You are on page 1of 8

Japanese Journal of Applied Physics 56, 07JF21 (2017) REGULAR PAPER

https://doi.org/10.7567/JJAP.56.07JF21

Multifunctional pulse generator for high-intensity focused ultrasound system


Satoshi Tamano1,3, Shin Yoshizawa1, and Shin-Ichiro Umemura2
1
Graduate School of Engineering, Tohoku University, Sendai 980-8579, Japan
2
Graduate School of Biomedical Engineering, Tohoku University, Sendai 980-8579, Japan
3
Healthcare Business Unit, Hitachi, Ltd., Kokubunji, Tokyo 185-0014, Japan

Received November 26, 2016; revised April 11, 2017; accepted April 16, 2017; published online June 21, 2017

High-intensity focused ultrasound (HIFU) can achieve high spatial resolution for the treatment of diseases. A major technical challenge in
implementing a HIFU therapeutic system is to generate high-voltage high-current signals for effectively exciting a multichannel HIFU transducer at
high efficiencies. In this paper, we present the development of a multifunctional multichannel generator/driver. The generator can produce a long
burst as well as an extremely high-voltage short pulse of pseudosinusoidal waves (trigger HIFU) and second-harmonic superimposed waves for
HIFU transmission. The transmission timing, waveform, and frequency can be controlled using a field-programmable gate array (FPGA) via a
universal serial bus (USB) microcontroller. The hardware is implemented in a compact printed circuit board. The test results of trigger HIFU reveal
that the power consumption and the temperature rise of metal–oxide semiconductor field-effect transistors were reduced by 19.9% and 38.2 °C,
respectively, from the previous design. The highly flexible performance of the novel generator/driver is demonstrated in the generation of second-
harmonic superimposed waves, which is useful for cavitation-enhanced HIFU treatment, although the previous design exhibited difficulty in
generating it. © 2017 The Japan Society of Applied Physics

A highly negative acoustic pressure is expected to be


1. Introduction effective for inducing cavitation. An asymmetric pressure
Since the 1950s, therapeutic ultrasound has been applied to waveform with either its negative or positive peak emphasized
the treatment of various diseases. The rapid technological can be produced by superimposing a second-harmonic wave
progress over the last two decades has extended the clinical onto a fundamental wave. Such second-harmonic super-
applications of high-intensity focused ultrasound (HIFU) imposed waves were proven to be useful for not only gen-
from benign prostatic hyperplasia1) and uterine fibrosis2) to erating scattered cavitation microbubbles but also promoting
malignant tumors. Currently, HIFU treatment is practiced as or inhibiting their evolution into a large cavitation cloud.13)
a method to thermally ablate solid tumors, such as those of A major technical challenge in implementing a HIFU
prostate cancer,3) liver cancer,4) bone metastases,5) breast therapeutic system is to generate high-voltage high-current
cancer,6) brain tumor,7) and pancreatic cancer.8) signals for the effective excitation of a multichannel HIFU
Ultrasonic therapies9) can be categorized into those transducer at a high efficiency. High efficiency is particularly
inducing thermal or nonthermal effects on the living body important for reducing the volume of the entire circuit driving
upon ultrasonic exposure. Ultrasonic energy is converted to a large number of transducer elements to an acceptable level
heat upon its in biological tissue, and the tissue can be because heat generation functions against the reduction in
irreversibly changed following a sufficient increase in its volume. As the conventional class A amplifier design does
temperature. On the other hand, shock wave lithotripsy, not agree with this requirement, a number of class D
histotripsy, fracture treatment, and sonodynamic therapy are amplifier designs have been proposed.
based on the nonthermal effects of ultrasound. Umemura and Cain14) developed a unipolar power trans-
A long treatment time is a practical problem of HIFU mission circuit using FETs for driving a 32-element HIFU
treatment. When using a typical equipment for the HIFU transducer. The maximum transmission voltage was +40 V.
treatment of prostate cancer, a median operating time of In this transmission circuit, by connecting an inductor in
142 min (range, 35–390 min), which includes the cooling series with the output of the FETs, and further connecting an
time between consecutive HIFU exposures, is required for adjustable capacitor parallel to the transducer, a narrow-band
treating a median prostate volume of 21.9 cm3 (range, 4.6– transmission waveform was generated.
68.8 cm3).1) A novel HIFU treatment technique is required for Furthermore, Brown and Lockwood15) developed a uni-
coagulating a larger volume in a shorter time. polar ultrasonic wave transmission circuit using FETs. The
Acoustically driven microbubbles can accelerate ultrasonic maximum transmission voltage was −350 V, the transmitting
heating. The trigger HIFU method10,11) was proposed for frequency was approximately 2 MHz, and a single negative
utilizing them. By using this method, one can generate pulse was generated with a duration of 500 ns. However, its
cavitation microbubbles selectively in a focal region using a application to HIFU with longer exposure durations was not
short ultrasonic pulse at an extremely high amplitude. The considered.
generated microbubbles can enhance the heating effect of a Xu et al.16) proposed a high-frequency (60 MHz or more)
burst of ultrasonic waves at a moderately high amplitude, transmission circuit with a bipolar power supply, a trans-
immediately following the pulse. A significant reduction in former, and FETs. The transmission of a single pulse at
treatment time was thereby achieved. 35 MHz and 160 V peak to peak (Vpp) was possible.
Cavitation microbubbles are also be used in therapeutic However, when the transmission wave was extended to
methods for eroding calculi and soft tissue, known as seven wavelengths, the maximum transmission voltage
lithotripsy and histotripsy, respectively. Furthermore, they decreased to 75 Vpp.
can induce sonochemical reactions, which can be utilized for Lewis and Olbricht17,18) developed a HIFU transmission
therapeutic purposes in the form of sonodynamic therapy.12) circuit for a single transducer with using multiple of parallel-
07JF21-1 © 2017 The Japan Society of Applied Physics
Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

