Professional Documents
Culture Documents
SECTION A
Introduction: Challenges Using Conventional
Techniques for Volatile Compound Analysis
SECTION B
Routine Laboratory Applications of SIFT-MS
SECTION C
Applying Routine Analysis Procedures to SIFT-MS
SECTION D
Development and Validation of SIFT-MS Methods
SECTION E
Simple, Comprehensive and Broad-Spectrum
SIFT-MS Analysis
SECTION F
Summary: SIFT-MS and Routine Analysis
Sponsored by
This custom ebook is sponsored by Syft Technologies, Anatune, and Gerstel and presented in partnership with LCGC.
OVERVIEW
T
esting laboratories frequently face sample preparation and throughput
challenges for volatile compound analyses when using conventional
chromatographic techniques. This is especially the case for reactive and
polar species of low molecular weight, and for samples with significant
water content. The adoption of a newer direct mass spectrometry technique, such
as selected ion flow tube mass spectrometry (SIFT-MS), can resolve these issues.
This ebook demonstrates how readily routine analysis procedures developed for
conventional methods can be applied to SIFT-MS. The ebook:
• Introduces the SIFT-MS technique, its automation, and the throughput benefits
that its adoption confers for testing laboratories;
• Outlines how routine analytical procedures—such as sample scheduling,
automated calibration and the use of standard additions—can be applied to
SIFT-MS;
• Describes three detailed case studies that demonstrate successful application
of standard method validation procedures to conventional and unconventional
SIFT-MS methods; and
• Illustrates how SIFT-MS simplifies the analysis of important chromatographically
challenging volatiles, such as formaldehyde, volatile fatty acids, organosulfur
compounds and amines.
SECTION D:
SECTION A:
SECTION B:
Routine
SECTION C:
Analysis
Method
SECTION E:
Comprehensive
SECTION F: Overview
Intro to SIFT-MS Development, Summary
Applications Procedures Analysis
Validation
S P ONS OR E D C ONT E NT
SECTION A
Challenges with
Conventional
Volatile Compound
Analysis
Techniques
Darren Baker/stock.adobe.com
T
his section first highlights 1. Challenges Facing Laboratories
the key challenges faced by and Contract Research
routine testing laboratories and Organizations (CROs)
contract research organizations Competition in the routine testing
(CROs) using conventional techniques for laboratory and CRO markets, coupled
volatile compound analysis. Direct mass with increased awareness of health
spectrometry, in the form of selected and environmental impacts of VOCs,
ion flow tube mass spectrometry (SIFT- mean that there are more samples than
MS), is then introduced. Automated ever to be analyzed and the cost per
sample must be reduced. However,
SIFT-MS addresses the challenges,
conventional chromatographic methods
providing high-throughput analysis of
for routine VOC analysis—predominantly
volatile organic compounds (VOCs) and gas chromatography (GC)-based, but
many inorganic gases to sub-part-per- also utilizing liquid chromatography
billion concentrations (by volume; ppbV) (LC) for certain analytes—are limited in
without preconcentration, derivatization their ability to accommodate increased
or drying. capacity per instrument (Figure 1). With
only the chromatographic techniques in
Subsequent sections of this ebook build
a lab’s instrument portfolio, the solution
on these basic advantages of SIFT-MS, is to hire more staff, purchase or lease
showing practical, proven strategies for more instruments and accessories,
adopting automated SIFT-MS in routine and expand the lab’s floor area to
analysis workflows. accommodate them.
Ammonia
Benzene
Formaldehyde
Hydrogen sulfide
Ethanol
Pyridine
Methyl bromide
Carbon dioxide
Hydrogen chloride
Sulfur dioxide
“fingerprints” of each compound simplifies analysis because:
can be used to identify and quantify 1. Discrimination imposed by the
the compound. chromatographic column is
In Figure 7, the same mixture is irrelevant
analyzed using the three positively 2. USCI ionizes the compound directly
charged SIFT-MS reagent ions. By in the gas phase, without any need
applying multiple, rapidly switchable for derivatization (e.g. aldehydes
reagent ions that ionize compounds and volatile fatty acids are analyzed
in different ways, SIFT-MS can resolve directly)
all 15 compounds in real-time without 3. SIFT-MS is robust to moisture,* so
using chromatography. drying of samples is unnecessary
Chromatography-free USCI-SIFT-MS 4. Lower inlet temperatures are used,
Figure 6: 70-eV EI-MS (a) without and (b) with gas chromatography.
