Professional Documents
Culture Documents
ASiR Upgrade Installation Manual - 5496303-1EN - 4
ASiR Upgrade Installation Manual - 5496303-1EN - 4
5496303-1EN
Rev 4
© 2019, GE Company,
All Rights Reserved.
GE COMPANY
DIRECTION 5496303-1EN, REVISION 4 ASIR UPGRADE INSTALLATION MANUAL
Page 2
GE COMPANY
DIRECTION 5496303-1EN, REVISION 4 ASIR UPGRADE INSTALLATION MANUAL
LEGAL NOTES
TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright © 2019 by GE Company, Inc. All rights reserved.
IMPORTANT PRECAUTIONS
LANGUAGE
警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他
形式的伤害。
警告 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯
服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性
或其他的危險。
警告 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服
務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其
他危險而受傷。
UPOZORENJE Ovaj servisni priručnik dostupan je na engleskom jeziku.
(HR) • Ako davatelj usluge klijenta treba neki drugi jezik, klijent je dužan
osigurati prijevod.
• Ne pokušavajte servisirati opremu ako niste u potpunosti pročitali i
razumjeli ovaj servisni priručnik.
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge,
operatera ili pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.
警告 • このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳
作業はその業者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わ
ないでください。
• この警告に従わない場合、サービスを担当される方、操作員あるいは
患者さんが、感電や機械的又はその他の危険により負傷する可能性が
あります。
경고 • 본 서비스 지침서는 영어로만 이용하실 수 있습니다 .
(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우 , 번역 서비스
를 제공하는 것은 고객의 책임입니다 .
• 본 서비스 지침서를 참고했고 이해하지 않는 한은 해당 장비를 수리하려
고 시도하지 마십시오 .
• 이 경고에 유의하지 않으면 전기 쇼크 , 기계상의 혹은 다른 위험으로부
터 서비스 제공자 , 운영자 혹은 환자에게 위해를 가할 수 있습니다 .
BRDINJUMS • Ðî apkalpes rokasgrâmata ir pieejama tikai angïu valodâ.
• Ja klienta apkalpes sniedzçjam nepiecieðama informâcija citâ valodâ,
(LV) nevis angïu, klienta pienâkums ir nodroðinât tulkoðanu.
• Neveiciet aprîkojuma apkalpi bez apkalpes rokasgrâmatas izlasîðanas
un sapraðanas.
• Ðî brîdinâjuma neievçroðana var radît elektriskâs strâvas trieciena,
mehânisku vai citu risku izraisîtu traumu apkalpes sniedzçjam,
operatoram vai pacientam.
ÁSPËJIMAS • Ðis eksploatavimo vadovas yra prieinamas tik anglø kalba.
• Jei kliento paslaugø tiekëjas reikalauja vadovo kita kalba – ne anglø,
(LT) numatyti vertimo paslaugas yra kliento atsakomybë.
• Nemëginkite atlikti árangos techninës prieþiûros, nebent atsiþvelgëte á
ðá eksploatavimo vadovà ir já supratote.
• Jei neatkreipsite dëmesio á ðá perspëjimà, galimi suþalojimai dël elek-
tros ðoko, mechaniniø ar kitø pavojø paslaugø tiekëjui, operatoriui ar
pacientui.
ADVARSEL • Denne servicehåndboken finnes bare på engelsk.
• Hvis kundens serviceleverandør trenger et annet språk, er det kundens
(NO) ansvar å sørge for oversettelse.
• Ikke forsøk å reparere utstyret uten at denne servicehåndboken er lest
og forstått.
• Manglende hensyn til denne advarselen kan føre til at servicelev-
erandøren, operatøren eller pasienten skades på grunn av elektrisk støt,
mekaniske eller andre farer.
OSTRZEŻENIE • Niniejszy podręcznik serwisowy dostępny jest jedynie w języku angiel-
skim.
(PL) • Jeśli dostawca usług klienta wymaga języka innego niż angielski,
zapewnienie usługi tłumaczenia jest obowiązkiem klienta.
• Nie próbować serwisować wyposażenia bez zapoznania się i zrozumie-
nia niniejszego podręcznika serwisowego.
• Niezastosowanie się do tego ostrzeżenia może spowodować urazy
dostawcy usług, operatora lub pacjenta w wyniku porażenia elektrycz-
nego, zagrożenia mechanicznego bądź innego.
IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be
used.
