Professional Documents
Culture Documents
5433542-1EN Rev 5
5433542-1EN Rev 5
OPERATING DOCUMENTATION
5433542-1EN
Revision 5
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
2
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
3
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
This page left intentionally blank.
4
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Important Information
LANGUAGE
ПРЕДУПРЕЖДЕНИЕ Това упътване за работа е налично само на английски език.
(BG) • Ако доставчикът на услугата на клиента изиска друг език, задължение на
клиента е да осигури превод.
警告 本维修手册仅提供英文版本。
(ZH-CN) • 如果客户的维修服务人员需要非英文版本,则客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修服务人员、操作人员或患者造成电击、机械伤害或其他形式的
伤 害。
警告 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其他的危
險。
警告 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危險而受
傷。
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge, operatera ili
pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.
Important Information 5
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a
pochopili jeho obsah.
• Do not attempt to service the equipment unless this service manual has been con‐
sulted and is understood.
• Failure to heed this warning may result in injury to the service provider, operator or
patient from electric shock, mechanical or other hazards.
6 Important Information
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt tämän
huolto-ohjeen.
• Ne pas tenter d'intervenir sur les équipements tant que ce manuel d’installation et
de maintenance n'a pas été consulté et compris.
• Versuchen Sie nicht diese Anlage zu warten, ohne diese Serviceanleitung gelesen
und verstanden zu haben.
• Μην επιχειρήσετε την εκτέλεση εργασιών σέρβις στον εξοπλισμό αν δεν έχετε
συμβουλευτεί και κατανοήσει το παρόν εγχειρίδιο σέρβις.
Important Information 7
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Reynið ekki að afgreiða tækið nema að þessi þjónustuhandbók hefur verið skoðuð
og skilin.
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda
eða sjúklings frá raflosti, vélrænu eða öðrum áhættum.
• Il mancato rispetto della presente avvertenza potrebbe causare lesioni all'addetto al‐
la manutenzione, all'operatore o ai pazienti provocate da scosse elettriche, urti mec‐
canici o altri rischi.
警告 このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者 さんが、
感電や機械的又はその他の危険により負傷する可能性があります。
8 Important Information
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Jei nepaisysite šio įspėjimo, galimi paslaugų tiekėjo, operatoriaus ar paciento suža‐
lojimai dėl elektros šoko, mechaninių ar kitų pavojų.
• Manglende hensyn til denne advarselen kan føre til at serviceleverandøren, oper‐
atøren eller pasienten skades på grunn av elektrisk støt, mekaniske eller andre
farer.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual
de assistência técnica.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual
de assistência técnica.
Important Information 9
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Ne pokušavajte da opravite uređaj ako niste pročitali i razumeli ovo servisno uputst‐
vo.
• La no observancia del presente aviso puede dar lugar a que el proveedor de servi‐
cios, el operador o el paciente sufran lesiones provocadas por causas eléctricas,
mecánicas o de otra naturaleza.
10 Important Information
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Försök inte utföra service på utrustningen om du inte har läst och förstår den här
servicehandboken.
• Om du inte tar hänsyn till den här varningen kan det resultera i skador på servicete‐
knikern, operatören eller patienten till följd av elektriska stötar, mekaniska faror eller
andra faror.
Important Information 11
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
This page left intentionally blank.
12 Important Information
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Revision History
Revision Date Reason for Change
5 4-May-2017 Updated for PARC4 (Q.Core Power) upgrade option. Chapter 1: Updated Table 1-2 to add
PARC4 (Q.Core Power) size and weight dimensions and revise dolly width and length dimen‐
sions per Engineering. Added section 4.2.2.6 for PARC4 (Q.Core Power) shipping methods. Up‐
dated section 4.4.4 for unpacking PARC4 (Q.Core Power). Chapter 2: Updated Table 2-1 to add
PARC4 (Q.Core Power) system component weight load. Updated Table 2-2 for PARC4 (Q.Core
Power) system component dimensions. Added Illustration 2-4 and table below for PARC4
(Q.Core Power) dimensions. Added Illustration 2-13 and table below for PARC 4 center-of-gravi‐
ty. Updated table below Illustration 2-16 typical scan suite layout for PARC4 (Q.Core Power).
Added section 3.8 for PARC4 (Q.Core Power) placement requirements. Updated section 4.1.1
for PARC4 (Q.Core Power) non-seismic anchoring statement. Chapter 4: Updated section 1.1.2
for Q.Core/PARC4 (Q.Core Power). Renamed section 1.2 from “System Cooling Requirements”
to “Heat Output” to match other products. Added heat output text for PARC4 (Q.Core Power).
Updated Table 4-3 with PARC4 (Q.Core Power) heat output. Updated Table 4-3 with Q.Core
(Recon Cabinet) heat output.Chapter 5: Updated sections 1.4.4 and 1.9 for Continuous (average)
power demand at maximum duty cycle value. Updated table below Illustration 5-2 system ground
map for PARC4 (Q.Core Power). Updated Table 5-8 for PARC4 (Q.Core Power) component des‐
ignators. Updated text in section 3.1 for PARC4 (Q.Core Power) upgrade cable kits. Added Table
5-17 and 5-18 for PARC4 (Q.Core Power) upgrade cable kits.
4 9-Oct-2014 HCSDM00314963: Chapter 5: Updated section 3.1 for new cable catalogs P5051TE, P5051TF,
P5064TE, and P5064TF.
3 24-May-2013 HCSDM00196987: Chapter 1: Updated section 4.1.1. Table 1-2: Corrected width for PET Image
Ring with dollies and PET Source Ring/Trailer with dollies. Updated Illus 1-3 and 1-4. Chapter 2:
Corrected weights for Console, Monitor, and Workspace Table in Table 2-1. Corrected height for
Q.Core and Operator Console in Table 2-2, Illus 2-3, and Illus 2-11. Corrected weights for Illus
2-4, and Illus 2-6 through 2-9.
2 26-Mar-2013 Front/back cover updates. HCSDM00179998: Chapter 1, section 2: Corrected “Floor Specifica‐
tion” minimum concrete thickness; Table 1-2: Weights corrected for CT Gantry (16 slice) and PET
Image Ring. Chapter 2, Table 2-1: Weights corrected for CT Gantry (16 slice), Patient Table, and
PDU; section 3: Removed VQC phantom from Tables 2-5 and 2-6. Chapter 3: Added section
1.2.6 – PET Alignment (VQC) Phantom. Chapter 5, Tables 2-9 and 2-10: Corrected cable length
for Run # 56. Entire manual: Moved descriptive text in illustrations to tables.
Revision History 13
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
This page left intentionally blank.
14 Revision History
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Table of Contents
Chapter 1 GENERAL REQUIREMENTS............................................................................................... 23
1 Introduction....................................................................................................................................23
1.1 Objective and Scope of this Manual..................................................................................... 23
1.2 Responsibility of the Customer.............................................................................................23
1.3 Site Project Coordinator....................................................................................................... 24
2 System Siting Requirements......................................................................................................... 25
2.1 Project Manager (PM) Tasks................................................................................................25
2.2 Customer Requirements for Site Readiness........................................................................ 25
3 Regulatory Requirements..............................................................................................................30
3.1 Building Codes, Regulations and Permits............................................................................ 30
3.2 Clearance Regulations......................................................................................................... 30
3.3 Codes, Clearances, and Service Space Regulation.............................................................30
4 Delivery and Handling................................................................................................................... 31
4.1 Project Manager Tasks.........................................................................................................31
4.1.1 Minimum Doorway Opening........................................................................................ 31
4.1.2 Minimum Hallway Width.............................................................................................. 31
4.2 Shipping Dimensions and Weight.........................................................................................31
4.2.1 Delivery Sizes and Weights......................................................................................... 31
4.2.2 Shipping Methods (Dollies, Skids)............................................................................... 32
4.2.2.1 CT Gantry........................................................................................................... 32
4.2.2.2 PET Components............................................................................................... 33
4.2.2.3 Patient Table.......................................................................................................34
4.2.2.4 PDU.................................................................................................................... 35
4.2.2.5 Q.Core................................................................................................................ 35
4.2.2.6 PARC4 (Q.Core Power) Reconstruction Cabinet............................................... 35
4.2.2.7 Operator Console............................................................................................... 36
4.3 System Lifting and Rigging Restrictions............................................................................... 36
4.4 Shipping and Receiving........................................................................................................36
4.4.1 Handling Restrictions...................................................................................................36
Table of Contents 15
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
4.4.2 Dollies.......................................................................................................................... 37
4.4.3 Delivery Temperature and Humidity Tolerance........................................................... 37
4.4.4 Unpacking System.......................................................................................................37
4.5 Storage................................................................................................................................. 38
16 Table of Contents
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Table of Contents 17
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
18 Table of Contents
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Table of Contents 19
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
20 Table of Contents
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Table of Contents 21
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
This page left intentionally blank.
