OEC 9900 Reference Monitor Service Manual Update
OEC 9900 Reference Monitor Service Manual Update
UPDATE
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 1 of 33
Table of Contents
Table of Contents ................................................................................................................................................................................2
Service disclaimers ............................................................................................................................................................................3
Overview .............................................................................................................................................................................................. 12
Product Lines ................................................................................................................................................................................ 12
Tools and Test Equipment ...................................................................................................................................................... 12
Reference Documents.............................................................................................................................................................. 12
Safety................................................................................................................................................................................................ 12
Hazard Alerts ................................................................................................................................................................................ 12
Illustrated Parts Manual ............................................................................................................................................................... 13
Planned Maintenance ................................................................................................................................................................... 16
Service Instructions ........................................................................................................................................................................ 17
C-Arm Covers................................................................................................................................................................................ 17
Reference Monitor Assembly ..................................................................................................................................................... 18
Manpower Requirements ....................................................................................................................................................... 18
Removal of Reference Monitor Assembly ....................................................................................................................... 18
Installation of new Reference Monitor Assembly ....................................................................................................... 21
24VDCPower Supply....................................................................................................................................................................... 25
Manpower Requirements ....................................................................................................................................................... 25
Power Supply Removal ............................................................................................................................................................ 25
Install Power Supply .................................................................................................................................................................. 26
Balun ...................................................................................................................................................................................................... 27
Manpower Requirements ....................................................................................................................................................... 27
Balun Removal ............................................................................................................................................................................. 27
Balun Install ................................................................................................................................................................................... 28
LEMO Adaptor Cable ...................................................................................................................................................................... 29
Set-up ............................................................................................................................................................................................... 29
Removal........................................................................................................................................................................................... 29
Installation .......................................................................................................................................................................................... 31
Test ......................................................................................................................................................................................................... 31
Finish ...................................................................................................................................................................................................... 33
Revision History ................................................................................................................................................................................ 33
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 2 of 33
Service disclaimers
WARNING This Service Manual is available in English only.
(EN)
If a customer's service provider requires a language other than English, it is
the customer's responsibility to provide translation services.
Do not attempt to service the equipment unless this service manual has
been consulted and is understood.
Failure to heed this warning may result in injury to the service provider,
operator, or patient from electric shock, or from mechanical or other
hazards.
警告 本维修手册仅提供英文版本。
(ZH-CN) 如果客户的维修服务提供者需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
不遵守该警告可能会导致维修提供者、操作员或患者因触电、机械或其他
危害造成的损伤。
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 3 of 33
警告 本維修手冊僅有英文版本。
(ZH-TW) 如顧客的維修服務提供者要求英文以外的語言,顧客有責任提供翻譯的服
務。
除非已查閱並瞭解本維修手冊,否則請勿嘗試維修設備。
未注意本警告可能導致維修服務提供者、操作人員或病患遭受到觸
電、由機械或其他因素造成的危害。
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 4 of 33
WAARSCHUWING DEZE ONDERHOUDSHANDLEIDING IS ALLEEN VERKRIJGBAAR IN HET ENGELS.
(NL)
ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL NODIG HEEFT, DAN IS
DE KLANT VERANTWOORDELIJK VOOR DE VERTALING VAN DEZE
HANDLEIDING.
PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING IS GERAADPLEEGD EN BEGREPEN.
INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, KAN HET
ONDERHOUDSPERSONEEL, DE BEDIENER OF EEN PATIËNT LETSEL OPLOPEN
ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK OF DOOR MECHANISCHE OF
ANDERE RISICO'S.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 5 of 33
Warnung Diese Serviceanleitung existiert nur in englischer Sprache.
(DE)
Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe
des Kunden, für eine entsprechende Übersetzung zu sorgen.
Versuchen Sie nicht, diese Anlage zu warten, ohne dass die Serviceanleitung
gelesen und verstanden wurde.
Bei Nichtbeachtung dieser Warnung kann es zu Verletzungen des
Kundendiensttechnikers, des Bedieners oder des Patienten durch
Stromschläge, mechanische oder sonstige Gefahren kommen.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 6 of 33
AVVERTENZA Il presente Manuale di assistenza è disponibile soltanto in inglese.
(IT)
Qualora il tecnico del servizio di assistenza del cliente richieda una lingua
diversa dall’inglese, è responsabilità del cliente fornirne la traduzione.
