AG600, AGZ600
AG100-AG500,
AGZ100-AGZ500
AG1000-AG4000
AGZ1000-AGZ4000,AGZ10
USER MANUAL
AG and AGZ series
Contents:
1. General description.................................................................................................. 3
2. Completeness .......................................................................................................... 3
3. Technical data ......................................................................................................... 4
4. General balance view .............................................................................................. 5
5. Keys and indicators ................................................................................................. 7
6. Safety rules.............................................................................................................. 8
7. Preparing working environment ............................................................................... 9
8. Preparing balance to work ..................................................................................... 10
9. General operation principles .................................................................................. 11
10. Start-up .................................................................................................................. 12
11. Internal calibration ................................................................................................. 12
12. Checking the balance ............................................................................................ 14
13. Connecting a computer, printer or label printer ...................................................... 14
14. Special functions.................................................................................................... 19
14.1 Autotaring function ............................................................................................... 20
14.2 Pieces counting function....................................................................................... 21
14.3 Unit selection function .......................................................................................... 23
14.4 Percent function ................................................................................................... 24
14.5 Function for calibration with external weight / calibration options.......................... 25
14.6 Function for setting serial port .............................................................................. 29
14.7 Print settings function ........................................................................................... 31
14.8 Time and date setting function.............................................................................. 34
14.9 LCD settings function ........................................................................................... 35
14.10 Recipe function ................................................................................................... 36
14.11 Animals weighing function .................................................................................. 37
14.12 Tare setting function ........................................................................................... 38
14.13 Force unit function .............................................................................................. 39
14.14 Maximum and minimum value indication function ............................................... 40
14.15 Anti-disturbance filter function............................................................................. 41
14.16 Language selection function ............................................................................... 42
14.17 Function of comparing with preset threshold values ........................................... 43
14.18 Funkcja sumowania ważeń TOTAL .................................................................... 47
14.19 Statistical calculations function ........................................................................... 48
14.20 Funkcja wyliczenia gramatury papieru ................................................................ 50
14.21 Density measurement function (option) .............................................................. 50
14.21.1 Solid body density determination ..................................................................... 52
14.21.2 Liquid density determination ............................................................................ 54
14.21.3 Density measurement report ........................................................................... 56
15. Procell program description (demo version)........................................................... 57
16. Troubleshooting and maintenance ......................................................................... 58
Declaration of Conformity
USER MANUAL 3
____________________________________________________________________________________________________________________
1. General description
AG and AGZ series balances are destined for high accuracy weighing in
laboratory practice. AG series balances are equipped with internal calibration
system for accuracy control during balance operation. AGZ series balances do
not have internal calibration system and can be used when legal verification is
not required.
All balances are metrologically tested. According to an order balances can be
calibrated or legally verified.
Balances with legal verification are marked with the following legal and
securing items:
- green metrological mark placed on the balance name plate,
- notified body stamp (number of notified body) on the balance name plate,
- protective seals placed on: an edge of balance name plate, the casing mounting
screw and in the place of access to adjustment switch,
Renewing of balance legal verification is required when protective seal is
violated or after period of 3 years starting from 1 st December of year when first
legal verification was performed. In order to renew legal verification please
contact AXIS authorised service centre.
In AGZ series balances all functions connected with internal calibration are
removed (chapter 11 and 14.5).
Balance classification according to PKWiU: 33.20.31.
Certificates:
3. Technical data
AG100 AG200 AG300 AG500
Type AGZ100 AGZ200 AGZ300 AGZ500
Capacity (Max) 100g 200g 300g 500g
Min load (Min) 0,02g 0,02g 0,02g 0,02g
Reading unit (d) 0,001g 0,001g 0,001g 0,001g
Verification unit (e) 0,01g 0,01g 0,01g 0,01g
Tare range -100g -200g -300g -500g
Accuracy class II
Working temperature +18 ÷ +33°C
Weighing time <3s
Pan dimension 115mm
Balance dimension 215(235 with legs)x345x90mm
(with legs)
Balance weight 5kg
Power supply ~230V 50Hz 6VA / =12V 800mA
Recommended external
calibration weight F2 100g F2 200g F2 200g F1 500g
(OIML)
Caution:
F1 and F2 are international symbols of calibration weight classes according to
O.I.M.L. Some requirements for weight accuracy are connected with those
classes.