22 Table of Contents
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
This manual guides you through the pre-installation siting and regulatory requirements. Keep in
mind, this manual cannot address or answer each and every site specific question or concern.
Contact your GE Healthcare Project Manager (PM) for answers to any additional questions or
concerns not addressed in this manual. Prior to any construction or approval, General Electric
Headquarters Architectural Planning must review all PET/CT preliminary concepts, site plans,
and final working drawings associated with the installation of the system. Contact your GE PM
or complete information regarding your site-specific room layout.
1.2 Responsibility of the Customer
It is the responsibility of the customer (buyer/purchaser) to prepare the site in accordance with
all the specifications provided in this manual and in conjunction with site-specific drawings and
applicable regulations. It is essential to verify all aspects of the site configuration before
construction has begun, as subsequent changes can be costly or impractical. A detailed pre-
installation checklist is provided in this manual. It is the responsibility of the customer to ensure
all requirements on the checklist are fulfilled and that the site conforms to all specifications and
requirements detailed in this manual.
Pre-Installation requirements shall include the procurement and installation of all required
materials and services necessary to prepare the room to be ready for installation of the PET/CT
system. The customer is responsible for all aspects of site preparation, including:
• Assigning a project coordinator.
• Planning and construction requirements for the installation of the PET/CT system in
accordance with all national, state, or local regulatory requirements for the country in which
the installation occurs, for example:
• All design work associated with preparing the installation site for the PET/CT system and all
architectural, mechanical, and electrical drawings associated with the design of the site.
• All alterations or modifications to products not specifically included in the sales contract.
• A clean and safe work environment for installation of the PET/CT system.
• A location with proper lighting, a level finished floor, finished walls, and a finished ceiling.
• A support structure in the floor, walls, and ceiling suitable for mounting all system
components as specified in the site design.
• Installation of all required conduit, ducts, and raceways to safely route all cables.
• Supplying electrical power of the required voltage, all necessary power supply cables and
grounds, all necessary power cables and grounds to the PDU and system cabinet, and an
Emergency-Off switch in the scan room.
• Installation of all properly-sized junction boxes, outlets with covers, line safety switches, and
fittings installed at the locations specified in the site design.
○ The electrical contractor shall ring out and tag all wires at both ends.
○ Wires shall be continuous and without splices.
○ Ground wires shall conform to product requirements.
○ Color-coded wires shall be used whenever possible, to enable easier identification.
• All work shall conform to IBC (International Building Code) and local building and safety
codes.
NOTE: GE Healthcare does not provide or install the wires, conduits, junction boxes, or
ducting illustrated in this publication, unless specifically stated.
1.3 Site Project Coordinator
The site project coordinator is the primary contact and liaison between GE Healthcare and all
site related functions, between the purchaser, the construction planners, architects, contractors,
and any other site administrative personnel.
To ensure a successful installation, it is recommended that a single/individual site project
coordinator manage the entire project. Ideally, the project coordinator is a person familiar with
all phases of pre-installation and installation of similar medical device construction projects, from
conceptual planning through to system start up. The site project coordinator shall be responsible
for working closely with GE Healthcare to ensure the client (buyer/purchaser) upholds all
requirements in this manual.
24 1 Introduction
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Floor Specification
The floor shall have a minimum concrete thickness of 127 mm (5 in.).
The floor shall be no greater than 6 mm (0.250 in.) out of level over a 3048 mm (10 ft.)
range, with level defined as the horizontal surface between the highest and lowest points.
NOTE: If the concrete floor has a floor covering installed over it (such as floor tile), 17 or
more openings 101.6 mm (4 in.) in diameter will be cut into the floor covering to
ensure the table and gantry rest on the concrete. (Openings are cut during
installation.)
Shims shall not be used to level the gantry or patient table.
To ensure timely system delivery and installation, the customer shall complete all necessary
work listed in this Pre-Installation Manual and schedule a site-ready Project Manager (PM)
visit prior to system delivery.
• Pre-installation Checklist
The customer shall also verify site readiness by filling out and signing the following Pre-
Installation Checklist. The checklist shall be completed six weeks prior to scheduled delivery
date.
Hospital Name:
AW
Camera
PACS
Other ___________________________
Other ___________________________
Commitment Dates:
Weekdays
Weekend
Quick Install
Equipment Compatibility:
Remote Monitors
AW Relocation
Cardiac Option
Injectors
EKG Monitor
Injector Control
Laser camera
UPS
2nd Monitor
Respiratory Gating
Will the existing network, broadband, and camera ca‐ List any issues or concerns:
ble drops reach all required locations for the PET/CT
scanner?
Network Installation:
DICOM Print
AW
GoldSeal
3 Regulatory Requirements
The following codes, standards, and laws are referenced in this section:
• OSHA 29 CFR 1910
30 3 Regulatory Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Site Dimensions PM shall measure and verify all site dimensions to ensure the facility can ac‐
commodate the delivery of the system (and any related components or
equipment), from the delivery drop-off point to the scan suite.
NOTE: The minimum unobstructed doorway opening to accommodate
delivery of the system is 1067 mm (42 in.). The minimum un‐
obstructed hallway width to accommodate delivery of the sys‐
tem is 1803 mm (71 in.).
Delivery Type PM shall determine type of delivery: ground level, loading dock, or tilt-bed
truck.
Delivery Equipment PM shall determine if delivery requires special dollies, lifting crates, or rig‐
gers. PM shall order any additional delivery equipment and all necessary de‐
livery personnel.
NOTE: The CT gantry cannot be lifted or transported by any means
other than the GE support cradle and dolly system on which is
was shipped. Otherwise, serious damage to the gantry could
result. The CT gantry is not designed to be lifted by special rig‐
ging.
Identify Delivery Route PM shall identify the delivery route, which may include any elevators, door‐
ways, and hallways necessary to accommodate the delivery of all system
components.
Non-Construction-Zone Route to Scan Suite PM shall verify an accessible, dust-free, non-construction-zone delivery
route to the scan suite.
Packaging Requirements PM shall order any construction site packaging requirements prior to ship‐
ment. Packaging cannot be modified once the system is shipped.
Floor Protection PM shall determine if floor protection is required along facility delivery route
and communicates requirement to delivery company/personnel.
CT Gantry (64 Slice) with Dollies On, Side Rails On 1955 (77) 1295 (51) 2438.4 (96) 1932 (4260)
CT Gantry (64 Slice) with Dollies On, Side Rails Off 1955 (77) 970 (38.25) 2438.4 (96) 1914 (4220)
CT Gantry (16 slice) with Dollies On, Side Rails On 2000 (79) 1290 (51) 2810 (111) 1982 (4360)
CT Gantry (16 slice) with Dollies On, Side Rails Off 2000 (79) 1067 (42) 2810 (111) 1967 (4310)
PET Base and Retractor Assembly with Dollies 990 (39) 1054 (41.5) 2286 (90) 678 (1495)
PET Image Ring with Dollies On, Side Protective 1880 (74) 1041.4 (41) 2667 (105) 1282 (2820)
Braces On
PET Image Ring with Dollies On, Side Protective 1880 (74) 985.52 (38.8) 2667 (105) 1282 (2820)
Braces Off
PET Image Ring without Dollies 1819 (72) 720 (28) 1931 (76) 954 (2110)
PET Image Ring Dolly (assembled) — 1041.4 (41) 2667 (105) 327 (720)
PET Source Ring and Trailer with Dollies 1092 (43) 1041.4 (41) 2387.6 (94) 608 (1340)
Table (Blue Dollies On) 1410 (55.5) 864 (34) 3836 (151) 1241 (2736)
Table (Blue Dollies Off, Red Castors On) 1410 (55.5) 1016 (40) 3048 (120) 1295 (2856)
Table (Tilting Dollies On) 1778-2032 965 (38) 2489-2921 1147 (2530)
(70-80) (98-115)
Power Distribution Unit (with cardboard packaging) 1092 (43) 584 (23) 762 (30) 413 (910)
Q.Core (on skid) 1067 (42) 635 (25) 864 (34) 87 (192)
PARC4 (Q.Core Power) Reconstruction Cabinet (on 1655 (65.2) 1480 (58.3) 980 (38.6) 304 (670)
skid)
Console (on skid) 1067 (42) 635 (25) 864 (34) 87 (192)
Illustration 1-2: PET Base and Retractor Assembly, with Shipping Dollies
Illustration 1-3: PET Gantry Image Ring, with Shipping Dollies and Side Rails
Illustration 1-4: PET Source Ring and Trailer, with Shipping Dollies and Side Rails
Red caster towers ship attached to the ends of the dollies (see Illustration 1-5). They are used
for fitting the Table in an elevator and for final positioning of the Table in front of the Gantry (see
Illustration 1-6).