Control board EEP EEP Transmission boards


ROM ROM

To
MOSFET N Channel
From/to PC Transducers
USB Control TX Driver MOSFET
μPC FPGA FPGA MOSFET P Channel
Driver MOSFET
7 levels
32 channels
X-tal

Fig. 1. Block diagram of multifunctional pulse generator=driver for HIFU system.

connected N-channel=P-channel integrated FETs. Their aim trates the transmission board control bus as well as the func-
was to increase transmission efficiency using FETs with tions described previously. The dimensions of the printed-
lower on-resistance than using single FET and a bipolar circuit board (PCB) are 382 mm wide and 308 mm high.
HIFU transmission circuit. 2.1.2 Transmission circuit. The transmission board in-
Moro and coworkers19–21) proposed a high-voltage stair- cludes a transmission FPGA for generating a timing signal
case drive circuit for HIFU transmission. For this purpose, for ultrasonic transmission based on a command transmitted
four staircase voltage levels and metal–oxide semiconductor from the control FPGA, RAMs that store transmission focus
field-effect transistors (MOSFETs) were used. The reduction data, N- and P-channel MOSFETs that generate ultrasonic
of odd harmonic transmission components of the pseudosi- transmission signals, and MOSFET drivers. The dimensions
nusoidal waveform was achieved. However, the device of the PCB are the same as those of the transmission control
overheating problem with long exposure duration remained. circuit.
This problem was solved by Tamano et al.22) using both N- 2.1.3 Transmission FPGA. In the circuit design described
and P-channel MOSFETs for each intermediate voltage level. in Sect. 2.1.7, it is essential to output 12 MOSFET driver
Seip et al.23) proposed a HIFU transmission circuit with signals per channel. To control 16 channels per transmission
digital pulse-width-modulation (PWM) control class D FPGA, it is essential to use a FPGA with a large number
amplifiers, which can adjust transmission amplitude. How- of pin configurations; not only MOSFET driver signals but
ever, in order to reduce high-order harmonics, a fifth-order also communicates with the memory that stores the focus
low-pass filter is used. Moreover, we required a transformer data and with the control FPGA. On the basis of these
for impedance matching. For this reason, in the case of requirements, we selected 5CEBA7F31C8 (Intel) as the
multichannel integration, a large-scale circuit configuration is transmission FPGA. This FPGA has 5640 logical resources
required. composed of a combination circuits, sequential circuits, and
In this paper, we report the development of a novel 156 DSP blocks. Furthermore, as it can use 480 pins as an
multifunctional, reconfigurable generator=driver system input=output terminal, it is the most suitable device for this
using a field-programmable gate array (FPGA) and high- study. As this device uses a Taiwan semiconductor
voltage MOSFETs. The system can produce pseudosinusoi- manufacturing company’s (TMSC) 28 nm process, low-
dal waves, trigger HIFU mode ultrasonic waves, and second- power consumption operation is possible even for large-
harmonic superimposed waves. scale integration.
2.1.4 Transmission focusing RAM. Transmission focus-
2. Materials and methods ing RAM stores transmission focus data generated using an
2.1 System description external PC. For setting 16,384 focus points, CY7C137226)
In Fig. 1, a block diagram of the proposed generator=driver (Cypress) was selected.
system is illustrated. The generator circuit is configured to 2.1.5 MOSFET driver. In general, the FPGA single-ended
transmit pseudosinusoidal waves, trigger HIFU waves, or external output is 2.5 V CMOS level. On the other hand, a
second-harmonic superimposed waves. MOSFET gate requires a threshold voltage (Vgs(th)), which is
2.1.1 Transmission control circuit. The generator unit generally higher than 5.0 V, and a high peak current for
is interfaced with a personal computer (PC) and receives proper switching operation. Therefore, the MOSFET driver is
commands, waveform data, and focusing delay data through a a level shift amplifier that converts a low-voltage low-current
high-speed USB 2.0 microcontroller (Cypress CY7C68013A- signal from the FPGA to a MOSFET gate signal. For the
128AXC24)). The control FPGA (Intel 5CEBA7F31C825)) transmission board of this study, 384 MOSFET drivers are
stores the instruction information from the PC in an internal required per board. To mount the drivers in a small area,
register, transmits the data of the transmission waveform and MD182027) (Microchip) with four-channel MOSFET drivers
focuses to the transmission FPGA. For data communication per IC in a 3 mm square was selected. It has a rise and falls
between the control FPGA and the transmission FPGA, time of 6.0 ns to drive the MOSFETs at a high speed. Table I
information is transferred using a dedicated local bus. shows representative specifications of the MOSFET drivers.
Ultrasonic waves are transmitted from 32 channels by one 2.1.6 High-voltage MOSFET. In the trigger HIFU mode, a
transmission board. For a HIFU transducer consisting of 128 short pulse at an extremely high amplitude is essential for
elements, four transmission boards are required. The control generating cavitation in living bodies. In a previous study,22)
board generates actual ultrasonic wave sequences and arbi- we examined the use of SiHFR9310-GE328) (Vishay) as a
07JF21-2 © 2017 The Japan Society of Applied Physics
Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

