You are on page 1of 10

Magnetic Resonance Imaging 24 (2006) 103 – 112

Original contributions
Effects of static and radiofrequency magnetic field inhomogeneity in
ultra-high field magnetic resonance imaging
Trong-Kha Truonga,1, Donald W. Chakeresa, David Q. Beversdorf b,
Douglas W. Scharreb, Petra Schmalbrocka,4
a
Department of Radiology, The Ohio State University, Columbus, OH 43210, USA
b
Department of Neurology, The Ohio State University, Columbus, OH 43210, USA
Received 21 September 2004; accepted 18 September 2005

Abstract
To characterize the severe static (B 0) and radiofrequency (B 1) magnetic field inhomogeneity in ultra-high field ( z 7 T) magnetic
resonance imaging, gradient echo (GE) and spin echo (SE) images of in vivo and postmortem human brains were acquired. The B 0 and B 1
inhomogeneity were experimentally mapped and/or numerically simulated, and correlated with the image artifacts. Whereas B 0
inhomogeneity affects predominantly GE images near air/tissue interfaces, B 1 inhomogeneity affects SE images more severely and shows
nonintuitive patterns. Mapping of the B 0 and B 1 inhomogeneity is important in characterizing image artifacts. This will help develop better
B 0 and B 1 inhomogeneity correction methods.
D 2006 Elsevier Inc. All rights reserved.
Keywords: Static magnetic field; Radiofrequency magnetic field; Ultra-high field MRI; Human brain

1. Introduction in image artifacts, specifically signal loss due to intravoxel


dephasing in gradient echo (GE) imaging, as well as
Ultra-high field (z 7 T) magnetic resonance imaging
geometric distortions in both GE and spin echo (SE)
(MRI) offers many advantages over MRI at clinical field
imaging. Moreover, severe B 1 inhomogeneity leads to
strengths up to 3 T, particularly an increased signal-to-noise
substantial variations of the flip angle and receive sensitivity
ratio (SNR), allowing higher spatial resolution and/or faster
throughout the image volume, resulting in variable SNR and
imaging, a greater spectral dispersion, as well as an
image contrast. In addition, B 0 inhomogeneity may also
enhanced sensitivity to magnetic susceptibility. Its potential
alter the effective flip angle [7].
has been clearly demonstrated for high-resolution imaging
Overall, both B 0 and B 1 inhomogeneity result in regional
of the brain microvasculature [1,2], brain spectroscopy [3]
signal variation and signal loss, and it is often not intuitive
and functional MRI (fMRI) [4].
whether B 0 or B 1 effects are dominant. The objective of this
However, ultra-high field MRI also suffers from more
work is to characterize these effects on representative GE
severe static magnetic field (B 0) [5] and radiofrequency
and SE images of the human brain acquired at ultra-high
(RF) magnetic field (B 1) [6] inhomogeneity. Susceptibility
field strength and relate them to quantitative measures of the
differences at tissue interfaces, most prominently air/tissue
B 0 and B 1 inhomogeneity. This knowledge is expected to
interfaces, cause macroscopic B 0 inhomogeneities that result
ultimately help devise better strategies for correction of
image artifacts due to B 0 and B 1 inhomogeneity seen at
Preliminary results of this work were presented at the 12th Scientific ultra-high field strength.
Meeting and Exhibition of the International Society for Magnetic
Resonance in Medicine, Kyoto, Japan. 15 – 21 May 2004, p. 2170.
4 Corresponding author. Tel.: +1 614 293 4139; fax: +1 614 293 8129. 2. Methods
E-mail address: schmalbrock.1@osu.edu (P. Schmalbrock).
1
Present address: Brain Imaging and Analysis Center, Duke University, The studies were performed on an 8 T/80 cm bore
Durham, North Carolina 27710, USA. human whole-body MRI system (Magnex-General Electric,
0730-725X/$ – see front matter D 2006 Elsevier Inc. All rights reserved.
doi:10.1016/j.mri.2005.09.013
104 T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112