USER MANUAL 5
____________________________________________________________________________________________________________________
8 1
2
3 1 – pan
2 – pan support
3 – pan ring
4 4 – graphical display
5 – keys
5 6 – rotating legs
7 – water level
7 8 – draft shield
Note:
AG600 does not
6
have the pan ring
and the draft shield.
1
2
5 1 – pan
2 – nuts (under pan)
7 4 – graphical display
5 – keys
6 6 – rotating legs
7 – water level
6 USER MANUAL
____________________________________________________________________________________________________________________
Connectors view:
Printer Keyboard
Feeder or computer
~230V/50Hz
6. Safety rules
+33°C
+18°C
Location for the balance should be chosen with care in order to limit influence
of the factors that can interrupt working balance. This location has to maintain
proper temperature for working balance and necessary space for its operating.
The balance should stay on stable table made of material that does not influence
magnetically on the balance.
Rapid air blast, vibrations, dust, rapid temperature changes or air humidity over
90% are not allowed in balance surrounding. The balance should be far from
heat sources and devices emitting strong electromagnetic or magnetic fields.
10 USER MANUAL
____________________________________________________________________________________________________________________
4. (for AG100-AG600)
Gently insert the mandrel
of pan support 2 into
balance mechanism
7 socket through the pan
ring 3 and the pan 1 on
(AG600 balances have
not the pan ring).
6
1
5. (for AG1000-AG4000)
Place nuts 2 on mandrels
2 that are visible in
balance cover holes, put
the pan 1 on nuts.
For transportation take off the pan and pan support (AG(Z)100-
AG(Z)600 balances) or nuts (AG(Z)1000-AG(Z)4000 and AGZ10
balances), on which the pan is placed.
10. Start-up
Plug feeder into 230V power supply socket and feeder output connector into
12V socket at back of the balance.
2
6 When cooperating with computer, the
RxD (receive) 7 scale sends weighing result after initialize
3
TxD (transmission)
8
signal from computer or after pressing
4 key on the scale.
9 +5V(option)
5 When cooperating with a printer data is
GND
send automatically after result
housing stabilisation, but next transmission is
possible after removing previously
weighted sample.
When cooperating with label printer after pressing key, the scale sends
instructions set for the label printer. Label number 0001, hour, data (if the clock
USER MANUAL 15
____________________________________________________________________________________________________________________
is installed and on) and nett weight. During transmission LabEL communicate is
displayed.
The way of sending data and transmission parameters is set using SErIAL
special function.
Set of send data is set using special function PrInt.
The following data can be send:
- Header (scale type, Max, d, e, serial number),
- Operator identification number,
- Successive printout number (measurement),
- Identification number or product bar code,
- Number of pcs (PCS function only),
- Single detail mass (PCS function only),
- Nett weight,
- Tare (package mass),
- Gross weight,
- Total mass (Total function only).
Computer must have a special program for cooperation with data from a scale.
Dedicated programs are also offered by AXIS.
Except RS232C joint, the scale can be equipped with USB or Wi-Fi interface.
Needed controllers and instruction can be found on a CD supplied with Axis
scales.
Detailed protocol description
LonG protocol
Communication parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps,
Attention:
Network number different than zero (Port/nr function) changes scale working
mode: communication with a computer is possible after logging the scale in with
02h scale number command. To log the scale out use 03h command.
For example: Using a program to test RS232 interface ( program is available in
www.axis.pl / programy komputerowe ) for scale number 1 please write: $0201
to log in, then SI, and write: $03 to close communication.
in order to set low threshold 100kg in scale B150 (d=50g) the following
order should be sent:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
Setting high threshold value (option):
Computer Scale: S H D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah),
D1...DN – threshold value (see )
Scale Computer: without response.
BALANCE PRINTER
ELtron protocol
US - Steering instruction
FR"0001" - Label number define instruction
? - Instruction that starts list of variable signs
mm:gg - 5 signs: minutes:hour
rrrr.mm.dd - 10 signs: year.month.day
masa - 10 signs: scale indication+ mass unit
P1 - Steering instruction
Attention:
1. Except variable signs constant signs also can be inscribed for example factory
name, product name and so on.