The hypothetical 15-component sample is derived from the US EPA
Compendium Method TO-15 and generated from mass spectra in the
NIST library (http://webbook.nist.gov/chemistry).
* SIFT-MS is robust to moisture both in the sample introduction phase (slow, continuous sample introduction eliminates the issue caused in GC inlets due to
the high coefficient of expansion of water), and in the ionization phase (USCI accommodates the presence of water much better than EI ionization).
ashtproductions/stock.adobe.com
S P ONS OR E D C ONT E NT
SECTION B
Routine Laboratory
Applications of
SIFT-MS
megaflopp/stock.adobe.com
T
his section surveys various [PET]) using static headspace-SIFT-MS
applications to demonstrate (Figure 9). These samples were analyzed
how automated SIFT-MS at a rate of 1 sample per minute (or 60
revolutionizes routine analysis samples per hour).
of volatile compounds by resolving the
throughput challenges that confront b. Multiple headspace extraction (MHE)
chromatographic techniques. Analysis
MHE enables matrix-independent
of volatiles in diverse matrices, from air
concentrations of VOC content in solids to
to the headspace of polymers, soil, and
water, is illustrated. be obtained. SIFT-MS revolutionizes the
3. Continuous Headspace
Analysis (CHA)
Chromatographic techniques are highly
impractical probes of dynamic processes,
because they require that multiple
instantaneous samples be taken, followed
by prolonged analysis of each sample. With
delay while the chromatographic column
SIFT-MS, monitoring dynamic processes is
separates the compounds in time.
simple due to the high time resolution, high
Figure 12a illustrates residual solvent sensitivity, and comprehensive analysis via
analysis of powdered paracetamol tablets multiple rapidly switchable reagent ions.
(1 g incubated for 15 minutes at 60 °C) When coupled with automation, samples
using a static headspace approach. can be run 24 hours/day, speeding up
Applying MHE, however, enables direct research and development or QA testing.
analysis of the total solvent content in the This application is illustrated in
tablets. Figure 12b shows MHE results Figure 13, where the release of
7. Automated Thermal
Desorption Tube Analysis
SIFT-MS combined with GERSTEL
automation and the GERSTEL TD 3.5+
headspace measurements (Figure 16). thermal desorber greatly simplifies
analysis of thermally desorbed or
With current automation technology,
thermally extracted volatile compounds.
which was optimized for GC/MS, we see Figure 18 illustrates the fundamental
a 3.5-fold increase in sample throughput difference for thermal desorption-SIFT-
when compared to the standard MS (TD-SIFT-MS) analysis versus TD-GC/
S P ONS OR E D C ONT E NT
SECTION C
Applying
Routine Analysis
Procedures to
SIFT-MS
H
akf/stock.adobe.com
and permeation tubes being most and dimethylsulfoxide (DMSO) have all
commonly used), automation enables been utilized successfully.
more cost-effective, simpler solution-
Calibration standards can be prepared
based standards to be utilized. The
automatically on the autosampler
most important consideration here
platform if a compatible GERSTEL
is that because SIFT-MS is a direct
MultiPurpose Sampler is used (Robotic-
analysis method, the solvent suitability
Pro with tool change, or dual-head XT).
is different compared with GC/MS.
The GERSTEL Maestro PrepAhead
The criterion to be assessed is that the
functionality enables this preparative
solvent must be non-reactive or react
work to be done extremely efficiently,
only very slowly with one or more SIFT-
in addition to the calibration run on the
MS reagent ions. In contrast to GC/MS,
SIFT-MS instrument itself (Figure 24).
water is the preferred solvent for SIFT-
MS. However, in various applications Analogous to GC/MS, multiple-point
methanol, chloroform, dichloromethane, calibration curves covering the range of
Figure 26: Data from (a) ten and Table 4: Concentrations for
(b) three replicate analyses for replicate calibration standards
benzene, ethylene, isobutane, and then spiked soil samples
and toluene of the “Syft Calibrant (including recoveries) analyzed
Standard.” RSDs are annotated using automated methanolic
in red. extraction and SIFT-MS. Data are
organized per the sequence table
for the run and illustrate the use
of a QCC. See the text here and
in Section D.4 for
more details.