Abingdon, UK) equipped with a Bruker Avance console obtained by segmentation of computed tomography (CT)
(Billerica, MA, USA). The images were acquired using a images of the head, neck and thorax [5].
transverse electromagnetic (TEM) RF head coil [8] with For both the experimental measurements and the
16 struts and four excitation ports that was individually numerical simulations, maps of the B 0 gradient along the
tuned for each study. We studied a total of 15 postmortem x-, y- and z-axes (i.e., the change of the B 0 field along these
unembalmed human subjects (seven male and eight axes) were computed. Note that the B 0 gradient in the slice
female, 57 –85 years old) with various pathologies, and direction (i.e., orthogonal to the slice) contributes the most
12 healthy volunteers (seven male and five female, 20 – 53 to susceptibility artifacts in two-dimensional (2D) imaging,
years old) who gave informed consent to the experimental because the slice thickness is typically much larger than the
protocol approved by our institutional review board. in-plane resolution.
Postmortem studies were carried out initially to develop To quantify the B 1 inhomogeneity, maps of the local flip
the acquisition protocols. Representative cases of these in angle and receive sensitivity were experimentally measured
vivo and postmortem studies were evaluated in the work as follows. Two series of low-resolution GE images were
presented here. acquired with TR H T 1 and nominal flip angles of a 0 and
Because of the substantial flip angle variability prevalent 2a 0, in which case the signal intensity at pixel (x, y) can be
at ultra-high field strength, we defined a bnominal Q flip expressed as:
angle as the average flip angle in a 1-cm3 region of interest,
typically located in the hippocampus or the central cingulate Sa ð x; yÞ ¼ qð x; yÞrð x; yÞsin½að x; yÞ ð1Þ
gyrus. The transmit power level resulting in a nominal flip and
angle of 908 was first determined using a voxel-selective
stimulated echo pulse sequence with an 8-ms sinc RF pulse. S2a ð x; yÞ ¼ qð x; yÞrð x; yÞsin½2að x; yÞ ð2Þ
This value was then used together with the calibration of the respectively, where q is the proton density, r the local
RF amplifier as a reference to set the transmit power level receive sensitivity and a the local flipangle. A flip angle
corresponding to a specified nominal flip angle in subse- map can be computed from the signal intensity ratio of the
quent acquisitions. two series of images as follows [10]:
High-resolution GE images were acquired with the  
following parameters: 8 ms sinc RF pulse, repetition time 1 S2a ð x; yÞ
að x; yÞ ¼ cos : ð3Þ
(TR) 600 ms, echo time (TE) 12 ms, nominal flip angle 238 2Sa ð x; yÞ
or 488, receiver bandwidth (BW) 50 kHz, field-of-view
Then a map of the product of the proton density and the
(FOV) 1818 cm2 or 1813.5 cm2, matrix size (MTX)
receive sensitivity can be computed from the flip angle map
1024512 or 1024384, slice thickness (ST) 2 mm, and
using Eq. (1):
number of excitation (NEX) 1. A series of axial, coronal and
sagittal images were acquired with the number of slices and Sa ð x; yÞ
slice gap adjusted to allow acquisition of representative qð x; yÞrð x; yÞ ¼ : ð4Þ
sin½að x; yÞ
images throughout the brain.
High-resolution SE images were acquired with TR 1500ms, Finally, a measure of the receive sensitivity can be
TE 40, 50 or 70 ms, nominal flip angle 908/1808, MTX generated from this map by low-pass filtering (with a 2D
512256 (in vivo), 512384 or 256256 (postmortem), ST boxcar filter) and normalizing between 0% and 100%. Note
3 mm, NEX 1 (in vivo) or 2 (postmortem), and otherwise iden- that since the spatial frequency bands of the proton density
tical parameters. and receive sensitivity variations may overlap, such a low-
To quantify the B 0 inhomogeneity, a B 0 map was pass filtering could possibly result in residual proton density
experimentally measured as follows. Two series of low- weighting (with low spatial frequencies) and/or an attenu-
resolution GE images were acquired at two different TEs ation of the receive sensitivity variations (with high spatial
using a three-dimensional (3D) dual-echo GE pulse se- frequencies). An optimal filter size of 1616 was deter-
quence [9]. The two series of phase images were subtracted, mined empirically to minimize these effects.
and the phase difference unwrapped and divided by cDTE For this B 1 mapping, the following parameters were
(where c is the gyromagnetic ratio and DTE the interecho used: 8 ms sinc RF pulse, TR 4000 ms (in vivo) or 5000 ms
spacing) to yield a B 0 map. The following parameters were (postmortem), TE 7 ms, BW 50 kHz, FOV 1818 cm2,
used: 0.5 ms sinc RF pulse, TR 20 ms, TE 1.2 and 3.0 ms, MTX 25664 (in vivo) or 256 128 (postmortem),
nominal flip angle 108, BW 100 kHz, FOV 181818 cm3, ST 3 mm, slice gap 1 mm, axial or coronal acquisition
MTX 969696 (resulting in an isotropic resolution of plane, and NEX 1. Nominal flip angles of 608 and 1208
(1.9 mm)3, axial acquisition plane, NEX 1 and 50% were used since they produce maximal signal spread and,
oversampling of the slice encoding to reduce slice aliasing. therefore, most accurate flip angle measurements.
In addition, for two subjects, the B 0 field was also The flip angle and receive sensitivity maps were
numerically simulated at a resolution of (4 mm)3 using a finite interpolated and reformatted to yield maps corresponding
difference method and magnetic susceptibility distributions to the axial, coronal and sagittal GE and SE images. In
T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112 105

Fig. 1. Axial high-resolution GE (TR/TE 600/12 ms, nominal flip angle 238, MTX 1024384) (A) and SE (TR/TE 1500/40 ms, nominal flip angles 908/1808,
MTX 256256) (B) images of a healthy 32-year-old male through the inferior frontal lobes. While both GE and SE images provide excellent depiction of
anatomical structures in some regions (e.g., the vascular and microvascular branches, putamen, globus pallidus and splenium of the corpus callosum), other
regions are completely degraded by signal loss. On the SE image, the signal intensity in GM is higher than that in WM (solid arrow). The SE image shows
some chemical shift artifacts (dashed arrow, frequency encode direction anterior/posterior). The corresponding flip angle and receive sensitivity distributions
(in degrees and %, respectively) are shown as color maps (C, D) as well as contour lines superimposed on the SE image (E, F) to facilitate direct visual
correlation. In the left posterior cortex, both flip angle and receive sensitivity are very low, accounting for the low overall signal. Note that the accuracy of the
receive sensitivity map in that region is affected by the low SNR and low flip angle (see text). Also note some right/left asymmetry between the flip angle and
receive sensitivity maps in the central and anterior regions. The corresponding map of the experimental B 0 gradient along the slice direction (superior/inferior,
in ppm/cm) (G) shows severe B 0 inhomogeneity in the central inferior frontal region due to the sinuses below, accounting for the signal loss seen in that region
on the GE image but not on the SE image.