NOTE: The patient table ships to international sites in a crate. The installation team uncrates
the table and attaches the dollies at the site.
Illustration 1-5: Patient Table with Shipping Dollies
4.2.2.4 PDU
The PDU is shipped on a skid. Do not remove the PDU from the skid until it is in the room ready
for installation.
4.2.2.5 Q.Core
The Q.Core is shipped on a skid. Do not remove the Q.Core from the skid until it is in the room
ready for installation.
4.2.2.6 PARC4 (Q.Core Power) Reconstruction Cabinet
The PARC4 (Q.Core Power) is shipped on a skid. Do not remove the PARC4 (Q.Core Power)
from the skid until it is in the room ready for installation.
• Shock Restrictions: The system cannot tolerate shock or vibration. System components
cannot be tipped, dropped, or hoisted. The PM shall communicate these restrictions to
everyone involved with handling the system components.
Floor Protection
To protect the floor during delivery, floor protection shall be used along the entire delivery path
and throughout the scan suite, where necessary.
4.4.2 Dollies
• U.S. Installations – Shipments within the United States typically involve the use of dollies for
moving the gantry and table. After completing the installation, return all dollies and the
gantry shipping cage to GE using the shipping document found in Box #1.
• Zero Clearance Dollies (Mini) (For CT Gantrys only) – Deliveries involving small elevators
with a depth of at least 2692 mm (106 in.) require zero clearance dollies. Zero clearance
dollies allow movement of the gantry in tight areas; avoid using them for normal dock or van
deliveries. To order zero clearance dollies, go to: http://www.umi-dollyshop.com.
• Tilting Dollies (for Patient Table) – Deliveries involving small elevators with a depth of at
least 2438 mm (96 in.) require tilting dollies. If storing the patient table prior to installation,
do not order tilting dollies as there is a limited number of dollies available. If you are unable
to obtain tilting dollies, substitute riggers in place of the dollies to deliver the table. To order
tilting dollies for the patient table, go to: http://www.umi-dollyshop.com.
• Installations Outside U.S. – For shipments outside the United States, customers may
purchase dollies at: http://www.umi-dollyshop.com. DO NOT return dollies or the gantry
shipping cage to the U.S. Instead, forward dollies and cage to the local GE office or
warehouse.
4.4.3 Delivery Temperature and Humidity Tolerance
NOTICE
Failure to adhere to temperature requirements during delivery and storage will
likely result in equipment damage.
After delivery to the scan suite and before unpacking any system components, allow 12 hours
for the equipment to adjust to room temperature.
4.4.4 Unpacking System
Do not remove any protective wrapper or packaging from any system component until all
construction is complete and all construction dust is removed from the installation site.
Retain the packaging surrounding the scanner desktop and UPS.
Do not remove the Console or Q.Core/PARC4 (Q.Core Power) from their shipping skids until
after they have been delivered to the CT equipment room.
4.5 Storage
If storing a system prior to installation, the system shall be stored in its original packaging in a
temperature and humidity controlled environment protected from water and dust. It is advised
that storage of the system be no longer than six months. If storage is going to exceed six
months, contact your PM for long-term storage procedures.
Table 1-3: Humidity and Ambient Temperatures for Storage*
Ambient temperature shall be maintained within a range of: +4 to +27° C (+40 to +80° F)
Maximum rate of change in the temperature shall be no greater than: 3° C per hour (+37° F)
Maximum rate of change in the relative humidity shall be no greater than: 5% RH per hour
NOTICE
Storage exceeding six months is not advised.
System Compo‐ Net Weight Maximum Maximum Load Pattern mm (in.) Normal Method of Mounting mm
nent kg (lb) Uplift Load N Compressive (in.) (GE-supplied1)
(lb) load N (lb)
CT Gantry 1810 (3990) 0 4588 (1031) CT effective load area is 700 x Hilti Kwik-Bolt II 12.7 mm (1/2
(64 slice) 2013 (27.6 x 79.25) with four in.) diameter x 254 mm (10 in.)
round pads, each 63.5 (2.5) in long per P/N 2106573-2 at four
contact with the floor. leveling pads into concrete
floor.
CT Gantry 1682 (3700) 0 4895 (1100) Rectangular base plate 700 x Hilti Kwik-Bolt II 12.7 mm (1/2
(16 slice) 1966 (28 x 77) with four round in.) diameter by 203 mm (8 in.)
pads, each 63.5 (2.5) in contact long per P/N 2106573 at four
with floor. Individual pad load‐ leveling pads into the concrete
ings are 910 lb., 960 lb., 1040 floor.
lb., and 1090 lb.
PET Gantry 1968 (4339) 0 6101 (1250) While in the imaging position, Hilti Kwik-Bolt II 12.7mm (1/2
the effective PET load area is in.) diameter by 8 in. (203mm)
398 x 645 (15.7 x 25.4) with 7 long per P/N 2106573 at seven
pads each 63.5 (2.5) as well as leveling pads into concrete
2 pads that do not get anchored floor.
(support only)
Patient Table 1049 (2308) 890 (200) 4926 (1107) Rectangular base 550 x 2134 Hilti Kwik-Bolt II 12.7mm (1/2
Includes (21.7 x 84.0) with 6 round pads, in.) diameter per 8 in. (203 mm)
227 (500) each 63.5 (2.5) in contact with long per P/N 2106573 at four
Patient the floor. leveling pads into concrete
floor.
Power Distribu‐ 370 (813) 0 1070 (240) Four Casters support area of Casters are for positioning and
tion Unit (PDU) 711 x 559 (28 x 22). service. See Note 2.
Q.Core 87 (192) 0
PARC4 (Q.Core 246 (540) 0 737 (166) Rectangular base with four cas‐ Casters are for positioning and
Power) tors. service. See Note 2.
Optional Components
Universal Power 281 (619) 0 689 (155) Rectangular base 305 x 813 Casters are for positioning. Set
Supply (UPS) (22 x 32) with four castors, on floor. Adjust the six leveling
each in contact with the floor. pads on the floor.
Notes:
1.) Use the GE-supplied mounting hardware only if anchoring the system to 127 mm (5 in.) concrete floors.
2.) Seismic angle brackets are included and shipped with the PDU and PARC4 (Q.Core Power).
System Component A B C
PET-CT Gantry (overall) without Trailer 2235 (88) 1473 (58) 1930 (76)
Table (at max elevation; 1" [25 mm] below Gantry ISO 660 (26) 3454 (136) 1067 (42)
center)
Power Distribution Unit (PDU) 711 (28) 559 (22) 1067 (42)
PARC4 (Q.Core Power) Reconstruction Cabinet 616 (24.3) 1257 (49.5) 1422 (56)
Console Workspace Table (adjustable height) 1350 (53) 741 (29) 683-912 (27-36)
40 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
4 Front
4 Clearance (minimum)
42 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
1 76 kg (167 lb)
44 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
3 Minimum thickness
1 Center of gravity weight = 1049 kg (2308 lb); includes 227 kg (500 lb) patient
2 Minimum thickness
1 Center of gravity weight = 1049 kg (2308 lb); includes 227 kg (500 lb) patient. Y = 21.3”
46 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
4 Clearance (minimum)
4 Clearance (minimum)
48 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
B Side View
3 Minimum thickness
3 L 3" x 3" x 1/4" x 14" bracket mounted to cabinet frame with 4 - #12 S.M. screws (each
side)
50 1 System Components
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
2 Room Layout
2.1 Scan Suite Configuration
A scan suite, which includes a control room and a scan room, requires a minimum room size to
safely support all PET/CT service activities. An example of a typical system configuration is
detailed in Illustration 2-16.
Illustration 2-16: Typical Scan Suite Layout Configuration
The enclosure (Patient Touch) Leakage Envelope, as detailed in Illustration 2-17, defines a
zone in the Scan Room only, where the enclosure leakage must be tested. Areas that fall
outside of this envelope DO NOT need to be tested. The intent of this graphic is to provide the
PM with a view of potential electrical devices, plumbing fixtures, hospital gas outlets, and metal
surfaces that may fall within this scan room envelope, which may require additional grounding
prior to customer turnover. [Height of envelope from floor-to-ceiling: IEC-2 = 2286.0 mm (7.5
Ft.), IEC-3 = 1829.0 mm (6.0 Ft.). UL60601-1 (2.12.20 DV Addition) and GE Healthcare
requirement.]