a. Method
Figure 33: Calibration curves
for (a) GC/MS and (b) SIFT-MS Porcine blood plasma samples (1 mL)
detection of the solvent and its were provided in 20-mL headspace
metabolite. vials, which were then analyzed by
headspace GC/MS and headspace SIFT-
MS. The concentration of the solvent
and metabolite were determined in the
samples by both analysis methods by
quantifying against calibration standards
from 2-1000 ppm prepared in 0.1 M
sodium chloride solution in MilliQ
ultrapure water. Samples and standards
were heated for 15 minutes at 40 °C
in an agitator prior to injection. The
calibration curves obtained for GC/MS
and SIFT-MS are shown in Figure 33.
The sample sequence included a 200-
ppm quality control check standard after
every six plasma samples. This standard
contained both target compounds and
the results are summarized in Table 6.
The repeatability of the QCC standards
is excellent for SIFT-MS, outperforming
GC/MS.
b. Results
Faster detection of co-formulants could The results obtained for the plasma
aid earlier identification of pesticide samples are shown in Figure 34 for the
poisoning and faster intervention, solvent and its metabolite. Each sample
reducing mortality. The aim of this was analyzed using both analytical
techniques and the correlation is
study was to compare the analytical
excellent.
performance of headspace-SIFT-MS
with gold-standard headspace-GC/ c. Study conclusion
MS, because SIFT-MS provides higher Excellent correlation between gold-
sample throughput. standard headspace-GC/MS method and
GC-MS/ SIFT-MS/
Standard GC-MS/ppm SIFT-MS/ppm
ppm ppm
Replicate 1 199 199 199 203
Replicate 2 196 191 196 188
Replicate 3 202 194 220 207
Replicate 4 199 195 227 212
Replicate 5 192 198 206 210
Replicate 6 197 190 195 188
Replicate 7 222 202 258 211
Replicate 8 201 176 206 205
Replicate 9 195 196 214 208
Average 200 193 213 203
Standard Deviation 8.8 7.4 19.9 9.3
%RSD 4.4 3.8 9.3 4.6
the new headspace-SIFT-MS method in sample throughput can be achieved
was demonstrated in this study even compared to the headspace-GC/MS
though SIFT-MS is theoretically not as method.
selective as GC/MS. SIFT-MS analysis
benefited from 50% lower carryover 8. SIFT-MS is an Excellent Fit for
than GC/MS, and the repeatability of Routine Analysis
QCCs was also higher. By implementing SIFT-MS fits very comfortably in the
headspace-SIFT-MS, a four-fold increase routine testing laboratory and CRO,
Standard additions
Internal standards
Method development
Method validation
S P ONS OR E D C ONT E NT
SECTION D
Development and
Validation of
SIFT-MS Methods
T
Africa Studio/stock.adobe.com
Table 8: Target compounds for the method, plus reagent and product
ions used to quantify them.
1,2,4-trichlorobenzene (180)
1,4-dichlorobenzene (148)
1,4-dichlorobenzene (146)
chlorobenzene (112)
chlorobenzene (114)
* International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human Use (ICH). Validation of Analytical Procedures: Text and
Methodology Q2(R1). 1996.
Table 12: Final parameters for headspace (HS) analysis of water using
HS-SIFT-MS. The method was developed to use both positively and
negatively charged reagent ions.
iii. LOD/LOQ and range (e.g., benzene and DCM) are clearly
The LOD/LOQ approach for SIFT-MS is measurable to lower levels, but defining
modified compared to chromatographic a single LOQ for the method gives
methods, because there is no baseline 10 ppb in solution. This means that the
noise from which to calculate signal-to- range of the method is 10–2000 ppb.
noise ratios. The approach taken was to
iv. Precision
determine the LOQ based on %RSDs of
low-concentration standards. Table 14 summarizes the system
precision data obtained for the method
The LOQ was estimated as lying within
at three concentrations (six replicates
the 1–10-ppb concentration range in
each). Typical acceptance criteria (RSD
solution. Triplicate measurements were
<10%) are met easily, since the %RSD
made at four concentrations across
this range (Table 13). Compounds that is less than 9% in all cases.
partition very well into the headspace Analytical precision data are shown in
Table 13: LOQ data for the method, with the mean and standard
deviation shown in parts-per-billion by volume measured in the
headspace. The data for the 10-ppb solution meet acceptance criteria,
so the general method LOQ is stated as 10 ppb in solution.