addition, to facilitate direct visual correlation between these flip angle and at 25%, 50% and 75% for the receive
maps and the anatomical images, contour lines were sensitivity) and superimposed on the corresponding
generated from these maps (at 608, 908 and 1208 for the SE images.
106 T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112

3. Results region of increased signal with loss of detail. In addition,


severe banding artifacts are seen in that region as well as
3.1. Signal variation and signal loss due to B 0 and
near the frontal sinus. Artifacts due to B 0 inhomogeneity
B 1 inhomogeneity
also affect axial GE images in the regions just superior to
Fig. 1 shows in vivo axial GE and SE images as well as the temporal bones (not shown), whereas SE images are
the corresponding maps of the flip angle, receive sensitivity significantly less affected. On both the GE and SE images,
and experimental B 0 gradient in the slice direction (since CT regions of signal loss are seen in the left posterior and right
images were not acquired for this subject, B 0 numerical frontal lobes and correspond to regions with low flip angle
simulations are not available). On the GE image, a central and/or receive sensitivity. Signal loss due to B 1 inhomoge-
region of signal loss is seen in the inferior frontal lobes neity is more severe on the SE image than on the GE image.
superior to the planum sphenoidale and corresponds to a Fig. 2 shows in vivo midsagittal GE and SE images as
region of high B 0 inhomogeneity (0.4 ppm/cm along the well as the corresponding maps of the flip angle, receive
slice direction as compared to 0.2 ppm/cm in most other sensitivity, experimental and numerical B 0 gradient in the
regions of the brain). There is also a surrounding large slice direction. Note the excellent depiction of vasculature in

Fig. 2. Midsagittal high-resolution GE (TR/TE 600/12 ms, nominal flip angle 238, MTX 1024512) (A) and SE (TR/TE 1500/70 ms, nominal flip angles 908/
1808, MTX 512256) (B) images of a healthy 53-year-old male. Both GE and SE images provide excellent depiction of the vasculature and the more anterior
brain anatomy, particularly the most anterior portion of the posterior fossa. However, there is severe signal loss in the posterior portions of the supra- and infra-
tentorial compartments. On the GE image, signal loss is also seen in the inferior frontal lobes, whereas the SE image better shows the optic nerve (yellow solid
arrow) as well as the CSF just superior to the planum sphenoidale (yellow dashed arrow). The SE image is markedly distorted by CSF flow artifacts. The
corresponding flip angle and receive sensitivity distributions (in degrees and %, respectively) are shown as color maps (C, D) as well as contour lines
superimposed on the SE image (E, F). Variability in local flip angle and receive sensitivity accounts for the signal loss in the posterior region as well as the
excellent depiction of the brainstem. Note that the receive sensitivity map is contaminated by residual proton density weighting, especially in the ventricles
(black arrow). The corresponding maps of the experimental (G) and numerical (H) B 0 gradient along the slice direction (right/left, in ppm/cm) show minimal B 0
inhomogeneity throughout the brain, although the experimental map shows some variability due to noise in the experimental data. In contrast to Fig. 1, the B 0
variations are smaller and the associated susceptibility artifacts less severe.
T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112 107

Fig. 3. Midsagittal high-resolution GE (TR/TE 600/12 ms, nominal flip angle 488, MTX 1024512) (A) and SE (TR/TE 1500/50 ms, nominal flip angles
908/1808, MTX 256256) (B) images of a postmortem unembalmed 57-year-old female with Huntington’s disease. The GE image shows exquisite
depiction of vascular and brain anatomy. On the other hand, the SE image better shows the region adjacent to the planum sphenoidale, clivus (yellow solid
arrow) and soft palate (yellow dashed arrow), which are all more severely affected by signal loss on the GE image. Regions with low signal correspond to
regions of high B 0 and/or B 1 inhomogeneity, as seen on the flip angle and receive sensitivity distributions (in degrees and %, respectively) shown as color
maps (C, D) and as contour lines superimposed on the SE image (E, F), as well as the map of the experimental B 0 gradient in the slice direction (right/left,
in ppm/cm) (G). Note that the depicted flip angle and receive sensitivity maps in the central region are incorrect. The true flip angle exceeds 1808 in that
region, leading to a signal void in the SE image (see text). Also note that the receive sensitivity map is contaminated by residual proton density weighting,
especially in the ventricles (black arrow). The greatest B 0 inhomogeneity is observed near the planum sphenoidale.

the interhemispheric fissure on both GE and SE images. On because the B 0 gradient in the left/right direction is not as
the GE image, regions of signal loss and banding artifacts large as the B 0 gradient in the inferior/superior direction.
are seen in the inferior frontal lobes superior to the sphenoid Imaging in the sagittal plane thus allows reduction of
sinus and posterior to the frontal sinus and are due to B 0 susceptibility artifacts in this region. Regions of signal loss
inhomogeneity. However, these susceptibility artifacts are are also seen in the occipital lobe and cerebellum on both
not as severe as on the axial image (Fig. 1A). This is GE and SE images, as well as in the parietal lobe on the SE
108 T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112