NOTE: The enclosure leakage envelope has nothing to do with Regulatory Work Space
Clearance or Safe Egress requirements for Service Personnel (NFPA 70E).
1 = PDU, A = Patient Care Perimeter Envelope, B = Equipment Perimeter, C = Width and Height
— Width Height
52 2 Room Layout
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Working Space: Work space for equipment operating at 600 Volts, nominal, or less, to ground,
and likely to require examination, adjustment, servicing, or maintenance while energized. Refer
to the conditions in Table 2-3 and Table 2-4.
IEC/NFPA 70e (Table 110.26) 2011 Edition GE Healthcare requires the following minimum work
space requirements for the safe servicing of the product.
Terms Defined for “Work Space Conditions”
• Grounded Surface/Wall: Made of concrete, masonry, brick, ceramic tile, or a wall that
contains surface mounted electrical boxes, conduits, or ducting.
• Ungrounded Surface/Wall: Made of wood or other insulated construction material that will
not create a path to ground when touched.
• Obstructions: Surface mounted floor ducts or other trip hazards, walls, pilasters, support
columns, and equipment covers stored temporarily that would block direct access to an exit
from the room.
• Head Clearance: Head clearance represents the height dimension of the work space, as
measured from the floor directly in front of the equipment to the ceiling (or overhead
obstruction). It requires a minimum of 1981 mm (78.0 in.), or the height of the equipment,
whichever is greater.
• Powered On Service – Work Space Egress - 712.0 mm (28.0 in.): Any work space around
the perimeter of the system or subsystem, shall have at least one unobstructed route to a
direct exit of the room. The width of the exit route shall not be less than 712.0 mm (28.0 in.)
along the entire length of the route. This emergency egress route must be free of
obstructions and trip hazards, including equipment covers that may have been removed for
service.
• Small Room – (Not Recommended): A condition of installation where the gantry may be
placed a minimum of 356.0 mm (14.0 in.) from a wall where access to electrical power or the
wall is not required. (Limited to the side of the gantry, opposite the Tube-Change side of the
gantry.)
If the depth of the working space is directly facing an ungrounded sur‐ 914.0 (36.0)
1 face or wall without live voltage panels (600V or less) and without sur‐
face mounted ducts or conduits.
If the depth of the working space is directly facing a grounded surface 1067.0 (42.0)
Length/Depth 2
or wall.
If the depth of the working space is directly facing a surface or wall with 1219.0 (48.0)
3 live voltage panels (600V or less), grounded surface mounted ducts, or
conduits.
Minimum width of the working space in front of the electrical equipment, 762.0 (30.0)
unless the width of the equipment is larger.
If the equipment is wider than 762.0 mm (30.0 in) the width of the Size of Equipment
Width 4
equipment shall become the width of working space.
Minimum Height of the working space shall be clear and extend from 2000.0 (78.0)
the grade (floor), unless the height of the equipment is higher.
Height 5
If the equipment is taller than 2000.0 mm (78.0 in.), the required height Height of Equipment
of the working space shall become the height of the equipment.
Minimum distance required on the side of the gantry, opposite the Tube-Change side of the gantry. 356.0 (14.0)
1 Effectively insulated
2 Scanner or subsystem
1 CT Gantry
2 Table
3 Clearance radius
Storage Cabinet 46 x 91 x 107 cm (18" D x 36" W x 42" H) 45.3 kg (100 lb) (approximately)
Install Support Kit (box) 30 x 30 x 38 cm (12" x 12" x 15") 9.1 kg (20 lb)
Tube Hoist Kit 77 x 8 cm and 38 x 15 cm (30" x 3" and 15" x 6") 13.6 kg (30 lb)
Table 2-6: Optima 560, Discovery 610 and 710 (16 slice) Storage Cabinet and Equipment
Storage Cabinet 46 x 91 x 107 cm (18" D x 36" W x 42" H) 45.3 kg (100 lb) (approximately)
Install Support Kit (box) 30 x 30 x 38 cm (12" x 12" x 15") 9.1 kg (20 lb)
Tube Hoist Kit 77 x 8 cm and 38 x 15 cm (30" x 3" and 15" x 6") 13.6 kg (30 lb)
4 Anchoring
4.1 Anchoring Requirements – Non-Seismic Installation
WARNING
POTENTIAL FOR PATIENT INJURY!
AN IMPROPERLY SECURED TABLE MAY TIP, DISLODGING THE PATIENT.
PATIENT SAFETY DURING SYSTEM OPERATION REQUIRES PROPER
ANCHORING OF SYSTEM COMPONENTS.
4.1.1 Gantry, Patient Table, and Trailer Anchoring – Non-Seismic
The Gantry, Patient Table, and Trailer shall be securely anchored to the floor (see Illustration
2-21). The scanner desktop, Power Distribution Unit, and system recon cabinet (Q.Core or
PARC4 (Q.Core Power)) do not require anchoring to the floor in a non-seismic installation. Use
the floor template (p/n 5322810) or its dimensions to locate the Gantry, Patient Table, and
Trailer support positions within the scan room, making sure that any anchors that pass through
the supports clear all structural beams and interferences in the floor.
It is the responsibility of the buyer/purchaser of the system to have a licensed structural
engineer work in conjunction with a qualified contractor to use either the GE-supplied floor
anchor hardware or provide an equivalent anchoring system to mount the gantry and patient
table to the floor.
The buyer/purchaser shall consult a licensed architect, licensed structural engineer, qualified
contractor, or the PM to resolve all anchoring issues.
62 4 Anchoring
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
1 Anchor Assembly
3 Floor depth
5 Drill depth
64 4 Anchoring
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Equipment placement
• Materials used for construction of walls, floors, ceiling, doors, and windows
• Equipment in surrounding scan room areas (such as film developer, film storage)
• For small and medium filter survey, the 20 cm water phantom should be placed on the
phantom headholder inserted into the end of the patient table.
The four scatter surveys depict measured radiation levels within the scanning room at the
indicated distances, while scanning a 16cm CTDI phantom for the Head Scan mode and 32cm
CTDI phantom for the Body Scan Mode. The mAs, kV and aperture scaling factors are provided
in Table 3-1 and they can be utilized to adjust the exposure levels to the typical usage at the
site.
For example: The exposure level for a 120kV, 800 mA, 1sec scan at 50" (127 cm) away from
the scan plane is: 10.4 μGy (from Illustration 3-4) × 0.71 (from Table 3-1) × 800/100 (from Table
3-1) = 59.1 μGy.
NOTE: Actual measurements can vary. Expected deviation equals ±15%, except for the 5mA
and 1.25mm techniques, where variation may be greater (up to a factor of 2), due to
the inherent deviation in small values. The maximum deviation anticipated for tube
output equals ±40%. Isocurves already include the deviation margins.
Table 3-1: Shielding Requirements Scaling
40 mm aperture 1.00
Scale units: centimeters (inches). ISO-contour level units: µGy/scan. Technique: 140 kV, 100 mA, 1 sec, 4x5.00 mm
66 1 Radiation Protection
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Illustration 3-2: Typical Scatter Survey (Large Filter) - 16 Slice Optima 560, D610, D710
Scale units: centimeters (inches). ISO-contour level units: µGy/scan. Technique: 140 kV, 100 mA, 1 sec, 4x5.00 mm
Illustration 3-3: Typical Scatter Survey (Small & Medium Filter) - 64 Slice D610, D710
Scale units: centimeters (inches). ISO-contour level units: µGy/scan. Technique: 140 kV, 100 mA, 1 sec, 40 mm aperature
68 1 Radiation Protection
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Illustration 3-4: Typical Scatter Survey (Large Filter) - 64 Slice D610, D710
Scale units: centimeters (inches). ISO-contour level units: µGy/scan. Technique: 140 kV, 100 mA, 1 sec, 40 mm aperature
operational, unstable material may be on the premises. It is very important to recognize that
clear and significant hazards from ionizing radiation may exist at the site, as it is undergoing
preparation. Other equipment may be in place and operational at this time. This may include
such equipment as X-ray systems and CT scanners (other than the CT Gantry within the
PET/CT system). Calibration source may be on the site at some time during the preparation
process, as well as after the PET imager has been put into operation. A cyclotron may be
operational at the site. Definite steps should be taken to ensure the safety of workers, patients,
and visitors, during all phases of the construction, installation and operation of the facility.
NOTE: By the time the site is ready to have radioactive material brought in, the licensing
process must be complete. The site must be properly licensed before receiving
radioactive material.
