Professional Documents
Culture Documents
http://dx.doi.org/10.7567/APEX.7.062401
To evaluate the X-ray-induced afterglow phenomenon, we developed an ionizing-radiation-induced luminescence characterization system
equipped with a pulse-width-tunable X-ray source. The system consists of a pulse X-ray tube and a detector system based on photon counting.
The excitation pulse width was tunable from nano- to millisecond ranges, and the dynamic range of the X-ray-induced afterglow was 106.
Conventional scintillators for X-ray CT or security systems, namely, Bi4Ge3O12, CdWO4, Tl-doped CsI, and Tb and Pr-codoped Gd2O2S, were
evaluated for the performance test. Results show that the afterglow time profiles of these scintillators are consistent with generally known results
with high accuracy. © 2014 The Japan Society of Applied Physics
norganic scintillators with ceramic or single-crystalline detectors, it would be fruitful to develop some other scientific