Professional Documents
Culture Documents
Troubleshooting
Compressed air leakage noise Defective pneumatic brake Switch off the valve to shut down com-
pressed air supply and contact Maquet
technical
service personnel or authorized service
partner
The distribution module not pivoting Distribution module overloaded Check and if necessary lighten the distribu-
properly tion module
The distribution module is knocking Stoppers set incorrectly Set / adjust the stoppers
against surrounding objects
Paint damaged or flaking Ceiling supply units knocking against Set / adjust the stoppers
surrounding objects
Touch up any chipped paint.
LED lights are not lighting on Electrical defect of lights Contact Maquet technical service person-
nel or authorized service partner
Drawer light’s brightness is Power of battery is low. Replace batteries
decreased or not on
Moisture in the system Gas leaks or problem with gas seal Contact the gas supplier
The image is not showing on screen Interface connection is loosen. Check and tighten the interfaces
or not clear connection
Interfaces of optical fiber are swapped Swap the interfaces of optical fiber and
to plug then plug in