Table I. MOSFET driver specifications. Table II. MOSFET specifications.


Manufacturer Microchip Type P N
Product name MD1820 Manufacturer On Semiconductor On Semiconductor
Peak output sink current Isink (A) 2 Product name FQD3P50 FDD5N50NZ
Peak output source current Isource (A) 2 Drain-to-source voltage Vds (V) −500 500
Propagation delay tP (ns) 6.5 Continuous drain current Id (A) −2.1 4
Output risetime tr (ns) 7 Static drain-to-source
3.9 1.4
Output fall time tf (ns) 7 on-resistance Rds(on) (Ω)
Propagation delay matching Δtdm (ns) ±2.0 Total gate charge Qg (nC) 18 9
Peak diode recovery
−4.5 10
dV=dt (V=ns)
Repetitive avalanche
5 6.2
energy EAR (mJ)
P-channel MOSFET that actually generates ultrasound
Turn off fall time tf (ns) 45 21
transmission waveforms. As the total gate charge (Qg) of
SiHFR9310-GE3 is as small as 13 nC, the heat generation of
the MOSFET could be suppressed. However, the maximum
drain–source breakdown voltage (BVdss) is limited to shows the representative specifications of the MOSFETs used
−400 V. In the trigger HIFU mode, an ultrasonic pulse in this study.
should be irradiated at 400 Vpp. Furthermore, the transient 2.1.7 Transmission circuit schematics. In a previous
voltage may exceed 400 Vpp. Therefore, in this study, study,22) we proposed the use of a pair of P- and N-channel
FQD3P5029) (On Semiconductor) with a BVdss of −500 V MOSFETs for the intermediate power supply voltage trans-
was used to reduce the risk of device damage. As Qg is mission circuit of the staircase drive circuit using multiple
increased from 13 to 18 nC, the temperature of the device power supplies, as shown in Fig. 2(a). In particular, the circuit
will increase. It was used in combination with an N-channel configuration of positive and negative two-stage power
MOSFET, FDD5N50NZ30) (On Semiconductor). Table II supplies has been described in detail. We have discussed that

(a) (b)

HV2 > HV1>GND


Fig. 2. Transmission circuit diagram: (a) previous design of HIFU transmission circuit and (b) proposed HIFU transmission circuit. Compared with the
circuit diagram in (a), that in (b) has additional transmission circuits accompanying one intermediate power supply level and ground level. On the other hand,
the highest power supply level N-channel MOSFET, series diode, and the lowest power supply level P-channel MOSFET, series diode were removed from the
previous transmission circuit.