Fig. 4. In vivo coronal high-resolution GE (TR/TE 600/12 ms, nominal flip angle 238, MTX 1024512) (A) and SE (TR/TE 1500/70 ms, nominal flip angles
908/1808, MTX 512256) (B) images of the same subject depicted in Fig. 2. On the SE image, the signal intensity in GM is lower than that in WM (yellow
arrow). The SE image also shows chemical shift artifacts with the fat scalp displaced inferiorly over the skull, as well as flow artifacts from the CSF in the
prepontine cistern. The corresponding flip angle and receive sensitivity distributions (in degrees and %, respectively) are shown as color maps (C, D) as well as
contour lines superimposed on the SE image (E, F). The flip angles and receive sensitivity are very low in the lateral inferior temporal lobes, leading to very
poor image quality in these regions. Note that there are spatial differences between the flip angle and receive sensitivity distributions (black dashed arrows).
Also note that the receive sensitivity map is contaminated by residual proton density weighting, especially in the ventricles (black solid arrow). The
corresponding experimental (G) and numerical (H) maps of the B 0 gradient along the slice direction (anterior/posterior, in ppm/cm) have a good correlation and
show severe B 0 inhomogeneity in the inferior temporal lobes. Without B 1 mapping, one might assume that the signal loss in that region is entirely due to B 0
inhomogeneity from air spaces in the mastoid and middle ear. However, concurrent B 0 and B 1 mapping show that both B 0 and B 1 inhomogeneity contribute to
the signal loss.

image, and correspond to regions with low flip angle and/or on the GE image. Note the exquisite anatomical detail seen in
receive sensitivity. the posterior fossa. Furthermore, both veins and arteries
Fig. 3 shows postmortem midsagittal GE and SE images appear dark in this postmortem study because of the increased
as well as the corresponding maps of the flip angle, receive deoxyhemoglobin as compared to in vivo studies. Conversely
sensitivity and experimental B 0 gradient in the slice to the GE image, the SE image shows a central dark area in the
direction (since CT images were not acquired for this region with the largest flip angles. In this region, flip angles
subject, B 0 numerical simulations are not available). Again, are much larger than the optimal 908/1808, and the spins are
there is a good correlation between regions of signal loss boverflipped,Q thus causing the signal void. Due to B 0
and the measured B 0 and B 1 inhomogeneity. In this inhomogeneity, the margin of the inferior anterior cranial
particular case, the region with the largest flip angles and fossa is better seen on the SE image than on the GE image,
receive sensitivity coincides with the lateral ventricles, thus and the clivus and soft palate are visible on the SE image but
giving the rest of the brain a fairly homogeneous appearance not on the GE image. Note the difference in B 0 and B 1
T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112 109

