Professional Documents
Culture Documents
SOP 010 - A PSA Cone Multi-Pass Procedure For CVD Applications KP Revised
SOP 010 - A PSA Cone Multi-Pass Procedure For CVD Applications KP Revised
Revision 3
DRAFT
Updated: Feb 2, 2009 Written By: Toni-Lynn Card
1.0 Scope
1.1 The Knelson MD3 Concentrator (KC-MD3) is used to determine the
amenability of an ore type to gravity concentration.
1.2 Multi-pass tests are used to generate mass pull/recovery profiles,
more commonly for CVD amenability testing. The number of
passes is dependent on numerous variables including client
requests, sample type and composition, particle size, etc. Typically,
four to five passes are used or until the target mineral is exhausted.
2.0 Summary
2.1 Multi-pass test program may differ, depending on the clients’
needs. Generally, the procedure will follow these laboratory
processing steps
2.1.1 Client inquiry
2.1.2 Determine the test program
2.1.3 Sample receiving
2.1.4 Prepare sample to the appropriate processing size (as
required).
2.1.5 Perform test program as per flow sheet/proposal.
2.1.6 Perform calculations including mass balances, as required.
2.1.7 Compose report (if required).
2.1.8 Sample storage
2.1.9 See below for the flow sheet. This is to be used as a
guideline for processing. The number of stages may differ.
With the low yield cone, panning the concentrates is
required.
KC Tails 2
KC Tails 3
KC Tails 4
Assay for target
element
Tails (Stored) 3
months
Tails Disposed
6.0 Procedure
6.1 Sample Preparation- perform steps as required.
6.1.1 Please refer to SOP 004: Sample Preparation, if sample
preparation is required.
6.1.2 Split sample into the appropriate size for testing purposes.
6.1.2.1 Multi-pass samples requires 1kg. The mass as
well as the particle size (P80) used for testing
should be determined with the Territory Manager
or Lab Supervisor/Manager and the client as
needed.
6.5 References
6.5.1 General Operating Guidelines and Basic Laboratory Test
Procedures for a KC-MD3 Knelson Concentrator. Rev.
October’99.