Professional Documents
Culture Documents
HHS
Tools and Miscellaneous Test Procedures
OPERATING DOCUMENTATION
46-017560
Revision 14
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
2
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
Important Information
LANGUAGE
ПРЕДУПРЕЖДЕНИЕ Това упътване за работа е налично само на английски език.
(BG) • Ако доставчикът на услугата на клиента изиска друг език, задължение на клиента
е да осигури превод.
警告 本维修手册仅提供英文版本。
(ZH-CN) • 如果客户的维修服务人员需要非英文版本,则客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修服务人员、操作人员或患者造成电击、机械伤害或其他形式的
伤 害。
警告 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其他的危險。
警告 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危險而受傷。
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge, operatera ili
pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.
Important Information 3
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a
pochopili jeho obsah.
• Do not attempt to service the equipment unless this service manual has been con‐
sulted and is understood.
• Failure to heed this warning may result in injury to the service provider, operator or
patient from electric shock, mechanical or other hazards.
4 Important Information
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt tämän
huolto-ohjeen.
• Ne pas tenter d'intervenir sur les équipements tant que ce manuel d’installation et de
maintenance n'a pas été consulté et compris.
• Versuchen Sie nicht diese Anlage zu warten, ohne diese Serviceanleitung gelesen
und verstanden zu haben.
• Μην επιχειρήσετε την εκτέλεση εργασιών σέρβις στον εξοπλισμό αν δεν έχετε
συμβουλευτεί και κατανοήσει το παρόν εγχειρίδιο σέρβις.
Important Information 5
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Reynið ekki að afgreiða tækið nema að þessi þjónustuhandbók hefur verið skoðuð og
skilin.
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda eða
sjúklings frá raflosti, vélrænu eða öðrum áhættum.
• Il mancato rispetto della presente avvertenza potrebbe causare lesioni all'addetto alla
manutenzione, all'operatore o ai pazienti provocate da scosse elettriche, urti mecca‐
nici o altri rischi.
警告 このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者 さんが、
感電や機械的又はその他の危険により負傷する可能性があります。
6 Important Information
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Jei nepaisysite šio įspėjimo, galimi paslaugų tiekėjo, operatoriaus ar paciento suža‐
lojimai dėl elektros šoko, mechaninių ar kitų pavojų.
• Manglende hensyn til denne advarselen kan føre til at serviceleverandøren, oper‐
atøren eller pasienten skades på grunn av elektrisk støt, mekaniske eller andre farer.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual
de assistência técnica.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual
de assistência técnica.
Important Information 7
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Ne pokušavajte da opravite uređaj ako niste pročitali i razumeli ovo servisno uputstvo.
• No se deberá dar servicio técnico al equipo, sin haber consultado y comprendido este
manual de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor de servicios,
el operador o el paciente sufran lesiones provocadas por causas eléctricas, mecáni‐
cas o de otra naturaleza.
8 Important Information
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
• Försök inte utföra service på utrustningen om du inte har läst och förstår den här
servicehandboken.
• Om du inte tar hänsyn till den här varningen kan det resultera i skador på servicete‐
knikern, operatören eller patienten till följd av elektriska stötar, mekaniska faror eller
andra faror.
Important Information 9
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
This page left intentionally blank.
10 Important Information
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
Revision History
Rev Date Reason for change Pages
2 Dec. 1, 1997 Revised to include hardcopy form references for third party users. -
4 April 23, 2004 Fluke87 multimeter added to list of tools in section 1, BUCge93442. -
6 February 15, 2006 Added: Innova 3100–IQ and Innova 4100–IQ information -
Added: Innova 2100–IQ suitcase
7 July 20, 2007 Added: Innova 2121–IQ and Innova 3131–IQ information. -
9 March 17, 2008 Chapter 2 section 1, Chapter 5 sections 1 and 2 (REG0093) updated with new 48
Victoreen tool
Chapter 5, Radiation Instruments, updated with a new section “4 KVp & HVL
Measuring Instruments”
11 July 27, 2011 Updated to introduce Innova IGS 520, 530, 540 48
14 May 29, 2012 Chapter 2, section “Introduction” has been updated to introduce Optima CL320i/ 48
CL323i
Revision History 11
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
This page left intentionally blank.
12 Revision History
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
Table of Contents
Chapter 1 SAFETY.................................................................................................................................15
1 Safety............................................................................................................................................15
1.1 X-Ray Protection..................................................................................................................15
1.2 Certified Electrical Contractor Statement.............................................................................16
1.3 Damage in Transportation....................................................................................................16
Table of Contents 13
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
14 Table of Contents
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
Chapter 1 Safety
1 Safety
1.1 X-Ray Protection
WARNING
X-RAY EQUIPMENT IF NOT PROPERLY USED MAY CAUSE INJURY.
ACCORDINGLY, THE INSTRUCTIONS HEREIN CONTAINED SHOULD BE
THOROUGHLY READ AND UNDERSTOOD BY EVERYONE WHO WILL USE
THE EQUIPMENT BEFORE YOU ATTEMPT TO PLACE THIS EQUIPMENT IN
OPERATION. THE GENERAL ELECTRIC COMPANY, MEDICAL SYSTEMS
GROUP, WILL BE GLAD TO ASSIST AND COOPERATE IN PLACING THIS
EQUIPMENT IN USE.
ALTHOUGH THIS APPARATUS INCORPORATES A HIGH DEGREE OF
PROTECTION AGAINST X-RADIATION OTHER THAN THE USEFUL BEAM, NO
PRACTICAL DESIGN OF EQUIPMENT CAN PROVIDE COMPLETE
PROTECTION. NOR CAN ANY PRACTICAL DESIGN COMPEL THE OPERATOR
TO TAKE ADEQUATE PRECAUTIONS TO PREVENT THE POSSIBILITY OF ANY
PERSONS CARELESSLY EXPOSING THEMSELVES OR OTHERS TO
RADIATION.
IT IS IMPORTANT THAT EVERYONE HAVING ANYTHING TO DO WITH X-
RADIATION BE PROPERLY TRAINED AND FULLY ACQUAINTED WITH THE
RECOMMENDATIONS OF THE NATIONAL COUNCIL ON RADIATION
PROTECTION AND MEASUREMENTS AS PUBLISHED IN NCRP REPORTS
AVAILABLE FROM NCRP PUBLICATIONS, 7910 WOODMONT AVENUE, ROOM
1016, BETHESDA, MARYLAND 20814, AND OF THE INTERNATIONAL
COMMISSION ON RADIATION PROTECTION, AND TAKE ADEQUATE STEPS
TO PROTECT AGAINST INJURY. THE EQUIPMENT IS SOLD WITH THE
UNDERSTANDING THAT THE GENERAL ELECTRIC COMPANY, MEDICAL
SYSTEMS GROUP, ITS AGENTS, AND REPRESENTATIVES HAVE NO
RESPONSIBILITY FOR INJURY OR DAMAGE WHICH MAY RESULT FROM
IMPROPER USE OF THE EQUIPMENT. VARIOUS PROTECTIVE MATERIAL
AND DEVICES ARE AVAILABLE. IT IS URGED THAT SUCH MATERIALS OR
DEVICES BE USED.
CAUTION
United States Federal law restricts this device to use by or on the order of a
physician.
If you have any comments, suggestions or corrections to the information in this document, please
write them down, include the document title and document number, and send them to:
Chapter 1 Safety 15
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
AMERICAS W–622
MILWAUKEE, WI 53201–0414
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after
damage is found. At this time be ready to supply name of carrier, delivery date, consignee name,
freight or express bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure
Bulletins (6/17/94).
16 1 Safety
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
NOTE: Tools must be properly calibrated and in good working order. Calibration data must be
recorded on data form which includes required GE calibration frequencies.
1. Field Laptop Computer with Excel spreadsheet versions of HHS Data Sheets. These are found
in the XR TOOLS window. Alternately, the sample forms in 46–017566, HHS FORMS,
COMPLETING THE FDA 2579 AND GE DATA RECORD FORMS can be photocopied.
2. HHS Field Test Kit 46–303879G1, or old style kit, 46–177372G1 with upgrade 46–315614G1.
○ (For Innova 2000/2100-IQ/2121-IQ / Innova IGS 520 / Innova IGS 620) use the 2280567
Lead Sheet suitcase.
○ (For Innova 3100/3100-IQ/3131-IQ / Innova IGS 530 / Innova IGS 630 / Discovery IGS
730 / Optima CL320i/CL323i) use the 2408517 HHS test kit.