inhomogeneity between Figs. 2 and 3 due to differences in the (i.e., low flip angle and/or receive sensitivity), motion and
subjects’ anatomy and head tilt [5]. signal loss due to intravoxel dephasing in regions with high
Fig. 4 shows in vivo coronal GE and SE images and the B 0 inhomogeneity (although to a lesser extent than the
corresponding maps of the flip angle, receive sensitivity, depicted GE images due to the short TE used for the
experimental and numerical B 0 gradient in the slice B 0 mapping). Visual inspection of Figs. 2G,H and 4G,H
direction. On the GE image, regions of signal loss in the shows that there is generally a good agreement between the
inferior temporal lobes near the mastoid and middle ear air experimental and numerical B 0 gradient maps. Small
spaces are due to both B 0 and B 1 inhomogeneity, as shown discrepancies may be attributed to the limitations inherent
by the B 0 gradient maps as well as the flip angle and receive to experimental B 0 mapping listed above. Furthermore,
sensitivity maps. Again, signal loss due to B 1 inhomogene- differences in head orientation between the MR images used
ity is more severe on the SE image than on the GE image for the experimental mapping and the CT images used for
and, in this particular case, affects the right side to a greater the numerical simulations may cause differences in the field
extent than the left. The combination of all these effects maps [5]. Such differences in head orientation also
leads to particularly low SNR in these regions. complicate direct quantitative comparison between the
3.2. Image contrast experimental and numerical B 0 maps, as an accurate 3D
image registration between the MR and CT images would
As was previously reported [1], GE images provide be required, which is beyond the scope of this paper.
excellent depiction of the vasculature. For in vivo studies, The method used to acquire flip angle maps from two
the veins appear dark due to the blood oxygen level- series of images with nominal flip angles of a 0 and 2a 0 is
dependent (BOLD) effect (Figs. 1A, 2A and 4A), whereas theoretically accurate for local flip angles a ranging between
for postmortem studies, both arteries and veins appear dark 08 and 1808. However, in practice, low image SNR prevents
because of the increase in deoxyhemoglobin [11] (Fig. 3A). accurate computation of flip angles close to 08 or 1808.
Cerebrospinal fluid (CSF) has the highest signal intensity Examples of problems with low flip angles are seen in the
on GE images, followed by gray matter (GM) and white
posterior cerebral hemispheres on axial images (Fig. 1C)
matter (WM), attesting to the predominant proton density
and in the inferior temporal lobes on coronal images
contrast resulting from the choice of acquisition parameters.
(Fig. 4C). If the local flip angle exceeds 1808, this method
Such contrast combined with high spatial resolution can
can no longer determine the correct flip angle (Fig. 3C).
lead to excellent delineation of the anatomy (e.g., in the
Finally, flip angle maps may become inaccurate in regions
cerebellum; Fig. 3A), provided flip angles are not too low.
with high B 0 inhomogeneity (e.g., superior to the sphenoid
Due to T 2* effects, structures with high iron content have
sinus; Fig. 1C). To address these limitations, other studies
the lowest signal intensity with excellent delineation from
were performed in which images were acquired using up to
the surrounding brain (e.g., the red nuclei and substantia
18 different nominal flip angles and then fitted to a sine
nigra; Fig. 4A). Vascular structures are sometimes seen on
curve pixel-by-pixel to determine the flip angle map
SE images, but are generally not as well depicted as on GE
images. On SE images with TE b 50 ms, the signal intensity [14 –16]. This approach generally decreases the noise in
in GM is higher than that in WM (Fig. 1B), whereas this the flip angle maps; however, the overall pattern is
contrast is inverted for TE N70 ms (Fig. 4B). This behavior independent of the number of nominal flip angles used.
is different from the familiar appearance seen on T 2- More specifically, signal loss in regions with a flip angle
weighted images acquired at 1.5 T and was observed in a near zero cannot be recovered.
number of in vivo and postmortem studies [12,13]. Errors in the local flip angle measurement propagate into
the receive sensitivity maps (e.g., in the left posterior cortex
3.3. Other artifacts in Fig. 1D). Furthermore, the latter are contaminated by
The in vivo images, particularly the SE images, are also residual proton density weighting, especially in large
affected by severe artifacts in the phase encode direction anatomical structures such as the ventricles. Nevertheless,
resulting from pulsating CSF flow. These artifacts are most overall patterns of receive sensitivity can be assessed and
prominent at the posterior fossa level (Figs. 2B and 4B). show differences with the flip angle distributions. Finally,
Further studies are needed to evaluate the effectiveness of while the flip angle and receive sensitivity patterns are
flow compensation methods. Chemical shift artifacts are similar among subjects, differences arise from variable
also more pronounced in SE images, misplacing the signal subject anatomy and/or RF coil tuning.
from the scalp in the frequency encode direction poste-
riorly (Fig. 1B) or inferiorly (Figs. 2B – 4B). In GE images, 4. Discussion
fat/water intravoxel dephasing nulls the fat signal, thus
avoiding pronounced chemical shift artifacts. The examples shown in this paper clearly demonstrate
that ultra-high field MRI is plagued by substantial signal
3.4. Accuracy of B 0 and B 1 mapping
variation and signal loss. Correlation between B 0 and B 1
Unlike numerical simulations, experimental B 0 mapping mapping and the routine magnitude images shows that these
is affected by noise and artifacts due to B 1 inhomogeneity effects can be fully explained by B 0 and/or B 1 inhomoge-
110 T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112

neity. Because there are different sources of signal variation, subject’s anatomy (i.e., the spatial distribution of tissues
B 0 and B 1 mapping add confidence and give important with different dielectric and conductive properties), the RF
insights into the interpretation of the magnitude images well coil design, as well as the coil tuning, but the overall
beyond what is possible from assessing signal variations patterns are relatively similar. Furthermore, the experimental
alone. For example, signal loss in the lateral inferior maps are similar to numerical simulations based on realistic
temporal lobes is due to both B 0 and B 1 inhomogeneity head models [17], and differences between the flip angle
(Fig. 4). Without B 1 mapping, one might assume that the and receive sensitivity maps follow numerical simulations
signal loss is only due to B 0 inhomogeneity. of the clockwise- and counterclockwise-rotating compo-
As expected, severe signal loss due to B 0 inhomogeneity is nents of the B 1 field [18,19]. Such differences are also
seen on GE images. The regions affected are relatively theoretically expected [20,21] and were experimentally
restricted to the vicinity of air/tissue interfaces (approximate- confirmed in other studies [14 –16].
ly within 1 cm) and are most prominent in the inferior frontal Spin echo images are more severely affected by signal
and temporal lobes, as predicted by the experimental and loss due to B 1 inhomogeneity than GE images. This is
numerical B 0 gradient maps. These artifacts are more severe because the signal intensity in SE images (for TR~T 1) is
than what is routinely observed at lower field strength. For given by:
example, the effects of the skull base air spaces extend well 
S ð x; yÞ ¼ qð x; yÞrð x; yÞsin3 að x; yÞ
into the brain. Susceptibility artifacts can be somewhat
reduced by optimal selection of the acquisition plane [5].     
 1  exp  TR=T 1 ðx; yÞ gexp TE=T 2 ðx; yÞ ; ð6Þ
This may be an essential strategy at ultra-high field strength,
even though it limits certain regions to be evaluated using whereas the signal intensity in GE images is given by:
only certain acquisition planes. On the other hand, SE images
are significantly less affected by signal loss due to B 0 S ð x; yÞ ¼ qð x; yÞrð x; yÞsin½að x; yÞ
inhomogeneity than GE images, but the effect is still more  
1  exp  TR=T 1 ðx; yÞ
severe than what is seen on SE images acquired at lower field     
1  cos að x; yÞ exp  TR=T 1 ðx; yÞ
strength. For example, in Fig. 3, the central nasal cavity
 