L’intervento sull’apparecchiatura è consentito solo dopo aver consultato e
compreso questo manuale di assistenza.
La mancata osservanza di questa avvertenza può provocare lesioni al
tecnico dell’assistenza o al paziente, causate da scosse elettriche oppure da
pericoli di natura meccanica o altro.
警告 このサービスマニュアルは英語でのみ提供されています。
(JA)
お客様のサービス担当者が英語以外の言語を必要とする場合、翻訳サー
ビスの提供はお客様の責任となります。
このサービスマニュアルを熟読し理解してから、装置のサービスを行な
ってください。
この警告に従わない場合、感電または機械的もしくはその他の危険要因
により、サービス担当者、操作員、または患者が負傷する原因となりま
す。
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 7 of 33
ĮSPĖJIMAS ŠIS EKSPLOATAVIMO VADOVAS PATEIKIAMAS TIK ANGLŲ KALBA.
(LT)
JEI KLIENTO PASLAUGŲ TIEKĖJAS REIKALAUJA VADOVO KITA NEIANGLŲ
KALBA PASIRŪPINTI JO TEKSTO VERTIMU YRA KLIENTO ATSAKOMYBĖ.
NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS, NEBENT SKAITĖTE ŠĮ
EKSPLOATAVIMO VADOVĄ IR JĮ SUPRATOTE.
JEI NEATKREIPSITE DĖMESIO Į ŠĮ PERSPĖJIMĄ, PASLAUGŲ TEIKĖJAS
OPERATORIUS AR PACIENTAS GALI BŪTI SUŽALOTIDĖL ELEKTROS SMŪGIO,
MECHANINIO AR KITOKIO PAVOJAUS.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 8 of 33
ATENŢIE ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
(RO)
DACĂ UN FURNIZOR DE SERVICII AL CLIENTULUI NECESITĂ O ALTĂ LIMBĂ
DECÂT ENGLEZA, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT DACĂ AŢI CONSULTAT ŞI
AŢI ÎNŢELES ACEST MANUAL DE SERVICE.
IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, A OPERATORULUI SAU A PACIENTULUI ÎN URMA
ELECTROCUTĂRII SAU A PERICOLELOR MECANICE SAU DE ALTĂ NATURĂ.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 9 of 33
Opozorilo Ta servisni priročnik je na voljo le v angleščini.
(SL)
Če strankin ponudnik storitve potrebuje drug jezik, razen angleščine, potem
je stranka odgovorna, da zagotovi prevajalske storitve.
Ne poskusite opravljati servisiranja, če niste prebrali in razumeli tega
servisnega priročnika.
Če ne upoštevate tega opozorila, lahko pride do poškodb ponudnika storitve,
operaterja ali bolnika zaradi električnega udara ali mehanskih in drugih
nevarnosti.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 10 of 33
DİKKAT BU SERVİS KILAVUZUNUN SADECE İNGİLİZCESİ MEVCUTTUR.
(TR)
MÜŞTERİNİN HİZMET SAĞLAYICISI, İNGİLİZCE DIŞINDA BİR BAŞKA DİLDE
KILAVUZ TALEP EDERSE, TERCÜME HİZMETLERİNİN SAĞLANMASI MÜŞTERİNİN
SORUMLULUĞUDUR.
SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARIN BAKIMINI
YAPMAYIN.
BU UYARIYA DİKKAT EDİLMEMESİ, ELEKTRİK, MEKANİK VEYA DİĞER
TEHLİKELERDEN DOLAYI HİZMET SAĞLAYICININ, OPERATÖRÜN VEYA
HASTANIN YARALANMASINA YOL AÇABİLİR.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 11 of 33
Overview
This service manual supplement is for 9900 systems that have a reference monitor installed on the
C-Arm. The reference monitor replicates the Workstation left monitor image. Follow the instructions
in this document when installing, performing planned maintenance, or servicing 9900 systems with
the reference monitor.
Product Lines
• GE OEC 9900 Elite C-Arms excluding motorized systems.
Reference Documents
• 9900 Service Manual
• Reference Monitor User Manual
Safety
Note: Use proper PPE per the 9900 Service manual.
CAUTION Follow all LockOut/TagOut (LOTO) procedures as described in the service manual.