○ (For Innova 4100/4100-IQ / Innova IGS 540) use the 2370573 HHS test kit)
3. Keithley 35080 HV Divider non-invasive (other kVpmeters (refer to Chapter 5, Radiation
Measuring Instruments, section 4, KVp & HVL Measuring Instruments).
4. Radiation meter with rate and integrate modes (Refer to Chapter Radiation Instruments).
5. Victoreen 470A survey meter, or MDH meter with earphone, or Victoreen 451B (5270475) for
vascular.
6. Mammo probe for rate meter.
7. mAs meter.
8. Light meter, Digaphot 3300/3303 or equivalent, must have an accuracy of 5% or better.
9. Fluke 87 series 3 multimeter.
10. Digital Voltmeter: Fluke 8030A or Beckman 3030 RMS or equivalent.
11. Film marking pencils or felt tip pens. The pencils are available from the supplies section (Cat.
No. E7005A for box of 12).
12. Additional instruments described in the individual service manuals for functional check.
13. Long Afterglow Phosphor tool, 2120565, for light to X–ray field alignment test and spot–film
device X–ray field size test.
NOTE: If the system fails the SID test of 46–017564 HHS COLLIMATOR TESTS, it may be
necessary to obtain a Focal Spot Locator Kit, 46–250080G1.
3 Calibration Services
All services described here must be traceable to the National Institute of Standards & Technology
(NIST).
1. The Manufacturers Factory Authorized Service Centers for all equipment except radiation
meters by Capintec (models 192 and 174) and MDH (2025).
2. Reputable local metrology labs and test equipment service vendors as determined by Region
management for all equipment except radiation meters by Capintec (models 192 and 174) and
MDH (2025).
NOTE: Radiation meter calibration must include calibration of the electrometer and
calibrating of all ion chamber probes. For X–Ray service applications, the ion
chamber should be calibrated at one point only. The optimum calibration point for
X–Ray service use is 100 kVp/HVL 5 mm Al. A vendor may offer to furnish and charge
an additional amount for ion chamber calibration at more than one calibration point.
Make sure that you are not paying for unnecessary calibration at more than one
point.
3. GE Instrumentation & Computer Services (GE-ICES) repairs and calibrates all types of
instruments except radiation meters by Capintec (models 192 and 174) and MDH (2025).
4. Capintec and MDH meter should be sent to GE Electronic Services in Kansas City, with one
exception: As of this writing, GEES cannot calibrate a mammo probe for the MDH meter. If
your MDH meter has a mammo probe, refer to the following step.
5. Radcal Corp. Division of MDH for MDH 2025 radiation meters with mammo probes.
Instrument Manufacturer Repair Service Mid Digits for 631 and 641
NOTE: All maintenance of expense tools and equipment should be charged to 632-00-xx.
It is important that the correct account numbers be utilized so that equipment calibration
and repair costs may be accumulated in order to measure equipment performance and
for blanket repair/calibration contract negotiation purposes.
5 Shipment to Vendors
When shipping test equipment to vendors for repair and calibration, use the following guidelines:
1. All shipments are to be on a purchase order; this assures return of the instruments.
2. Transportation to the vendor should be prepaid. The return freight will be prepaid by the vendor
unless specified otherwise on the purchase order.
3. Premium transportation charges are the district’s responsibility. Authorization for non-surface
transportation is necessary.
1. An RGD form for return to the Parts and Repair Operation Service Shop, Waukesha,
Wisconsin.
2. An RGD form for return of High Voltage Bleeders.
7 Calibration Responsibility
Each Regional Manager shall assign an employee in the Region to be responsible for:
• Insuring that a calibration program is maintained in accordance with this P&P Bulletin.
• Insuring that calibration records are maintained in the Region for all instruments covered by
this P&P Bulletin.
24 7 Calibration Responsibility
HHS Tools and Miscellaneous Test Procedures
GE Healthcare Direction 46-017560, Revision 14
2.2 Overview
Use this procedure if you cannot easily run a tape measure between the focal spot decal and the
film plane, or if there is a chance that the decal may be misplaced.
2.3 Procedure
1. Select one of the HHS kit Plexiglas plates with the 1 cm pattern and slide the plate, pattern
side up, onto the collimator rails.