structures are not seen at all on both the SE and GE images.  exp  TE=T 2 4ðx; yÞ : ð7Þ
Signal misregistration and geometric distortions may occur in
both GE and SE imaging, and such artifacts may be present in Furthermore, GE images were acquired with nominal flip
the images shown here, but are not very obvious without angles at or below the Ernst angle where the signal curve
direct comparison to images acquired at lower field strength [Eq. (7)] is fairly flat, so that the GE signal intensity does
or using different modalities such as CT. not vary strongly with flip angle, except in regions with very
In addition to causing signal loss and geometric small flip angles ( b108). In the examples shown here, local
distortions, B 0 inhomogeneity also influences the effective flip angles range from 08 to 308 (Figs. 1A, 2A and 4A) and
excitation field [7]: 208 to 908 (Fig. 3A), whereas the Ernst angle is 468 and 508
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi for GM and WM, respectively, assuming TR =600 ms and
Beff ¼ B21 þ DB20 ; ð5Þ T 1 = 1650 and 1345 ms, respectively [22]. Conversely, the
SE signal intensity not only decreases in regions with small
and thus the effective flip angle. In Eq. (5), DB 0 represents flip angles, but also in regions with flip angles larger than
the relative static magnetic field with respect to the field 908/1808 where the spins are overflipped (Fig. 3B). The
produced by the magnet. For the 8-ms sinc RF pulses used predicted signal variability with flip angle correlates well
in this work, a 908 flip angle corresponds to a field strength with measured image SNR. For example, on the postmortem
B 1 of 4 AT. Experimental and numerical B 0 maps show sagittal GE image acquired with TR/TE 600/12 ms and a
that the magnitude of DB 0 is smaller than 0.5 ppm or nominal flip angle of 488 (Fig. 3A), we selected one region
4 AT in most regions of the brain, but can reach up to in the parietal lobe and one region in the occipital lobe
3 ppm or 24 AT in the vicinity of air/tissue interfaces, where the true flip angles are 578 and 258, respectively, and
therefore significantly altering the effective flip angle and the receive sensitivity is 50% and 49%. On the
exaggerating the signal loss and banding artifacts in corresponding SE image acquired with nominal flip angles
these regions. of 908/1808 (Fig. 3B), the true flip angles are 1078/2148 and
While susceptibility artifacts occur predominantly in GE 478/948, respectively. From these values, we calculated the
images and are mostly restricted to the vicinity of air/tissue signal intensity ratios for GM/WM in these two regions
interfaces, both GE and SE images show substantial signal using Eq. (6) for the SE image and Eq. (7) for the GE image.
variability due to B 1 inhomogeneity. Regions of low signal We assumed a proton density of 0.80 and 0.65 for GM and
correspond to regions with low flip angles and/or receive WM, respectively. We neglected the T 2 and T 2* terms for
sensitivity. The flip angle and receive sensitivity maps show this estimation, even though T 2 and T 2* effects contribute to
nonintuitive complex 3D spatial distributions. There are image contrast. The results are shown in Table 1 and
notable differences between subjects depending on the correlate well with the measured SNR ratios for GM/WM in
T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112 111

Table 1
Comparison between calculated signal intensity ratios and measured SNR ratios for GM/WM regions in the parietal and occipital lobes on corresponding GE
and SE images
GE image SE image
Calculated signal intensity Measured SNR Calculated signal intensity Measured SNR
GM/WM ratio (parietal lobe) 0.17/0.15 = 1.17 37.4/30.7 = 1.22 0.21/0.18 = 1.15 18.7/15.7 = 1.19
GM/WM ratio (occipital lobe) 0.13/0.12 = 1.14 20.1/19.0 = 1.06 0.087/0.089 = 0.98 11.6/12.1 = 0.95

the same regions. SNR variability with flip angle and for a specific experimental case (i.e., same subject anatomy,
receive sensitivity is similar on the in vivo images. Note, RF coil tuning, etc.). Consequently, numerical B 0 maps will
however, that SNR differences also scale with voxel size, likely play an important role in the conceptual evaluation of
matrix size, receiver bandwidth and NEX, as usual, and that susceptibility artifact correction methods as well as the
in vivo SNR may be further decreased by motion. development of those methods that do not rely on specific
The flip angle variability affects not only the SNR but B 0 maps (e.g., gradient compensation or passive shimming),
also the image contrast. However, because of the choice of whereas experimental B 0 maps will likely be required for
acquisition parameters, this effect is minimal on the GE and the development of other methods such as active shimming
SE images shown here. The GE images have mixed T 2*/ or post-processing.
proton density weighting, with relatively little dependence Experimental B 1 mapping requiring a long TR (i.e.,
of image contrast on T 1, and thus flip angle. Similarly, the TR H T 1) is too slow for clinical applications. However,
SE images are proton density- and T 2-weighted, so that the knowledge of the overall 3D characteristics of the B 1 field
image contrast varies minimally with flip angle. Conversely, helps select optimal flip angles. For example, if a pathology
the image contrast would be more severely affected by flip is suspected in the temporal lobe, the nominal flip angle can
angle variability on T 1-weighted images. be determined using a voxel-selective stimulated echo pulse
Finally, it should be noted that B 1 inhomogeneity and the sequence with the voxel placed in that region. On the other
resultant spatial variability of the flip angle become hand, B 1 mapping is crucial for the understanding of B 1
progressively more problematic in multi-pulse sequences. inhomogeneity patterns and for quantitative studies such as
For example, the inversion recovery spin echo sequence T 1 and T 2 measurements [22,29,30]. Experimental B 1
shows very complex signal patterns and artifacts, especially mapping will also play an important role in ultra-high field
when stimulated and higher order echoes are incompletely RF coil design and for studying the effects of coil tuning.
suppressed [22]. Furthermore, spatially selective presatura- Methods for correction of B 1 inhomogeneity include
tion pulses for suppression of vascular flow artifacts would simulations of TEM coils with multi-port RF excitation
be compromised. For example, presaturation pulses in the using numerically optimized amplitudes and phases to
region inferior to the pons where severe flow artifacts are achieve more homogeneous excitation profiles [17]. Anoth-
observed (Figs. 2B and 4B) would be inefficient, because er approach involves the use of tailored RF pulses for
local flip angles are very low in this region even though it is suppression of B 1 inhomogeneity [31]. Both methods will
within the RF coil. require knowledge of the B 1 transmit and receive field.
The discussion above demonstrates the utility of B 0 and In conclusion, artifacts due to both B 0 and B 1
B 1 mapping for the assessment of image artifacts in ultra- inhomogeneity are severe in ultra-high field MRI, and
high field MRI of the human brain. Because of the severity experimental and/or numerical mapping of the B 0 and B 1
of signal and contrast variability, correction methods for inhomogeneity is important in identifying their origin.
both B 0 and B 1 inhomogeneity are needed to fully take The problems due to severe B 0 and B 1 inhomogeneity de-
advantage of the increased SNR available at ultra-high scribed in this paper can be limited by focusing on localized
field strength. studies. This approach has been very successfully em-
A variety of susceptibility artifact correction methods ployed in localized spectroscopy [3] and fMRI in the
have been proposed, including gradient compensation or occipital lobe [4]. However, if ultra-high field MRI is to
z-shim techniques [23,24], tailored RF pulses [25], active become useful for a broad range of biomedical applications
[26] and passive [27] shimming, and post-processing [28]. including survey exams covering most of the head
Implementation and optimization of all of these methods structures in one exam, correction methods for B 0 and B 1
require knowledge of the B 0 field. B 0 numerical simulations inhomogeneity need to be developed. Detailed character-
and experimental mapping both have advantages and ization of the effects of B 0 and B 1 inhomogeneity is an
limitations and are therefore complementary. Numerical important first step in that direction.
simulations have the advantage of being more flexible and
unaffected by noise and artifacts due to B 1 inhomogeneity
Acknowledgments
(i.e., low flip angle and/or receive sensitivity), motion or
intravoxel dephasing. On the other hand, experimental B 0 This research was supported by funding from The
mapping is fast and can provide more detailed information Department of Radiology of the Ohio State University. We
112 T.-K. Truong et al. / Magnetic Resonance Imaging 24 (2006) 103 – 112

