s, 1024x1024 pixel) is used to keep track of the impact position.
For impact and puncture test according to ISO 2919 and for percussion test according to
the IAEA regulations for special form radioactive material the same equipment is used
by BAM. (Fig.1)
This test device has also a magnetic release system to assure a momentum free drop,
and is suitable for hammer masses up to 5 kg. Impact tests with higher hammer masses
are performed at the 9 m drop facility.
Most of the different steel hammers for the tests in different classes (Table 1) are
available (Fig.2).
For percussion test with special form designs the specimen shall be placed on a sheet of
lead and struck by a steel bar of 1.4 kg from a height of 1 m. A fresh flat surface of lead
shall be used for each impact. The bar shall strike the specimen so as to cause
maximum damage.
As an example, in case of small welded Ir-192 sources used for the brachy therapy the
bar should strike the weld-seam with its rounded edge to cause maximal bending stress
by pressing the source only partly into the soft lead (Fig. 1 b).
For ISO 2919 impact test class 4 the specimen is placed on a steel anvil with a mass of
at least 20 kg, and is struck by a hammer of 2 kg from a height of 1 m. In most cases
strains are much higher than in case of the IAEA percussion test, were an only 1.4 kg
hammer drives the specimen in the soft lead sheet, excepting cases with additional
bending stress as described above. Beside the striking position, also the form of
simulated content can have an influence on damage results. For example Ir-192 sources
could be filled with one long iridium wire or several shorter iridium-cylinders. Under
impact the edges of the cylinders can partly push trough the thin capsule wall.
Steel hammers for puncture test bear a fixed pin with a diameter of 3 mm and a height
of 6 mm and can also be attached at the magnetic release device .The mass depends on
the test-class and varies between 1 g and 1 kg. If source design has more than one
vulnerable area, tests have to be carried out on each of them.
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
3
Poster presentations
Rolle,A., Neumeyer,T., Droste,B.
Testing of sealed radioactive sources at BAM
a) b)
Fig.1: Percussion test on a Ir-192 source for brachy therapy (1.4 kg from 1 m on a sheet of lead) with
different impact positions a) less damage , maximum bending stress in the flexible wire; b) maximum
bending stress in weld seam of the source
a) b)
Fig.2: Examples for steel hammers for a) puncture and b) impact or percussion test
Temperature tests
An approval for special form radioactive material requires a heat test in air at a
temperature of 800°C for 10 min.
Temperature tests according to ISO 2919 include a high–temperature test at
temperatures of 80 to 800°C for 1h and a low-temperature test to -40°C for 20 min. The
class 6 test (-40°C for 20 min, 800°C for 1 h) as an adequate alternative to the IAEA
heat test (800°C for 10 min) is often employed by BAM.
BAM uses for upper temperature test a furnace with a capacity of 10 kW for a
maximum temperature of 1200°C to be reached in 2.5 h. Attention should be paid to
positioning the specimen in the middle of the furnace and to fitting the thermocouple
directly on the specimen surface. In case of designs with a low-melting content, for
example distance pieces made of aluminium, an eutectic reaction can happen and can
cause a damage of the capsule material and in worst a leakage.
A climate chamber (-80°C until 190°C) is used for low temperature test.
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
4
Poster presentations
Rolle,A., Neumeyer,T., Droste,B.
Testing of sealed radioactive sources at BAM
Bending test
Bending tests shall apply only for long, slender sources. IAEA regulations require this
for sources with both a minimum length of 100 mm and a length to minimum width
ratio of not less than 10. The free end of a rigidly damped specimen has to be struck by
a 1.4 kg steel bar. BAM uses a common clamping tool for fixing the specimen and a
tripod to adjust the drop height.
Bending tests according to ISO 2919 shall apply to sealed sources having a length to
minimum width ratio of 15 or greater. Test procedure is different from this described
above. BAM has up to now never carried out this ISO bending test.
Vibration test
BAM features an electro-dynamic shaker with up to 26 kN peak force and frequencies
of 5-2000 Hz with acceleration and force control for performing the vibration test
according to ISO 2919. Up to now none of the tested sealed sources have shown any
signs of damage after these tests done by BAM. Often applicants abstain from this test.
Table 2.: Minimum void in capsule for different volumetric leak test methods
Leak test method Minimum void in capsule
[mm³]
Vacuum bubble
- glycol or isopropyl alcohol 10
- water 40
- Pressurized bubble with isopropyl alcohol 10
Liquid nitrogen bubble 2
Helium pressurization 10
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
5
Poster presentations
Rolle,A., Neumeyer,T., Droste,B.
Testing of sealed radioactive sources at BAM
Conclusions
The International Standards ISO 2919 and the IAEA Safety Standard TS-R-1 provide a
set of mechanical and temperature tests for assessing the safety of sealed radioactive
sources in transport and use. BAM possesses comprehensive knowledge, experience
and equipment for performing nearly all of these tests.
Since BAM is the competent authority for approval of special form radioactive material
tests according to the IAEA regulations have always the priority. But depending for
their work load test labs are always ready to carry out also the classification tests
according to ISO 2919. About four weeks after incoming of specimens, a test program
for one source design can be finished in general.
BAM actively works in further development of test methods and regulations and its ´
representatives take part in national and international working groups and boards
dealing with this topic.
References
/1/Rolle,A,Droste,B.Safety during whole life time: important aspect in safety assessment of sealed
radioactive sources. Packaging, Transport, Storage & Security of Radioactive Material 2008; Vol 19
(No 3):151-154
2/ Regulations for the Safe transport of Radioactive material, 2005 Edition, International Atomic Energy
Agency (IAEA) No. TS-R-1
/3/ International Standard ISO 2919:1999, Radiation protection- Sealed radioactive sources-General
requirements and classification
/4/ International Standard ISO 9978:1992, Radiation protection-Sealed radioactive sources-Leakage test
methods
Proceedings of Third European IRPA Congress 2010 June 14−16, Helsinki, Finland
6
View publication stats