1.2.1 Protection of Equipment
It is important that background radiation be kept to a minimum. The coincidence detection used
in a PET system allows a moderate amount of external singles events. The PET/CT system has
been found to have less than 1% deadtime if the external field is below 1 mR/hr from a single
source. Because area background can be more general than a single source, a lower limit is
appropriate. If the area dose rate is maintained to less than 0.2 mR/hr (due to 511 or lower
energy gamma rays) at the covers, detector deadtime should not exceed 1%.
Radioactive sources must be stored in approved shielded containers. It is recommended that
any radioactive source not specifically designed to be housed in the gantry’s lead storage
container be stored in a separate room (hot lab) adjacent to, and accessible from, the Scan
Room. This hot lab should be near the cyclotron (if used). Doses should be prepared in the
same area.
Consideration should be given to the placement of the gantry in relation to existing X-ray,
Magnetic Resonance, or Nuclear diagnostic equipment. Magnetic interference above 1.0 gauss,
at the surface of PET components, can adversely affect the image quality. Good shielding
techniques must be implemented in order to avoid this type of interference.
Some procedures involve the use of radioactive water. This will result in the patient exhaling
radioactive carbon dioxide. This carbon dioxide must be contained in order to avoid adversely
affecting the image quality. Some PET procedures require the use of radioactive gases. This too
can result in compromising image quality if not properly controlled.
1.2.2 Protection of Personnel
The escape of radioactive gases, if not properly confined, can cause unnecessary exposure to
clinical staff. All sources must be properly stored in appropriate enclosures to provide adequate
protection to all in the suite.
1.2.3 Barriers, Partitions and Shielding
Appropriate barriers such as walls, lead-shielded glass, lead shields etc. must be installed to
protect staff from unnecessary exposure to radiation. A qualified radiological health physicist
must be consulted in the design of walls and safety barriers to assure proper attenuation.
Keep in mind that patients become significant sources of radioactivity. Consideration should be
given to maximize the distance between the patient and operator during the uptake and
acquisition phases of scan procedures.
70 1 Radiation Protection
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
The PET scanner uses a radioactive source pin that contains Ge68, an isotope with a half-life of
270.8 days. The radioactive source pin is referred to as “low activity pin.” The pin has an initial
activity level of 10MBq (0.27mCi) ± 20% for Optima 560 and Discovery 610, and 18.5MBq
(0.5mCi) ± 20% for Optima 560 FX, and Discovery 710.
Refer to Illustration 3-5. During normal system operation, the radioactive source pin resides in a
storage container, located inside the PET Trailer, at the rear of the PET/CT gantry.
Illustration 3-5: Source Pin Storage Location
When the pin is in use, it is located inside the gantry near the wall of the patient port. Depending
on the task the pin is performing, it may be held in a fixed location or rotated around the
circumference of the patient port at a speed of up to 20 revolutions per minute (one revolution
per three seconds). The pin is transferred from the storage container to its position near the
patient port, and returned to the storage container after use, by a mechanical system under
software control. Radiation indicators are displayed on both the gantry control panels and the
operator’s keypad when the pin source leaves the storage container.
1.2.5.1 Dose Rates with Pin Source Stored
When the radioactive source pin is stored in the lead container, and no other sources are
present in the scanner room, the maximum dose rates on the PET/CT system are directly over
the source loader. The PET/CT system uses one pin with an initial activity level of 10MBq
(0.27mCi) ± 20% or less for Optima 560 and Discovery 610, and 18.5MBq (0.5mCi) ± 20% for
Optima 560 FX, and Discovery 710. The exposure rate at the cover is specified to equal 2mR/hr
or less.
1.2.5.2 Dose Rates with Pin Source in Use
The dose rates were measured in the following conditions:
• Pin source rotating around the patient port.
• Patient table lowered to its lower limit (55 cm above the floor).
• All measurements were taken at a height equal the center of the patient port.
The results of these experiments, measured along the central axis of the scanner, are
summarized in Table 3-2. (Distances are measured from the front-most imaging slice; positive
distance is in the direction towards the scanner table.)
Table 3-2: Dose Rate at Specified Distances
Dose rate per mCi mR/hr/mCi 0.07 0.14 0.75 5.46 0.68 0.68
72 1 Radiation Protection
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• EMI from a three-phase transformer is much less than a bank of three single-phase
transformers of equivalent power.
• Consider and measure EMI where the facility power is running near the scan room.
• Pay attention to power substations and high-voltage power lines near the scan facility.
• If you have any concerns, measure for all EMI to confirm the site meets all required
specifications.
Electrostatic discharge ± 6 kV contact ± 6 kV contact Floors shall be concrete. If floors are covered with synthet‐
(ESD) IEC 61000-4-2 ± 8 kV air ± 8 kV air ic material, the relative humidity should be at least 30%.
Electrical fast transient/ ± 2 kV for power ± 2 kV for power sup‐ Mains power quality should be a typical commercial or
burst IEC 61000-4-4 supply lines ply lines hospital environment.
± 1 kV for input/ out‐ ± 1 kV for input/ out‐
put lines put lines
Surge IEC 61000-4-5 5 ± 1 kV differential ± 1 kV differential Mains power quality should be a typical commercial or
mode mode hospital environment.
± 2 kV common ± 2 kV common
mode mode
Voltage dips, short inter‐ < 5 % UT (> 95% dip < 5 % UT (> 95% dip Mains power quality should be a typical commercial or
ruptions and voltage var‐ in UT) for 5 sec in UT) for 5 sec hospital environment. If the user of the system requires
iations on power supply continued operation during power mains interruptions, it is
input lines IEC 61000-4- recommended that the system is powered from an uninter‐
11 ruptible power supply or a battery.
Power frequency (50/60 3 A/m 3 A/m Power frequency magnetic fields should be at levels char‐
Hz) magnetic field IEC acteristic of a typical location in a typical commercial or
61000-4-8 hospital environment.
Conducted RF IEC 3 VRMS 150 kHz to 3 V 150 kHz to 80 Do not use portable and mobile RF communications
61000-4-6 80 MHz MHz equipment closer to any part of the system, including ca‐
Radiated RF IEC 3 V/m 150 kHz to 80 3 V/m 150 kHz to 80 bles, than the recommended separation distance calculat‐
61000-4-3 (Alternative MHz MHz ed from the equation appropriate for the frequency of the
method: Full range IEC transmitter.
61000-4-21 test in lieu of Recommended Separation Distance:
Large, Permanently- In‐ See Table 3-4 where P is the maximum output power rat‐
stalled Equipment ex‐ ing of the transmitter in watts (W), according to the trans‐
emption) mitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site sur‐
vey,a should be less than the compliance level in each fre‐
quency range.b Interference may occur in the vicinity of
equipment marked with the following symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, ama‐
teur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electro‐
magnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the system is used exceeds the applicable RF compliance level above, the system should be ob‐
served to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorient‐
ing or relocating the system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
For transmitters rated at a maximum output power not listed above, the separation distance can
be estimated using the equation in the corresponding column, where power (P) is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz to 800 MHz, the separation distance for the higher frequency range
applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
As an example, keep a 1 W mobile phone (800 MHz to 2.5 GHz carrier frequency) at
least 2.3 m from the PET/CT system (to avoid image interference risks).
Limitations Management:
Adhering to the distance separation recommended in (150 KHz to 2.5 GHz) reduces
disturbances recorded at the image level, but may not eliminate all disturbances. However,
when installed and operated as specified, the system maintains its essential performance by
continuing to acquire, display, and store diagnostic quality images safely.
2.1.6 Cable Shielding and Grounding
All interconnect cables to peripheral devices must be shielded and properly grounded, except
when technologically prohibited. Use of cables not properly shielded and grounded may result in
the equipment causing radio frequency interference.
GE Healthcare is not responsible for any interference caused by using other than recommended
interconnect cables or panels, or by unauthorized changes or modifications to this equipment.
Unauthorized changes or modifications could void the users' authority to operate the equipment.
2.2 Electromagnetic Emission
This equipment complies with IEC 60601-1-2 Edition 3 (2007) EMC standard for medical
devices.
NOTE: This system complies with the EMC standard when used with supplied cables. If
cables of different lengths are required, contact your PM. Cables cannot be cut,
shortened, lengthened, or spliced.
The system is suitable to be used in an electromagnetic environment, in compliance with the
limits and recommendations provided in Table 3-5.
RF emissions CISPR 11 Group 1 The system uses RF energy only for its internal function. Therefore, its RF
emissions are very low and not likely to cause interference in nearby electronic
equipment.