07JF21-3 © 2017 The Japan Society of Applied Physics


Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

in combination. In repetitive transmission in this mode, more


power is consumed for transmitting the heating bursts than
trigger pulses. Therefore, the reduction in power consump-
tion has to be the first concern in heating burst transmission.
When a seven-step staircase amplifier with three positive and
negative steps is used in the trigger HIFU mode, the five
lower steps with two positive and negative steps among the
seven steps can be used for heating burst transmission. In
contrast, three voltage steps with only one positive and
negative step among the five steps can be used for heating
burst transmission when a five-step staircase amplifier with
Fig. 3. (Color online) Measurement results of transducer electrical two positive and negative steps is used in the trigger HIFU
properties. The center frequencies of the transducer were 1.1 and 2.1 MHz. mode.
The blue line shows impedance and the red line shows the phase.
Figure 4 shows the driving waveforms of the prototype
amplifier in the trigger HIFU mode at 1 MHz, with the trigger
pulse and heating burst driving voltages of 400 and 100 Vpp,
it is possible to generate the desired staircase waveform respectively. The power spectra of the trigger pulse and
without depending on the imaginary part of load impedance. heating burst are also shown. The upper figures show those
That is, in a conventional class D amplifier system, an with two positive and negative voltage steps of 200.3 and
impedance matching circuit23,31) is required according to the 49.8 V. In this case, all the four power supply voltages are
load characteristics; however, in the circuit in our previous used for trigger pulse generation, and the two lower supply
study, the impedance matching circuit is not required. Even if voltages are used for heating burst generation. The lower
reflections of an electric signal occur owing to the impedance figures show those with three positive and negative voltage
mismatch, the excessive electric charge should be absorbed steps of 200.3, 49.8, and 20.1 V. In this case, all the six
by the power supply capacitor immediately and reused later. power supply voltages are used for trigger pulse generation,
Therefore, it is not necessary to consider signal deterioration and the four lower supply voltages are used for heating burst
due to electric signal reflection. Furthermore, in the previous generation. The lower and upper supply voltages correspond
study, we examined the relationship between the number of to the proposed and previous circuits, respectively.
staircase steps and power consumption. As a result, it was The harmonic spectral levels corresponded with those of
confirmed by simulation that the power consumption becomes the fundamental levels are listed in Table III. The third and
minimum when the positive and negative power supply fifth harmonic levels of the heating burst generated by the
voltage levels are three to four steps. proposed circuit were 10 and 47 dB lower than those of the
On the basis of the results of these studies and the previous circuit, respectively. This resulted in a significantly
preliminary consideration of an actual circuit scale size, a smaller temperature rise of the MOSFETs and lower power
seven-stage staircase, with three positive and negative power consumption of the proposed circuit than of the previous
supply voltages and the ground level, was selected for this circuit, as shown in Table III. The power consumption of
study. The circuit should exhibit flexibility in driving not the proposed circuit was approximately 80% of that of the
only pseudosinusoidal waveforms of the trigger pulse and previous circuit while the fundamental power of the proposed
heating burst in the trigger HIFU mode but also second- circuit was approximately 90% of that of the previous circuit.
harmonic superimposed waveforms. A diagram of the newly The reason why the reduction in the MOSFET temperature
proposed circuit is shown in Fig. 2(b). rise by the proposed circuit was considerably more significant
2.2 HIFU transducer than the reduction in the power consumption is probably
The prototype transmission circuit was tested and evaluated because the major portion of the consumed electric power
using a prototype dual frequency transducer at 1 and may have been converted to acoustic power by the driven
2 MHz.32) It consists of seven electrically independent square array transducer and only the remaining electric power may
elements with a width and a height of approximately 8.8 mm, have contributed to the heating of MOSFETs. This character-
and a resulting area of 77.4 mm2. The electrical properties of istic is very important when a large number of such circuits in
the central element are shown in Fig. 3. It shows that the a compact form are used to drive an array transducer.
impedance decreases and the phase approaches zero near 3.2 Second-harmonic superimposed wave mode
both 1 and 2 MHz. This indicates that the ultrasonic waves A second-harmonic superimposed pressure waveform at a
can be efficiently transmitted at these frequencies. This low intensity can be written as13)
transducer element was used for evaluating the circuit
p ¼ p1 cos !t þ p2 cosð2!t þ ’Þ; ð1Þ
performance on the basis of the measurements in the
following experiments. where p1 and p2 are the amplitudes of the fundamental and
second-harmonic superimposed pressure waveform, respec-
3. Results and discussion tively, and φ is the phase of the second-harmonic super-
3.1 Trigger HIFU mode imposed pressure waveform relative to the fundamental
In the trigger HIFU mode, a trigger pulse at a high peak superimposed pressure waveform. φ is set at π and zero for
power corresponding to a high voltage (>300 Vpp) for a short the peak negative-enhanced (PNE) and peak positive-
duration of 5–200 µs and a heating burst with a medium enhanced (PPE) waves, respectively. The ratio of the peak
output (<100 Vpp) for a long duration up to 1–10 s are used positive to the negative pressure of the PPE waveforms,
07JF21-4 © 2017 The Japan Society of Applied Physics
Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