Hazard Alerts
This document uses three kinds of hazard alerts:
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 12 of 33
Illustrated Parts Manual
GE Part Vendor Part
Description Image
Number Number
5477289 20432-001 REFERENCE MONITOR
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 13 of 33
GE Part Vendor Part
Description Image
Number Number
5477293 20465-001 CONTROL PANEL COVER FOR REFERENCE
MONITOR
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 14 of 33
GE Part Vendor Part
Description Image
Number Number
20296-005 #10 FLAT WASHER
QTY 4
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 15 of 33
Planned Maintenance
1. Connect the interconnect cable from the workstation and start up the system.
2. Verify that the system starts up with no errors, and that the reference monitor turns on and
displays the same image as the left monitor displays.
3. Verify that the control panel console keys, X-ray button, up-down buttons, and control panel
display function properly.
4. Verify that the reference monitor swivels left to right and pivots up and down smoothly.
5. Swivel can be adjusted by loosening or tightening the set screw located at the lower rear of
the curved arm.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 16 of 33
6. Tilt can be adjusted by loosening or tightening the two (2) socket head cap screws located at
the upper rear of the curved arm. If the tilt screws are adjusted, both screws should be
adjusted the same amount.
Service Instructions
C-Arm Covers
1. Raise the vertical column 4 to 8 inches to make it easier to remove the C-Arm covers.
2. Power down the system.
3. Unplug the system and disconnect the interconnect cable.
Note: Remove the appropriate covers per the procedures in the following sections.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 17 of 33
Note: Remove the cross arm brake handle and control panel cover per the procedures in the
following sections.
Manpower Requirements
1 person, 1.5 hours
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 18 of 33
3. Remove Cross Arm brake handle.
4. Remove the four (4) 10-32 head cap screws and #10 flat washers at the base of the
reference monitor assembly.
5. Remove the two screws from the underside of the Control Panel cover.
6. Lift the control panel cover and reference monitor assembly from the control panel
chassis.
7. Disconnect the three (3) control panel cover internal cables from the three (3) switches.
8. Slide the entire reference monitor cover assembly into the mounting grooves on the left
side of the control panel chassis.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 19 of 33
RJ45
Receptacle
12. Remove the right Control Panel (when viewing from the rear of the C Arm).
13. Remove the Phillips 10-32 pan head screw and star washer from the control panel
chassis at the lower right corner that secures the ground wire.
14. Remove the four (4) 4-40 X ¼ flat head screws from the cover plate on the underside of
the control panel cover and remove the cover plate.
Cover plate
4-40 x 1/4
Flat Head
Torx Pan
head screws
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 20 of 33
Tie
wraps
16. Remove the Control Panel cover and Reference Monitor assembly from the system.
17. Remove the four (4) 10-14 torx pan head screws from the underside of the Control Panel
cover and remove Reference Monitor assembly from the control panel cover.
18. Route the monitor cables through the top cut out of the control panel cover.
Reference monitor
assembly
19. Make sure the reference monitor is oriented such that it can be fully rotated when facing
the X-ray switch, and then attach the reference monitor assembly to the control panel
housing using the four (4) 10-14 pan head screws removed earlier.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 21 of 33
10-14 Pan
head screws
20. Ensure that the monitor cable, ground wire and X-ray switch wires are routed within the
monitor mount channels and that there is sufficient length at both ends such that the
wires do not bind or become pinched when the cover plate is installed.
21. Secure the monitor cable, ground wire and X-ray switch wires to the underside of the
control panel cover using the cover plate and the four (4) 4-40 X ¼ flat head screws as
shown below.
Cover
plate
Monitor
channels
22. Install the coiled cord strain relief to the slot on the cover and secure to the two (2) cable
anchors using tie wraps.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 22 of 33
Tie
wraps
23. Slide the entire reference monitor cover assembly into the mounting grooves on the left
side of the control panel chassis.
CAUTION: Be very careful to secure the cover in the grooves so the assembly does not fall during
installation and get damaged.
24. Connect the ground wire from the reference monitor assembly to the control panel
chassis at the lower right corner with the Phillips 10-32 pan head screw and star washer.
Ground
25. Reinstall the right control panel with the four (4) screws previously removed.
26. Reconnect the three (3) control panel cover internal cables to the three (3) switches.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 23 of 33
27. Lower the reference monitor assembly onto the control panel chassis. Carefully dress all
cables inside the cover to avoid crimping, cutting, or short circuit.