RF emissions CISPR 11 Class A When installed in such a shielded location, the scanner is suitable for use in all
establishments other than domestic, and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
3 Vibration Isolation
3.1 Scanning Facility Vibration Isolation
The scanning facility shall be isolated from vibration such as; hospital power plants, pumps,
motors, air handling equipment, air conditioning units, nearby rooms with exercise equipment or
where exercise occurs, hallway foot traffic, elevators, parking lots, roads, subways, trains, and
heliports; otherwise, vibration will affect the image quality of the scanner.
3.2 Frequency/Vibration Range
CT systems are sensitive to vibration and may display limited performance if exceeding the
vibration limits listed below. The band of frequencies in which systems exhibit the most
sensitivity appears at or near the resonant frequencies of the gantry and the patient table, the
latter of which varies depending on patient mass and location. These frequencies fall within the
following ranges:
• Patient Table: 2 – 10 Hz
• Gantry: 8 – 14 Hz
Floor vibration from any source shall not exceed the levels detailed in Illustration 3-7 and
Illustration 3-8, as represented by the solid line labeled CT Scanner/Table. These illustrations
compare this limit to the limits of what the AISC (American Institute of Steel Construction) and
the ISO (International Organization for Standardization) call Class A (VC-A) and Class B (VC-
B).
78 3 Vibration Isolation
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Illustration 3-7: Allowable floor vibration in acceleration units compared to ISO class A and B
limits
80 25
Illustration 3-8: Allowable floor vibration in velocity units compared to ISO class A and B limits
2 CT Scanner/Table 12.5 40
80 50
80 3 Vibration Isolation
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
WARNING
THE SILVER, COPPER, GOLD FILMS USED IN THE CT SYSTEM ARE
ESPECIALLY SENSITIVE TO CHEMICAL CONTAMINATION.
THE PRESENCE OF SULFIDE, CHLORIDE AND NITRATE CONTAMINATES
(WITH SULFUR BEING THE MOST DAMAGING), CAN DAMAGE THE CT
SYSTEM.
IF HIGH LEVELS OF CONTAMINATES EXIST, CONSIDER INSTALLING AN
APPROPRIATE AIR FILTRATION SYSTEM.
The scanner shall not be installed in the same room with a wet film processor. Certain scanner
components could become contaminated by the chemicals contained in the processor.
Ensure any sulfide, chloride, or nitrate contaminate levels are at acceptable levels (Class 1).
See IEC 60654-4 for air quality guidelines.
4.4 Finished Wall Requirement
4.4.1 Wall Paint
The scan and control room walls shall be painted prior to the system installation.
4.4.2 Wall Paint - Exception
A primer coat of paint is acceptable for system installation. After the system is installed, any final
coats of paint shall be applied by brush. Spray painting in not permitted as it can seriously
damage CT system components.
NOTE: Spray painting in not permitted. Spray painting can seriously damage CT system
components.
4.5 Option Requirements
4.5.1 Non-GE Installed Options
Buyer/purchaser shall confirm all non-GE installed options have been reviewed and final
locations determined. Prior to system installation, the buyer/purchaser shall be responsible for
pre-installing all ceiling mounting plates/pedestals for non-GE installed options prior to system
delivery.
4.5.2 GE Options
Buyer/purchaser shall confirm all GE-installed options have been reviewed and final locations
determined.
• drywall dust
Failure to protect the CT system from these types of dust may result in damage to the system
and early system failure.
1.1.1.2 Chemical Contamination Concerns
WARNING
THE SILVER, COPPER, GOLD FILMS USED IN THE CT SYSTEM ARE
ESPECIALLY SENSITIVE TO CHEMICAL CONTAMINATION.
THE PRESENCE OF SULFIDE, CHLORIDE AND NITRATE CONTAMINATES
(WITH SULFUR BEING THE MOST DAMAGING), CAN DAMAGE THE CT
SYSTEM.
IF HIGH LEVELS OF CONTAMINATES EXIST, CONSIDER INSTALLING AN
APPROPRIATE AIR FILTRATION SYSTEM.
The scanner shall not be installed in the same room with a wet film processor. Certain scanner
components could become contaminated by the chemicals contained in the processor.
It is the responsibility of the buyer/purchaser to insure any sulfide, chloride, or nitrate
contaminate levels are at acceptable, low levels (Class 1). See IEC 60654-4 for air quality
guidelines.
1.1.2 Temperature and Humidity Requirements
Ensure the site provides an HVAC system capable of maintaining the temperature and humidity
requirements as specified in Table 4-1 and Table 4-2. The environmental conditions at the site
shall be maintained at all times (including overnight, weekends, and holidays). Environmental
conditions apply to the Table, Gantry, Power Distribution Unit, Q.Core/PARC4 (Q.Core Power)
and scanner desktop.
NOTE: Exceeding the environmental specifications may adversely affect system operation
and image quality.
Table 4-1: System Temperature Limits
NOTE: Be certain to account for ANY cooling equipment cycle-control range, ensuring that
the maximum and minimum ambient room temperatures do not exceed those shown
in Table 4-1 during room thermal cycling. For example, if the HVAC is capable of ± 2º
C control, then the limits would be 20º C - 24º C to maintain absolute limits.
Table 4-2: Humidity (Scan and Control Rooms)
• PARC4 (Q.Core Power) air INTAKE occurs along the FRONT of the PARC4 (Q.Core
Power). PARC4 (Q.Core Power) air EXHAUST occurs at the TOP of the PARC4 (Q.Core
Power).
84 1 HVAC Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Scan Room:
Control Room:
* Does not include heat load from room lighting, non-PET/CT equipment, personnel, etc.
1.3 Air-Handling System Initial Start-Up Considerations
Prior to the initial start-up of the scan suite air-handling system, ensure the air-handling system
ducts and filters are thoroughly clean of dust and other potential airborne contaminants.
After new construction or scan suite renovation, the air-handling ventilation system, on initial
startup, could blow dust and other airborne contaminates throughout the scan suite, potentially
damaging the PET/CT scanner.
86 1 HVAC Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
88 1 Power Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
(For 16 Slice Optima 560, Discovery 610, Discovery 710) The minimum recommended size for
a dedicated distribution transformer is: 112.5 kVA, rated 2.4% regulation at unity power factor.
Resultant maximum allowable feeder regulation is 3.4%.
1.8.3 Using an Existing Distribution Transformer
Do not use an existing distribution transformer to power a system if other X-ray equipment,
using rapid film changers, is connected to the existing transformer.
1.9 System Power Requirements (64 Slice Discovery 610, Discovery 710)
The customer shall ensure the site meets all minimum system power requirements listed below
before installation can begin.
• Maximum power demand = 150kVA @ 0.85 PF: at a Selected Technique of 140 kV, 715
mA.
Minimum Recommended Circuit Protection Rating 150 150 150 125 125 125
Table 5-2: Minimum Feeder Wire Size (64 Slice Discovery 610, Discovery 710)
Feeder Length (Power Substation Minimum Feeder Wire Size, AWG or MCM (sq. mm)/ VAC
to A1 Disconnect)
380 VAC 400 VAC 420 VAC 440 VAC 460 VAC 480 VAC
15 m (50 ft) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
30 m (100 ft) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
46 m (150 ft) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
61 m (200 ft) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
76 m (250 ft) 2/0 (70) 2/0 (70) 1/0 (55) 1/0 (55) 1 (45) 1 (45)
91 m (300 ft) 3/0 (85) 3/0 (85) 2/0 (70) 2/0 (70) 1/0 (55) 1/0 (55)
107 m (350 ft) 4/0 (100) 3/0 (85) 3/0 (85) 2/0 (70) 2/0 (70) 1/0 (55)
122 m (400 ft) 250 (125) 4/0 (100) 3/0 (85) 3/0 (85) 3/0 (85) 2/0 (70)
NOTE: In all cases the recommended ground wire is a 1/0 (55 sq. mm) ground wire.
90 1 Power Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
Table 5-3: Minimum Sub-Feeder Wire Size (64 Slice Discovery 610, Discovery 710)
Sub-feeder Length (A1 to PDU) Minimum Sub-feeder Wire, AWG or MCM (sq. mm)
380 VAC 400 VAC 420 VAC 440 VAC 460 VAC 480 VAC
9.75 m (32 ft) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
The information in Table 5-1, Table 5-2, and Table 5-3 (above) assumes the use of copper
wire, rated 75° C and run in steel conduit. All ampacity is determined in accordance with the
National Electrical Code (NFPA 70), Table 310-16 (2002). The ampacity of the circuit
protection device listed above determines the minimum feeder size, except where total
source regulation limits require a larger size.
NOTE: Power feeders running under the scan room floor, as well as power vault
substations under the floor, above the scan suite, or in adjacent rooms, may
cause excessive EMI fields. The responsibility for meeting all site EMI
requirements rests with the customer.