Table III. Trigger HIFU heating mode transmission waveform comparison.


Circuit configuration
Previous circuit Proposed circuit
Power supply steps 4 6
Triggered-HIFU transmission mode Triggered pulse Heating burst Triggered pulse Heating burst
Power consumption (W) — 17.6 — 14.1
MOSFET temperature rise (°C) — 85.1 — 46.9
Third harmonics amplitude relative to the fundamental wave (dB) −22.1 −17.5 −23.0 −27.4
Fifth harmonics amplitude relative to the fundamental wave (dB) −48.6 −32.4 −53.2 −79.5

(a-1) (b-1) (c-1)

Voltage [V]

Power [dB]
Voltage [V]

Time [μs] Frequency [MHz]


Time [μs]
(d-1) (e-1)
Voltage [V]

Power [dB]
Time [μs]
Frequency [MHz]

(a-2) (b-2) (c-2)


Voltage [V]

Power [dB]
Voltage [V]

Time [μs] Frequency [MHz]


Time [μs]
(d-2) (e-2)
Power [dB]
Voltage [V]

Time [μs] Frequency [MHz]

Fig. 4. (Color online) Previous and proposed circuit trigger HIFU waveforms and power spectra at 1 MHz transmission: Upper figures (a-1)–(e-1) show
experiment results of the previous circuit with four power voltage steps. (a-1) actual 400 Vpp trigger pulse and 100 Vpp heating burst waveform, (b-1) trigger
pulse waveform, (c-1) trigger pulse power spectrum, (d-1) heating burst waveform, and (e-1) heating burst power spectrum. Lower figures (a-2)–(e-2) show
experimental results of the proposed circuit with six-power-voltage-step. (a-2) actual 400 Vpp trigger pulse and 100 Vpp heating burst waveform, (b-2) trigger
pulse waveform, (c-2) trigger pulse power spectrum, (d-2) heating burst waveform, and (e-2) heating burst power spectrum. In the heating burst waveform, the
proposed circuit is 9.9 and 47.1 dB smaller than the previous circuit at the third and fifth harmonics, respectively.

p+=∣p−∣, becomes maximum when p2=p1 is 0.5. Similarly, the bubbles. Therefore, both PNE and PPE waveforms should
ratio of the peak negative to the positive pressure of the be useful as the trigger pulse in the trigger HIFU mode.
PNE waveforms, ∣p−∣=p+, becomes maximum when p2=p1 is Such waveforms can be generated easily by a class A linear
0.5. The peak positive pressure and the absolute peak nega- amplifier such as E&I 100A2.33) However, class A linear
tive pressure both become maximum when p2=p1 is 1 at the amplifiers are not suitable for driving a multichannel
same total acoustic power of the fundamental and second- transducer in the compact amplifier form because they
harmonic. themselves generate heat; at least they can generate acoustic
A high peak negative pressure13) is a key factor for power. On the other hand, conventional class D amplifiers do
generating acoustic cavitation. A high peak positive pres- not generate heat themselves considerably and thus cannot
sure13) also has been reported to be effective for generating a generate complicated waveforms such as those with second-
large cavitation cloud from a small number of cavitation harmonic superimposition. The staircase voltage amplifiers
07JF21-5 © 2017 The Japan Society of Applied Physics
Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

(a-1) (b-1) (c-1)

Voltage [V]

Power [dB]
Amplitude

Time [μs] Frequency [MHz]


Time [μs]
(d-1) (e-1)

Power [dB]
Voltage [V]

