Component
TH
Update Instructions TH004/13/S
Installation of SW Release 4.2.110
Harry Wassermann
TH004/13
Siemens
Safety Update
Perform: Within 03 months
Intranet download available: Q Yes H No
Remote update: Q Yes, w/o CSE Q Yes, with CSE H No
(Category 2, 4, 5) (Category 1, 3)
Remark: n.a.
Scope
2013
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2013“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA
and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are not entitled to use this documentation
without prior written authority.
Systems/Products Affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
This update applies to the syngo RT Therapist systems listed below. . . . . . . . . . . . . 4
Reason for the Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Special tools / documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Material Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Return of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
WEEE statement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 _______ Update_________________________________________________________ 8
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1. Version check for RT Therapist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2. Back up RT Therapist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3. Update processing for Therapist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4. Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Customer Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Customer application training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
The current information regarding the affected systems can be found on the Med CS Intra-
net / Update Handling Database. The list of affected sites is included in the modification
release letter.
This update is intended to provide a SW update for the syngo RT Therapist / syngo RT
Therapist Connect/MOSAIQ products, which are running SW version 4.2.108. This SW
update contains fixes for the safety-related issues which are described in the customer
safety advisory notices distributed as TH009/12/S (position verification beams in an abso-
lute setup) and TH018/12/S (immediate resumption of an interrupted beam). The solution
of TH009/12/S consists of a preventive warning message and an Addendum to the User
Manual that also addresses several document improvements.
Prerequisites 0
The CSE should be familiar with the ARTISTE linear accelerator along with the CC node
within the control console cabinet, as well as the RT Therapist Connect.
Verify the following:
Check the Med CS intranet or extranet for the latest version of this document (For Service
-> Product Information) or contact your local Siemens office if you do not have access to
the Siemens intranet or extranet.
• Standard CSE tool kit
• Service laptop computer
• syngo RT Therapist VB 10 (available on the CS Intranet TH11-400.815.xx.xx.xx)
• Blank DVD-R discs, 4.7 GB, as needed
• External USB hard disk
Ordering information
The following update kits must be ordered from CSML (SAP Distribution Channel, plant
2050/6050) or BU Logistics (factory BU) (only with an order via BU Logistics).
Please note that different order numbers apply for the different languages in which the kits
are available: English, German, Japanese, Polish, and Swedish, as well as an "interna-
tional language" kit.
Return of Parts 0
There are no parts to return. Except for dead on arrival, return parts will be handled in
accordance with the ARTD "Principles of Return Goods Management" using the parts
return form.
Please return them using the material number of the update kit.
Returns must be packaged correctly.
No credit will be issued for returns that have been damaged as a result of incorrect han-
dling.
Safety disclaimer
Service personnel must be aware of potential safety hazards and take all necessary mea-
sures to prevent them. Before installation, refer to the safety precautions section of the cur-
rent accelerator installation manual, which describes the hazards associated with servicing
the accelerator.
WEEE statement 0
1. Log in to Local Service via the syngo main menu Options>Service>Local Service.
2. Select Configuration>Users and uncheck Enable Autologin.
2. Back up RT Therapist 0
• References
- syngo RT Therapist Connect / MOSAIQ TH11-400.815.09.xx.02
- syngo RT Therapist 4.2 (VB10) Startup TH11-400.815.12.xx.02
• Before starting, perform a full image backup of the system partition.
NOTE Connect the external USB drive after the reboot from the
COBR CD has completed.
b) For SOURCE, select Internal Disk -> select DISK0 [MED_SYSTEM] -> (C:).
c) For DESTINATION, select CD/DVD writer -> and CD/DVD device or USB hard
disk.
d) Click OK.
e) If necessary, insert another blank DVD.
f) Reboot system after backup is complete.
5. Click the hyperlink Click here to install to start the install process.
- If an Internet Explorer message appears first, click Yes.
NOTE The system will appear to freeze up. The autoscript contin-
ues installing in the background.
The upgrade takes about 20 to 30 minutes.
8. After the patch installation is complete, click the upper "here" in the Readme.htm win-
dow to open the install log file for further information.
- Scroll down to the end of the log file and verify 4.2.110 Patch installation script fin-
ished.
15. Click Next and confirm the upcoming System Type message with OK.
16. Reboot the RTT
4. Final steps 0
1. Update the revision level, the user documents (if applicable), and the technical docu-
mentation.
2. Fill out and make a copy of the attached Completion Protocol and file it in the corre-
sponding binder.
3. The update has to be reported by authorized employees, either via:
- automated data transfer (only in countries connected to SAP P42 via an Update Han-
dling Interface)
- The “ Update Handling “ on the Siemens Healthcare Intranet/Extranet -> Customer
Services application (for countries NOT connected to SAP P42 via the Update Han-
dling Interface)
- the Completion Protocol (for countries that do not have one of the connections indi-
cated above)
4. This update includes IVKs. Changes in the IVK structure must be reported to the man-
agement system group to update the installed volume.
5. This update resolves the problem described in the letters attached to TH009/12/S and
TH018/12/S. The letters are now invalid, so the following steps need to be performed:
- Open the Digital Linear Accelerator System Owner Manual, chapter “Safety Advisory
Letters” and find the letters related to the UIs TH009/12/S and TH018/12/S
"INVALID, problem is fixed with UI TH004/13/S," and then sign this text.
n.a.
n.a.
This update is intended to provide a SW update for syngo RT Therapist / syngo RT Thera-
pist Connect/MOSAIQ products that are running SW version 4.2.108. This SW update con-
tains fixes for some safety-related issues that are described in the customer safety advisory
notices distributed as TH009/12/S (position verification beams in an absolute setup) and
TH018/12/S (immediate resumption of an interrupted beam). The solution of TH009/12/S
consists of a preventive warning message and an addendum to the user manual that also
addresses several document improvements. In addition, complaint 12-07265-OCS-0353
has been addressed.
Protocol 0
Customer: ...........................................................................................................
Notification: ....................................
Remark: ...........................................................................................................
NOTE After the update has been completed, please make a copy of
this page, fill it out, and file the protocol in the corresponding
binder.
Options: