Chapter 10: Servicing/repairs

AquaBplus Disinfection Report

Responsible organization: Serial number: Disinfectant:

Address: Equipment code: Type/manufacturer:

City: Date: Batch number:

Phone: – Storage life:

No. Disinfection process Unit Measured value 

1 Preparations

1.1 Test of the disinfection process

Duration of the last disinfection h 

Required quantity of disinfectant Containers 

kg

Completed Disinfection and Settings Reports are available. –  Yes 

1.2 All supply ports disconnected. 

1.3 Permeate conductivity prior to disinfection µS/cm 

1.4 Safety

Protective equipment prepared (goggles, rubber gloves, protective clothing as necessary). 

Necessary test equipment and accessories prepared. 

Required quantity of disinfectant prepared: Number of Target: 


containers Actual:
 Puristeril plus  Puristeril 340
kg Target:
Actual:

1.5 AquaBplus switched to STANDBY. 

2 Disinfection procedure

2.0 Preparations completed. 

2.1 Time (start of process) o'clock : 

2.2 Protective equipment donned. 

2.3 Full disinfectant container opened and container connector attached. 

Fresenius Medical Care AquaBplus SM-EN 9A-2019 10-19


Chapter 10: Servicing/repairs

No. Disinfection process Unit Measured value 

2.4 Starting the Disinfection mode 


Display: Start disinfection
Confirmed with the Enter and Menu keys.

2.5 Display: Phase 0/6 Preparation Please add disinfectant 

Filter in the disinfection connector cleaned. 

Disinfection connector attached to container. 

Preparations confirmed by selecting Confirm. 

2.6 Display: Terminate disinf. supply! 


Disinfection connector disconnected from container and connected to the reverse osmosis system.

2.7 Repeat disinf. supply prompt 


Yes: continue with step 2.5
No: continue with the next step 2.8

2.8 Time (start of dwell phase) o'clock : 

3 Residual test

3.1 Display: Check for residual disinfectant! (in phase 6/6 Rinse phase 2) 
Residual test carried out at all supply ports.

3.2 Residual test correct? prompt

Are all supply ports free of residual disinfectant? –  Yes 

Yes: If the  key is pressed, disinfection is finished. Continue with item 3.3. 
No: If another key is pressed, additional rinse programs are required. Continue with item 3.2

3.3 Disinfection program (including rinse programs) finished 

3.4 Time (program end) o'clock : 

4 Finishing the disinfection

4.1 Conductivity after disinfection (start Supply) µS/cm 

4.2 Final assessment/release

Value same as before the disinfection?  Yes  No 

4.3 Time (end of disinfection) o'clock : 

Total disinfection time h 

Comments:

Date: Technician’s signature: Customer's signature:

#FMESIG_5## #FMESIG_6##

10-20 Fresenius Medical Care AquaBplus SM-EN 9A-2019