Time [μs] Frequency [MHz]

(a-2) (b-2)
(c-2)
Amplitude

Voltage [V]

Power [dB]
Time [μs] Time [μs] Frequency [MHz]
(d-2)
(e-2)
Voltage [V]

Power [dB]

Time [μs] Frequency [MHz]

Fig. 5. (Color online) Ideal and experimental results of second-harmonic superimposed PNE, PPE waveform, and a power spectrum when p2=p1 is 1: The
upper figures (a-1)–(e-1) show the ideal PNE waveform and experimental results. (a-1) ideal waveform, (b-1) three-power-voltage-step waveform, (c-1) three-
power-voltage-step power spectrum, (d-1) five-power-voltage-step waveform, and (e-1) five-power-voltage-step power spectrum. In the three-power-voltage-
step transmission circuit and the five-power-voltage-step transmission circuit, there is no large difference in the second-harmonic generation level. However, in
the five-power-voltage-step transmission circuit, the third and fourth harmonics are smaller by 22.6 and 13.0 dB, respectively than in the three-power-voltage-
step circuit. The lower figures (a-2)–(e-2) show the ideal PPE waveform and experimental results. (a-2) ideal waveform, (b-2) three-power-voltage-step
waveform, (c-2) three-power-voltage-step power spectrum, (d-2) five-power-voltage-step waveform, and (e-2) five-power-voltage-step power spectrum. In the
three-power-voltage-step transmission circuit and the five-power-voltage-step transmission circuit, there is no large difference in the second-harmonic
generation level. However, in the five-power-voltage-step transmission circuit, the third and fourth harmonics are smaller by 45.1 and 9.5 dB, respectively, than
in the three-power-voltage-step circuit.

with multiple voltage levels proposed in this study may transmissions, positive and negative voltage steps of 200.3
potentially generate such complicated waveforms without and 101.6 V, respectively, and ground potential were used.
heating themselves considerably. Their power spectra are shown in Figs. 5(c-1), 5(e-1), 5(c-2),
We designed and evaluated three- and five-level trans- and 5(e-2). Because of the infinite rise and fall times, the
mission waveforms so that the acoustic powers of 1 MHz second-harmonic amplitudes were 1–2 dB lower than the
fundamental and 2 MHz second-harmonic are the same. corresponding fundamental amplitudes. These can be ad-
Figures 5(a-1) and 5(a-2) show the ideal PNE and PPE justed to be closer to 0 dB if necessary. The unwanted third
second-harmonic superimposed waveforms, respectively, and fourth harmonics are also significant in the waveforms
when p2=p1 is 1. Figures 5(b-1) to 5(e-1) and 5(b-2) to with three voltage levels while they are not more than −20 dB
5(e-2) show the exact PNE and PPE drive waveforms and in the five voltage levels. We also evaluated the PNE and
power spectra, at an either negative or positive peak voltage PPE waveform power consumptions and the MOSFET
of 200 V, respectively. Only three voltage levels were used temperature rise, at an either negative or positive voltage of
for the waveforms shown in Figs. 5(b-1) and 5(b-2) while 50 V, to prevent MOSFET damage due to excessive heat
five voltage levels were used for those in Figs. 5(d-1) and generation. In this case, the five-voltage-level transmission
5(d-2). In the case of three-voltage-level transmissions, circuit reduced power consumption by 40.7% or more from
positive and negative voltage steps of 200.3 V and ground the three-voltage-level transmission circuit and suppressed
potential were used. In the case of five-voltage-level the device temperature rise beyond 51.3 °C. These confirm
07JF21-6 © 2017 The Japan Society of Applied Physics
Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

Table IV. Second-harmonic superimposed wave mode transmission Consequently, we confirmed that the proposed circuit is
waveform comparison.
effective for second-harmonic superimposed PPE and PNE
Circuit configuration waveform generation. By combining our proposed circuit, a
Three-level Five-level multichannel transducer, and our proposed HIFU trans-
transmission transmission mission techniques, we can expect to generate cavitation
Generating waveform PNE PPE PNE PPE bubbles efficiently and accurately at the desired target
Second harmonics amplitude relative location and time. Consequently, this method could be
−1.4 −2.1 −1.8 −1.5
to the fundamental wave (dB) efficiently used for HIFU treatment, sonodynamic therapy,
Third harmonics amplitude relative
−6.5 −5.6 −29.1 −50.8
lithotripsy, and histotripsy.
to the fundamental wave (dB)
Fourth harmonics amplitude relative 4. Conclusions and future works
−8.0 −8.7 −21.0 −18.2
to the fundamental wave (dB)
We modified the HIFU ultrasonic transmission circuit
Power consumption (W) 148.9 169.3 88.3 88.6
previously reported and fabricated a HIFU transmission unit
Maximum temperature rise (°C) 81.4 93.9 30.1 31.8
with 128 channels integrated. In particular, by increasing the
transmission voltage level from four to seven, the heat
generation of the MOSFET device could be considerably
reduced. Simultaneously, we achieved a significant reduction
PNE in power consumption. As a result, the risk of MOSFET
damage is reduced, and HIFU can be transmitted stably and
Voltage [V]