1.10 System Power Requirements (16 Slice Optima 560, Discovery 610, Discovery
710)
The customer shall ensure the site meets all minimum system power requirements listed below
before installation can begin.
• Maximum power demand = 90kVA @ 0.85 PF: at a Selected Technique of 140 kV, 380 mA.
Table 5-4: Nominal Line Voltage Ranges (16 Slice Optima 560, Discovery 610, Discovery 710)
Nominal Line Voltage 200 220 240 380 400 420 440 460 480
Hi-Line Limit, +10% 220 242 264 418 440 462 484 506 528
Lo-Line Limit, -10% 180 198 216 342 360 378 396 414 434
Momentary Line Current 260 236 217 137 130 124 118 113 108
Maximum Line Current 289 262 241 152 144 137 131 126 120
Minimum Recommended Circuit Protection 150 150 150 110 110 100 100 90 90
Rating
Table 5-5: Minimum Feeder Wire Size (16 Slice Optima 560, Discovery 610, Discovery 710)
Feeder Length (Power Minimum Feeder Wire Size, AWG or MCM (sq. mm)/ VAC
Substation to A1 Dis‐
connect) 200 VAC 220 VAC 240 VAC 380 VAC 400 VAC 420 VAC 440 VAC 460 VAC 480
VAC
15 m (50 ft) 1/0 (55) 1/0 (55) 1/0 (55) 2 (35) 2 (35) 3 (30) 3 (30) 3 (30) 3 (30)
30 m (100 ft) 2/0 (70) 1/0 (55) 1/0 (55) 2 (35) 2 (35) 3 (30) 3 (30) 3 (30) 3 (30)
46 m (150 ft) 4/0 (100) 3/0 (85) 2/0 (70) 2 (35) 2 (35) 3 (30) 3 (30) 3 (30) 3 (30)
Feeder Length (Power Minimum Feeder Wire Size, AWG or MCM (sq. mm)/ VAC
Substation to A1 Dis‐
connect) 200 VAC 220 VAC 240 VAC 380 VAC 400 VAC 420 VAC 440 VAC 460 VAC 480
VAC
61 m (200 ft) 5/0 (125) 4/0 (100) 4/0 (100) 2 (35) 2 (35) 3 (30) 3 (30) 3 (30) 3 (30)
76 m (250 ft) 6/0 (170) 5/0 (125) 5/0 (125) 1 (45) 1 (45) 2 (35) 2 (35) 2 (35) 3 (30)
91 m (300 ft) 7/0 (215) 6/0 (170) 5/0 (125) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 2 (35) 2 (35)
107 m (350 ft) 8/0 (275) 7/0 (215) 6/0 (170) 2/0 (70) 1/0 (55) 1/0 (55) 1 (45) 1 (45) 1 (45)
122 m (400 ft) 8/0 (275) 7/0 (215) 7/0 (215) 2/0 (70) 2/0 (70) 1/0 (55) 1/0 (55) 1/0 (55) 1 (45)
NOTE: In all cases the recommended ground wire is a 1/0 (55 sq. mm) ground wire.
Table 5-6: Minimum Sub-Feeder Wire Size (16 Slice Optima 560, Discovery 610, Discovery
710)
9.75 m (32 ft) 1/0 (55) 1/0 (55) 1/0 (55) 2 (35) 2 (35) 3 (30) 3 (30) 3 (30) 3 (30)
The information in Table 5-4, Table 5-5, and Table 5-6 (above) assumes the use of copper wire,
rated 75° C and run in steel conduit. All ampacity is determined in accordance with the National
Electrical Code (NFPA 70), Table 310-16 (2002). The ampacity of the circuit protection device
listed above determines the minimum feeder size, except where total source regulation limits
require a larger size.
NOTE: Power feeders running under the scan room floor, as well as power vault substations
under the floor, above the scan suite, or in adjacent rooms, may cause excessive EMI
fields. The responsibility for meeting all site EMI requirements rests with the customer.
92 1 Power Requirements
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
2 Grounding
The design of the scanner uses an equal potential grounding system. Illustration 5-2 and Table
5-7 detail the required ground system. Three primary grounding points exist, they include:
• A system power ground point located in the PDU.
• A reference ground point located between the gantry and the table base.
The electrical contractor shall ground ALL patient-accessible metal surfaces to the same
potential as the A1 Disconnect. The electrical contractor shall bond the ground wire to any
intermediate distribution panel the ground wire passes through, in accordance with all local
codes.
Illustration 5-2: System Ground Map
Bonding Power from A1 to PDU The metal conduit, raceway, or armored cabling used to run power from the A1
Disconnect to the PDU shall be bonded in accordance to the NEC.
Dedicated Ground A dedicated 1/0 (55 mm2), or larger, insulated copper ground wire shall be instal‐
led between the main distribution panel and the PDU, in accordance with the
NEC.
Grounding Power, A1, and PDU All three-phase wires with ground running between the power source, the A1
Disconnect, and the PDU shall be installed in accordance to the NEC.
Maximum Resistance Between PDU and Facility The resistance between the PDU ground and the facility Earth ground shall not
Ground exceed 0.5 ohm.
Maximum Resistance Between PDU and Earth The resistance between the PDU ground and Earth ground shall not exceed 2
ohms.
Cable Shielding and Grounding All interconnect cables to peripheral devices shall be shielded and properly
grounded, except where technologically prohibited.
94 2 Grounding
Optima PET/CT 560, 560 FX; Discovery PET/CT 610, 710 Pre-Installation Manual
Direction 5433542-1EN, Revision 5
• Table 5-9 details the Standard-Length Cable Kit 5491000-3 (P5051TE) – Supplied by GE
Healthcare.
• Table 5-10 details the additional Standard-Length Cable Kit 5491000-10 (P5064TE) for 64-
slice systems – Supplied by GE Healthcare.
• Table 5-11 details the Long-Length Cable Kit, Optional 5491000-4 (P5051TF) – Supplied by
GE Healthcare.
• Table 5-12 details the additional Long-Length Cable Kit, Optional 5491000-20 (P5064TF) for
64-slice systems – Supplied by GE Healthcare.
• Table 5-17 details the PARC4 (Q.Core Power) Upgrade Mandatory Cable Kit, Optional
5768790 (P3200ZH) for 64-slice systems – Supplied by GE Healthcare.
• Table 5-18 details the PARC4 (Q.Core Power) Upgrade Relocation Cable Kit, Optional
5768683 (P3200PT) for 64-slice systems – Supplied by GE Healthcare.
56 25.5 5339979- Console GND 1238 VW-1 600 0 105 11.9 1 2 12.2 (0.5)
(83.7) 3 to Raceway (FT-1) (0.47)
[22.1 (73)] GND
102 26.4 2373436- Gantry to RG-2 FT-4 1900 <30 5.9 8 24 15 (0.6)
(86.6) 2 Console LAN 2 VDC (0.23)
[22.9 (75)]
101 26.4 5419981 Console to RG-2 FT-4 300 <30 80 11.2 25 22 17x58
(86.6) MSUB J9 2 VDC (0.44) (0.7x2.3)
[22.9 (75)] 19x51
(0.7x2.0)
200 30.5 (100) 5313938- J7 to Console, UL FT-4 300 <30 60 6.8 4 pair 24 13 (0.5)
[22.9 (75)] 6 Respiratory VDC (0.26)
XX 30.5 (100) 5169456 Gantry to In‐ 2464 FT-4 300 80 6.6 22 40 (1.5)
[27.7 (91)] jector (0.26)
52A 8.6 (28.2) 2343528- PDU to Gan‐ 2587 FT-4 600 208Y/1 90 13.8 5 8 56.4 (2.2)
[6.1 (20)] 2 try 120VAC 20 (0.54)
52A 8.6 (28.2) 2343528- PDU to Gan‐ 2587 FT-4 600 208Y/1 90 13.8 5 8 56.4 (2.2)
Note [6.1 (20)] 4 try 120VAC 20 (0.54)
1 (Not Used)
50A 8.6 (28.2) 2343529- HVDC from 2587 FT-4 600 350 90 19 3 (2) 4, (1) 22 (0.9)
[6.1 (20)] 2 PDU to Gan‐ VDC (0.75) 8
try
51A 8.6 (28.2) 2343530- Axial Drive 2587 FT-4 600 440Y/2 90 12.3 4 14
[6.1 (20)] 2 Power PDU to 54 (0.48)
Gantry
55A 8.6 (28.2) 5339979- Raceway 1238 VW-1 600 0 105 11.9 1 2 12.2 (0.5)
[4.3 (14)] 2 GND to PDU - (FT-1) (0.47)
GND
100A 9.9 (32.5) 5120646- PDU to MSUB FT-4 300 <30 80 11.2 25 22 17x58
[6.1 (20)] 2 J11 VDC (0.44) (0.7x2.3)
19x51
(0.7x2.0)
209A 13 (42.6) 5339979- Q.Core GND 1238 VW-1 600 0 105 11.9 1 2 12.2 (0.5)
[9.6 (32)] 6 to Raceway (FT-1) (0.47)
GND
203 13 (42.6) 5313938- SBA J7 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
[9.9 (33)] 7 Q.Core J6 VDC (0.26)
201 13 (42.6) 5313938- Q.Core J4 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
[8.4 (28)] 8 Switch Port 5 VDC (0.26)
202 13 (42.6) 5313938- PARC J5 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
Note [8.4 (28)] 9 Switch Port 7 VDC (0.26)
1 (Not Used)
203 13 (42.6) 5313941– PDU TS5 to 2587 FT-4 600 208Y/1 60 19 5 10 25 (1.0)
Note [9.6 (32)] 2 PARC Bulk‐ 20 (0.75)
1 head (Not
Used)
203 13 (42.6) 2343531- Q.Core Power 2587 FT-4 600 120 90 11.7 3 10 56.4 (2.2)
[9.6 (32)] 4 from PDU, VAC (0.46)
short
053A 19.9 2343531- PDU TS5 to 2587 FT-4 600 120 90 12.2 3 10 56.4 (2.2)
(65.3) 2 Console Pow‐ VAC (0.48)
[16.6 (54)] er
Note 1: Extra Cable. Not used for Optima 560, Discovery 610, and Discovery 710 systems.