Table 14: System precision data for the method, with the mean and
standard deviation shown in parts-per-million by volume measured in
the headspace.
Table 15: Amount of sample recovered (in ppm) for each of the
triplicate measurements at 500, 1000, and 1500 ppb concentration
in water.
were met, including those for accuracy SIFT-MS instrument is ca. 25 standard
and recovery, which is where failure is cubic centimeters per minute (sccm) at
often experienced in the intermediate a temperature of 150 °C. The robustness
precision phase. study therefore investigated the effect
on linearity and repeatability with the
vii. Robustness following changes using the 1-ppm
Ensuring consistency of sample flow calibration standard:
rate into the SIFT-MS instrument is very
• Inlet temperature decrease (due to
important for repeatable quantitation in
heater or control fault): 140 °C,
SIFT-MS and has some susceptibility to
short-term drift (e.g., due to blockage of • Inlet flow reduction (due to
the flow-controlling capillary or change in blockage): 20 sccm, and
inlet temperature). The factory inlet flow • Inlet flow increase (due to cooler
on a Syft Technologies’ Voice200ultra inlet): 30sccm.
Table 17: Target compounds for the method, plus reagent and product
ions used to quantify them.
Table 18: Detection parameters for BTEX using the positively charged
SIFT-MS reagent ions. Product ions from reaction with named analyte
are shown as ion formula (m/z, in Daltons), and product ion branching
ratio as a percentage.
1. The three xylene isomers (m-, o-, and p-) all react similarly with these reagent ions within 5% relative product abundance.
Note that for clarity Figure 17 is plotted analyses, without the need to fully
in terms of the spiked individual redevelop the GC/MS method. Secondly,
concentrations of ethylbenzene and it achieved a two-fold increase in sample
m-xylene, but only the total of the two is throughput for SIFT-MS when compared
measured (as noted above). with GC/MS analysis using the current
Table 4 shows the results obtained for sequential sample preparation/analysis
calibration standards and the various method. (It should be possible to further
spiked soil samples in the sequence increase this by utilizing the multiple
that they were run. Spike levels in ppb positions available with the GERSTEL
are applicable to all samples; the RSDs modules and by batch-preparing the
for calibration standard measurements samples.) Finally, the method has
were excellent (1.6 to 2.3% with no
easily met all acceptance criteria during
internal standard used), and recoveries
validation, as shown in Table 19. As
of 75-111% are all acceptable. For the
discussed in the first case study, these
experimental conditions used here, the
LOQ and LOD are 40 ng g-1 and 12 ng g-1, parameters are often used in a wide
respectively. There is potential to achieve variety of applications.
lower LOQs and LODs through better For more information on this study,
optimized extraction conditions. refer to the Syft application note “Rapid
This study demonstrated firstly that Analysis of BTEX Residues in Soil using
SIFT-MS can replace GC/MS for certain Automated SIFT-MS” (APN-043).
Table 19: Detection parameters for BTEX using the positively charged
SIFT-MS reagent ions. Product ions from reaction with named analyte
are shown as ion formula (m/z, in Daltons), and product ion branching
ratio as a percentage.
Accuracy (% RSD)
• 0.8 – 7.4%
• 500 ppb
<10% • 1.0 – 4.0%
• 1000 ppb
• 0.2 – 4.2%
• 1500 ppb
For more information on this study, refer to the Syft application note “Rapid Analysis of BTEX Residues in Soil using Automated SIFT-MS” (APN-043).
Intermediate precision
Robustness
When required, standard and sample
stability tests are also made with ease.
Africa Studio/stock.adobe.com
S P ONS OR E D C ONT E NT
SECTION E
Simple,
Comprehensive
and Broad-
Spectrum SIFT-MS
Analysis
A
Mariusz Blach/stock.adobe.com
Table 20: Linearity and recovery data for direct formaldehyde analysis.
See the text for full details.