safely. Our newly proposed circuit was demonstrated to be


effective for transmission of pseudosinusoidal waves not only
in the trigger HIFU mode but also in the second-harmonic
superimposed wave mode. The applicability of our proposed
PPE circuit has expanded. In the future, we will evaluate the
transmission waveform acoustic characteristics of the circuit
by connecting the circuit to a transducer that supports 128-
Time [μs] channel dual-frequency transmission. Moreover, we are
planning to apply our newly developed HIFU unit in the
Fig. 6. (Color online) Actual drive waveforms for the 400 Vpp trigger
pulse exposure with PNE waves immediately followed by PPE waves at 1
trigger HIFU mode and the second-harmonic superimposed
and 2 MHz transmissions, using five-voltage-level circuit. When using the wave mode.
ground-level clamping circuit, PG and NG in Fig. 2(b), the drive voltage
settled to the ground level immediately after the transmission. Acknowledgment
This research was partially supported by the Japan Agency
for Medical Research and Development, AMED.
that more than three-voltage-level are necessary to suppress
unwanted harmonics below a reasonable level and to prevent
MOSFET damage owing to excessive MOSFET heat 1) T. Uchida, S. Shoji, M. Nakano, S. Hongo, M. Nitta, A. Murota, and Y.
generation. These results are shown in Table IV. Nagata, Int. J. Urol. 16, 881 (2009).
Yasuda et al.34) found that HIFU exposure with PNE 2) S. Vaezy, V. Y. Fujimoto, C. Walker, R. W. Martin, E. Y. Chi, and L. A.
Crum, Am. J. Obstet. Gynecol. 183, 6 (2000).
waves immediately followed by PPE waves leads to the 3) T. Uchida, H. Ohkusa, Y. Nagata, T. Hyodo, T. Satoh, and A. Irie, BJU Int.
generation of cavitation bubbles most efficiently in the 97, 56 (2006).
polyacrylamide gel phantom. 4) R. Yang, C. R. Reilly, F. J. Rescorla, P. R. Faught, N. T. Sanghvi, F. J. Fry,
In a three-voltage-level transmission circuit, it was T. D. Franklin, Jr., L. Lumeng, and J. L. Grosfeld, Arch. Surg. 8, 126
(1991).
impossible to generate a waveform with the unwanted 5) M. Huisman, M. K. Lam, L. W. Bartels, R. J. Nijenhuis, C. T. Moonen,
harmonics suppressed to a reasonable level. However, by F. M. Knuttel, H. M. Verkooijen, M. Vulpen, and M. A. Bosch, J. Ther.
using a five-voltage-level transmission circuit, we were able Ultrasound 2, 16 (2014).
to generate the waveforms with almost equal fundamental 6) F. Wu, G. Haar, and W. Chen, Expert Rev. Anticancer Ther. 7, 823 (2007).
7) J. E. Kennedy, Nat. Rev. Cancer 5, 321 (2005).
and second-harmonic acoustic powers and suppress un- 8) A. Sofuni, F. Moriyasu, T. Sano, K. Yamada, F. Itokawa, T. Tsuchiya, S.
wanted harmonics below a reasonable level. As a result, by Tsuji, T. Kurihara, K. Ishii, and T. Itoi, J. Hepato-Biliary-Pancreatic Sci. 18,
using the five-voltage-level transmission circuit, we were 295 (2011).
able to generate PNE waves immediately followed by PPE 9) K. Tachibana, Innervision 31, 33 (2016) [in Japanese].
10) R. Iwasaki, R. Nagaoka, R. Takagi, K. Goto, S. Yoshizawa, Y. Saijo, and S.
waves at 1 and 2 MHz at a waveform voltage of 400 Vpp, Umemura, Jpn. J. Appl. Phys. 54, 07HF11 (2015).
as shown in Fig. 6. Therefore, we can expect to produce 11) K. Goto, R. Takagi, T. Miyasaka, H. Jimbo, S. Yoshizawa, and S.
cavitation bubbles efficiently. Umemura, Jpn. J. Appl. Phys. 54, 07HF12 (2015).
12) S. Umemura, K. Kawabata, and K. Sasaki, J. Acoust. Soc. Am. 101, 569
Yasuda et al.34) also suggested that the second-harmonic
(1997).
superimposed ultrasound exposure cannot only generate 13) M. Osuga, J. Yasuda, H. Jimbo, S. Yoshizawa, and S. Umemura, Jpn. J.
cavitation bubbles efficiently but also control them spatially Appl. Phys. 55, 07KF18 (2016).
and temporally. Our proposed circuit is easily applicable to 14) S. Umemura and C. A. Cain, IEEE Trans. Ultrason. Ferroelectr. Freq.
Control 39, 32 (1992).
multichannel arrays using multifocus transmission beam-
15) J. A. Brown and G. R. Lockwood, IEEE Trans. Ultrason. Ferroelectr. Freq.
forming, three-dimensional cavitation generation position, Control 49, 848 (2002).
and the time of occurrence could be easily controlled. 16) X. Xu, J. T. Yen, and K. K. Shung, IEEE Trans. Ultrason. Ferroelectr. Freq.