Table 5-10: Additional Standard-Length Cable Kit 5491000-10 (P5064TE) for 64-slice Systems
– Supplied by GE Healthcare
053A 19.9 5121809- PDU TS5 to 2587 FT-4 600 208Y/1 90 12.3 4 10 56 (2.2)
Note (65.3) 2 Console Pow‐ 20 (0.48)
1 [16.6 (54)] er
56 25.5 5339979-3 Console GND 1238 VW-1 600 0 105 11.9 1 2 12.2 (0.5)
(83.7) to Raceway (FT-1) (0.47)
[22.1 (73)] GND
102 26.4 2373436-2 Gantry to RG-2 FT-4 1900 <30 5.9 8 24 15 (0.6)
(86.6) Console LAN 2 VDC (0.23)
[22.9 (75)]
101 26.4 5419981 Console to RG-2 FT-4 300 <30 80 11.2 25 22 17x58
(86.6) MSUB J9 2 VDC (0.44) (0.7x2.3)
[22.9 (75)] 19x51
(0.7x2.0)
200 30.5 (100) 5313938-6 J7 to Con‐ UL FT-4 300 <30 60 6.8 4 pair 24 13 (0.5)
[22.9 (75)] sole, Respira‐ VDC (0.26)
tory
XX 30.5 (100) 5169456 Gantry to In‐ 2464 FT-4 300 80 6.6 22 40 (1.5)
[27.7 (91)] jector (0.26)
52 19.4 2343528 PDU to Gan‐ 2587 FT-4 600 208Y/1 90 13.8 5 8 56.4 (2.2)
(63.6) try 120VAC 20 (0.54)
[17.2 (56)]
52 19.4 2343528-3 PDU to Gan‐ 2587 FT-4 600 208Y/1 90 13.8 5 8 56.4 (2.2)
Note (63.6) try 120VAC 20 (0.54)
1 [17.2 (56)] (Not Used)
50 19.4 2343529 HVDC from 2587 FT-4 600 350 90 19 3 (2) 4, 22 (0.9)
(63.6) PDU to Gan‐ VDC (0.75) (1) 8
[17.2 (56)] try
55 19.4 5339979 Raceway 1238 VW-1 600 0 105 11.9 1 2 12.2 (0.5)
(63.6) GND to PDU - (FT-1) (0.47)
[15.1 (50)] GND
209 25.5 5339979-5 Q.Core GND 1238 VW-1 600 0 105 11.9 (. 1 2 12.2 (0.5)
(83.6) to Raceway (FT-1) 47)
[22.1 (73)] GND
203 30.5 (100) 5313938 SBA J7 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
[27.4 (90)] Q.Core J6 VDC (0.26)
201 30.5 (100) 5313938-2 Q.Core J4 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
[25.9 (85)] Switch Port 5 VDC (0.26)
202 30.5 (100) 5313938-3 PARC J5 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
Note [25.9 (85)] Switch Port 7 VDC (0.26)
1 (Not Used)
203 19.4 5313941 PDU TS5 to 2587 FT-4 600 208Y/1 60 19 5 10 25 (1.0)
Note (63.6) PARC Bulk‐ 20 (0.75)
1 [15.8 (52)] head (Not
Used)
203 19.4 2343531-3 Q.Core Power 2587 FT-4 600V 120 90 11.7 3 10 56.4 (2.2)
(63.6) [16 from PDU, VAC (0.46)
(52.6)] long
053 24.5 2343531 PDU TS5 to 2587 FT-4 600 120 90 12.3 3 10 56.4 (2.2)
(80.4) Console Pow‐ VAC (0.48)
[21.2 (69)] er
Note 1: Extra Cable. Not used for Optima 560, Discovery 610, and Discovery 710 systems.
Table 5-12: Additional Long-Length Cable Kit 5491000-20 (P5064TF) for 64-slice Systems –
Supplied by GE Healthcare
053A 24.5 5121809 PDU TS5 to 2587 FT-4 600 208Y/1 90 12.3 4 10 56 (2.2)
Note (80.4) Console Pow‐ 20 (0.48)
1 [21.2 (69)] er
110 14 (46) [13.7 5169224 A1 to UPS 2587 FT4 600 208Y/ 90 14 5 18 25 (1.0)
(45)) 120 (0.54) Dia.
PDU Model No. Maximum Nomi‐ Required Mains Disconnect (A1) Catalog No. Optional Partial UPS Kit
nal kVA Rating Catalog No. (See Note 2)
Europe and Asia (380-400V or North America (440V or
420V) (See Note 1) 460-480V)
Note 1: Additional A1 Disconnects for Europe available through European Sales Team
Note 2: REQUIRES one of the A1 mains disconnect detailed at left, or equivalent.
Customer Installed Wiring Description Cables Supplied Plug Pulling Di‐ Wire and Cable Pig‐
mensions tails m (ft)
Qty Size AWG (mm2) Part No Length Dia. in. From To From To
m (ft) (mm)
RUN NO. 1 FROM PRIMARY POWER SOURCE TO FACILITY DISCONNECT (POWER SOURCE - A1)
Maximum Run Length *
RUN NO. 2 FROM FACILITY DISCONNECT TO POWER DISTRIBUTION UNIT (A1 - PM)
Maximum Run Length *
RUN N0. 3 FROM FACILITY DISCONNECT TO SYSTEM EMERGENCY OFF (A1 - SEO)
RUN NO. 4 POWER DISTRIBUTION UNIT TO WARNING LIGHT CONTROL (PDU - WL)
CONTROL TS6 1, 2, 3, 4, 5, 6,
7, 8
RUN NO. 5 POWER DISTRIBUTION UNIT TO SCAN ROOM DOOR INTERLOCK (PDU - DOOR SWITCH)
System Reference Associated Material/Labor Supplied by Customer USA Vendor / CAT No. GE Catalog
Equipment Contractor
Table 5-17: PARC4 (Q.Core Power) Upgrade Mandatory Cable Kit 5768790 (P3200ZH) (64-
slice)
210 30.5 (100) 5313938- PARC J7 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
[22.9 (75)] 14 Switch J7, VDC (0.26)
Long
Table 5-18: PARC4 (Q.Core Power) Upgrade Relocation Cable Kit 5768683 (P3200PT) (64-
slice)
25.5 5339979- PARC GND to 1015, VW-1 600 0 105 11.9 (. 1 2 12.2 (0.5)
(83.6) 5 Raceway 1063, (FT-1) 47)
209
[21.8 GND Bar 1284,
(71.5)] 1283
30.5 (100) 5313938- RRR J4 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
201 [24.3 2 Switch Port 5 VDC (0.26)
(79.7)]
30.5 (100) 5313938 SBA J7 to UL FT-4 300 <30 60 6.6 4 pair 24 13 (0.5)
203
[27.4 (90)] RRR J6 VDC (0.26)
A PDU 5 Line
Illustration 5-5: TS6 Room Door Interlock Connections – with Door Interlock
www.gehealthcare.com