CAUTION: The cables inside the cover are a tight fit. Dress them carefully.
28. Secure the reference monitor assembly to the inside cover mounting bracket with four (4)
10-32 head cap screws and #10 flat washers. Tighten the screws ½ turn past snug.
29. Reinstall the two screws to the underside of the Control Panel cover.
30. Reinstall the cross arm brake handle.
31. Install the coil cord strain relief to the front cover.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 24 of 33
32. Route the RJ45 end of the coil cord through the mainframe circuit area down to the
balun in the base.
24VDCPower Supply
Manpower Requirements
1 person, 1.0 hours
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 25 of 33
4. Remove the power supply retention clip.
AC
DC
Power Supply
Retention Clip
DC
Note: The power supply is mounted with double sided tape. Clean any tape residue from the base
wall after removing.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 26 of 33
2. Slide the metal retainer clip over the power supply about halfway up the body of the
power supply. Do not let the clip stick to the tape.
3. Mount the power supply flush with the inside left wall of the C-Arm base (as viewed from
the rear of the mainframe) with the clip over the outside of the base wall. Press the power
supply firmly to bond the adhesive tape to the wall.
Balun
Manpower Requirements
1 person, 1.0 hours
Balun Removal
1. Power down the system.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 27 of 33
2. Unplug the system and disconnect the interconnect cable.
3. Remove the front top cover and front leg cover of the C-Arm.
4. Remove the balun retention clip.
5. Disconnect the black and white wires from the balun.
Note: The balun is mounted with double sided tape. Clean any tape residue from the base wall after
removing.
Balun Install
1. Remove the release liner from the double-sided tape on the new balun.
2. Slide the balun retainer clip over the balun. Do not let the clip stick to the tape.
3. With the BNC connector towards the front of the C-Arm, mount the balun to the inside
right wall of the C-Arm base (as viewed from the rear of the mainframe) with the clip over
the outside of the base wall. Press firmly to bond the adhesive tape to the wall.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 28 of 33
Power
Balun
Supply
4. Reconnect the black & white wires to the balun in the same receptacles from which they
were removed.
5. Reconnect the RJ45
6. Reconnect the BNC connector.
7. Proceed to the test section of this document.
Set-up
1. Power down the system.
2. Unplug the system and disconnect the interconnect cable.
3. Remove the front top cover, right side cover, rear cover, and front leg cover.
Removal
4. Remove the cable tie from P1 LEMO wires to gain access to the P1 socket.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 29 of 33
5. Use a LEMO pin extractor to remove the two pins 23 and 21 (center and just above
center) from the interconnect receptacle.
6. Disconnect the other end of the LEMO Adapter cable BNC connector from the balun BNC
connector.
Balun BNC
Connector
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 30 of 33
Installation
1. Obtain the new LEMO adapter cable and insert the white lead LEMO pin into P1 pin 21
and ground (white/black stripe) lead to P1 pin 23.
2. Install a tie wrap that was removed earlier to the P1 LEMO wires, including the new LEMO
adaptor cable.
3. Route the BNC connector end of the LEMO adaptor cable down through the side hole in
the chassis, taking care to keep the cable away from the high voltage lines, and down
behind the round bar into the base space.
Video
cable
i
Test
WARNING C-Arm covers are removed. Dangerous voltages are present when the covers are
removed. Take appropriate safety precautions.
1. With the covers still off, reconnect the interconnect cable from the workstation and start
up the system.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 31 of 33
2. Verify that the system starts up with no errors, and that the reference monitor turns on
and displays the same image as the left monitor displays.
3. Verify that the control panel console keys, X-ray button, up-down buttons, and control
panel display function properly.
4. Verify that the reference monitor swivels and pivots up and down smoothly.
5. Swivel can be adjusted by loosening or tightening the set screw located at the lower rear
of the curved arm.
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page
Made in the USA 5479608-1EN-01 1 32 of 33
6. Tilt can be adjusted by loosening or tightening the two (2) socket head cap screws
located at the upper rear of the curved arm. If the tilt screws are adjusted, both screws
should be adjusted the same amount.
Finish
1. Properly shut down the system and unplug it.
2. Re-install all covers.
Revision History
Rev # Date Description of Change
1 April 2013 Initial release
© 2013 GE OEC Medical Systems, Inc. - All Rights Reserved. Part Number Rev Page