07JF21-7 © 2017 The Japan Society of Applied Physics


Jpn. J. Appl. Phys. 56, 07JF21 (2017) S. Tamano et al.

Control 54, 443 (2007). 26) CY7C1372 datasheet (Cypress) [http://www.cypress.com/file/226436/


17) G. K. Lewis, Jr. and W. L. Olbricht, Rev. Sci. Instrum. 79, 114302 (2008). download].
18) G. K. Lewis, Jr. and W. L. Olbricht, Rev. Sci. Instrum. 80, 114704 (2009). 27) MD1820 datasheet (Microchip) [http://ww1.microchip.com/downloads/en/
19) K. Moro, S. Yoshizawa, and S. Umemura, Proc. 30th Symp. Ultrasonic DeviceDoc/md1820.pdf].
Electronics, 2009, p. 333. 28) SiHFR9310-GE3 datasheet [http://www.vishay.com/docs/91284/
20) K. Moro, S. Yoshizawa, and S. Umemura, Jpn. J. Appl. Phys. 49, 07HF02 sihfr931.pdf].
(2010). 29) FQD3P50 datasheet (On Semiconductor) [https://www.fairchildsemi.co.jp/
21) K. Takada, J. Okada, K. Nakamura, S. Yoshizawa, and S. Umemura, Jpn. J. datasheets/FQ/FQD3P50.pdf].
Appl. Phys. 51, 07GF23 (2012). 30) FDD5N50NZ datasheet (On Semiconductor) [https://
22) S. Tamano, H. Jimbo, T. Azuma, S. Yoshizawa, K. Fujiwara, K. Itani, and www.fairchildsemi.co.jp/datasheets/FD/FDD5N50NZ.pdf].
S. Umemura, Jpn. J. Appl. Phys. 55, 07KF17 (2016). 31) M. M. El-Desouki and K. Hynynen, Sensors 11, 539 (2011).
23) R. Seip, W. Chen, J. Tavakkoli, L. A. Frizzel, and N. T. Sanghvi, Proc. 3rd 32) Z. Zaini, M. Osuga, H. Jimbo, J. Yasuda, R. Takagi, S. Yoshizawa, and S.
Int. Symp. Therapeutic Ultrasound, 2003, p. 423. Umemura, Jpn. J. Appl. Phys. 55, 07KF15 (2016).
24) CY7C68013A-128AXC datasheet (Cypress Semiconductor) 33) 100A2, Power Amplifier datasheet (Electronics & Innovation) [http://
[http://www.cypress.com/file/138911/download]. www.bellnw.com/downloads/datasheet/eandi/100A2datasheetRevB.pdf].
25) 5CEBA7F31C8 datasheet (Intel) [https://www.altera.co.jp/content/dam/ 34) J. Yasuda, A. Asai, S. Yoshizawa, and S. Umemura, Jpn. J. Appl. Phys. 52,
altera-www/global/en_US/pdfs/literature/hb/cyclone-v/cv_51002.pdf]. 07HF11 (2013).

07JF21-8 © 2017 The Japan Society of Applied Physics

You might also like