2. In standard one label pattern is possible to prinout (number 0001). Using
bigger amount of patterns (other label numbers) is possible thanks to LAbEL
special function.
3. To achieve label printout, label printer must have inscribed label pattern (label
pattern is created on computer and using computer it is saved to label printer
memory). Label pattern is designed by ZEBRA DESIGNER program which is
supplied together with label printer.
Scales parameters and transmission protocol must corespond to label printer
type.
USER MANUAL 19
____________________________________________________________________________________________________________________
14.Special functions
All balances, beside basic metrological functions like weighing and taring, are
equipped with set of special functions. Basic set of special functions is shown
below. In respect of metrology calibration with external weight is important
special function.
To avoid showing too many functions in menu, user can choose which functions
should be displayed in menu and should not. Menu settings function is used for
this.
This function automatically keeps zero indication when a pan is empty or zero
indication was forced with T key.
Max ... Min ... e= ... d= .. Indications that are near to zero will
AUT be automatically corrected and zero
indication will be maintained despite
0.000 g changing environmental conditions
0% 100%
(temperature, air humidity, etc.)
To stop the function press Menu key to enter user functions menu, choose
Autotaring function and then choose Deactivate option.
USER MANUAL 21
____________________________________________________________________________________________________________________
USER FUNCTIONS&OPT.
Press Menu key to enter the user
Autotaring function menu, choose PCS counting
PCS counting with the cursor and press ENTER key.
. . .
To stop the function press Menu key, choose PCS counting function and then
choose Deactivate option.
Note:
In order to show mass value temporarily and return to pieces indication press
key.
22 USER MANUAL
____________________________________________________________________________________________________________________
Note: To correct wrong digits when entering unit weight, press < key to delete the
last number or CLR to leave the function and proceed from the beginning.
INSTRUKCJA OBSŁUGI 23
USER FUNCTIONS&OPT.
Autotaring
PCS counting Press Menu key, choose Unit
Unit selection selection with the cursor and press
. . . ENTER key.
Exit
USER FUNCTIONS&OPT. \ UNIT SELECTION Using > and < keys choose proper
unit and press ENTER key.
1. Unit : < gram >
2. Exit
Press < key or choose Exit option to
g return to weighing.
24 USER MANUAL
____________________________________________________________________________________________________________________
USER FUNCTIONS&OPT.
Autotaring Press Menu key, choose Percent with
PCS counting the cursor and press ENTER key.
Unit selection
Percent
. . .
Comments:
1. In order to show mass value in grams temporarily and return to percentage
indication press key.
2. According to a mass referenced to 100% percentage indications will be
displayed in different formats. For reference mass in range 0÷3,5% of balance
capacity, displayed result will be shortened of two decimal places, in range
3,5÷35% - shortened of one decimal place, above 35% - in full accuracy.
USER MANUAL 25
Adjustment Mark
switch
ON OFF
USER FUNCTIONS&OPT.
Autotaring
Press Menu key, choose Calibration
PCS counting
option with the cursor and press
. . .
Calibration ENTER.
. . .
CALIBRATION
Put calibration weight matching
External calibration:
Put external load _200.000 g
displayed value on.
CALIBRATION
External calibration:
take off ext. load
Take off the calibration weight.
CALIBRATION
Wait until internal calibration is
Internal calibration:
finished.
please wait ...
USER FUNCTIONS&OPT.
Autotaring Press MENU key, choose Calibration
PCS counting option with the cursor and press
. . . ENTER.
Calibration
. . .
-----------Calibration report-----------
Scale type:
Date of adjustation:
Temperature of adjust.:
Factory external weight: ... - external weight value registered during factory calibration
Factory internal weight: ... - internal weight value registered during factory calibration
Current external weight: ... - external weight value registered during last calibration
Current internal weight: ... - internal weight value registered during last external calibration
Weight difference: ... - difference between internal weight values: factory value–current value
USER MANUAL 29
USER FUNCTIONS&OPT.
Autotaring Press MENU key, choose Serial Port
PCS counting with the cursor and press ENTER key.
. . .
Serial port
. . .