Revision History
Revision History
Revision Date Reason for change
1 June 16, 2015 Initial release.
2 September 28, 2015 Add product name for Revolution ACT and Revolution ACTs.
3 August 11, 2017 Add Z440 information.
4 August 21, 2019 Update ASiR option label & Console Rating Plate
Table of Contents
Chapter 1
Overview................................................................................................................. 17
Section 1.0
Introduction ...................................................................................................... 17
1.1 Overview.......................................................................................................................... 17
1.2 Flowchart ......................................................................................................................... 17
Table of Contents
Section 2.0
Material List ...................................................................................................... 18
Chapter 2
Installation Procedure........................................................................................... 19
Section 1.0
Overview ........................................................................................................... 19
Section 2.0
Procedure.......................................................................................................... 19
2.1 Preparation ...................................................................................................................... 19
2.2 Power-Off (Shutdown) ..................................................................................................... 19
2.3 GPU Card Installation ...................................................................................................... 20
2.3.1 Z420 Host Computer .......................................................................................... 20
2.3.2 Z440 Host Computer .......................................................................................... 22
2.4 Check GPU Card ............................................................................................................. 24
2.5 Console Cover Installation............................................................................................... 24
2.6 Power On......................................................................................................................... 24
2.7 Recon Enhancement Option Installation ......................................................................... 25
2.8 ASIR Application Installation............................................................................................ 25
Section 3.0
Finalization........................................................................................................ 25
Chapter 1
Overview
Section 1.0
Introduction
1.1 Overview
This procedure defines how to install GPU card and option application.
1 - Introduction
1.2 Flowchart
The illustration below shows the main flow of the installation procedures of this option.
Start
Preparation
Power Off
Power On
Finalization
End
Section 2.0
Material List
Verify that the contents of option kit is correct according to the Material List below.
Chapter 2
Installation Procedure
Section 1.0
Overview
This procedure describes the ASIR Upgrade Installation. The main components are listed as
follows:
• ASiR software option without GPU card and Recon enhancement
- ASiR Software
2 - Introduction
• ASiR software option with GPU card and recon enhancement
- ASiR Software
- Z420 Host PC: AMD FirePro V3900 1GB GFX Graphic Card
Z440 Host PC: AMD W4300 with qualified vbios Graphic Card
- Recon Enhancement Software
Section 2.0
Procedure
2.1 Preparation
1.) Select one of the following methods to Power OFF the Operator Console if the Power is ON.
a.) If Applications are up, click on the [Shut Down] button and select [Shutdown].
b.) If Applications are down, open a Unix Shell from pop-up menu. Enter (type) the following
command, at the indicated prompts.
ctuser@hostname: su -
Password: #bigguy
root@hostname: halt
2.) Power OFF the SVCT PDU.
NOTICE FOLLOW ALL REQUIRED SAFETY AND PPE PROCEDURES CUSTOMARY FOR YOUR
ORGANIZATION WHEN WORKING ON THIS PRODUCT.
2 - Introduction
Figure 2-3 Remove PCI Slot Plate
5.) Align the connectors on the card with the slot on the system board.
6.) Lower the card down over the slot on the system board.
7.) Press down on the card until it is fully seated.
8.) Rotate the PCI retention clamp down over the PCI bulkhead.
9.) Press the two green PCI retention clamp tabs on the rear of the chassis until they lock into
place.
2 - Introduction
Figure 2-7 Press Two Tabs of PCI Retention Clamp
5.) Align the connectors on the card with the slot on the system board.
6.) Lower the card down over the slot on the system board.
7.) Press down on the card until it is fully seated.
a.) Power on console and Open a shell, then type: glxinfo | grep render
b.) Check the result as below figures.
NOTICE AMD uses the same set of drivers for W5100 and W4300 cards. Hence W4300 Card is
detected and reported by OS as W5100. This behavior is known and is not an issue
because GEHC part number 8770000-131 for SVCT16 and SVCT8 is for W4300 card. GEHC
does not use W5100 in Revolution ACT/ACTs (SVCT16) and Revolution ACTs (SVCT8).
Hence “W5100” output of glxinfo is to be interpreted as “W4300”.
2.6 Power On
Power On the system and confirm the monitor display shows normal screen, and starts up without
any error.
1.) Un-install Neuro 3D Filter option. (This option is not a part of Std. option and conflicts with ASIR
Application.)
2.) Install ASIR Application software using Option Install menu.
Section 3.0
Finalization
2 - Introduction
1.) Execute the sanity scan. Confirm the scan is completed and image is generated without any
error.
2.) Attach the ASiR option label near the Console Port label in side console cover.
3.) Attach the console rating plate on the OC rating plate board in rear chassis.