2. Select the desired SID, auto. mode, and insert a 12” x 14” loaded cassette into the Bucky.
3. Select AEC mode, 60 kV and expose the film. If AEC is not available select about 2 to 5 mAs.
4. The spacing of the pattern image is independent of location on the film or orientation of the
tube axis. Therefore simply make the measurement in the sharpest part of the image, (anode
end). Mark the centers of two crosses at the edge of the film, measure the separation, and
divide by the number of spaces to obtain the unit spacing (S).
5. Use the formula engraved on the pattern to calculate the SID. For an Ultranet S or SA collimator
the formula is:
2.4 Finalization
No finalization steps.
3.2 Overview
If the collimator rotates on its axis it may be quicker to perform this test than the one in 46–017564
HHS Collimator Tests.
3.3 Procedure
1. Insert a loaded 14” x 17” cassette into the Bucky or cassette holder.
2. Select the desired SID, auto. mode, 60 kV, AEC mode and a low density. If AEC is not available
select 60 kV and about 2 mAs for a 44” SID.
3. Rotate the collimator 45° on its axis and expose the film. Remove the cassette and temporarily
set it aside.
4. Insert a medium cassette and allow the collimator to size down. Now set the collimator to
override. This will allow you to change cassettes without having the collimator re-size itself.
5. Re-insert the original loaded cassette, and make a second exposure on the same film.
7. Insert a small cassette and allow the collimator to size down. As before, set the collimator to
override and replace the small cassette with the original loaded large cassette.
8. Make a third exposure on the same piece of film. With a little practice you will be able to get
three readable images superimposed on one film as shown in the following illustration.
3.4 Finalization
No finalization steps.
4.2 Overview
For this alternative you will use the Plexiglas plate with the .10” grid pattern. We can’t think of a
situation when this procedure would be preferable to the one of 46–017564 HHS Collimator
Tests. Perhaps you just like doing this the hard way. The plate contains a .10” grid pattern. Films
are exposed simultaneously in the Bucky and in the cardboard cassette. The pattern on the two
films is used to determine the magnification ratio. Magnification is then used to determine the size
of subsequent exposures using film in the cardboard cassette and empty cassettes in the Bucky.
4.3 Procedure
1. Insert a 10” x 12” loaded cassette into the Bucky, and slide the Plexiglas plate, onto the
collimator rails.
2. Select the desired SID, auto. mode, 60 kV, and AEC mode. If AEC is not available select 60
kV and about 5 mAs for a 44” SID.
3. Load a 5” x 7” film into the cardboard cassette and attach this to the Plexiglas plate via the
strip magnets on each piece. If 5” x 7” film is not available, use the lead backed 10” x 12” vinyl
cassette from the HHS kit and attach this to the Plexiglas plate with tape or clips.
5. Determine the magnification by comparing the size of one large square on the Bucky film to
one large square on the collimator film.
NOTE: The magnification is not the pattern size on the Bucky film divided by the actual
size of the pattern (1.00”). Magnification is always determined by comparing two
film images.
6. Insert small and large empty cassettes into the Bucky and expose additional collimator films.
You may be able to offset the collimator film to capture two exposures on one film.
7. Multiply the dimensions of the images on the collimator films by the magnification ratio to
determine the size of the X–ray field at the Bucky film plane.
4.4 Finalization
No finalization steps.
5.3 Procedure
1. Beam angulation is irrelevant, set for convenience. Use common SID.
2. Insert a 12” x 14” loaded cassette into the bucky or cassette holder and center it.
3. Select manual collimator mode. On some systems (e.g. remote tables) this may require first
selecting AUTO, then one on one format, and then going into override or OBL mode.
4. Place the light to X–ray test pattern on the table (or tape to wall unit).
5. Using the collimator dials, adjust the light field to a 10” square as indicated on the test pattern.
Center the pattern as necessary.
6. Select 60 kV and AEC mode; if no AEC use 2 to 5 mAs. Expose the film.
7. Refer to the following illustration. Following the instructions above, the light field readings
should all be 5.00”. Determine the difference between X–ray and light fields by reading the X–
ray field numbers directly off the image using the scales printed on the image.