Theoretical Measured Concentration Calculated Concentration
Recovery (%)
Concentration (ppbV) (ppbV) (ppbV)
96 75 104 107.9
7272 6852
7217 6986
7158 6895
7275 7275
7156 6858
7077 6765
7110 6789
Test 2: Flow Tube 113 °C and Sample Inlet 120 °C Test 4: Flow Tube 120 °C and Sample Inlet 127 °C
Theoretical Concentration Measured Concentration Theoretical Concentration Measured Concentration
(ppbV) (ppbV) (ppbV) (ppbV)
7218 6706
7143 6689
7032 6707
7275 7275
6967 6620
6926 6696
6933 6634
c. Results
Figure 49 shows the results obtained
for several apple varieties, a pear
variety and a banana variety, which
used the lower incubation temperature
and shorter incubation time. Following
incubation, the SIFT-MS method enabled
each sample to be analyzed in less than
two minutes, and quantified ethylene
together with the characteristic aroma
Figure 50: GERSTEL Maestro
compounds.
software illustrates how rapidly
direct SIFT-MS analysis can Clearly there is opportunity to further
determine optimum headspace optimize the method—especially in
equilibration time. terms of shorter incubation times and
reduced temperature so as to reduce
temperature- and enzyme-induced
changes to the volatile profiles. Because
of the short injection and syringe flush
cycles with SIFT-MS, six incubation
cycles (in 5-minute steps from 5 to 30
minutes) can be evaluated in less than
37 minutes (Figure 50)! With GC/MS,
this evaluation will take 3 hours and 42
minutes, or six times longer.
limiting step: the long analysis time in
GC/MS. Under the one-hour sample d. Conclusion
incubation conditions, six samples can SIFT-MS provides simple and highly
be analyzed nearly 3.5x faster by using sensitive and selective analysis of low
SIFT-MS rather than GC/MS. Translating molecular weight volatiles from fruits,
this to 24-hour throughput, SIFT-MS can with no delays due to late elution of
analyze over six times more samples non-targeted, less volatile compounds.
Efficiency improvements vary depending column when only more volatile species
on the experimental approach, but this are actually targeted.
study demonstrates six-fold throughput
enhancements when utilizing automated
incubator hardware optimized for GC.
Not only does SIFT-MS deliver higher
sample throughputs for routine analysis,
but it can rapidly determine optimal
incubation times and temperatures in
the method development phase. For
another example of this, see the method
development portion of Case Study 1 in
Part 3 of this webinar series.
For more details on this study, see the
Syft Technologies application note High-
Throughput Analysis of Volatiles from
Fruit Using SIFT-MS (APN-031).
4. Summary
By applying ultra-soft chemical
ionization (USCITM), SIFT-MS greatly
simplifies analysis of important
chromatographically challenging volatiles,
such as formaldehyde, volatile fatty
acids, organosulfur compounds, and
amines. SIFT-MS eliminates the need
for derivatization, drying steps, or
selection of special columns, detectors
and/or analysis conditions. Additionally,
SIFT-MS resolves commonly perceived
limitations of direct analysis methods, in
that its unique USCI system can provide
isomer resolution. The elimination
of chromatographic separation also
means that there is no waiting for lower
volatility compounds to elute from the Mariuz Blach/stock.adobe.com
SECTION F
Summary: SIFT-
MS and Routine
Analysis
Anchalee/stock.adobe.com
T
his ebook series has demonstrated that SIFT-MS is the ideal analytical
technique to address throughput challenges for volatile compounds in
a wide range of matrices (Section B), while also extending analytical
capability through simultaneous analysis of chemically diverse and
chromatographically challenging species (Section E). SIFT-MS achieves all this while
comfortably accommodating routine analysis procedures (Section C) and analytical
method validation that conforms to regulatory submission requirements (Sections
D and E).
SIFT-MS is therefore the essential third technique to complement GC and LC in
routine testing laboratories and CROs.
74 AUG U ST 2 0 2 0 | L C G C
BIOGRAPHIES
Acknowledgments
• Drs Diandree Padayachee, Kalib Bell and Jing Ma – Syft Technologies, NZ
• Dr Colin Hastie – Anatune Limited, UK
• Dr Natalie Homer - Mass Spectrometry Core, for Cardiovascular Sciences,
Queen’s Medical Research Institute, Edinburgh, UK
• Drs Michael Eddleston and Adrian Thompson – University/BHF Centre for
Cardiovascular Sciences, Queen’s Medical Research Institute, UK