would like to thank Ryan Gilbert for optimizing the RF coil RF coils for high field imaging. Proc 12th Intl Soc Magn Reson Med.
tuning for these studies, as well as Dr. Roger Dashner for his Kyoto (Japan)7 ISMRM; 2004. p. 488.
[15] Ibrahim TS, Mitchell C, Schmalbrock P, Lee R, Chakeres DW.
contribution to the postmortem studies. Electromagnetic perspective on the operation of RF coils at 1.5–11.7
tesla. Magn Reson Med 2005;54:683 – 90.
References [16] Mitchell CA, Ibrahim TS, Chakeres DW, Gilbert RD, Schmalbrock
P. B 1 field mapping in-vivo for three TEM coil modes at 8 tesla.
[1] Abduljalil AM, Schmalbrock P, Novak V, Chakeres DW. Enhanced Proc 13th Intl Soc Magn Reson Med. Miami (FL)7 ISMRM; 2005.
gray and white matter contrast of phase susceptibility-weighted p. 928.
images in ultra-high-field magnetic resonance imaging. J Magn Reson [17] Ibrahim TS, Lee R, Baertlein BA, Abduljalil AM, Zhu H, Robitaille
Imaging 2003;18:284 – 90. PML. Effect of RF coil excitation on field inhomogeneity at ultra high
[2] Christoforidis GA, Bourekas EC, Baujan M, Abduljalil AM, Kangarlu fields: a field optimized TEM resonator. Magn Reson Imaging 2001;
A, Spigos DG, et al. High resolution MRI of the deep brain vascular 19:1339 – 47.
anatomy at 8 tesla: susceptibility-based enhancement of the venous [18] Hoult DI. The principle of reciprocity in signal strength calculations —
structures. J Comput Assist Tomogr 1999;23:857 – 66. a mathematical guide. Concepts Magn Reson 2000;12:173 – 87.
[3] Lei H, Zhu XH, Zhang XL, Ugurbil K, Chen W. In vivo P-31 [19] Ibrahm TS, Mitchell C, Gilbert R, Schmalbrock PM, Chakeres
magnetic resonance spectroscopy of the human brain at 7 T: an initial DW. The TEM resonator modes: a study at high field MRI. Proc
experience. Magn Reson Med 2003;49:199 – 205. 11th Intl Soc Mag Reson Med. Toronto (Canada)7 ISMRM; 2003.
[4] Yacoub E, Duong TQ, Van de Moortele PF, Lindquist M, Adriany G, p. 2394.
Kim S-G, et al. Spin-echo fMRI in humans using high spatial [20] Ibrahim TS. A new approach for an analytical formulation of the
resolutions and high magnetic fields. Magn Reson Med 2003; NMR signal using the principle of reciprocity. Proc 12th Intl Soc
49:655 – 64. Magn Reson Med. Kyoto (Japan)7 ISMRM; 2004. p. 1571.
[5] Truong T-K, Clymer BD, Chakeres DW, Schmalbrock P. Three- [21] Ibrahim TS. Analytical approach to the MR signal. Magn Reson Med
dimensional numerical simulations of susceptibility-induced magnetic 2005;54:677 – 82.
field inhomogeneities in the human head. Magn Reson Imaging 2002; [22] Mitchell CA, Truong T-K, Ibrahim TS, Schmalbrock P. Accurate T 1
20:759 – 70. measurements at 8 tesla despite radiofrequency inhomogeneity. Proc
[6] Ibrahim TS, Abduljalil AM, Baertlein BA, Lee R, Robitaille PML. 11th Intl Soc Mag Reson Med. Toronto (Canada)7 ISMRM; 2003.
Analysis of B 1 field profiles and SAR values for multi-strut transverse p. 1089.
electromagnetic RF coils in high field MRI applications. Phys Med [23] Yang QX, Williams GD, Demeure RJ, Mosher TJ, Smith MB.
Biol 2001;46:2545 – 55. Removal of local field gradient artifacts in T 2*-weighted images at
[7] Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic high fields by gradient-echo slice excitation profile imaging. Magn
resonance imaging: physical principles and sequence design. New Reson Med 1998;39:402 – 9.
York (NY)7 Wiley; 1999. p. 47. [24] Glover GH. 3D z-shim method for reduction of susceptibility effects
[8] Vaughan JT, Hetherington HP, Out JO, Pan JW, Pohost GM. High- in BOLD fMRI. Magn Reson Med 1999;42:290 – 9.
frequency volume coils for clinical NMR imaging and spectroscopy. [25] Stenger VA, Boada FE, Noll DC. Multishot 3D slice-select tailored RF
Magn Reson Med 1994;32:206 – 18. pulses for MRI. Magn Reson Med 2002;48:157 – 65.
[9] Truong T-K, Dashner RA, Chakeres DW, Schmalbrock P. Comparison [26] Jesmanowicz A, Starewicz P, Hyde JS. Determination of shims
between B 0 experimental mapping techniques and numerical simu- needed for correction of tissue susceptibility effects in fMRI. Proc 8th
lations at 8 T. Proc 11th Intl Soc Mag Reson Med. Toronto (Canada)7 Intl Soc Mag Reson Med. Denver (Colo)7 ISMRM; 2000. p. 1378.
ISMRM; 2003. p. 1048. [27] Wilson JL, Jenkinson M, Jezzard P. Optimization of static field
[10] Insko EK, Bolinger L. Mapping of the radiofrequency field. J Magn homogeneity in human brain using diamagnetic passive shims. Magn
Reson Ser A 1993;104:78 – 82. Reson Med 2002;48:906 – 14.
[11] Dashner RA, Chakeres DW, Kangarlu A, Schmalbrock P, Christoforidis [28] Kadah YM, Hu X. Algebraic reconstruction for magnetic resonance
GA, DePhilip RM. MR imaging visualization of the cerebral imaging under B 0 inhomogeneity. IEEE Trans Med Imag 1998;
microvasculature: a comparison of live and postmortem studies at 8 T. 17:362 – 70.
Am J Neuroradiol 2003;24:1881 – 4. [29] Schmalbrock P, Ibrahim TS, Mitchell CA. Comparison of methods for
[12] Whitaker CDS, Truong T-K, Dashner RA, et al. Assessment of iron imaging T 1 relaxation for 8 T whole-body MRI. Proc 10th Intl Soc
with 8 T MRI T 2 imaging in human brain. Proc 12th Intl Soc Mag Mag Reson Med. Honolulu (HI)7 ISMRM; 2002. p. 2293.
Reson Med. Kyoto (Japan)7 ISMRM; 2004. p. 186. [30] Whitaker CDS, Truong T-K, Ibrahim TS, Schmalbrock P. Accuracy of
[13] Truong T-K, Whitaker CDS, Schmalbrock P. Comparison between 8 T T 2 measurements: evaluation of errors due to RF inhomogeneity.
Hahn spin echo and Gradient Echo Sampling of the Spin Echo Proc 11th Intl Soc Mag Reson Med. Toronto (Canada)7 ISMRM,
(GESSE) T 2 measurements at ultra-high field strength. Proc 12th Intl 2003. p. 1098.
Soc Mag Reson Med. Kyoto (Japan)7 ISMRM; 2004. p. 2318. [31] Saekho S, Boada FE, Noll DC, Stenger VA. B 1 inhomogeneity
[14] Ibrahim TS, Mitchell C, Schmalbrock P, Chakeres DW. Computa- compensation using 3D tailored RF pulses. Proc 11th Intl Soc Mag
tional and experimental analysis of electromagnetic fields induced by Reson Med. Toronto (Canada)7 ISMRM; 2003. p. 717.

You might also like