NOTE: The fact that the test pattern and cassette are not in contact has no bearing on
this measurement. Since the pattern is in inches, you can use the image of the
scales to measure the misalignment regardless of any magnification. You get
the same relative image if the cassette is in the table or under it. If coins or Allen
wrenches were used for this test, there may be some concern about
magnification.
5.4 Finalization
No finalization steps.
The table top output is not to exceed 10 R/min as set with an instrument that has an accuracy of
+/- 5%. The meter accuracy for this reading would be 10 R/min +/- 0.5 R/min. Therefore, set the
maximum table top exposure rate to 9.5 R/min on the meter so that the actual table top exposure
rate is 9.5 R/min +/- 0.5 R/min and the maximum output will not exceed 10 R/min.
Application: Fluoro and Cine set up, Entrance exposure rates, Half-value layer measurements and
cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Accuracy: + 5%, including probe energy dependency (above 50 kVp). Refer to Probe Saturation
Effects.
In addition, the Model 2025/2026 is used with the Model 20X5–6M mammography probe for
measuring exposure rates and exposures from 0.001 R/min to 1000 R/min and 0.1 mR to 200 R,
full scale.
Application: Fluoro and Cine set up, Entrance exposure rates, Half-value layer measurements and
Radiographic exposures. (Must NOT be used for leakage and scatter measurements).
Accuracy: +/– 5%, including probe energy dependency (above 50 kVp). Refer to Probe Saturation
Effects.
Application: Leakage and scatter radiation measurements. (Must be used for Primary Barrier
Transmission Tests and for under–table tube leakage radiation tests.)
Accuracy: within +/- 10% of full–scale indication exclusive of energy response. Refer to Probe
Saturation Effects.
Application: Leakage and scatter radiation measurements. (Must be used for Primary Barrier
Transmission Tests and for under–table tube leakage radiation tests.)
1.5 Unfors Xi
This digital meter is equipped with single solid state probe for measuring exposure rates and
exposures from 140mR/min-7000R/min )(mode ”high”)and (70µR/min-7R/min)(mode ”low”); full
scale. and exposures from 1mR-9999R )(mode ”high”)and (1µR-9999R)(mode ”low”); full scale.
Application: Fluoro and Record set up, Entrance exposure rates, Half-value layer measurements
and cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Application: Fluoro and Record set up, Entrance exposure rates, Half-value layer measurements
and cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Accuracy: + 5%, including probe energy dependency (above 50 kVp). Refer to Section 3,Probe
Saturation Effects
Application: Fluoro and Record set up,, Entrance exposure rates, Half-value layer measurements
and cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Accuracy: + 5%, including probe energy dependency (above 50 kVp). Refer to Section 3,Probe
Saturation Effects.
Application: Fluoro and Record set up, Entrance exposure rates, Half-value layer measurements
and cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Application: Fluoro and Record set up,, Entrance exposure rates, Half-value layer measurements
and cadiographic exposures (Must NOT be used for leakage and scatter measurements).
Accuracy: + 5%, including probe energy dependency (above 50 kVp). Refer to Section 3,Probe
Saturation Effects.
2.2 Overview
There are five preliminary checks and setup procedures that must be followed prior to using a
meter to measure radiation.
2.3 Procedure
1. Check to see that meter bears a calibration sticker and that calibration is not past due.
2. Connect up all cables and set probe in proper orientation for measurement.
3. On the meter:
4. Verify proper operation of the meter using the calibration check source (if supplied with meter).
NOTE: However, for the Victoreen Model 470 A meter, you MUST use the meter with
the cap in place on the front of the meter, AND with the disk shaped cover affixed
to the end of the meter cap.
NOTE: However, for the Victoreen Model 451B meter, you MUST use the meter with the
slide open, to see the circular metallic entrance window of the survey meter
chamber. Otherwise low kV radiation will be strongly attenuated.
The meter is now ready to measure radiation. See 46–017563 HHS Fluoro System Tests
for equipment set-up.
2.4 Finalization
No finalization steps.
NOTE: A radiation meter may give no indication of probe saturation other than an incorrect
reading.
For single exposures measured on an integrating meter, the exposure rate is simply the R reading
divided by the exposure time.
Example: During beam quality tests, a 100 mR exposure is recorded; the exposure time is 20 ms.
What is the exposure rate?
Note that for the example, the Capintec 500 ml and 30 ml probes and the MDH 180 ml probe are
all inadequate and will provide incorrect (low) readings. Whenever the intensity limit of a probe is
exceeded, the probe will read low.