Check if port interface settings are
USER FUNCTIONS&OPT./SERIAL PORT compliant with parameters of
connected external device. If not,
1. Port 1 using ∨ and ∧ keys move the cursor on
2. Port 2 parameter that should be set and press
3. Wyjście ENTER key.
1. Address : 2345
2. Exit
USER MANUAL 31
USER FUNCTIONS&OPT.
Autotaring Press MENU, choose Print settings
PCS counting with the cursor and press ENTER key.
. . .
Print settings
. . .
Header - the entrance to the printout header definition menu; sign indicates if at
least one option in the Header definition menu is checked.
blank line
--------------- WEIGHING --------------- mode name
Date : 2000-04-25 Time : 22:32 date and time
Scale type : AGNZ200 balance type
Serial number : 123456 serial number
ID1 string ID1
ID2 string ID2
ID3 string ID3
Signature signature
.....................................
blank line
ID1 string ID1
ID2 string ID2
ID3 string ID3
Measurement number : 1 measurement number
T 0.0000 g tare
N 66.7425 g net
B 66.7425 g gross
LCD 333.7125 ct LCD result
USER MANUAL 33
____________________________________________________________________________________________________________________
Footer - the entrance to the printout footer definition menu; sign indicates if at
least one option in Footer definition menu is checked.
blank line
------------ PCS COUNTING ------------- mode name
Date : 2000-04-25 Time : 23:05 date and time
Scale type : AGNZ200 balance type
Serial number : 123456 serial number
ID1 string ID1
ID2 string ID2
ID3 string ID3
Signature signature
.....................................
--------------------------------------- separating line
3 empty lines
ID1, ID2, ID2 - strings (max. 20 characters) typed using PS2 keyboard or balance
numeric keypad, which works similarly to mobile keyboard (characters coupled
with relevant key appear in first line of the display after the key is pressed); typed
string is approved with ENTER key, last character can be deleted using < key.
34 USER MANUAL
____________________________________________________________________________________________________________________
USER FUNCTIONS&OPT.
Autotaring Press MENU, choose Time&date
PCS counting setting with the cursor and press
. . . ENTER key.
Time&date setting
. . .
USER FUNCTIONS&OPT.
Autotaring
PCS counting
Press MENU, choose LCD settings
. . . with the cursor and press ENTER key.
LCD settings
. . .
USER FUNCTIONS&OPT.
Press MENU key to enter the user
functions menu, select Recipe with
Autotaring cursor and press ENTER.
PCS counting
. . .
Recipe Using ∨ and ∧ keys, move cursor to
Activation and press ENTER.
USER FUNCTIONS&OPT. \ RECIPE
The balance is ready for weighing the
1. Activate
2. Reset successive ingredients, and after each
3. Exit ingredient it is necessary to press T
key. It will zero the balance indications.
On the left side the sum of previously
Max ... Min ... e= ... d= ..
=0.000g
weighed ingredients () and their
n=0 SUM number (n) is displayed.
To read the total mass of all weighed
0.000 g ingredients use key (pressing it
0% 100% again will cause return to ingredient
weighing).
USER
To finish the FUNCTIONS&OPT.
function operation,\ press
RECIPE
MENU to enter the user functions menu, select Recipe
function, and select Deactivation option.
1. Deactivate
2. Reset
3. Exit
Total weight: 0.000g
Number of ingred.: 0
USER MANUAL 37
____________________________________________________________________________________________________________________
Result 10.123 g
0% 100%
38 USER MANUAL
____________________________________________________________________________________________________________________
USER FUNCTIONS&OPT.\ STRENGHT UNIT Press MENU key. Using ENTER key
choose Strenght unit function.
1. Activate Choose Activate option and confirm
2. Exit using ENTER.
N
Attention:
Units converting from mass (kg) to force (N) is made for acceleration of gravity
(g=9,80665m/s2):
1N0,101971 kg
40 USER MANUAL
____________________________________________________________________________________________________________________
This function allows for language selection for messages and print purposes.
USER FUNCTIONS&OPT.
Press MENU key, select Language
Autotaring selection with cursor and press
PCS counting ENTER.
. . .
Language
Exit
.
