Professional Documents
Culture Documents
I-P-1.01-W-2 (Packer Spouts)
I-P-1.01-W-2 (Packer Spouts)
Instrumentation Department
Issue No:02 Rev No:01 Effective Date: 01.07.2016 I-P-1.01-W-2
CALIBRATION OF PACKER SPOUTS.
1.1.2 Check mechanical alignment of bag chairs and load cell assembly.
1.1.3 Select the tare symbol on spout controller and make it zero kgs.
1.1.4 Apply upward and downward force on the bag chair, Note the readings after applying &
before applying the force. If any variation observed between the two readings, calibration
has to be done with respective parameters .
1.2.3 Observe the display of respective spout for 50 Kg and remove test weight from the weigher
and check for zero if any variation observed repeat the procedure point no-1.1 and 1.2.