For repeated pulse exposures made on an integrating meter, the peak exposure rate is the R
reading divided by the product of the number of exposures and the pulse width, i.e.:
For repeated pulse exposures made on an exposure rate meter, the peak exposure rate is the rate
meter reading divided by the X-ray duty cycle (where duty cycle is pulse width x frame rate), i.e.:
NOTICE
Avoid condensation on radiation meters.
Meters subjected to condensation may require hours (in some cases days) to dry
out and stabilize. If a meter will be exposed to extreme cold, store the meter, probes,
and cables in air-tight plastic bags and don’t open the bags unless the contents are
at room temperature. Failure to observe this caution could result in excessively
noisy or inaccurate readings.
The instruments must be returned to the manufacturer for calibration. In addition, any repair of
radiation instruments must be done through the manufacturer, since the instruments must be re-
calibrated after repair.
Radiation meters must carry a calibration sticker, indicating the dates of the last and next
recalibration.
Application: Fluoro and Record set up, Reproducibility, Accuracy of kVp, Half–value layer
measurements
Accuracy: + 2%.
• Reproducibility: ±1%
Application: Fluoro and Record set up , Reproducibility , Accuracy of kVp, Half–value layer
measurements
Accuracy: + 2%.
Application: Fluoro and Record set up , Reproducibility , Accuracy of kVp, Half–value layer
measurements
Accuracy: + 2%.
Application: Fluoro and Record set up , Reproducibility , Accuracy of kVp, Half–value layer
measurements
• Reproducibility: ±0.2kV
Application: Fluoro and Record set up , Reproducibility , Accuracy of kVp, Half–value layer
measurements
Accuracy:±1% or ±1kV
1.2 Overview
Excel 5 spreadsheets replace the traditional hard copy HHS data records. Spreadsheets are
available for form 3382 (basic xr, vas., and parts replacement), 3382MM (mammo.) and 3382IN
for the Innova systems and Discovery IGS 730. For X-Ray tube or other certified part replacements,
use spreadsheet 3382. Spreadsheets are in the XRTOOLS window, along with a README file
which contains these instructions.
1.3 Procedure
1. Double click XRTOOLS window.
2. Double click desired spreadsheet. If using 3382, open and print README file; it contains a
convenient test pattern for use in performing overhead collimator light and X–Ray field tests.
3. After reading Intro sheet of the spreadsheet and familiarizing yourself with the Help sheet, click
on data entry sheet (click on the tab at bottom of screen that says [data entry]).
4. Click [Start] button near top of data entry sheet and follow screen instructions for entering basic
customer information. Use [Tab] key to skip over non–data cells. Be sure to click [done] button
after entering basic customer information (this action creates new customer file).
5. Follow screen instructions and enter test equipment information. When finished, click [Room
Equipment] button below test equipment chart.
6. Select room equipment, using any character to select. Never use backspace key to erase a
wrong entry, – use delete key or backspace and enter keys. Be sure to click [done] button after
selecting all room equipment. If this is not done then proper mA/kV accuracy chart will not be
set up.
7. When room equipment has been indicated and associated [done] button is clicked, program
will flag appropriate tests. User will be presented with an index sheet from where he/she can
select each required test.
8. Fill in test data for each test that has a req. flag.
9. Save your work often by clicking Save Customer File which is found in main menu item Cookie
Jar.
10. When finished with all tests you should e-mail your data to your Zone Compliance Engineer.
Select UPLOAD DATA from main menu item Cookie Jar. Follow screen instructions. Macro
routine will create a small text file with data which can be attached to e-mail.
11. If a printout is desired, you may use print options in cookie jar, or you may print an individual
test by using [Print] button found near the top of each test section.
NOTE: Pass/fail flags will appear for each test when required data has been entered.
For some tests, a yes/no or n/a selection MUST be made in order to set p/f flag.
Some cells contain default values, such as 60 kV in the technique accuracy
charts. Note that the “kV” part of the entry is actually part of the cell format. To
change the default value, simple type in a new number, e.g. 70, without the “kV”.
Similar examples of unit formats are found throughout spreadsheets. You should
always type only the number to change default value.
1.4 Finalization
No finalization steps.
www.gehealthcare.com