USER MANUAL 43
____________________________________________________________________________________________________________________
P1
zero Min
treshold thresholdI
P2
Min Max
threshold thresholdI
P3
Max
2. Impulse mode threshold
(approx. 2 sec.):
Outputs: 2s.
P1
Min
threshold
2s.
P2
Max
threshold
P3
zero
treshold
3. Level mode:
Outputs:
P1
zero Min
treshold threshold
P2
Min Max
threshold threshold
P3
Max
treshold
46 INSTRUKCJA OBSŁUGI
____________________________________________________________________________________________________________________
Scale Relay
P1
P2
P3
GND Imax < 25mA
Shield
INSTRUKCJA OBSŁUGI 47
____________________________________________________________________________________________________________________
AUTO 0.000 g
0% 100%
INSTRUKCJA OBSŁUGI 49
____________________________________________________________________________________________________________________
1 7,5476
2 7,5480 The individual measurements
3 7,1902 results are printed during
4 6,8227 performing (option: Print ON).
5 6,4719 After pressing F1 key the
------------------------------------------------------
statistical parameters are
Number of samples = 5
printed.
Average value = 7.11608 g
Stand. deviation = 0.93771 g
Relative deviation = 13.18 %
Minimal value = 6.4719 g
Maximal value = 7.5480 g
To finish the function operation, press MENU to enter the user functions menu,
select Statistics function, and select Deactivation option.
50 INSTRUKCJA OBSŁUGI
____________________________________________________________________________________________________________________
GRMG
0.000 g/m 2
0% 100%
INSTRUKCJA OBSŁUGI 51
____________________________________________________________________________________________________________________
This function allows for determination of solid body density, upon the basis of
weight in the air and weight of material immersed in liquid of known density,
according to the formula:
m1
ρ = * ρ liquid
m1- m2
where:
m1-mass in the air
m2-mass in the liquid
This function also allows for determination of liquid density, upon the basis of
plunger weight (with known density) in the air and tested liquid. The following
formula is used:
m1 - m2
ρ =
V
where:
m1-plunger mass in the air
m2-plunger mass in the liquid
V – plunger volume
USER FUNCTIONS
Suspend the trays and tare scale
Autotaring pressing button T.
PCS counting
. . . Press button Menu, select Density
Density measurement
measurement with cursor and press
. . .
ENTER..
1.Activate
2.Material type : solid Select Material type with cursor and
3.Liquid type : water 20.0oC press ENTER.. Use > and < keys to
4.Liquid density : 0.99820g/cm3 choose solid material and press ENTER
5.Report printout
6.Exit
>T< 0.0000 g
0% 100%
1.Activate
Select Material type with cursor and
2.Material type : solid press ENTER.
3.Liquid type : water 20.0oC
4.Liquid density : 0.99820g/cm3
5.Report printout
6.Exit
USER FUNCTIONS / DENSITY Use > and < keys to choose Plunger
1.Activate
volume and press ENTER
2.Material type : liquid
3.Plunger volume : 0.00000 cm3
4.Report printout
5.Exit
>T< 0.0000 g
0% 100%
ENTER 25.5462 g
0% 100%
To print the measurement results, connect the printer to RS232C connection of the
balance. The connection method is described in the instruction manual.
USER FUNCTIONS / DENSITY
Print example:
Measurement number =
Mass in the air = g
Mass in the liquid = g
Density … = g/cm3
Density with comp. = g/ cm3
Water density = g/cm3
o
Water temperature = C
USER MANUAL 57
____________________________________________________________________________________________________________________
In order to write a series of weighing results into Excel table proceed with the
following:
1. Connect the balance to RS232C computer serial port,
2. Run ProCELL,
3. During first start-up close registration window (or contact AXIS to register
the program),
4. Open spreadsheet (ProCELL will be covered by application window),
5. Set the cursor in desired place and press key on the balance several times.
The program automatically detects computer port (COM1-COM4), sets
communication parameters and hides into taskbar at the bottom of the screen (or
into systray). After that only Excel spreadsheet and key on the balance are
used.
58 INSTRUKCJA OBSŁUGI
_________________________________________________________________________________________________________________
___
Failure messages
Declaration of Conformity
We:
AXIS Spółka z o.o. 80-125 Gdańsk, ul. Kartuska 375B